IAAF Calibration and Testing Manual

IAAF Calibration and Testing Manual


Edition 2010

Edition 2010

IAAF Calibration and Testing Manual

CONTENTS

INTRODUCTION................................................................................................ 3
GENERAL ......................................................................................................... 5
MEASUREMENT OF DISTANCE ....................................................................... 8
MEASUREMENT OF MASS ............................................................................... 12
WIND GAUGES ................................................................................................. 12
SYNTHETIC TRACK SURFACE ......................................................................... 16
PHOTO FINISH ................................................................................................. 16
STOPWATCHES ................................................................................................ 16
TRANSPONDER TIMING .................................................................................. 16
FALSE START EQUIPMENT.............................................................................. 16
HURDLES ......................................................................................................... 16
SAFETY CAGES NETTING................................................................................. 20
HAMMER HANDLE ........................................................................................... 20
ATTACHMENTS:
A - STOPWATCH CALIBRATION ....................................................................... 21
B - HAMMER HANDLE TESTING ..................................................................... 25
C - PHOTO FINISH CALIBRATION.................................................................... 29

Edition 2010

IAAF Calibration and Testing Manual

INTRODUCTION

The sport of athletics relies basically on the measurement of performance by


time and distance. Over the past century there has been a marked
improvement in the accuracy of these measurements. However, it is not
possible or necessary to measure in our sport to the accuracy needed in
science.
Over time, the IAAF has determined the required specifications for
implements and equipment and all these require measurements of one or
more of length, mass and time.
The purpose of this Manual is to indicate how measurement equipment is to
be calibrated and the accuracy required for the resulting measurements.
Also, suggestions are made as to how federations might more cheaply check
the accuracy of their measurement instruments.
There are national and international standards for commonly used
measurement equipment and also for specific measurements. However,
different levels of measurement accuracy may be permitted to suit the nature
of particular activities. Our sport also uses athletics specific equipment such
as false start equipment which must also be calibrated. One problem relating
to this equipment is that different manufacturers use different measurement
techniques and the measurements resulting may not be consistent between
manufacturers. This equipment is at present under serious research study
and will be added progressively to the Manual once the Technical Committee
is satisfied with the recommended calibration process.
Most countries have a national weights and measures authority, and/or a
national accreditation organisation that determines the laboratories /
organisations within the country that meet the accreditation requirements for
particular measurement activities. These organisations may be a valuable
source of information and assistance to National Federations.
The IAAF should welcome feedback from federations, organisations and
individuals on the information presented in the Manual so that the subsequent
editions can incorporate improvements deemed necessary as the result of
using the methods proposed in the Manual.
The purpose of this Manual is to provide testing organisations, federations
and manufacturers with clear advice on the testing and calibration methods to
be adopted for measuring devices, some equipment and materials used in
IAAF competitions.
Where possible, international standards or national standards are referred to
and complemented as necessary.

Edition 2010

IAAF Calibration and Testing Manual

It is intended that the Manual will be updated and refined as experience is


gained in use of the methods listed.
Any comments on the Manual might be addressed to:
COMPETITIONS DEPARTMENT
INTERNATIONAL ASSOCIATION OF ATHLETICS FEDERATIONS (IAAF)
Attention: Technical Manager, E-mail: technicalofficer@iaaf.org
17, rue Princesse Florestine, BP 359 - MC 98007 Monaco Cedex
Tel: (+377) 93108888 - Fax: (+377) 93159515 - Direct Fax (+377) 93503263

Edition 2010

IAAF Calibration and Testing Manual

GENERAL
Purchasing Measurement devices
Federations are advised to purchase all measurement devices with a
calibration certificate provided by an accredited authority so that the
measurements can be traced back to a national or international
measurement standard as required by IAAF Competition Rule 148. This may
increase the initial cost of the device but the additional cost will usually be
considerably less than having it calibrated after purchase.
Appropriate Testing Organisations
Certification, inspection and testing work on measuring devices used in IAAF
competitions should be undertaken by organisations that have demonstrated
their competence by having ISO 17025 accreditation by a nationally recognised
accreditation body.
Nationally recognised accreditation bodies are listed on the web pages of the
International Laboratory Accreditation Cooperation (www.ilac.org) and the
International Accreditation Forum (www.iaf.nu). The European Cooperation
for Accreditation is also a source of information on European accreditation
organisations.
Bureau International des Poids et Measures (BIPM) is the International
Bureau of Weights and Measures. On its website www.bipm.org are listed the
members. If a federation is from a country that is not listed by BIPM, it is still
probable that your government has nominated an organisation responsible for
national weights and measures.
Test and Calibration Reports
The test and/or calibration reports should provide all the necessary
information for verifying that the measuring device is suitable for use in IAAF
competitions.
The following information, as required by the international standard ISO/IEC
17025, should be included in each report:

A title (e.g. Test Report or Calibration Certificate)


Name and address of the laboratory, and the location where the testing /
calibrations were carried out, if different from the address of the
laboratory.
Name of the national accreditation body that accredited the laboratory.
Unique identification of the test / calibration document, including on each
page an identification to ensure that the page is recognised as part of the
document and a clear identification of the end of the document.

Edition 2010

IAAF Calibration and Testing Manual

Name and address of the client.


Identification of the method used.
Description, condition and identification of the item tested or calibrated.
Date of receipt of the test / calibration item where applicable and the date
the work was carried out.
Reference to the sampling plan and procedures used by the laboratory or
other bodies where applicable.
Results with, where appropriate, the units of measurement.
Name, function and signature or equivalent identification of the person
authorising the test / calibration document.
A statement on the estimated uncertainty of measurement where
applicable
A statement to the effect that the results relate only to the item tested or
calibrated where appropriate.
Where necessary for the interpretation of the test results, the following
shall be included:
1. Deviations, additions or exclusions from the test method, and specific
test conditions, e.g. environmental conditions.
2. A statement of compliance / non-compliance with requirements and/or
specifications.
3. Additional information required by specific methods or clients.
4. Opinions and interpretations where appropriate and needed.
5. Additional information required by specific methods or clients.

Calibration Certificates in addition to the information listed above shall


include also where necessary for the interpretation of calibration results:

Conditions, e.g. environmental conditions during calibration that may have


an influence on the measurement results.
Evidence that the measurements are traceable.
If a statement of compliance with a specification is made, the clauses of
the specification, which are met or not met, must be identified.
Where a statement of compliance is made omitting the measurement
results and associated uncertainties, the laboratory must record and
retain those results.
The uncertainty of measurement must be taken into account when
statements of compliance are made.
The calibration results before and after adjustment or repair, if available,
must be reported.
Calibration certificates or labels should not contain any recommendation
on the calibration interval except when requested by the client.

When testing and calibration results are obtained from sub-contractors, the
results of tests performed by sub-contractors must be clearly identified.

Edition 2010

IAAF Calibration and Testing Manual

Where calibration work has been sub-contracted, the laboratory performing


the work must issue the calibration certificate to the contracting laboratory.
Amendments to a test report or calibration certificate after issue must be in
the form of another document and include reference to the original document.
If a new test report or calibration certificate is required, it must be uniquely
identified and include reference to the original it replaces.

Edition 2010

IAAF Calibration and Testing Manual

MEASUREMENT OF DISTANCE
The devices covered by these regulations include steel measuring tapes, steel
bars, video distance measuring (VDM) and electronic distance measuring
(EDM) apparatus.
Steel Measuring Tapes
Applicable Standards:
ISO 8322-2 1999 Building construction - Measuring instruments - Procedures
for determining accuracy in use - Part 2 Measuring tapes, or
BS 4035 Specification for linear measuring instruments for use on building
and civil engineering construction works. Steel measuring tapes and
retractable steel pocket rules, or
AS 1290.5-1999 Linear measuring instruments used in construction. Part 5:
Coated and etched steel measuring tapes.
Tape Tension:
50m or shorter: 50N;
longer than 50m: 100N or the tape manufacturer's specified tensile force.
Test Temperature:
The calibration test shall be undertaken at a temperature selected by the
laboratory within the range 20 to 23 degrees Celsius and stable to 1.0 degree
Celsius during the test.
Required 95% Confidence Limit Uncertainty of a Single Measurement with
controlled temperature and tape tension:
(0.5mm + 0.1mm/m)
For measurements made by truncating, the measurement to the nearest
whole centimetre (as required by IAAF Rules), with the correct tension applied
to the tape and with a tape temperature in the range 10 and 30 degrees
Celsius without temperature correction then the uncertainty of a
measurement at a 95% confidence limit is: +5mm (5.8mm + 0.154mm/m).
Reading
To the next lower graduation if it is not an even graduation.
Graduation Test Intervals:
Every 10m for initial and subsequent calibrations.
Recalibration Intervals:
Every four years or after damage / repair of the tape.
Limitation on Use:
The calibrated steel tape should not be used for other than measuring
records or checking the accuracy of other working tapes.

Edition 2010

IAAF Calibration and Testing Manual

Advice for National Federations


Steel measuring tapes might be certified by a practicing land surveyor using
his standard measure which is traceable back to the national standard
measurement. Such certification should be undertaken when the tape is
purchased if it has not been purchased with a measurement calibration
certificate, after repair of the tape and every year at a temperature close to 20
degree Celsius.
The tape should be check measured by the surveyor every 10 metres, and
either side of a repair if the tape has been broken, against the surveyor's
standard measure. The temperature of the tape and the tension applied to the
tape (a tension of 5kg (50N) for 50m or shorter tapes and 10kg (10N) for 100m
tapes would be appropriate) should be recorded on the certificate.
It should be recognised that temperature has a big effect on measurement
with a steel tape. A temperature change of 9 degrees Celsius makes a 1cm
difference per 100m. If your steel tape measures 100m against the standard
measure at 20 degrees Celsius or lower then for all competitions where the
temperature is higher the tape will measure short and you can be sure that
the distance measured is not less than that shown on the tape.
Your calibrated steel tapes should not be used for everyday competition
measurements but be kept for measuring records, check measuring when
EDM or VDM is being used, and checking the accuracy of fibreglass tapes and
subsidiary measuring devices like steel measuring bars. When measuring
apply the appropriate tension to the tape using a spring balance attached to a
tape puller. The steel tape should be kept in a cool place and only brought
outside when a measurement is to be taken so as to minimise temperature
errors.
Steel Measuring Bars
Applicable Standards:
BS 4372 1968 Specifications for engineer's steel measuring rules.
Test Temperature:
The calibration test shall be undertaken at a temperature selected by the
laboratory within the range 20 to 23 degrees Celsius and stable to 1.0
degrees Celsius during the test.
Required 95% Confidence Limit Uncertainty of a Single Measurement with
controlled temperature:
(0.5 mm + 0.1 mm/m).
Reading:
To the next lower graduation if it is not an even graduation.

Edition 2010

IAAF Calibration and Testing Manual

Graduation Test Intervals:


Every 1m for initial and subsequent calibrations.
Recalibration Intervals:
Every four years or after damage / repair of the bar.
Limitation on Use:
The calibrated steel tape should be used for checking the accuracy of other
working measuring bars.
Vernier Callipers and Micrometers
Applicable Standards:
ISO 6906:1984 Vernier calipers reading to 0.02 mm or
BS 887:2008 Precision Vernier calipers. Requirements and test methods or
AS 1984-1997: Vernier Callipers (metric series) or
JIS B 7507:1993 Vernier, dial and digital callipers or
DIN 862 (1988-12) Vernier calipers requirements and testing.
ISO 3611:1978 Micrometer callipers for external measurement or
AS 2102-1989 Micrometer callipers for external measurement or
BS EN ISO 3611 Micrometer callipers for external measurement or
BS 870:2008 Micrometer callipers for external measurement or
JIS B 7502:1994 Micrometer callipers for external measurement or
DIN 863-1 (1994-04) Micrometer callipers for external measurement.
Test Temperature:
The calibration test shall be undertaken at a temperature selected by the
laboratory within the range 20 to 23 degrees Celsius and stable to 1.0 degree
Celsius during the test.
Required 95% Confidence Limit Uncertainty of a Single Measurement:
0.05mm
Reading
To the nearest 0.1mm.
Recalibration Intervals:
Every four years or after damage / repair.
Electronic Distance Measuring
Applicable Standards:
Australia: Regulation 80 of the National Standard Measurement Act 1960
Required 95% Confidence Limit Uncertainty of a Single Measurement:
(6mm + 4ppm.)

Edition 2010

10

IAAF Calibration and Testing Manual

Recalibration Intervals:
Every year or after damage / repair of the instrument.
Calibration: The government authority responsible for surveying usually have
one or more EDM calibration baselines available for use by surveyors or
alternatively the calibration is undertaken by government surveyors. A typical
EDM Calibration Handbook is produced by the ACT Government and is
available at:
www.actpla.act.gov.au/__data/assets/pdf_file/.../edm-handbook.pdf
An EDM instrument is calibrated on a baseline with permanent monuments at
known elevations and distances apart to determine instrument constants and
errors. There is usually a computer program available to determine the
instrument constants errors and their associated uncertainties.
Notes for federations: Off the shelf EDM are supplied to an industry maximum
standard deviation of (3mm + 2ppm). A multiplying factor of 2.0 has been
used above to convert the standard deviation to a 95% confidence limit. This is
a simple interpretation of the ISO Guide to the Expression of Uncertainty of
Measurement.
If you have a Total Station EDM that reads angles electronically it could have a
check and adjust inbuilt program that should be used periodically.
Before each competition your EDM should be:
Checked for any obvious damage
Checked for horizontal angle measurement accuracy by:
1. Bisecting a clear object about 100m away and record the
horizontal angle.
2. Change face and bisect the same object and record the
horizontal angle.
3. The difference between the readings should be 1800 00 00 the
manufacturers tolerance.

Check for vertical angle measurement accuracy by:


1. Bisecting a clear object about 100 m away and record the
vertical angle.
2. Change face and bisect the same object and record the
vertical angle.
3. The difference between the readings should be 3600 00 00
the manufacturers tolerance.

If in doubt have your EDM checked by the manufacturers local agent.


Road Running and Race Walking Course Measurement
Refer to the IAAF/AIMS The Measurement of Road Race Courses second
edition 2004, updated 2008.

Edition 2010

11

IAAF Calibration and Testing Manual

MEASUREMENT OF MASS
Mechanical or electronic scales that meet the specifications below may weigh
the mass of implements, and other equipment such as relay batons and
crossbars.
Weighing Machines
Applicable Standards:
As laid down by the national laboratory accreditation authority with
measurements to be undertaken by a laboratory accredited by the national
accreditation authority.
Test Temperature:
The calibration test shall be undertaken at a temperature selected by the
laboratory within the range 20 to 23 degrees Celsius and stable to 1.0 degree
Celsius during the test.
Required 95% Confidence Limit Uncertainty of a Single Measurement:
0.1g.
Reading:
To the next lower graduation if it is not an even graduation.
Test Intervals:
Every 0.5kg from 0.5kg to 10kg for initial and subsequent calibrations.
Recalibration Intervals:
Every year or after damage / repair of the machine.
Limitation on Use:
The weighing machine should be located on a firm flat surface and not be
moved.
Comment:
When implements and other equipment are being check weighed at the
stadium the mass shall be recorded to the next lower gram unless the mass
measured is an even gram.
The current calibration certificate shall accompany record applications for the
weighing apparatus used to check the mass of the implement before and after
the competition.
WIND GAUGES
Applicable Standards:
ISO 17713-1:2007 Meteorology Wind measurements Part 1: Wind tunnel
test methods for rotating anemometer performance.

Edition 2010

12

IAAF Calibration and Testing Manual

ISO 16622:2002 Meteorology Sonic anemometers/thermometers Acceptance test methods for mean wind measurements.
AS 2923 1987 Ambient Air Guide for measurement of horizontal wind for
air quality applications.
Test Temperature:
The calibration test shall be undertaken at a temperature selected by the
laboratory within the range 20 to 23 degrees Celsius and stable to 1.0 degree
Celsius during the test.
Required 95% Confidence Limit Uncertainty of a Single Measurement
Reading:
To the next higher graduation if it is not an even graduation.
Recalibration Intervals:
Every four years or after damage / repair of the anemometer.
Limitation on Use:

IAAF Product Certificate Test


1. The gauge and its working shall be examined critically to confirm that it
complies with the intent of the IAAF Rules and construction regulations.
2. The manufacturer shall adjust the gauge so that it accurately measures a
velocity of 2m/s parallel to the measuring direction over an interval of 10
seconds.
3. Velocity calibration: +0.5, 1.0, 1.5, 2.0, 3.0, 4.0m/s for a 10 second time
interval parallel to the measuring direction.
4. Angle calibration: -60 to +60 at 30 intervals to the measuring direction
with wind tunnel air velocity being such that the velocity in the measuring
direction is +2m/s over a 10 s time interval.
-600 and +600 angles with the wind tunnel velocity 4m/s
-300 and +300 angles with the wind tunnel velocity 2.3m/s
5. Additional Time calibration: At timing intervals 5 and 13s at velocity +2m/s.
6. Anemometer Sensitivity test: Undertake the test in fluctuating light winds
of between 1 and 4m/s in an open area free of obstructions by comparing
the averaged readings from the instrument, orientated so that positive
readings are obtained, with those obtained by say a Laser Doppler
Anemometer with traceable calibration over a period of one hour.
7. Wind gauge calibration shall only be undertaken by a laboratory that is
accredited by the national laboratory accreditation organisation
recognised by the International Weights and Measures Organisation
(BIPM).

Edition 2010

13

IAAF Calibration and Testing Manual

Individual Wind Gauge Calibration


Each wind gauge on sale shall be supplied with a calibration certificate stating
the uncertainty of calibration for a test over a 10 second interval at a velocity
of + 2.0m/s. Such calibration certificates may be provided by the manufacturer
if their test facility has been accredited by the national accreditation authority.
Recalibration
Each wind gauge shall be tested over a 10 second interval at a velocity of
+2.0m/s every four years. Simple test procedures that federations may
consider are given hereunder.
Comparison Testing
A wind gauge that has a current traceable calibration certificate may be used
as a reference measure for other wind gauges. The wind gauges to be
checked and the reference wind gauge should be set up in an open field when
there are constant light winds with the measurement directions parallel to
each other and the general wind direction.. At least five comparison measures
of wind speed in the range of 1 to 3 metres per second but preferably as close
to 2 metres per second as possible over a 10 second interval should be taken.
Provided the average difference between the wind readings for the wind
gauge being checked compared with the wind speeds of the reference wind
gauge is not more than 0.1metres per second then the wind gauge being
tested may be certified as being in calibration. An appropriate sticker should
be attached to the wind gauge indicating the date of test, the testers name
and the reference wind gauges identification number. The test results should
be filed by the Technical Manager for future reference.

Edition 2010

14

IAAF Calibration and Testing Manual

Approximate Check 1 of Wind Gauge Accuracy


1.

Locate a long corridor or hall in which all the doors can be closed, there
is no ventilation air movement and is unoccupied for the period of the
testing.
2. Mark the start and finish of a 20 metres straight line on the floor.
3. Check that you can walk the 20 metres in 10 seconds by using a
stopwatch it is a fast walking pace. It does not matter if you walk the
distance slightly quicker than 10 seconds but it cannot be longer than 10
seconds. Therefore it is desirable that a second person time the 20m
walk every time to ensure that time taken is less than 10 seconds.
4. Set the wind gauge at the 10 second measuring interval.
5. Hold the wind gauge on your shoulder or under your arm with the wind
gauge parallel to the ground, vertically orientated as it is when in
athletics use and parallel to the straight line you have marked with the
arrow indicating wind direction towards you so that you are measuring
positive wind which is the one of most interest for record purposes It
would be helpful for another person to check the positioning of the wind
gauge for you
6. Start walking several metres before the start line to get the gauge vanes
moving switch on the gauge at the Start line and walk from the start to
the finish of the 20 metres at a brisk walking pace and stop dead on the
20m mark. The wind gauge will automatically record the wind speed
after 10 seconds. As you have walked 20 metres in 10 seconds the actual
average wind speed is 2.0 metres per second.
7. Read and record the wind speed indicated by the wind gauge.
8. Repeat the process walking in the opposite direction and record the wind
speed result.
9. The process may be repeated again and the average of all the wind
speeds measured should be recorded against the identification number
of the wind gauge.
10. If the average wind speed is between 1.9 and 2.1 metres per second then
prima facie the wind gauge is in calibration. If it measures between 1.8
and 2.2 metres per second then the use of the wind gauge should be
limited to less important meetings. Outside the latter range then the
wind gauge should be repaired if the cost is not prohibitive.
Check 2 of Wind Gauge Accuracy
As for Check 1 above but instead of hand carrying the wind gauge, install the
wind gauge on a flat bed trolley that can be pushed or pulled smoothly along a
track.
The timing can be done by stopwatch as above or by photo-electric cell timing.
A more elaborate test set-up might have the trolley driven by a battery
powered electric motor.

Edition 2010

15

IAAF Calibration and Testing Manual

Approximate Check of the Wind Gauge Timing Device


Older vane wind gauges of the Cantabrian type use resistor-capacitor
networks to time and these can drift with age. To test the gauge timing
accuracy press the wind gauge start button and start a stopwatch
simultaneously. Stop the watch when the LCD display on the wind gauge first
appears. The stopwatch time should be between 9.9s and 10.2s to allow for
the stopwatch operators reaction time.
SYNTHETIC TRACK SURFACE
The required test procedures are given in Chapter 3.1.3 of the IAAF Track and
Field Facilities Manual.
PHOTO FINISH
See Attachment C.
STOPWATCHES
See Attachment A.
TRANSPONDER TIMING
Still to be proposed.
FALSE START EQUIPMENT
Still under consideration.
HURDLES
Tilting Test
The hurdle tilting test shall be undertaken as per Rule 168.2 using either a
spring balance graduated in kgf or by applying the horizontal force by hanging
a 3.6kg weight over a pulley. The tilting force is the force just required to tilt
the feet of the hurdle from the level ground. Note that as the hurdle tilts
further the force required to continue the overturning becomes less. The
hurdle should be tested at each hurdle height with the foot weights adjusted
for the different heights.
The accuracy of the spring balance should be verified before commencing the
test by hanging weights of 3.0kg and 4.0kg separately on the spring balance
and noting the readings obtained by the spring balance.

Edition 2010

16

IAAF Calibration and Testing Manual

Top Bar Deflection Test

Performance Requirement
With the hurdle feet held to the ground, measure the horizontal deflection at
the centre of the top edge of the top bar at its mid-span, when applying 10kg
load horizontally at this point, taking up any slack in the uprights before the
zero measurement. For compliance the amount of this measured deflection
shall be less than the 35mm permitted by Rule 168.2.

Testing method
Alternative methods of testing are possible. One acceptable method is
detailed below:

Acceptable Method
Purpose: This in-situ hurdle testing procedure provides a simple technique to
ensure compliance with Rule 168.2. The intention of this in-situ procedure is
to measure the horizontal deflection of the top bar (including any deflection of
the vertical supports), in the assembled hurdle.

Equipment required

Two hurdles (one to support measuring set up)


Hand held spring balance to carry a minimum of 10kg load (as an
alternative a 10kg weight can be suspended over a pulley).
An appropriate connection for loading at the top edge of the top bar (a
block of wood shaped to slide over both sides of the top bar, with an
attached hook lining up with the top of the bar has been used in the
example illustrated).
Steel measuring tape.
A resistance load to stop the loaded hurdle from tipping over
(conveniently provided by one or two other people).
A bag to support a 10kg load.
A combination of certified shots and discus are suggested to provide a
10kg load.
Sticky tape to attach the measuring tape.

Edition 2010

17

IAAF Calibration and Testing Manual

Illustrations:

Edition 2010

18

IAAF Calibration and Testing Manual

Procedure:
1. Calibrate the spring balance in the vertical direction using the bag with
the 10kg load.
2. Set up two hurdles as illustrated. The testing shall be done in the
normal direction of hurdling. The hurdle being tested should have the
top bar at the Mens competition height. (There is no need to test
hurdles at a lower height.)
3. For the simple reading of the horizontal deflection the steel measuring
tape is fixed (sticky tape) to the top bar of the second hurdle in the
position as illustrated, and the second hurdle is set at a height that
allows the steel tape to directly line up with the connection block (or
similar arrangement),
4. Apply load stability (one or two people) to the feet of the hurdle being
tested.
5. Take up any slack in the uprights before the zero measurement.
6. Apply the 10kg load horizontally to the centre of the top edge of the top
bar at its mid-span.
7. Measure and record the resulting horizontal deflection using the
second hurdle as the fixed datum.
Results: Variations can be expected within a set of hurdles of similar
manufacture:

Edition 2010

19

IAAF Calibration and Testing Manual

At higher temperatures deflections may increase (Testing should be


done at temperature conditions typical for local competitions)
Material stiffness varies.
Assembly and bar fixing details have variations.

Procedure described by Don Mackenzie, Rod Syme and Alan Tucker (NZL).
SAFETY CAGES NETTING
Hammer
The netting mesh size shall be measured to check that the maximum mesh
size is 0.05m for steel wire netting and 0.044m for cord netting. The mesh
size, strand cross-sectional area and material composition shall be recorded.
A single strand of the netting tested in a tensile testing machine at a straining
rate of 50 mm/minute for fibre and 5mm/minute for steel wire at a
temperature of 21 degree Celsius shall have a minimum breaking strength of
300kgf (3000N).
Alternatively the energy absorption of the mesh shall meet the dynamic test
principles laid down in EN1263-1:1996 Safety nets Part 1: Safety
requirements, test methods such that the netting will withstand an object
100kg mass being dropped into the netting from a height of 7m. This equates
with the energy absorbed when a 7.26kg hammer travelling at 32m/s is
stopped by the netting allowing a safety factor and an ageing factor.
Discus
The netting mesh size shall be measured to check that the maximum mesh
size is 0.05m for steel wire netting and 0.044m for cord netting. The mesh
size, strand cross-sectional area and material composition shall be recorded.
A single strand of the netting tested in a tensile testing machine at a straining
rate of 50mm/min for cord and 5mm/min for steel wire at a temperature of 21
degree Celsius shall have a minimum breaking strength of 40kgf (400N).
Alternatively the energy absorption of the mesh shall meet the dynamic test
principles laid down in EN1263-1:1996 Safety nets Part 1: Safety
requirements, test methods such that the netting will withstand an object
15kg mass being dropped into the netting from a height of 7m. This equates
with the energy absorbed when a 2kg discus travelling at 25m/s is stopped by
the netting allowing a safety factor and an ageing factor.
HAMMER HANDLE
Testing (see Attachment B)

Edition 2010

20

IAAF Calibration and Testing Manual

ATTACHMENT A
STOPWATCH CALIBRATION
Introduction
The international standard for evaluating the accuracy of quartz watches is
given in ISO 10553:2003 Horology Procedure for evaluating the accuracy of
quartz watches. The procedure designed for quartz watches with an
indicated accuracy of less than 30 seconds per annum involves taking
measurements over three days with a three day stabilisation period before
and a twenty four rest period before the next stabilisation/test period. With
three separate three day test periods the total measurement period for this
test phase is 66 days and this is followed by a temperature simulation test
program over thirteen days. As a typical stop watch may have an indicated
accuracy of 6ppm i.e. about 0.5s per day such a test is not appropriate for
competition stopwatches apart from being too expensive.
The calibration test method given below was suggested by the Centre for
Sports Technology London and their permission to use their work is gratefully
acknowledged.
Procedure
Determine an appropriate verifiable source of UTC time. This could be from a
talking clock provided by the local telephone service, your computer or a GPS
timer. The detailed procedure that follows is for a telephone service.
Guidance for using other systems follows.
Reset the timer being checked to zero.
Whilst listening to the time on the telephone, start the timer at the third of a
set of strokes or when the timing source reaches a full minute. Note the time.
For long periods using the telephone time service, it is helpful to choose to
start the clock when a full minute is reached at the third stroke.
When the required time has elapsed, stop the timer.
Where long intervals of time are involved, the telephone may be replaced and
the call made again near to the time needed. The telephone will, in any case,
disconnect automatically after a maximum of one minute in Australia but at
different intervals elsewhere. Note that these days there is usually a charge
for the telephone time service.
Repeat the test at least three times but preferably four or more times for
each time interval so that the calculated standard deviation of the mean of the
errors will be reasonably accurate.

Edition 2010

21

IAAF Calibration and Testing Manual

Record the time recorded by the timer for each of the tests.
It would also be acceptable to compare a timer against another timer that has
been calibrated professionally and is traceable back to a national or
international standard. This could be a photo-finish timing system. Both
timers would be started and stopped simultaneously, and the times
compared. As stated above preferably five or more separate readings would
be taken at the maximum time interval at which the timer being calibrated
records the time in 1/100ths of a second.
Calculations
The error is the difference between the stopwatch reading and the reference
time interval. Record the error of the time for each test interval including the
sign of the error, + or . Calculate the mean and the standard deviation for
each time interval. The mean time difference from the timing interval should
not exceed the equivalent of 1 second per day. If the time difference exceeds
this amount then return the watch to the supplier if it is under warranty. If the
warranty has expired do not use the watch for competition timing. The
standard uncertainty for the mean time is the calculated standard deviation
for the mean time interval divided by n where n is the number of readings
taken at the time interval. This is essentially the standard deviation of the
difference between the start and stop reaction times. In the following
uncertainty calculation a value of 0.050s has been used. The actual value
obtained from the measurements should be used. The standard uncertainty
value obtained should not exceed 0.010s. If it does exceed that figure then the
whole test procedure and particularly the start / stop procedure should be
reviewed or a different tester used.
Uncertainty (Precision and Accuracy)
The UK Speaking Clock is stated to be accurate to 5 ms and is corrected as
necessary, twice a day, by reference to the NPL atomic clock.
These values for accuracy refer to the difference between the time given by
the Speaking Clock and the true time, as indicated by the NPL Clock. Over
shorter intervals, provided the going rate of the clock is constant, errors
should be pro-rata.
In other countries the speaking clock accuracy may be less as it will depend
on the delay added by the telephone service as this can be variable as the
information is possibly routed differently each time you call for a time. As an
estimate the delay might be taken as 15 milliseconds and a semi-range of 15
milliseconds might be achievable in Australia.
Using a programme from Dimension 4 it is possible to synchronise your PCs
clock for Windows-based operating systems using a suitable source server
preferably as close to your computer as practicable to minimise the number
of routing nodes. The synchronising frequency can be set by you. Also note

Edition 2010

22

IAAF Calibration and Testing Manual

that the network delay is determined by a one-way measurement that


assumes that the send and return paths are the same. Due to the vagaries of
the software clocks and the internet connection the errors are likely to be of
the same order as for the talking clock but it is a cheaper option certainly in
Australia.
A GPS timer might also be used as the source timer if it has been
synchronised with an UTC source. However as cheap GPS receivers are
generally primarily concerned with position the time display could be delayed
by a variable amount.
The uncertainty of the time recorded will depend upon:
variations in the speed of reaction of start and stop of the user (you will
recall that the standard adjustment of stopwatch manual times for
races 400m and less for comparing with automatic times is +0.24 s to
allow for operator reactions);
at least one digit with digital timers;
the constancy of the speaking clock or other timing reference going
rate.
An uncertainty budget for the calibration using the talking clock in Australia is
given below.
UNCERTAINTY BUDGET FOR AN ELECTRONIC STOPWATCH CALIBRATED AGAINST A TALKING CLOCK IN
AUSTRALIA
Component
number

Uncertainty
component

Nominal
Value

Type

Uncertainty
comment

Distribution

Standard
uncertainty

Variance

Speaking
clock
accuracy

15 ms

Uncertaint
y
is
independe
nt of timed
interval =
8.66 ms

Rectangula
r

0.00866

0.0000750

Start/stop
error

50 ms

sd
of
difference
in reaction
times

Normal

0.05

0.0025

Digitisation
error
stopwatch

10 ms

Resolution
is 0.01 s

Rectangula
r

0.002887

8.333E-06

Timebase
variations

i.e.
crystal
oscillator
frequency.

Assume 10
ppm

For 100 s =
0.001 s

Rectangula
r

0.00057735

3.333E-07

Combined
standard
uncertainty

0.051

Uncertainty
95% CI

0.102

Edition 2010

23

IAAF Calibration and Testing Manual

It can be seen that the largest effect by far on the uncertainty of measurement
is operator start / stop error cf. its standard uncertainty (0.050ms) with the
combined standard uncertainty of 0.051ms.
Start / stop errors will vary from person to person. For one person the errors will
tend to be constant either always early or always late so there is a cancellation
effect.
Mechanical watches and clocks going rate most probably will vary with
temperature, the degree to which the watch is wound and physical movement of
the watch particularly rotational. Mechanical timers must always be handled with
great care, be full wound before use and be kept at constant temperature. These
old technology timing devices should no longer be used for competition timing.
With electronic timers, the time-base is provided by a quartz crystal oscillator.
These oscillators are selected for their stability and constant frequency, though
they can be affected by temperature. Significant temperature variations cannot
occur over short periods since the crystal is physically located within the clock or
watch. If the frequency is constant over the calibration period a valid calibration
can be carried out using periods much longer than the periods the watch is used
to measure. Since as shown above the significant errors listed above are mostly
constant and give rise to a constant-time uncertainty in the measured time, the
longer the calibration period, the smaller the calibration uncertainty expressed
as a percentage of the measured time. Therefore it is recommended that the
calibration test interval be 24 hours or the maximum time at which the timer
reads to 1/100ths of a second with the time being read and recorded to 1/100ths
of a second. Federations are advised to purchase digital watches that continue to
time in 1/100ths of a second even if the hours digit returns to zero after 10 hours.
A summary of the results achieved in the testing should always be kept in the
Federation Equipment Manual recording the errors indicated for future
reference.
Check the results of the measurements against the acceptability criteria for the
timer being calibrated.

Edition 2010

24

IAAF Calibration and Testing Manual

ATTACHMENT B
HAMMER HANDLE TESTING
Test Requirements
The IAAF handle test requirements are:
Hammer handles shall have load applied in a universal-testing machine at
a uniform rate with the extension recorded until destruction.
The gripping of the handle shall simulate in simple terms the grip by a
thrower.
A load extension curve shall be produced for each handle tested, the
ultimate strength recorded, and the point and type of failure recorded.
The physical details of the handle shall be recorded including handle make
and model, length and width of the handle, the weight and the material
composition of the handle if known.
Handle Grip

Brace

Thimble
Loop

Figure 1 Generic Hammer Handle Showing Nomenclature Used


Test Rig Design
The test rig used at the School of Aeronautical, Civil and Mechanical
Engineering (SACME) ADFA Campus of the University of NSW to meet the
above requirements was designed to load the hammer handles in uniaxial
tension on a hydraulic universal testing machine (250kN capacity Instron in
SACME). The test rig consisted of two parts: an upper component on which
the handle grip is mounted and a lower mount to which the loop of the handle
is attached. The component parts of the mounts are shown in Figure 2.

Edition 2010

25

IAAF Calibration and Testing Manual

Interface to
Instron grips
for thin
Handles

Large diameter
pin with larger
interface to for
Thicker
Handles

Loading
hook

Rollers for Thin


Handle Grips

Modified
loading pin
connecting to
handle eyelet

Rollers for
Thick Handle
Grips

Figure 2. Components of Modified Test Fixture


A loading pin is inserted through the thimble or the loop in the brace at the
apex of the hammer handle and mounted in a yoke that fits in the Instron
lower grip.
The centrifugal pull from the handle wire is transmitted through the two side
braces as lateral forces at the two ends (or the brace joints) of the handle
grip. In the real situation these lateral forces on the handle grip are reacted
by a distributed load provided by the fingers and the palm of the thrower. The
actual distribution of the load on the handle grip may not be uniform but is
likely to vary from person to person depending on individual variations in
gripping. However, it appears reasonable that the two outer fingers are likely
to provide greater support than the inner two fingers in a natural grip with the
four fingers spread symmetrically along the handle grip. This situation is
simulated in the loading rig by employing two rollers to provide simple
support symmetrically at quarter points along the span of a 130mm wide
handle grip. Note that the same test rig can be employed for testing hammers
with straight handles as well as curved handles. The positioning of a hammer
grip over the rollers is shown in Figure 3.

Edition 2010

26

IAAF Calibration and Testing Manual

Figure 3. Handle in the Modified Test Fixture


(Note the thick ridge along the base of the grip)
The loading of the handle grip is shown schematically in Figure 4 along with
the bending moment distribution along the span of the hammer grip. It may
be noted that the bending moment due to the lateral load is then uniformly
distributed across the central half of the span and has a value of PL/8, where
L is the length of the handle grip (nominally 130mm). This situation is less
severe but more realistic than the bending moment distribution provided by a
point load at the centre, and at the same time easier to simulate than a
uniformly distributed support along the mid-span of the beam.

Edition 2010

27

IAAF Calibration and Testing Manual


P/2

P/2

P/2

P/2

L/4

L/4

CLsym L/4

L/4

Bending Moment Distribution

PL/8

Figure 4. Schematic of Loading of Handle Grip and bending Moment


Distribution
Test Procedure
The Hammer Handle is mounted on the upper and lower mounts as described
above which were inserted into the jaws of the Instron Hydraulic Test
machine. The test data consisting of the applied tensile load and the axial
displacement (elongation of the hammer handle) are recorded digitally via a
data acquisition board into a PC for processing. All specimens were initially
loaded up to about 4kN (approximately the working load, i.e. the maximum
load expected to be exerted on the hammer handle when used in
competitions) at a slow loading rate of 1mm/min, and the load and
displacement data collected at a sampling rate of about 4 samples/s. The
specimens are then unloaded and inspected for any permanent deformations.
An intermediate load step of up to 8kN is employed if the handle is capable of
withstanding this load. Finally the specimen is loaded until failure occurs. For
the final tests (loading up to failure) a typical displacement rate of 5 mm/min
and sampling rate of 20Hz (20 samples/s) is employed. The failure mode of
the hammer handle is observed and recorded photographically.
Performance Requirements
Using the above test procedure the total deformation of the handle in the
direction of the load under a tension of 3.8kN shall not exceed 3mm and the
minimum handle breaking strength shall be 8kN.

Edition 2010

28

IAAF Calibration and Testing Manual

ATTACHMENT C
PHOTO FINISH CALIBRATION
Introduction
The following calibration procedure is proposed to show that the equipment
design meets the IAAF Product Certificate requirements. A shorter test may
be appropriate for the initial calibration of individual items that have a generic
Product Certificate and for subsequent calibrations.
The calibration shall only be undertaken by a laboratory that is accredited by
the national organisation recognised by the International Laboratory
Accreditation Cooperation (ILAC).
The photo-finish equipment and its working shall be examined critically to
confirm that it complies with the intent of the IAAF Rules and construction
regulations.
Recommended Calibration Procedure
The calibration test shall be undertaken at a temperature selected by the
laboratory within the range 20 to 23 degrees Celsius and stable to 1.0 degree
Celsius during the test.
An electrical pulse referenced to a traceably calibrated frequency standard
shall simultaneously start the clock of the photo-finish equipment and
operate an electronic flash or LED located 1.50 m in front of the camera. The
flash or LED is to operate at the end of the selected test time interval.
The flank of the flashed light shall be evenly distributed over the line sensor
or equivalent. The duration of the flash shall be such as to reproduce as a
vertical line 10 pixels wide. For example, at an acquisition rate of 1000 lines
per second this is 1/100 second.
The frequency standard shall produce timing pulses that shall have an error
not exceeding 1 in 109 compared to International Atomic Time. This is to be
verified with an uncertainty not exceeding 1 in 1010. Tests shall be done every
15 minutes for 24 hours and shall have errors in the generated times of no
more than 0.25ms. The times shall be recorded to a resolution of 0.5 ms or
better. The calibration certificate shall give the generated time intervals and
corresponding measured time intervals given by the equipment timer with
uncertainties that are determined using the procedures given in the ISO Guide
to the Expression of Uncertainty in Measurement and shall include a
component for the stability of the equipment under test.
The time interval error shall not exceed 40ms for the 24 hour interval and
proportionally less for shorter times. The uncertainty of the tests (95%

Edition 2010

29

IAAF Calibration and Testing Manual

confidence interval) shall not exceed 0.051ppm of the time interval or 0.6
milliseconds in 15 minutes whichever is the greater.
This should ensure that the uncertainty does not exceed 5 ms for times up to
the maximum 24 hour interval. The allowable error reaches 1.2 ms at a time
interval of 1 hour.
Alternative methods of calibration may be proposed to suit the available
laboratory equipment to meet the uncertainty expectations above including:
Generate a train of impulses, at TTL level into a 50 ohm load, of duration 900 s
and multiples of this time, synchronised on the negative going edge by a 1
PPS signal referenced to UTC. Use these pulses to generate test time
intervals and to operate a LED as well as initiating the sequence of time
recording of the photo-finish camera.
The operation of the LED will be filmed by the photo-finish camera which
records the time of each impulse with a resolution of 0.5ms. Each camera
image has a time mark.
To extract the results, the time of the trigger for each flash of the LED may be
manually placed on the image record using software provided by the photofinish supplier.
Additional Calibration at Varied Temperature
Additional calibration shall be taken at approximately 10, 30 and 40 degrees
Celsius stable within 1 degree Celsius during the test to determine how well
the camera oscillator is temperature compensated.
After the test temperature has stabilised the camera shall be tested at ten
minute intervals for at least one hour using 0.001 second resolution.
PHOTO-FINISH COMPARISON TESTING
Introduction
The following procedure might be used by federations to verify that their
photo-finish equipment is operating correctly and are of acceptable accuracy.
This procedure can be used for comparing two or more cameras. It would be
preferable if at least one of the cameras has a current calibration certificate.
Procedure
Starting Signal Synchronisation
The starting signal times for all cameras should be synchronised so that the
times match. If over the period of say one hour from receiving the same start

Edition 2010

30

IAAF Calibration and Testing Manual

signal the cameras being compared the times have drifted apart by more than
the manufacturers drift specification of say 1ppm i.e. 3.6ms in one hour then
it indicates a faulty piece of equipment that must be identified and rectified.
It is worth noting that there can be a problem with the timing system even
though it has satisfied a zero test.

Edition 2010

31