WORK INSTRUCTION FOR

MICROMETER CALIBRATION
S.NO.

DESCRIPTION

CLEAN FACES AND ALL EXTERIOR SURFACES WITH CLOTH,


DAMPENED WITH CLEANING SOLUTION.

CHECK MEASURING SCREW FOR WEAR BY PUSHING THIMBLE TO


AND FOR IN DIRECTION OF MEASURING SCREW AXIS.

PRESS ORIGIN BUTTON FOR COUNTER ZERO.

PLACE BLOCK OR BLOCK COMBINATIONS BETWEEN ANVIL AND


SPINDLE OF THE INSTRUMENT UNDER TESTING.

CLOSE THE INSTRUMENT UNDER TESTING USING RATCHET OR


FRICTION STOP.

USE BLOCK COMBINATIONS TO CHECK THE ACCURACY OF THE


INSTRUMENT IN POSITIONS SO AS TO COVER TOTAL THIMBLE
RANGE.

NOTE THE READINGS OBSERVED AT EVERY POSITION OF


INSTRUMENT UNDER TESTING.

AFTER THAT CALCULATE THE ERROR (DIFFRENCE OF MASTER


INSTRUMENT READING & UNDER CALLIBRATION INSTRUMENT
READING ).

REQUIRED ACCURACY 3%.

10

AFTER THAT CALCULATE THE ACCURACY.

11

AFTER THAT FILL CALIBRATION TAG OF THE INSTRUMENT


UNDER OBSERVATION.

PREPARED BY:

Doc. No. 10-WI-29


Rev. No-00
Deptt.-Std. Room (Quality)

RESPONSIBILITY

RECORDS

STD. ROOM
ENGINEER

APPROVED BY: