Professional Documents
Culture Documents
SERVICES:
Flowmeter HIRE-easy to use, non-invasive
ON-SITE flow measurements
Flowmeter DEMO - see how to do it!
FLUXUS® G
______
Non-Invasive
Gas Flow Measurement
Gas Storage Facilities
______
Gas Transmission
______
Chemical Industry
______
Petrochemical Industry
______
Offshore
______
Manufacturing Industry
______
Non-Invasive
Gas Flow Measurement
with FLUXUS® G
_____
© BEB
Our tradition is innovation
FLEXIM's ultrasonic gas flow meters use the proven ______
clamp-on transit-time correlation technique also Ever since its inception, FLEXIM has
been among the pioneers in the field of
employed for the F series liquid meters. ultrasonic flow measurement. Many years
Special ultrasonic transducers are simply clamped of application experience combined with
innovative concepts and solutions have
onto the outside of the pipe and never come in contact
allowed FLEXIM to extend the scope
with the gas. of the ultrasonic technology to non-
invasive measurement of gases.
2
FLEXIM_Fluxus_G_en_V5_091217 23.12.2009 12:56 Uhr Seite 3
© Pål Rønning
© BEB
3
FLEXIM_Fluxus_G_en_V5_091217 23.12.2009 12:56 Uhr Seite 4
Superior
Functionality
_____
4
FLEXIM_Fluxus_G_en_V5_091217 23.12.2009 12:56 Uhr Seite 5
5
FLEXIM_Fluxus_G_en_V5_091217 23.12.2009 12:56 Uhr Seite 6
Technical Data
_____
Ultrasonic flow meters natively measure The Transit Time Difference Correlation Principle makes use of the fact
the actual volumetric flow. that the time of flight of an ultrasonic signal is affected by the flow
The transmitters feature an integrated velocity of the carrier medium. Like a swimmer working his way across
flow computer which makes use of a flowing river, an ultrasonic signal travels slower upstream than down-
external temperature and pressure data stream.
to convert the actual volumetric flow into Our instruments work according to this transit-time principle: an ultra-
standard volume (mass) flow based on sonic pulse is sent downstream through the medium, another pulse is
ISO 12213-1. sent upstream. By measuring the transit-time difference, the flow
velocity can be determined. The volumetric flow is then calculated out
In principle, flow measurement is possible of the flow velocity and the pipe parameters.
on all types of gases. However, when
measuring on metal pipes, a certain
minimum pressure is required, depending
on the composition of the gas and the
pipe material. Plastic pipes will support General Technical Data
operation at atmospheric pressure. __________________________________________________________________________________________
Measuring principle: Transit time difference correlation principle
Our application engineers will be __________________________________________________________________________________________
pleased to advise you. Flow velocity: (0.01 to 35) m/s, pipe size dependent
__________________________________________________________________________________________
Repeatability: 0.15 % of reading ± 0.01 m/s
__________________________________________________________________________________________
Accuracy (given a fully Volume flow:
developed rotationally ± 1% to 3 % of reading ± 0.01 m/s, depending on the application
symmetric flow profile) ± 0.5 % of reading ± 0.01 m/s with field calibration
__________________________________________________________________________________________
Measurable gases: Ratio of the characteristic acoustic impedance of the
pipe wall material to that of the gas < 3000
__________________________________________________________________________________________
Operating pressure: Measurement does not restrict
the maximum operating pressure
__________________________________________________________________________________________
Pipe size range: (7 to 1600) mm
__________________________________________________________________________________________
Gas temperature: (–40 to +200) °C, depending on the transducer type
__________________________________________________________________________________________
6
FLEXIM_Fluxus_G_en_V5_091217 23.12.2009 12:56 Uhr Seite 7
Applications
_____
______ ______
Measurement on natural gas pipelines Measurement of synthesized gas
______ ______
Measurement in natural gas storage Gas measurement in the plastics
installations (storage and extraction) production process (high pressure)
______ ______
Management of gas storage facilities Measurement of ethylene, helium
(efficiency maximization, performance and nitrogen
optimization, salt cavern leaching) ______
______
Regulation and control of drilling sites
in natural gas extraction
______
Control of compressor stations Manufacturing industry
______
Dimensioning of extraction sites and ______
drilling probes (efficiency maximization) Measurement of compressed air
______ ______
Measurement of ethane, propane and
hydrogen
______
Measurement of sour gas
______ Service
Measurement for the gas supply industry
______ ______
Measurement of injection gas Supervision of permanently
in the petroleum industry installed meters
______ ______
Allocation and checkmetering Error diagnostics
______ ______
Inventory control / Service and maintenance
Lost and unaccounted for gas analysis ______
______
Measurement of injection media in
gas processing (monoethylamine, TEG, etc.)
Gas storage facility Kalle,
______
7
FLEXIM_Fluxus_G_en_V5_091217 23.12.2009 12:56 Uhr Seite 8
The
Transmitters
_____
Technical Data
__________________________________________________________
Operating temperature:
__________________________________________________________
Measuring channels:
__________________________________________________________
__________________________________________________________
Qutputs:
__________________________________________________________
8
FLEXIM_Fluxus_G_en_V5_091217 23.12.2009 12:56 Uhr Seite 9
The Portable All-round Meter The Multi-function Meters The Explosion-proof Experts
FLUXUS® G601 FLUXUS® G704 / FLUXUS® G709 FLUXUS® G800 / FLUXUS® G801
Unmatched in performance, this handy and Permanently installed and freely configurable: FLUXUS® G800 and G801 are ATEX certified
versatile instrument is ideal in support of The FLUXUS® G704 is designed for permanent for use in hazardous areas. The especially
service and maintenance activities. It may installation. With its range of electrical inputs robust G801 is completely seawater-resistant
also be used for the control or even the and outputs, it can handle a variety of process and therefore ideal for offshore applications.
temporary replacement of permanently parameters. Although technically identical to
installed meters. the G704, FLUXUS® G709 is designed for
permanent installation in 19" rack systems.
_________________________________________________________________________________________________________________________________________________________________________
–10 °C to 60 °C –20 °C to 60 °C –20 °C to 50 °C
_________________________________________________________________________________________________________________________________________________________________________
2 1 or 2 1 or 2
_________________________________________________________________________________________________________________________________________________________________________
Standard: 2 x current, 2 x binary A variety of combinations are available from 1 current output, 1 binary output (OC).
Multifunctional: 4 x current, 2 x binary, the following: current (0/4 mA to 20 mA), Also available: 1 or 2 binary outputs (relay),
1 x frequency voltage, frequency, pulse, alarm 1 binary output (OC) and 1 current output
_________________________________________________________________________________________________________________________________________________________________________
9
FLEXIM_Fluxus_G_en_V5_091217 23.12.2009 12:56 Uhr Seite 10
10
FLEXIM_Fluxus_G_en_V5_091217 23.12.2009 12:56 Uhr Seite 11
11