Professional Documents
Culture Documents
CS-200
July 2011
Thank-you for purchasing LECO analytical equipment. Installation by LECO service personnel is
included as part of your purchase. The following information is provided so that the installation site can be
properly prepared prior to the arrival of LECO personnel. Please follow the information for your new
instrument carefully in order to assure proper site preparation. If any questions arise during site
preparation, contact the LECO Service Department (Phone: 269-982-5497 Fax: 269-983-1477).
If the above steps are not followed, LECO will not be responsible for delays, cancellations, or
rescheduling of installations.
CS-200
FLOOR SPACE
Before your instrument arrives, make sure you know where it will be installed. When factoring overall
floor space requirements, a minimum 6-inch (15 cm) access area should be factored in around each unit.
COUNTER TOP
23.5-inches
(60 cm)
MONITOR
PC
ANALYZER
PRINTER
OPTIONAL
BALANCE
63-inches
(160 cm)
FLOOR
16.5-inches
(42 cm)
14-inches
(36 cm)
27.5-inches
(70 cm)
*
9-inches
(23 cm)
8-inches
(20 cm)
93-inches
(236 cm)
Height
Width
Depth
Weight
Analyzer
Computer
PC Monitor
Printer (optional)
20 lbs. (9 kg)
Balance (optional)
Not available
Autoloader (optional)*
CS-200
Computer ...........................................
Monitor ...............................................
Printer (optional).................................
NOTE:
DETERMINATOR
POWER CORD
BLACK
POWER
220 VAC
RED
NEUTRAL
0 VAC
GREEN
GROUND
CHASSIS
(CABINET)
GROUND
POWER
Y
G
GREEN
GREEN G
115 VAC
230 VAC
POWER
RED
115 VAC
GROUND
RED
WHITE *
BLACK
WHITE
X
BLACK
CHASSIS
(CABINET)
GROUND
765-978 RECEPTACLE
* WHITE WIRE IS CUT OFF INSIDE THE UNIT AND IS NON-FUNCTIONAL
POWER
Y
115 VAC
GREEN
GREEN G
115 VAC
POWER
WHITE *
BLACK
0 VAC
NEUTRAL
RED
230 VAC
GROUND
RED
WHITE
X
CHASSIS
(CABINET)
GROUND
BLACK
765-978 RECEPTACLE
* WHITE WIRE IS CUT OFF INSIDE THE UNIT AND IS NON-FUNCTIONAL
CS-200
GAS SUPPLIES
Gas supplies will require regulators.
Refer to the Gas Regulator chart. OXYGEN IS NOT
RECOMMENDED FOR USE IN THE PNEUMATIC SYSTEMS OF LECO INSTRUMENTS.
NOTE:
Supply or supply lines must be within 6 feet of the instruments designated area.
Carrier Gas .............................................. Oxygen, 99.5% pure, 40 psi (2.8 kg/cm2) 10%.
Pneumatic Gas........................................ Compressed Air or Nitrogen, 40 psi (2.8 kg/cm2) 10%.
Source must be oil and water free.
GAS REGULATORS
GAS
LECO
PART NO.
VICTOR
PART NO.
DELIVERY
RANGE
INLET
CONNECTION
Oxygen
501-291
VTS450D-540-0781-3905
0 to 125 psi
CGA 540
Nitrogen
764-216
VTS450D-580-0781-3909
0 to 125 psi
CGA 580
Compressed Air
766-036
VTS450D-346-0781-4013
0 to 125 psi
CGA 346
OUTLET
CONNECTION
9
5
/8 - 18 Female R.H.
9
CS-200
ENVIRONMENTAL CHARACTERISTICS
Nominal Operating Conditions
Operating Temperature .................................
32 to 122oF (0 to 50oC)
Cleanliness ....................................................
Circulation ......................................................
Ventilation ......................................................
CS-200
FLOOR SPACE
93-inches x 63-inches (236 cm x 160 cm)
POWER REQUIREMENTS
Printer and Balance - 115 VAC 10%, 50/60 Hz, single phase earthed supply
Analyzer - 230 VAC 10%, 50/60 Hz, single phase earthed supply, 20 amps
Power Purifier (optional) - Model 621-325 - 208/240 VAC, 60 Hz, single phase, 60 amp
GENERAL
Unit has arrived at installation site, been inspected and missing parts (if any) ordered
_________________________________________________________________________________
City: _________________________________
Phone: (______)________________
State: ________________
Fax: (______)_______________
Zip: __________________
e-mail: ________________________________
By signing below I certify that site preparation is complete. I understand that if the site is not prepared properly and
completely, installation and training will have to be rescheduled. I also understand that if installation and training must
be rescheduled due to incomplete site preparation, my company will be billed at normal LECO service rates for
installation and training.
Signature: _________________________________________
Date: ___________________________
CS-200