Professional Documents
Culture Documents
Child Dental Health, School of Oral and Dental Science, University of Bristol, Bristol UK
Medical Physics Department, RUH NHS Trust, Bath, UK
c Queen Alexandra Hospital, Cosham, Portsmouth, UK
d Consultant orthodontist, Kings Mill Hospital, Manseld, United Kingdom
b
a r t i c l e
i n f o
a b s t r a c t
Article history:
Objectives. Light cured materials are increasingly used in orthodontic clinical practice and
concurrent with developments in materials have been developments in light curing unit
technology. In recent years the irradiances of these units have increased. The aim of this
28 March 2013
study was to determine the safe exposure times to both direct and reected light.
Methods. The weighted irradiance and safe exposure times of 11 dental curing lights (1
plasma arc, 2 halogen and 8 LED lights) were determined at 6 distances (260 cm) from the
light guide tip using a spectroradiometer. In addition, using the single most powerful light,
Keywords:
the same two parameters were determined for reected light. This was done at a distance
Light curing
of 10 cm from the reected light, but during simulated bonding of 8 different orthodontic
Weighted irradiance
brackets of three material types, namely stainless steel, ceramic and composite.
Results. The results indicate that the LED Fusion lamp had the highest weighted irradiance and the shortest safe exposure time. With this light the maximum safe exposure time
without additional eye protection for the patient (at 10 cm), the operator (at 30 cm) and the
assistant (at 60 cm) ranged from 2.5 min, 22.1 min and 88.8 min respectively. This indicates a
relatively low short term risk during normal operation of dental curing lights. For reected
light at a distance of 10 cm the risk was even lower, but was affected by the material and
shape of the orthodontic bracket under test.
Signicance. The short term risks associated with the use of dental curing lights, halogen,
LED or plasma, appear to be low, particularly if as is the case adequate safety precautions
are employed. The same is true for reected light from orthodontic brackets during bonding.
What is still unclear is the potential long term ocular effects of prolonged exposure to the
blue light generated from dental curing lights.
2013 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
1.
Introduction
The use of light curing as a means of initiating the polymerization of orthodontic adhesives has gained in popularity
Corresponding author at: School of Oral and Dental Science, University of Bristol, Bristol, UK. Tel.: +44 0117 342 4355.
E-mail address: tony.ireland@bristol.ac.uk (A.J. Ireland).
0109-5641/$ see front matter 2013 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
http://dx.doi.org/10.1016/j.dental.2013.03.023
e86
d e n t a l m a t e r i a l s 2 9 ( 2 0 1 3 ) e85e90
2.
e87
d e n t a l m a t e r i a l s 2 9 ( 2 0 1 3 ) e85e90
Table 1 Light units under test, type, wavelength and irradiances derived from the manufacturers information.
Light unit
Apollo 95 E
Cromalux
CU80
DEMIOrtho
Elipar Freelight 2
Fusion
Mini LED
Mini LED2
Ortholux
Smartlite
Starlight
Type
Wavelength (nm)
Irradiance mW/cm2
Plasma arc
Halogen
Halogen
LED
LED
LED
LED
LED
LED
LED
LED
460490
400500
380510
420465
430480
385430
420480
420480
430480
450490
440480
1600
650800
600
11001330
1000
1500
1250
2000
1600
950
100
and 800 nm in order to identify the output spectrum. Subsequence measurements were made only at the relevant
wavelengths. The combined tolerance of the measurement
was typically less than 5%, including temporal variation due
to continuous battery discharge condition. Care was taken to
ensure each of the lights was tested with a fully charged battery and a new bulb was placed in each of the quartz tungsten
halogen lights prior to testing. The ambient light was minimized during testing by closing the blinds and turning off
indoor uorescent lighting. In the second part of this study,
determining the effect of different brackets and bracket materials (metal and esthetic see Table 2) on reected light and
therefore the maximum daily exposure limits, a phantom
head unit with acrylic teeth was used, as it was felt these were
less variable than using extracted human incisor teeth. All
the measurements for reected light were performed with the
light curing unit that had been found to produce the greatest
spectral output in the rst part of the experiment, namely the
Fusion LED curing light. The spectroradiometer tip was placed
at a distance of 10 cm from the brackets and was held in place
using the custom made jig (Fig. 1). The brackets were placed
on upper canine to canine with Transbond XT (3 M) adhesive
and the light tip held at a distance of 0.5 cm from the central
incisor teeth.
3.
Results
Bracket type
Trade name
Manufacturer
Ceramic
Ceramic
Stainless steel
Stainless steel
Ceramic
Composite
Stainless steel
Stainless steel
Clarity
Clarity SL
Victory
Microarch
Encore
Tiger
Mini Ovation
TOC bracket
3M Unitek
3M Unitek
3M Unitek
GAC International
Ortho Technology
TOC
GAC International
TOC
e88
d e n t a l m a t e r i a l s 2 9 ( 2 0 1 3 ) e85e90
Table 3 Weighted intensity mW/cm2 at each distance (cm) for each of the 11 lights under test.
Weighted intensity mW/cm2 at each distance (cm)
Light unit
2 cm
Apollo95E
Cromalux
CU80
DEMIOrtho
Eipar freelight2
Fusion
MiniLED
MiniLED2
Ortholux LED
SmartLite
Starlight
9.90
26.01
27.43
181.02
42.54
266.68
62.01
36.33
55.77
22.36
44.26
10 cm
20 cm
30 cm
40 cm
50 cm
1.47
1.25
4.26
2.03
6.57
2.43
1.59
2.74
1.09
2.01
0.35
0.29
1.07
0.47
1.68
0.65
0.40
0.67
0.29
0.52
0.15
0.16
0.46
0.20
0.75
0.28
0.17
0.27
0.13
0.22
0.10
0.12
0.27
0.12
0.42
0.15
0.09
0.16
0.08
0.13
0.06
0.09
0.17
0.07
0.25
0.10
0.06
0.10
0.05
0.08
60 cm
0.04
0.05
0.12
0.05
0.19
0.07
0.05
0.07
0.04
0.06
for the Fusion light, to 11.77 min for the Apollo 95E plasma arc
lamp. The maximum permissible exposure time is the theoretical time that it would take for a lesion to form on the retina.
If the distance from the light tip to the patients eye is approximately 10 cm, to the operator approximately 30 cm and to the
assistant 60 cm, then direct exposure at these distances would
cause damage to the retina in approximately 2.5 min, 22.1 min
and 88.8 min respectively. This is assuming that no protection
was being used and that the direct exposure was continuous.
The second part of this study looked at the effect of
orthodontic brackets and therefore reected light on the
weighted intensities and the safe exposure times using the
Fusion LED light at 10 cm. The initial reading was a baseline reading of the reected light from the acrylic teeth, with
no orthodontic brackets. The subsequent readings measured
the reected light with orthodontic brackets placed on upper
canine to canine, again measured at 10 cm. The results are
illustrated in Table 4. Looking at both the measured weighted
intensities and safe exposure times of reected light using
the Fusion LED with the various brackets, the stainless steel
Mini Ovation bracket gave the highest value with the Tiger
bracket second. This perhaps suggests that both bracket shape
and material are important for the reected light. However,
in all cases the measured weighted irradiances were low and
so the maximum safe daily exposures were relatively high.
As the values were so low for reected light at 10 cm it was
felt that any results measured at distances of greater than
10 cm would be too low to be of any clinical signicance and
Table 4 The calculated safe exposure times (min) using the ICNIRP guidelines at each of the distances under test and for
each of the light curing units (calculated using the formula LB t 100).
Light unit
Apollo95E
Cromalux
CU80
DEMIOrtho
Eipar freelight2
Fusion
MiniLED
MiniLED2
Ortholux LED
SmartLite
Starlight
11.77
4.49
4.25
0.64
2.74
0.19
1.88
3.21
2.09
5.22
2.64
10 cm
20 cm
30 cm
40 cm
50 cm
60 cm
11.3
13.4
3.9
8.2
2.5
6.9
10.5
6.1
15.3
8.3
47.2
57.1
15.6
35.5
9.9
25.8
42.2
25.1
57.9
32.2
113.5
104.8
36.3
81.8
22.1
60.3
100.9
61.2
123.5
76.1
175.3
144.8
62.4
144.6
40.1
114.2
176.0
105.7
211.3
126.7
301.3
194.6
97.6
229.6
66.6
173.0
265.2
166.4
326.7
202.0
432.9
312.7
138.5
322.3
88.8
248.2
365.6
226.1
450.9
301.7
e89
d e n t a l m a t e r i a l s 2 9 ( 2 0 1 3 ) e85e90
Cromalux
CU80
DEMIOrtho
Eipar_freelight2
Fusion
MiniLED
MiniLED2
Ortholux_LED
SmartLite
Starlight
1000000.00
10000000.00
100000.00
10000.00
1000.00
100.00
10
20
30
40
50
60
70
Distance [cm]
Fig. 2 Weighted intensities (mW/cm2 ) of the different curing lights at 2, 10, 20, 30, 40, 50 and 60 cm.
4.
Discussion
The rst part of this study looked at the maximal permissible daily exposure times for light curing units at a variety
of distances from the light source ranging from 2 to 60 cm,
with 2 cm being the minimum permissible distance at which
the spectroradiometer would work. It can be seen that a
direct exposure at 2 cm with the most intense light, namely
the Fusion LED, would produce a retinal lesion in just 11.4 s
(0.19 min) based on the ICNIRP guidelines [21]. This may be
related to the collimation of the light beam, which has been
shown to be greater in the case of the Fusion LED than some
other curing lights [22]. An earlier pilot study had measured
the light tip to eye distance when bonding an upper incisor
bracket and for the patient was measured as approximately
10 cm; for the orthodontist approximately 30 cm and to the
orthodontic assistant approximately 60 cm and is similar to
the distances used in previous studies [23]. Using these same
distances it can be estimated from our results that the time
taken to form a lesion on the retina with a direct exposure
would be 2.5 min for the patient, 22.1 min for the orthodontist
and 88.8 min for the assistant, depending on which guidelines
are used. Whichever guidelines are used, all of the gures
assume a direct exposure with no safety procedures in place.
If safety precautions are used, such as orange lters close
to the light guide tip, an orange paddle shield or orange spectacles, then the light produced by each of light curing units
under test is unlikely to reach the maximum daily exposures
e90
d e n t a l m a t e r i a l s 2 9 ( 2 0 1 3 ) e85e90
degeneration, in 838 bay workers. They found that in agematched controls, patients with macular degeneration had
signicantly higher exposure to blue or visible light, but were
not different with respect to exposure to UVA or UVB. It was
concluded that high levels of exposure to blue or visible light
may cause ocular damage, especially later in life, and may be
related to the development of age-related macular degeneration. It should also be remembered that people who have had
cataract surgery are at greater ocular risk to blue light exposure and this may be later in life, when the risks of exposure
to blue light are already greater [27].
5.
Conclusions
The short term risks associated with the use of dental curing
lights, halogen, LED or plasma, appear to be low, particularly
if, as is the case, adequate safety precautions are employed.
The same is true for reected light from orthodontic brackets
during bonding. What is still unclear are the potential long
term ocular effects of prolonged exposure to the blue light
generated from dental curing lights.
Acknowledgements
Based on a thesis submitted to the School or Oral and Dental Science, University of Bristol, in partial fulllment of the
requirements for the DDS degree
references