White Paper
Sept 2013
This paper describes theoretical and practical aspects of sample preparation for Light Sheet Fluorescence
Microscopy (LSFM). We present general rules for sample handling and mounting, as well as guidelines with
respect to the best preparative technique to use, taking into account sample type, structure and properties.
Step by step protocols and recommended materials for ZEISS Lightsheet Z.1 samples are included.
These protocols cover sample preparation ranging from micrometer-sized fluorescent beads to millimeter-sized
insects, providing detailed information relating to preparation and observation techniques. Finally, this paper
identifies the main artifacts and problems that can result from the preparation techniques.
CONTENTS (click text to go to page):
1 INTRODUCTION
PREPARATION
22
22
3.1
2.1
3.2
2.2
(Oryza latypes)
2.2.1
Embedded Samples
3.3
2.2.2
Hanging Samples
12
(Drosophila melanogaster)
2.2.3
Enclosed Samples
13
3.4
2.2.4
FEP Tubing
15
(Arabidopsis thaliana)
24
2.3
15
3.5
25
2.3.1
Sample Chambers
15
3.6
2.3.2
16
2.3.3
Sample Holder
18
3.7
2.3.4
18
(Anopheles gambiae)
2.3.5
Hydrogel Preparation
19
2.4
20
2.5
20
ADDITIONAL INFORMATION
2.5.1
20
4.1 Tips
28
2.6
Antifading Agents
21
4.2 Troubleshooting
28
2.7
21
4.3
30
4.4
31
23
5 Index
26
27
28
33
White Paper
1. INTRODUCTION
A microscope generally performs best on suitable samples,
and when the samples are optimally prepared for the imag-
deeply and does not change its properties and shape along
issues.
White Paper
Topic
Subtopic
Sample/Model Organism
Technique/ LSFM
implementation
Reference
Physics
Fluorescent beads
Mouse cochlea
Fluorescent beads
Caenorhabditis elegans
live sea urchin embryo, live
Danio rerio embryo
MISERB
sTSLIM
SPIM
DSLM
LSFM
Biochemistry
Laser Microsurgery
Microtubule dynamic instability
mRNA nuclear export
In vitro microtubules
In vitro microtubules
Chironomus tentans Salivary
Glands
MCDK cells, Drosophila
melanogaster
Drosophila melanogaster
SPIM
SPIM
SPIM
LSFM (FCS)
Capoulade, 2011
SPIM
Heterochromatin dynamics
Imaging of engineered gene
expression
Microbiology
Marine microbiology
LSFM
Cell biology
Tumour spheroids
waoSPIM
Arabidopsis thaliana
Arabidopsis thaliana
DSLM
SPIM
SPIM
SPIM
mSPIM
iSPIM
Drosophila melanogaster
SPIM
Physiology
(HR) OPFOS
OPFOS; LSFM
miniSPIM
OPFOS
Mouse brain
Mouse brain
Mouse cochlea/zebrafish inner
ear, brain/ rat brain
Mouse vomeronasal cells
Mouse
LSFM
Ultramicroscope
TSLIM
OCPI
SPIM
Meriones unguiculatus
cochlea, Hippocampus reidi
head, Xenopus laevis
OPFOS
Ormia ochracea;
Emblemasoma auditrix
Drosophila melanogaster
Calanus pacificus
Thermocyclops consimilis
LSP
Ultramicroscope
PLIF
LSFM
Plant Biology
Large organism
general biology
White Paper
image quality.
over, it usually allows sample rotation to record multiple image stacks that are acquired independently along different
the sample chamber and the system cavity for the sample
aggressive chemicals.
sample from one side (Fig. 1/A and B). The sample is not
Fig. 1 Relations between the sample and the objective in microscopy. Samples are traditionally isolated from the objective by a glass coverslip (A and B)
limiting access to one side only. (C). In LSFM, the illumination is positioned at 90 compared to the detection axis and can be set up in a sideways geometry
(horizontal microscope).
White Paper
Fig. 2 Different sample types. (A) very large objects (cm), (B) large objects (mm), (C) medium size samples (100 m) and small samples (10 m) (D).
Each is represented in relation to the following one to allow size comparison.
(Fig. 2) and you can check their size relationship (Fig. 2) and
White Paper
Fig. 3 Sample positioning in LSFM. The sample can be held in front of the objective (A) embedded in gel, (B) by a clip, (C) in a container, or on a coverslip
(D; Note: ZEISS Lightsheet Z.1 is not designed to support this way of mounting the sample). (E to H) show an eye bird view of the mounting (A to D).
lens. Gel embedding (Fig. 3/A and E) is usually safe but the
capillary that holds the gel can potentially touch the detection
the sample chamber and this could affect the imaging process.
White Paper
Fig. 4 Embedded samples. Large samples, such as an adult Drosophila melanogaster, can be embedded in a large gel tube or a cut 1 ml syringe (A),
intermediate size samples, such a Medaka or Zebrafish embryo can be prepared by using either a cut 1 ml plastic syringe (see also Fig. 5) or a glass
capillary (C and D) and small samples such as Drosophila melanogaster embryos or early stage cell clusters can be prepared using a smaller capillary (D).
of water. The object can be any size, as the gel can be molded
White Paper
Fig. 5 Preparing a sample embedding container with a syringe. Many tubular objects can be used to make embedded samples. A simple technique is illustrated here. The tip of a syringe is cut away (A, B), agarose can then be easily pumped in using the plunger (C), and the sample can be positioned within the
agarose tube (D). After polymerization the sample can be pushed out of the syringe for imaging (E).
sample holders to limit waste. We have found that homemade sample embedding container using 1 ml syringes
the material.
Make sure you use the right capillary for your sample.
The sample size should not be more than 2/3 of the final
ensure that you use the right plunger for you particular
The first is to directly mix the object with the agarose then
White Paper
Fig. 6 Preparing a sample embedding container with a capillary. A glass capillary can be used. For ZEISS Lightsheet Z.1 they come from the manufacturer
(Brand) in just the right length (A). Other capillaries can be cut to an appropriate length (B), the agarose with the object can be easily pumped in using the
suitable plungers with Teflon tips (C). If no such plunger is available it can be made, for example, from a piece of electrical wire (D). To avoid leakage, such a
plunger can be sealed with nail polish (E) once the sample is pushed out. The sample can then be imaged (F).
is pulled back into the syringe and is covered with more molten
wise the cooling agarose would become diluted and lose its
with the gelling agent, then to place the object within the gel
10
White Paper
Fig. 7 Basic principles of sample embedding. A cylinder with a suitable plunger is used as a mounting device (A). The 1 % low melting-point agarose is
melted, then brought to 37 C, then pumped into the cylinder. (B) The object is then introduced to the agarose with a needle or forceps. (C) Once solidified,
the embedded sample can be pushed out and imaged (D). Alternatively, the object, devoid of water, or other solution, is added to a solution of 1 % low
melting-point agarose at just above gelling temperature (typically 40 C) and sucked into the cylinder (E) and then allowed to polymerize. The embedded
sample can then be pushed out and imaged (F).
Fig. 8 Aligning an embedded sample. The sample can be aligned in a particular orientation to allow the details of interest to be close to the outer surface of
the agarose. The solidified agarose is pushed out the syringe a few millimeters and a small v-groove is cut into the cylinder to take up the sample (A).
The sample is placed into the v-groove (B). The sample on the agarose is pulled back into the syringe and more agarose is added (C). After the cylinder has
completely solidified the sample is pushed out of the syringe allowing free sight on to the sample (D). The same approach can be used to carve a central tunnel
in the middle of the agarose to align the sample along the agarose tube axis.
11
White Paper
Fig. 9 Different ways of hanging a sample. Samples can be either hooked or deposited on a bent glass capillary (A, D), glued to a rod or capillary (B),
or clamped on a syringe tip using the plunger and the syringe body as a holder (C).
12
White Paper
The container can be easily molded using a gelling agent specifically chosen for its stiffness and transparency. The custom-made
use for molding (Fig. 11/A). The plunger is pulled into the syringe
body and filled with molten gelling agent. The lunger is further
pulled to create the bottom part of the container (Fig. 11/B and
object but also for the imaging chamber and the sample
Fig. 10 Enclosed chambers for LSFM. Incubation chambers can be made by molding an agarose beaker that can be mounted on a simple plastic holder and
loaded with the sample prior to imaging (A). Another solution is to create a chamber with a specific polymer with a refractive index close to water (for example
PFTE or FPE) using heat or glue to seal the chamber to the needed size and volume, and attach it to a suitable holder(syringe, capillary etc) (B). A PDMS,
FPE tube or glass chamber can also be considered (C).
Fig. 11 Making and mounting an agarose incubation chamber. A modified syringe plunger is made by inserting a smaller diameter cylinder on to the
plunger (A, B). The tube is molded by simply pouring the molten agent into this device (C). Once removed, the open end of the container (D) can be closed with
agarose (E).
13
White Paper
2007).
or the buffer used during the experiment. The polymer is usually used as a sheet that can be formed as required. The other
possibility is to approach commercial manufacturers to make
polymer chambers at the specific sizes and lengths required.
Fusing polymer sheets can be done using a welding iron with
controlled temperature or a welding device use for melting
Fig. 12 Making and mounting an incubation polymer foil chamber. A piece of polymer foil (A) is folded and either heat- or glue-sealed to generate a
tube of a predefined size (B). Excess foil can be removed or used to glue the tube on to a specific holder (capillary,thread, metal rod) (D). One side of the tube
can be then glued or heat-sealed to close the chamber (C). The polymer used must be suitable for microscopy and easy to seal. The chamber can be glued to a
support, held by forceps, or inserted into a slit rod.
14
White Paper
points:
treatment (cost)
Fitting: objective, illumination position, stage, heaters
Size
Cost
Carl Zeiss Microscopy GmbH (hereinafter we) hereby informs you that we will warrant the specified and agreed performance of Lightsheet Z.1 system only if sample
chambers are applied and used that either are delivered or explicitly approved by us.
1
The sample chamber design has been optimized to ensure the most established applications of Light Sheet Fluorescence Microscopy. Exceptional applications may
require a slightly modified sample chamber design. In order to enable customized modifications of the existing sample chamber we also provide the corresponding
CAD file and a technical drawing. We explicitly advise you that already minor deviations of the dimensions and tolerances specified in these documents will cause a
significant loss of image quality and can potentially result in a liquid leakage. Therefore, you will not hold us or one of our affiliates liable for any damages caused
by the employment of self-built or third-party-built sample chambers, the use of such self-built or third-party-built sample chambers will be solely on your own risk.
Furthermore we want to inform you, that we will not render any assistance relating to the production and application of such self-built or third-party-built sample
chambers.
15
White Paper
6. Sample chamber
4. Securing screw
Plastic syringes exist in various sizes (0.2 ml, 0.3 ml, 0.5 ml,
16
White Paper
17
White Paper
particular mount:
stage holder.
Agarose
Cost.
LSFM the object must be held from above via the sample hold-
er. Depending on the size of the sample there are two different
and syringes (Fig. 14/C-F and I). Always use the minimal cylinder
from the top with a plunger that can be operated once the
used for injection fit the upper part of the holder. In this way
Gelrite
and Co, Inc (Rahway, NJ), Kelco Division, USA. One advantage
will slide when wet. Please handle them with care and
rose gel with the same stability. It has a higher index of refraction but less scattering compared to agarose. Gelrite has a
18
White Paper
required for the same purpose. It hydrates rapidly and gel set-
fore a gel made with Gelrite and pure water is unstable com-
before use.
pared to a PBS- (buffer) based gel. Polymerisation is faster compared to agarose, which might be advantageous for some
and store them in a cool and dry place. In this case, each
aliquot can be liquefied using a heating block (80 C 90 C)
Galactan
Agar
Gelatin
Carrageenan
Alginate
Phytagel
Agargel
Transfergel
We will not try to cover every single one of these but rather
PBS) to the flask. Swirl to mix the solution. Place the flask in
allow the agarose to boil over as this will affect the final aga-
reuse it.
retardation.
Wear heat-protective gloves when handling the flask.
19
White Paper
antibody reactions.
for each new sample type will ensure that the best balance
20
White Paper
to use a beaker filled with the right buffer. The samples are
at all time. They are stable and can be easily move from the
ever, you can rinse the sample within the capillary or syringe
with water or your specific buffer after imaging if the chamber
drugs etc.).
filled with the appropriate medium, e.g. water, PBS (Fig. 15).
Fig. 15 Supports for sample embedding containers. Support are used to hold three dimensional objects that cannot be held flat easily. Moreover,
embedded samples need to be kept in buffer to avoid gel shrinkage or sample damage. A simple system is to use a beaker filled with PBS and place plasticine on
the upper border to support the sample embedding container (A). More elaborate supports can be made using clips of different sizes for holding syringes (B) or
even capillaries (C).
21
White Paper
Fluorescent beads
you store many sets of samples in the same fridge day after
day.
Method
water tight container where the samples are kept with enough
water not to dry out. One point to consider is the way the
sample was prepared. Embedded samples may weaken with
time as some gels may not maintain their strength over time
plunger.
10. Make sure that only a very short part of the agarose
acquisition.
With Lightsheet Z.1, a two channel system, one can use one
channel for the beads (e.g., red) and one channel for the
22
White Paper
Heating block
Method
the species used (frog, fish, fly, worm, etc.). This protocol
Fig. 16 Mounting an Oryza latypes embryo. (A) The embryo is prepared (labelling, drug treatment, dissected...) (B) The embryo is deposited on the side of the Eppendorf tube and the excess of water is removed with a pipette. (C) The embryo is dropped into the agarose and pumped into the capillary. (D) The embryo can be imaged.
23
White Paper
that are difficult to image fully and are usually dissected and
scope.
1 ml syringes
Arabidopsis thaliana seeds
Heating block
2. To allow sample preparation the pupa must be submerged in agarose by pouring it directly on top of it in
a large drop of molten low melting point agarose.
3. The pupa can then be pumped into a capillary as previously described.
4. The insect can then be imaged.
24
White Paper
Fig. 17 Mounting a Drosophila pupa. (A) The pupa is prepared (labeling, drug treatment, dissected...) (B) The pupa is deposited in a watch glass and
covered by melted agarose to embed the hydrophobic pupa. (C) The pupa is then pumped into the capillary. (D) The pupa can be imaged.
Fig. 18 Mounting an Arabidopsis thaliana root. (A) An Arabidopsis thaliana seed is positioned at the bottom of an agarose cylinder (B).
After a few days of development, the root can be seen in the agarose cylinder bottom. (C) The agarose cylinder is pushed out of the syringe for imaging.
Modified plunger
Sealing device
Slitted capillary
1 ml syringe
Capillary holder
Heating block
Method
an extracellular matrix.
Samples.
ExtraCell etc.)
25
White Paper
Method
1. MDCK cell cysts grown in extracellular matrix are collected
shearing damage.
gel.
11. The gel pellet is washed twice with PBS (500-1000 rpm, 5
minutes).
12. The cysts can be stained at this stage with Hoechst to label the nuclei.
13. The gel is pelleted and as much of the supernatant as
possible is removed.
14. The gel pellet is mixed with low melting point
PBS
Triton X-100
Heating block
26
White Paper
Method
(Anopheles gambiae)
type of plankton.
Ethanol (70 %)
Glycerol (50 %)
Sucrose
Heating block
Fig. 19 Mounting a complete adult Anopheles gambiae. (A) The insect is paralyzed. (B) The chitin surface is treated using 70 % ethanol. (C) The insect is
positioned within the melted agarose. (D) The mounted insect is ready for imaging.
27
White Paper
4.2 Troubleshooting
ADDITIONAL INFORMATION
Common problems
4.1 Tips
LSFM is a fluorescence microscopy technique so many troubleshooting guidelines from other microscopy techniques apply
Protocols
contrast.
see if you can find a tip that you may apply to solve your prob-
lem. Be creative.
Safety
Lightsheet Z.1.
effect.
The light sheet comes from the side and any obstacle
illuminated.
imaging chamber in regard to the sample preparation technique, especially for embedding as gelling agents are used.
28
White Paper
The light sheet comes from the side and any obstacle
tube the more unstable it will be. You can also tighten
ing is active.
The cables are not properly secured with the cable holders
Optical aberrations
but this may decrease the overall signal. You can also use
stripes.
The use of image processing methods such as deconvolution may help to get rid of those aberrations.
29
White Paper
Pipettes
2 ml serological pipette, FALCON, BD Labware,
Chemicals
Agarose:
Molecular biology grade, for routine use, SIGMA,
Ref.35 7507
1 ml serological pipette, FALCON, BD Labware,
Ref.35 7521
Ref. A-9539
Type VII, low gelling temperature, SIGMA,
Ref. A-4018-50G
Agarose Low Melt (no 6351.1) from Carl Roth
Equipment
For the following, please refer to your local lab supply
companies:
Heating blocks
brunschwig@brunschwig.nl)
Stereomicroscope
Scalpels
Tweezers
Dissection Needles
Watch glass
Sonicator
30
White Paper
31
White Paper
Kalinka, A.T. et al., 2010. Gene expression divergence recapitulates the developmental hourglass model.
Nature, 468(7325), pp.811814.
Karakyl, E.M. et al., 2009. Copepod feeding quantified by planar laser imaging of gut fluorescence.
Limnology and Oceanography: Methods, 7, pp.3341.
Kaufmann, A. et al., 2012. Multilayer mounting enables long-term imaging of zebrafish development in a light sheet
microscope. Development, 139(17), pp.32423247.
Keller, P.J. et al., 2008. Three-Dimensional Microtubule Behavior in Xenopus Egg Extracts Reveals Four Dynamic States
and State-Dependent Elastic Properties. Biophysical journal, 95(3), pp.14741486.
Lorenzo, C. et al., 2011. Live cell division dynamics monitoring in 3D large spheroid tumor models using light sheet microscopy.
Cell Division, 6(1), p.22.
Maizel, A. et al., 2011. High-resolution live imaging of plant growth in near physiological bright conditions using light sheet
fluorescence microscopy. The Plant Journal, 68(2), pp.377385.
Mertz, J. & Kim, J., 2010. Scanning light-sheet microscopy in the whole mouse brain with HiLo background rejection.
Journal of biomedical optics, 15(1).
Olarte, O.E. et al., 2012. Image formation by linear and nonlinear digital scanned light-sheet fluorescence microscopy
with Gaussian and Bessel beam profiles. Biomedical Optics Express, 3(7), pp.14921505.
Pampaloni, F., Reynaud, E.G. & Stelzer, E.H.K., 2007. The third dimension bridges the gap between cell culture and
live tissue. Nature reviews Molecular cell biology, 8(10), pp.839845.
Ritter, J.G. et al., 2008. High-contrast single-particle tracking by selective focal plane illumination microscopy.
Optics express, 16(10), pp.71427152.
Rubio-Guivernau, J.L. et al., 2012. Wavelet-based image fusion in multi-view three-dimensional microscopy.
Bioinformatics, 28(2), pp.238245.
Santi, P.A. et al., 2009. Thin-sheet laser imaging microscopy for optical sectioning of thick tissues.
Biotechniques, 46(4), p.287.
Scherz, P.J. et al., 2008. High-speed imaging of developing heart valves reveals interplay of morphogenesis and function.
Development, 135(6), pp.11791187.
Schrter, T.J. et al., 2012. Scanning thin-sheet laser imaging microscopy (sTSLIM) with structured illumination and
HiLo background rejection. Biomedical Optics Express, 3(1), pp.170177.
32
White Paper
5 Index
A
Antifading Agents
21
F
FEP Tubing
15
20
L
LSFM
Sample Mounting
22
M
Materials and Equipment
Sample Chambers
15
16
Sample Holder
18
18
Hydrogel Preparation
19
S
Sample
Holding
Sample Preparation
Preparation of Fluorescent Beads
22
23
24
24
Samples
Embedded
Hanging
12
Enclosed
13
21
20
33