CALIBRATION REPORT

OEG INDIA PVT. LTD.

SITE: OPML- AMLAI


Equipment: 25MW Alternator CW I/L temp. Gauge

Calibrated By: A.G

Make: Thermostatic

Verified By: D.A

Range: 0 to 100 deg C

Calibration date: 11.04.2013

Serial No:

Next Calibration due date: 11.10.2013

Tag No: 40-TG-543

Scale Division value: 1 deg C

Master Calibrator Details


Description

Make

Temperature Calibrator

Feed (Unit)
Deg C
Deg C
Deg C
Deg C
Deg C
Deg C
Deg C

Nagman Instrument

Master Calibrator reading


Rising
Falling
40
100
50
90
60
80
70
70
80
60
90
50
100
40

Model

Calibration validity

650H

18.01.2014

Test Instrument reading


Rising
Falling
41
100
50
91
62
80
71
69
80
60
90
51
100
40

Error
1/0
0/1
2/0
1/1
0/0
0/1
0/0

Observation Report
Sl No

1
2
3
4
5
6

Details
Body
Dial
Needle
Condition
Terminals
Parts missing

OK
Ok
Ok
Ok
Ok
NA
NA

Defect

For Control & Instrument

CALIBRATION REPORT
OEG INDIA PVT. LTD.

SITE: OPML- AMLAI


Equipment: 30 MW oil cooler inlet oil pr. gaugeCalibrated By: DD
Make: General

Verified By: MRD

Range: 0 to 20 kg/cm2

Calibration date: 12.04.2013

Serial No:

Next Calibration due date: 12.10.2013

Tag No: 51-PG-504

Scale Division value: 0.5 kg/cm2

Master Calibrator Details


Description

Make

Pressure Calibrator

Feed (Unit)
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2

Nagman Instrument

Master Calibrator reading


Rising
Falling
2.5
20.0
5.0
17.5
7.5
15.0
10.0
12.5
12.5
10.0
15.0
7.5
17.5
5.0
20.0
2.5

Model

Calibration validity

MPC-E-P-35

12.01.2014

Test Instrument reading


Rising
Falling
2.5
20.0
5.0
17.5
7.5
15.0
10.0
12.5
12.5
10.0
15.0
7.5
17.5
5.0
20.0
2.5

Error
0/0
0/0.1
0/0
0/0.1
0/0
0/0
0/0
0/0

Observation Report
Sl No

1
2
3
4
5
6

Details
Body
Dial
Needle
Condition
Terminals
Parts missing

OK
Ok
Ok
Ok
Ok
NA
NA

Defect

For Control & Instrument

CALIBRATION REPORT
OEG INDIA PVT. LTD.

SITE: OPML- AMLAI


Equipment: Air Dryer inlet pressure gauge

Calibrated By: DD

Make: Micro

Verified By: DA

Range: 0 to 14 kg/cm2

Calibration date: 12.04.2013

Serial No:

Next Calibration due date: 12.10.2013

Tag No:

Scale Division value: 0.2 kg/cm2

Master Calibrator Details


Description

Make

Pressure Calibrator

Feed (Unit)
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2

Nagman Instrument

Master Calibrator reading


Rising
Falling
2.0
14.0
4.0
12.0
6.0
10.0
8.0
8.0
10.0
6.0
12.0
4.0
14.0
2.0

Model

Calibration validity

MPC-E-P-35

12.01.2014

Test Instrument reading


Rising
Falling
2.0
14.0
4.2
12.0
6.0
10.2
8.0
8.0
10.2
6.0
12.0
3.8
14.0
2.0

Error
0/0
0.2/0
0/0.2
0/0
0.2/0
0/0.2
0/0

Observation Report
Sl No

1
2
3
4
5
6

Details
Body
Dial
Needle
Condition
Terminals
Parts missing

OK
Ok
Ok
Ok
Ok
NA
NA

Defect

For Control & Instrument

CALIBRATION REPORT
OEG INDIA PVT. LTD.

SITE: OPML- AMLAI


Equipment: 30mw Condenser CW I/L pr. gauge

Calibrated By: DD

Make: Thermostatic

Verified By: DA

Range: 0 to 10kg/cm2

Calibration date: 12.04.2013

Serial No:

Next Calibration due date: 12.10.2013

Tag No: 40-PG-510A

Scale Division value: 0.1 kg/cm2

Master Calibrator Details


Description

Make

Pressure Calibrator

Feed (Unit)
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2

Nagman Instrument

Master Calibrator reading


Rising
Falling
1.0
10.0
2.0
9.0
3.0
8.0
4.0
7.0
5.0
6.0
6.0
5.0
7.0
4.0
8.0
3.0
9.0
2.0
10.0
1.0

Model

Calibration validity

MPC-E-P-35

12.01.2014

Test Instrument reading


Rising
Falling
1.0
10.0
2.1
9.0
3.0
8.0
4.1
6.9
5.1
6.0
6.0
5.0
7.0
4.0
8.1
2.9
9.0
2.0
10.0
1.0

Error
0/0
0.1/0
0/0
0.1/0.1
0.1/0
0/0
0/0
0.1/0.1
0/0
0/0

Observation Report
Sl No

1
2
3
4
5
6

Details
Body
Dial
Needle
Condition
Terminals
Parts missing

OK
Ok
Ok
Ok
Ok
NA
NA

Defect

For Control & Instrument

CALIBRATION REPORT
OEG INDIA PVT. LTD.

SITE: OPML- AMLAI


Equipment: 30mw Control oil pressure switch low

Calibrated By: AG

Make: Switzer

Verified By: MRD

Range: 1.5 to 16 kg/cm2

Calibration date: 09.04.2013

Serial No:

Next Calibration due date: 09.10.2013

Tag No: 51-PSL-401

Scale Division value: NA

Master Calibrator Details


Description

Make

Pressure Calibrator

Feed (Unit)
Kg/cm2

Nagman Instrument

Master Calibrator reading


Rising
Falling
9.3
8.5

Model

Calibration validity

MPC-E-P-35

12.01.2014

Test Instrument reading


Rising
Falling
Reset
Set

Error

Observation Report
Sl No

1
2
3
4
5
6

Details
Body
Dial
Needle
Condition
Terminals
Parts missing

OK
Ok
Ok
Ok
Ok
Ok
NA

Defect

For Control & Instrument

CALIBRATION REPORT
OEG INDIA PVT. LTD.

SITE: OPML- AMLAI


Equipment: LP dosing pump-1 discharge. pr gauge

Calibrated By: DD

Make: General

Verified By: MRD

Range: 0 to 25 kg/cm2

Calibration date: 13.04.2013

Serial No: DAP-03899 LFBSFG-V-U/1187472

Next Calibration due date: 13.10.2013

Tag No: 41-PG-011

Scale Division value: 0.5 kg/cm2

Master Calibrator Details


Description

Make

Pressure Calibrator

Feed (Unit)
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2

Nagman Instrument

Master Calibrator reading


Rising
Falling
0.0
25.0
5.0
20.0
10.0
15.0
15.0
10.0
20.0
5.0
25.0
0.0

Model

Calibration validity

MPC-E-P-35

12.01.2014

Test Instrument reading


Rising
Falling
0.0
25.0
5.5
20.0
10.0
14.5
14.5
9.8
20.2
5.0
25.0
0.0

Error
0.0/0.0
0.5/0.0
0.0/0.5
0.5/0.2
0.2/0.0
0.0/0.0

Observation Report
Sl No

1
2
3
4
5
6

Details
Body
Dial
Needle
Condition
Terminals
Parts missing

OK
Ok
Ok
Ok
Ok
NA
NA

Defect

For Control & Instrument

CALIBRATION REPORT
OEG INDIA PVT. LTD.

SITE: OPML- AMLAI


Equipment: 25MW Bleed to LPH steam pr. gauge

Calibrated By: AG

Make: Siemens

Verified By: DA

Range: -1 to 1.5 kg/cm2

Calibration date: 08.04.2013

Serial No:

Next Calibration due date: 08.10.2013

Tag No: 52-PG-105

Scale Division value: 0.05

Master Calibrator Details


Description

Make

Pressure Calibrator

Feed (Unit)
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2

Nagman Instrument

Master Calibrator reading


Rising
Falling
-1.0
1.5
-0.5
1.0
0.0
0.5
0.5
0.0
1.0
-0.5
1.5
-1.0

Model

Calibration validity

MPC-E-P-35

12.01.2014

Test Instrument reading


Rising
Falling
-1.0
1.5
-0.5
1.05
0.05
0.5
0.55
0.0
1.0
-0.55
1.5
-1.0

Error
0.0/0.0
0.0/0.05
0.05/0.0
0.05/0.0
0.0/0.05
0.0/0.05

Observation Report
Sl No

1
2
3
4
5
6

Details
Body
Dial
Needle
Condition
Terminals
Parts missing

OK
Ok
Ok
Ok
Ok
NA
NA

Defect

For Control & Instrument

CALIBRATION REPORT
OEG INDIA PVT. LTD.

SITE: OPML- AMLAI


Equipment: 25mw turbine rear bearing inlet
lube oil pr. gauge

Calibrated By: DD

Make: Siemens

Verified By: MRD

Range: 0 to 4 kg/cm2

Calibration date: 15.04.2013

Serial No:

Next Calibration due date: 15.10.2013

Tag No: 52-PG-508

Scale Division value: 0.1 kg/cm2

Master Calibrator Details


Description

Make

Pressure Calibrator

Feed (Unit)
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2
Kg/cm2

Nagman Instrument

Master Calibrator reading


Rising
Falling
0.0
4.0
1.0
3.0
2.0
2.0
3.0
1.0
4.0
0.0

Model

Calibration validity

MPC-E-P-35

12.01.2014

Test Instrument reading


Rising
Falling
0.0
4.0
1.0
3.0
2.1
2.0
3.0
1.0
3.9
0.0

Error
0.0/0.0
0.0/0.0
0.1/0.0
0.0/0.0
0.1/0.0

Observation Report
Sl No

1
2
3
4
5
6

Details
Body
Dial
Needle
Condition
Terminals
Parts missing

OK
Ok
Ok
Ok
Ok
NA
NA

Defect

For Control & Instrument

CALIBRATION REPORT
OEG INDIA PVT. LTD.

SITE: OPML- AMLAI

Equipment: Belt weigher

Calibrated By: AKG

Make: Schenck

Verified By: MRD

Range: 0 to 75 Tph

Calibration date: 12.04.2013

Serial No: 6111-1442-2-2-001

Next Calibration due date: 12.05.2013

Tag No: 42-belt-weigher

Scale Division value: N/A

Observation Report

Given weight at
weighing zone
(Kg)

Display at Intecont

Error

Q (Kg/Mtr)

I (tph)

0.0

0.0

0.0

0.0

20

8.69

14.2

0.0

NB: weighing zone length = 2.3 mtr

For Control & Instrument