Professional Documents
Culture Documents
DIGITAL THEODOLITE
DT-101
Foreword
Thank you for purchasing the TOPCON DIGITAL THEODOLITE, DT-101. For the best
performance of the instruments, please carefully read these instructions and keep them
in a convenient location for future reference.
CONTENTS
1
SELECTING MODE...............................................................................................7-1
7.1
7.2
8
9
ANGLE MEASUREMENT......................................................................................3-1
3.1
3.2
3.3
3.4
3.5
9.1
9.2
9.3
9.4
9.5
9.6
11
12
13
14
15
16
17
Sighting collimator
Connector (to be
connected to EDM)
Objective lens
Instrument
center mark
Horizontal motion
clamp
Horizontal
tangent screw
Optical plummet
telescope
Display
Lower horizontal
motion clamp
1-1
Hand grip
Vertical motion
clamp
Plate level
Vertical tangent
screw
Operating keys
Circular level
Base
Leveling screw
Tribrach fixing lever
1-2
1.2 Display
Display marks
Display
Contents
Display
Contents
V-angle
HR
H-angle right
EDM working
HL
H-angle left
Horizontal distance
Ht
Relative elevation
NAVG
Slope distance
REP
N coordinate
TILT
E coordinate
Z coordinate
Percent grade
SO
Stake-out measurement
FUNCTION
Keys
FUNCTION
Keys
R/L
REC
HOLD
REP
FUNC
Illumination
ON/OFF
ANG%
V-angle display
Selection for vertical angle/percent
display
0 SET
of
display and
1-3
reticle
Output
Angle mode
V, HR (or HL)
Coordinate mode
N, E, Z, HR (or HL)
Hm(Average), Hn(Total)
The details necessary for the connection with the DT-101 are obtained from its Interface
Manual which is optionally available. Please refer to the manual.
1-4
Leveling screw C
90
Leveling
screw B
Leveling screw A
Leveling screw C
"
Point
Center mark
Leveling screw A
Leveling
screw B
2-1
Confirm the battery power remaining display. Confirm the battery power remaining display.
Replace with battery when battery level is low. Refer to Chapter 2.3 Battery Power
Remaining Display" .
Measurement is possible.
2-2
Zenith
Inclination of the standing
axis in the X direction
V
HR
b
TILT
12030'40"
$
$
To set auto tilt correction from the moment that power is on, refer to chapter 7" Vertical
Angle Tilt correction ( Tilt ON/OFF)".
The display of Vertical or Horizontal angle is unstable when instrument is on an unstable
stage or a windy day. You can turn off the auto tilt correction function of V angle in this
case.
2-3
3 ANGLE MEASUREMENT
ANGLE MEASUREMENT
3.1 Measuring Horizontal Angle Right and Vertical Angle
Make sure the mode is in Angle measurement.
Operating procedure
Operation
Display
Collimate A
[0 SET]
Twice
9030'40"
HR
2030'40"
9030'40"
HR
# Collimate the 2nd target (B).
Collimate B
The required V/H angle to target B will be
displayed.
V
HR
000'00"
9840'40"
12345'20"
Reference
How to Collimate (Reference)
! Point the telescope toward the light. Turn the diopter ring and adjust the diopter so that the cross
hairs are clearly observed.
(Turn the diopter ring toward you first and then backward to focus.)
" Aim the target at the peak of the triangle mark of the sighting collimator. Allow a certain space
between the sighting collimator and yourself for collimating.
# Focus the target with the focusing knob.
*
3-1
Function knob
Telescope eyepiece(Diopter ring)
3 ANGLE MEASUREMENT
Operation
Display
Collimate A
V
HR
angle
Right
9010'20"
12030'40"
[R/L]
(HR)
V
H L
9010'20"
23929'20"
# Measure as HL mode.
Operation
angle,
Display
9010'20"
13040'20"
[HOLD]
V
HR
Collimate
9010'20"
13040'20"
Blinks
[HOLD]
V
HR
9010'20"
13040'20"
*1) To return to the previous mode, press any key except the [HOLD] and the [FUNC+FUNC]
key.
3-2
3 ANGLE MEASUREMENT
Operation
Display
9010'20"
HR
" Press the [ANG%] key. *1)
12030'40"
[ANG/%]
%
V
-0.30
HR
12030'40"
*1) Every time pressing the [ANG/%] key, the display mode switches.
$ When the measurement is carried out over than45 (100%) from the horizontal, the
display shows [- - - - -] .
Operation
Display
[FUNC]
F
V
9010'20"
HR
" Press the [REP] key.
[REP]
12030'40"
REP
Ht
0
000'00"
REP
Ht
0
000'00"
H
# Collimate the target A. and press the [0SET] Collimate A
key twice.
[0SET]
[0SET]
H
% Collimate the target B using the horizontal Collimate B
clamp and tangent screw.
[HOLD]
Press the [HOLD] key.
REP
Ht
13025'20"
1 AVG
H
13025'20"
REP
Ht
26050'40"
2 AVG
H
13025'20"
2 measurements
3-3
3 ANGLE MEASUREMENT
( Repeat & to ' to measure the desired
number of repetitions.
REP
Ht
52141'20"
4 AVG
H
13025'20"
[FUNC]
[HOLD]
[Example] 4 measurement
3-4
4 DISTANCE/COORDINATE MEASUREMENT
DISTANCE/COORDINATE MEASUREMENT
4.1 Combined Use with EDM
Before measuring distance, set the atmospheric correction and the prism constant of
EDM.
For details, see the EDM Instruction manual.
PC-5S
4-1
4 DISTANCE/COORDINATE MEASUREMENT
DT-101
Operation
Collimate
" Set
the
instrument
for
distance
) pressing the
measurement mode (HR,
[FUNC] key and the [0SET] key to accept
slope distance data from EDM.
[FUNC]
[0SET]
Display
m
HR
12030'40"
*
123.456
HR
12030'40"
4-2
4 DISTANCE/COORDINATE MEASUREMENT
Operation
Collimate
" Set
the
instrument
for
distance
) pressing the
measurement mode (HR,
[FUNC] key and the [0SET] key to accept
slope distance data from EDM. *1)
[FUNC]
[0SET]
Display
m
HR
12030'40"
*
123.456
1
HR
12030'40"
123.456
3
HR
12030'40"
3 measurements
*1) In the number of distance measurement, the measured value will not display on the DT101 combined with other EDMs except DM-A5. Press [REC] key of EDM to display the data
EDM at that time.
4-3
4 DISTANCE/COORDINATE MEASUREMENT
Prism (n,e,z)
z
Instrument point
(0,0,0)
E
e
Operation
Display
angle
Display
V
HR
9010'20"
13040'20"
[HOLD]
V
HR
9010'20"
13040'20"
[FUNC]
[ANG%]
' EDM
switch
ON,
the
coordinate
EDM
measurement starts.
N coordinate ( ) and E coordinate ( ) is Power ON
displayed.
( Press the [FUNC] key and the [ANG%] key to
display Z coordinate ( )*1)
[FUNC]
[ANG%]
*
123.456
23.456
3.456
*
HR
9530'40"
*1) Pressing the [FUNC] key and the [ANG%] key, N ( ) , E ( ) and Z ( ) data is displayed
alternately.
$ Z coordinate measured by DT-101 is the difference in height between the horizontal
position of the instrument telescope and target.
4-4
4 DISTANCE/COORDINATE MEASUREMENT
10mm
Operating procedure
Operation
Collimate
" Set
the
instrument
for
distance
) pressing the
measurement mode (HR,
[FUNC] key and the [0SET] key to accept
slope distance data from EDM.*1)
[FUNC]
[0SET]
Display
m
HR
12030'40"
*
123.456
HR
12030'40"
*1) To measure the relative elevation ( ) and slope distance ( ), press the [FUNC] and the
[0SET] key.
*2) At the condition of more than 2,000 m of measuring distance, the displayed value by the
tracking mode is :
Displayed value = Measured distance -2,000m
In this case measurement must be done without atmospheric correction and correction
for refraction calculation and correction for refraction and earth curvature way.
4-5
Stadia hairs
L=100 l
5-1
6.3 Stake-out
The difference between the measured distance and the distance preset by data collector
operation is displayed.
Displayed value=Measured distance - Standard ( Preset) distance
! Stake-out operation can be performed for horizontal distance ( ), relative elevation
( ) or slope distance ( ).
Procedure
" Set the standard distance from Data collector.
# Press the [FUNC] and the [0SET] key of the DT-101.
$ Press the [FUNC] and [0SET] key, to select one of the distance measurement mode
( , , )
$ Set the EDM to the distance measurement mode in tracking.
The difference between the measured distance and the standard distance is
displayed.
! To return to normal distance measurement mode, reset the standard distance to 0"
or turn the power switch off once.
! Stake-out operation can be carried out in the continuous measurement mode
according to the similar operation procedure.
6-1
Instrument point C
Origin(0,0,0)
e
Procedure
" Input the coordinate of instrument point from Data collector.
# Collimate target A.
$ Set the direction angle from Data collector.
% Collimate target B using horizontal tangent screw and horizontal motion clamp.
& Set the instrument for coordinate measurement mode pressing the [FUNC] key and
the [ANG%] key to accept slope distance data from EDM.
' Set the EDM to the slope distance mode.
N coordinate ( ) and E coordinate ( ) from the set origin are measured repeatedly.
Pressing the [FUNC] key and [ANG%] key , Z coordinate ( ) key is measured
repeatedly.
! Pressing the [FUNC] key and the [ANG%] key, N ( ) , E ( ) and Z ( ) data is
displayed alternately.
! To return to the coordinate measurement with the origin at the instrument point, reset
the coordinate of instrument point at (0,0,0).
6-2
7 SELECTING MODE
SELECTING MODE
By operating the keys, the following modes are available.
Items
Contents
Setting value
0
Setting value
1
Angle unit
DEG/GON
DEG
GON
Angle unit
MIL
DEG/GON
MIL
V angle
Z0 / H0
Zenith 0
Horizontal 0
Minimum
angle unit
5"
1"
Tilt correction
ON/OFF
OFF
ON
OFF
ON
REC TYPE
REC-A
REC-B
CRLF
OFF
ON
Selecting Mode 2
Digit
No.
Items
Contents
Setting value
0
Setting value
1
Distance unit
meter
feet
Distance
display order
H.I. error
correction
OFF
ON
W-Correction
ON/OFF
OFF
ON
W-Correction
Coefficient
K=0.14
K=0.20
NEZ
MEMORY
OFF
ON
NEZ / ENZ
NEZ
ENZ
Unused
7-1
7 SELECTING MODE
Selecting Mode 1
Operation
Display
S E L E C T
0 0 1 1 1 0 0 0
Digit No.8
# Let the digit No.4 to be set blink by pressing
the [ ] key.
Digit No.1
[ ]
S E L E C T
0 0 1 1 1 0 0 0
Blinking
$ Press the [
]
S E L E C T
0 0 1 1 0 0 0 0
% Let the digit No.5 to be set blink by pressing
the [ ] key.
[ ]
S E L E C T
0 0 1 1 0 0 0 0
Blinking
& Press the [
]
S E L E C T
0 0 1 0 0 0 0 0
' Press the [0 SET] key.
[0 SET]
S E L E C T
S E T
( Turn the power OFF.
!
Power OFF
7-2
7 SELECTING MODE
Selecting Mode 2
Operation
Display
S E L E C T
0 0 0 0 0 0 1 0
Digit No.8
# Let the digit No.3 to be set blink by pressing
the [ ] key.
Digit No.1
[ ]
S E L E C T
0 0 0 0 0 0 1 0
Blinking
$ Press the [
]
S E L E C T
0 0 0 0 0 1 1 0
% Let the digit No.6 to be set blink by pressing
the [ ] key.
[ ]
S E L E C T
0 0 0 0 0 1 1 0
Blinking
& Press the [
]
S E L E C T
0 0 1 0 0 1 1 0
' Press the [0 SET] key.
[0 SET]
S E L E C T
S E T
( Turn the power OFF.
!
Power OFF
7-3
Replace of battery(DB-35)
Hole
Convex
Lid
Hook
Lid
" Push the hook downward and take the lid out.
# Take out the old batteries and put new batteries as illustration shows in direction of
.
$ Insert a convex in a upper hole. Click to close the lid by pressing it.
Replace all 4 batteries to new ones at the same
time.
Don't mix the old batteries to the new ones.
For installing
Place the base of the on-board battery into the main body, push the on-board battery
toward the instrument side till the battery clicks into position.
8-1
9.
Adjustment screw
Adjustment screw
Leveling screw
Holding ring
9-1
Set screw
Check
" Place the plate level parallel to a line
running through the centers of two
leveling screws, say, A and B. Use
these two leveling screws only and
place the bubble in the center of the
plate level.
# Rotate the instrument 180 or 200g
around the vertical axis and check
bubble movement of the plate level. If
the bubble has been displaced, then
proceed with the following adjustment.
180
Plate level
Leveling
screw A
Adjustment
" Adjust the level adjustment capstan
Level adjustment
screw, with the accessory adjusting capstan screw
pin and return the bubble towards the
center of the plate level. Correct only
one-half of the displacement by this
method.
# Correct the remaining amount of the
bubble displacement with the leveling
screws.
$ Rotate the instrument 180 or 200g
around the vertical axis once more
and check bubble movement. If the
bubble is still displaced, then repeat
the adjustment.
Leveling
screw B
Half amount
displacement
Check
" Carefully level the instrument with the plate level only. If the bubble of the circular
level is centered properly, adjustment is not required. Otherwise, proceed with the
following adjustment.
Adjustment
" Shift the bubble to the center of the circular level, by adjusting three capstan
adjustment screws on the bottom surface of the circular level, with the accessory
adjusting pin.
Bottom of tribrach
9-2
Check
" Set the instrument up the tripod and carefully level it.
# Sight the cross-hairs on a well defined Point A at a distance of, at least, 50 meters (
160ft.) .
$ Next swing the telescope vertically using the vertical jog and shuttle, and check
whether the point travels along the length of the vertical cross-hair.
% If the point appears to move continuously on the hair, the vertical cross-hair lies in a
plane perpendicular to the horizontal axis ( and adjustment is not required ).
& However, if the point appears to be displaced from the vertical cross-hair, as the
telescope is swung vertically, adjustment is required in the reticle plate.
Adjustment
" Unscrew the cross-hair adjustment section cover, by revolving it in the
counterclockwise direction, and take it off. This will expose four eyepiece section
attachment screws.
# Loosen all four attachment screws slightly with the accessory screw-driver (while
taking note of the number of revolutions).
Then revolve the eyepiece section so that the vertical cross-hair coincides to Point A'.
Finally, re-tighten the four screws by the amount that they were loosened.
$ Check once more and if the point travels the entire length of the vertical cross-hair,
further adjustment is not required.
Eyepiece section
attachment screws
Eyepiece section
attachment screws
Eyepiece
9-3
50m
9-4
50m
Adjustment
" Unscrew the cross-hair adjustment section cover.
# Find Point D at a point between Points C and B,
which should be equal to 1/4th the distance
between Points B and C and measured from Point
C. This is because the apparent error between
Points B and C is four times the actual error since
the telescope has been reversed twice during the
checking operation.
Capstan
adjustment
screws
50m
Eyepiece
Capstan
adjustment
screws
Note1) : First, loosen the capstan adjustment screw on the side to which the vertical crosshair line must be moved. Then tighten the adjustment screw on the opposite side by
an equal amount which will leave the tension of the adjustment screws unchanged.
Revolve in the counterclockwise direction to loosen and in the clockwise direction to
tighten, but revolve as little as possible.
Note2) : Perform following adjustment after complete above adjustment . Chapter 9.6
"Adjustment of Vertical Angle 0 Datum".
Check
" Coincide the center mark and the point.
# Rotate the instrument 180 or 200g around the vertical axis and check the center
mark.
If the point is properly centered in the center mark, adjustment is not required.
Otherwise, adjust in the following manner.
Adjustment
" Take off the adjustment section robber cap of the optical plummet telescope
eyepiece, by revolving it in the counterclockwise direction, and take it off. This will
expose four capstan adjustment screws which should be adjusted with the accessory
adjusting pin to shift the center mark to the point. However, correct only one-half of
the displacement in this manner.
Capstan
adjustment
screws
Eyepiece
1/2 of displacement
# Use the leveling screws and coincide the point and center mark.
$ Rotate the instrument 180 or 200g around the vertical axis once more and check the
center mark. If it is coincided to the point, then further adjustment is not required.
Otherwise, repeat the adjustment.
Note : First, loosen the capstan adjustment screw on the side to which the center mark must
be moved. Then tighten the adjustment screw on the opposite side by an equal amount
which will leave the tension of the adjustment screws unchanged.
Revolve in the counterclockwise direction to loosen and in the clockwise direction to
tighten, but revolve as little as possible.
9-5
Operation
Display
[0SET]
+
Power ON
Turn
Telescope
0 SET
SETP 1
TILT
[0SET]
V
SETP 2
TILT
[0SET]
Power
OFF
SET
If there is any misoperation during checking, display for error appears. Then repeat the
above procedure from the start.
Check that the instrument works properly by sighting a target at normal and reverse
telescope settings and check if the amount of normal and reverse setting is 360.
9-6
10.1
Distance Calculation Formula; with correction for refraction and earth curvature taken into
account. Follow the Formula below for converting horizontal and vertical distances.
Horizontal distance D=AC() or BE()
Vertical distance Z =BC() or EA()
D=L{cos-(2 -)sin}
Z=L{sin-( -)cos}
=Lcos/2R
Earth curvature
correcting item
=KLcos/2R
Atmospheric
refraction
correcting item
K=0.14 or 0.2
Coefficient
of
refraction
R=6372km
Radius of earth
( or )
Altitude angle
L
Slope distance
E
D
-Z
D
C
2
Radius of earth
R=6372Km
The conversion formula for horizontal and vertical distances is as follows when
correction for refraction and earth curvature is not applied.
D=Lcos
Z=Lsin
Note : The coefficient of the instrument has been set at 0.14 before shipment (K=0.14).
if the K" value is to be changed, refer to "13. Selecting mode".
10-1
Prism center
EDM
H.I.=166mm
S
DT-101
S' :
S :
:
S 2 - ( 0.166 sin )
11-1
12 DETACH/ATTACH OF TRIBRACH
12 DETACH/ATTACH OF TRIBRACH
The instrument is easily detached or attached to the tribrach, with a tribrach fixing lever
loosened or tightened for this purpose.
!
Detachment
" Loosen the tribrach fixing lever, by revolving it 180 or 200g in the
counterclockwise direction (which will point the triangle mark upwards).
# Grip the carrying handle firmly with one hand while holding the tribrach with the
other. Then lift the instrument straight upwards and off.
Attachment
" Hold the instrument by the carrying handle, with one hand, and carefully lower it
on top of the tribrach while, at the same time, coinciding the alignment piece with
the tribrach alignment groove on the instrument and tribrach respectively.
# When fully seated, revolve the tribrach fixing lever 180 or 200g clockwise ( which
will point the triangle mark downwards again).
Alignment piece
Alignment groove
Securing screw
12-1
13 PRECAUTIONS
13 PRECAUTIONS
1.
For transportation, hold by the handle or yoke of the instrument. Never hold by the
lens barrel as it can affect the fixing bracket inside and reduce the accuracy of the
instrument.
2.
Never expose the instrument without a filter to direct sunlight. It may damage the
components inside the instrument.
3.
4.
5.
6.
When opening the carrying case and taking out the instrument, place the case
horizontally, then open the case.
7.
When storing the instrument in the case, always align the white indication marks on
the instrument, and set the telescope at the vertical position.
8.
9.
For cleaning the instrument after use, remove dust using a cleaning brush, then
wipe off with a cloth.
10.
For cleaning the lens surface, use a cleaning brush to remove the dust, then use a
clean lintless cotton cloth. Moisten it with alcohol ( or mixture with ether ) to wipe
gently in a rotational motion from the center out.
11.
12.
To remove the dust on the case, never use thinner or benzine. Use a clean cloth
moistened with neutral detergent.
13.
Check each part of the tripod after extended use. Parts (screws or clamps) may
work themselves free.
13-1
14 OPTIONAL ACCESSORIES
14 OPTIONAL ACCESSORIES
14-1
14 OPTIONAL ACCESSORIES
Data collector
Suitable for systemization of measuring instrument. Measuring data will be automatically stored and
transferred to a computer system, making measuring operations more efficient and saving time and
effort in such operation.
14-2
14 OPTIONAL ACCESSORIES
Trough Compass-5
The trough compass is simply mounted on top of the carrying handle
Diagonal Eyepiece
Observation in an easy posture will be provided up to the zenith position
14-3
14 OPTIONAL ACCESSORIES
Prism sets
See the description on Chapter 15 "TARGET (PRISM) SYSTEM".
Eyepiece prism
The eyepiece prism is supplied with a single dark filter glass for use on the telescope and is very
convenient when tilting the telescope up to 40 from the zenith. The images are observed upside
down, but not reversed laterally.
symbol 129 \f "Wingdings" \s 10"# Simply screw the eyepiece prism directly into the female
screw mounts on the eyepiece of the telescope.
symbol 130 \f "Wingdings" \s 10$# Use the dark filter glass ( with spring clips ) on the eyepiece
prism fixed to the telescope eyepiece, when observing the sun directly, otherwise take it off for
normal use.
14-4
14 OPTIONAL ACCESSORIES
Flat head 5/8" symbol 180 \f "Symbol" \s 10 11 threads with adjustable legs.
Solar filter
A filter designed exclusively for direct collimation of the sun.
14-5
Target pole-2
Prism-2
Prism-2
Tilting prism
holder-2
Triple prism
holder-2
Adjustable target-2
Prism
adaptor-2
Pole adaptor-F2
Tribrach
15-1
Prism
adaptor-S2
Optical plummet
tribrach
16 ERROR DISPLAYS
16 ERROR DISPLAYS
Display
b
Contents
Countermeasure
E01
key,
returns
to
E02
Displayed when the telescope rotated Press [0SET] key, and after display
abnormally.
of "0 SET", set vertical angle zero set
by rotating telescope.
E03
E04
Displayed when the discrepancy value Press [0SET] key, measure again from
of each measuring is more than 30" the beginning.
while repetition angle measurement is
operated.
E51
Displayed when measurement carried Set correction for refraction and earth
out within 9 from zenith or nadir at curvature mode OFF or measure out
the Earth curvature and refraction of 9 from the zenith or Nadir.
correction mode is ON.
E70
E80s
Mainly at the time data transmission Confirm the connection cables and the
between
DT-101
and
external operation procedure are correct.
instrument.
E99
If errors still persist after attempting to clear them, contact your local TOPCON dealer or
TOPCON Head office.
16-1
17 SPECIFICATIONS
17 SPECIFICATIONS
Length
149mm
Objective lens
45mm
Magnification
30
Image
Erect
Field of view
130'
Resolving power
2.5"
Minimum focus
90cm
Stadia ratio
100
Stadia constant
Electronic Angle
Method
Incremental reading
Measurement
Detecting
Horizontal: 2side
Telescope
Vertical: 2side
Minimum reading
1"/5"(0.2mgon/1mgon)
Accuracy *1)
2"
Diameter circle
71mm
Display
Unit
2side
Illumination
Display
Reticle
Tilt sensor
Automatic vertical
compensator
Compensating range
3'
Optical Plummet
Magnification
Telescope
Field of view
Focusing
0.5m~
Plate level
30" / 2mm
Circular level
10' / 2mm
Power Supply
Battery
4 AA batteries
Operating Time
12 hours
Size
Instrument height
176mm (6.93in)
Dimension
DWH(mm)
149180313
(5.877.0912.3 in)
Weight
(Including batteries)
Compensator
Level Sensitivity
17-1
TOPCON CORPORATION
75-1 Hasunuma-cho, Itabashi-ku, Tokyo, 174 Japan.
Phone: 3-3967-1101 Fax: 3-3960-4214 Telex:0727-2384
TOPCON CALIFORNIA
3380 Industrial BLVD, Suite 105, West Sacramento, CA. 95691, U.S.A.
Phone: 916-374-8575 Fax: 916-374-8329
TOPCON MIDWEST
1728 West Algonquin Road, Arlington Heights, Illinois 60005, U.S.A.
Phone: 708-818-9188 Fax: 708-818-9342
TOPCON DENVER
4895 Joliet Street ,Suite E Denver Co 80239 U.S.A
Phone: 303-373-0990 Fax: 303-373-0998
TOPCON S.A.R.L.
104/106, Rue Rivay, 92300 Levallois-Perret, FRANCE
Phone: 1-4106 9494 (MEDICAL) 1-4106 9490 (TOPOGRAPHIE)
Fax: 1-47390251 Telex: 620287
MADRID OFFICE
Avenida Ciudad de Barcelona 81,1 Planta 28007, Madrid, SPAIN
Phone: 1-552-4160 Fax: 1-552-4161