JPD-100A

Fetal doppler
OPERATE MANUAL

Jumper Medical Co., Ltd

Copyright

Copyright belongs to Jumper Medical Co.,Ltd,


our company can reserve all the copyright of expert puissance according to
the legality.With out the company script permission,no one is allowed to increase
and decrease,brief,pirate,modifyorcopy.

2005

Table fo content

Open and check............................................. ........1


After sales service company...................... ............1
Notice,caution and suggestive instruction.............. .......2

Specification...........................................................2
Main configuration..................................................3
Function keys introduction......................................4
Operate manual......................................................5
Stock and transport................................................6
Cleaning.................................................................6
Safety sort..............................................................6
Apply range............................................................6
In case of difficulty..................................................7
Circuitry principium................................................8

Unpacking

Unpacking the unit carefully and make sure that all accessories are
put aside so they will not be lost.Examine the unit for any possibility
of shipping damage.If your unit is damage or fails to operate,notify
our dealer or our company immediately.Please offer the model number,
S/N,date of buying and contact information when you contact us.

Standard packing list


1.mainframe(including Probe)...................1 set
2.adapter ...................................................1 pc
3. Specification.............................................1 pc
4. Warranty card..........................................1 pc
5. Qualify card..............................................1 pc

If you have speical request,please see the contract.

After service company


After service company:Jumper Medical Co., Ltd
Add: 2nd Floor, No. 5 Bldg., Nanshui Industry Village,
5th Industry Rd., Shenzhen 518067 P.R. China
Tel:86-755-83547883 Zip:518067

Notice,caution and suggestive instruction


1.There is criticallycircuitry inside ,do not open it by yourself.
2.You can use the litmusless cleanser to clean the surface,do not put it in high
temperature or use corrosive cleanser ,you can use a few of alcohol to clean the
probe
3.When you finish using it,use the cloth to wipe it.
4.We suggest using the high-powered alkalescent battery,the tinpot battery may
creepage and damage the equipment .
5.If you will not use it for a long time,take off the battery ,please dispose the used
battery according to the local governments request.
6.Do notcollided the equipment lest damage it.
7.Please keep the equipment waterproof and dampproof.

Specification
Ultrasound operating frequency
Integrative sensitivity

2.0MHz

3%

200mm far from the probe

Space pink value press

17.38mW/cm

no less than 90db

Valid area of the ultrasound energy-transfer component


5mW/cm2

Ultrasound output power


FHR range: 50
Power supply
Dimensions

1cm2

210bpm
DC 7.2V

240

160

115mm

Mainframe N.W. 0.25Kg


Speed

2
Gel request Density 1.0g/cm

Impedance

1.7

1.7m/s

Attenuation

10 g/cm.s

0.02db/mm

Note:Jumper follows a policy of continuous advancements in development.


For this reason specification may be changed without notice.

Main structure
Head-on of the mainframe

Adapter

ositio
m this p
Insert fro

10

Back of the mainframe


Probe
LCD display
11

Switch
Turn up button
Turn down button
Manual calculation start and stop button

12

Monitor mode switch button


Manual calculation start and stop button
9

Loudhailer hole

10

Charge junction

11

Battery storehouse

12

Label

Function introduce

13

Symbol of monitor model

13

14

FHR display area

14

MODEL

Monitor mode switch button,it has 3 models and sign by1,2,3


Model 1:instantaneous auto calculate FHR
Model 2:calculate average FHR per 10 seconds(easy to check the
baseling of FHR)
Model 3:Manual calculate FHR

Press the button can switch the monitor model,when you turn
on the equipment,acquiescence model 1.
Turn down button,press it,the volume reduced until no voice,
when the volume is minimum,press it will without voice.
Turn up button,press it ,the volume increased,when the volume
is maximun,press it will without voice.
Manual calculation start and stop button,the two buttons with
the same function and only valid in model 3.when it is in model
3,press the button and wait until 10 times fetal heart beat,then
press it again,FHR display area will display the manual calculate
result.

Operate manual
Auto measure the HR
1.Turn on
Press

For 2 seconds ,the equipment will turn on and self-check,when the

HR appear ---,,the self-check is finish and get into use,model 1 is default.


2.Detect the FHR
Put the Gel on the detect position of the pregnant womansabdomen ,then
press the probe ensure it is good connected ,when the probe is on the right
position,you can hear the fetal heart beat,wait until the voice is clearly,the LCD
will show the right FHR.
3.Adjust the volume
The volume will increase,reduce when you press the
During the FHR is showed,if the volume is in maximum or minimum,continue to
press it will useless ,and without voice.
4.Turn off
When you do not use it, press
turn off if it is no signal,press

for 2 seconds to turn off it. The equipment will auto

to restart it.

Handmade measure HR
1.Turn on the equipment and press ,choose the model 3--handmade
measure HR
2.Detect the FHR
Put the Gel on the detect position of the pregnant womansabdomen ,then
press the probe ensure it is good connected ,when the probe is on the right
position,you can hear the fetal heart beat,press

or , According to the voice

from the Loudspeaker 10 times later,press

again,the LCD will show the

or

. average FHR,repeatit if you want to continue it


3.Change to auto measure.
Press

can change to auto measure model.

4.Turn off
When you do not use it, press
will autoturn off if it is no signal,press

for 2 seconds to turn off it. The equipment


to restart it.

Stock and transport


Stock environment:temperature 5
Humidity

--45

9%--45%

The product can be sent by air ,sea and land-carriage,but cant place it
in the open air.

Cleaning

Mainframe:use the litmusless cleanser or alcohol to clean the surface,dont


let the liquid get into the inside of the device,especially the LCD display
windows.
Probe:use the litmusless cleanser or alcohol to clean the surface,the line and
sheath can use the litmusless cleaser to clean it.

Safety sort
The safety sort of the product is :II Before

Apply range
Suit for fetus more than 16 weeks old

In case of difficulty

What apprance to be a malfunction may not always serious,if your unit


should not perform as expected,consult the table below to see if the
problems can be corrected before seeking help from your dealer or service
representative.

Symptom
Audio scream

Remedy

Cause
Battery is exhausted
Audio volume too high
Too much Gel on the
probe surface

Audio volume too low


Weak sound output

Low sensitivity

Insufficient Gel
Battery is exhausted

Change the battery


Turn down the volume

Use less Gel


Turn up the volume
Add Gel
Change the battery

Incorrect probe position

Locate the correct position

Insufficient Gel

Add Gel

circuit diagram