Professional Documents
Culture Documents
nova
ANALYZE
CURSOR
ANALYZE
nova
CURSOR
F, G, H, I, J, K Reference Manual
Preface
1-800-458-5813
FAX:
Technical Assistance
For technical assistance, call Nova Biomedical Technical Services at:
Telephone:
1-800-545-NOVA or
(617) 894-0800
FAX:
(617) 894-0585
Copyright
Printed in the U.S.A. Copyright 1997, Nova Biomedical, Waltham, MA 02254.
Preface
ii
NCCLS Cross-Reference
NCCLS Guideline
1.
Principle of Test
Appendix B
2.
Specimen required
3.
Appendix C
4.
5.
Step-by-Step Directions
Analysis
Chapter 3
Maintenance
Chapter 4
Troubleshooting
Chapter 5
6.
Calculations
7.
8.
Expected values
Appendix D
9.
Linearity limits
Appendix A
10.
Limitations (interferences)
11.
References
Appendix D
12.
Effective Date
Table of Contents
13.
Distribution
14.
Author
Alan J. Mannarino
NOTE: This manual complies with the NCCLS guidelines Vol. 4, Section 2.1. As a user, you
can copy and compile these above sections into one compact NCCLS Stat Profile Ultra (F,
G, H, I, J, and K) Manual or use the cross-reference to find the appropriate section in the
Stat Profile Ultra (F, G, H, I, J, and K) Reference Manual.
Table of Contents
Contents
1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1
1.2
1.3
1.4
2.3
2.4
TOC-1
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1
3.2
3.3
3.4
3.5
3.6
3.7
3.8
3.9
3.10
3.11
3.12
3.13
3.14
3.15
TOC-2
Table of Contents
3.16
3.17
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.1
4.2
4.3
4.4
TOC-3
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.1
5.2
5.3
TOC-4
Table of Contents
5.4
6.3
6.4
6.5
TOC-5
Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
B.1
B.2
B.3
B.4
TOC-6
Reference Values . . . . . . . . . . . . . . . . . . . . . . . . . . .
Screen Maps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ordering Information . . . . . . . . . . . . . . . . . . . . . . .
Shutdown/Reinstallation Procedures . . . . . . . . . . . .
G.1
G.2
D
E
F
G
D-1
E-1
F-1
G-1
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-1
PN 16794 Rev. B 6/97
1 Introduction
Introduction
This manual provides all necessary instructions for the routine operation and maintenance of
the Stat Profile Ultra Analyzer. Please read this manual carefully. It has been prepared to help
you attain optimum performance from your Stat Profile Ultra Analyzer.
WARNING: Blood samples and blood products are potential sources of hepatitis and
other infectious agents. Handle all blood products and flowpath components (wasteline, septum assembly, probe, flow cell, etc.) with care. Gloves and protective
clothing are recommended.
This section introduces the Stat Profile Ultra Analyzer and covers requirements, tests
performed, procedural limitations, clinical utility, and sample handling.
1.1
Installation
This section covers the installation requirements and assembly procedures for the Stat Profile
Ultra Analyzer.
NOTE: Under the Ultra Warranty, a Nova service representative will install
this equipment for you.
1.1.1 Requirements
Working Area Requirements:
The analyzer should be above any instrument, apparatus, or the like that could damage it by
fluid from a spill. Keep the working area around the system free of dirt, corrosive fumes,
vibration, and excessive temperature changes. Ambient operating temperature is 16 C to
30 C (61F to 86 F).
Electrical Requirements:
A grounded, 3-wire receptacle within 5 feet of the system is required for operation. The U.S.
models require a 120 Volt AC line at 50/60 Hz frequency. The analyzer can be operated at
90 - 264 Volt AC 50/60 Hz.
Fuse requirements:
1-1
1. Intro.
1. Intro.
1.2
Hct
Na
Cl
BUN
Hct
Na
Cl
Glu
BUN
Hct
Na
Cl
Glu
Lact BUN
Hct
Na
iCa
Hct
Na
iCa
Glu
BUN
Hct
Na
iCa
Glu
Lact BUN
1-2
BUN
1 Introduction
1. Intro.
P50
Normalized Calcium (nCa) - only analyzers that measure ionized calcium
Anion Gap (An. Gap)
Alveolar Oxygen (A)
Arterial Alveolar Oxygen Tension Gradient (AaDO2)
When a flashing result is displayed on the results screen and printed with a question mark, this
is an indication that a minor error has occurred during the analysis. Refer to the displayed error
code to interpret whether to accept or reject the result.
Clinical Utility1
The following list includes the clinical utility information for each of the analytes measured
on the Stat Profile Ultra Analyzer.
Blood Gases:
(PCO2, PO2,
and pH)
O2 Saturation
Lactate
Sodium
Potassium
1-3
1. Intro.
1.3
Chloride
Ionized
Calcium
Hematocrit
Glucose
Measurement of glucose is used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal
hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell
carcinoma.
Urea Nitrogen
Ref.
The Sample
This section covers sample requirements and reference values for the Stat Profile Ultra
Analyzer.
1-4
1 Introduction
1. Intro.
PCO2*
PO2*
Sodium
Potassium
Chloride
iCa
Glucose
Lactate
BUN
Hematocrit
SO2*
Samples
Whole blood
Whole blood, expired gas
Whole blood, expired gas
Whole blood, plasma, or serum
Whole blood, plasma, or serum
Whole blood, plasma, or serum
Whole blood, plasma, or serum
Whole blood, plasma, or serum
Whole blood, plasma, or serum
Whole blood, plasma, or serum
Whole blood
Whole blood
*Anaerobic samples
Capillary Samples
Capillary sampling requires a capillary tube with a minimum delivery volume of 225 L.
Adapt the capillary tubes to the instrument with Nova Capillary Adapters (PN 06529) as
described in Capillary Sample Analysis, Section 3.17.2.2.
National Committee for Clinical Laboratory Standards. Considerations in the Simultaneous Measurement of Blood Gases, Electrolytes, and Related Analytes in Whole Blood; Proposed Guideline.
NCCLS Document C32-P. Vol. 13, No. 17.
1-5
1. Intro.
NOTE: All other parameters should be obtained from the undiluted, unmodified blood sample.
1. Mix the sample in the syringe. Pipet 0.5 mL into a conical sample cup.
2. Pipet 1.0 mL of isotonic saline into this sample cup. Mix well.
3. Aspirate this diluted sample into the Ultra to measure the glucose concentration.
Multiply the observed concentration by 3.0 to establish the initial concentration.
NOTE: All other parameters should be obtained from the undiluted, unmodified blood sample.
1. Centrifuge the whole blood. Pipet 0.5 mL of serum/plasma into a conical sample cup.
2. Pipet 1.0 mL of isotonic saline into this sample cup. Mix well.
3. Aspirate this diluted sample into the Stat Profile to measure the lactate concentration.
Multiply the observed concentration by 3.0 to establish the initial concentration.
1-6
1 Introduction
1. Intro.
NOTE: All other parameters should be obtained from the undiluted, unmodified blood sample.
1. Centrifuge the whole blood. Pipet 0.5 mL of serum/plasma into a conical sample cup.
2. Pipet 0.5 mL of isotonic saline into this sample cup. Mix well.
3. Aspirate this diluted sample into the Stat Profile Ultra M Analyzer to measure the
BUN concentration. Multiply the observed concentration by 2.0 to establish the
initial concentration.
1-7
1. Intro.
SO2% Interferences:
High levels of COHb and MetHb will increase the reported SO2% value.
MetHb values above 15% will interfere with the SO2% value.
Bilirubin values greater than 10 mg/dL will interfere with the SO2% value.
Lipid solutions of greater than 50 mg/dL will interfere with the SO2% value.
Hemolysed samples will interfere with the SO2% value.
Hematocrit (Hct) Interference:
White Blood Count (WBC) greater than 50,000 WBC/L may increase the hematocrit value.
1.4
1-8
1. Intro.
SO2% Interferences:
High levels of COHb and MetHb will increase the reported SO2% value.
MetHb values above 15% will interfere with the SO2% value.
Bilirubin values greater than 10 mg/dL will interfere with the SO2% value.
10% Intralipid solutions will interfere with the SO2% when the blood concentrations of
Intralipid are >500 mg/dL.
Hemolysed samples will interfere with the SO2% value.
Hematocrit (Hct) Interference:
White Blood Count (WBC) greater than 50,000 WBC/L may increase the hematocrit value.
1.4
1-8
2 Setup
2
2.1
2.2
Adapting the Program to Your Clinical Requirements with the Set Up Menu
Use the Set Up Menu to adapt the analyzer to your requirements. This menu is used if the
analyzer loses power and the internal battery is dead (Section 2.5) or if the barometric pressure
module is removed. The menu is:
A password, which safeguards the setup parameters, is required to enter the Set Up Menu. Set
up the system program as follows:
1. From the Ready For Analysis screen, press MENU to access the Main Menu screen.
2. Press 4 to access the Set Up Menu screen. The Enter System Password screen appears
first. The screen displays a flashing cursor next to the point for keying in the
password.
3. Key in the password (a number) and press ENTER. The analyzer is factory-set with
0 as the default password. Upon entry of the correct password, the Set Up Menu
screen is displayed.
2-1
2. Setup
1
2
3
4
5
6
30 Apr 95
11:45:45
2-2
2 Setup
2.2.1 Operation Configuration
1. From the Set Up Menu, press 2 to display the Operation Configuration screen.
2. Press the number before the desired option.
3. Press CLEAR to return to the Set Up Menu screen.
OPERATION CONFIGURATION
1
2
3
4
5
6
Analysis Configuration
Calibration Configuration
Barometric Pressure
Set Date And Time
Set Analyzer Identification
Sample Number Counter
On
11:45:45
2.2.1.1Analysis Configuration
1. From the Operation Configuration Menu, press 1 to display the Analysis Configuration screen.
2. Press the number before the desired option to change the mode.
3. Press 5 to display the Analysis Reporting Options screen.
4. Press CLEAR to return to the Operation Configuration screen.
2-3
2. Setup
This allows you to access the menus and screens to adapt your analysis and calibration options,
set the barometric pressure, set the date/time, set the instrument identification, display reagent
remaining information.
Full Panel
2. Setup
Single
Auto
Enabled
30 Apr 95
11:45:45
5. In the Analysis Output Options screen, press the number to change the mode of the
option.
6. Press CLEAR, CLEAR to return to the Operation Configuration screen.
ANALYSIS OUTPUT OPTIONS
1
2
3
4
Analysis
Transmit
Transmit
Transmit
Transmit
Analysis
Analysis
Analysis
Mode
mV Data
Drift Data
Errors
5
6
7
30 Apr 95
11:45:45
2-4
Manual
Off
Off
Off
Auto
On
On
2 Setup
2.2.1.2Calibration Configuration
CALIBRATION CONFIGURATION
30 minutes
30 Apr 95
11:45:45
2-5
2. Setup
Access this screen to select the frequency of Gas Cal I. A Gas Cal I is a single point calibration
for PCO2 and PO2 electrodes. A Gas Cal I cannot be aborted. The message, Gas Cal I Pending,
will be posted on the Ready For Analysis screen 5 minutes before it starts.
This screen is also used to allow or not allow the transmitting and/or printing of calibration
information.
1. From the Operation Configuration screen, press 2 to display the Calibration Configuration screen.
2. Press 1 to set the Cal I Frequency time of 30 or 45 minutes.
3. Press 2 to display the Calibration Output Options screen. Then press the number
before the desired option to change the mode.
4. Press CLEAR, CLEAR to return to the Operation Configuration Menu.
2. Setup
This screen allows you to change the units for barometric pressure and update the barometric
pressure.
1. From the Operation Configuration screen, press 3 to display the Barometric Pressure screen.
2. Press 1 and/or 2 to change the mode.
BAROMETRIC PRESSURE
Update Barometric Pressure
Units of Measurement
mmHg
Method of Determination
Measured
30 Apr 95
11:45:45
3. Press 3 to display the Enter Barometric Pressure screen. Change the pressure, if
necessary, and press CLEAR, CLEAR to return to the Operation Configuration screen.
ENTER BAROMETRIC PRESSURE
760.0 mmHg
30 Apr 95
11:45:45
2-6
2 Setup
2.2.1.4Set Date and Time
1. Press 2 from the Set Up Menu. Then press 4 from the Operations Configuration Menu
to display the Set Date/Time screen.
2. Key in the new date and press ENTER to enter new date. Or, press ENTER
immediately to skip to the time entry.
3. Key in the new time and press ENTER to enter the new time.
4. Press CLEAR to exit this screen.
DATE/TIME
Date:
30/04/95
(DD/MM/YY)
Time:
11:45:45
(HH:MM:SS)
30 Apr 95
11:45:45
2-7
2. Setup
The Stat Profile Ultra contains an internal battery-backed clock and calendar that has been set
at Nova Biomedical for Eastern Standard Time. To set the date and time for your location
proceed as follows:
2. Setup
1. From the Operation Configuration screen, press 5 to display the Analyzer Identification screen.
2. Enter the analyzer's identification number (2-digits) and press ENTER to enter the
number or press CLEAR to cancel the change.
3. Press CLEAR to return to the Operation Configuration screen.
ANALYZER IDENTIFICATION
Analyzer Identification #:
nn
30 Apr 95
11:45:45
2-8
2 Setup
SAMPLE NUMBER COUNTER
123
___
2. Setup
30 Apr 95
11:45:45
Electrolyte Resolutions
Units of Measurement
Electrode Offsets
Results Reporting
11:45:45
2-9
2. Setup
This screen sets the number of decimal places for Na+ and K+ display or printout. Values are
rounded to the decimal place shown. (The values shown in the screen are for illustration only.)
Set electrolyte resolution as follows:
1. Press 1 to display the Set Electrolyte Resolution screen.
2. Press the number before the desired electrolyte to change the resolution.
3. Press CLEAR to return to the Results Configuration screen.
ELECTROLYTE RESOLUTION
Na+ Result
xxx.
K+ Result
xx.xx
Cl- Result
xxx.x
30 Apr 95
11:45:45
2-10
2 Setup
UNITS OF MEASUREMENT
Temperature
Blood Gas Values
Ca++ Concentration
Glu Concentration
Lac Concentration
BUN/Urea Concentration
Hemoglobin
C
kPa
mmol/L
mg/dL
mmol/L
mg/dL BUN
mmol/L (Hb/4)
2. Setup
1
2
3
4
5
6
7
30 Apr 95
15:45:45
2.2.2.3Electrode Offsets
The values entered in this screen offset the measured results by the selected percentage. This
is used to correlate the results to the other methodologies currently being used in your
laboratory.
NOTE: Use great care in establishing the offset values for your lab. Contact
Nova Technical Service if you need assistance to determine offset values.
Enter electrode offset values as follows:
1. From the Results Configuration screen, press 3 to access the Electrode Offsets
screen.
2. Press ENTER repeatedly to move the cursor through the available positive or
negative values.
3. Key in the desired offset value to 1 decimal place next to the original parameter. The
offset ranges are as follows in Table 2.1.
2-11
2. Setup
Electrode Offset
pH
+ 0.0 mpH
PCO2
+ 0.0 %
PO2
+ 0.0 %
+ 0.0 %
SO2%
Hct
+ 0.0 %
Na+
+ 0.0 %
K+
+ 0.0 %
+ 0.0 %
Ca++/ClGlu
+ 0.0 %
Lac
+ 0.0 %
BUN
+ 0.0 %
Press ENTER to Skip Entry.
Press CLEAR to Exit.
30 Apr 95
15:45:45
____________________________________________________________________________________
Table 2.1 Offset Limits
PO2 10%
PCO2 10%
pH
Hct
Na+
K+
ClCa++
SO2
Glu
Lac
BUN
4. Press ENTER to replace the original offset percent value with the new value. To
cancel an entry press CLEAR before pressing ENTER.
5. Repeat Steps 2 through 4 for each value.
6. Press CLEAR to return to the Results Configuration screen.
2-12
2 Setup
2.2.2.4 Results, Measured Results, and Calculated Results Reporting Screens
Results Reporting screen - This screen selects from 2 display and print options. Selecting Off
(suppressed) causes the display and printing of results at patient temperature only. Selecting
On causes the display and printing of results at both the patient temperature and 37.0 C
(98.6F). If this screen is not accessed during setup, results will be displayed and printed at the
patient temperature only.
Measured Results Reporting screen - If sodium, a measured parameter, is Off (suppressed) in this
screen, it cannot be selected On during an analysis and cannot be displayed or printed out for
an analysis. If sodium is On, it can be turned On or Off during an analysis. For the effect of
measured result reporting changes on the calculated results reported, refer to Table 2.3.
Calculated Results screens - If BE-ECF, a calculated parameter, is Off (suppressed) in this screen,
it cannot be selected On during an analysis and cannot be displayed or printed out for an
analysis. If BE-ECF is On, it can be turned on or off during an analysis. For calculated result
reporting changes which affect the calculated results reported, refer to Table 2.3.
Reversible Suppression (On/Off) - On the Measured Results Reporting screen, the analytes are
displayed highlighted and underlined (default). To suppress the analyte (to eliminate unwanted
testing), press the corresponding number and the analyte becomes normal intensity (remains
underlined). The analyte is Off or suppressed. Any errors during calibration or analyses will not
be displayed for the Off analyte. Any measured result that depends on the suppressed analyte
will report as long as the suppressed analyte is calibrated and has no errors. To reverse
suppression, press the corresponding number of the analyte.
On the Calculated Results Reporting screens, the parameters are highlighted when On
(unsuppressed). To suppress or turn Off, press the corresponding number and the parameter
becomes normal intensity. To turn On, press the corresponding number of the parameter
(highlighted).
Blank Electrode Suppression - From the Measured Results Reporting screen, you can permanently
remove available analytes to eliminate unwanted tests. Press and hold the FUNCTION key
then press the corresponding analyte number key. The analyte changes to normal intensity with
no underline. Errors will not be reported. Also, any measured result that depends on this analyte
will be suppressed. To reverse blank electrode suppression, press and hold the FUNCTION
key then press the corresponding analyte number key. The analyte changes to highlighted
intensity with underline. After you turn it On (unsuppress the analyte), you must calibrate the
electrode.
2-13
2. Setup
The Results Reporting function in the Set Up Menu allows access to 4 screens. The Results
Reporting, Measured Results Reporting, ABG Calculated Results Reporting, and Electrolyte
Calculated Results Reporting screens, described as follows:
2. Setup
Suppressed Parameter
Na+
pH
PCO2
PO2
SO2
Hct
Hb, SO2**
Glu, BUN
Osmolality
2-14
2 Setup
Table 2.4 Tests from Test Panel
Full
Panel
Blood
Gas
Blood
Gas +
Pulmonary
Electrolyte
Metabolite
Expired
Gas
X
X
X
X
X
X
X
X
X
X
X
X
................................................................................. X ........................................
................................................................................. X ........................................
................................................................................. X ........................................
................................................................................. X ........................................
................ X ................. X ................. X ..............................................................
................ X ................. X ................. X ............................................................ X
................ X ................. X ................. X ............................................................ X
...................................... X ................. X ..............................................................
...................................... X ................. X ..............................................................
....................................................................................................... X ..................
............................................................ X ....................................... X ..................
....................................................................................................... X ..................
Calculated Values*
Base Excess Blood:
Base Excess ECF**:
Bicarbonate:
Std. Bicarb.Conc.:
Total CO2:
O2 Content:
Normalized Calcium:
Hemoglobin:
Anion Gap:
Osmolality:
A:
AaDO2:
a/A:
P50:
PO2/FI:
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
2. Setup
Measured Tests
Sodium:
Potassium:
Chloride:
iCalcium:
pH:
PCO2:
PO2:
SO2:
Hct:
Glucose:
Lactate:
BUN:
2-15
2. Setup
1. From the Set Up Menu, press 1 then 4 to display the Results Reporting screen.
2. Press 1 to turn Results Measured On or Off.
3. Press TEST SELECT to display the Measured Results Reporting screen.
RESULTS REPORTING
On
30 Apr 95
11:45:45
Results Reporting:
Press Number
Hilite - Reporting
Configure Electrodes:
Press Function + Nbr
Underline - Installed
pH
1
PCO2
2
PO2
3
SO2%
4
Hct
5
Na+
6
K+
7
Ca++/Cl8
Glu
9
Lac
.
BUN
0
11:45:45
4. Press the test number to suppress or unsuppress the test or press FUNCTION and the
test number to blank electrode suppress or unsuppress the test.
2-16
2 Setup
Results Reporting:
Press Number
Hilite - Reporting
Hb
1
BE-ECF
2
BE-B
3
SBC
4
HCO35
TCO2
6
O2ct
7
A
8
AaDO2
9
a/A
.
P50
0
PO2/FI
-
11:45:45
Results Reporting:
Press Number
Hilite - Reporting
nCa++
1
Osm
2
11:45:45
2-17
2. Setup
5. Press TEST SELECT again to display the Gas Calculated Results Reporting screen
or press CLEAR to return to the Results Configuration screen.
6. Press the test number to suppress or unsuppress the test.
7. Press TEST SELECT again to display the Electrolyte Calculated Results Reporting
screen or press CLEAR to return to the Results Configuration screen.
8. Press the test number to suppress or unsuppress the test.
9. Press CLEAR to return to the Results Configuration screen.
2. Setup
Results Reporting:
Press Number
Hilite - Reporting
Osm
1
An.Gap
2
11:45:45
pH
PCO2
PO2
SO2%
Hct
Na+
K+
Ca++/ClGlu
Lac
BUN
REFERENCE
LOW
HIGH
0.000
0.000
0.0
0.0
0.0
0.0
0.0
0.0
0.
0.
0.0
0.0
0.00
0.00
0.00
0.00
0.0
0.0
0.0
0.0
0.0
0.0
PANIC
LOW
0.000
0.0
0.0
0.0
0.
0.0
0.00
0.00
0.0
0.0
0.0
HIGH
0.000
0.0
mmHg
0.0
mmHg
0.0
0.0
%
0.0
mmol/L
0.00 mmol/L
0.00 mmol/L
0.0
mg/dL
0.0
mmol/L
0.0
mg/dl
11:45:45
2-18
2 Setup
2.3.3 Quality Control Setup
1
2
3
4
5
6
7
Lot Number
2316.10.94
3524.12.94
3525.12.95
2822.05.95
2823.05.95
3705.02.95
Status
Expired
Active
Active
Active
Due
Incomplete
Empty
30 Apr 95
11:45:45
1 QC Lockout Mode
Disabled
Auto
Manual
4
5
6
7
Manual
Auto
Manual
Manual
Print
Print
Print
Print
30 Apr 95
11:45:45
2-19
2. Setup
QC Lockout Mode
Disabled
Reminder Only A warning message, "QC Analysis Due," is displayed on the Ready for
Analysis screen 30 minutes prior to the programmed analysis time (see
Step 8) up until the QC sample has been run.
Mandatory QC A warning message, "QC Analysis Due," is displayed on the Ready for
Analysis screen 30 minutes prior to the programmed analysis time (see
Step 8) up until the programmed analysis time. A new message, "QC
Required," is displayed at the programmed analysis time and remains
until the QC sample has been run. Results for the tests that require QC
will not be displayed for any patient samples in this mode.
NOTE: Analysis is not prevented in this mode, but result displays are.
Therefore, if QC is required for all tests and an analysis is run, the sample
will be analyzed but no results will be reported. After the analysis is run,
there is no method for retrieving the results. (See Operation, Override
Mandatory QC Analysis.)
Auto(matic) operation of daily functions occurs at 23:40 hours each day. Automatic operation
of monthly functions occurs at 23:40 hours on the last day of each month.
2-20
2 Setup
Move Daily Data to Month-to-Date
Daily data is moved to Month-to-Date data for each control if daily data
exists. Daily data is then erased.
2. Setup
#1
Lot Number:
Expiration Date:
oooooooooo
dd/mm/yy
(DD/MM/YY)
hh:mm
Analysis Frequency
hh
(HH:MM)
hours.
11:45:45
2-21
2. Setup
8. Enter or change information as desired. Press ENTER to input information into the
position indicated by the reverse video cursor bar. Move the cursor with the up and
down arrow keys. "Initial Analysis Time" is the time of day you desire a particular
control to be run. "Analysis Frequency" determines how often the control is run after
the "Initial Analysis Time".
The following warning appears if an attempt is made to enter a new Lot Number for
a control that is not Empty or is Incomplete:
WARNING:
Changing the Lot Number will
delete all data for the current
control.
Press ENTER to Confirm Change.
Press CLEAR to Cancel Change.
The following warning appears if an attempt is made to delete the Lot Number (by
entering the Erase Code Characters: _ (Dot, Dash, Dot) for a control that is either
Active or Expired:
WARNING:
Deleting the Lot Number will
delete the data for the current
control.
Press ENTER to Confirm Change.
Press CLEAR to Cancel Change.
The following warning appears if an attempt is made to change the Expiration Date
for a control that is either Active or Expired:
WARNING:
The Expiration Date must match
the date specified by the
manufacturer.
Press ENTER to Confirm Change.
Press CLEAR to Cancel Change.
When changing lots of controls, all data from the existing lot should be printed or
recorded for laboratory records prior to entering a new lot number. At the time a new
lot number is entered, all stored data for the previous lot number is erased.
One method of performing crossovers is to enter the new lot information and track
the existing lot manually for the crossover period.
2-22
2 Setup
pH
n.nnn
PCO2
nnn.n
PO2
nnn.n
Na+
nnn.n
K+
nn.nn
Ca++/Cl- nn.nn
Glu
nnn.n
Lac
nn.nn
BUN
nn.nn
High
n.nnn
nnn.n
nnn.n
nnn.n
nn.nn
nn.nn
nnn.n
nn.nn
nn.nn
mmHg
mmHg
mmol/L
mmol/L
mg/dL
mg/dL
mmol/L
mmol/L
11:45:45
2-23
2. Setup
2. Setup
1. From the Set Up Menu, press 4 to display the Communications Setup screen.
2. Press CLEAR to return to the Set Up Menu screen.
COMMUNICATIONS SETUP
Results Transmission
COM1
Ticket Printer
COM3
Remote Diagnostic
COM2
30 Apr 95
11:45:45
2-24
2 Setup
TRANSMISSION PORT CONFIGURATION
Warning, Port Previously Assigned !
Port
COM1:
Baud
4800
Parity
Odd
Word Length
Stop Bits
ACK Required
Enabled
2. Setup
30 Apr 95
11:45:45
Patient Temperature
37.0
Hemoglobin
14.3
g/dL
FIO2
(Disabled)
20.9
30 Apr 95
11:45:45
2-25
2. Setup
NOTE: For an individual test you may enter patient temperature or hemoglobin in the Patient Data screen.
2-26
From the Set Up Menu, press 6 to display the Password Options screen.
Press 1, 2, or 5 to change the options from enabled to disabled, etc.
Press 3 to display the Operator Passwords: #1 or Remote Review.
Press 4 to display the Operator Passwords: #2 or No Remote Review.
Press 6 to display the Technician Passwords screen.
Press 7 to display the Supervisor Password screen.
Press 8 to set all passwords to zero.
Press CLEAR to return to the Set Up Menu screen.
2 Setup
PASSWORD OPTIONS
Enabled
2
3
4
Enabled
#1
#2
Enabled
6
7
Passwords Entry
Password Entry
---
2. Setup
Technician
Supervisor
11:45:45
Password Inputs
1. From the Password Options screen, press 3 to display the Operator Passwords screen
for Group #1 or Remote Review (if enabled), or press 4 to display the Operator
Passwords screen for Group #2 or No Remote Review (if enabled).
2. Press ENTER until you come to the desired operator number. Then enter a password
of up to 4 digits. Press ENTER to continue to the next operator number.
3. Press CLEAR to return to the Password Options screen.
OPERATOR PASSWORDS
Remote Review (or Group #1)
OprPswdOprPswdOprPswd
101
nnnn
116
nnnn
131
nnnn
102
nnnn
117
nnnn
132
nnnn
103
nnnn
118
nnnn
133
nnnn
104
nnnn
119
nnnn
134
nnnn
105
nnnn
120
nnnn
135
nnnn
106
nnnn
121
nnnn
136
nnnn
107
nnnn
122
nnnn
137
nnnn
108
nnnn
123
nnnn
138
nnnn
109
nnnn
124
nnnn
139
nnnn
110
nnnn
125
nnnn
140
nnnn
111
nnnn
126
nnnn
141
nnnn
112
nnnn
127
nnnn
142
nnnn
113
nnnn
128
nnnn
143
nnnn
114
nnnn
129
nnnn
144
nnnn
115
nnnn
130
nnnn
145
nnnn
Press ENTER to Skip Entry, CLEAR to Exit
2-27
2. Setup
OPERATOR PASSWORDS
No Remote Review (or Group #2)
OprPswdOprPswdOprPswd
201
nnnn
216
nnnn
231
nnnn
202
nnnn
217
nnnn
232
nnnn
203
nnnn
218
nnnn
233
nnnn
204
nnnn
219
nnnn
234
nnnn
205
nnnn
220
nnnn
235
nnnn
206
nnnn
221
nnnn
236
nnnn
207
nnnn
222
nnnn
237
nnnn
208
nnnn
223
nnnn
238
nnnn
209
nnnn
224
nnnn
239
nnnn
210
nnnn
225
nnnn
240
nnnn
211
nnnn
226
nnnn
241
nnnn
212
nnnn
227
nnnn
242
nnnn
213
nnnn
228
nnnn
243
nnnn
214
nnnn
229
nnnn
244
nnnn
215
nnnn
230
nnnn
245
nnnn
Press ENTER to Skip Entry, CLEAR to Exit
4. From the Password Options screen, press 6 to display the Technician Passwords
screen.
5. Press ENTER until you come to the desired operator number. Then enter a password
of up to 4 digits. Press ENTER to continue to the next operator number.
6. Press CLEAR to return to the Password Options screen.
TECHNICIAN PASSWORDS
OprPswdOprPswdOprPswd
301
nnnn
316
nnnn
331
nnnn
302
nnnn
317
nnnn
332
nnnn
303
nnnn
318
nnnn
333
nnnn
304
nnnn
319
nnnn
334
nnnn
305
nnnn
320
nnnn
335
nnnn
306
nnnn
321
nnnn
336
nnnn
307
nnnn
322
nnnn
337
nnnn
308
nnnn
323
nnnn
338
nnnn
309
nnnn
324
nnnn
339
nnnn
310
nnnn
325
nnnn
340
nnnn
311
nnnn
326
nnnn
341
nnnn
312
nnnn
327
nnnn
342
nnnn
313
nnnn
328
nnnn
343
nnnn
314
nnnn
329
nnnn
344
nnnn
315
nnnn
330
nnnn
345
nnnn
Press ENTER to Skip Entry, CLEAR to Exit
7. From the Password Options screen, press 7 to display the Supervisor Passwords
screen. Then enter a password of up to 4 digits.
8. Press CLEAR to return to the Password Options screen.
2-28
2 Setup
SUPERVISOR PASSWORD
__
2. Setup
Password
30 Apr 95
15:45:45
2.3
From the Main Menu, press 1 to display the Patient Recall Options screen.
30 Apr 95
15:45:45
2-29
2. Setup
12 Records Transmitted.
15:45:45
4.
5.
Time
Accession #
Patient I.D.
10 Mar 95
17:47:30
12345678901234567890
12345678901234567890
10 Mar 95
17:47:30
12345678901234567890
12345678901234567890
10 Mar 95
17:47:30
12345678901234567890
12345678901234567890
10 Mar 95
17:47:30
12345678901234567890
12345678901234567890
Fnc/-PgUp
-LnDn
6. To view the patient data, enter the number on the left of the screen (1-4) associated
with the patient.
2-30
2 Setup
7. From the Patient Data Recall Menu, press 4 or 5 to delete all stored or uploaded stored
results.
8. Press ENTER when ready to delete the results.
9. Press CLEAR to exit.
2. Setup
Warning !
30 Apr 95
15:45:45
2.4
Setup Values
Calibration Gas Values
Barometer Offset
Percent Reagent Remaining in the Reagent Pack
Previous Results
Calibration slopes, pending analysis results and patient data, operation menu values (other than
the barometer offset), and error codes will be lost. Upon return of power, run a calibration, enter
Operation Menu choices, and resume operation.
If the analyzer has been set up previously but upon power loss Set Up Required appears on the
Ready For Analysis screen, the internal battery has died and no setups are saved. The password
is reset to 0 (zero). After accessing the Set Up Menu and performing the setup procedure,
normal operation can resume. Contact Nova Technical Service to have a Service Representative replace the battery.
2-31
3 Operation
Operation
The Stat Profile Ultra Components are shown in Figure 3.1. These components are described
in the following sections.
3
6
3. Operation
5
1
DWG #7-M-003A
2
MENU
CALIBRATE
TEST SELECT
CURSOR
QC
ANALYZE
PATIENT DATA
ENTER
STATUS
CLEAR
FUNCTION
1.
2.
3.
4.
5.
6.
7.
CRT Display
Keypad
Printer
Septum Assembly/Probe Port
Analytical Compartment
Fluids Deck (Reagent Flow Controllers)
Reagent Pack
3-1
CRT Display
3. Operation
The CRT Display displays screens of information on system functions, how to access system
functions, and output patient results. If the instrument is idle for more than 15 minutes, the
computer will blank the CRT to minimize wear. In this situation, pressing the . key will recall
the screen that was blanked.
NOTE: Pressing keys other than . will recall the previous screen as well, but
may activate menu items on these screens.
3.2
CALIBRATE
QC
MENU
CURSOR
TEST SELECT
ANALYZE
ENTER
PRINT
FUNCTION
PATIENT DATA
CLEAR
STATUS
NOTE: If the instrument is idle for more than 15 minutes, the computer will
blank the CRT to minimize wear. In this situation, the pressing of any key will
recall the screen that was blanked, but will not execute the function described
in Table 3.1. Any subsequent key press will execute its associated function.
Table 3.1 Key Functions
3-2
Key
Function
CALIBRATE
Starts a full 2-point calibration. This key is inoperative if the system is busy
with another sequence.
QC
From the Aspirate Analysis Sample screen, displays the QC Selection screen
to allow the operator to specify the current analysis as a proficiency analysis
(e.g., an analysis for a CAP survey) or as an analysis of one of the specified
QC control materials.
Manually prints the information displayed on the current screen. If the printer
is already printing, the PRINT key is disabled.
3 Operation
Table 3.1 Key Functions (continued)
Function
FUNCTION
Use the FUNCTION Key with the other keys to alter the functionality of other
keys, such as:
1.
To cause the ANALYZE Key to start an analysis, use the alternate
analysis mode.
2.
To cause the CALIBRATE Key to display the Calibration Options
screen, you can execute partial as well as full calibrations.
ENTER
CURSOR
CLEAR
MENU
Displays the Main Menu. This allows you to use the support functions of the
software. As you become more experienced, you can bypass the MENU key
(and the Main Menu) by pressing the digit key corresponding to the support
functions.
3-3
3. Operation
Key
Key
Function
TEST SELECT
1.
3. Operation
2.
3.
PATIENT DATA
1.
2.
3.
From the Ready For Analysis screen, display the Test Panels screen for
establishing which of the 6 panel options are to be reported for the next
analysis.
From the Patient Data screen, displays the first of the Results Reporting
screens, for establishing which analytical test results are to be
reported for the currently executing or just completed analysis.
From the Results Reporting Configuration screen, displays the first of
the Global Results Reporting screens, for establishing which of the
analytical test results are to be reported for all analyses.
The Patient Data screen cannot be accessed if the CRT is displaying a Menu
or Status screen or the results of a QC analysis.
STATUS
ANALYZE
Starts an analysis in the Default Analysis mode. This key is inoperative if the
system is busy with another sequence.
3-4
1.
2.
Numeric entry.
Menu Selection.
3 Operation
Table 3.1 Key Functions (continued)
Key
- (Minus Sign)
Printer
The Printer uses heat-sensitive paper and a 100 dot thermal print head.
3.4
Septum Assembly
The septum assembly is described in the next section, Analytical Compartment.
3.5
Analytical Compartment
The Analytical Compartment contains the electrodes and the components of the flow path, i.e.,
the Sample Probe, Septum Assembly, Sample Preheater, Flow Cell, and tubing harnesses
which bring the samples and standards to the electrodes. The Barometric Pressure Module, Gas
Humidifier Wells, Reference Electrode, Reagent Preheater, Fiber Optic Module, and temperature sensors are also in this compartment. A thermostat maintains the compartment temperature at 37 C.
3-5
3. Operation
3.3
Function
1
2
3. Operation
8
8
7
11
12
4A
6
4B
5
S
DWG #7-11-004C
4C
1.
2.
3.
4a.
4b.
4c.
5.
3-6
Reagent Harness
Reagent Preheater
Gas Humidifier Wells
Septum Harness
Sample Probe
Septum Assembly
Electrode Rack Assembly
6.
7.
8.
9.
10.
11.
12.
Sample Preheater
Flow Cell
Electrodes
Reference Electrode
W/R-Harness
Fiber Optic Module
Flow Cell Light
3 Operation
3.5.1 Reagent Harness
The Reagent Harness transports fluids from the reagent pack to the analytical components.
The Reagent Preheater equilibrates the reagent pack fluids and the gases to 37 C.
3-7
3. Operation
3.5.8 Electrodes
The Electrodes housed in the flow cell are the core of the Stat Profile Ultra.
Electrodes are flow-by to minimize sample size. The methodology used by each electrode is
shown Table 3.2.
____________________________________________________________________________________
Table 3.2 Stat Profile Ultra Electrode Methodologies
Electrode
Methodology
Sodium
Potassium
Chloride
iCa
pH
PCO2
Severinghaus-type electrode
PO2
Hct
Impedance electrode
Glucose
Enzymatic electrode
Lactate
Enzymatic electrode
BUN
Enzymatic electrode
SO2
____________________________________________________________________________________
3-8
3 Operation
The electrodes clip into the flow cell (the Hct electrode is part of the sample preheater), and
the cables plug into the electrode rack. The voltages read by the Na+, K+, Cl-, iCa, BUN, and
pH electrodes are related to a reference voltage measured at the reference electrode. Any
reference dependent electrode has a single wire lead. Voltages from the PCO2, glucose, lactate,
and PO2 electrodes are measured by internal reference electrodes. These electrodes have 2 wire
leads. The fiber optic module is used to measure voltages for SO2% calculations.
The Reference Electrode is mounted on the sample preheater above the flow cell. It is a solidstate electrode continuously supplied with reference solution to provide the reference voltage
for comparison to sample voltages. The exit port of the flow path is located on this electrode.
3-9
3. Operation
3. Operation
1. Pinch Valve
2. Pump
3. Pump Bypass Valve
W R
1
W R
3
C
DWG #5-10-003B
3-10
3 Operation
3.6.2 Pump
The Pump uses peristaltic action to draw the fluids through the tubing and the flow path. A
stepper motor drives the pump and moves at low, medium low, medium high, and high speeds.
The pump rollers contact the waste line to draw standards or samples through the W-line and
simultaneously supply reference solution.
The Pump Bypass Valve is a pinch valve which allows gas to bypass the pump (gas cannot flow
by peristaltic action) and flow through the bypass tubing segment in the W/R-harness.
3.7
Reagent Pack
The Reagent Pack contains standards for pH, electrolyte (Na+, K+, and Cl-), hematocrit, iCa,
glucose, BUN, and lactate electrode calibration, flush solution for washing the flow path, and
reference solution for establishing a reference voltage at the reference electrode. The pack also
has a Reagent Management System cartridge (see Section 3.13) and a waste bottle to collect all
system wastes. SO2 is calibrated separately using Oxygen Saturation Calibrator solutions.
3-11
3. Operation
Fluidic Diagram
FLUSH
3. Operation
REAGENT
PACK
REF
WASTE
WASTE
Pump
R W
REFERENCE
ELECTRODE
ELECTRODES
PINCH
VALVE
SAMPLE PREHEATER
WITH AIR DETECTOR
GASES
GAS
HUMIDIFIER
WELLS
SAMPLER PROBE
GAS
DWG #5-10-004A
FLUSH
C
D
A
B
3-12
SEPTUM ASSEMBLY
3 Operation
3.9
Operation Overview
3-13
3. Operation
The Stat Profile Ultra Analyzer is an automatic, microcomputer-based system for analyzing
samples for blood gas, electrolytes, iCa, glucose, lactate, BUN, SO2, and hematocrit. System
access is via the keypad, allowing entry of data and instructions in response to information
displayed on the CRT.
After the instrument is set up and calibrated, the CRT displays the Ready For Analysis screen
signifying that the instrument is ready. The operator presses ANALYZE to extend the probe
into the sample, immerses the probe, then presses ANALYZE again to begin the analysis cycle.
After the sample is drawn into the flow path to the electrodes, the electrode millivolt values are
measured (error messages are generated if any problems are detected). Patient data is entered
by the operator, calculations for measured parameter results are performed, and these results
are in turn used for computing the calculated results. The measured and calculated results are
displayed on the CRT and printed out.
A 2-point calibration establishes the electrode slopes used for result calculations. The
instrument performs a 2-point calibration every 2, 4, or 6 hours, depending on the number of
analyses performed. See Section 3.15 for details. A 1-point calibration (Cal 1) is performed at
30 or 45 minute intervals and is independent of the sample cycle. Patient results are calculated
using the millivolt difference between the patient sample and the 1-point calibration standard
together with the electrode slope from the last 2-point calibration. The 1-point calibration is
also used to check for unacceptable electrode drift (electrode drift is defined as a small change
in electrode signal over time, the slope remaining constant).
The following description of the 2-point calibration can also serve as an example of what
happens during a 1-point calibration and sample cycle. Refer to Figure 3.5. When calibration
starts, the pinch valve opens reagent line B. The probe tip moves to the Standard B position
in the septum assembly, the pump turns, Standard B is aspirated through the sample preheater
and the air detectors which check for proper flow. During aspiration, the pump simultaneously
pumps reference solution into the reference electrode. The pump stops, positioning the
standard in front of the electrodes. The iCa, pH, and hematocrit electrode millivolt readings are
taken. (The second point for iCa is taken when flush is pulled through the system.) This
sequence is repeated for Standard A, establishing the second of the 2 calibration points for the
pH electrode (the second point for the hematocrit electrode is taken when Standard C is read).
The pump bypass valve then opens and the probe tip moves to the gas position in the septum
assembly. An internal gas valve meters Gas A into the system. The gas is humidified before
entering the flow path where the PCO2 and PO2 millivolt readings are taken. This sequence
is repeated for Gas B, establishing the second of the 2 calibration points for the PCO2 and PO2
electrodes. Gas is then purged out of the system and the pump bypass valve closes.
Finally, the sequence for the Na+, K+, Cl-, glucose, BUN, and lactate electrodes, using Standard
D for the first calibration point then Standard C for the second calibration point, is run.
A barometer provides continuous pressure readings which are used in the calculation of blood
gas results.
3. Operation
Analyze a sample
Calibrate the instrument
Perform maintenance
Set up instrument and test parameters
Check instrument status and troubleshoot
Analyze a QC sample
Initiate a printout
Select tests
Enter patient data
All screens give information on how to move to the next screen or how to exit the screen. The
screen that most sequences start from is the Ready For Analysis screen. An introduction to this
screen, and the Status and Main Menu screens, follows.
3-14
3 Operation
3.11 Ready for Analysis Screen
The Ready For Analysis screen tells whether the instrument is ready to perform an analysis.
When the Not Ready version of this screen is displayed, see Section 3.11.1.
3. Operation
30 Apr 95
11:45:45
Analysis mode
When ready for an analysis, information on how to begin the analysis
Uncalibrated electrodes (if any)
Auto-Cal Pending if less than 10 minutes remain to a scheduled calibration
Percent reagent remaining in the reagent pack (if enabled)
Date and time
3-15
3. Operation
NOT READY
30 Apr 95
11:45:45
3-16
3 Operation
3. Press 3 to display the Calibration Option Menu screen.
4. Press 7 to display the Calibration Gases screen.
5. Set the gas calibration values as printed on the gas cylinders as described
in Section 4.1.4 (see also Appendix A).
6. Press CLEAR, CLEAR, CLEAR, to return to the Not Ready screen.
3-17
3. Operation
The reagent pack and RMS pod must be installed and the Reagent Pack screen is needed
to respond to this prompt. Refer to Section 4.4.11.
3. Operation
Sensor Status
Air Detector Status
Sequence Errors
System Errors
System Status
These screens cycle. The screens can be displayed at any time during an analysis, calibration,
or idle sequence. Press ENTER to go to the next Status screen and CLEAR to exit.
SENSOR STATUS
System Idle
SensorSlope
pH
+10.2
+10.2
PCO2
PO2
+10.2
SO2%
+10.2
Hct
+10.2
Na+
+10.2
K+
+10.2
Ca++/Cl- +10.2
Glu
+10.2
Lac
+10.2
BUN
+10.2
Errors Detected
Status Offset
C OK
+ 0.0 mpH
C OK
+ 0.0 %
C OK
+ 0.0 %
C OK
+ 0.0 %
C OK
+ 0.0 %
C OK
+ 0.0 %
C OK
+ 0.0 %
C OK
+ 0.0 %
C OK
+ 0.0 %
C OK
+ 0.0 %
C OK
+ 0.0 %
Reporting
On
On
On
On
On
On
On
On
On
On
On
3-18
Sequence in progress
Time to completion of sequence
Slope of each electrode
3 Operation
Status of each electrode minor or major error occurrence for last calibration or
analysis cycle
-
Reporting status of electrode if results are being reported for the particular
parameter
Electrode offsets (if used)
System Idle
Errors Detected
STATUS
AD2
+123.4
OK
AD3
+123.4
OK
This screen lists air detector errors (which are mainly electrode or fluidic in nature) from the
last analysis (labeled Analysis) and from calibrations and other operations (labeled Others).
The screen is updated with each calibration and analysis. Refer to Section 5.1 for a discussion
of troubleshooting with this screen.
3-19
3. Operation
3. Operation
Analysis
25 Na+ INSTAB C
This screen lists sequence errors (which are mainly electrode or fluidic in nature) from the last
analysis (labeled Analysis) and from calibrations and other operations (labeled Others). The
screen is updated with each calibration and analysis. Refer to Section 5.1 for a discussion of
troubleshooting with this screen.
3-20
3 Operation
3.12.4 System Errors
The System Errors screen is next displayed by pressing ENTER.
SYSTEM ERRORS
A5 AB TEMP TOO HIGH
3. Operation
System errors are hardware or software related. These codes may occur during calibration,
analysis, or during an idle sequence. System errors are cleared at the start of a full calibration.
Refer to Section 5.1 for a discussion of troubleshooting with this screen.
3-21
3. Operation
Barometric Pressure:
37.0
37.1
C
C
756.4 mmHg
3-22
3 Operation
3.13 Reagent Management System Overview
The Reagent Management System consists of the reagent pack cartridge and the analyzer
program that accepts the information from the cartridge.
The Reagent Management System:
:
:
:
:
12345678
4206
Jun 95
84%
Standard
7.384
140.0
4.0
1.0
80.0
2.0
10.0
Concentration
6.840
75.0
10.0
2.0
200.0
6.0
50.0
3-23
3. Operation
3. Operation
MAIN MENU
Maintenance Menu
Setup Menu
Service Menu
30 Apr 95
11:45:45
By pressing 1, 2, 3, 4, or 5, you select the Patient Recall Menu, Maintenance Menu, Quality
Control Menu, Setup Menu, or Service Menu. Each of these menus displays a list of numbered
functions for selection. Each menu and its function are explained in this manual.
The Main Menu (and the MENU key) can be bypassed by pressing the appropriate digit (1 for
Patient Recall Menu, 2 for Maintenance Menu and so forth) from the Ready For Analysis
screen.
3-24
3 Operation
3.15.1 Two-Point Calibration (Automatic and Manual)
3-25
3. Operation
3. Operation
Definitions:
Daily Statistics - All QC samples stored since the last Move Daily Data to Month-to-Date was
performed. A day may represent a shift or multiple days. Setting Move Daily Data to Monthto-Date to auto (in QC Setup) insures that the daily data accumulation begins at midnight of
the current day.
Month-to-Date (M-T-D) Statistics - Cumulative statistics since the last Erase Month-to-Date Data
was performed. Month-to-Date statistics are simply cumulative statistics that can be cleared
on demand. Setting Erase Month-to-Date Data to auto (in QC Setup) insures that the Monthto-Date accumulated data begins on the first day of the current month.
Cumulative Statistics - A running accumulation of mean, SD (standard deviation), CV%
(coefficient of variance in percent), expected range, and n (number of samples) for all stored
QC samples. The Cumulative Statistics cannot be reset except by changing to a new control.
A. To access the Quality Control Reports screen from the Main Menu, press 3,1, or from
the Ready for Analysis screen, press the QC key.
NOTE: If passwords are enabled, the Password Required screen is displayed first.
A complete screen map of all QC screens is in Appendix E.
B. From this screen you have 9 options:
1. Display Summary Statistics for Test - This report will display the Summary
Statistics screen, which contains Daily, Month-to-Date, and Cumulative
Statistics for a single control and test.
a. Press 1.
b. Choose a control by pressing the corresponding number on the
keypad.
c. If blood gas or SO2% control is selected, the next screen is
displayed. Select a test to be viewed.
d. The Summary Statistics screen appears after a test is selected or
directly after a hematocrit control is selected. This screen gives
the Daily, Month-to-Date, and Cumulative Statistics for a
single control and test.
e. From this screen, you can view different controls for the test
selected by pressing the up () or down () arrow key. To view
a different test for a selected control, press the right () or left
() arrow key.
f. To print the results of the selected control and test, press PRINT.
3-26
3 Operation
3. Display Cumulative Statistics for Control - This report will display the
Cumulative Statistics for Control screen, which contains cumulative mean,
SD, CV%, expected range, and n for all tests for the selected control.
a. Press 3.
b. Choose a control by pressing the corresponding number on the
keypad.
c. From this screen, you can view different control results by
pressing the up () or down () arrow key. All applicable tests
are displayed for the selected control.
d. To print the Cumulative Statistics of the selected control, press
PRINT.
4. Print Daily QC Analysis Log - Press 4 for a printout of all QC samples (not
statistics) stored since the last Move Daily Data to Month-to-Date. Includes
date, time, results, and expected ranges. Maximum storage limit is 169 QC
samples. If the storage area is not cleared (by moving daily data to Monthto-Date), QC sample #170 will erase QC sample #1 etc.
5. Print Daily QC Summary Graphs - This report is printed from the Daily QC
Summary screen.
a. Press 5.
b. Choose all tests for which you want results printed by pressing
the corresponding numbers on the keypad.
c. Press ENTER to print the Daily QC Summary Graphs, a
graphical printout of the mean and range of all QC samples not
yet moved to the Month-to-Date statistics.
On the printout, a left or right arrow indicates a data point
outside of the printable range. Left and right brackets indicate
the lower and upper results recorded (range). The asterisk
indicates the mean of the recorded results. If a range indicator
(bracket) and the mean indicator (asterisk) have the same value,
only the asterisk will be printed. An asterisk without any
brackets indicates a single data point (n=1).
3-27
3. Operation
2. Display Cumulative Statistics for Test - This report will display the Cumulative
Statistics for Test screen, which contains the cumulative mean, SD, CV%,
expected range, and n for all controls for that test.
a. Press 2.
b. Choose a test by pressing the corresponding number on the
keypad.
c. To view results of a different test, press the right () or left ()
arrow key.
d. To print the Cumulative Statistics of the selected test, press
PRINT.
3. Operation
7. Print Month-to-Date Statistics by Test - This report will print for all controls
the Month-to-Date statistics for all tests selected.
a. Press 7.
b. Choose all tests for which you want results printed by pressing
the corresponding numbers on the keypad.
c. Press ENTER to print the Month-to-Date Cumulative Statistics
report. The printout includes mean, SD, CV%, n, and expected
range for each test (all controls).
8. Print Levey-Jennings Graphs - to display the Levey-Jennings screen.
a. Press 8.
b. Choose all tests for which you want graphs of the daily mean
printed by pressing the corresponding numbers on the keypad.
c. Press ENTER to print the Levey-Jennings Graphs. This printout includes expected range, cumulative mean, SD, and 2SD
range around the cumulative mean for each selected test.
0. Cancel Print-out - Press 0 to cancel printing.
3-28
Press ANALYZE.
Press QC to display the QC Analysis screen.
Select the appropriate control level.
Present the sample to the probe.
Status Indicators:
Empty - Control not in use.
Incomplete - Lot#, valid expiration date, and at least one range must be set
up to be complete.
Active - Control is properly set up.
3 Operation
Due - In Reminder mode, this indicates that control needs to be run. In
Mandatory mode, this indicates the 30 minute period prior to causing QC
lockout.
Required - In Mandatory mode only, this indicates that QC must be run in
order to display results.
3-29
3. Operation
5. After QC sample analysis, results are displayed. Choose to save all, selected, or no
results.
6. Enter the number next to each test to change the storage selection from STORE to
DISCARD.
7. Press ENTER to store selected results in Daily Statistics. Press CLEAR to discard all
results. Results remain available until ANALYZE is pressed again.
3. Operation
Full Panel
Blood Gas
Blood Gas Plus
Pulmonary
Electrolyte
Metabolite
All tests
pH, PCO2, and PO2
pH, PCO2, and PO2, SO2%, and Hct
pH, PCO2, and PO2, SO2%, Lac and Hct
Na, K, Ca/Cl
Glu, BUN, and Lac
Press the appropriate panel number. Only the panel selected is reported, the others are
suppressed. Any test that was suppressed in setup blinks on/off if it is included in the selected
panel. This alerts you that this test is suppressed and cannot be unsuppressed with this screen.
Analysis begins when the ANALYZE button is pressed. Once the sample is accepted, the test
panel is frozen and cannot be unsuppressed during the analysis.
TEST PANELS
1
2
3
4
5
6
Full
Blood Gas
Blood Gas Plus
Pulmonary
Electrolyte
Metabolite
pH
PCO2
PO2
SO2%
Hct
Na+
K+
Ca++/Cl-
Lac
BUN
3-30
Glu
3 Operation
3.17.2 Three Ways of Running Samples, Based On Sample and Container Type
Samples can be analyzed from syringes, vacuum tubes, sample cups, capillaries, or expired air
bags as follows. A minimum of 225 L of sample is required for whole blood, serum, and
plasma samples.
1. Press ANALYZE. The sample probe extends and the CRT displays the Position
Sample for Aspiration screen.
2. Immerse the probe in the sample (syringe shown in Figure 3.6) and press ANALYZE
a second time to aspirate. Do not allow the Sample probe to touch the base of the syringe
plunger or the bottom of the sample cup or tube.
3-31
3. Operation
3. Operation
11:45:45
DWG #7-0-035A
PROBE
SYRINGE
3. When the probe retracts, withdraw the sample (if the sample is from a syringe, expel
the bubble of air from the syringe and cap it). The CRT displays the Sequence Start
screen.
3-32
3 Operation
NOTE: If no sample is detected or if a major flow error occurs during analysis, the system aborts the analysis and begins a flush cycle. If an insufficient
sample is detected, the system will finish the analysis and generate error
code 9C, Insufficient Sample.
PATIENT DATA
Analysis in Progress
nnn Seconds to Completion
Accession #
Patient I.D.
xxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxx
Temperature
xxxxx
C
Hemoglobin
xxxx
g/dL
Type
0-Arterial 1-Venous 2-Capillary
Time Drawn
xxxxxxxxx
xxxxx
%
FIO2:
e. Repeat Steps a through d as needed. Then press CLEAR to exit the screen
or PATIENT DATA to access the Measured Results Reporting screen as
in Step 5.
The Patient Data screen shows the state of the analysis (In Progress, Completed, Aborted, etc.)
and the time remaining to completion. If the analysis has finished when you exit Patient Data,
the current Patient Data values will be used to compute the Corrected and Calculated results,
and the Measured Results screen will be displayed. If the Auto Results Print mode is selected,
the results will be printed.
5. From the Patient Data screen, press TEST SELECT to access the Measured Results
Reporting screen. This screen allows you to change the setting on any of the measured
parameters or calculated results for a single patient test.
3-33
3. Operation
4. About 10 seconds after the Sequence Start screen is displayed, press PATIENT
DATA (if the screen was not set up to be automatically displayed) to access the
Patient Data screen. Enter the patient data, and, if necessary, change the Individual
and Calculated Results reporting as follows.
a. Press ENTER repeatedly to cycle through the choices to the desired choice.
b. Key in the appropriate value using the keypad.
c. Press ENTER to enter the value and also to move the cursor to the next
choice.
d. Press CLEAR to delete.
3. Operation
NOTE: Test Select and Set Up Menu restriction functions define which tests
can be changed.
c. Repeat Step 5 until all desired parameters are changed.
MEASURED RESULTS REPORTING
Analysis in Progress
nnn Seconds to Completion.
Accession #
xxxxxxxxxxxxxxxxxxx
pH
1
PCO2
2
PO2
3
Patient I.D.
xxxxxxxxxxxxxxxxxxx
SO2%
4
Hct
5
Na+
6
K+
7
Ca++/Cl-
8
Glu
9
Lac
.
BUN
0
Hb
1
BE-ECF
2
BE-B
3
Patient I.D.
xxxxxxxxxxxxxxxxxxx
SBC
4
HCO3
5
TCO2
6
O2Ct
7
A
8
AaDO2
9
a/A
.
P5O
0
PO2/FI
-
3-34
3 Operation
6. Press TEST SELECT to display the Gas Calculated Results Reporting screen.
7. Repeat Step 5 until all desired tests are changed.
8. Press TEST SELECT to display the Electrolyte Calculated Results Reporting screen.
9. Repeat Step 5 until all desired tests are changed.
10. Press CLEAR to exit this screen and the Patient Data screens.
11. Measured results are displayed upon sequence completion, and measured and
calculated results are printed out. See Figure 3.7.
nCa++
1
Osm
2
Patient I.D.
xxxxxxxxxxxxxxxxxxx
Osm
1
An.Gap
2
Patient I.D.
xxxxxxxxxxxxxxxxxxx
3-35
3. Operation
NOTE: Any results suppressed in the Set Up Menu or the Measured Results
Reporting screen will not be displayed or printed.
MEASURED RESULTS
nnnnnnnnnnnnnnnnnnnnn
Sample # nnnnn
iiiiiiiiiiiiiiiiiiiii
3. Operation
Corrected to 123.4
pH
7.369
PCO2
43.3
mmHg
PO2
100.1
mmHg
SO2%
98
Hct
48.0
%
123.4
mmol/L
Na+
K+
1.2
mmol/L
mmol/L
Ca++/Cl- 1.27
Glu 123.4
mg/dL
Lac
1.2
mmol/L
BUN
1.2
mmol/L
Press ENTER for Calculated Results.
Press CLEAR to Exit.
+12.3
+12.3
12.3
12.3
12.3
12.3
123.4
123.4
123.4
12.3
12.34
12.3
mmol/L
mmol/L
mmol/L
mmol/L
mmol/L
mL/L
mmol/L
mmol/L
mmol/L
mg/dL
mmol/L
mmol/L
3-36
3 Operation
nCa++
Osm
12.34
123.
mmol/L
mOsm
3. Operation
12.
13.
14.
3-37
3. Operation
Sample #
Accession #
Patient I.D.
Time of Analysis
Time Drawn
Sample Type
Patient Temperature
30 Apr 96
12:05
Arterial
37.0 C
pH
PCO2
PO2
7.373
39.1
mmHg
159.3
mmHg
12:15
Arterial
37.0 C
Measured to
Corrected to 37.0 C
pH
PCO2
PO2
3
021296
82.5901
30 Apr 96
37.0 C
7.357
42.6
mmHg
96.3
mmHg
Corrected to 37.0 C
SO2%
Hct
? 2.
Error
Na+
K+
Ca++
Glu
Lac
BUN
159.4
5.71
1.57
80.
1.0
10.
mmol/L
mmol/L
mmol/L
mg/dL
mg/dL
mg/dL
Calculated Results
BE-ECF
BE-B
SBC
HCO3TCO2
O2Ct
A
AaDO2
a/A
P50
Hb
nCa++
Osm
- 6.4
- 3.7
22.3
24.2
73.4
16.5
0.8
? 0.7
1.39
314.
mmol/L
mmol/L
mmol/L
mmol/L
mmol/L
mL/dL
mmHg
mmHg
7.357
42.6
mmHg
96.3
mmHg
SO2%
Hct
? 3.
Error
Na+
K+
Ca++
Glu
Lac
BUN
139.4
3.76
1.07
194.
2.8
27.
mmol/L
mmol/L
mmol/L
mg/dL
mg/dL
mg/dL
Calculated Results
mmHg
g/dL
mmol/L
mOsm/kg
ERRORS
pH
PCO2
PO2
BE-ECF
BE-B
SBC
HCO3TCO2
O2Ct
A
AaDO2
a/A
P50
Hb
nCa++
Osm
- 1.7
- 0.4
24.0
25.3
95.6
mmol/L
mmol/L
mmol/L
mmol/L
mmol/L
mL/dL
mmHg
mmHg
1.0
? 0.9
1.05
289.
mmHg
g/dL
mmol/L
mOsm/kg
ERRORS
64 Hct TOO Low
95 SO2% CALC. Hct
3-38
3 Operation
3.17.2.2 Running Capillary Samples
NOTE: Use a capillary tube of at least 225 microliters delivery capacity.
DWG #7-0-034A
CAUTION: In the following step, ensure that the capillary-adapter unit remains in line with the probe. Misalignment may cause poor aspiration.
4. Press ANALYZE. When the sample probe retracts, allow the adapter to slide off the
sample probe. The CRT displays the Analysis in Progress screen.
3-39
3. Operation
1. Remove the mixing flea then fit the Capillary Adapter to the tip of the capillary tube,
sliding the base of the adapter over one end of the capillary tube. See Figure 3.8.
2. Press ANALYZE. The sample probe extends and the CRT displays the Sample
Analysis screen.
3. Fit the adapter-capillary unit to the sample probe tip by allowing the capillary tube
and sample probe to meet flush. Hold the capillary in line with the sample probe.
Remove the end cap. Continue to hold the capillary for the next step.
3. Operation
1. From the Ready For Analysis screen, press 2 (or Menu, 2) to display the Maintenance
Menu.
2. Press 1 to choose Alternate Analysis (Expired Gas).
The Expired Gas Analysis screen appears.
The probe extends.
3. Slide the bag over the tip of the sample probe, maintain slight pressure on the bag,
and proceed to Step 4.
4. Press ANALYZE. When the probe retracts, allow the probe to slide off the bag. The
Analysis in Progress screen is displayed.
NOTE: Press ANALYZE within 45 seconds after choosing Expired Gas (from
Step 2).
CAUTION: Do not touch the analyzers compartment door during an analysis,
electrode instability could result.
5. Continue with Step 4 of the Running Syringe, Vacuum Tube, or Sample Cup
Sampling procedure (Section 3.17.2.1).
NOTE: The Expired Gas Measured and Calculated results will not include
electrolytes, hematocrit, pH, glucose, and the associated results from the
Calculated Results screen.
PATIENT RESULTS
Temp
37.0
PCO2
PO2
35.3
65.4
mmHg
mmHg
3-40
4 Maintenance
4
Maintenance
The following sections provide detailed information and directions for these areas:
WARNING: Blood samples and blood products are potential sources of hepatitis and
other infectious agents. Handle all blood products and flow path components (wasteline, septum assembly, probe, flow cell, etc.) with care. Gloves and protective
clothing are recommended.
4.1
From the Ready for Analysis screen, press MENU to display the Main Menu
screen.
Press 2 to display the Maintenance Menu screen. If the system is busy, access is not
allowed until the system is idle.
4-1
4. Maint.
A Maintenance Log (PN 10794) that includes performance records and a maintenance checklist
is supplied with your instrument. Use this log to record data for long-term performance
verification and to document maintenance.
1
2
3
4
5
6
7
4. Maint.
30 Apr 95
11:45:45
3.
See Sections 4.1.1 to 4.1.7 to display menu screens. After displaying a screen, press
ENTER (or ANALYZE for the Flow Cell Conditioning and Sample Preheater
Cleaning functions) to initiate the function. If, before pressing ENTER, you wish
to exit any maintenance function, press CLEAR to return to the Maintenance Menu
screen. Press CLEAR again to return to the Main Menu screen. Press CLEAR once
more to return to Ready For Analysis screen.
4-2
4 Maintenance
4.1.3 Calibration Options
Use this screen to perform full or individual calibrations. As a time saver, a 2-point calibration
of gas, pH/hematocrit, electrolytes, metabolites (glucose, lactate, BUN), and SO2 subsystems
can be done whenever you do not need calibration of all electrodes. From the Calibration
Options screen press the appropriate number for a subsystem calibration. A subsystem
calibration is commonly used when maintenance is performed on an electrode, when new
reagents or gases have been introduced to the system, for electrode performance verification,
or for troubleshooting.
1
2
3
4
5
6
Full Calibration
Gas Calibration
pH + Hct Calibration
Electrolyte Calibration
Metabolite Calibration
SO2% Calibration
Calibration Gases
30 Apr 95
11:45:45
4-3
4. Maint.
1.
2.
Full Calibration
4. Maint.
Start Sequence?
30 Apr 95
15:43:10
4-4
4 Maintenance
Terminate a calibration as follows:
1.
2.
If the calibration progress screen is not displayed, press CLEAR repeatedly until
it is displayed.
Press CLEAR, ENTER to terminate the calibration.
4.1.3.3 Cal I
From the Calibration Options screen, press 6 to display the SO2% Calibration
screen.
Follow the directions on the screen.
Press ANALYZE when ready to aspirate Standard #1.
Press ANALYZE when ready to aspirate Standard #2.
4-5
4. Maint.
4. Maint.
3.
4.
Press 4 from the Maintenance Menu to display the Fluid/Gas Prime Options screen.
Press the number of the desired option. Press the FUNCTION key and the number
if a calibration is also desired. This starts the sequence. To abort press CLEAR.
A progress screen indicates the time remaining to completion.
After completion, the Ready for Analysis screen is displayed.
Fluid Prime
Gas Prime
11:45:45
4-6
4 Maintenance
FLOW PATH MAINTENANCE
30 Apr 95
11:45:45
4. Maint.
CAUTION: Remove the glucose, BUN, and lactate electrodes before cleaning
with preheater cleaning agent. They may be damaged by direct contact with
the cleaning solution.
1.
2.
Open the door. Remove the glucose, BUN, and lactate electrodes from the flow cell.
4-7
Immerse the probe into the cleaning solution sample and press ANALYZE. After
the probe retracts, the time remaining to completion is displayed. To stop the cycle,
press CLEAR.
4. Maint.
Insert electrode blanks in the glucose, BUN, and lactate chambers and close the
door.
Press ENTER to acknowledge that the blanks are in place and to extend the probe.
After completion of the cycle, remove the electrode blanks and reinstall the
glucose, BUN, and lactate electrodes. Close the analytical compartment door.
Press ENTER.
Calibration in progress. (Only if FUNCTION key was pressed at the start of the procedure.)
Open the door. Remove the BUN electrode from the flow cell.
Insert an electrode blank into the BUN chamber and close the door.
Press ENTER to extend the probe.
4-8
Immerse the probe into the cleaning solution sample and press ANALYZE. After
the probe retracts, the time remaining to completion is displayed. To stop the cycle,
press CLEAR.
4 Maintenance
*
After completion of the cycle, remove the electrode blank and reinstall the BUN
electrode. Close the analytical compartment door. Press ENTER.
Calibration in progress. (Only if FUNCTION key was pressed at the start of the procedure.)
1.
2.
3.
4.
5.
From the Maintenance Menu, press 3, 7 to display the Calibration Gases screen.
Press ENTER repeatedly to cycle the cursor (appearing to the right of the values)
through the allowed values.
Key in the gas composition.
Press ENTER to replace the original value with the new default value. To cancel
an entry, press CLEAR instead of pressing ENTER.
After changed, press CLEAR to return to the Maintenance Menu, press CLEAR
again to return to the Main Menu, then press CLEAR again to return to the Ready
for Analysis screen.
CALIBRATION GASES
5.0 %
20.0 %
10.0 %
0.0 %
30 Apr 95
11:45:45
4-9
4. Maint.
CAUTION: Correct calibration gas values are critical for valid calibration
and analysis.
Scheduled Maintenance
It is important to perform preventive maintenance as scheduled. The Maintenance Log (PN
10794) gives suggested schedules based on sample volume. Space is provided for slopes, flow
time, and control results in the Maintenance Log.
NOTE: Keep the analyzer compartment door closed as much as possible. This
will keep the components inside the analyzer compartment at 37 C and
minimize the time needed to recover temperature.
4.3
Maintenance Procedures
4. Maint.
Perform these procedures per the suggested schedule in the Maintenance Log and as needed.
Check off procedure performed in the Maintenance Log.
30 Apr 95
11:45:45
4-10
4 Maintenance
4.3.1.1 Flow Cell Conditioning
The flow cell is conditioned with whole blood after periods of inactivity, with infrequent
sample analysis, after cleaning, or after electrode maintenance. The instrument aspirates blood
into the flow cell, holds it for 5 minutes, then flushes the flow cell. Condition the flow cell as
follows:
1.
2.
3.
4.
From the Maintenance Menu screen, press 5 to display the Component Conditioning Menu.
Press 2 to display the Na+ Conditioning screen. (Press FUNCTION with 2 to
calibrate automatically after the procedure is completed.)
Remove Electrode. Condition Na+ 2 min. Install electrode & Plug in.
3.
4.
Unplug the Na+ electrode from the rack. Unclip the Na+ electrode and remove it.
Fill the Electrode Conditioning Holder (PN 09458) 1/4-full of Na/pH Conditioning
Solution (PN 06856).
4-11
4. Maint.
5.
Insert the Na+ electrode into the holder, make sure that the glass tip is immersed in
the conditioning solution.
Remove the electrode from the solution and rinse the tip with deionized water. Dry
the electrode with a lint-free tissue. Be careful not to break the glass tip.
Rinse the electrode conditioning holder thoroughly with deionized water.
Install the Na+ electrode by sliding the electrode body into the flow cell until the
electrode clips into place.
Plug the Na+ electrode into the rack and press ENTER.
Calibration in progress. (Only if FUNCTION key was pressed at start of the procedure.)
4. Maint.
10.
DWG #5-0-054B
Whenever necessary, clean the septum assembly inlet port using a clean swab moistened with
deionized water. See Figure 4.1.
4-12
4 Maintenance
4.3.3 Gas Humidifier Well Filling
Fill the humidifier chambers with deionized water to the fill lines on the chamber. See Figure 4.2.
When replacing the caps, ensure that they are correctly seated and finger-tighten to prevent gas leakage.
Do not overtighten or gas will leak out of caps.
If the water becomes cloudy, withdraw the water with a syringe, clean the chamber walls with
a swab, then refill the chambers.
In unusual circumstances, bacteria may grow in the chamber. If so, you may clean the chamber
walls with a swab moistened in alcohol.
4. Maint.
DWG #5-0-033B
FILL TO HERE
4-13
Component Replacement
The electrodes, tubings, and reagent pack are replaced through the Component Replacement
screen. From the Maintenance Menu, press 7 to display this screen. Select the component to
be replaced and proceed.
COMPONENT REPLACEMENT
4. Maint.
30 Apr 95
11:45:45
4-14
Remove electrode & washer. Dry chamber. Condition, remembrane or replace electrode as required.
Replace washer, reinstall electrode & plug in. Close door.
4 Maintenance
2.
3.
4.
Guide Pin
Electrode Chamber
Na
Ca
BUN
Remove the washer by hooking it with the hook end of the Washer Removal Tool
(PN 07157) and, with a lint-free tissue or swab, dry the electrode chamber and
washer thoroughly. If bubble hang up at the electrode tip was a problem or if other
problems suggest a bad washer, replace the washer. The sodium, glucose, and
lactate chambers do not contain washers.
Washer
Type
Thin,
Large Hole
No Washer
DWG #7-F-005A
Thin,
Large Hole
Part
Number
Electrode
Type
PN 08153
K+ / Cl-
Na+
PN 08153
Ca++ / K+
W
R
No Washer
BUN
No Washer
Glu
No Washer
Lac
Thin,
Small Hole
PN09796
pH
Thick,
Small Hole
PN 09797
PCO2
Thick,
Large Hole
PN 09798
PO2
Thick,
Large Hole
PN 09798
SO2
Washer
Washer
Removal
Tool
Flow Cell
Figure 4.4 Flow Cell Washers and Insertion
4-15
4. Maint.
5.
7.
Position the correct electrode washer (from Kit PN 08153 for K+, Ca++, or Cl-) on
the broad end of the washer removal tool, see Figure 4.4, and insert it into the flow
cell, seating it against the back wall of the chamber.
Close the door.
NOTE: In the following steps, keep the door closed as much as possible to
minimize time needed to recover operating temperature.
8.
Precondition the specific electrode or, for the chloride electrode, replace the cap as
follows.
b.
c.
Chloride
a.
b.
Unscrew the old chloride cap from the chloride body and dispose of the cap.
Check the washer on the body of the electrode for damage and replace with
the spare washer from the chloride cap kit if necessary.
Fill a new Chloride Cap (PN 08129) to the top edge with Chloride Internal
Filling Solution (PN 00099). See Figure 4.5.
1
Electrode
Body
Chloride
Cap
DWG #7-M-006A
Chloride
Internal
Filling
Solution
4-16
4 Maintenance
c.
d.
Insert the Electrode Body (PN 06005) straight down into the filled cap and
screw the cap onto the electrode body.
Dry the electrode with a lint-free tissue, take care to avoid touching the tip.
Check that there are no bubbles in the electrode. If there are bubbles,
unscrew the cap, add more filling solution, and repeat Steps c and d.
Insert the electrode into the flow cell chamber by sliding the electrode body
into the flow cell until the electrode clips into place.
Plug the cable into the rack.
e.
f.
Sodium
a.
b.
c.
d.
e.
f.
9.
*
Press ENTER.
NOTE: All electrodes must be in the flow cell during a Flow Cell Conditioning cycle.
a.
b.
c.
*
Calibration in progress.
11.
From the Maintenance Menu, press 6 to display the Flow Path Maintenance screen.
Then select 3, Flow Path Maintenance.
Open the door and locate the pH electrode.
Unplug the electrode.
Unclip the electrode and remove it from the flow cell. See Figure 4.6.
4-17
4. Maint.
Fill the Electrode Conditioning Holder (PN 09458) 1/4 full with Na/pH
Conditioning Solution (PN 06856).
Insert the electrode into the holder, make sure that the glass tip is fully
immersed in the conditioning solution. Wait 2 minutes.
Remove the electrode from the solution and rinse the tip with deionized water.
Dry the electrode with a lint-free tissue. Be careful not to break the glass tip.
Rinse the holder thoroughly with deionized water.
Install the Na+ electrode by sliding the guide pin and electrode body into the
flow cell until the electrode clips into place.
Plug the electrode cable into the electrode rack.
K
BUN
Glu
pH
Lac
DWG #7-J-003A
PCO2
Guide Pin
Electrode Chamber
PO 2
SO2
4. Maint.
5.
Remove the washer by hooking it with the hook end of the Washer Removal Tool
(PN 07157) and, with a lint-free tissue or swab, dry the electrode chamber and
washer thoroughly. If bubble hang up at the electrode tip was a problem or if other
problems suggest a bad washer, replace the washer.
Washer
Type
Thin,
Large Hole
No Washer
DWG #7-F-005A
Thin,
Large Hole
Part
Number
Electrode
Type
PN 08153
K+ / Cl-
Na+
PN 08153
Ca++ / K+
No Washer
BUN
No Washer
Glu
No Washer
Lac
Thin,
Small Hole
PN09796
pH
Thick,
Small Hole
PN 09797
PCO2
Thick,
Large Hole
PN 09798
PO2
Thick,
Large Hole
PN 09798
SO2
Flow Cell
Figure 4.7 Flow Cell Washers and Insertion
4-18
W
R
Washer
Washer
Removal
Tool
4 Maintenance
6.
7.
Position the electrode washer from Kit PN 09796 for pH on the broad end of the
washer removal tool, see Figure 4.7, and insert it into the flow cell, seating it against
the back wall of the chamber.
Close the door.
NOTE: In the following steps, keep the door closed as much as possible to
minimize time needed to recover operating temperature.
8.
9.
Press CLEAR, ENTER to terminate the Flow Path Maintenance sequence.
10. From the Maintenance Menu, press 5, Component Conditioning Menu. Then press
1, Condition Flow Cell.
NOTE: All electrodes must be in the flow cell during a Flow Cell Conditioning cycle.
a.
b.
c.
11. From the maintenance Menu, press 3, Calibration Options. Then press 3, pH
Calibration.
4. Maint.
For a pH electrode which has been in use, condition with Na/pH Conditioning
Solution as follows:
a.
Fill the Electrode Conditioning Holder (PN 09458) 1/4 full with Na/pH
Conditioning Solution (PN 06856).
b.
Place the Electrode (PN 06013) into the cup so that the tip of the electrode
is immersed in the Na/pH Conditioning Solution.
c.
Wait 15 minutes.
d.
Remove the electrode from the cup and rinse with deionized water.
e.
Dry the electrode with a lint-free tissue. Be careful not to break the glass tip.
f.
Insert the electrode into the flow cell by sliding the guide pin and electrode
body into the flow cell until the electrode clips into place.
g.
Plug the cable into the rack.
Remove electrode & dry chamber. Remembrane or replace electrode as required. Install electrode.
Close door.
4. Maint.
2.
3.
4.
Glu
c.
Gently polish the electrode tip on the paper, move the tip in a circular
fashion for about 10 seconds. See Figure 4.8.
5.
Soak the electrode in Electrode Soaking Solution (PN 13409) for 2 minutes.
6.
Remove the electrode and rinse the electrode with deionized water. Then blot dry
the body of the electrode with lint-free tissue. Do not wipe the tip of the electrode.
7.
Gently push a new enzyme-membrane cap onto the electrode body without twisting
the cap (Figure 4.9).
8.
Dry the flow cell chamber with a lint-free tissue or swab.
9.
Insert the electrode assembly into the glucose flow cell; plug the cable into the rack.
10. Close the door and press ENTER.
*
4-20
4 Maintenance
*
Calibration in progress.
DW
Glu
G
#5
0
-0
-M
3B
4. Maint.
Remove electrode & dry chamber. Remembrane or replace electrode as required. Install electrode.
Close door.
2.
3.
4-21
Lac
DWG
#5-M
-004
4. Maint.
5.
6.
7.
8.
9.
Soak the electrode in Nova Cleaning Solution (PN 11272) for a minimum of
5 minutes. Remove the electrode and rinse the tip thoroughly with deionized water.
Then blot dry with a lint-free tissue.
Gently push a new enzyme-membrane cap onto the electrode body. See Figure
4.11.
Dry the flow cell chamber with a lint-free tissue or swab.
Insert the electrode into the flow cell, plug the cable into the rack.
Close the door and press ENTER.
Calibration in progress.
DW
La
G
B
05
-0
#5
4-22
4 Maintenance
4.4.5 BUN Electrode Membraning and Replacement
The following are more detailed instructions than displayed on the screen.
Remove electrode & dry chamber. Remembrane or replace electrode as required. Install electrode.
Close door.
2.
3.
4.
5.
6.
DW
G#
5-1
0-0
05B
BUN
7.
8.
*
Insert the electrode into the flow cell, plug the cable into the rack.
Close the door and press ENTER.
NOTE: This is not necessary if only the BUN membrane was replaced. To
eliminate, press CLEAR.
*
Calibration in progress.
4-23
4. Maint.
4. Maint.
1.
Remove electrode & washer. Dry chamber. Polish if PO2. Remembrane or replace electrode as
required. Replace washer, reinstall electrode. Close door.
2.
3.
4.
5.
6.
4-24
From the Maintenance Menu, press 7 to display the Component Replacement Menu.
Press 1 to display the Electrode Replacement screen. Then press 1 to display the
PCO2/PO2 Electrode Change Procedure screen.
Thin,
Large Hole
PN 08153
Ca++
Thin,
Large Hole
PN 08153
Mg++
No Washer
Glu
No Washer
Lac
Thin,
Small Hole
PN09796
pH
Thick,
Small Hole
PN 09797
PCO2
Thick,
Large Hole
PN 09798
PO2
Thick,
Large Hole
PN 09798
SO2
W
R
Washer
Washer
Removal
Tool
4. Maint.
DWG #7-11-002D
4 Maintenance
Flow Cell
Figure 4.13 Flow Cell Washers and Insertion
7.
Remove the PCO2 membrane cap unit from the electrode body by carefully pulling
it straight off (see Figure 4.14).
O2
DW
pC
-M
#5
06
-0
B
9.
4-25
4. Maint.
NOTE: When assembling the electrode in the following steps, touch the
membrane cap on the edges only; avoid contact with the measuring surface
of the membrane cap.
11.
Fill a new membrane cap so that the bottom of the meniscus of the internal filling
solution (IFS) is at the fill line (just above the vent hole). See Figure 4.15.
DWG #5
-M-007A
Fill
to here
CAUTION: In the following step, do not twist the electrode body or membrane
cap when assembling the electrode; slide the cap straight on in 1 quick
motion to avoid damaging the membrane.
12.
4-26
Align the groove on the membrane cap with the PCO2 label on the electrode (this
causes the vent hole to be on top when installed into the flow cell). Insert the
electrode body into the membrane cap and press down firmly to ensure a good seal.
See Figure 4.16. A small amount of filling solution may squirt out the vent hole.
If this happens, dry the outside of the cap thoroughly with a lint-free tissue or swab.
Carefully dry the measuring surface of the membrane with lint-free tissue.
You can extend the life of the PCO2 membrane by minimizing the evaporation of
IFS from the vent hole in the cap. Cover the vent hole of the cap with the adhesive
strip that Nova provides. Use a new adhesive strip after each membrane replacement.
When the electrode is held horizontally, a bubble extending almost the length of the
cap will appear. This bubble is necessary to allow for thermal expansion of the
filling solution after the electrode is inserted into the flow cell.
pCO 2
4 Maintenance
Firmly Insert
Electrode.
Dry Thoroughly.
Air Bubble
Vent Hole
pCO 2
pCO2
DWG #5-M-009B
13.
14.
15.
16.
4-27
4. Maint.
Mg
Glu
Lac
pH
PCO2
PCO2
PO2
Guide Pin
Electrode Chamber
4. Maint.
DWG #7-11-006C
SO2
NOTE: All electrodes must be in the flow cell during a Flow Cell Conditioning cycle.
a.
b.
c.
*
20.
4-28
4 Maintenance
c.
The following procedure explains how to membrane the electrode and how to polish the PO2
electrode.
1.
Remove electrode & washer. Dry chamber. Polish if PO2. Remembrane or replace electrode as
required. Replace washer, reinstall electrode. Close door.
2.
3.
4.
5.
6.
4-29
4. Maint.
NOTE: The screens are in a layered format. To see the entire format, pull
out the flow chart in Appendix E, 2.0 Maintenance.
4. Maint.
DWG #7-11-002D
PN 08153
Ca++
Thin,
Large Hole
PN 08153
Mg++
No Washer
Glu
No Washer
Lac
Thin,
Small Hole
PN09796
pH
Thick,
Small Hole
PN 09797
PCO2
Thick,
Large Hole
PN 09798
PO2
Thick,
Large Hole
PN 09798
SO2
Flow Cell
7.
Washer
Washer
Removal
Tool
Unscrew the used PO2 cap from the PO2 body and dispose of it. Check the washer
on the body of the electrode for damage and replace it with the spare washer from
the PO2 premembraned cap kit if necessary.
Polish the electrode as follows:
a.
Place a drop of PO2 Electrolyte Solution (PN 06554) on the polishing paper
(from kit PN 11099).
b.
Hold the PO2 polishing paper so that the tip of your index finger provides
light pressure against the back of the paper.
c.
Gently polish the electrode tip on the paper, move the tip in a circular
fashion for about 10 seconds. See Figure 4.19.
pO 2
8.
W
R
DWG
#5-M
-010
4-30
4 Maintenance
d.
Wipe the electrode tip with a lint-free tissue soaked in PO2 electrolyte
solution. See Figure 4.20.
pO 2
DWG #5-M-011B
Lint-free Tissue
Soaked in PO2 Solution
Fill a new PO2 Premembraned Cap (PN 11099) only to the bottom of the threads
with PO2 Internal Filling Solution (PN 06554). Tap it gently to dislodge any air
bubbles on the membrane. See Figure 4.21.
DWG #5
-M-008A
Fill
to here
Insert the Electrode Body (PN 11098) straight down into the filled cap and screw
the cap onto the electrode body. See Figure 4.22.
pO 2
DWG #5-M-012B
10.
11.
Dry the electrode with a lint-free tissue. Take care not to touch the tip.
4-31
4. Maint.
9.
Insert the electrode into the flow cell by sliding the electrode body into the flow cell
until the electrode clips into place. See Figure 4.23.
Mg
Glu
Lac
pH
PO2
PCO2
PO2
SO2
Guide Pin
4. Maint.
DWG #7-11-001C
Electrode Chamber
Figure 4.23
13.
14.
*
NOTE: All electrodes must be in the flow cell during a Flow Cell Conditioning cycle.
a.
b.
c.
*
4-32
If the electrode does not calibrate due to slope errors, remove air bubbles as follows:
a.
Open the door, remove the electrode, unplug the cable, and shake down the
electrode with a wrist-snapping motion to move air bubbles to the back of
the electrode.
b.
Reinsert the electrode into the flow cell, plug the electrode cable into rack,
and close the door.
c.
From the Maintenance Menu, press 3, 2 to start a gas calibration.
4 Maintenance
4.4.8 SO2 Maintenance (Periodically - every 3 months)
1.
2.
3.
4.
5.
6.
From the Maintenance Menu, press 7 to display the Component Replacement Menu.
Press 1 to display the Component Replacement screen. Then press 4 to display the
Reference Electrode Change Procedure screen. Wait for gas flow to stop.
Disconnect W/R lines from Ref electrode. Rotate latch above electrode. Replace electrode. Reposition
latch. Attach tubing.
2.
3.
4.
5.
6.
7.
4-33
4. Maint.
7.
8.
From the Maintenance Menu, press 6 to display the Flow Path Maintenance screen.
Then select 3, Flow Path Maintenance.
Open the door and locate the SO2 cable.
To remove the cable, push the knurled ring in and turn it toward you until the cable
can be guided out.
Check the washer (PN 09798). Replace the washer if it looks deformed, cracked,
or worn.
Clean the cable surface with a lint-free tissue that is soaked with bleach. Rinse with
deionized water and blot dry.
Reinstall the washer. With the slot at the cable end toward you, reinstall the fiber
optic cable and turn the knurled ring away from you.
Press CLEAR, ENTER to terminate the Flow Path Maintenance sequence.
From the maintenance Menu, press 3. Then select 6, SO2 Calibration.
DWG #7-0-013A
4. Maint.
8.
9.
10.
*
Close door.
11.
4-34
Loosen the retaining block thumbscrew, center the retaining block in the depression on top of the electrode, and reset the retainer.
Plug the cable into the rack.
Attach the W/R-lines to the reference electrode and press ENTER.
Press ENTER.
Calibration in progress.
4 Maintenance
4.4.10 Replace Pump Tubing
The following are more detailed instructions than displayed on the screen.
The tubing segments that go around the pump can be replaced without replacing the complete
W/R-harness. The following procedure is for replacing these segments (PN 07575). To replace
the complete W/R-harness, see Section 4.4.11.
1.
1
2
3
4
5
6
Replace
Replace
Replace
Replace
Replace
Replace
Pump Tubing
Reagent/Septum Harness
W/R Harness
S line / probe
Flow Cell
Preheater
30 Apr 95
11:45:45
Remove tubing from bypass valve. Disconnect R line from left pump fitting & W line from T connector.
2.
3.
4. Maint.
4. Maint.
DWG #5-0-041B
Figure 4.25 Removing the W-Bypass Tubing From the Pump Bypass Valve
4.
Remove the R-tubing segment from the tubing support block as follows (Figure
4.26):
a.
Gently stretch the R-tubing and slide it out of the horizontal slot.
b.
Remove the R-tubing from the vertical slot.
DWG #5-0-035B
Figure 4.26 Removing the R-Tubing Segment From the Tubing Support Block
5.
4-36
4 Maintenance
6.
Disconnect the R-tubing segment from the R-line at the second tubing collar. See
Figure 4.27.
R TUBING SEGMENT
4. Maint.
DWG #5-0-037B
R LINE
Remove the W-tubing from the tubing support block. See Figure 4.28.
a. Gently stretch the W-tubing and slide it out of the horizontal slot.
b. Remove the W-tubing from the vertical slot.
R
W
DWG #5-0-042C
7.
Figure 4.28 Removing the W-Tubing Segment From the Tubing Support Block
4-37
Disconnect the W-tubing segment from the W-line at the second tubing collar. See
Figure 4.29. Press ENTER.
4. Maint.
DWG #5-M-014B
W
R
DWG #7-0-017A
11.
*
4-38
Pull W/R tubes and seals from rear of analytical chamber. Remove tubing from pump & discard.
4 Maintenance
12.
Loosen the white plugs and pull the W/R-tubing out of the back of the analyzer
compartment. See Figure 4.31.
DWG #5-0-043B
4. Maint.
13.
Install new W pump tubing with seal at LEFT. Place section between T connectors in bypass valve,
Feed W tubing into analytical compartment & seat seal. Attach T connector to labeled W line.
14.
Press ENTER.
Install R line with seal at RIGHT. Feed R tubing into compartment & seat seal. Connect labeled R line
into pump fitting. Open door, attach tubing to reference. Close door.
15.
16.
R
W
DWG #5-0-060B
Figure 4.32 Inserting the New Tubing Segments into the Back of the Analyzer
4-39
Open the door and attach the W and R-tubes to the reference electrode. See Figure 4.33.
4. Maint.
DWG #7-0-016A
18.
19.
DWG #5-0-038C
CAUTION: Do not overstretch the tubing segments in the steps that follow.
Overstretched tubing segments will give inadequate suction and may require replacement of the segment.
4-40
4 Maintenance
If you look carefully at the tubing segments you will notice a seam running
lengthwise. Make sure that the seam of the tubing is not twisted. It is important to keep the orientation of this line straight to avoid premature failure of
the segment.
20.
c
b
R
W
4. Maint.
DWG #5-0-040B
b.
c.
Attach the exposed black fitting of the R-tubing segment to the R-line. See Figure 4.36.
R TUBING SEGMENT
R
W
DWG #5-M-015B
21.
Slide the W-segment connector nearest the front of the analyzer into the
front, vertical slot of the support block.
Without overstretching the W-tubing, pull it around the pump and slide the
second connector into the bottom horizontal slot.
Position both tubing connectors (oblong shaped) so that the widest portion
is horizontal and the R-line connectors can fit above them.
R LINE
4-41
Without overstretching the R-segment, pull it around the pump and insert the small
connector into the top horizontal slot of the support block. See Figure 4.37.
NOTE: Make sure the R-tubing is installed in the correct orientation. Incorrect installation of the R-tubing may lead to contamination of the reference
solution.
a
b
R
DWG #5-0-039B
4. Maint.
23.
*
Press ENTER
Press CLEAR, CLEAR, CLEAR to return to the Ready for Analysis screen.
4-42
4 Maintenance
To replace the tailpiece, first, disconnect the tailpiece from the front section of the
waste pump tube (Figure 4.39). Next, disconnect the rear section of the waste pump
tube and remove the tubing segment from the bypass valve. Finally, disconnect the
tailpiece from the reference electrode W-port (Figure 4.40). Discard the tailpiece.
R
W
4. Maint.
DWG #5-0-015B
DWG #7-0-014A
1.
Figure 4.40 Disconnecting the Tailpiece from the Reference Electrode W-Port
4-43
Connect the new tailpiece to the front segment of the W-pump tube (Figure 4.41).
Next, connect the rear segment of the waste pump tube and install the bypass tubing
section into the bypass valve. Finally, connect the tailpiece segment to the W-port
of the reference electrode (Figure 4.42).
DWG #5-0-018B
4. Maint.
DWG #7-0-015A
4-44
4 Maintenance
4.4.11 W/R-Harness Replacement
The following are more detailed instructions than displayed on the screen.
Replace the complete W/R-harness (including tubing segments) as follows. To replace the
tubing segments only, refer to Section 4.4.10.
1.
Press ENTER.
Remove 3 W/R tubing clips. Disconnect R line from pump fitting, W line from T connector. Discard tubing.
3.
4.
5.
6.
7.
8.
9.
10.
*
Open the top cover. Remove the W-bypass tubing segment from the pump bypass
valve.
Open the door and disconnect the W/R-tubings from the reference electrode. Close
the door.
Loosen the white plugs and remove the tubing from the back of the analytical
compartment.
Stretch outward and lift the W/R-tubing collars from the horizontal slots in the
tubing support block. Remove the remaining tubing collars from the tubing support
block. The tubing segments should now be free of the pump.
Remove the W/R-tubing harness manifolds from the mounting pins.
Remove the W-line from the reagent pack waste bottle.
The old W/R-tubing harness is now free from the instrument and can be discarded.
Press ENTER.
Install new harness. Connect 3 tubing clips. Connect W line to T connector, R line to pump fitting. Insert
R/W straws in bottles.
4-45
4. Maint.
DWG #5-0-045C
4. Maint.
11.
12.
Insert the W-bypass tubing segment into the pump bypass valve.
NOTE: Be sure the W-bypass tubing segment is seated in the groove which is
at the bottom of the slot.
13.
Position the new R-pump segment around the upper portion of the pump assembly.
Stretch the tubing slightly so that the larger collar fits into the front retainer slot and
the smaller collar fits into the rear retainer slot.
NOTE: Make sure the R-tubing is installed in the correct orientation (see
Figure 4.43). Incorrect installation of the R-tubing may lead to contamination of the reference solution.
14.
15.
16.
4-46
Press the 3 small W/R-tubing manifold bars onto their pins, make sure that the
tubing feeds around smoothly and is not twisted.
Insert the W-line into the waste container opening and the R-straw into the R-bottle
opening, make sure the reagent line fittings seat snugly onto the reagent bottles.
Thread the other ends of the tubing harness through the appropriate openings into
the analyzers compartment and secure the white plugs. See Figure 4.44.
4 Maintenance
DWG #5-0-060B
Open the analyzers door. Connect the W/R-lines to the W and R-ports on the
reference electrode. Close the door. Press ENTER.
Press CLEAR, CLEAR, CLEAR to return to the Ready (Not Ready) for Analysis
screen. Close the cover.
4-47
4. Maint.
17.
Purging
Open pinch valve bar, remove tubing. Detach labeled tubing clips from pins.
5.
6.
7.
4. Maint.
Press ENTER.
Open the valve pinch bar and pull the tubing out from the valve tubing clips.
Remove the 5-tube manifold bar, located in front of the valve, by lifting it off the
mounting pins.
Press ENTER.
Open Door. Disconnect reagent tubing manifold from top of reagent preheater. Pull harness through
rear of analytical chamber. Discard tubing.
8.
Open the analyzers door and locate the manifold connector mounted on top of the
reagent preheater. See Figure 4.45. Lift the connector release lever to remove the
connector.
REAGENT PREHEATER
SOCKET
DWG #5-M-016A
CONNECTOR
RELEASE LEVER
HARNESS
COLLAR
MANIFOLD
CONNECTOR
9.
10.
*
Disconnect septum harness from bottom of reagent preheater. Turn septum 1/4 turn clockwise and
remove with tubing. Unplug manifold connector & discard tubing.
11.
12.
13.
4-48
Locate the rubber harness collar in the back wall of the analytical compartment.
Pull the rubber harness collar out of the analyzers back wall and pull the tubing
harness out through the slot. Press ENTER.
Remove the septum harness manifold connector on the bottom of the preheater.
Disconnect the tubing manifold from the septum assembly.
Press ENTER.
4 Maintenance
*
Install new reagent harness through rear of analytical compartment & connect manifold to reagent
preheater. Reinstall tubing seal in opening.
14.
Feed the 5-tube manifold into the analyzers compartment through the slot in the
back of the compartment. Firmly seat the rubber harness collar into the analyzers
compartment slot.
Position the C, D, A, and B-tubings on the pinch valve assembly and close the pinch
bar. Pull out on the knob and release the knob to lock the bar in place. Ensure that
the tubing is correctly positioned on the pinch valve assembly.
PINCH BAR
-023A
DWG #5-0
MANIFOLD BAR
16.
17.
*
Press the 5-tube manifold onto the mounting pins that are located between the
reagent pack and the pinch valve. Press the 5-tube manifold bar onto the mounting
pins located between the pinch valve and the analyzers compartment.
Press ENTER.
Connect notched end of septum harness to bottom of reagent preheater, other end to septum. Install
septum & tubing onto probe turning counterclockwise so tubing forms downward loop.
4-49
4. Maint.
PINCH VALVE
DWG #5-0-019D
REAGENT
PREHEATER
MANIFOLD
CONNECTOR
WITH NOTCH
KEYED
MANIFOLD
CONNECTOR
SEPTUM
HARNESS
SEPTUM
HARNESS
4. Maint.
18.
19.
20.
*
Wipe reagent straws with alcohol, dry and insert into proper bottle.
21.
22.
23.
The septum harness part number is PN 06517. Plug the keyed manifold connector
of the new harness into the septum assembly fitting.
Loop the septum harness below the sampler assembly and connect the other end to
the bottom of the reagent preheater. The manifold connector is notched with a
projection to allow only the correct fit into the reagent preheater.
Close door and press ENTER.
Wipe reagent straws with alcohol, dry, and insert the F, C, D, A, and B reagent
straws into the appropriate openings in the reagent pack. See Figure 4.50.
Press down on the reagent line bottle fittings to ensure a snug fit onto the reagent
bottles.
Close the cover and press ENTER.
4-50
4 Maintenance
1.
14678
4206
MAY 95
15%
CALIBRATION STANDARD
pH
7.384
Na+ 140.0
K+
4.0
Ca++/Cl- 1.0
Glu
80.0
Lac
2.0
BUN
10.0
CONCENTRATIONS
6.840
75.0
10.0
2.0
200.0
6.0
50.0
4. Maint.
4.
5.
6.
7.
8.
9.
Then press ENTER to display the Reagent Pack Change Procedure screen.
Open the top cover.
Remove the reagent straws from the reagent pack. Dry the outside of each straw as
it is pulled from the pack.
Remove the RMS pod. Remove and discard the used reagent pack.
Remove the septum assembly by giving the assembly a 1/4-turn clockwise and by
pulling it out.
Disconnect the septum harness manifold connector from the septum assembly.
Plug the manifold connector into the new septum assembly.
Install the septum assembly by sliding the assembly over the end of the probe. See
Figure 4.48. The septum harness manifold connector should be to the lower left.
Turn the assembly 1/4-turn counterclockwise until snug. The septum harness
manifold should be at the lower right of the assembly (see Figure 4.53).
DWG #7-0-019A
2.
3.
4-51
Shake the new Reagent Pack and position the pack in the reagent pack bay so that
the RMS pod is in the back.
Plug the RMS pod into the socket located in the upper rear of the Reagent Pack Bay.
See Figure 4.49.
REAGENT MANAGEMENT
POD
47B
#5-0-0
DWG
SOCKET
4. Maint.
ATTACHED TO
REAGENT PACK
12.
13.
C
F
B
D
DWG #7-11-008C
14.
15.
NOTE: If the message, Check Septum Assembly, appears on the screen, see
Section 5.2.2, Troubleshooting.
4-52
4 Maintenance
4.4.14 Sample Probe/S-line Replacement
The following are more detailed instructions than displayed on the screen.
Disconnect the probe/S-line from the sample preheater by pulling the polyurethane tubing segment (on the end of the S-line) off of the sample preheater. See
Figure 4.51.
DWG #7-0-018A
S-Line
Polyurethane Tubing
7.
Press ENTER.
4-53
4. Maint.
Using hex key, loosen probe socket screw. Remove probe by pulling up and out.
8.
Use the hex key to loosen the socket screw about 2-1/2 turns at the top of the probe. Then
grasp the probe mounting block and remove the probe/S-line unit. See Figure 4.52.
S-LINE
SOCKET SCREW
HEX KEY
4. Maint.
ADJUSTMENT
PLATE
SEPTUM
ASSEMBLY
OPENING
DWG #5-0-064A
MOUNTING
BLOCK
SAMPLE
PROBE
9.
*
Press ENTER.
Insert tip of probe through opening in sampler and position mounting block behind hex screw. Verify
block is aligned with back plate & not angled. Tighten hex screw.
10.
Replace the Probe/S-line by first sliding the probe tip through the septum assembly
opening all the way to the mounting block. Then move the block back and around
the socket screw to hook it in place. Finally, use the hex key to tighten the socket
screw. Keep the probe aligned while you tighten it to the adjustment plate.
NOTE: Store the hex key inside the analyzers compartment under the sampler for future use.
11.
*
Install septum & tubing onto probe, turning counterclockwise so tubing forms downward loop.
12.
13.
*
4-54
Press ENTER.
To replace the septum assembly (with attached septum harness), slide the assembly
over the end of the probe so that the rectangular piece on the end fits into the septum
assembly opening in the baseplate of the sampler assembly. The septum harness
manifold connector should be to the lower left. Turn the assembly 1/4-turn
counterclockwise until snug. The septum harness manifold connector should now
be at the lower right of the assembly. See Figure 4.53.
Press ENTER.
Verify probe alignment using gauge. Refer to reference manual for details.
4 Maintenance
14.
DWG #5-0-026B
4. Maint.
15.
*
Press ENTER.
Connect the probe/S-line to the connector on the bottom of the sample preheater.
Close door.
17.
18.
19.
Check for leaks around the S-line connection. If a leak occurs, reseat the S-line on
the connector.
Reattach the guard onto the sampler.
Close the analyzers door.
4-55
1.
Open door. Remove W/R lines from Ref electrode. Rotate latch above electrode. Unplug & remove
all electrodes. Remove SO2 cable. Loosen flowcell retaining screws by turning counterclockwise &
remove flowcell. Do not lose interconnect tubing on either end.
2.
3.
4.
5.
6.
7.
8.
4-56
4 Maintenance
Retaining
Block
W
Reference
Electrode
Flow Cell
for Models
F, G, H
Electrode
Rack
Assembly
Cl
Na
K
BUN
Glu
Top Slotted
Retainer
Fastener
PCO2
PO2
SO2
Retainer
Screw
Flow Cell For Models
I, J, K
K
Electrical
Connector
Na
Ca
BUN
Glu
Lac
pH
SO2
Sample
Preheater
DWG #7-F-001A
PCO2
PO2
9.
10.
*
If the Flow Cell (PN 15857) is being replaced, remove the sodium electrode and
insert it into the new flow cell. Repeat this procedure for the other electrodes and
washers, avoiding any mix-up by replacing each washer and electrode immediately
upon removal.
Check the new flow cell for intact interconnect tubings. Press ENTER.
Install new flowcell seating interconnect tubing into opening at top of sample preheater. Tighten
retaining screws by turning clockwise.
4-57
4. Maint.
Lac
pH
12.
*
Attach reference electrode, positioning interconnect tubing carefully. Reposition latch. Reattach W/
R tubing.
4. Maint.
13.
14.
15.
*
Reattach the flow cell to the sample preheater back plate as follows:
See Figure 4.54.
a.
Set the flow cell into the recess on the sample preheater, align the right
surface of the flow cell flush with the right side of the back plate. Ensure
that the flow cell interconnect tubing is seated properly on the sample
preheater top connector.
b.
Turn the top retainer screw 1/4-turn clockwise.
Plug all cables into the rack. Press ENTER.
Replace the reference electrode on top of the flow cell by placing the electrode on
top of the flow cell, align the electrode sides with the backplate sides. Insure that
the reference electrode connector is seated properly on the flow cell interconnect
tubing.
Center the retaining block in the depression on top of the electrode.
Attach W/R lines to the reference electrode. Press ENTER.
NOTE: All electrodes must be in the flow cell during a Flow Cell Conditioning cycle.
a.
b.
c.
*
4-58
Calibration in progress.
4 Maintenance
4.4.16 Preheater Replacement
The following are more detailed instructions than displayed on the screen.
Open door. Remove W/R lines from Ref electrode. Rotate latch above electrode. Unplug all
electrodes. Remove SO2 cable. Loosen flowcell retaining screw(s) by turning counterclockwise.
Remove flowcell. Do not lose tubing on either end.
2.
3.
4.
5.
6.
8.
9.
10.
11.
*
Press ENTER.
Reinstall flowcell seating interconnect tubing into opening at top of sample preheater. Tighten
retaining screw(s) by turning clockwise.
4-59
4. Maint.
1.
14.
*
Reattach the flow cell to the sample preheater back plate as follows:
See Figure 4.54.
a.
Set the flow cell into the recess on the sample preheater, align the right
surface of the flow cell flush with the right side of the back plate. Ensure
that the flow cell interconnect tubing is seated properly on the sample
preheater top connector.
b.
Turn the top retainer screw 1/4-turn clockwise.
Press ENTER.
Attach reference electrode, positioning interconnect tubing carefully. Reposition latch. Reattach
W/R tubing. Plug in all electrodes. Reinstall SO2 cable. Close door.
Replace the reference electrode on top of the flow cell by placing the electrode on
top of the flow cell, align the electrode sides with the backplate sides. Ensure that
the reference electrode connector is seated properly on the flow cell interconnect
tubing.
16. Center the retaining block in the depression on top of the electrode.
17. Replace the preheater-flow cell-reference electrode assembly by aligning the guide
pin and by pressing the assembly onto the electrical connectors. Then turn the
bottom retainer screw 1/4-turn clockwise to engage the electrode rack assembly.
18. Plug all cables into the rack.
19. Plug the SO2 cable into the flow cell.
20. Reattach the S-line to the sample preheater.
21. Attach the W/R-lines to the reference electrode. Press ENTER.
4. Maint.
15.
NOTE: All electrodes must be in the flow cell during a Flow Cell Conditioning cycle.
a.
b.
c.
*
Calibration in progress.
4-60
4 Maintenance
5A
-05
3-0
G#
DW
UP
Open the printer platen. Gently move the lever to its opposing position and remove
the old roll of paper.
641
#10
log
Cata
per
a
P
ent
em
lac
Rep
1A
-05
3-0
G#
DW
Remove the paper holder from the old roll of paper. Discard the used roll.
Insert the paper holder into a new roll of paper (PN 10641). The loose end of the
paper should feed from the bottom of the roll.
DWG #3-0-016
3.
4.
4-61
4. Maint.
2.
UP
UP
5.
6.
641
#10
log
ata
rC
ape
nt P
e
em
lac
Rep
641
#10
log
ata
er C
p
a
P
ent
em
lac
Rep
3A
-05
3-0
G#
52
-0-0
DW
3
G#
4. Maint.
DW
7.
8.
9.
10.
UP
11.
Center the paper and close the printer platen by moving the lever back to the original
position. See Figure 4.59.
Press the PAPER ADVANCE button on top of the analyzer. Tear off the paper.
Position the printer cover on top of the printer so that the cover's front rim touches
the front of the PAPER ADVANCE button.
Slide the cover forward (about 1/2-inch) until it falls into place. The PAPER
ADVANCE button will come through the square opening and the cover's tongue
will drop behind the silver bar.
Press the PAPER ADVANCE button to verify that paper advances through the
cover.
641
#10
log
ata
rC
ape
P
t
en
em
lac
Rep
1A
-05
4A
3-0
G#
DW
-05
3-0
G#
DW
12.
13.
4-62
From the Ready For Analysis screen, press MENU, 5, 7, 4, 1 to obtain a character
printout to verify proper operation.
Press CLEAR, CLEAR, CLEAR to return to the Ready For Analysis screen.
4 Maintenance
4.4.18 Gas Cylinders Replacement
1.
Remove the regulators from the cylinders and remove the empty cylinders from the
stands.
Place the new cylinder(s) (Cal. Gas A, PN 06586; Cal. Gas B, PN 06587) in the
Stands (PN 06546).
6.
7.
8.
NOTE: If the gas compositions of the new gas cylinders are different from the
previous cylinders, set new gas cal values. See Section 4.1.6.
9.
Turn the central knob on the low pressure gauge (left valve): clockwise will
increase the pressure and counterclockwise will decrease the pressure.
From the Maintenance Menu, press 4 then 2 to perform a gas prime. Reverify low
gas pressure value during prime sequence.
Readjust as necessary.
4-63
4. Maint.
5.
Attach the regulator to the cylinder. Be sure to include the washer at the valvecylinder junction. Repeat for the other cylinder.
Turn the gas cylinder on with the Gas Wrench (PN 06547). The gas delivery must be
at 5 to 7 psig. If gas pressure is not in this range, adjust the regulators per the Gas
Regulator Adjustment procedure.
Check all junctions on the cylinder valve fittings for leaks with leak detecting agent.
If leaks are detected, turn off the tank, check the washer at the valve-cylinder
junction for proper seal, reposition the regulator on the tank, and tighten the fitting.
From the Maintenance Menu, press 4 then 2 to perform a gas prime. Reverify low
gas pressure value during prime sequence.
Verify that both humidifier wells exhibit bubbling.
Shut off the gas supply of the filter to be replaced. Disconnect the hose from the gas
supply side of the filter. See Figure 4.60.
A
4. Maint.
B
Label Side
Towards Analyzer
24A
0-0
G
DW
#5-
Filter
Assembly
To Gas Supply
2.
3.
4.
5.
6.
4-64
Place the gas tanks approximately 5 feet or less from the gas input ports.
4 Maintenance
2.
Attach the gas regulators to the tanks and secure the tanks with stands.
Slide an end of a hose over the fitting on the gas tank regulator and secure the hose.
Repeat for the other hose and regulator.
Connect the hoses to the input sides of the filter assemblies.
Attach filter assemblies to the 2 gas ports on the right side of the instrument making
sure to attach hose A to fitting A and hose B to fitting B.
DWG #7-0-003A
1.
2.
3.
4.
Remove the fuses and check to see if one or both are blown. If fuse(s) is blown, note
the line voltage indicated on the serial number plate located next to the line cord and
replace with the fuse(s) indicated in Table 4.1. If the fuses are intact, contact Nova
Technical Service for assistance.
Unplug the glucose and lactate electrode cables from the rack by lifting the latch
up with 1 hand while pulling the connector out with the other hand.
4-65
4. Maint.
If the instrument status light is off and the instrument does not respond, but the cooling fan is
running, check the external fuse as follows:
Line Voltage
Fuse
100
4A SLO-BLO
120
4A SLO-BLO
220
2A SLO-BLO
240
2A SLO-BLO
____________________________________________________________________________________
4. Maint.
5.
6.
7.
NOTE: If the cooling fan is not running and the instrument status light is off,
the instrument is not receiving power from the receptacle.
4-66
5 Troubleshooting
5
Troubleshooting
This section describes the status screens, error codes, and Service Menu and explains the
troubleshooting procedures for the Stat Profile Ultra.
WARNING: Blood samples and blood products are potential sources of hepatitis and
other infectious agents. Handle all blood products and flow path components (wasteline, septum assembly, probe, flow cell, etc.) with care. Gloves and protective
clothing are recommended.
5.1
SEQUENCE ERRORS
Analysis
25 Na+ INSTAB C
Others
A5 AB TEMP TOO HIGH
The Sequence Errors screen displays errors which occur during an analysis or calibration cycle
and are mainly electrode or flow related. Press ENTER to display the System Errors screen.
5-1
5. T. Shoot
The system monitors and self-diagnoses the instrument for errors. If an error occurs, an error
code and the message, Check Status, will be displayed. There are 2 status screens, explained
as follows, for displaying error codes: the Sequence Errors screen and the System Errors
screen. From the Ready For Analysis screen, press STATUS and ENTER until the Sequence
Errors screen is displayed.
5. T. Shoot
5.2
Troubleshooting
The Stat Profile troubleshooting section is organized so that sequence error codes are listed
numerically and system error codes are listed alphabetically.
The recommended troubleshooting procedures use the most logical and direct steps to resolve
the error code. The solutions are set up in a block format which lists groups of steps to perform
in order to restore operation. The steps are also organized to prevent unnecessary parts
replacement, such as electrodes and tubing, until the more common causes for an error have
been checked.
In the case of multiple error codes, those errors which apply to flow are at the top of the
hierarchy. In most cases, when you resolve the flow error codes, the other errors will be
resolved as well.
At the bottom of each of the error code pages are important and helpful notes, which explain
some of the rationale for an error codes appearance and additional information on how the Stat
Profile functions.
The pages that follow also outline the Stat Profile Ultra 2-point calibration and the function of
each of the fluids and gases.
If the recommendations given here do not resolve the problem, contact Nova Technical
Services for troubleshooting assistance. It is helpful to have printed or written down the error
codes, flow times, and slope performance numbers.
5 Troubleshooting
5.2.1 Stat Profile Ultra 2-Point Calibration Sequence
The following is a summary of the Stat Profile Ultra 2-point calibration. A basic understanding
of the functions of each standard and gas and of the 2-point calibration sequence itself will be
especially helpful when you troubleshoot flow related problems.
The functions of each standard and gas are listed in Table 5.1.
Function
Standard A
Standard B
Standard C
Standard D
Gas A
Gas B
Flush
When you troubleshoot flow problems, it is important to view the system as a whole. This
means that you must consider all of the various components that interact in order to transport
fluids throughout the system. These components include the pump, sampler probe, bypass
valve, pinch valve, septa assembly, and tubing.
The information in this section, along with an understanding of the flow path components and
their functions, will give you the knowledge and tools you will need to resolve most of the
problems you will encounter.
The calibration sequence is outlined in Table 5.2 according to the order in which the various
calibration standards and gases are brought into the system for slope determination, flow
checks, membrane checks, and so on. The times that are listed are approximate.
5-3
5. T. Shoot
Standard/Gas
5. T. Shoot
Time Remaining
(seconds) Fluid/Gas
584
............................
515
Std B ...................
448
Flush ...................
376
Std A ...................
314
Gas A ..................
5-4
235
157
Gas B ..................
Std D ...................
46
Std C ...................
20
............................
Function
Sequence Start
Standard B reading for AD2, AD3, Hct, iCa, and pH
Flush readings for iCa
Standard A reading for pH, Lac, and BUN linearity check
Gas A reading for PO2 and PCO2, sample preheater
check AD2, AD3
Gas B reading for PO2 and PCO2
Calibration flow check and Standard D reading
for electrolytes, glucose, lactate, BUN
Standard C reading for electrolytes, glucose, lactate, BUN
AD2, AD3, and Hct
Electrode Slopes are calculated
AD thresholds are calculated
Cal Drifts are calculated
Initialization
End Calibration
5 Troubleshooting
Rapid Reference Guide for Resolving Problems with Stat Profile Ultra Results
pH
Results High
Results Low
pO2
Results High
Results Low
ClResults High
Results Low
pCO2
Results High
5. T. Shoot
Results Low
Na+
Results High
Results Low
Glucose/Lactate/BUN
Results High
Results Low
Replace membrane
K+
Results High
Results Low
Ca++
Results High
Results Low
Combination Problems
pH and Na+ both out
Check barometer
Check system Gas Cal values
5-5
Table 5.3 Stat Profile Ultra Error Codes and Page Numbers
5. T. Shoot
Error
Code Description
Page
5-6
Error
Code Description
Page
5 Troubleshooting
Table 5.3 (continued) Stat Profile Ultra Error Codes and Page Numbers
Error
Code Description
Page
Page
5-7
5. T. Shoot
Error
Code Description
Error
Code Description
Page
5. T. Shoot
5-8
5 Troubleshooting
Error Codes
01 pH Overload A/C
11 pH Overload B/S
Description:
A pH electrode overload occurs when the voltage of the pH electrode exceeds the limits
of the analyzer.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
An electrode overload is the second highest level error code in the troubleshooting hierarchy. If
the overload code is accompanied by a flow error code, resolve the flow error first and the electrode
overload should disappear.
Electrode slope error codes will also appear when an electrode overload exists. Once you resolve
the electrode overload error, the electrode slope errors should disappear.
If an electrode overload exists on more than 1 electrode, the problem may stem from an improperly
attached rack. Remember to power the system down before you reattach the rack.
Error code 30, Math Error, may appear when an overload error occurs. Once you resolve the
electrode overload, the math error should disappear.
Electrode overload errors typically stem from either a faulty electrode or an electronic problem.
21 pH Instability A/C
31 pH Instability B/S
Description:
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
If the instability code is accompanied by flow error codes or slope error codes, resolve these error
codes first and the instability codes should disappear.
Instabilities on patient samples may occur randomly from time to time as a result of interfering
species within the sample, such as drugs.
The pH washer is thin with a small hole.
Make sure reference pump tubing is correctly installed.
5 Troubleshooting
Error Code
41 pH Slope Error
Description:
A pH slope error occurs when the difference (delta) between the calibration Standard
A and Standard B readings is not within the acceptable limits of the analyzer.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
An electrode slope error is the third highest level error code in the troubleshooting hierarchy. If the
slope code is accompanied by a flow error code or overload error, resolve these errors first and the
electrode slope error code should disappear.
If multiple slope errors occur and no flow problems are evident, the problem may stem from an
R-line flow problem.
If the slope sign appears converse (i.e. -10.0 instead of +10.0), the problem may stem from reversed
A and B-lines in the fluid pack bottles, or from reversed A and B-bottles in the fluid pack.
pH relies on the reference electrode.
5. T. Shoot
51 E-Zero Drift pH
Description:
An E-zero drift error occurs during the time between a 1-point calibration check and
the last 2-point calibration, when the electrode millivolts for Standard C drift more than
is allowed by the limits of the electrode.
Solution Block A
Solution Block B
1. Replace the pH electrode (after you condition it according to the procedure in Section
4, Maintenance).
2. Run a flow cell conditioning cycle.
3. If the problem persists after running samples,
contact Nova Technical Services.
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
An E-zero drift is the second highest level error code in the troubleshooting hierarchy. If the E-zero
drift appears with a flow error code, resolve the flow error code first and the E-zero drift error
should disappear.
The amount of drift will be printed in mph units for Standard C if this calibration option was
selected under the Operation Menu.
pH relies on the reference electrode.
Nova Technical Services 1-800-545-NOVA
5-12
5 Troubleshooting
Error Code
Description:
This error code appears when a given sample analyzed falls outside the measurable
range of the analyzer for the analyte in question (pH).
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5. T. Shoot
71 A to A Drift pH
Description:
A pH A to A (analysis to analysis) drift error occurs when the electrodes voltage for
Standard A drifts more than the analyzers limits.
Solution Block A
Solution Block B
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
A pH A to A drift is the second highest level error code in the troubleshooting hierarchy. If the A
to A drift appears with a flow error code, resolve the flow error first and the A to A drift error should
disappear.
pH is reference electrode dependent.
5 Troubleshooting
Error Codes
22 PO2 Instability A
32 PO2 Instability B/S
Description:
An instability error occurs during calibration or analysis cycles if the PO2 electrodes
response fluctuates excessively while the reading is being taken.
Solution Block A
Solution Block B
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
If the instability code is accompanied by flow error codes or slope error codes, resolve these error
codes first and the instability codes should disappear.
The PO2 washer is thick with a large hole (PN 09798).
PO2 is temperature sensitive, and it may take 10 to 15 minutes for the internal filling solution and
electrode to equilibrate to 37C.
Description:
A PO2 slope error occurs when the difference (delta) between the calibration Gas A and
Gas B readings is not within the acceptable limits of the analyzer.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
PO2 is temperature sensitive, and it may take 10 to 15 minutes for the internal filling
solution and electrode to equilibrate to 37C.
The PO2 electrode requires a temperature equilibration time of 5 to 10 minutes.
An electrode slope of approximately 0.0 indicates a faulty base electrode.
A PO2 slope error that occurs along with a PCO2 slope error could indicate that there is a
problem in the gas cylinders or that the gas lines are backwards.
PO2 does not rely on the reference electrode.
Be sure not to overfill the membrane cap with PO2 solution.
5 Troubleshooting
Error Code
Description:
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
An E-zero drift is the second highest level error code in the troubleshooting hierarchy. If the E-zero
drift appears with a flow error code, resolve the flow error code first and the E-zero drift error
should disappear.
Failure to allow enough time for the electrode to temperature equilibrate after servicing is a
common cause of E-zero drift for PO2.
PO2 is temperature sensitive, and it may take 10 to 15 minutes for the internal filling solution and
the electrode to equilibrate to 37C.
The amount of drift will be printed in mmHg for Gas A if this calibration option was selected under
the Operation Menu.
The PO2 electrode does not rely on the reference electrode.
Description:
This error code appears when a sample is outside the measurable range of the analyzer
for PO2.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5 Troubleshooting
Error Codes
Description:
A PO2 membrane overload error occurs when the membrane checks that occur during
the calibration and analysis sequences detect the presence of moisture in the flow cell
chamber.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
The PO2 electrode membrane test helps you determine if a membrane has ruptured. The membrane
test fails, however, if the flow cell chamber is wet when the electrode is inserted into it or if a salt
bridge develops within the chamber.
To avoid problems with membrane failure codes, be sure to keep the flow cell chamber clean and
dry before you insert the PO2 electrode.
Error code 30, Math Error, may appear when a PO2 membrane error occurs. Resolve the PO2
membrane error first and the math error should disappear.
Be careful not to overfill the membrane cap with PO2 solution.
23 PCO2 Instability A
33 PCO2 Instability B/S
Description:
An instability error occurs during calibration or analysis cycles if the PCO2 electrodes
response fluctuates excessively while the reading is being taken.
Solution Block B
1. Replace the PCO2 membrane cap and perform flow cell conditioning with whole
blood as described in Section 4, Maintenance.
2. Shake the electrode down to remove any
bubbles from behind the membrane.
3. Clean and dry the PCO2 flow cell chamber
and washer or replace the washer.
4. Allow the necessary electrode equilibration time.
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
If the instability code is accompanied by flow error codes or slope error codes, resolve these error
codes first and the instability codes should disappear.
The PCO2 washer is thick with a small hole.
PCO2 is temperature sensitive, and it may take 10 to 15 minutes for the internal filling solution and
the electrode to equilibrate to 37C.
PCO2 does not use the reference electrode.
5 Troubleshooting
Error Code
Description:
A PCO2 slope error occurs when the difference (delta) between the calibration Gas A
and Gas B readings is not within acceptable limits of the analyzer.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5. T. Shoot
and the last gas calibration I or the last full gas calibration (whichever occurred last).
An E-zero drift PCO2 error indicates that the electrode millivolts for Gas A have drifted
outside the acceptable limits of the analyzer.
Solution Block A
Solution Block B
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
An E-zero drift is the second highest level error code in the troubleshooting hierarchy. If the E-zero
drift appears with a flow error code, resolve the flow error code first and the E-zero drift error
should disappear.
Failure to allow enough time for the electrode to temperature equilibrate after servicing is a
common cause of E-zero drift for PCO2.
PCO2 is temperature sensitive, and it may take 10 to 15 minutes for the internal filling solution and
the electrode to equilibrate to 37C.
The amount of drift will be printed in millimeters of Hg for Gas A, if this calibration option was
selected under the Operation Menu.
PCO2 is not reference dependent.
5 Troubleshooting
Error Code
Description:
This error code appears when a sample is outside the measurable range of the electrode
for PCO2.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Description:
A PCO2 membrane overload error occurs when the membrane checks, which occur in
the calibration and analysis sequences, detect the presence of moisture on the flow cell
chamber.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
The PCO2 electrode membrane test helps you know if a membrane has ruptured. The membrane
test fails, however, if the flow cell chamber is wet when the electrode is inserted into it or if a salt
bridge develops within the chamber.
To avoid problems with membrane failure codes, always make sure that the flow cell chamber is
clean and dry before you insert the PCO2 electrode into it.
Error code 30, Math Error, may appear when a PCO2 membrane error occurs. Resolve the PCO2
membrane error first and the math error should disappear.
5 Troubleshooting
Error Code
Description:
At the normal Gas A calibration read time, the software determines if additional time
is needed to reach the electrode endpoint. If required, Gas A will continue to flow until
the endpoint is reached. If no endpoint is reached after 60 seconds of additional gas, this
error code is displayed and the PCO2 electrode becomes uncalibrated.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Description:
This overload error occurs if the millivolt readings exceed the acceptable limit when
the analyzer is taking readings for SO2% Calibrator Level 1 or flush.
Solution Block A
Solution Block B
5. T. Shoot
Notes:
5 Troubleshooting
Error Code
Description:
This overload error occurs if the millivolt readings exceed the acceptable limit when
the analyzer is taking readings for SO2% Calibrator Level 2 or the sample.
Solution Block A
Solution Block B
5. T. Shoot
Notes:
Description:
An instability error occurs during calibration or analysis cycles when the electrode's
response fluctuates while the electrode response reading is being taken.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5 Troubleshooting
Error Code
Description:
No mV difference was observed between the calibrator standards or between the blank
and standard readings.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Description:
An E-zero drift occurs if the mV for flush drift more than allowed by the limits of the
SO2% channel.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
If the fiber optic cable is removed from the flow cell or if it is disturbed by moving the flow cell/
preheater/rack, the flush mV will show drift and SO2% MUST be recalibrated.
5 Troubleshooting
Error Code
Description:
This error code appears when a sample is outside the measurable range of the analyzer
for SO2%.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Description:
This error code appears when a sample falls above the allowable lower limit for SO2%.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5 Troubleshooting
Error Code
94 Hi K+ Hemolysis
Description:
This error code appears when K+ is too high for a concentration to be calculated by
SO2%.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
H8 SO2% Chan/Sample
Description:
This error indicates a problem with the LED during sample analysis, but does not
indicate an error when running controls.
Solution Block A
Solution Block B
5. T. Shoot
Notes:
5 Troubleshooting
Error Code
Description:
A hematocrit slope error occurs when the difference (delta) between the calibration
Standard B and Standard C readings is not within the acceptable limits of the hematocrit
sensor.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Hematocrit slope errors typically result from a flow problem in the system. If the analyzer displays
flow error codes, resolve the flow errors first and the hematocrit slope error should disappear.
Hematocrit slope errors could result from coated air detectors in a dirty sample preheater.
A hematocrit slope error that appears alone usually stems from a failing sample preheater or
an electronic problem.
24 Hct Instability C
34 Hct Instability B/S
Description:
An instability error occurs during calibration or analysis cycles when the electrodes
response fluctuates excessively while the reading is taken.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Error code 34 often occurs with error code 9C, Insufficient Sample. Resolve the error code,
Insufficient Sample, first and the Hct Instability error code should disappear.
5 Troubleshooting
Error Code
Description:
An E-zero drift Hct error occurs during the time between a 1-point calibration check
and the last 2-point calibration. An E-Zero Drift Hct error indicates that the Hct
impedance electrode (air detector #3 in the sample preheater) millivolts for Standard
C have drifted outside the acceptable limits of the analyzer.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
An E-zero drift is the second highest level error code in the troubleshooting hierarchy. If the E-zero
drift appears with a flow error code, resolve the flow error code first and the E-zero drift error
should disappear.
74 A to A Drift Hct
Description:
A Hct A to A (analysis to analysis) drift error occurs when the Hct impedance electrode,
air detector (oscillator) #3, located in the sample preheater, drifts in millivolts for
Standard C more than the analyzers acceptable limits. The analyzer periodically
performs a 1-point calibration to check the Standard C millivolts for drift.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Since the Hct impedance electrode, air detector #3, located in the sample preheater, performs best
when the preheater is kept clean, the frequency of sample preheater cleaning based on sample
volume may prevent this error code from occurring.
Refer to Section 4, Maintenance, or to the Stat Profile Maintenance Logbook for the sample preheater
cleaning frequency.
5 Troubleshooting
Error Code
Description:
This error code appears when a sample is below the analyzers lower limit for Hct.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Description:
This error code appears when a sample is above the analyzers upper limit for Hct.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5 Troubleshooting
Error Code
Description:
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
If the error is accompanied by a Na+ error code, resolve this error code first and the Hct error
condition might also be resolved.
The measurable range for Na+ is 80 to 200 mmol/L.
25 Na+ Instability C
35 Na+ Instability D/S
Description:
An instability error occurs during calibration or analysis cycles when the electrodes
response fluctuates excessively while the electrode response reading is being taken.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
If the instability code is accompanied by flow error codes or slope error codes, resolve these error
codes first, and the instability codes should disappear.
Na+ electrode uses the reference electrode.
5 Troubleshooting
Error Codes
66, 67, 68, 69 Na+, K+, Cl-, Ca++ Out of Range (Sample Analysis)
Description:
These error codes appear when a sample is outside the measurement range of the
electrode for Na+, K+, Cl-, Ca++.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
75 A to A Drift Na+
Description:
A sodium A to A (analysis to analysis) drift error occurs when the voltage for Standard
C drifts outside the analyzers acceptable limits. The analyzer periodically performs a
1-point calibration to check the Standard C millivolts for drift.
Solution Block B
1. Open the analytical compartment door, access the System Test screen, and extend the
probe to position #9. Remove the W-tube.
Use a syringe to flush deionized water
through the W-port of the reference electrode; fluid should exit from the probe tip.
2. Remove and flush the reference electrode
as described in Section 5, Troubleshooting,
then perform an Electrolyte Calibration. If
the error persists after the next Cal I, continue with Step 3.
3. Replace the W/R-tubing segments as described in Section 4, Maintenance. Perform a
full calibration. If the error code persists
after the next Cal I, continue with Step 4.
4. Replace the sodium electrode as described
in Section 4, Maintenance, and perform an
Electrolyte Calibration.
5. Contact Nova Technical Services.
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5 Troubleshooting
Error Code
Description:
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5. T. Shoot
Solution Block A
Description:
An instability error occurs during calibration or analysis cycles when the electrodes
response fluctuates while the reading is being taken.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
If the instability code is accompanied by flow error codes or slope error codes, resolve these error
codes first and the instability codes should disappear.
Instabilities on patient samples may occur randomly from time to time as a result of interfering
species within the sample, such as drugs.
This electrode relies on the reference electrode.
5 Troubleshooting
Error Codes
Description:
A slope error occurs when the difference (delta) between the calibration Standard C and
Standard D readings is not within acceptable limits of the analyzer.
Solution Block B
1. If the slope error recurs for sodium, potassium, or calcium, replace the appropriate
electrode. If the slope error recurs for chloride, replace the membrane cap.
2. If multiple electrode slope errors occur, the
reference electrode might be at fault.
a. Check the R-solution level in the reagent pack window. Replace the pack
if necessary.
b. Flush the reference electrode as described in Section 5, Troubleshooting.
c. Calibrate the analyzer.
3. If multiple electrode slope errors recur,
replace the reference electrode.
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
An electrode slope error is the third highest level error code in the troubleshooting hierarchy.
If the slope code is accompanied by a flow error code or an overload error code, resolve the
flow or overload error codes first and the electrode slope error code should disappear.
If multiple slope errors appear and no flow problems are evident, the problem may stem from
a faulty R-line, an R-line flow problem, or a faulty fluids pack.
If the slope sign appears negative, the problem may stem from reversed C and D-lines in the reagent
bottles or from reversed bottles in the fluids pack.
The sodium electrode does not have a flow cell chamber washer.
The potassium and calcium electrodes require a thin washer with a large hole (PN 08153).
5. T. Shoot
Solution Block A
5. T. Shoot
Description:
An E-zero drift error occurs during the time between a 1-point calibration check and
the last 2-point calibration. An E-zero drift error indicates that the electrode millivolts
for Standard C have drifted more than the acceptable operational limits of the electrode.
Solution Block A
Solution Block B
1. If the error code persists on only the chloride electrode, replace the chloride cap as
described in Section 4, Maintenance. If the
error code persists only on the Na+, K+, or
Ca++ electrode, replace the electrode as described in Section 4, Maintenance.
2. If E-zero drifts occur for multiple electrodes, the reference electrode might be at
fault.
a. Check the R-solution level in the
reagent pack and replace the pack if
necessary.
b. Flush the reference electrode as
described in Section 5, Troubleshooting.
c. Calibrate the analyzer.
3. If multiple E-zero drift errors recur after the
next Cal I, replace the reference electrode.
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
An E-zero drift is the second highest level error code in the troubleshooting hierarchy. If the E-zero
drift appears with a flow error code, resolve the flow error code first and the E-zero drift error
should disappear.
The amount of drift will be printed in milligrams per deciliter for Standard C, if this calibration
option was selected under the Operation Menu.
These electrodes are all reference dependent.
5 Troubleshooting
Error Codes
0A, 1A, 09, 19, 0C, 1C Lactate, Glucose, BUN Overload (A/C, D/S)
Description:
A lactate overload occurs when the voltage for Standards A, C, D, or S exceeds the
limits of the analyzer or if the sample concentration is very high.
A glucose overload occurs when the millivolt readings for Standard C, D, or S exceeds
the limits of the analyzer.
A BUN overload occurs when the voltage for Standards A, C, or D exceeds the
Solution Block A acceptable limits of the analyzer.
Solution Block B
Solution Block B
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
An electrode overload is the second highest level error code in the troubleshooting hierarchy. If
the overload code is accompanied by a flow error code, resolve the flow error first and the electrode
overload should disappear.
Electrode slope error codes will also appear when an electrode overload exists. Once you resolve
the electrode overload error, the electrode slope errors should disappear.
If an electrode overload exists on more than 1 electrode, the problem may stem from an improperly attached
rack. Remember to power the system down before you reattach the rack. The BUN electrode uses the
reference electrode common to pH, Na+, K+, and Cl-. If any of these channels are also experiencing
problems, flush or replace the Reference electrode as described in Section 5, Troubleshooting.
Error code 30, Math Error, may appear when an overload error occurs. Once you resolve the
electrode overload, the math error should disappear.
Electrode overload errors typically stem from either a faulty electrode or an electronic problem.
Lactate is not dependent on the reference electrode.
Nova Technical Services 1-800-545-NOVA
5-49
5. T. Shoot
Solution Block A
2A, 3A Lactate Instab (A/C, D/S); 29, 39 Glucose Instability (C, D/S);
2C, 3C BUN Instab (A/C, D/S)
Description:
An instability error occurs during calibration or analysis cycles when the glucose.
lactate, or BUN electrodes response fluctuates excessively while the reading is taken.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
If the instability error code is accompanied by flow error codes or slope error codes, resolve these
errors first and the instability codes should disappear.
The glucose electrode may exhibit an unstable reading on samples that have a glucose content
greater than the 500 mg/dL measurable range.
The BUN electrode uses the reference electrode common to pH, Na+, K+, and Cl-. If multiple
channels are experiencing problems, flush or replace the reference electrode.
Glucose does not rely on the reference electrode.
5 Troubleshooting
Error Codes
Description:
A lactate or glucose slope error occurs when the difference (delta) between the
calibration Standard C and Standard D readings is not within acceptable limits of the
analyzer.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Glucose electrode slope errors typically occur either when a membrane cap is new and needs to
equilibrate as described in Section 4, Maintenance, or when the membrane is new and needs to be
replaced.
Nova Biomedical recommends that the analyzer be operated in single mode for 1 hour whenever
the electrode or membrane is replaced.
The lactate and glucose electrodes do not depend on the reference electrode, so you should not
suspect the reference electrode as the cause of a glucose slope error.
4C BUN Slope
Description:
A BUN slope error occurs when the difference (delta) between the calibration Standard
C and Standard D voltages is not within the analyzers acceptable limits.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
The BUN electrode uses the reference electrode common to pH, Na+, K+, and Cl-. If multiple
channels are experiencing problems, flush or replace the reference electrode as described in Section
4, Maintenance.
5 Troubleshooting
Error Codes
Description:
An E-zero drift error occurs during the time between a 1-point calibration check and
the last 2-point calibration. An E-zero drift error indicates that the electrode millivolts
for Standard C have drifted outside the acceptable limits of the analyzer.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
An E-zero drift is the second highest level error code in the troubleshooting hierarchy. If the E-zero
drift appears with a flow error code, resolve the flow error code first and the E-zero drift error
should disappear.
The amount of drift will be printed in milligrams per deciliter for Standard C, if this calibration
option was selected under the Operation Menu.
The glucose electrode does not depend on the reference electrode.
These error codes may occur for very high lactate or glucose samples (outside the measuring
range). Resolve the out of range error first, and the E-zero drift error should disappear.
Description:
An out of range error occurs when an analysis is performed on a sample and the result
for an analyte is calculated to be above or below the acceptable range for the electrode.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
The measurable range for lactate is 0 to 20 mmol/L. Lactate values between 20-23 will be displayed
with a question mark. Results greater than 23 will not be displayed and sample dilution is required.
An estimated lactate value will be displayed in reverse video on the Metabolites screen (MENU,
5,1,3). Use extreme caution when you view this value and use it only as a guideline for dilution
purposes. Dilution procedure is described in Section 1, Introduction.
5 Troubleshooting
Error Code
Description:
A result too high error occurs when an analysis is performed on a sample and the result
is calculated to be above or equal to 23 mmol/L.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Description:
A result too high error occurs when an analysis is performed on a sample and the result
is calculated to be greater than 110 mg/dL.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5 Troubleshooting
Error Codes
Description:
A lactate, BUN, or glucose A to A (analysis to analysis) drift error occurs when the
voltage for Standard C drifts more than the analyzers acceptable limits. The analyzer
periodically performs a 1-point calibration to check the Standard C millivolts for drift.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Lactate and glucose electrodes A to A drifts typically occur either when a membrane cap is new
or when the membrane is failing and needs to be replaced.
Nova Biomedical recommends that you operate the analyzer in single mode for 1 hour whenever
the membrane or electrode is replaced.
Description:
5. T. Shoot
Solution Block A
Solution Block B
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5 Troubleshooting
Error Code
Description:
Lactate A solution high errors usually occur with a new lactate membrane that needs
additional equilibration. Recovery on Standard A (15 mmol/L) is high.
Solution Block A
Solution Block B
1. If the error code persists after 2 full calibrations, replace the membrane as described in
Section 4, Maintenance.
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Description:
This error usually occurs when a lactate membrane is no longer linear. Recovery on
Standard A (15 mmol/L) is low.
Solution Block B
1. Polish the electrode and replace the membrane as described in Section 4, Maintenance.
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5 Troubleshooting
Error Codes
Description:
A PO2 electrode overload occurs when the voltage readings exceed the acceptable
limits of the analyzer.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
A PO2 electrode overload is the second highest level error code in the troubleshooting hierarchy.
Resolve the overload error code and all of the other PO2 error codes (such as slope error codes)
should disappear.
Electrode slope error codes will also appear when an electrode overload exists. Once you resolve
the electrode overload, the electrode slope errors should disappear.
If the problem is occurring on multiple electrodes, it may result from an improperly attached rack.
Remember to power the system down before you reattach the rack.
Error code 30, Math Error, may appear when a PO2 overload error occurs. The math errors should
disappear when you resolve the electrode overload errors.
Electrode overload errors typically stem from either a faulty electrode or an electronic problem.
The oxygen electrode does not rely on the reference electrode.
Nova Technical Services 1-800-545-NOVA
5-61
Description:
A PCO2 overload occurs when the voltage readings exceed the acceptable limits of
the analyzer.
Solution Block A
Solution Block B
1. Replace the PCO2 electrode.
2. Condition the new electrode as described
in Section 4, Maintenance.
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
A PCO2 overload is the second highest level error code in the troubleshooting hierarchy. If the
overload code is accompanied by a flow error, resolve the flow error and the overload error should
disappear.
Electrode slope error codes will also appear when an electrode overload exists. Once you resolve
the overload, the slope errors should disappear.
If the problem is occurring on multiple electrodes including the PCO2, it may result from an
improperly attached rack. Remember to unplug the power cord before you reattach the rack.
Error code 30, Math Error, may appear when a PCO2 overload occurs. The math errors should
disappear when you resolve the electrode overload.
Electrode overload errors typically stem from either a faulty electrode or an electronic problem.
The PCO2 electrode is not dependent on the reference electrode.
Nova Technical Services 1-800-545-NOVA
5-62
5 Troubleshooting
Error Codes
Description:
An Hct overload occurs when the voltage readings exceed the acceptable limits of the
analyzer.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
The Hct impedance electrode is located in the sample preheater and is also an air detector or air
oscillator #3.
This error code typically results from an electronic problem.
If the error code is accompanied by a flow error code, resolve the flow error code first and the
electrode overload error should disappear.
Description:
A sodium electrode overload occurs when the voltage readings exceed the acceptable
limits of the analyzer.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
An electrode overload is the second highest level error code in the troubleshooting hierarchy. If
the overload code is accompanied by a flow error code, resolve the flow error first and the electrode
overload should disappear.
Electrode slope error codes will also appear when an electrode overload exists. Once you resolve
the overload, the electrode slope errors should disappear.
If an electrode overload exists on more than 1 electrode, the problem may stem from an improperly
attached rack.
Error code 30, Math Error, may appear when a Na+ overload error occurs. Once you resolve the
overload, the math error should disappear.
5 Troubleshooting
Error Codes
Description:
Electrode overload error occurs when the voltage readings exceed the acceptable limits
of the analyzer.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
An electrode overload is the second highest level error code in the troubleshooting hierarchy.
If the overload code is accompanied by a flow error code, resolve the flow error code and the
overload error should disappear.
Electrode slope errors will appear if an overload occurs.
If the problem occurs on multiple electrodes, it may stem from an improperly attached
rack. Remember to unplug the power cord before reattaching the rack.
Error code 30, Math Error, may appear when an electrode overload occurs. When you resolve the
overload, the math error should disappear.
Description:
An out of range error occurs when an analysis is performed on a sample and the
result for an analyte is calculated to be above or below the acceptable range of the
electrode.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5 Troubleshooting
Error Code
Description:
A result too high error occurs when an analysis is performed on a sample, and the result
is calculated to be above or equal to 700 mg/dL.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Description:
An out of range error occurs when an analysis is performed on a sample and the result
for an analyte is calculated to be above or below the acceptable linear range of the
electrode.
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5 Troubleshooting
Error Code
9C Insufficient Sample
Description:
This error code occurs when air detector 3 (also called air oscillator 3), located in the
sample preheater, detects air at the trailing edge of the sample during an analysis. This
error code usually stems from a short sample or flow problem, but could also stem from
a high hematocrit sample (>70%).
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
If the error code is accompanied by a flow error code, resolve the flow error code first, and the
insufficient sample code should disappear.
This error code may appear with high Hct samples (greater than 70%) because the air detector 3
(which is also the Hct impedance electrode) confuses the sample with air.
Results generated with this error code will print out along with a question mark.
If this error code persists after multiple samples, it is most likely caused by a dirty preheater. See
Section 4, Maintenance (Sample Preheater Cleaning), for preheater cleaning procedure.
60 Bypass Valve
Description:
The bypass valve monitors the potassium channel to make sure the reference solution
does not backflow into the flow cell.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
When the bypass valve fails to pinch off the bypass tubing completely, reference solution can
migrate down the flow path into the flow cell chamber and affect the K+ results.
When you position the bypass valve tubing, make sure it is seated inside the notch at the bottom
of the tubing slot. On some bypass valve assemblies, this notch is located half way down the tubing
slot.
5 Troubleshooting
Error Codes
Description:
A delta millivolt error code appears when the measured millivolt difference between
the standards used to calibrate the air detectors (Standard A and flush) is not within
acceptable limits.
Solution Block B
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
Most of the problems you encounter with delta millivolt errors will stem from restrictions in the
flow path. Over time, the sample preheater may develop a protein buildup which you should be
able to remove with the bleach solution as described above (Solution Block B, Step 1).
88
89
90
91
G0
Solution Block A G1
R0
R1
R3
R4
R5
5. T. Shoot
Solution Block A
Solution Block B
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
5 Troubleshooting
Error Codes
Solution Block A
5. T. Shoot
Solution Block B
Notes:
H3 Barometer Failure
Description:
This error code usually stems from a hardware failure of the barometer.
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
When a hard failure of the barometer occurs, the message, Check Barometric Pressure, appears in
the Ready/Not Ready for Analysis screen.
A hard failure of the barometer causes the barometric pressure section of the operation configuration (in the Set Up Menu) to automatically switch to Entered.
If a hard failure of the barometer occurs, the analyzer can still be run by manually entering the
barometric pressure in the Operation Configuration Menu.
If you have to manually enter the barometric pressure in the Operation Menu, Nova recommends
that you check the barometric pressure during each shift and adjust it accordingly until the new
barometric pressure module is installed.
5 Troubleshooting
Error Code
Description:
Solution Block A
Solution Block B
5. T. Shoot
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
After you unplug the analyzer and reattach the rack and sample preheater, it may take several
minutes for the air bath and sample preheater to reach 37C.
The power should always be removed from the analyzer before you reattach the rack, to prevent
the sensor board from being damaged.
Description:
Solution Block B
5. T. Shoot
Solution Block A
If these solutions do not resolve the problem, call Nova Technical Services
Notes:
This is an internal check for BUN membrane linearity during calibration. Therefore, it is not
necessary to run BUN over-range solution.
5 Troubleshooting
5.2.2 The Stat Profile Ultra Flow Test
WARNING: Blood samples and blood products are potential sources of hepatitis and
other infectious agents. Handle all blood products and flow path components (wasteline, septum assembly, probe, flow cell, etc.) with care. Gloves and protective
clothing are recommended.
The Stat Profile Ultra flow test is used to troubleshoot error codes such as:
4.
5.
Using the menu button, access the service screen (press MENU then 5, 1).
5-77
5. T. Shoot
1.
2.
3.
SYSTEM TEST #1
Revision 123.456
pH
+ 047.43
PCO2
-082.12
PO2
-0051.20
1
2
3
4
5
6
7
8
9
0
Na+
+008.56
K+
+017.35
++
Ca /Cl +074.43
Millivolt Display
PCO2 Membrane
O2 Gain
Air Oscillators
Pump
Sampler Position
Valve Position
Gas
Pump Bypass Valve
SO2% LEDs
Glu
Lac
BUN
+042.32
-023.23
-047.76
Multiple
Off
Low
Off
0-Off
0
Std C
0-Off
Closed
Off
5. T. Shoot
2.
3.
4.
5.
6.
5-78
5 Troubleshooting
If no suction is observed here, the problem is with the pump, bypass valve, or W-line. Correct the
problem as outlined below.
Correcting Pump Problems:
a.
b.
Check that the W-line bypass tubing segment is correctly positioned in the
bypass valve. Incorrectly positioned tubing will prevent the pump from
developing any draw. Split tubing or a pinhole leak can also cause this
problem.
Check that the bypass valve is closed. Press 9 repeatedly on the system test
screen and watch to see that the bypass valve actually opens and closes.
(You should see the valve piston move in and out.) Contact Nova service
if the bypass valve has failed.
Flush or replace the W-pump tubing segment and the waste line. The waste
path from the W-line to the waste bottle may be clogged.
c.
If the W-line is pulling water and the system failed the water test (no suction at the probe), there
is a blockage or leak in the flow path. Proceed with the Flow Path Flush Procedure below.
To flush the entire flow path with air or water, perform Steps 1 through 6 of the Flow Path Flush
procedure. For a flush of the flow path starting at the sample preheater, open the analyzers
door, detach the S-line from the sample preheater, attach the probe cleaning syringe tubing
directly to the bottom of the sample preheater, access the System Test #1 screen, and continue
with Step 4. When the procedure has been completed, reattach the S-line to the sample
preheater.
For individual components of the flow path, perform the appropriate procedure per the
Individual Component procedures.
CAUTION: Do not open or close the analyzers door when the probe is in
position 9 (extended). Probe damage will occur.
Flow Path Flush
1.
2.
3.
4.
5.
5-79
5. T. Shoot
DWG #5-0-046B
5. T. Shoot
6.
If other components need flushing, continue with the Individual Component Flush
as follows. If no other components need flushing, press 9, 6, 0, CLEAR, CLEAR
to return to the Main Menu and close the analyzers door.
3.
4.
5.
5-80
Open the analyzers door. Remove the S-line from the sample preheater.
From the Ready For Analysis screen, press MENU, 5, 1, 6, 9 to move the sample
probe out of the septum assembly. Attach the flexible tubing of a syringe filled with
deionized water to the sample probe. Position a beaker underneath the S-line.
Push the deionized water through the probe.
If little or no flow occurs, there is probably a blood clot in the sample probe. Force
deionized water through the probe to dislodge the blood clot. If this is not
successful, draw some Cleaning Solution (PN 11802) into the syringe and after
letting the solution sit on the clot, force the remaining solution through the clot.
Again attach a deionized water-filled syringe to the probe and push water through
the probe.
If, after performing Step 5, a steady stream of water does not leave the S-line,
remove the probe, see Section 5.3.14, and inspect for a bent sample probe. A bent
probe will reduce flow and should be replaced.
5 Troubleshooting
6.
7.
If no other components need flushing, reattach the S-line to the sample preheater.
Press 0, CLEAR, CLEAR, CLEAR to return to the Main Menu and close the
analyzers door.
Press CAL, ENTER to calibrate.
2.
3.
4.
7.
8.
9.
5-81
5. T. Shoot
5.
6.
From the Ready For Analysis screen press MENU, 5 to access the Service Menu.
Press 6 to display the Fluid/Gas Prime Options screen. Then press 8 to purge the
flow cell. Wait for the gas flow to stop.
Clamp and disconnect the W/R-lines from the reference electrode.
Unplug all electrode cables from the electrode rack assembly.
Remove the glucose, lactate, and BUN electrodes from the flow cell. Position the
electrodes so that the membrane will not be damaged. Insert the electrode blanks
in both the glucose, lactate, and BUN channels.
Disconnect the S-line from the sampler preheater.
Remove the preheater-flow cell-reference electrode assembly by first turning the
bottom retainer screw 1/4-turn counterclockwise, then pull the assembly straight
off the guide pin and electrical connectors.
Loosen the thumbscrew located on top of the reference electrode, turn the retaining
block to the side, and retighten the thumbscrew.
Lift the reference electrode up and out of the way of the flow cell.
If the blocked component is known, flush it according to the proper procedure.
Otherwise, continue with Step 10.
Retaining
Block
W
Reference
Electrode
Flow Cell
for Models
F, G, H
Electrode
Rack
Assembly
Cl
Na
K
BUN
Glu
Lac
pH
Top Slotted
Retainer
Fastener
PCO2
PO2
SO2
5. T. Shoot
Retainer
Screw
Flow Cell For Models
I, J, K
K
Electrical
Connector
Na
Ca
BUN
Glu
Lac
pH
PO2
Sample
Preheater
SO2
DWG #7-F-001A
PCO2
5-82
Attach a syringe containing deionized water to the W-port and, while covering the
R-port, push water through the electrode so that the water flows out of the flow cell
port. See Figure 5.3.
5 Troubleshooting
Water
R
-0
-0
#5
W Port
(R Port Is
Covered)
A
55
DW
11.
12.
Repeat Step 10, attach the syringe to the R-port and cover the W-port.
If flushing the flow cell or the sample preheater is not needed, continue with Step
20. Otherwise, turn the top thumbscrew 1/4-turn counterclockwise and lift the flow
cell with the electrodes still inserted off the sample preheater connector. Continue
with the appropriate component.
13.
Insert the blunt tip needle (from probe cleaning kit) of a syringe filled with
deionized water into an interconnect tubing on the flow cell and push deionized
water through the flow cell. See Figure 5.4.
Cl
K
Na
Na
K
N
BU
u
Gl
c
La
pH
O
PC
2
PO
N
BU
u
Gl
c
La
pH
O2
PC
2
PO
2
SO
2
SO
02
01
-0
-0
F
7-
Ca
J
7-
DW
DW
5-83
5. T. Shoot
Repeat Step 13, flush the flow cell in the opposite direction by inserting the syringe
into the other flow cell interconnect tubing.
Inspect the flow cell for cracks and other physical damage.
Check the interconnect tubing attached to the flow cell for leaks or breaks. Replace
it if necessary (PN 07161).
DWG #5-0-057A
18.
If the flow cell is still on the sample preheater, remove by turning the top retainer
screw 1/4-turn counterclockwise.
Attach the tubing of a cleaning syringe filled with deionized water to the top
connector and push water through the sample preheater. Then attach the tubing to
the bottom connector and flush the preheater in the opposite direction. See Figure
5.5.
5. T. Shoot
WATER
19.
5-84
If additional cleaning is required, flush the sample preheater with 50% bleach and
let sit for 10 minutes. Then rinse the preheater with deionized water.
5 Troubleshooting
Reassembly Procedure
20.
5-85
5. T. Shoot
Reattach the flow cell to the sample preheater back plate as follows:
a.
Set the flow cell into the recess on the sample preheater, align the right
surface of the flow cell with the right side of the back plate. Ensure that the
flow cell interconnect tubing is seated properly on the sample preheater top
connector.
b.
Turn the top retainer screw 1/4-turn clockwise.
21. Replace the reference electrode on top of the flow cell by placing the electrode on
top of the flow cell, align the electrode sides with the backplate sides. Ensure that
the reference electrode connector is seated properly on the flow cell interconnect
tubing.
22. Loosen the retaining block thumbscrew, center the retaining block in the depression on top of the electrode, and retighten the thumbscrew.
23. Replace the preheater-flow cell-reference electrode assembly by aligning the guide
pin, pressing the assembly onto the electrical connectors, then turning the bottom
retainer screw 1/4-turn clockwise to engage the electrode rack assembly.
24. Remove the electrode blanks and insert the glucose, lactate, and BUN electrodes
into the flow cell.
25. Plug all cables into the rack assembly.
26. Reconnect the S-line to the sample preheater.
27. Attach the W/R-lines to the reference electrode.
28. Close the analyzers door and wait 5 minutes for the compartment to come to
temperature.
29. Verify flow of solution as follows:
a.
From the Ready For Analysis screen press Menu, 5, 6, 1; immediately
observe the waste-line flow for 5 seconds.
b.
If the flow is impeded, check the interconnections between the S-line
preheater, preheater flow cell, and flow cell reference electrode, and
check that the W/R-tubings are properly connected.
30. Once the analyzers compartment has stabilized to 37 C for at least 10 minutes,
press CALIBRATE to perform a 2-point calibration. Press CALIBRATE again to
perform a second 2-point calibration.
6.
5. T. Shoot
7.
Using the menu button, access the system test screen by pressing MENU, 5, 1.
Open the analytical compartment door.
Select 8, Gas. (Visually confirm that the bypass valve is closed.)
Select 3, A-Hi.
Look at the gas humidifier. See if bubbles start to flow through the humidifier. If
no bubbles appear, listen to make sure the gas metering module is running as
evidenced by a continuous ticking sound. If the gas meter module is not ticking, a
hardware failure has occurred (this is rare). Contact Nova Technical Services.
If the gas metering module is running, but no gas appears in the humidifier, check
the gas supply and the lines connecting the gas supply to the analyzer. If these are
OK, the gas connection to the humidifier may be loose. Remove the humidifier and
check. If this connection is OK, an internal line has come loose (rare). Contact Nova
Technical Services is this occurs.
Within about 30 seconds, the bubbles should get smaller and slow down to almost
a complete stop (1 bubble every 5 seconds or so is acceptable) as the entire system
pressurizes to 5 psig (the regulator setting). Failure of the gas to slow down indicates
a flow path leak. Stop here and isolate the leak using the hemostat or clamp.
a.
Clamp the W-line just above the reference electrode and check the bubbles.
If the bubbles stop, the leak exists between the pump, bypass valve, or waste
line. If the bubbles do not stop, proceed with b.
b.
Clamp the soft tubing at the end of the S-line where it connects to the
preheater and observe whether the bubbles stop. If the bubbles stop, the leak
is somewhere between the reference electrode and the preheater.
8.
9.
If the bubbling still has not stopped, clamp the rearmost tubing on the septum
harness and observe whether the bubbles stop. If the bubbles stop here, a faulty
septum assembly or a faulty connection between the septum harness and the septum
assembly is the problem.
If the bubbling does not stop when the rearmost line of the septum harness is
clamped off, the following must be checked.
5-86
Confirm that the humidifier caps are finger tight, and that the o-ring on the
cap is not missing or damaged.
5 Troubleshooting
Remove the humidifier and check to make sure the gas line tubing is
securely connected to the humidifier ports and the T-fitting at the humidifier outlet.
Virtually all flow problems are detected by the above procedures. However, if these tests fail
to locate the problem, it will also be necessary to do a probe alignment check. See the probe
alignment procedure in Section 5.3.1.5.
NOTE: The composition of room air (20.9% oxygen and 0.03% carbon
dioxide) results in a PO2 of about 150 mmHg and a PCO2 of nearly
0 mmHg.
The net effect of room air is to raise control PO2 values and depress the PCO2 value. As carbon
dioxide is lost, the pH will begin to rise. Eventually the pH rise will reduce ionized calcium.
Storage Temperature:
Recovery of gas values (and to some extent pH results) in aqueous controls is affected by their
temperature. The area where the controls are stored prior to analysis must be temperature
controlled. Stat Profile Ultra controls are manufactured for use at 25 C. Your results will be
noticeably different if the temperature of the control is significantly above or below this
reference point. Therefore, DO NOT store controls on top of the analyzer, near a heating or airconditioning duct, radiator, window, etc.
Oxygen shows the effects of temperature most clearly. For example, the recovered PO2 will
vary 1% for every degree centigrade change away from 25 C. The change will be in the
OPPOSITE direction from the temperature shift. As the temperature goes up, the PO2 falls, and
vice versa, since the solubility of gases increases as the temperature of the solvent decreases.
The pH FOLLOWS temperature. In an extremely cool room (like some O.R.s) pH will recover
low compared to the assay range.
Variation in temperature can account for poor precision in your quality assurance results versus
peer group. Equilibrium of control to a stable temperature will eliminate this problem.
5-87
5. T. Shoot
Once opened, an ampule of control material must be analyzed immediately. In less than 1 minute,
equilibrium with room air begins to affect PO2, PCO2, pH, and ionized calcium results.
5. T. Shoot
Altitude contributes to the recovery of gas results (most notably oxygen) in controls. Stat
Profile Ultra controls are assayed at sea level. Therefore, recovery of PO2 will decrease as
elevation increases. This is generally not noticeable until elevation exceeds 3000 feet, but a
convenient rule of thumb is to expect a 0.5% decrease in recovered oxygen for every 1000 feet
of elevation. If your lab is above 3000 feet, you should take this factor into consideration when
evaluating PO2 performance.
Similarly, if an incorrect barometric pressure has been entered on the analyzer, it can
compromise both control and patient blood gas results. This can be easily checked by comparing
the barometric pressure that appears on the printout or the Status screen with the analyzers
internal barometric output. To check the barometric output first press the MENU key then
select the Service screen (#3). From the Service screen, select Analog Input Subsystem.
Channel #13 is the barometer output. If the reading from this channel is significantly different
from the barometric pressure displayed on the Status screen, an offset has probably been
entered by an operator. Before correcting the value (under the Operation Menu selection),
check the barometric pressure with a reference such as a mercury barometer.
5-88
Several electrodes (pH, Na+, K+, Ca++, Cl-) depend on the common reference
electrode. If several of these analytes are simultaneously out of control, clean and
inspect the reference electrode. Also confirm that reference (R) solution is being
delivered to the reference electrode. Poor reference solution flow can cause
calibration problems with all of the electrodes mentioned, but not usually all of
them simultaneously. It is common to see 1 or 2 on this list involved at first, and
5 Troubleshooting
2.
3.
4.
5.
if the situation is not corrected, other electrodes will go in and out of calibration or
control.
Recall the function of the reagents in the pack: Standards A and B are for pH;
Standards C and D are for electrolytes/glucose/lactate/BUN; B and Flush are for
iCa. Determine whether problems arose upon installation of a new reagent pack,
or if some other component was changed which could explain the problem.
If one or both of the gas channels is involved, check for leaks in the gas path (Gas
Test). Look for a loose connection to the preheater, loose humidifier caps, etc.
Check for incorrectly entered calibration gas values which may be causing the
problem.
Unimpeded flow of reagents, gases, and specimens through the analyzer is critical.
Check and track calibration and analysis flow rates. If they are changing rapidly or
are near the extremes of their ranges, this is a signal for preemptive maintenance
(usually tubing replacement or cleaning).
1.
2.
3.
4.
5.
6.
7.
The life of an electrode can vary significantly with each analyte. Check the
electrode warranty information or call Nova if you have a question.
Check how much time has elapsed or how many samples have been run using the
sensor since maintenance was last performed. Refer to Chapter 4, Maintenance, for
corrective maintenance steps involving individual parameters. If the suggestive
maintenance procedures fail to correct the problem, contact Nova for advice.
Any sensor close to the limits of its slope range or a sensor that is exhibiting unstable
slope (varies greatly from calibration to calibration) indicates a need for maintenance.
Foreign material present in the flow path will adversely impact performance.
Inspect the flow path for evidence of micro-clots, fibrin, bacterial growth, or
coating on the sensors by other materials such as the dyes present in some controls.
Check or change the washer in use with the sensor. Confirm that it is the correct
thickness and diameter. Replace any washer that is torn, has a ragged edge, or is
misshaped.
A newly remembraned gas electrode MUST be conditioned with whole blood and
must be allowed time for temperature equilibration before calibrating and running
controls. ALWAYS TAKE THESE STEPS.
The use of offsets to correlate the Stat Profile Ultra analyzer with some other
instrument in the laboratory will also affect QC ranges. If you use offsets, you must
either recalculate the QC ranges to take this into account or designate every control
sample as a QC Sample Type on the Patient Data screen.
5-89
5. T. Shoot
Troubleshooting a single parameter failure uses a different approach. A number of factors need
to be considered.
5. T. Shoot
Whole blood is significantly more viscous than the aqueous solutions used for analyzer
calibration and QC. The higher the patients hematocrit, the more viscous the sample, and the
slower the flow through the analyzer. This is why the range for analysis flow is greater than
that allowed for calibration. Flow times will change with every sample.
To understand how this can impact results, consider a situation where a tiny clot or
accumulation of fibrin occurs around the tip of an electrode. Calibration may still be successful,
and reasonable aqueous QC results can be obtained if the blockage is not severe. Whole blood
results may be first to reflect a disturbance of this type. To avoid this type of problem, make
visual inspection of the flow cell a routine part of maintenance.
Matrix Effects/Interferences:
Nova instruments are designed for clinical environments to analyze actual patient specimens,
not modified blood samples. Specimens removed from the patient, anticoagulated appropriately, and promptly analyzed are the only type of sample for which where the measurement
results will be reliable. Matrix effects/interferences can occur when patient specimens are
removed from the body, modified and then measured on a Nova instrument. For example,
matrix effects have been seen on Nova analyzers when attempting to analyze samples collected
from cell savers used in various surgical procedures. Also, evaluation laboratories run
specimens from patients with a wide variety of pathologies and from patients who are being
treated with a broad spectrum of therapeutic and pharmacological agents. Despite extensive
clinical trials, it is not possible to anticipate every possible combination of transfused blood
products, crystalloids, and drugs (or their metabolites) that may be present in a blood sample.
As a result, some users have found that their particular patient mix has necessitated making
adjustments to maintenance. For example, a high number of cardiopulmonary bypass pump or
ECMO (extracorporeal membrane oxygenation) samples result in a need for increased
analyzer maintenance. If you are experiencing excessive downtime, you may need to modify
your own maintenance schedules. Novas Clinical Applications Group will assist you in
tailoring a maintenance program to meet these needs.
5-90
5 Troubleshooting
5.2.5 Sample Handling
Contamination by Room Air:
As previously noted, room air has a PO2 of about 150 mmHg and PCO2 of nearly 0 mmHg.
Should a large air bubble or several smaller ones remain in an ABG collection syringe for any
length of time, the sample blood gases will equilibrate with the air bubble. The patients gas
values may be raised or lowered, and the measured values will not be representative of the
patients in vivo state.
Delays in Transit:
1.
2.
3.
For all the above reasons, it is always preferable to analyze the specimen quickly. If a delay
cannot be avoided, icing of the syringe is the next best choice.
Amount and Type of Anticoagulant Used:
For whole blood and plasma analysis, sodium heparin at a level of 15 to 20 I.U. per mL of blood
collected is optimal for most patient populations. Some patient subpopulations may require
slightly more heparin (such as neonatal capillary samples), and some may require slightly less
(preheparinized blood in cardiopulmonary bypass). At levels much higher or lower than
suggested above, the sample and its results may be compromised.
Too little heparin results in a variety of problems from buildup of protein on sensors to microclots in the system to outright blockage of the flow path. This has been observed in O.R.s
where the patients blood has been partially heparinized for cardiopulmonary bypass. These
samples must still be collected with heparinized syringes to avoid clotting problems.
Too much heparin will bind ionized calcium in the patients blood and lead to falsely low
results. This is commonly seen when a high dose of lyophilized heparin is used in an ABG
syringe, and the syringe is subsequently under filled.
If preheparinized syringes are in use, select the least amount of heparin necessary and make
sure that all staff members have been educated in the correct fill volume, as well as mixing and
capping techniques.
5-91
5. T. Shoot
4.
Glucose will drop by about 5% per hour when a heparinized whole blood specimen
is stored at room temperature.
Potassium can increase by as much as 0.2 mmol/L in less than an hour when an
ABG specimen is held on ice.
Oxygen, carbon dioxide, and pH are all susceptible to changes based on room air
contact which can diffuse through the walls of a plastic syringe or even to
metabolism taking place within the specimen with long delays.
Red cells will settle out of the plasma as a function of time. In many disease states
this process is accelerated. For accurate measurement of hematocrit, it is imperative that the ABG syringe be carefully and thoroughly mixed prior to analysis.
5. T. Shoot
5-92
5 Troubleshooting
The message, Clean Sample Preheater, will clear from the Ready for Analysis screen only
following a full 2-point calibration sequence (successful). If the message recurs, clean the
preheater with deproteinizing solution (see Section 4.1.5.2).
5.3
1 System Test
2 Analog Input Signal Data
3 Analysis/Calibration Signal Data
4 Temperature Control Signal Data
5 System Error Log
6 Fluid/Gas Prime Menu
7 I/O Devices Menu
11:45:45
5-93
5. T. Shoot
1.
2.
3.
From the Service Menu, press 1 to access the System Test #1 screen. The System
Test #1 screen displays multiple millivolt readings for electrodes and air detectors
at the top of the screen, in addition to displaying the system test functions.
Version 123
SYSTEM TEST #1
Revision 123.456
pH
+ 047.43
PCO2
-082.12
PO2
-0051.20
1
2
3
4
5
6
7
8
9
0
Na+
+008.56
K+
+017.35
Ca++/Cl- +074.43
Millivolt Display
PCO2 Membrane
O2 Gain
Air Oscillators
Pump
Sampler Position
Valve Position
Gas
Pump Bypass Valve
SO2% LEDs
Multiple
Off
Low
Off
0-Off
0
1
0-Off
Closed
Off
5-94
Glu
Lac
BUN
+042.32
-023.23
-047.76
5 Troubleshooting
2.
To monitor a single analog channel continuously, the System Test screen exists in
a second form. The System Test #2 screen, which is accessed from the System Test
#1 screen, displays mV readings from 1 of 32 channels. Press 1 to change the
millivolt display to the System Test #2 screen, as shown below.
To move to higher channel numbers, press ENTER repeatedly. To move to lower
channel numbers, press the dash (-) repeatedly.
Version 123
SYSTEM TEST # 2
Revision 123.456
Channel 1
- 008.61 mV
Millivolt Display
PCO2 Membrane
O2 Gain
Air Oscillators
Pump
Sampler Position
Valve Position
Gas
Pump Bypass Valve
SO2% LEDs
Multiple
Off
Low
Off
0-Off
0
1
0-Off
Closed
Off
3.
4.
5-95
5. T. Shoot
Shorting
Jumper
R
Cl
Na
K
BUN
Glu
Lac
pH
PCO2
PO2
SO2
d
e
e
REF
Note:
Electrode cables and
tubing are
not shown here
for clarity.
K
Na
Ca
BUN
Glu
Lac
pH
PCO2
PO2
SO2
PCO2
REF
PCO2
PO2
PO2
Glu
Glu
Electrode
Rack
Assembly
Electrode
Rack
Assembly
S
DWG #7-J-002A
5. T. Shoot
DWG #7-F-003A
CAUTION: Do not plug the shorting jumper into the PO2, glucose, or lactate
sockets.
3.
4.
5.
6.
7.
Plug the Shorting Jumper (PN 09931) into all sockets (excluding PO2 , glucose, and
lactate) in the rack as shown in Figure 5.6.
From the System Test #1 screen (press MENU, 5, 1), press CALIBRATE. Ignore
all error codes, except for instability codes. If any of these error codes are displayed,
contact Nova Technical Services.
The millivolt readings at the top of the screen should stay between -2.0 and
+2.0 mV for Na +, K + , Cl -, iCa, BUN, pH, and P CO 2. If any readings are
outside this range, contact Nova Technical Services.
Unplug the shorting jumpers. Connect the electrode cables to the rack.
Wait 5 minutes then press CALIBRATE. After the first calibration is finished,
press CALIBRATE again.
5-96
5 Troubleshooting
1.
2.
3.
4.
5.
0 - Off
1 - Low
2 - MLo
3 - MHi
4 - High
2.
3.
4.
5.
6.
7.
5-97
5. T. Shoot
From the System Test #1 screen (press MENU, 5, 1), press 5 to display all 5 speeds
next to the pump function. (The current state of the pump is shown before pressing
5.) These speeds are:
PROBE POSITIONS
5. T. Shoot
DWG #5-0-048A
SEPTUM
ASSEMBLY
GAS
FLUSH
STD C
STD D
STD A
STD B
5
6
SEPTUM
AIR
AIR
SAMPLE
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
5-98
From the System Test #1 screen (press MENU, 5, 1), press 6 to display all 10
options next to the Sampler function. (The current state of the sample probe is
shown before pressing 6.)
Press 0. The sample probe does not move (Gas position).
Press 6, 1. The sample probe moves DOWN to position 1 (Flush).
Press 6, 2. The sample probe moves DOWN to position 2 (Std./C).
Press 6, 3. The sample probe moves DOWN to position 3 (Std./D).
Press 6, 4. The sample probe moves DOWN to position 4 (Std. A).
Press 6, 5. The sample probe moves DOWN to position 5 (Std. B).
Press 6, 6. The sample probe moves DOWN to position 6. (Probe tip is buried in
the bottom septum of the septum assembly.)
Press 6, 7. The sample probe moves DOWN to position 7. (Probe tip is just visible
in the depression at the bottom of the septum assembly.)
Press 6, 8. The sample probe moves DOWN to position 8. (Probe tip extends
1/8 inch (3mm) out of septum retainer.)
Press 6, 9. The sample probe moves DOWN to position 9. (Probe tip extends
11/2 inch (38 mm) out of septum assembly for sample aspiration.)
Press 6, 0. The sample probe moves UP to position 0, the home position.
Perform the Probe Septum Assembly Placement Test as follows.
5 Troubleshooting
Probe Placement Test
Proper probe position in the septum assembly is critical for error-free reagent delivery and
analysis. Verify proper probe position as follows:
1.
From the System Test #1 screen (press MENU, 5, 1), press 6, 9 to move the probe
to position 9.
CAUTION: Do not open or close the door when the probe is in position 9.
Probe damage will occur.
2.
Slide the probe adjustment tool (See Figure 5.8) over the probe and hold it firmly
to the base of the sampler assembly so that the probe tip extends toward the stepped
end. The probe tip should be flush with the beginning of the step. If the probe tip
is not flush, adjust as follows.
Loosen the screw that holds the sample probe adjustment plate to the sampler
assembly. See Figure 5.8.
5. T. Shoot
DWG #5-0-026B
5-99
7.
Move the adjustment plate as needed to position the sample probe flush with the
beginning of the adjustment step. See Figure 5.8.
Tighten the adjustment plate screw to lock the sample probe into position.
Verify that the probe tip is properly adjusted by moving the sampler position to 0
then back to position 9. Use the sample probe adjustment probe to verify the correct
positioning of the sample probe. Repeat the procedure if positioning is not correct.
Press 6, 0 to return the probe to the home position.
1.
From the System Test #1 screen (press MENU, 5, 1), press 7 to display the 8 valve
options. See Figure 5.9.
Position 1
Position 2
Position 3
B
Position 4
1
49A
DWG
#5-0-0
2.
3.
4.
5.
6.
5-100
Press 1. The motor does not move, and the valve stays in position 1.
Press 7, 2. The motor moves in the forward direction, moving the valve to position 2.
Press 7, 3. The motor moves in the forward direction, moving the valve to position 3.
Press 7, 4. The motor moves in the forward direction, moving the valve to position 4.
Press 7, 5. The motor moves in the reverse direction, moving the valve to position 1.
5 Troubleshooting
7.
8.
9.
10.
Press 7, 6. The motor moves in the reverse direction, moving the valve to position 2.
Press 7, 7. The motor moves in the reverse direction, moving the valve to position 3.
Press 7, 8. The motor moves in the reverse direction, moving the valve to position 4.
Press 7, 1. The motor moves in the forward direction, moving the valve to position 1,
the idle position.
CAUTION: The pump bypass valve must always be opened when the gas is
manually turned on (Step 2 below and see Section 5.3.1.8) for gas to flow.
Press 9 to open the pump bypass valve.
Press 8, displaying the 7 gas choices.
Press 0. The gas is off.
Press 8, 1. Gas A bubbles through the humidifier at a slow rate.
Press 8, 2. Gas A bubbles through the humidifier at a medium rate.
Press 8, 3. Gas A bubbles through the humidifier at a fast rate.
Press 8, 4. Gas B bubbles through the humidifier at a slow rate.
Press 8, 5. Gas B bubbles through the humidifier at a medium rate.
Press 8, 6. Gas B bubbles through the humidifier at a fast rate.
Press 8, 0 to shut off the gas.
Press 9 to close the pump bypass valve.
5. T. Shoot
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
2.
3.
From the System Test #1 screen (press MENU, 5, 1), press 9 while observing the
bypass valve pinch bar. The pinch bar should retract, opening the W-bypass tubing
segment. See Figure 5.10
Press 6, 9 to extend the sample probe.
Pull back on the plunger of a syringe fitted with a length of tubing and attach the
tubing to the sample probe tip.
5-101
Push air through the flow path. If air does not flow freely, remove the tubing from
the probe cleaning syringe, disconnect the W-line from the reference electrode, and
repeat Steps 3 and 4 with the probe cleaning syringe attached to the W-tubing. If
the air now flows freely, clean the flow path per Section 4.1.5. If the air did not flow
freely, contact Nova Technical Services for assistance.
5. T. Shoot
DWG #5-M-020B
5.
6.
Press 9 again to extend the pinch bar and pinch the W-bypass tubing closed. If the
valve does not function properly, contact Nova Technical Services for assistance.
Press 6, 0 to return the probe to the home position.
3.
4.
5-102
Press MENU, 5, 1, 6, 0 to enter system test and move the probe to the air position.
Open the door and disconnect the R-line from the reference electrode. Hold the
R-line over a piece of gauze or a sample cup to catch any solution that may flow
in the following steps.
Press 5, 4 to operate the pump at high speed.
Reference solution should flow from the R-line tubing at a minimum of 4 drops per
10 seconds.
5 Troubleshooting
5.
6.
7.
From the Service Menu, Press 2 to display the Analog Input screen.
Press CLEAR to return to the Ready For Analysis screen.
ANALOG INPUT
0
1
2
3
4
5
6
7
8
9
10
mV
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
Chan
11
12
13
14
15
16
17
18
19
20
21
mV
Chan
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
22
23
24
25
26
27
28
29
30
31
5. T. Shoot
Chan
mV
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
11:45:45
5-103
From the Service Menu, Press 3 to display the Analysis/Calibration Signal Data
screens.
Press CLEAR, CLEAR, CLEAR to return to the Ready For Analysis screen.
5. T. Shoot
Stability of the PO2, PCO2, and pH electrode readings during analysis and
calibration as shown by the Stability (Stab) Factors. Significant increases
in the factors (towards the high single numbers) indicate that electrode
maintenance is needed.
Millivolt readings for the last calibration and the last analysis.
Drift from the calibration value.
Results (Conc) of the last analysis (not temperature corrected). In the event
of a major error, concentration is displayed in reverse video here, but not
displayed on the results screen. Extreme caution is advised should the
choice be made to utilize these results.
Slope Performance Numbers for the last calibration. The Slope Performance Limits are listed in Table 5.4.
pH
9.1 - 11.2
PCO2
7.9 - 12.2
PO2
(-15.0) - (-1.6)
This screen is useful for diagnosing a poorly performing electrode. Access this screen as
follows:
1.
2.
5-104
From the Service Menu, press 3. Then continuously press ENTER until the Blood
Gases and pH screen appears.
Press CLEAR to leave this screen.
5 Troubleshooting
BLOOD GASES & pH
Gas B/Std B
Gas A/Std A/Std C
mVStabmVStab Slope
pH
-21.6
0.4
-53.3
0.4
+10.1
pH
-53.3
0.4
+10.1
+127.6
0.4
+111.3
1.3
+9.3
PCO2
PO2
-4.0
0.0
-224.1
0.0
-4.0
Sample
Gas A/Std C
mVStabmVStabConc
pH
-64.7
0.4
-57.1
0.4
7.568
PCO2
-91.9
0.9
+110.5
8.3
18.8
-228.4
0.0
-231.5
0.0
155.1
PO2
Press ENTER for Next Screen.
Press CLEAR to Exit.
30 Apr 95
11:45:45
This screen is useful for diagnosing a poorly performing electrode and allows you to check:
Stability of Na+, K+, Cl-, and iCa electrode readings for analysis and
calibration as shown by the Stability (Stab) Factors. Significant increases
in the factors (towards the high single numbers) indicate that electrode
maintenance is needed.
Millivolt readings for the last calibration and the last analysis.
Drift from the calibration value.
Results (Conc) of the last analysis. In the event of a major error, concentration is displayed in reverse video here, but not displayed on the results
screen. Extreme caution is advised should the choice be made to utilize
these results.
Slope Performance Numbers for the last calibration. The Slope Performance Limits are listed in Table 5.5.
Table 5.5 Electrolyte Slope Limits
Na+
K+
Ca++
Cl-
8.8
8.6
8.9
7.7
10.7
10.9
11.6
12.2
5-105
5. T. Shoot
From the Service Menu, press 3. Then continuously press ENTER until the
Electrolytes screen appears.
Press CLEAR to exit this screen.
ELECTROLYTES
Std D/Std B
Std C/Flush
mVStabmVStabSlope
Na+
-180.3
2.9
-165.5
1.8
+9.2
+42.1
0.4
+0.6
0.0
+10.2
K+
Cl-/Ca++ +0.6
0.1
-8.5
0.3
+10.4
Sample
Std C/Flush
mVStabmVStabConc
Na+
-166.3
2.5
-160.5
0.2
133.0
K+
+0.2
0.1
+0.3
0.3
3.94
Cl-/Ca++ -5.5
0.2
-8.5
0.2
1.27
5. T. Shoot
Electrolytes Screen
5-106
Stability of glucose, lactate, and BUN electrode readings for analysis and calibration as shown by the Stability (Stab) Factors. Significant increases in the factors
(towards the high single numbers) indicate that electrode maintenance is needed.
Millivolt readings for the last calibration and the last analysis.
Drift from the calibration value.
Results (Conc) of the last analysis. In the event of a major error, concentration is
displayed in reverse video here, but not displayed on the results screen. Extreme
caution is advised should the choice be made to utilize these results.
Slopes from the last calibration. The slope limits are listed in Table 5.6.
5 Troubleshooting
Table 5.6 Glucose , Lactate, and BUN Slope Limits
Glucose
5.0
45.0
Lactate
6.0
100.0
BUN
8.9
12.2
From the Service Menu, press 3. Then continuously press ENTER until the
Metabolites screen appears.
Press CLEAR to exit this screen.
METABOLITES
5. T. Shoot
Std D
Std C/Std A
mVStabmVStabSlope
Glu
+33.9
0.0
+12.6
0.5
+17.8
Lac
+39.9
0.1
+14.8
0.3
+41.7
Lac
+70.2
0.6
BUN
-9.0
0.8
-39.1
0.2
+10.6
BUN
+3.3
0.8
Sample
Std C
mVStabmVStabConc
Glu
+13.7
0.0
+12.4
0.4
86.4
Lac
+5.8
0.1
+13.4
0.0
0.7
BUN
-37.7
0.1
-39.1
0.3
10.6
Press ENTER for Next Screen.
Press CLEAR to Exit.
30 Apr 94
11:45:45
Metabolites Screen
5-107
Slope Performance Numbers for the last calibration. The Slope Performance Limits are listed in Table 5.7.
Hct
40.0
98.0
From the Service Menu, press 5. Then continuously press ENTER until the Air
Detectors & Hct screen appears.
Press CLEAR to exit this screen.
AIR DETECTORS & Hct
5. T. Shoot
Std BStd C
AD2
+196.5
+126.4
AD3
+314.3
+273.0
mV
+70.1
+41.3
Std C
Std B
mVStabmVStabSlope
Hct
+314.3
0.4
+273.0
0.1
+66.4
Sample
Std C
mVStabmVStab Conc
Hct
+270.2
0.0
+272.1
0.5
48
11:45:45
5-108
5 Troubleshooting
5.3.3.5 Oxygen Saturation
This screen is useful for diagnosing a poorly performing SO2% and allowing you to check:
From the Service Menu, press 3. Then continuously press ENTER until the Oxygen
Saturation screen appears.
Press CLEAR to exit this screen.
OXYGEN SATURATION
Std 1
mVStab
LED 1 +123.4
12.3
LED 3 +123.4
12.3
LED 1
LED 3
Dark
mVStab
+123.4
12.3
+123.4
12.3
5. T. Shoot
Flush
mVStab
LED 1 +123.4
12.3
LED 3 +123.4
12.3
Std 2
mVStabSlope
+123.4
12.3
+12.3
+123.4
12.3
+12.3
Sample
mVStabConc
+123.4
12.3
123.4
+123.4
12.3
5-109
From the Service Menu, press 4, Temperature Control Signal Data, to display the
Temperature Control screen.
Press CLEAR to leave this screen.
TEMPERATURE CONTROL
AB BLR MTR
SP REMOVED
Temperature
Air Bath
37.0 C
Preheater
37.1 C
Thermistor
Current
+0004.30 mV
+ 566.68 mV
+0080.10 mV
+ 50.99 mV
DAC
5. T. Shoot
DOOR OPEN
Controller
119
On
Off
134
On
Off
30 Apr 95
11:45:45
5-110
5 Troubleshooting
5.3.5 Fluid/Gas Prime
This screen allows you to prime all the solutions and gases individually. Also the flow path can
be purged with this screen displayed. Display the screen as follows:
1.
2.
3.
From the Service Menu, press 6 to display the Fluid/Gas Prime Options screen.
Press the number corresponding to the solution or gas that you want to prime or
pump or press 8 to purge the flow path.
Press CLEAR to exit this screen.
Prime
Prime
Prime
Prime
Prime
Prime
Prime
Purge
Flush
Std. A
Std. B
Std. C
Std. D
Gas A
Gas B
Flow Path
5. T. Shoot
1
2
3
4
5
6
7
8
30 Apr 95
11:45:45
From the Service Menu, press 7 to display the I/O Devices Menu.
Press 1 to 2 to transmit last results or calibration data.
To access options 3 through 7 see Sections 5.3.6.1 to 5.3.6.4.
Press CLEAR to exit the I/O Devices Menu.
5-111
4
5
6
7
Printer Subsystem
External Parallel Port
Communications Tests
Disk Drive Utilities
30 Apr 95
11:45:45
5. T. Shoot
5.3.6.1
This screen is used primarily by Nova Service Technicians to test the bar code wand port. A
test fixture is necessary to test the port.
Access this screen as follows:
1.
From the Service Menu, press 7 to display the I/O Devices Menu.
2.
Press 3 to display the Bar Code Wand Port Test screen.
3.
If performing the test, install the test fixture onto the port and press ENTER.
If not performing the test, press CLEAR to exit the I/O Devices Menu.
BAR CODE WAND PORT TEST
30 Apr 95
11:45:45
5-112
5 Troubleshooting
BAR CODE WAND PORT TEST
Data
Timeouts
Parity
Framing
Overrun
Errors
1
0
0
103
196
30 Apr 95
11:45:45
Printer Subsystem
From the Service Menu, press 7 to display the I/O Devices Menu.
Press 4 to display the Printer Subsystem screen.
Press 1 to print a character set (prints all possible characters).
Press 2 to print system dump (yields all millivolt, slope, setup, and other system
parameters).
Press 3 to print the setup parameters (causes a printout of all setup parameters,
modes, and values).
Press 4 to print the system error log.
Press 5 to abort the printout (stops any printing operation).
Press 6 to enable or disable the printer.
Press CLEAR to exit the I/O Devices Menu.
5-113
5. T. Shoot
5.3.6.2
2
3
4
Cancel Print-out
Printer Enabled
30 Apr 95
11:45:45
5. T. Shoot
30 Apr 95
11:45:45
5-114
5 Troubleshooting
EXTERNAL PARALLEL PORT TEST
30 Apr 95
11:45:45
5. T. Shoot
Test results may be transmitted manually at any time after the analysis has been completed and
the results have been displayed. If the Manual Transmit mode has been selected in the Set Up
Menu, this is the only way to transmit results (see Section 2.2.1). The Transmitter Test and Endof-Line Delay Adjust functions are useful for setting up and troubleshooting communications
links. Note that once the End-of-Line Delay has been determined on this screen, the value
should be programmed into the system as a setup parameter. For more information on the
Communications Subsystem, see Section 6.
Access this screen as follows:
1.
2.
3.
4.
5.
From the I/O Devices Menu, press 6 to display the Communications Test screen.
Press 1 to select communication port test.
Press 2 to run a Transmitter Test, printing out an ASCII character sequence for
transmitter verification.
Press 3 to do a loop-back test.
Press CLEAR to leave this screen.
5-115
COM1
30 Apr 95
11:45:45
TRANSMITTER TEST
5-116
5 Troubleshooting
Press 3, to display the Digital Loop-Back Test screen.
DIGITAL LOOP-BACK TEST
30 Apr 95
11:45:45
5. T. Shoot
Errors
TXD
RXD
DSR
DTR
DCD
CTS
RTS
RI
123
123
123
123
123
30 Apr 95
11:45:45
5-117
Errors
TXD
RXD
DSR
DTR
DCD
CTS
RTS
RI
123
123
123
123
123
30 Apr 95
11:45:45
5. T. Shoot
5-118
5 Troubleshooting
DISK OPERATIONS
SYSTEM BUSY
Program Reload
30 Apr 95
11:45:45
30 Apr 95
11:45:45
6.
7.
5-119
5. T. Shoot
4.
5.
5.4
Review the setup parameters of the previous software from the printout.
Make the appropriate setup selections.
Calibrate the analyzer.
Upon completion of the QC setup, an optional printout of the setup parameters may
be made and saved for future reference. The software disk may remain in the
analyzer or stored in a secure location for emergency use.
Analyzer Reset
Can occur under some circumstances if the laboratory has a power supply
or grounding problem.
Can also be initiated by the key operator (if the lab has forgotten the setup
password or needs to solve a rare software-related condition).
The analyzer will lose all battery-protected setup selections and certain
operations menu selections as well.
All selected options revert to the factory defaults.
5. T. Shoot
5-120
6 Computer/Peripheral
6.0 Stat Profile Ultra Communications Interface
The following are guidelines for the implementation of the Nova Stat Profile Ultra Communications System to the ASTM E 1394-91 and ASTM E 1381-91 Standards.
6.1 Introduction
This chapter describes how the Ultra transmits data to an external computer. Data transmission
involves a low-level protocol and a high-level protocol.
The low-level protocol is concerned with establishing communication, detecting errors, and
sending and receiving messages. It is not concerned with message content.
The high-level protocol is concerned with message content.
The protocols used by Ultra are designed to conform to specifications published by the
American Society for Testing and Materials (ASTM). Copies of the specifications can be
obtained by contacting ASTM:
ASTM
100 Barr Harbor Drive
West Conshohocken, PA USA 19428-2959
USA
6. Computer
Phone:
FAX:
(610) 832-9585
(610) 832-9555
NOTE: You can configure the Ultra so that ACK/NAK responses are not
required. When Ultra is configured in this way, the frame structure conforms to ASTM E1381-91, but the low-level protocol as a whole does not.
The Ultra can receive only ACK and NAK responses from the external computer; it cannot
receive messages from the external computer. In other words, the Ultra is always the sender.
In the description of the low-level protocol, italic numbers in parentheses refer to sections of
the ASTM E1381-91 specification. For example, (6.5.2.4) means that more information can
be found in section 6.5.2.4 of ASTM E1381-91.
6-1
IDLE State
TRANSFER State
6. Computer
6-2
6 Computer/Peripheral
6.2.1.3 Transfer State
During the Transfer State, the Ultra transmits messages to the receiver. The Ultra continues in
this state until all messages are transmitted or an abnormal exit condition arises.
The Transfer State is entered by both instruments after a successful link bid occurs. After
entering this state, the Ultra has 30 seconds to send the first frame. If no frame or <EOT> is
received within the specified time period, the receiver should proceed to the Link Termination
State. (6.5.2.4)
After receiving a frame, the receiver has 15 seconds to respond with an acknowledgment
control character (ACK/NAK). If the receiver responds with a <NAK> control character, Ultra
will transmit the same frame again within a 30 second time period. Ultra will transmit the same
frame 6 times before giving up and entering the Link Termination State. If the receiver
responds with an <ACK> control character, the next frame should be expected to arrive within
a 30 second time period. (6.5.2.3, 6.5.2.4)
Messages are sent in frames. Frames are described in sections 6.3.1 through 6.3.3.3 of
ASTM E1381-91. Ultra does not transmit messages of more than 240 characters; this means
that Ultra does not use the <ETB> transmission control character.
Frame acknowledgment is described in sections 6.3.4 through 6.3.4.4 of ASTM E1381-91.
Ultra does not honor receiver interrupt requests. (6.3.4.4 ,6.3.5.3)
6-3
6. Computer
The Link Termination State returns the data link to the Idle State. The Ultra transmits an
<EOT> character to notify the receiver that all frames have been sent and the session is
terminated. Upon reception of the <EOT> character, the receiver should immediately enter the
Idle State.
6. Computer
ULTRA IMPLEMENTATION
7.1.1
Record Type ID
Single character: H
7.1.2
Delimiter Definition
7.1.3
Message Control ID
Not Used
7.1.4
Access Password
Not used
7.1.5
Sender Name or ID
NOVA^ULTRA x^vv..vv^dd
Where x is A for Ultra A, B for Ultra B,
etc., vv..vv is version of Ultra software, and
dd is an Analyzer Identification # set by the
Ultra operator.
NOTE: For early versions of Ultra C, "ULTRA11"
will be sent instead of ULTRA C.
7.1.6
Not used
7.1.7
Reserved
Not used
7.1.8
Not used
7.1.9
Characteristics of sender
Not used
7.1.10
Receiver ID
Not used
7.1.11
Not used
7.1.12
Processing ID
Not used
7.1.13
Version No.
7.1.14
Date and time at which message was transmitted. (This is not the time of the analysis or
calibration.)
6-4
6 Computer/Peripheral
6.3.2 Patient Information Record
ULTRA IMPLEMENTATION
8.1.1
Record Type
Single character: P
8.1.2
Sequence Number
8.1.3
Not used
8.1.4
Patient ID # if available; otherwise, blank. (Patient ID # is available if and only if it was entered
by Ultra operator.) For cals, this will be blank.
For QC samples, this will be the lot number of
the control material. NOTE: This is the last field
that is used in this record.
8.1.5
Patient ID No. 3
Not used
8.1.6
Patient Name
Not used
8.1.7
Not used
8.1.8
Birthdate
Not used
8.1.9
Patient Sex
Not used
8.1.10
Not used
8.1.11
Patient Address
Not used
8.1.12
Reserved Field
Not used
8.1.13
Not used
8.1.14
Attending Physician ID
Not used
8.1.15
Special Field 1
Not used
6. Computer
6-5
6. Computer
ULTRA IMPLEMENTATION
8.1.16
Special Field 2
Not used
8.1.17
Patient Height
Not used
8.1.18
Patient Weight
Not used
8.1.19
8.1.20
Not used
8.1.21
Patients Diet
Not used
8.1.22
Not used
8.1.23
Not used
8.1.24
Not used
8.1.25
Admission Status
Not used
8.1.26
Location
Not used
8.1.27
Not used
8.1.28
Not used
8.1.30
Marital Status
Not used
8.1.31
Isolation Status
Not used
8.1.32
Language
Not used
8.1.33
Hospital Service
Not used
8.1.34
Hospital Institution
Not used
8.1.35
Dosage Category
Not used
6-6
6 Computer/Peripheral
6.3.3 Test Order Record
ULTRA IMPLEMENTATION
9.4.1
Record Type ID
Single character: O
9.4.2
Sequence Number
9.4.3
Specimen ID
9.4.4
Instrument Specimen ID
9.4.5
Universal Test ID
Not used
9.4.6
Priority
Not used
9.4.7
Not used
9.4.8
6-7
6. Computer
6. Computer
ULTRA IMPLEMENTATION
9.4.9
Not used
9.4.10
Collection Volume
Not used
9.4.11
Collector ID
Not used
9.4.12
Action Code
Not used
9.4.13
Danger Code
Not used
9.4.14
Not used
9.4.15
Not used
9.4.16
Specimen Descriptor
9.4.17
Ordering Physician
Not used
9.4.18
Not used
9.4.19
Not used
9.4.20
Not used
9.4.21
Not used
9.4.22
Not used
9.4.23
9.4.24
9.4.25
Instrument Section ID
6-8
Not used
6 Computer/Peripheral
ASTM REF ASTM NAME
ULTRA IMPLEMENTATION
9.4.26
Report Types
9.4.27
Reserved Field
Not used
9.4.28
9.4.29
Not used
9.4.30
Specimen Service
Not used
9.4.31
Specimen Institution
Not used
6. Computer
ULTRA IMPLEMENTATION
10.1.1
Record Type
Single character: R
10.1.2
Sequence Number
10.1.3
Universal Test ID
6-9
6. Computer
ULTRA IMPLEMENTATION
10.1.4
10.1.5
Units
Abbreviation of units.
10.1.6
Reference Ranges
Not used
10.1.7
6-10
6 Computer/Peripheral
ULTRA IMPLEMENTATION
10.1.8
Not used
10.1.9
Result Status
10.1.10
Not used
10.1.11
Operator Identification
User ID#
NOTE: If more than one result is sent for an order
(as will typically be the case), this field will be
used only in the first result record.
10.1.12
10.1.13
Not used
10.1.14
Instrument Identification
Not used
6-11
6. Computer
6. Computer
Comment Record(s) will immediately follow a Test of Record if and only if there were any
errors during sample analysis. There will be one Comment record for each error.
ASTM REF ASTM NAME
ULTRA IMPLEMENTATION
11.1.1
Record Type
Single character: C
11.1.2
Sequence Number
11.1.3
Comment Source
Single character: I
11.1.4
Comment Text
11.1.5
Comment Type
Single character: I
ULTRA IMPLEMENTATION
13.1.1
Record Type
Single character: L
13.1.2
Sequence Number
13.1.3
Termination Code
6-12
6 Computer/Peripheral
6.4 Parameter Names
The parameter names listed here are used as Manufacturers or Local Code (see ASTM E139491 sec. 6.6.1.4). They are used to create the Universal Test ID field of the result record (see
Result Record above and see ASTM E1394-91 sec. 10.1.3).
Only a subset of these parameters will be transmitted. No single Ultra model reports all of the
parameters listed here. Also, during setup the operator can disable transmission of any or all
of the following:
Description
Na+
mmol/L
Sodium Concentration
Na+_M1
mV
Sodium Millivolts #1
Na+_M2
mV
Sodium Millivolts #2
Na+_SL
None
Sodium Slope
Na+_D1
None
Sodium Drift #1
Na+_D2
None
Sodium Drift #2
Na+_V1
None
Na+_V2
None
K+
mmol/L
Potassium Concentration
K+_M1
mV
Potassium Millivolts #1
6. Computer
Parameter Name
6-13
6. Computer
Units
Description
K+_M2
mV
Potassium Millivolts #2
K+_SL
None
Potassium Slope
K+_D1
None
Potassium Drift #1
K+_D2
None
Potassium Drift #2
K+_V1
None
K+_V2
None
Cl-
mmol/L
Chloride Concentration
Cl-_M1
mV
Chloride Millivolts #1
Cl-_M2
mV
Chloride Millivolts #2
Cl-_SL
None
Chloride Slope
Cl-_D1
None
Chloride Drift #1
Cl-_D2
None
Chloride Drift #2
Cl-_V1
None
Cl-_V2
None
Ca++
mmol/L, mg/dL
nCa
mmol/L, mg/dL
Ca++_M1
mV
Ca++_M2
mV
Ca++_SL
None
Ca++_D1
None
6-14
6 Computer/Peripheral
Units
Description
Ca++_D2
None
Ca++_V1
None
Ca++_V2
None
Mg++
mmol/L, mg/dL
nMg
mmol/L, mg/dL
Mg++_M1
mV
Mg++_M2
mV
Mg++_M3
mV
Mg++_SL
None
Mg++_D1
None
Mg++_D2
None
Mg++_V1
None
Mg++_V2
None
Hct
Hematocrit (vol %)
Hct_M1
mV
Hematocrit Millivolts #1
Hct_M2
mV
Hematocrit Millivolts #2
Hct_SL
None
Hematocrit Slope
Hct_D1
None
Hematocrit Drift #1
Hct_D2
None
Hematocrit Drift #2
pH
None
pH
6-15
6. Computer
Parameter Name
6. Computer
Units
Description
pHTC
None
pH_M1
mV
pH Millivolts #1
pH_M2
mV
pH Millivolts #2
pH_M3
mV
pH Millivolts #3
pH_SL
None
pH Slope
pH_D1
None
pH Drift #1
pH_D2
None
pH Drift #2
pH_V1
None
pH_V2
None
PO2
mmHg, kPa
PO2
PO2TC
mmHg, kPa
PO2_M1
mV
PO2 Millivolts #1
PO2_M2
mV
PO2 Millivolts #2
PO2_SL
None
PO2 Slope
PO2_D1
None
PO2 Drift #1
PO2_D2
None
PO2 Drift #2
PO2_V1
None
PO2_V2
None
PCO2
mmHg, kPa
PCO2 Concentration
PCO2T
mmHg, kPa
Corrected PCO2
6-16
6 Computer/Peripheral
Units
Description
PCO2_M1
mV
PCO2 Millivolts #1
PCO2_M2
mV
PCO2 Millivolts #2
PCO2_SL
None
PCO2 Slope
PCO2_D1
None
PCO2 Drift #1
PCO2_D2
None
PCO2 Drift #2
PCO2_V1
None
PCO2_V2
None
GLU
mg/dL, mmol/L
Glucose Concentration
GLU_M1
mV
Glucose Millivolts #1
GLU_M2
mV
Glucose Millivolts #2
GLU_SL
None
Glucose Slope
GLU_D1
None
Glucose Drift #1
GLU_D2
None
Glucose Drift #2
GLU_V1
None
GLU_V2
None
LAC
mmol/L, mg/dL
Lactate Concentration
LAC_M1
mV
Lactate Millivolts #1
LAC_M2
mV
Lactate Millivolts #2
LAC_M3
mV
Lactate Millivolts #3
LAC_SL
None
Lactate Slope
6. Computer
Parameter Name
6-17
6. Computer
Units
Description
LAC_D1
None
Lactate Drift #1
LAC_D2
None
Lactate Drift #2
LAC_V1
None
LAC_V2
None
BUN
mg/dL, mmol/L
BUN Concentration
BUN_M1
mV
BUN Millivolts #1
BUN_M2
mV
BUN Millivolts #2
BUN_M3
mV
BUN Millivolts #3
BUN_SL
None
BUN Slope
BUN_D1
None
BUN Drift #1
BUN_D2
None
BUN Drift #2
Urea
mg/dL, mmol/L
Urea Concentration
Urea_M1
mV
Urea Millivolts #1
Urea_M2
mV
Urea Millivolts #2
Urea_M3
mV
Urea Millivolts #3
Urea_SL
None
Urea Slope
Urea_D1
None
Urea Drift #1
Urea_D2
None
Urea Drift #2
Creat
mg/dL, mol/L
Creatinine Concentration
Creat_M1
mV
Creatinine Millivolts #1
6-18
6 Computer/Peripheral
Units
Description
Creat_M2
mV
Creatinine Millivolts #2
Creat_SL
None
Creatinine Slope
Creat_D1
None
Creatinine Drift #1
Creat_D2
None
Creatinine Drift #2
SO2%
SO2 Percent
SO2%_M1
mV
SO2 Millivolts #1
SO2%_M2
mV
SO2 Millivolts #2
SO2%_SL
None
SO2 Slope
SO2%_D1
None
SO2 Drift #1
SO2%_D2
None
SO2 Drift #2
S1_M1
mV
S1_M2
mV
S1_M3
mV
S1_M4
mV
S1_M5
mV
S3_M1
mV
S3_M2
mV
S3_M3
mV
S3_M4
mV
S3_M5
mV
6. Computer
Parameter Name
6-19
6. Computer
Units
Description
FIO2
BP
mmHg
Barometric Pressure
TempM
deg C, deg F
Measurement Temperature
TempP
deg C, deg F
Patient Temperature
Hb
g/dL
Hemoglobin
BE-ecf
mmol/L
BE-b
mmol/L
SBC
mmol/L
Standard Bicarbonate
HCO3-
mmol/L
TCO2
mmol/L
Total CO2
O2Ct
Vol/100mL
Oxygen Content
mmHg, kPa
Alveolar Oxygen
a/A
None
AaDO2
mmHg, kPa
P50
mmHg
O2CAP
Vol/100mL
Oxygen Capacity
Osm
mOsm/kg
Osmolarity
RI
None
Respiratory Index
Gap(K)
None
PO2/FI
mmHg, kPa
PO2/FIO2
6-20
6 Computer/Peripheral
6.5 Examples
NOTE: The examples that follow were created by a Stat Profile Ultra C
Analyzer. Data sent by other Stat Profile Ultra analyzers are similar but not
identical to the data in these examples.
Example #1
This example includes both low level and high level protocol. It also shows <ACK> responses
received by the Ultra. The responses are shown in double square brackets ([[<ACK>]]). This
example is for an expired gas sample. Expired gas was chosen for this example because the
expired gas message is relatively short and a short message is desirable when observing the low
level protocol.
The following parameters apply to this example:
ANALYSIS OUPUT OPTIONS:
Analysis Transmit Mode:
Transmit Analysis mV Data:
Transmit Analysis Drift Data:
Transmit Analysis Errors:
Auto
Off
Off
On
6. Computer
INSTRUMENT:
ULTRA, software version 99.0.23
IDENTIFIERS ENTERED DURING SET UP:
Analyzer Identifier: 64
IDENTIFIERS ENTERED FOR SAMPLE:
Patient ID:
0987
Accession #: 00123
Sample Type: Expired Gas
OTHER IDENTIFIERS:
Sample #:
401
Operator ID#: 102
DATE and TIME:
Analyzed :
June 8, 1995 at 13:26:46
Transmitted : June 8, 1995 at 13:28:00
6-21
6. Computer
Example #2
This example shows only the characters of the high-level protocol records. The following
parameters apply to this example:
ANALYSIS OUPUT OPTIONS:
Analysis Transmit Mode:
Transmit Analysis mV Data:
Transmit Analysis Drift Data:
Transmit Analysis Errors:
Auto
Off
Off
On
INSTRUMENT:
ULTRA , software version 99.0.23
IDENTIFIERS ENTERED DURING SET UP:
Analyzer Identifier: 64
IDENTIFIERS ENTERED FOR SAMPLE:
Patient ID:
987214
Accession #: 550663
Sample Type: Arterial
Time Drawn: 13.05 (this is an operator-entered string)
6-22
6 Computer/Peripheral
OTHER IDENTIFIERS:
Sample #:
407
Operator ID#: 102
DATE and TIME:
Analyzed :
June 8, 1995 at 14:17:59
Transmitted : June 8, 1995 at 14:20:25
6. Computer
H|\^&| | | NOVA^ULTRA11^99.0.23^64||||||||1|19950608142025
P|1| | 987214
O|1|550663|407| | | | 13.05| | | | | | | | Arterial| | | | | | | | | F
C|1|I|64 Hct TOO LOW|I
C|2|I|9D Lac BACKGROUND HI|I
C|3|I|81 Mg++ OUT OF RANGE|I
R|1|^^^pHTC^T|7.164| | | | | F||102|19950608141759
R|2|^^^PCO2TC^T| 56.7|mmHg| | | | F
R|3|^^^PO2TC^T| 70.1|mmHg| | | | F
R|4|^^^SO2%^M| uc| | | | | F
R|5|^^^Hct^M|? 4.|%| | | | F
R|6|^^^Na+^M|158.|mmol/L| | | | F
R|7|^^^K+^M| 5.9|mmol/L| | | | F
R|8|^^^Ca++^M| 1.53|mmol/L| | | | F
R|9|^^^Mg++^M|? 0.05|mmol/L| | | | F
R|10|^^^GLU^M| uc|mg/dL| | | | F
R|11|^^^LAC^M|ERR|mmol/L| | | | F
R|12|^^^BE-ecf^C|- 8.3|mmol/L| | | | F
R|13|^^^BE-b^C|- 5.6|mmol/L| | | | F
R|14|^^^HCO3-^C|20.5|mmol/L| | | | F
R|15|^^^TCO2^C|22.2|mmol/L| | | | F
R|16|^^^A^C| 76.7|mmHg| | | | F
R|17|^^^AaDO2^C| 6.6|mmHg| | | | F
R|18|^^^a/A^C| 0.9| | | | | F
R|19|^^^Hb^C|? 1.3|g/dL| | | | F
R|20|^^^nCa^C| 1.35|mmol/L| | | | F
R|21|^^^nMg^C|? 0.05|mmol/L| | | | F
R|22|^^^BP^M|739.3|mmHg| | | | F
R|23|^^^TempP^E| 37.1|deg C
L|1|N
6-23
Auto
Off
On
On
INSTRUMENT:
ULTRA , software version 99.0.23
IDENTIFIERS ENTERED DURING SET UP:
Analyzer Identifier: 64
IDENTIFIERS ENTERED FOR SAMPLE:
Patient ID:
1473
Accession #: 5055223
Sample Type: Arterial
6. Computer
OTHER IDENTIFIERS:
Sample #:
408
Operator ID#: 102
DATE and TIME:
Analyzed :
June 8, 1995 at 14:21:55
Transmitted : June 8, 1995 at 14:24:20
6 Computer/Peripheral
R|5|^^^Hct^M|? 4.|%| | | | F
R|6|^^^Na+^M|157.|mmol/L| | | | F
R|7|^^^K+^M| 5.9|mmol/L| | | | F
R|8|^^^Ca++^M| 1.52|mmol/L| | | | F
R|9|^^^Mg++^M|? 0.05|mmol/L| | | | F
R|10|^^^GLU^M| uc|mg/dL| | | | F
R|11|^^^LAC^M|ERR|mmol/L| | | | F
R|12|^^^BE-ecf^C|- 7.9|mmol/L| | | | F
R|13|^^^BE-b^C|- 5.2|mmol/L| | | | F
R|14|^^^HCO3-^C|20.8|mmol/L| | | | F
R|15|^^^TCO2^C|22.6|mmol/L| | | | F
R|16|^^^A^C| 76.7|mmHg| | | | F
R|17|^^^a/A^C| 1.7| | | | | F
R|18|^^^Hb^C|? 1.2|g/dL| | | | F
R|19|^^^nCa^C| 1.35|mmol/L| | | | F
R|20|^^^nMg^C|? 0.05|mmol/L| | | | F
6. Computer
R|21|^^^BP^M|739.3|mmHg| | | | F
R|22|^^^TempP^E| 37.0|deg C
R|23|^^^pH_D1|-0.006|mV
R|24|^^^Na+_D1|- 0.0|mV
R|25|^^^K+_D1|- 0.0|mV
R|26|^^^Ca++_D1|- 0.0|mV
R|27|^^^Mg++_D1|- 0.0|mV
R|28|^^^LAC_D1|+ 0.0|mV
R|29|^^^pH_V1|7.384
R|30|^^^Na+_V1|140.0
R|31|^^^K+_V1| 4.0
R|32|^^^Ca++_V1|1.0
R|33|^^^Mg++_V1| 0.5
R|34|^^^LAC_V1| 2.0
L|1|N
6-25
Auto
On
Off
Off
6. Computer
6-26
6 Computer/Peripheral
H|\^&| | | NOVA^ULTRA11^99.0.23^64| | | | | | | | 1|19950608134609
P|1| | 6931
O|1|482108|404| | | | | | | | | | | | Venous| | | | | | | | | F
R|1|^^^pHTC^T|7.386| | | | | F||102|19950608134344
R|2|^^^PCO2TC^T| 36.5|mmHg| | | | F
R|3|^^^PO2TC^T| 98.6|mmHg| | | | F
R|4|^^^SO2%^M| uc| | | | | F
R|5|^^^Hct^M|? 3.|%| | | | F
R|6|^^^Na+^M|138.|mmol/L| | | | F
R|7|^^^K+^M| 3.7|mmol/L| | | | F
R|8|^^^Ca++^M| 0.93|mmol/L| | | | F
R|9|^^^Mg++^M| 0.53|mmol/L| | | | F
R|10|^^^GLU^M| uc|mg/dL| | | | F
R|11|^^^LAC^M|ERR|mmol/L| | | | F
R|12|^^^BE-ecf^C|- 3.2|mmol/L| | | | F
R|13|^^^BE-b^C|- 2.1|mmol/L| | | | F
6. Computer
R|14|^^^HCO3-^C|22.1|mmol/L| | | | F
R|15|^^^TCO2^C|23.2|mmol/L| | | | F
R|16|^^^A^C|101.0|mmHg| | | | F
R|17|^^^AaDO2^C| 2.4|mmHg| | | | F
R|18|^^^a/A^C| 1.0| | | | | F
R|19|^^^Hb^C|? 1.1|g/dL| | | | F
R|20|^^^nCa^C| 0.93|mmol/L| | | | F
R|21|^^^nMg^C| 0.53|mmol/L| | | | F
R|22|^^^BP^M|738.8|mmHg| | | | F
R|23|^^^TempP^E| 36.7|deg C
R|24|^^^pH_SL|+10.6| | | N
R|25|^^^PCO2_SL|+10.3| | | N
R|26|^^^PO2_SL|- 6.5| | | N
R|27|^^^Hct_SL|+68.9| | | N
R|28|^^^Na+_SL|+10.0| | | N
6-27
R|45|^^^K+_M2|- 15.5|mV
R|46|^^^Ca++_M1|- 5.0|mV
R|47|^^^Ca++_M2|- 4.2|mV
R|48|^^^Mg++_M1|+ 18.3|mV
R|49|^^^Mg++_M2|+ 17.9|mV
R|50|^^^Mg++_M3|+ 30.3|mV
R|51|^^^GLU_M1|+ 5.9|mV
R|52|^^^GLU_M2|+ 4.6|mV
R|53|^^^LAC_M1|+ 19.8|mV
R|54|^^^LAC_M2|+ 16.3|mV
R|55|^^^LAC_M3|+ 40.9|mV
R|56|^^^S1_M5|+100.7|mV
R|57|^^^S3_M5|+226.1|mV
L|1|N
6-28
6 Computer/Peripheral
Example #5
This example shows only the characters of the high-level protocol records. This example is for
a calibration. The following parameters apply to this example:
CALIBRATION OUPUT OPTIONS:
Calibration Transmit Mode:
Transmit Calibration Slope and mV Data:
Transmit Calibration Drift Data:
Transmit Calibration Errors:
Auto
Off
Off
Off
INSTRUMENT:
ULTRA , software version 99.0.23
IDENTIFIERS ENTERED DURING SET UP:
Analyzer Identifier: 64
DATE and TIME:
Transmitted : June 8, 1995 at 13:09:24
Example #6
This example shows only the characters of the high-level protocol records. This example is for
a calibration. The following parameters apply to this example:
CALIBRATION OUPUT OPTIONS:
Calibration Transmit Mode:
Transmit Calibration Slope and mV Data:
Transmit Calibration Drift Data:
Transmit Calibration Errors:
Auto
On
Off
On
INSTRUMENT:
ULTRA , software version 99.0.23
IDENTIFIERS ENTERED DURING SET UP:
Analyzer Identifier: 64
DATE and TIME:
Transmitted : June 8, 1995 at 13:57:40
6-29
6. Computer
6. Computer
6-30
Appendix A Specifications
A
Instrument Specifications
This section includes instrument specifications for the Stat Profile Ultra Analyzer.
A.1
Instrument Specifications*
Measurement Range:
pH
PCO2
PO2
Na+
K+
CliCa
Glucose
Lactate
BUN
SO2 %
Hct
6.50 - 8.00
3.0 - 200 mm Hg
0 - 800 mm Hg
80 - 200 mmol/L
1.0 - 20.0 mmol/L
50 - 200 mmol/L
0.1 - 2.7 mmol/L
15 - 500 mg/dL
0.3 - 20.0 mmol/L
3.0 -100.0 mg/dL
30 - 100 %
12 - 70 %
Acceptable Samples:
Whole Blood (for pH, PCO2, PO2, SO2, Hct, Na+, K+, Cl-,
iCa, Glu, Lac, BUN)
Plasma (for Na+, K+, Cl-, iCa, Glucose, Lactate, BUN)
Serum (for Na+, K+, Cl-, iCa, Glucose, Lactate, BUN)
Expired Gas (for PCO2 and PO2)
Measuring Technology:
A. Specs.
Throughput Time
2.6 minutes
1.3 minutes
2.9 minutes
1.3 minutes
2.8 minutes
1.5 minutes
3.1 minutes
1.5 minutes
A. Specs.
Single Mode
Normal Mode
Single Mode
Normal Mode
Sample Volume:
Barometer:
Reagents:
Slope Limits:
pH
PO2
PCO2
Na+
K+
CliCa
Glu
Lac
BUN
SO2
Hct
9.1 - 11.2
-15.0 - (-1.6)
7.9 - 12.2
8.8 - 10.7
8.6 - 10.9
7.7 - 12.2
8.9 - 11.6
5.0 - 45.0
6.0 - 100.0
8.0 - 12.2
7.0 - 20.0
40.0 - 98.0
A-2
Appendix A Specifications
Analytical Specifications for Imprecision
Analyte
pH
PCO2
PO2
Na+
K+
CliCa
Glu
Lac
BUN <60 mg/dL
BUN >60 mg/dL
SO2
Hct
Within-run
Precision
n=20
Day-to-Day
Precision
n=20
Whichever is Greater
CV%
SD
0.005
3.0
1.0 (mmHg)
3.0
1.5 (mmHg)
1.0
1.0 (mmol/L)
1.5
0.15 (mmol/L)
1.5
2.0 (mmol/L)
2.0
0.05 (mmol/L)
3.0
2.0 (mg/dL)
3.0
0.3 (mmol/L)
3.0
0.5 (mg/dL)
3.0
0.5 (mg/dL)
1.0 SO2%
1.0 Hct%
Whichever is Greater
CV%
SD
0.013
5.0
2.0 (mmHg)
5.0
3.0 (mmHg)
1.5
1.0 (mmol/L)
3.0
0.15 (mmol/L)
2.5
2.0 (mmol/L)
4.0
0.05 (mmol/L)
5.0
3.0 (mg/dL)
6.0
0.3 (mmol/L)
4.0
0.5 (mg/dL)
4.5
0.5 (mg/dL)
1.5 SO2%
1.0 Hct%
Mode
Number of
Samples N
Correlation
Coefficient R
Reference
Instrument
pH
Blood
90
0.997
6.88 - 7.44
STAT Plus 9
PCO2
Blood
74
0.994
28 - 112 mmHg
STAT Plus 9
PO2
Blood
101
0.999
18 - 414 mmHg
STAT Plus 9
SO2
Blood
100
0.999
12.6% - 99.0%
Corning 2500
Co-oximeter
Na+
Plasma
Blood
54
87
0.998
0.969
86 - 151 mmol/L
125 - 149 mmol/L
STAT Plus 9
STAT Plus 9
K+
Plasma
Blood
44
87
0.997
0.998
STAT Plus 9
STAT Plus 9
Cl-
Plasma
Blood
60
60
0.997
0.994
89 - 145 mmol/L
91 - 133 mmol/L
STAT Plus 10
STAT Plus 9
iCa
Plasma
Blood
69
95
0.999
0.998
STAT Plus 9
STAT Plus 9
Glu
Plasma
Blood
63
77
0.999
0.998
30 - 552 mg/dL
50 - 708 mg/dL
STAT Plus 9
STAT Plus 9
Lac
Plasma
Blood
68
92
0.997
0.994
STAT Plus 9
STAT Plus 9
BUN
Plasma
Blood
60
63
0.997
0.994
5 - 96 mg/dL
7 - 93 mg/dL
STAT Plus 10
STAT Plus 10
Hct
Blood
83
0.996
16% - 68%
STAT Plus 9
A. Specs.
Range
PCO2 mmHg
PO2 mmHg
mmol/L
Na+
K+
mmol/L
Cl-(a) mmol/L
BUN mg/dL
GLU(b) mg/dL
LAC(c) mmol/L
Hct
%
Mean
Level 1
SD
%CV
Mean
Level 2
SD
%CV
Mean
Level 3
SD
%CV
7.166
64.9
55.0
158.7
5.96
126.1
10.0
80
0.8
33
0.003
1.62
1.25
0.38
0.07
0.81
0.28
1.58
0.05
0.39
7.361
42.8
95.7
139.1
3.91
99.6
29
200
2.9
53
0.004
1.13
2.15
0.38
0.05
0.85
0.92
3.09
0.09
0.55
7.588
21.7
136.7
121.8
1.93
85.5
50
300
7.6
-
0.006
0.57
4.32
0.71
0.02
1.39
1.43
8.37
0.19
-
2.50
2.30
0.20
1.20
0.60
0.70
2.00
6.10
-
A. Specs.
A-4
2.60
2.20
0.30
1.20
0.90
3.20
1.50
2.90
-
2.60
3.20
0.60
1.20
1.60
2.90
2.80
2.50
-
Appendix A Specifications
Stat Profile Plus 10 Patient Comparison Results
Point-of-Care Study
Patient Comparison
Analyte
Number of
Samples N
Correlation
Coefficient R2
Slope
Y-Intercept
pH
PCO2
PO2
Na+
K+
Cl-(a)
BUN
GLU(b)
LAC(c)
Hct
50
50
50
50
50
50
49
50
46
50
0.994
0.986
0.995
0.983
0.987
0.992
0.995
0.995
0.995
0.841
0.991
0.920
1.008
0.973
0.864
0.942
0.985
0.989
0.987
0.886
+0.066
+3.455
+0.414
+3.889
+0.609
+6.12
+0.191
+0.939
+0.039
+5.40
Range
7.079 - 7.536
13.5 - 135.6 mm Hg
10.8 - 352.8 mm Hg
134.2 - 192.1 mmol/L
2.94 - 15.23 mmol/L
92.5 - 168.2 mmol/L
5 - 86 mg/dL
40 - 424 mg/dL
0.6 - 17.0 mmol/L
12 - 55 %
Gas Standard A
Carbon Dioxide
Oxygen
Nitrogen
Specified Value
5.0%
20.0%
Balance
Accuracy
0.03%
0.03%
Gas Standard B
Carbon Dioxide
Nitrogen
Specified Value
10.0 %
Balance
Accuracy
0.03%
A. Specs.
Electrical Compliance:
Height:
Width:
Depth:
Weight:
A-5
FDA Labeling:
Clean the display screen with a damp (not wet) lint-free cloth.
For a heavy buildup, use a liquid glass cleaner sprayed onto a
lint-free cloth first, never spray directly onto the CRT display.
If available, prepackaged screen wipes can also be used.
Clean the cabinet with a damp (not wet) lint-free cloth. Do not
use aerosol sprays, solvents, or abrasives that might damage the
finish.
CAUTION: Never use the control panel or doors (open or closed) to assist
you in lifting the analyzer. They cannot support the weight of the analyzer.
A. Specs.
Warranty:
A-6
Appendix B Theory
B
Theory
This section explains instrument theory of the Stat Profile Ultra Analyzer.
= SODIUM
= POTASSIUM
= CHLORIDE
SOLUTION A
0.2 M KCl
SOLUTION B
0.05 M NaCl
SOLUTION A
0.04 M NaCl
0.04 M KCl
DWG #5-0-070B
B.1
SOLUTION B
0.04 M NaCl
0.04 M KCl
SIMPLE
PERMEABLE
MEMBRANE
BEFORE EQUILIBRATION
AFTER EQUILIBRATION
B-1
B. Theory
The concentration gradient creates a driving force to send ions in solution from the cell where
they are most concentrated across the membrane to the cell where they are less concentrated.
This force decreases as the concentration equalizes on both sides of the membrane. Equilibration results in equal salt concentrations on both sides of the membrane and equal rates of
migration across the membrane.
[Cl ]o
[Cl ]i
[Na ]o
[Na ]i
+
[K ]o
[K ]i
+
Equation 1
SOLUTION A
0.2 M KCl
SOLUTION B
0.05 M NaCl
DWG #5-0-071B
SOLUTION A
0.2 M KCl
SOLUTION B
0.05 M NaCl
ION
SELECTIVE
MEMBRANE
++++
----
B. Theory
B-2
RT
a
ln o
ai
nF
Equation 2
Appendix B Theory
where Em is the membrane potential, a is the activity of the potassium ion, R is the universal
gas constant, T is the temperature in degrees Kelvin, F is Faradays constant, and n is the charge
of the measured ion: +1 for sodium and potassium, +2 for calcium, -1 for chloride. (The activity
of an electrolyte can be called a measure of its effective concentration in solution. It is equal
to the concentration multiplied by the ions activity coefficient, f; that is, a = f * c. The activity
coefficient describes an ions ability to react in solution.)
When the glass tube in Figure B.2 is filled with a salt solution of constant composition, called
the internal filling solution, the electric potential of the membrane depends only on the solution
outside the membrane as follows:
Em = Eo +
RT
ln a o
nF
Equation 3
where Eo is a constant that includes a term for ai, the activity of the ion in the IFS.
We can convert from the natural logarithm (ln) in Equation 3 to a base 10 logarithm and an
equivalent equation:
E m = E o + 2.303
RT
log a o
nF
Equation 4
In order to measure the potential on the ion selective membrane, a complete electrical circuit
is needed. A typical ion selective electrode measuring circuit is shown in Figure B.3. There are
2 electrodes here: a reference electrode and the ion selective electrode. The potential of the
reference electrode is constant; that of the ISE varies, depending on the activity of the ion of
interest in the test solution. A voltmeter measures the potential difference between the ion
selective electrode and the reference electrode.
VOLTMETER
ION - SELECTIVE
ELECTRODE
REFERENCE
ELECTRODE
INTERNAL
FILLING
SOLUTION
B. Theory
LIQUID/LIQUID
JUNCTION
SAMPLE SOLUTION
ION - SELECTIVE
MEMBRANE
B-3
Equation 5
The junction potential is the potential developed at the liquid/liquid junction between the
reference electrode and the test solution. The magnitude of the liquid junction potential which
arises as a result of the interdiffusion of the ions at the liquid/liquid junction is inconsequential.
RT
By setting S = 2.303
and substituting Equation 4 for Em,
nF
Equation 5 can be rewritten:
Ecell = Eo + S log ao - Er - Ej
Equation 6
E std
E std
C
log
D
Equation 7
When the unknown activity of the ion of interest is measured in a test solution, the potential
of the test solution, Ex, is compared with that of a standard solution, Estd, in which the activity
of the ion of interest is known. Most of the terms cancel out as follows:
B. Theory
minus
Ex
= Eo + S log ax - Er - Ej
Equation 8
Estd
= Eo + S log astd - Er - Ej
___________________
Equation 9
E = E x E std = S Log
ax
a std
Equation 10
Thus, the difference in potential between the 2 solutions, E, is dependent only on the ratio of
the activity of the ion of interest in the test solution, ax, and the activity of the ion of interest
in the standard solution, astd.
B-4
Appendix B Theory
B.2
Electrode Calibration
B.3
Principles of Measurement
B-5
B. Theory
The electrolyte electrodes require a sensing (ion selective) membrane, a reference electrode,
and an internal filling solution. Internal filling solution provides a fixed concentration of the
ion and the sample provides a second ion concentration.
A suitable electrode membrane is selectively permeable to a single ion of interest in the sample.
This selective permeability establishes an electrical potential as the charge associated with the
ion leaves its counter-ion behind in solution. The magnitude of this electrical potential is
determined by the concentration difference between the 2 sides of the membrane.
Equation 11
Equation 12
The total ionic strength of whole blood, plasma, and serum is relatively constant over the
physiological range.1 As a result, the activity coefficients of sodium, potassium, calcium, and
magnesium can be assumed to be constant. The internal standards are formulated to reflect the
same ionic strength as that of whole blood. Therefore, a given ions activity coefficient can be
assumed to be equal in the standard and sample. The activity coefficient terms in Equation 12
cancel out resulting in:
Cx
E = Ex - Estd = S log ____
Cstd
Equation 13
By holding Cstd in Equation 13 constant, E is dependent on only 1 variable, Cx, the concentration
of the ion of interest in the sample. Equation 14 can be rearranged to isolate this variable:
Cx = (Cstd) 10(E/S)
Equation 14
B. Theory
B.3.2 pH Electrode
Principle of pH Measurement
pH is measured using a hydrogen ion selective glass membrane. One side of the glass is in
contact with a solution of constant pH. The other side is in contact with a solution of unknown
pH. A change in potential develops which is proportional to the pH difference of these
solutions. This change in potential is measured against a reference electrode of constant
potential. The magnitude of the potential difference is a measure, then, of the pH of the
unknown solution.
B-6
Appendix B Theory
Definition of pH
The pH is defined as the negative logarithm of the hydrogen ion activity:
pH = -log10 aH+
Equation 15
Thus, when negative pH is substituted for the logarithm of the activity in Equation 10, the
equation becomes:
E = Ex - Estd = -SpHx - (-SpHstd)
Equation 16
Equation 16 can be rearranged so that pH, the pH of the unknown sample, can be calculated
by the microcomputer using 2 calibrating solutions of known pH:
pH x = pH std
where Slope =
E x E std
Slope
Equation 17
E x E std A
pH std A pH std
PCO2 is measured with a modified pH electrode. Carbon dioxide in the unknown solution
CO 2 + H 2 O H 2 CO3 H + + HCO3
Equation 18
The electrolyte solution behind the membrane is in contact with a glass hydrogen ion selective
electrode. The change in hydrogen ion activity in the electrolyte solution produces a potential
which is measured against the internal filling solution. This change in potential is measured
against the constant potential of the reference electrode half cell and is logarithmically related
to the PCO2 of the unknown sample.
B-7
B. Theory
B.3.5 Hematocrit
Hematocrit is defined as the percentage of red blood cells to the total blood volume and can
be obtained by measuring electrical resistance of the blood sample. Two standard solutions are
used to calibrate the hematocrit electrode and to obtain the slope. The analyzer then measures
the electrical resistance of the blood sample to obtain the hematocrit value. Next, the
hematocrit value obtained is corrected for the concentration of the sodium ion. The correction
equation is as follows:
Na +
140.0
B. Theory
B-8
Equation 19
Appendix B Theory
B.3.6 Glucose
Glucose measurement is based on the level of H2O2 produced during the enzymatic reaction
between glucose and oxygen molecules in the presence of the glucose oxidase enzyme. The
reaction is described by the following equation:
Glu cos e
Equation 20
At a constant potential of 0.70 volts, electroactive H2O2 is oxidized at the surface of the
platinum anode as follows:
H 2O2
2H + + O 2 + 2e
Equation 21
The current generated by the flow of electrons at the surface of the platinum electrode is
proportional to the glucose concentration of the sample.
B.3.7 Lactate
Lactate measurement is based on the level of H2O2 produced during the enzymatic reaction
between lactate and oxygen molecules in the presence of the lactate oxidase enzyme. The
reaction is described by the following equation:
Lactate
Oxidase
Lactate + O 2
Pyruvate acid + H 2 O 2
Equation 22
At a constant potential of 0.70 volts, electroactive H2O2 is oxidized at the surface of the
platinum anode as follows:
Equation 23
The current generated by the flow of electrons at the surface of the platinum electrode is
proportional to the lactate concentration of the sample.
B-9
B. Theory
H 2O2
2H + + O 2 + 2e
NH 3 + H +
NH +4
Equation 25
An ammonium ion selective electrode is used to detect the ammonium formed by the above
reactions. This measurement is then related to the concentration of urea present in the original
sample via the Nernst equation.
B. Theory
On the Stat Profile Ultra analyzer, a reflectance photometry system is used to measure the color
of the whole blood to determine the level of oxygen saturation. Using a multiple branch bundle
of optical fibers, the whole blood sample is illuminated by multiple wavelengths of light. A
portion of each wavelength of light is reflected onto one of the optical fiber bundles and in turn
onto a photo detector. By relating the signal observed to those of known standards, the color
of the blood sample is measured, and the level of oxygen saturation is determined.
B.4
Calculated Values
The analyzers microcomputer uses the measured results to calculate other clinically valuable
parameters. This section outlines the equations used to calculate these values.
B-10
Appendix B Theory
B.4.1 Temperature Correction for Measured Values*
The Stat Profile Ultra analyzer allows you to enter the patient temperature when this differs
from 37 C, as for example in patients having surgery under hypothermia. The pH, PCO2, and
PO2 sample values, at the patients actual temperature, are then calculated as follows:
pH(corrected) = pH + [- 0.0147 + 0.0065 (7.400 - pH)](T - 37)
Equation 26
Equation 27
Equation 28
)
)
5.49 10 11 Y + 0.071
9.72 10 Y + 2.30
Equation 29
Equation 30
B-11
Equation 31
Hemoglobin
The hemoglobin is calculated based on the following calculation:
Hemoglobin g/dL = Measured Hematocrit 3.0
Equation 32
Equation 33
B. Theory
B-12
Equation 34
Appendix B Theory
[Hb] = The hemoglobin value which is measured, manually entered, or is the 14.3 g/dL
default value
Base Excess Extracellular Fluid (BE-ECF)*
The Base Excess Extracellular fluid is a corrected form of the Base Excess Blood in which
allowance has been made for the fact that blood is only approximately 37% of the extracellular
fluid volume. Base excess is calculated as follows:
BE-ECF = [HCO3-]- 25 + 16.2 (pH - 7.40)
Equation 35
SO 2
therefore O 2 Ct = (1.39 [Hb]) 100 + (0.0031 [ PO 2 ])
Equation 36
Equation 37
B. Theory
B-13
1 %FIO2
%FIO2
%FIO2
100
A=
(B.P. 0.045T 2 + 0.84T 16.5)* *PCO 2
+
100
0.8
100
Equation 38
where
T = patient temperature
B.P. = barometric pressure
%FIO2 = fraction inspired oxygen, as a percent
Equation 39
B. Theory
B-14
Equation 40
Appendix B Theory
P50 or PO2 (0.5)*
The P50 is defined as the PO2 of a sample at which the hemoglobin is 50% saturated with
oxygen at pH 7.4, 37C, and 40 mmHg PCO2 for SO2 values between 40% and 80%.
SO 2
100 SO 2
Equation 41
PCO2
If an anaerobic sample is not available, by measuring the actual pH of the sample at which the
ionized calcium concentration was measured normalized ionized calcium can be calculated.
The normalized ionized calcium represents what the ionized calcium concentration would
have been if the initial pH was 7.40 (the midpoint of the pH reference range). The equation used
for this calculation is:
log [Ca++] 7.4 = log [Ca++]X - 0.24 (7.4 - X)
where
Equation 43
B-15
[Glu] + [BUN ] + 9
18
2.8
Equation 44
Sodium units are mmol/L, glucose units are mg/dL, and BUN units are mg/dL.
Anion Gap
Anion gap is the difference between the sum of the sodium and potassium concentrations (the
cations) and the sum of the chloride and bicarbonate concentrations (the anions), as follows:
Anion Gap = (Na+ + K+) - (Cl- + [HCO3-})
Equation 45
No anion gap is reported if any of the 4 concentrations are not reported. Any calculated anion
gap less than 1 nM is not reported.
PaO2/FIO2 = PaO2/(%FIO2/100%)
Equation 46
B. Theory
The Oxygen tension based index of PaO2/FIO2 is used in the estimation of the intrapulmonary
shunt fraction when pulmonary artery blood samples are not available. The Estimated Shunt,
an oxygen content based index, is derived by mathematical manipulation of the Classic Shunt
Equation5. In assessing the lung as an Oxygenator, the calculation of the intrapulmonary shunt
fraction at maintained inspired oxygen concentrations (Qsp/QT or Q va /QT ) is generally
recognized as the most reliable way to quantitate disruption of pulmonary oxygen transfer and
therefore the extent to which pulmonary disease is contributing to arterial hypoxemia.
B-16
Appendix B Theory
The most significant factor limiting the widespread clinical use of shunt fractions is that the
calculation requires oxygen analysis of pulmonary artery blood. The Estimated Shunt
calculation is based on use of an assumed Q(a-v)O 2 of 3.5 mL/dL which has been shown to be
a representative mean for large samples of critical ill patients with clinically adequate
perfusion states6,7,8.
The Estimated Shunt has been demonstrated to be far superior to oxygen tension based indices
in reliably reflecting changes in the Qsp/QT9. Monitoring of the tcPO2 Index and the P(a-et)CO2
should allow for verification of the adequacy of cardiac output and peripheral perfusion,
thereby confirming the reliability of the Estimated Shunt to quantitate changes in Qsp/QT.
References:
1.
2.
3.
4.
5.
6.
7.
8.
9.
Mohan, M.S. and Bates, R.G. 1977. Blood pH, Gases and Electrolytes. NBS Special Publication, 450. U.S.
Government Printing Office.
National Committee for Clinical Laboratory Standards. 1982. Tentative Standard for Definitions of
Quantities and Conventions Related to Blood pH and Gas Analysis. NCCLS 2:10.
Williams, W.J., Beutler, E., Ersley, A.J., and Rundles, R.W. 1977. Hematology. 2nd ed. McGraw-Hill Co.
Tietz, Norbert W., ed. 1986. Textbook of Clinical Chemistry. W.B. Saunders Co. Philadelphia, Penn.
Shapiro, B.A., Harrison, R.A., Cane, R.D., Templin, R. 1989. Clinical Application of Blood Gases. 4th ed.
Year Book Medical Publishers. Chicago. pp 158-9.
Harrison, R.A., Davidson, R., Shapiro, B.A., Myer, N.S. 1975. Reassessment of the assumed A-V oxygen
content difference in the shunt calculation. Anesth Analg. Vol 54 No 198.
Suter, P.M., Fairley, H.B., Isenberg, M.D. 1975. Optimum end expiratory pressure in patients with acute
pulmonary failure. N Engl J Med. Vol 292 No 84.
Suter, P.M., Fairley, H.B., Schlobohm, R.M. 1975. Shunt, lung volume, and perfusion during short periods
of ventilation with oxygen. Anesthesiology. Vol 43 No 617.
Cane, R.D., Shapiro, B.A., Templin, R., et al. 1988. The unreliability of oxygen tension based indices in
reflecting intrapulmonary shunting in critically ill patients. Crit Care Med. Vol 16 No1243.
B. Theory
B-17
B. Theory
B-18
Appendix B Theory
P50 or PO2 (0.5)*
The P50 is defined as the PO2 of a sample at which the hemoglobin is 50% saturated with
oxygen at pH 7.4, 37C, and 40 mm Hg PCO2 for SO2 values between 40% and 80%.
SO 2
100 SO 2
Equation 41
PCO2
If an anaerobic sample is not available, by measuring the actual pH of the sample at which the
ionized calcium concentration was measured normalized ionized calcium can be calculated.
The normalized ionized calcium represents what the ionized calcium concentration would
have been if the initial pH was 7.40 (the midpoint of the pH reference range). The equation used
for this calculation is:
log [Ca++] 7.4 = log [Ca++]X - 0.24 (7.4 - X)
where
Equation 43
B-15
[Glu] + [BUN ] + 9
18
2.8
Equation 44
Sodium units are mmol/L, glucose units are mg/dL, and BUN units are mg/dL.
References:
1.
2.
3.
4.
B-16
Mohan, M.S. and Bates, R.G. 1977. Blood pH, Gases and Electrolytes. NBS Special Publication, 450. U.S.
Government Printing Office.
National Committee for Clinical Laboratory Standards. 1982. Tentative Standard for Definitions of
Quantities and Conventions Related to Blood pH and Gas Analysis. NCCLS 2:10.
Williams, W.J., Beutler, E., Ersley, A.J., and Rundles, R.W. 1977. Hematology. 2nd ed. McGraw-Hill Co.
Tietz, Norbert W., ed. 1986. Textbook of Clinical Chemistry. W.B. Saunders Co. Philadelphia, Penn.
C.1
11.
12.
C. Reagents
13.
Gas A: PN 06586
Gas B: PN 06587
C-1
NOTE: Refer to the Nova Stat Profile Control insert for storage requirements
for these controls. Store Chem Set B Linearity Solution/External Standard
at 2 to 8 C and store all other Nova Stat Profile reagents, standards, and
solutions at 15 to 30 C.
C.2
Reagent Pack
The concentrations of the internal standards* are printed on the reagent pack. In addition to the
reagents and solutions, the reagent pack has a self-contained waste bottle for safe disposal of waste.
*Standards A and B are directly traceable to the National Bureau of Standards Buffers for
standardization of pH measurements, cited as follows:
Durst, R. A. 1975. Standardization of pH Measurements, NBSS Publication
PB248:127, National Bureau of Standards, National Technical Service.
C.3
Calibrating Gases
Nova calibrating gases contain mixed gas for calibrating the PO2 and PCO2 electrodes on the Stat
Profile Ultra analyzer (see Table C.1)*. Always check the gas composition values and specifications
on newly replaced gas cylinders against the Set Calibration Gases screen (Section 5.1.6).
____________________________________________________________________________________
C. Reagents
Gas Standard A
Carbon Dioxide
Oxygen
Nitrogen
Gas Standard B
Carbon Dioxide
Nitrogen
C-2
Specified Value
Accuracy
5.0%
20.0%
Balance
0.03%
0.03%
Specified Value
Accuracy
10.0%
Balance
0.03%
0.03%
C.4
NOTE: In order to minimize the time for daily analyzer performance when
running all controls together, Nova recommends that the controls be run in
the following order:
1. Ultra Multi Control Level 1 with Glucose Over-Range
2. Ultra Multi Control Level 2
3. Level 1, Level 2, and Level 3 STP Controls
C. Reagents
C-3
C. Reagents
C-4
Press ANALYZE, present the control, and press ANALYZE again to initiate
aspiration.
Compare the results with those listed in the assay data sheet included with the
controls.
C. Reagents
C-5
Appendix D Reference
D
Reference Values
Each laboratory should establish and maintain its own reference values. The values given here
should be used only as a guide.
____________________________________________________________________________________
Table D.1 Reference Values1,2
Test
Value
pH
7.35 - 7.45
PCO2
35 - 45 mmHg
HCO3-
21 - 28 mmol/L
(-2)-(+3) mmol/L
PO2
83 - 108 mmHg
95 - 98%
Hemoglobin (Hb)
(Male)
(Female)
Hematocrit (Hct)
(Male)
(Female)
43 - 51%
38 - 46%
____________________________________________________________________________________
D. Reference
D-1
Test
Value
Sodium2
Potassium2
Chloride2
98 - 106 mmol/L
Glucose2
65 - 95 mg/dL
Lactate2
iCa3
BUN2
7 - 18 mg/dL
Osmolality2
____________________________________________________________________________________
References:
D. Reference
1. Statland, Bernard. 1987. Clinical Decisions Levels for Lab Tests, Medical Economics Books.
2. Tietz, Norbert W., ed. 1987. Textbook of Clinical Chemistry, W.B. Saunders Co.,
Philadelphia, Penn.
3. Kost, G.T. 1993. The Significance of Ionized Calcium in Cardiac and Critical Care.
Arch. Pathol. Lab Med. Vol. 117: pp 890-896.
4. Altura, Burton M. and Lewenstam, Andrzej, ed. 1994. The Scandinavian Journal of
Clinical & Laboratory Investigation. Vol. 54: Supplement 217.
D-2
E. Maps
E-1
3
4
5
6
7
MENU
30 Aug 96
REAGENT PACK
30 Aug 96
ENTER
11:45:45
TUBING REPLACEMENT
*
1
Pump Tubing
Reagent / Septum Harness
W/R Harness
S line / Probe
Flowcell
Preheater
Replace
Replace
Replace
Replace
Replace
Replace
30 Aug 96
1
2
3
4
5
6
CONCENTRATIONS
6.840
12.3
12.3
1.2
123.4
123.4
123.4
11:45:45
123456789
1234567
mm yy
100%
CALIBRATION STANDARD
pH
7.384
Na+
123.4
K+
1.2
Ca++/Cl- 1.2
Glu
12.3
Lac
12.3
BUN
12.3
Alternate Analysis ( E x p i r e d G a s )
Flow Cell Light
Off
1
2
COMPONENT REPLACEMENT
MAINTENANCE MENU
11:45:45
30 Aug 96
30 Aug 96
11:45:45
11:45:45
Calibration to Follow.
123 Seconds to Completion.
30 Aug 96
ELECTRODE REPLACEMENT
11:45:45
Calibration in Progress.
Press ENTER When Ready.
123 Seconds to Completion.
30 Aug 96
11:45:45
11:45:45
Calibration in progress.
Calibration Gases
Fluid Prime
Gas Prime
11:45:45
FLUID
GAS
FULL
Prime in Progress
Close door.
Press ENTER When Ready.
123 Seconds to Completion.
ENTER
SO2% CALIBRATION
Calibration in progress.
M E T A B O L I T E E L E C T R O D E CHANGE PROCEDURE
M E T A B O L I T E E L E C T R O D E CHANGE PROCEDURE
M E T A B O L I T E E L E C T R O D E CHANGE PROCEDURE
Calibration in progress.
30 Aug 96
Standard
# 1:
Standard
# 2:
51
PREHEATER REPLACEMENT P R O C E D U R E
E
*
Close door.
Calibration in progress.
Press ENTER When Ready.
123 Seconds to Completion.
PREHEATER REPLACEMENT P R O C E D U R E
E
E
PREHEATER REPLACEMENT P R O C E D U R E
PREHEATER REPLACEMENT P R O C E D U R E
E
PREHEATER REPLACEMENT P R O C E D U R E
PREHEATER REPLACEMENT P R O C E D U R E
E
PREHEATER REPLACEMENT P R O C E D U R E
E
PREHEATER REPLACEMENT P R O C E D U R E
Calibration in progress.
Press ENTER When Ready.
123 Seconds to Completion.
11:45:45
*
Assay Values
95
ENTER
SO2% CALIBRATION
30 Aug 96
REAGENT
/SEPTUM HARNESS REPLACEMENT PROCEDURE
REAGENT/SEPTUM
/SEPTUM HARNESS REPLACEMENT PROCEDURE
REAGENT/SEPTUM
REAGENT
REAGENT
/SEPTUM HARNESS REPLACEMENT PROCEDURE
REAGENT/SEPTUM
REAGENT
/SEPTUM HARNESS REPLACEMENT PROCEDURE
REAGENT/SEPTUM
Assay Values
95
51
Purging
*
Position Standard # 2 for aspiration
Press ANALYZE when ready.
30 Aug 96
11:45:45
REAGENT
/SEPTUM HARNESS REPLACEMENT PROCEDURE
REAGENT/SEPTUM
PROCEDURE
/SEPTUM HARNESS REPLACEMENT PROCEDURE
REAGENT/SEPTUM
REAGENT
REAGENT
/SEPTUM HARNESS REPLACEMENT PROCEDURE
REAGENT/SEPTUM
11:45:45
*
ENTER
*
30 Aug 96
10.00 %
0.00 %
14:31:43
11:45:45
ENTER
5.00 %
20.00 %
REAGENT
/SEPTUM HARNESS REPLACEMENT PROCEDURE
REAGENT/SEPTUM
PROCEDURE
/SEPTUM HARNESS REPLACEMENT PROCEDURE
REAGENT/SEPTUM
REAGENT
REAGENT
/SEPTUM HARNESS REPLACEMENT PROCEDURE
REAGENT/SEPTUM
CALIBRATION GASES
ANALYZE
11:45:45
30 Aug 96
30 Aug 96
M E T A B O L I T E E L E C T R O D E CHANGE PROCEDURE
11:45:45
Calibration to Follow.
Press CLEAR to terminate Sequence.
# 1:
# 2:
Standard
Standard
E
FLOW CELL REPLACEMENT P R O C E D U R E
E
FLOW CELL REPLACEMENT P R O C E D U R E
E
FLOW CELL REPLACEMENT PROCEDURE
30 Aug 96
E
FLOW CELL REPLACEMENT PROCEDURE
Start Sequence?
1
2
ENTER
30 Aug 96
-------- Calibration
Calibration in progress.
11:45:45
30 Aug 96
Full Calibration
Gas Calibration
pH + Hct Calibration
Electrolyte Calibration
Metabolite Calibration
SO2% Calibration
1
2
3
4
5
6
CALIBRATION OPTIONS
S A M P L EP R E H E A T E RD E P R O T E I N I Z I N GP R O C E D U R E
S A M P L EP R E H E A T E RD E P R O T E I N I Z I N GP R O C E D U R E
S A M P L EP R E H E A T E RD E P R O T E I N I Z I N GP R O C E D U R E
S A M P L EP R E H E A T E RD E P R O T E I N I Z I N GP R O C E D U R E
S A M P L EP R E H E A T E RD E P R O T E I N I Z I N GP R O C E D U R E
*
Remove Electrode.
Condition Na+ for 2 min.
Install electrode & plug in.
N a+ C O N D I T I O N I N G
N a+ C O N D I T I O N I N G
N a+ C O N D I T I O N I N G
30 Aug 96
N a+ C O N D I T I O N I N G
*
ENTER
30 Aug 96
Calibration in Progress
Press CLEAR to Terminate Sequence.
MENU
QC
MAIN MENU
PASSWORD REQUIRED
1
Maintenance Menu
Password: _
Setup Menu
Service Menu
30 Aug 96
SETUP MENU
MENU
PASSWORD REQUIRED
PASSWORD REQUIRED
Password: _
Password: _
1
2
3
4
5
6
30 Aug 96
11:45:45
11:45:45
30 Aug 96
30 Aug 96
11:45:45
Print
Print
Print
Print
Print
Stop Printing
4
5
6
7
8
1
2
3
4
QC Reports Menu
Move Daily Data to Month-to-Date
Erase Daily Data
Erase Month-to-Date Data
2
3
WARNING:
the effects of this operation
cannot be undone.
Press ENTER to Confirm Operation.
Press CLEAR to Cancel Operation.
2
3
4
11:45:45
QUALITY CONTROL
QUALITY CONTROL
30 Aug 96
11:45:45
1
2
3
4
5
6
7
Lot Number
2316.10.94
3524.12.94
3525.12.95
2822.05.95
2823.05.95
3705.02.95
Status
Expired
Active
Active
Active
Due
Incomplete
Empty
30 Aug 96
11:45:45
30 Aug 96
11:45:45
11:45:45
30 Aug 96
0
MOVE DAILY DATA TO MONTH-TO-DATE
ERASE DAILY DATA
ERASE MONTH-TO-DATE DATA
SUMMARY STATISTICS
ENTER
Type
Blood Gas Ctrl #1
Blood Gas Ctrl #2
Blood Gas Ctrl #3
Hematocrit Ctrl #1
Hematocrit Ctrl #2
SO2% Control #1
SO2% Control #2
Lot Number
2316.10.95
3524.12.95
3525.12.95
2822.05.95
2823.05.95
3705.02.95
Status
Expired
Active
Active
Active
Due
Incomplete
Empty
Select Test
to be displayed
pH
1
PCO2
2
PO2
3
SO2%
4
Hct
5
Na+
6
K++
7
Ca++/Cl8
Glu
9
Lac
.
BUN
0
1 QC Lockout Mode
1
2
3
4
5
6
7
11:45:45
1
2
3
4
5
6
7
Type
Blood Gas Ctrl #1
Blood Gas Ctrl #2
Blood Gas Ctrl #3
Hematocrit Ctrl #1
Hematocrit Ctrl #2
SO2% Control #1
SO2% Control #2
Lot Number
2316.10.94
3524.12.94
3525.12.95
2822.05.95
2823.05.95
3705.02.95
Status
Expired
Active
Active
Active
Due
Incomplete
Empty
Auto
Auto
4
5
6
7
Manual
Manual
Manual
Manual
Print
Print
Print
Print
Disabled
11:45:45
30 Aug 96
11:45:45
11:45:45
SUMMARY STATISTICS
PCO2
2
PO2
3
Ctrl
If SO2%
selected
2
Hct
5
Na+
6
K++
7
Ca++/Cl8
Glu
9
Lac
.
BUN
0
SO2%
4
Mean
SD
xxx.x
xx.x
xxxx
xxx.x - xxx.x
SO2 2
xxx.x
xxx.x
xx.x
xxxx
xxx.x - xxx.x
Lot Number
2316.10.94
3524.12.94
3525.12.95
2822.05.95
2823.05.95
3705.02.95
30 Aug 96
CV%
If Hct Ctrl
selected
xxx.x
xxx.x
xx.x
xxx
M-T-D (nn/32)
xxx.x
xxx.x
xxxx
xxxx
Cumulative
xxx.x
xxx.x
xx.x
xxxx
hh:mm
Analysis Frequency
hh
(HH:MM)
hours.
30 Aug 96
11:45:45
30 Aug 96
Ctrl
Mean
SD
Hct
%
CV%
BG 1
BG 2
BG 3
xx.
xx.
xx.
xx.
xx.
xx.
xx.x
xx.x
xx.x
xxxx
xxxx
xxxx
xx.x
xx.x
xx.x
xx.
xx.
xx.
Hct 1
Hct 2
xx.
xx.
xx.
xx.
xx.x
xx.x
xxxx
xxxx
xx.x
xx.x
xx.
xx.
11:45:45
11:45:45
If
SO2% control
CONTROL RANGE LIMITS
SO2% Control #1
Expected Range
High
Lot Number
cccccccccc
Exp. Date
Daily
oooooooooo
dd/mm/yy
(DD/MM/YY)
Low
If Hct
selected
SD
dd/mm/yy
mmHg
Lot Number:
Expiration Date:
TEST SELECT
PCO2
nnn.n - nnn.n
Status
Expired
Active
Active
Active
Due
Incomplete
Empty
#1
11:45:45
Mean
Type
Blood Gas Ctrl #1
Blood Gas Ctrl #2
Blood Gas Ctrl #3
Hematocrit Ctrl #1
Hematocrit Ctrl #2
SO2% Control #1
SO2% Control #2
Expected Range
Expected Range
xxx.x
SUMMARY STATISTICS
Lot #
SO2 1
CV%
1
2
3
4
5
6
7
Select Test
to be displayed
3
4
SO2%
11:45:45
If Blood
Gas or
SO2% Ctrl
selected
30 Aug 96
30 Aug 96
SO2%
nn.
nn.
dd/mm/yy
5
Press or to select test.
Press CLEAR to Exit.
30 Aug 96
11:45:45
30 Aug 96
11:45:45
If Blood Gas
or electrolyte
test selected
If Hematocrit control
CUMULATIVE TEST STATISTICS
PCO2
mmHg
2
3
Low
Ctrl
Mean
SD
CV%
Expected Range
Lot Number
cccccccccc
BG 1
xxx.x
xxx.x
xx.x
xxxx
xxx.x - xxx.x
Exp. Date
BG 2
xxx.x
xxx.x
xx.x
xxxx
xxx.x - xxx.x
dd/mm/yy
BG 3
xxx.x
xxx.x
xx.x
xxxx
xxx.x - xxx.x
Hct
6
7
8
DAILY CONTROL SUMMARIES
MONTH-TO-DATE STATISTICS FOR TEST
LEVEY-JENNINGS GRAPHS
30 Aug 96
0
PO2
3
SO2%
4
Hct
5
Na+
6
++
++
K
7
Ca
8
Lac
.
BUN
0
Glu
9
Press ENTER to
initiate printing.
11:45:45
Type
Blood Gas Ctrl #1
Blood Gas Ctrl #2
Blood Gas Ctrl #3
Hematocrit Ctrl #1
Hematocrit Ctrl #2
SO2% Control #1
SO2% Control #2
1
2
3
4
5
6
7
11:45:45
PCO2
2
Lot Number
2316.10.95
3524.12.95
3525.12.95
2822.05.95
2823.05.95
3705.02.95
Lot Number
cccccccccc
Status
Expired
Active
Active
Active
Due
Incomplete
Empty
Exp. Date
dd/mm/yy
pH
n.nnn
PCO2
nnn.n
POP2
nnn.n
Na+
nnn.n
K+
nn.nn
Ca++/Cl- nn.nn
Glu
nnn.n
Lac
nn.nn
BUN
nn.nn
High
n.nnn
nnn.n
nnn.n
nnn.n
nn.nn
nn.nn
nnn.n
nn.nn
nn.nn
30 Aug 96
If Daily QC
Summary graphs
11:45:45
Hematocrit Ctrl #1
If Hct control
Lot #
cccccccccc
Exp. Date
dd/mm/yy
selected
Summary by Test
Test
Mean
SD
CV%
Hct
xx.
xx.
xx.x
xxxx
Expected Range
4
If Levey5
xx.
xx.
If Blood Gas
control
selected
Test
1
2
3
pH
PCO2
PO2
Hct
Na+
K+
Ca++/ClGlu
Lac
BUN
SD
CV%
x.xxx
xxx.x
xxx.x
xx.
xxx.x
xx.xx
xx.xx
xxx.x
xxx.x
xxx.x
x.xxx
xxx.x
xxx.x
xx.
xxx.x
xx.xx
xx.xx
xxx.x
xxx.x
xxx.x
xx.x
xx.x
xx.x
xx.x
xx.x
xx.x
xx.x
xx.x
xx.x
xxxx
xxxx
xxxx
xxxx
xxxx
xxxx
xxxx
xxxx
xxxx
xxxx
dd/mm/yy
Expected Range
x.xxx
xxx.x
xxx.x
xx.
xxx.x
xx.xx
xx.xx
xxx.x
xxx.x
xxx.x
x.xxx
xxx.x
xxx.x
xx.
xxx.x
xx.xx
xx.xx
xxx.x
xxx.x
xxx.x
SO2%
control
selected
If Hct or
Jennings graphs
30 Aug 96
If Month-to-Date
11:45:45
.
pH
1
nn.
nn.
SYSTEM BUSY
High
SO2% Control #1
Lot #
cccccccccc
Exp. Date
Test
Mean
SD
CV%
SO2%
xxx.x
xxx.x
xx.x
xxxx
Hct
xx.
xx.
xx.x
xxxx
dd/mm/yy
Expected Range
7
xxx.x - xxx.x
xx.
xx.
11:45:45
mmHg
mmHg
mmol/L
mmol/L
mg/dL
mg/dL
mmol/L
mmol/L
PASSWORD REQUIRED
SET UP MENU
Password: ______
1
2
3
4
5
6
MENU
30 Aug 96
2
3
4
5
Electrolyte Resolutions
37.0 C
Hemoglobin
14.9
mmol/L (Hb/4)
FIO2
20.9
Units of Measurement
Electrode Offsets
BUN
1
2
3
4
5
6
Analysis Configuration
Calibration Configuration
Barometric Pressure
Set Date and Time
Set Analyzer Identification
Sample Number Counter
COMMUNICATIONS SETUP
PASSWORD OPTIONS
11:45:45
30 Aug 96
Enabled
Results Transmission
COM11
Ticket Printer
COM13
2
3
4
Remote Diagnostic
COM12
6
7
Passwords Entry
Password Entry
2.34 mg/dL
30 Aug 96
Patient Temperature
11:45:45
OPERATION CONFIGURATION
DEFAULT PARAMETER
Results Reporting
30 Aug 96
RESULTS CONFIGURATION
1
11:45:45
On
---
Enabled
R.R.
*
No R.R.
Enabled
Technician
Supervisor
11:45:45
30 Aug 96
11:45:45
11:45:45
30 Aug 96
11:45:45
ELECTROLYTE RESOLUTION
ANALYSIS CONFIGURATION
OPERATOR PASSWORDS
Cl- Result
xxx.x
5
5
6
Auto
On
On
Single
Auto
Enabled
30 Aug 96
11:45:45
30 Aug 96
30 Aug 96
30 Aug 96
11:45:45
30 Aug 96
TESTSELECT
Configure Electrodes:
Press Function + Nbr
Underline - Installed
SO2%
4
Hct
5
Na+
6
K+
7
Ca++/Cl8
Glu
9
Lac
.
BUN
0
Word Length
Stop Bits
ACK Required
Enabled
11:45:45
Units of Measurement
mmHg
Method of Determination
Measured
30 Aug 96
30 Aug 96
11:45:45
11:45:45
Date:
30/04/95
(DD/MM/YY)
Time:
11:45:45
(HH:MM:SS)
Password
30 Aug 96
30 Aug 96
11:45:45
PANIC
LOW
HIGH
0.000
0.000
0.0
0.0
mmHg
0.0
0.0
mmHg
0.0
0.0
0.
0.0
%
0.0
0.0
mmol/L
0.00
0.00 mmol/L
0.00
0.00 mmol/L
0.0
0.0
mg/dL
0.0
0.0
mmol/L
0.0
0.0
mg/dL
Hb
1
BE-ECF
2
BE-B
3
SBC
4
HCO35
TCO2
6
O2Ct
7
A
8
AaDo2
9
a/A
.
P50
0
PO2/FI
-
30 Aug 96
11:45:45
Analyzer Identification #:
nn
Results Reporting:
Press Number
Hilite - Reporting
REFERENCE
LOW
HIGH
pH
0.000
0.000
PCO2
0.0
0.0
PO2
0.0
0.0
SO2%
0.0
0.0
Hct
0.
0.
Na+
0.0
0.0
K+
0.00
0.00
Ca++/Cl- 0.00
0.00
Glu
0.0
0.0
Lac
0.0
0.0
BUN
0.0
0.0
ANALYZER IDENTIFICATION
11:45:45
30 Aug 96
11:45:45
TESTSELECT
Ultra I, J, and K
Ultra F, G, and H
Results Reporting:
Press Number
Hilite - Reporting
nCa++
1
Osm
2
Results Reporting:
Press Number
Hilite - Reporting
Osm
1
An.Gap
2
123
___
11:45:45
11:45:45
30 Aug 96
__
TESTSELECT
SUPERVISOR PASSWORD
DATE/TIME
11:45:45
OprPswdOprPswdOprPswd
301
nnnn
316
nnnn
331
nnnn
302
nnnn
317
nnnn
332
nnnn
303
nnnn
318
nnnn
333
nnnn
304
nnnn
319
nnnn
334
nnnn
305
nnnn
320
nnnn
335
nnnn
306
nnnn
321
nnnn
336
nnnn
307
nnnn
322
nnnn
337
nnnn
308
nnnn
323
nnnn
338
nnnn
309
nnnn
324
nnnn
339
nnnn
310
nnnn
325
nnnn
340
nnnn
311
nnnn
326
nnnn
341
nnnn
312
nnnn
327
nnnn
342
nnnn
313
nnnn
328
nnnn
343
nnnn
314
nnnn
329
nnnn
344
nnnn
315
nnnn
330
nnnn
345
nnnn
On
Odd
30 Aug 96
PO2
3
Parity
TECHNICIAN PASSWORDS
11:45:45
PCO2
2
11:45:45
pH
1
4800
BAROMETRIC PRESSURE
Results Reporting:
Press Number
Hilite - Reporting
Measured
or
Entered
11:45:45
30 minutes
Data
Auto
Off
Off
Baud
OPERATOR PASSWORDS
Offset
+ 0.0 mpH
+ 0.0 %
+ 0.0 %
+ 0.0 %
+ 0.0 %
+ 0.0 %
+ 0.0 %
+ 0.0 %
+ 0.0 %
+ 0.0 %
+ 0.0 %
RESULTS REPORTING
Manual
On
Off
Off
5
6
7
30 Aug 96
ELECTRODE OFFSETS
Mode
Slope & mV Data
DRIFT Data
Errors
11:45:45
Electrode
pH
PCO2
PO2
SO2%
Hct
Na+
K+
Ca++/ClGlu
Lac
BUN
Calibration Transmit
Transmit Calibration
Transmit Calibration
Transmit Calibration
30 Aug 96
11:45:45
1
2
3
4
C
kPa
mmol/L
mg/dL
mmol/L
mg/dL BUN
mmol/L (Hb/4)
Temperature
Blood Gas Values
Ca++ Concentration
Glu Concentration
Lac Concentration
BUN/Urea Concentration
Hemoglobin
COM1:
CALIBRATION CONFIGURATION
UNITS OF MEASUREMENT
1
2
3
5
6
7
8
30 Aug 96
Port
11:45:45
xx.xx
K+ Result
Full Panel
Manual
Off
Off
Off
Mode
mV Data
Drift Data
Errors
xxx.
Transmit
Analysis
Analysis
Analysis
Na+ Result
Analysis
Transmit
Transmit
Transmit
1
2
3
4
11:45:45
11:45:45
SERVICE MENU
1 System Test
11:45:45
SYSTEM TEST #1
Version 123
DISK OPERATIONS
Revision 123.456
SYSTEM BUSY
Millivolt Display
PCO2 Membrane
PO2 Gain
Air Oscillators
Pump
Sampler Position
Valve Position
Gas Flow
Pump Bypass Valve
SO2% LED's
+042.32
-023.23
-023.23
Multiple
Off
Low
Off
0-Off
0
0
0-Off
Closed
Off
1
2
3
4
5
6
7
Printer Subsystem
External Parallel Port
Communications Tests
Disk Drive Utilities
Program Reload
Glu
Lac
BUN
1
2
3
4
5
6
7
8
9
0
Na
+008.56
K+
+017.35
Ca /Cl +074.43
++
pH + 047.43
PCO2 -082.12
PO2 -0051.20
Version 123
COMMUNICATIONS TEST
PROGRAM RELOAD
Revision 123.456
-0008.6 mV
Multiple
Off
Low
Off
0-Off
0
0
0-Off
Closed
Off
Gas B/Std B
mV Stab
-21.6
0.4
pH
pH
PCO2
PO2
+127.6
-4.0
pH
PCO2
PO2
Sample
mV Stab
-64.7
0.4
-91.9
0.9
-228.4
0.0
0.4
0.0
Channel 1
1
2
3
4
5
6
7
8
9
0
11:45:45
SYSTEM TEST #2
11:45:45
mV
Chan
30 Aug 96
11:45:45
30 Aug 96
11:45:45
STD C/FLUSH
mV Stab
-165.5
1.8
+0.6
0.0
+68.3
0.3
Sample
mV Stab
Na+
-166.3
2.5
K+
+0.2
0.1
Cl-/Ca++ +68.6
0.2
Std C/FLUSH
mV Stab
-160.5
0.2
+0.3
0.3
+66.9
0.2
TRANSMITTER TEST
aaaaaaaaaaaaaaaaaaaaaaaaaa
Errors 123
@@@@@@@@@@@@@@@@@@@@@@@@
Press ENTER for Next Screen.
Press CLEAR to Exit.
30 Aug 96
11:45:45
ENTER
PROGRAM RELOAD
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
Std D/Std B
mV Stab
Na+
-180.3
2.9
K+
+42.1
0.4
Cl-/Ca++ +77.8
0.1
11:45:45
22
23
24
25
26
27
28
29
30
31
mV
ENTER
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
30 Aug 96
11
12
13
14
15
16
17
18
19
20
21
ELECTROLYTES
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
+1234.5
ANALOG INPUT
0
1
2
3
4
5
6
7
8
9
10
ENTER
Chan
30 Aug 96
mV
Chan
30 Aug 96
11:45:45
ENTER
TEMPERATURE CONTROL
AB BLR MTR
Preheater
37.1 C
+0080.10 mV
+ 50.99 mV
119
Controller
On
134
Off
On
Glu
Lac
BUN
Off
-9.0
0.8
Sample
mV Stab
+13.7
0.0
+13.7
0.0
+13.7
0.0
STD C/Std A
mV Stab
+12.6
0.5
+12.6
0.5
+70.2
0.6
-39.1
0.2
+3.3
0.8
Std C
mV Stab
+12.4
0.4
+12.4
0.4
+12.4
0.4
Slope
+17.8
+17.8
+10.6
Conc
86.4
86.4
86.4
2
3
4
Cancel Printout
Printer Enabled
+0004.30 mV
+ 566.68 mV
Std D
mV Stab
+33.9
0.0
+33.9
0.0
DAC
Air Bath
37.0 C
Glu
Lac
Lac
BUN
BUN
Thermistor
Current
SP REMOVED
Temperature
DOOR OPEN
PRINTER SUBSYSTEM
METABOLITES
30 Aug 96
11:45:45
30 Aug 96
11:45:45
11:45:45
30 Aug 96
11:45:45
ENTER
95 8:40:28
95 8:40:28
95 9:40:28
95 10:40:28
95 11:40:28
95 8:40:28
95 8:40:28
95 8:40:28
95 8:40:28
95 8:40:28
95 8:40:28
95 8:40:28
95 8:40:28
95 8:40:28
LED 1
LED 3
Flush
mV Stab
+123.4
+12.3
+123.4
+12.3
Dark
mV Stab
+123.4 12.3
+123.4 12.3
LED 1
LED 3
Std 1
mV Stab
+123.4
+12.3
+123.4
+12.3
Std 2
mV Stab
+123.4 12.3
+123.4 12.3
Sample
mV Stab
+123.4 12.3
+123.4 12.3
LED 1
LED 3
Slope
+12.3
Errors
30 Aug 96
ENTER
Std C
+126.4
+273.0
mV
+70.1
+41.3
AD2
AD3
Flush
Std. A
Std. B
Std. C
Std. D
Gas A
Gas B
Flow Path
Hct
Std B
mV Stab
+314.3
0.4
STD C
mV Stab
+273.0
0.1
Slope
+66.4
Hct
Sample
mV Stab
+270.2
0.0
Std C
mV Stab
+272.1
0.5
Conc
48.
Errors 123
@@@@@@@@@@@@@@@@@@@@@@@@
Press CLEAR to Exit.
30 Aug 96
11:45:45
11:45:45
TRD
DCD
CTS
RIV
123
123
123
123
123
30 Aug 96
11:45:45
Prime
Prime
Prime
Prime
Prime
Prime
Prime
Purge
DSR
RTS
1
2
3
4
5
6
7
8
TXD
RXD
Errors
TXD
RXD
DSR
TRD
DCD
CTS
RTS
RIV
123
123
123
123
123
Conc
+123.4
ENTER
-LnDn
ENTER
Feb
Feb
Feb
Feb
Feb
Feb
Feb
Feb
Feb
Feb
Feb
Feb
Feb
Feb
Fnc/- PgUp
2
2
3
4
5
6
6
6
6
7
9
10
10
10
F4
S5
F4
F4
F4
F4
S5
F4
S5
F4
F4
F4
S5
S5
OXYGEN SATURATION
30 Aug 96
11:45:45
11:45:45
30 Aug 96
11:45:45
Ordering Information
Supplies and parts for the Stat Profile Ultra are available from Nova Biomedical.
DESCRIPTION
ITEM #
16017
16018
16019
16020
16016
16171
06007
06586
06587
06529
06005
00099
12704
08923
13409
09458
07158
16698
07161
17042
06588
06546
06548
08254
06545
06547
08897
16170
06555
06556
08161
09919
15507
F. Supplies
F-1
F. Supplies
F-2
DESCRIPTION
ITEM #
13511
13512
13513
13514
13515
13516
13517
13518
13519
13520
13521
13522
13523
13524
01498
06856
11272
17371
10641
07190
06857
07541
06553
07543
06013
11098
06554
11099
06009
11802
06516
06025
17393
02702
06523
06558
06517
ITEM #
Shorting Jumper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sodium Electrode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Stat Profile Ultra Reagent Pack F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Stat Profile Ultra Reagent Pack G . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Stat Profile Ultra Reagent Pack H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Stat Profile Ultra Reagent Pack I . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Stat Profile Ultra Reagent Pack J . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Stat Profile Ultra Reagent Pack K . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ultra Multi Control Level 1 with Glucose Over-Range . . . . . . . . . . . . . . .
Ultra Multi Control Level 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ultra Multi Control Multipack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
W/R-Tubing Harness Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
W/R-Tubing Segments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Washer Replacement Kit (ISE) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Washer Replacement Kit (pH) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Washer Replacement Kit (PCO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Washer Replacement Kit (PO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Washer Removal Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
09331
06011
18194
18195
18170
18196
18197
18198
18772
18773
18774
16824
07575
08153
09796
09797
09798
07157
F. Supplies
F-3
Appendix G Shutdown/Reinstallation
G
Shutdown/Reinstallation Procedures
G.1
Shutdown Procedure
1.
2.
3.
4.
5.
6.
7.
8.
9.
Clean the flow path with Cleaning Solution (PN 11802) per Section 4.1.5.1.
Perform a deionized water purge per steps 1 to 6 of Section 5.2.2.3 (also purging the
R-line with deionized water).
Cap the reagent pack bottles.
Remove the PCO2 and PO2 electrodes from the flow cell. Remove the membraned
caps from the PCO2 and the PO2 electrodes and drain the filling solution from
them. Rinse the parts of the electrodes that were in contact with the filling solution
with deionized water, replace the PCO2 and PO2 membraned caps onto the
electrodes for protection, see Sections 4.4.6 and 4.4.7, omitting replacement of the
filling solution. Replace the electrodes in the flow cell.
Unplug the glucose and lactate electrode cables from the rack by lifting the latch
up with 1 hand while pulling the connector out with the other hand.
Draw the water out of the gas humidifiers with a syringe.
Turn off the gases at the regulators.
Lift the pinch bar, located on the pinch valve, off the reagent harness tubing.
Unplug the instrument.
G. Shut/Rein
G-1
Reinstallation Procedure
For maximum efficiency follow this reinstallation procedure:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
G. Shut/Rein
14.
15.
16.
G-2
Appendix H Warranty
H
Warranty
Subject to the exclusions and upon the conditions specified below, Nova Biomedical or the
authorized Nova Biomedical distributor warrants that he will correct free of all charges
including labor, either by repair, or at his election, by replacement, any part of an instrument
which fails within one (1) year after delivery to the customer because of defective material or
workmanship. This warranty does not include normal wear from use and excludes: (A) Service
or parts required for repair to damage caused by accident, neglect, misuse, altering the Nova
equipment, unfavorable environmental conditions, electric current fluctuations, work performed by any party other than an authorized Nova representative or any force of nature; (B)
Work which, in the sole and exclusive opinion of Nova, is impractical to perform because of
location, alterations in the Nova equipment or connection of the Nova equipment to any other
device; (C) Specification changes; (D) Service required to parts in the system contacted or
otherwise affected by expendables or reagents not manufactured by Nova which cause
shortened life, erratic behavior, damage or poor analytical performance; (E) Service required
because of problems, which, in the sole and exclusive opinion of Nova, have been caused by
any unauthorized third party; or (F) Instrument refurbishing for cosmetic purposes. All parts
replaced under the original warranty will be warranted only until the end of the original instrument
warranty. All requests for warranty replacement must be received by Nova or their authorized distributor
within thirty (30) days after the component failure. Nova Biomedical reserves the right to change,
alter, modify or improve any of its instruments without any obligation to make corresponding
changes to any instrument previously sold or shipped. All service will be rendered during
Novas principal hours of operation. All requests for service outside Novas principal hours
of operation will be rendered at the prevailing weekend/holiday rates after receipt of an
authorized purchase order. Contact Nova for specific information.
The following exceptions apply:
1.
3.
H-1
H. Warranty
2.
The Glucose, Lactate, and BUN membranes are warranted as stated on the insert
that is shipped with the membranes, provided they are stored refrigerated and
placed into service prior to the expiration date on the packaging. This warranty is
invalid under the conditions specified after item 7.
The sodium, pH, PCO2, PO2, and reference electrodes, and the chloride, glucose,
and lactate electrode bodies are warranted for six (6) months from the date of
installation, provided they are stored at room temperature and placed into service
prior to the use before date on the packaging. In the event that an electrode does not
meet that use life, then Nova Biomedical will replace that electrode at no charge
under this warranty. This warranty is invalid under the conditions specified after
item 7.
Each potassium electrode that is purchased from Nova Biomedical will achieve
three (3) months of use life from the date of installation, provided they are stored
at room temperature and placed into service prior to the expiration date on the
4.
5.
6.
7.
packaging. A majority of the electrodes will achieve that use life without intervention from Nova. In the unlikely event that an electrode you purchase does not meet
that use life, Nova Biomedical will replace that electrode at no charge under this
warranty. This warranty is invalid under the conditions specified after item 7.
The chloride cap is warranted as stated on the insert that is shipped with the caps. This
warranty is invalid under the conditions specified after item 7.
The ionized calcium and BUN electrodes are warranted as stated on the insert that
is shipped with the electrode. This warranty is invalid under the conditions
specified after item 7.
Consumable items, including the reagent pack, calibration gases, replaceable
membranes, tubing and tubing harnesses, electrolyte solutions, external standards,
and septum assemblies are warranted to be free of defects at time of installation.
The item must be placed into service prior to the expiration date printed on the
packaging. All defects must be promptly reported to Nova Biomedical in writing.
This warranty is invalid under the conditions specified after item 7.
Freight is paid by the customer.
H. Warranty
H-2
Index
A
ABG Calculated Results Reporting screen 2-17, 3-34
Acceptable samples .............................. A-1, 1-3, 1-5
Activity coefficient ................................................B-3
Adapting the program ............................................ 2-1
Air Bath Not Ready ............................................. 3-17
Air detector status ................................................ 3-19
Air Detector Status screens ................................. 3-19
Air detector test ................................................... 5-96
Air Detectors and Hct screen ............................. 5-108
Air detectors, subsystem test ............................. 5-107
Air Detectors Uncalibrated .................................. 3-17
Alternate analysis (expired gas) ............................ 4-2
Altitude effects .................................................... 5-88
Alveolar oxygen (A) ............................................ B-14
Ammonium ion selective electrode ..................... B-10
Analog input data .............................................. 5-103
Analysis ............................................................... 3-30
Analysis configuration ........................................... 2-3
Analysis Configuration screen ............................... 2-4
Analysis Output Options screen ............................ 2-4
Analysis signal data ........................................... 5-104
Analytical compartment ................................ 3-5, 3-6
ANALYZE key ...................................................... 3-4
Analyzer identification .......................................... 2-8
Analyzer, lifting or moving .................................. A-6
Analyzer reset .................................................... 5-120
Analyzing samples ............................................... 3-30
Anion gap ............................................................B-16
Anticoagulant .............................................. 1-7, 5-91
Aqueous controls, problems with ........................ 5-87
Arterial alveolar oxygen tension gradient (AaDO2) B-14
Arterial alveolar oxygen tension ratio .................B-14
Auto-Cal .............................................................. 3-25
B
Bar code wand port test ..................................... 5-112
Barometer ............................................................. A-2
Barometric pressure ..................................... 2-6, 3-17
Barometric pressure effects ................................. 5-88
Barometric pressure module .................................. 3-9
Barometric Pressure screen ................................... 2-6
Base excess extracellular fluid (BE-ECF) ...........B-13
Battery ................................................................. 2-31
Bicarbonate concentration ...................................B-11
Blank electrode suppression ................................ 2-13
Blood Gas/Electrolyte/Chemistry Controls ........... C-5
Blood gases ............................................................ 1-3
Blood gases (slope limits) ................................. 5-104
Blood Gases and pH screen ............................... 5-105
C
Cabinet cleaning ................................................... A-6
Cal I ....................................................................... 4-5
Calcium ............................................ B-5, B-15, 4-16
Calcium electrode replacement ........................... 4-14
Calcium, ionized .................................................... 1-4
Calculated parameters .........................................B-11
Calculated results ......................................... 1-2, 3-13
Calculated Results screen .......................... 2-13, 3-37
Calculated values ...................................... B-10, 2-15
Calculating sample concentration ..........................B-6
CALIBRATE key .................................................. 3-2
Calibrating gases ................................................... C-2
Calibration ........................................................... 3-24
Calibration configuration ....................................... 2-5
Calibration gas values ................................ 3-16, 5-88
Calibration gases .......................................... A-5, 4-9
Calibration gases composition ...............................C-2
Calibration options ...................................... 3-25, 4-3
Calibration Options screen .................................... 4-3
Calibration signal data ....................................... 5-104
Calibration standards ............................................. 5-3
Calibrations, terminating ....................................... 4-4
Cancel Print-out ................................................... 3-28
Capillary samples ........................................ 1-5, 3-31
Capillary samples, running .................................. 3-39
Capillary-adapter fitting ...................................... 3-39
Carbon dioxide content (TCO2) ........................... B-12
Check septum assembly message ........................ 5-92
Chloride .............................................. B-5, 1-4, 4-16
Chloride cap ......................................................... 4-16
Chloride cap filling and installation .................... 4-16
Chloride electrode replacement ........................... 4-14
Clean sample preheater message ......................... 5-92
Cleaning, septum assembly inlet port .................. 4-12
Cleaning the cabinet ............................................. A-6
Cleaning the CRT ................................................. A-6
CLEAR key ........................................................... 3-3
Clinical utility ........................................................ 1-3
Communication states ........................................... 6-2
Communications interface ..................................... 6-1
Communications setup ........................................ 2-24
Communications Setup screen ............................ 2-24
Communications Test screen ............................. 5-116
I-1
Index
D
Daily Data to Month-to-Date, move .................... 2-21
Daily QC Analysis Log ....................................... 3-27
Daily QC Analysis Log, print .............................. 2-21
Daily QC Summary Graphs ................................. 3-27
Daily QC Summary Graphs, print ....................... 2-21
Date and Time screen ............................................ 2-7
Date, setting ........................................................... 2-7
Decimal Point key ................................................. 3-4
Default functions ................................................... 2-1
Default parameter ................................................ 2-25
Default Parameter screen ..................................... 2-25
Definition of PCO2 .................................................................................. B-7
Definition of pH .....................................................B-7
Definition of PO2 ...................................................................................... B-8
Degasing the electrode ......................................... 4-27
Delays in transit ................................................... 5-91
Deproteinizing solution, cleaning .......................... 4-8
Digit keys ............................................................... 3-4
Dimensions ........................................................... A-5
Disabled ............................................................... 2-20
Disk drive utilities ............................................. 5-118
Disk Operations screen ...................................... 5-119
E
Electrical compliance ........................................... A-5
Electrical requirements .......................................... 1-1
Electrode calibration .............................................. B-5
Electrode conditioning holder ............................. 4-19
Electrode drift ...................................................... 3-13
Electrode membrane cap insertion ...................... 4-27
Electrode offset .................................................... 2-11
Electrode Offset screen ........................................ 2-12
I-2
F
Fiber optic module ......................................... 3-6, 3-9
Fiber optics ............................................................ 4-5
Filling, gas humidifier well ................................. 4-13
Filling the PO2 membrane cap ............................ 4-31
FIO2 ......................................................................................................................... 2-26
Flow Cell ............................................................... 3-6
Flow cell ................................................................ 3-8
Flow cell conditioning ......................................... 4-11
Flow cell flush ..................................................... 5-83
Flow cell light ........................................ 3-6, 3-9, 4-2
Flow cell replacement .......................................... 4-56
Flow cell washers ................. 4-15, 4-18, 4-25, 4-30
Flow path air flush ............................................... 5-80
Flow path cleaning ................................................. 4-6
Flow Path Cleaning screen .................................... 4-7
Flow path flushing ............................................... 5-79
Flow path leaks .................................................... 5-86
Flow test .............................................................. 5-77
Fluid prime ........................................................ 5-111
Fluid prime options ................................................ 4-6
Fluid/Gas Prime Options screen ................ 4-6, 5-111
Fluidic diagram .................................................... 3-12
Fluidics ................................................................ 5-90
Full calibration ....................................................... 4-4
FUNCTION key .................................................... 3-3
Functions ............................................................... 3-2
Fuse replacement ................................................. 4-65
Fuse requirements .................................................. 1-1
Fuses .................................................................... 4-66
Index
G
Gas calibrating requirements ................................. 1-2
Gas cylinders filter replacement .......................... 4-63
Gas cylinders hose fitting .................................... 4-63
Gas cylinders regulator adjustment ..................... 4-63
Gas cylinders replacement ................................... 4-63
Gas filter replacement .......................................... 4-64
Gas hose connection ............................................ 4-64
Gas humidifier ..................................................... 5-86
Gas humidifier well fill levels ............................. 4-13
Gas humidifier well filling .................................. 4-13
Gas humidifier wells ...................................... 3-6, 3-7
Gas pressure test .................................................. 5-85
Gas prime ........................................................... 5-111
Gas prime options .................................................. 4-6
Gas regulator adjustment ..................................... 4-63
Gas test .............................................................. 5-101
Gas values ............................................................ 5-88
Gases ...................................................................... 5-3
Gases, calibrating .................................................. C-2
Glucose ......................................................... B-9, 1-4
Glucose electrode membraning ........................... 4-19
Glucose electrode polishing ................................ 4-19
Glucose electrode replacement ............................ 4-19
Glucose electrode tip polishing ........................... 4-20
Glucose, elevated ................................................... 1-6
Glucose Over-Range solution ................................C-3
Glucose slope limits .......................................... 5-107
Glucose, subsystem test ..................................... 5-106
H
Handling requirements .......................................... 1-5
Harness replacement ............................................ 4-47
Hct, subsystem test ............................................ 5-107
Hematocrit ........................................... B-8, C-4, 1-4
Hematocrit controls ............................................... C-1
Hematocrit slope limits ...................................... 5-108
Hemoglobin .............................................. B-12, 2-26
High-level protocol ................................................ 6-3
Humidifier leaks .................................................. 5-86
I
I/O devices ......................................................... 5-111
I/O Devices Menu screen .................................. 5-112
Idle state ................................................................. 6-2
Individual component flush ................................. 5-80
Inspired oxygen fraction ratio .............................B-16
Installation ............................................................. 1-1
Installing new software .......................................... 2-1
J
Junction potential ................................................... B-4
K
Keypad ................................................................... 3-2
L
Lactate .......................................................... B-9, 1-3
Lactate electrode membraning ............................ 4-21
Lactate electrode polishing .................................. 4-21
Lactate electrode replacement ............................. 4-21
Lactate electrode tip polishing ............................. 4-22
Lactate, elevated .................................................... 1-6
Lactate slope limits ............................................ 5-107
Lactate, subsystem test ...................................... 5-106
Levey-Jennings Graphs ....................................... 3-28
Levey-Jennings Graphs, print .............................. 2-21
Lifting the analyzer ............................................... A-6
Link establishment state ........................................ 6-2
Link termination state ............................................ 6-3
Loop-Back Test screen .......................... 5-117, 5-118
Low-level protocol ................................................. 6-1
M
Main menu ........................................................... 3-24
Main Menu screen ............................................... 3-24
Maintenance .......................................................... 4-1
Maintenance Menu ................................................ 4-1
Maintenance Menu screen ..................................... 4-2
Maintenance procedures ...................................... 4-10
Maintenance, scheduled ...................................... 4-10
Mandatory QC ..................................................... 2-20
Manual calibration ....................................... 3-25, 4-4
Matrix effects ....................................................... 5-90
Measured Results Reporting screen 2-13, 2-16, 3-34
Measured Results screen ..................................... 3-36
Measured tests ..................................................... 2-15
I-3
Index
N
Nernst equation ......................................................B-2
No Reagent/Gas When Required ......................... 3-17
Not Ready screen ................................................. 3-16
O
O2 saturation ......................................................... 1-3
Offset limits ......................................................... 2-12
One-point calibration .................................. B-5, 3-13
Operation ............................................................... 2-1
Operation configuration ......................................... 2-3
Operation Configuration screen ............................. 2-3
Operation overview ............................................. 3-13
Operator passwords ............................................. 2-27
Ordering information ............................................. F-1
Osmolality ...........................................................B-16
Over-Range solution .............................................. C-3
Override mandatory QC analysis ........................ 3-29
Oxygen content (O2Ct) ........................................B-13
Oxygen fraction ratio, inspired ............................B-16
Oxygen saturation ..................................... C-4, 5-109
Oxygen saturation calibration ................................ 4-5
Oxygen saturation controls ....................................C-1
Oxygen Saturation screen .................................. 5-109
Oxygenation index ...............................................B-16
P
Panic ranges ......................................................... 2-18
Paper installation ................................................. 4-62
Paper release ........................................................ 4-61
Partial pressure of carbon dioxide (PCO2) ............B-7
Partial pressure of oxygen (PCO2) ........................B-8
Parts ....................................................................... F-1
Password ................................................................ 2-1
Password options ................................................. 2-26
Password Options screen ..................................... 2-27
I-4
Index
Q
QC key ................................................................... 3-2
QC key (pressed in error) .................................... 3-29
QC lockout mode ................................................. 2-20
QC procedures .......................................................C-3
QC samples, running ........................................... 3-28
Quality Control .............................................C-4, C-5
Quality Control Configuration screen ................. 2-19
Quality Control Menu .......................................... 3-26
Quality control setup ........................................... 2-19
Quality Control Setup screen ..................... 2-19, 2-21
R
Ready for Analysis screen ................................... 3-15
Reagent flow controllers ...................................... 3-10
Reagent harness ............................................. 3-6, 3-7
Reagent harness maintenance .............................. 4-47
Reagent harness replacement .............................. 4-47
Reagent Management Pod installation ................ 4-52
Reagent Management System ............................. 4-50
Reagent Management System overview .............. 3-23
Reagent pack .............................................. C-2, 3-11
Reagent pack replacement ................................... 4-50
Reagent Pack screen ............................................ 3-23
Reagent Pack Status screen ................................. 4-51
Reagent pack tubing positions ............................. 4-52
Reagent preheater .......................................... 3-6, 3-7
Reagents ...................................................... A-2, C-1
Reference and Panic Ranges screen .................... 2-18
Reference electrode ....................................... 3-6, 3-9
Reference electrode flush .................................... 5-82
Reference electrode replacement ......................... 4-33
Reference ranges ................................................. 2-18
Reference solution flow check .......................... 5-102
Reference values ................................................... D-1
Reinstallation procedure .............................. G-1, G-2
Reminder only ..................................................... 2-20
S
S-line replacement ............................................... 4-53
Sample ................................................................... 1-4
Sample analysis printouts .................................... 3-38
Sample cup samples ............................................ 3-31
Sample handling .................................................. 5-91
Sample number counter ......................................... 2-8
Sample Number Counter screen ............................ 2-9
Sample preheater ........................................... 3-6, 3-7
Sample preheater flush ........................................ 5-84
Sample Preheater Not Ready ............................... 3-17
Sample probe ......................................................... 3-6
Sample probe flush .............................................. 5-80
Sample probe position test ................................... 5-97
Sample probe positions ........................................ 5-98
Sample probe replacement .................................. 4-53
Sample volume ..................................................... A-2
Sampler assembly .................................................. 3-7
Scheduled maintenance ....................................... 4-10
Sensor status ........................................................ 3-18
I-5
Index
I-6
T
Tailpiece, replacing .............................................. 4-42
Technician passwords .......................................... 2-28
Temperature .......................................................... A-5
Temperature control, subsystem ........................ 5-110
Temperature correction for measured values .......B-11
Terminating calibrations ........................................ 4-4
Test, air detector .................................................. 5-96
Test panel ............................................................. 2-15
Test Panels screen ................................................ 3-30
TEST SELECT key ............................................... 3-4
Test, shorting jumper ........................................... 5-95
Tests from test panel ............................................ 2-15
Tests performed ..................................................... 1-2
Theory .................................................................... B-1
Time, setting .......................................................... 2-7
Transfer state ......................................................... 6-3
Transmission port configuration .......................... 2-24
Transmission Port Configuration screen ............. 2-25
Troubleshooting ............................................. 5-1, 5-2
Troubleshooting results ....................................... 5-87
Two-point calibration ............... B-5, 3-13, 3-25, 4-4
Two-point calibration sequence ..................... 5-3, 5-4
U
Units Of measurement, set .................................. 2-10
Urea Nitrogen ........................................................ 1-4
V
Vacuum tube samples .......................................... 3-31
Valve positions ................................................... 5-100
Valve test ........................................................... 5-100
Vent hole .............................................................. 4-26
Verifying performance ...........................................C-3
W
W/R harness ........................................................... 3-9
W/R-harness .......................................................... 3-6
W/R-harness replacement .................................... 4-45
Warranty ...................................................... A-6, H-1
Water test ............................................................. 5-77
Weight ................................................................... A-5
Distributor
Address
City
State
Zip
FACTORY COPY
(to be sent to Waltham, Massachusetts)
Operator name
Laboratory
Company or institution
Address
City
Telephone No. (
State
Zip
State
Installed by:
Date
Zip
(Customer Signature)
NOVA BIOMEDICAL
CUSTOMERS WARRANTY
STAT PROFILE ULTRA (F, G, H, I, J, K) ELECTROLYTE ANALYZER
Subject to the exclusions and upon the conditions specified below, Nova Biomedical
or the authorized Nova Biomedical distributor warrants that he will correct free of
all charges including labor, either by repair, or at his election, by replacement, any
part of an instrument which fails within one (1) year after delivery to the customer
because of defective material or workmanship. This warranty does not include
normal wear from use and excludes: (A) Service or parts required for repair to
damage caused by accident, neglect, breakage, misuse, altering the Nova equipment,
unfavorable environmental conditions, electric current fluctuations, work performed by any party other than an authorized Nova representative, or any force of
nature; (B) Work which, in the sole and exclusive opinion of Nova, is impractical
to perform because of location, alterations in the Nova equipment or connection of
the Nova equipment to any other device; (C) Specification changes; (D) Service
required to parts in the system contacted or otherwise affected by expendables or
reagents not manufactured by Nova which cause shortened life, erratic behavior,
damage or poor analytical performance; (E) Service required because of problems,
which, in the sole and exclusive opinion of Nova, have been caused by any
unauthorized third party; or (F) Instrument refurbishing for cosmetic purposes. All
parts replaced under the original warranty will be warranted only until the end of the original
instrument warranty. All requests for warranty replacement must be received by Nova or their
authorized distributor within thirty (30) days after the component failure. Nova Biomedical
reserves the right to change, alter, modify or improve any of its instruments without
any obligation to make corresponding changes to any instrument previously sold or
shipped. All service will be rendered during Novas principal hours of operation. All
requests for service outside Novas principal hours of operation will be rendered at
the prevailing weekend/holiday rates after receipt of an authorized purchase order.
Contact Nova for specific information.
The following exceptions apply:
1. The Glucose, Lactate, and BUN membranes are warranted as stated on the insert
that is shipped with the membranes, provided they are stored refrigerated and
placed into service prior to the expiration date on the packaging. This warranty
is invalid under the conditions specified after item 7.
2. The sodium, pH, PCO2, PO2, and reference electrodes, and the glucose and
lactate electrode bodies are warranted for six (6) months from the date of
installation, provided they are stored at room temperature and placed into service
prior to the use before date on the packaging. In the event that an electrode does
not meet that use life, then Nova Biomedical will replace that electrode at no
charge under this warranty. This warranty is invalid under the conditions
specified after item 7.
3. Each potassium electrode that is purchased from Nova Biomedical will achieve
three (3) months of use life from the date of installation, provided they are stored
at room temperature and placed into service prior to the expiration date on the
4.
5.
6.
7.
packaging. A majority of the electrodes will achieve that use life without intervention from Nova. In the unlikely event that an electrode you purchase does not meet
that use life, Nova Biomedical will replace that electrode at no charge under this
warranty. This warranty is invalid under the conditions specified after item 7.
The chloride electrode and chloride caps are warranted as stated on the insert that
is shipped with the electrode and the caps. This warranty is invalid under the
conditions specified after item 7.
The ionized calcium electrode is warranted as stated on the insert that is shipped
with the electrode. This warranty is invalid under the conditions specified after
item 7.
Consumable items, including the reagent pack, calibration gases, replaceable
membranes, tubing and tubing harnesses, electrolyte solutions, external standards,
and septum assemblies are warranted to be free of defects at time of installation. The
item must be placed into service prior to the expiration date printed on the
packaging. All defects must be promptly reported to Nova Biomedical in writing.
This warranty is invalid under the conditions specified after item 7.
Freight is paid by the customer.
INSTRUCTIONS
NO POSTAGE
NECESSARY IF
MAILED IN THE
UNITED STATES
Nova Biomedical
NOVA
Biomedical
200 PROSPECT STREET
PO BOX 9141
WALTHAM, MA 02254-9825
1. Intro.
NOTE: All other parameters should be obtained from the undiluted, unmodified blood sample.
1. Mix the sample in the syringe. Pipet 0.5 mL into a conical sample cup.
2. Pipet 1.0 mL of isotonic saline into this sample cup. Mix well.
3. Aspirate this diluted sample into the Ultra to measure the glucose concentration.
Multiply the observed concentration by 3.0 to establish the initial concentration.
NOTE: All other parameters should be obtained from the undiluted, unmodified blood sample.
1. Mix the sample in the syringe. Pipet 0.5 mL into a conical sample cup.
2. Pipet 1.0 mL of isotonic saline into this sample cup. Mix well.
3. Aspirate this diluted sample into the Stat Profile to measure the lactate concentration.
Multiply the observed concentration by 3.0 to establish the initial concentration.
1-6
1 Introduction
1. Intro.
NOTE: All other parameters should be obtained from the undiluted, unmodified blood sample.
1. Mix the sample in the syringe. Pipet 0.5 mL into a conical sample cup.
2. Pipet 0.5 mL of isotonic saline into this sample cup. Mix well.
3. Aspirate this diluted sample into the Stat Profile Ultra M Analyzer to measure the
BUN concentration. Multiply the observed concentration by 2.0 to establish the
initial concentration.
1-7
1. Intro.
SO2% Interferences:
High levels of COHb and MetHb will increase the reported SO2% value.
MetHb values above 15% will interfere with the SO2% value.
Bilirubin values greater than 10 mg/dL will interfere with the SO2% value.
Lipid solutions of greater than 50 mg/dL will interfere with the SO2% value.
Hemolysed samples will interfere with the SO2% value.
Hematocrit (Hct) Interference:
White Blood Count (WBC) greater than 50,000 WBC/L may increase the hematocrit value.
1.4
1-8