Professional Documents
Culture Documents
Indira Soundiram
2012
GeneXpert
Module
2010 Cepheid
GX-XVI
GX-I
GX-II
GX-IV
Any Test
Any Time
Any Sample
Any Place
The GeneXpert
Family of Systems
GeneXpert
Module
GeneXpert
Cartridge
SmartCycler
System
ICORE
Module
2010 Cepheid
SmartCycler
Tube
ACCURATE
Results
4
2010 Cepheid
FAST
Answers
EASY
To Use
Cartridge
Self-Contained
Disposable
Computer System
Software
Barcode scanner
Optional Accessories
Printer
UPS
5
2010 Cepheid
GeneXpert Technology
Instrument platform
Cartridges
Protocols and Controls
2010 Cepheid
GeneXpert Technology
Employ random access fluidic systems that bridge microfluidics with macro
fluidics
specimen types
7
2010 Cepheid
2010 Cepheid
Cepheid Technology
GeneXpert
Washing
DNA/RNA
extraction
DNA or RAN
purification
Patient Sample
MasterMix
SmartCycler
Positive
Ngative
2010 Cepheid
Analysis
Detection
Amplification
in less than 2h
7. View/print
results about
50 minutes later!
10
2010 Cepheid
Platforms
SmartCycler Instrument
I-CORE module
GeneXpert Instrument
GX module
BDS
11
2010 Cepheid
12
2010 Cepheid
13
2010 Cepheid
Syringe Motor
Processing
Chambers
Motherboard
I-CORE
Module Door
Valve Drive Motor
14
2010 Cepheid
Reaction
tube
Valve
Body
Cepheid Technology
GeneXpert
15
2010 Cepheid
SmartCycler
Cepheid Cartridge
Syringe drive
I-CORE
Valve drive
16
2010 Cepheid
Ultrasonic horn
GeneXpert Module
Syringe Motor
Motherboard
Uniframe
I-CORE
Cartridge
Inserter
Ultrasonic
Horn
17
2010 Cepheid
Inside Chamber
Syringe drive shaft
ICORE tube insertion
inches
Ultrasonic horn
18
2010 Cepheid
GeneXpert Cartridge
GX cartridge has 11 chambers
Lower numbered chambers are used for sample
processing
Cartridge A
Cartridge C
2010 Cepheid
Pre-filter or
Funnel
Cartridge A
only
20
2010 Cepheid
Cartridge Body
(11 Fluid Chambers and
overmolded gasket)
Ports for
PCR tube
PCR
Tube
Valve body Ports
. .
Syringe Barrel
Rotary Valve/Filter and
Ultrasonic Lysis or SPB
Region
Cap/Ultrasonic Interface
Cartridge Foot
21
2010 Cepheid
22
2010 Cepheid
Cartridge C
The use of external controls will have limited use in assuring proper
assay performance
Cepheid designed specific molecular methods to include internal
Three elements of the molecular design were created to address all the
23
2010 Cepheid
Probe Check
What is it?
software.
What does it do?
Probe check verifies
Bead re-hydration
Probe integrity
24
2010 Cepheid
Primers: Can be the same as those for the target or IC or can be unique; present in
primer-probe bead
The SPC assesses the effectiveness of the sample processing steps, including and up
to reaction tube filling
25
2010 Cepheid
26
2010 Cepheid
27
2010 Cepheid
28
2010 Cepheid
29
2010 Cepheid
Interpretation of Results
Probe Check
Pass
Fail
Error
Target
Pos
No Value
SPC IC
Positive
Pos
Negative
30
2010 Cepheid
Test
Stopped
No Value
Invalid
No Value