The radiographic system described in this manual refers to a wall installation.

de Gtzen S.r.l. reserves itself the right to modify the production and the manual without notice on their
designated distribution.
de Gtzen S.r.l. shall not be liable for the incorrect use of the information contained in this
manual.
Copies, even excerpts, of this manual are allowed only for personal use and with written consent of
the manufacturer.

Edition: 01/2004

Drawn up by A. Colombo

Manual Revision
1.0

Approved by M. de Gtzen

Date
01/2004

Description
English Version

xgenus ac/dc

SUMMARY

INTRODUCTION
SUMMARY
PRELIMINARY INFORMATION
INFORMATION FOR THE INSTALLER
1.
2.
3.
4.
5.

6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.

xgenus ac/dc RADIOGRAPHIC SYSTEMS


PICTOGRAMS USED
OVERALL DIMENSIONS
INSTALLATION SPECIFICATIONS
INSTALLATION
5.1 UNPACKING
5.2 ASSEMBLING THE WALL PLATE
5.3 ASSEMBLING THE BRACKET/HORIZONTAL ARM
5.4 ASSEMBLING THE POSITIONER ARM
5.5 CONNECTION TO THE TERMINAL BOARD
5.6 ASSEMBLING THE RADIOGRAPHIC UNIT
5.7 BALANCING THE POSITIONER ARM
5.8 ASSEMBLING THE xgenus ac/dc X-RAY CONTROL
5.9 X-RAY CONTROL ELECTRICAL CONNECTION
X-RAY CONTROL PANEL
SYSTEM CONFIGURATION
CHANGING THE CONFIGURATION
START UP
CHECKING THE INSTALLATION
CHECKING THE EXPOSURE FACTORS
DIAGNOSTIC
TUBE HEAD CALIBRATION
ERROR MESSAGES
REPLACEMENT OF FUSES
MAINTENANCE
CLEANING THE OUTER SURFACES
IF A REPAIR SHOULD BECOME NECESSARY
ATTACHMENTS

xgenus ac/dc

page
page
page
page

2
3
4
5

page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page

6
7
9
13
15
15
16
18
20
21
23
26
29
31
33
35
38
40
41
43
46
47
48
49
50
51
51
52

PRELIMINARY INFORMATION
Before beginning to use the xgenus ac/dc radiographic systems, it is advisable to carefully read
and follow the instructions contained herein, so as to obtain the best possible performance.
Always pay close attention to the CAUTION, WARNING, and PLEASE NOTE messages when
operating the system.

LEGEND

CAUTION
The word CAUTION identifies those situations that might compromise the operator's personal
safety or cause injuries to others.

WARNING
The word WARNING identifies those occurrences which might compromise the xgenus ac/dc
x-ray systems performance.
PLEASE NOTE
The words PLEASE NOTE serves to draw your attention to special indications to facilitate
maintenance or make important information more clear.

xgenus ac/dc

INFORMATION FOR THE INSTALLER

CAUTION
The installer is responsible for the installation, calibration and testing with regard to
system safety and efficient operation.
For safe and reliable installation of the xgenus ac/dc radiographic systems, it is advisable to:

Check that the voltage mentioned on the rating plate matches the line voltage.

Install the xgenus ac/dc radiographic systems according to the procedures described
in this manual.

Provide the user with any information regarding the use of the xgenus ac/dc
radiographic systems according to that which is provided in the operators manual.

Certify the work done by completing the Declaration of Conformity form.

Return the completed warranty certificate to: TG Group.


Only when the completed warranty certificate is received is the warranty valid.

xgenus ac/dc

1. xgenus ac/dc RADIOGRAPHIC SYSTEMS


The xgenus ac/dc radiographic systems (Fig. 1) consists of:








TUBE HEAD XGG / TUBEHEAD XGDC


CONE
POSITIONER ARM (Scissor)
CONTROL XGT
WALL BOX
HORIZONTAL ARM

OPTIONAL ACCESSORIES :
long 12 (31cm) cone
short 8 (20cm) cone
long cone 12 (31cm) with a rectangular section sized 44x35mm
-

remote control button: xgenus ECB

Fig. 1




xgenus dc only

xgenus ac/dc

2. PICTOGRAMS USED

The identification tags on the tube heads XGG/XGDC , on the control XGT and on the cone indicate the
model name, the serial number, the manufacturing date and the symbols of the main technical
characteristics.

ID TAG OF THE
xgenus ac
TUBE HEAD XGG

ID TAG OF THE
xgenus dc
TUBE HEAD XGDC

ID TAG OF THE
xgenus
CONTROL XGT

ID TAG OF THE SHORT 8 CONE: BLD20 - 60mm (20 cm SSD)


ID TAG OF THE LONG 12 CONE: BLD31 - 60mm (31 cm SSD)
ID TAG OF THE RECTANGULAR CONE: BLD31SQ - 44x35mm (31 cm SSD)

GRADUATED SCALE TAG ON TUBE HEADS

xgenus dc only

xgenus ac/dc

SYMBOLS:
This symbol guarantees that the radiographic system complies with the regulations
contained in the European Directive EEC 93/42 regarding Medical Devices

The degree of protection against direct and indirect electric contacts is B type

Symbol indicating danger due to ionizing radiation

Refer to Manual's instructions

NEUTRAL" power supply wire

LIVE" power supply wire


Ground cable

xgenus ac/dc

3. OVERALL DIMENSIONS
Fig. 1A, 1B, 1C are the overall dimensions to accommodate varying installation situations:

SHORT ARM (optional)




length 41cm

16.2

STANDARD ARM

length 82,5cm

32.5

LONG ARM (optional)

length 110cm

43.5

xgenus ac/dc

SHORT ARM 41cm (16.2)

Fig. 1A

STANDARD ARM 82.5cm (32.5)

Fig. 1B

LONG ARM 110cm (43.5)

Fig. 1C

Fig. 2, 3A, 3B, 3C e 4 show the typical dimensions of the radiographic system:

STANDARD ARM 82.5cm (32.5) ONE INCH = 2.54 cm

Fig. 2

10

xgenus ac/dc

STANDARD ARM 82.5cm (32.5)

Fig. 3B
SHORT ARM 41cm (16.2)
ONE INCH = 2.54cm

xgenus ac/dc

Fig. 3A

Fig. 3C
LONG ARM 110cm (43.5)

11

ONE INCH = 2.54 cm

12

Fig. 4

xgenus ac/dc

4. INSTALLATION SPECIFICATIONS

WARNING
Prior to installing the radiographic system the Office Owner must be sure that the operatory and
its electrical services comply with all governing regulations and required installation specifications.
Otherwise one must bring them all to compliance.
1.

OPERATORY REQUIREMENTS
a. The operatory must be of suitable width to accommodate the sizes and overall
dimensions provided (refer to section 3 OVERALL DIMENSIONS).
Check that no obstacles are present while positioning the radiographic system.
b. The operatory must not be pressurized or constitute an explosion hazard.
c.

While operating, the ambient temperature must be within +5C and +40C

d. The storage temperature must be within -15C and +50C


e. The relative humidity must be within 25% and 75%
2.

REQUIREMENTS OF THE SUPPORTING WALL


a. The supporting wall for the radiographic system must be able to withstand 200Kg sheer
stress at every bolt.
b. The subject wall material and construction must be checked. If required, consult a
contractor to determine strength and stability.
c. If there is any doubt as to the strength or stability of the wall to support the x-ray,
reinforce by providing either a steel or double plywood plate. Refer to the Pre-installation
drawings.

3.

REQUIREMENTS OF THE ELECTRIC SYSTEM


a. The electric system must comply with all applicable regulatory bodies.
b. The electric system must be able to supply the power-voltage and current required by
the Manufacturer's rating plate of the radiographic system as per Chart 1

CHART 1

Model 

Input voltage of the radiographic systems

NOMINAL LINE VOLTAGE


MINIMUM LINE VOLTAGE
MAXIMUM LINE VOLTAGE
FREQUENCY
POWER DRAWN DURING EXPOSURE
(MAX. TIME = 3.9 SEC)

xgenus ac

xgenus dc

-5% +10%

115V

115V
15%

115V
109 Vac
126.5 Vac
50/60 Hz

115V
98 Vac
132 Vac
50/60 Hz

800 VA

1.4 kVA

PLEASE NOTE

the communication cables (C11, C12 C21, C22) between the timer and the tube head must be two-pole, twisted
and shielded with 0,25mm2 (ex.: Belden 9501 type).

xgenus dc only

xgenus ac/dc

13

4.

REQUIREMENTS OF THE ELECTRIC LINE


a.

The electric line must be "single-phase" type.

b.

Separate circuit of 15 amps to be supplied.

c.

Wire size according to AWG.

PLEASE NOTE
For longer lines, the conductor gauge must be increased in proportion.

Model 
Line voltage

MAXIMUM LINE VOLTAGE DROP


MINIMUM LINE APPARENT RESISTANCE

5.

xgenus ac

xgenus dc

-5% +10%

115V

115V
15%

1.2%
0.2

3%
0.2

ELECTRICAL CONNECTIONS

WARNING
Prior to installing the radiographic system, it is advisable that the electrical supply complies with
the recommended specifications in this manual.

X-ray control electrical connections

CAUTION
According to relevant and practical standards, the x-ray control must be installed in a
position allowing the operator to see and have full control of the radiographic
exposure.
On the x-ray control installation wall, suitable runs for the following electrical cables must be
provided, according to the electrical installation diagram (refer to section 18
ATTACHMENTS):
a.

X-ray control electrical cables


(refer to section 5.9 CONTROL ELECTRIC CONNECTIONS)

b.

Cables connecting the control and the radiographic unit


(refer to section 5.9 CONTROL ELECTRIC CONNECTIONS )

Radiographic unit electric connections


To the wall plate location on the installation wall, a suitable run for the cable connecting the
x-ray control and the radiographic unit must be provided
(refer to section 5.5 CONNECTION TO THE POWER SUPPLY TERMINAL BOARD)

14

xgenus ac/dc

5. INSTALLATION

CAUTION
The xgenus ac/dc radiographic systems must be installed by professionally trained
technicians, who must be able to certify their work.

CAUTION
Prior to proceeding with the installation, check that all requirements are met.

5.1 UNPACKING
The components of the xgenus ac/dc radiographic systems are delivered by the factory securely
packaged within a specifically designed box, as shown in the following sketch (Fig. 5):






DOCUMENTS: INSTRUCTIONS MANUAL, INSTALLATION & MAINTENANCE MANUAL, ANNUAL


MAINTENANCE INSTRUCTIONS AND WARRANTY
RADIOGRAPHIC UNIT PACKAGING
X-RAY CONTROL PACKAGING + EXTERNAL LAMP (IF PROVIDED)
POSITIONER ARM PACKAGING
HORIZONTAL ARM AND WALL BOX PACKAGING

PLEASE NOTE
Prior to installation, please check all components for shipping damage. Advise the carrier.
PLEASE NOTE
The carton materials and the polystyrene foam can be completely recycled or disposed of by
authorized recycling companies.
PLEASE NOTE
It is advisable to store the original packaging to return the goods for repairs if required.

xgenus ac/dc

15

5.2 ASSEMBLING THE WALL PLATE


ASSEMBLY INSTRUCTIONS (Fig. 6)
1. Take the wall plate out of the packaging (refer to Fig. 5) Use the drilling template provided

2. Position the drilling template  on the radiographic system installation wall, at the required
height (130 cm from the base is the suggested height)
3. Fix the template

 to the wall with adhesive tape

4. Check the holes for vertical and horizontal alignment with the floor, using a plumb line
5. Mark the plate fixing holes
6. Mark the holes for the electric cables connecting the x-ray control to the radiographic unit
PLEASE NOTE
To prevent any flaking of the white paint coat and to control the center distances between the
holes, it is advisable to start drilling with a tip 7, increasing this measure gradually.
7. Drill the plate fixing holes
8. If required, drill the holes for the electric cables connecting the x-ray control to the
radiographic unit
9. Remove the template
characteristics 
10. Unscrew the screw

 and insert the suitable anchor screws, according to the wall

 and remove the plug  from the wall plate

11. Withdraw the sliding cover

12. Attach the plate  to the wall and insert the screws
tighten alternately

 with the relevant washers, then

13. Check that the plate is securely fixed to the wall and is plumb and level.

PLEASE NOTE
If the wall is not perfectly leveled, put a suitable shim between the wall and the plate, so as to
prevent any possible warping of the mounting plate.

16

xgenus ac/dc








Fig. 6
xgenus ac/dc

17

5.3 ASSEMBLING THE BRACKET/HORIZONTAL ARM

PLEASE NOTE
The brackets come in the following lengths: 41cm (16.2) - 82.5cm (32.5) - 110cm (43.5)
The 82.5 cm and 110 cm brackets are provided with a stop key  (Fig. 7A and 7B) to prevent the
electric cable from twisting.
PLEASE NOTE
Generally, the stop key is installed so that the equipment position at rest is on the right side of a
possible viewer when standing in front of the wall plate (Fig. 7A).
Should the position at rest be on the left side, the stop key must be rotated by 180 (Fig. 7B).

Fig. 7A

Fig. 7B

WALL

18

xgenus ac/dc

After fixing the plate to the wall, assemble the bracket according to the following procedure (Fig. 8):
1. Take the bracket out of the packaging (refer to Fig. 5)
2. Insert the bracket pin

 into the wall plate  (upwards)

3. Insert the supporting rest

Fig. 8
PLEASE NOTE
Prevent any foreign material (drywall, dust, cement, etc. ) from settling on the pin seat.
The pin must slide freely in its seat. If required, thoroughly clean and lubricate with grease
Molikote D".
PLEASE NOTE
Carefully check the bracket is parallel to the floor by using a water level.

xgenus ac/dc

19

5.4 ASSEMBLING THE POSITIONER ARM

ASSEMBLY INSTRUCTIONS (Fig. 9)


1. Take out the positioner arm from the packaging (refer to Fig. 5)
2. Remove the arm and cap
3. Slide the arm guard slat
4. Insert the arm cable

 by unscrewing the fixing screw 

 through the washer  and then the positioner pin 









Fig. 9

20

5.

If required, clean the pin and the washer and lubricate with grease Molikote D".

6.

Insert the electric cable into the bracket housing

7.

Assemble the guard slat

8.

Insert the cable into the arm and push it until it reaches the pin outlet near the supply
terminal board

xgenus ac/dc

5.5 CONNECTION TO THE TERMINAL BOARD

CAUTION
For electrical safety, it is essential that the ground conductors be correctly connected.

WARNING
The electric cable length is suitable for 82.5cm long arms.
DOT NOT tamper with the terminals arrangement when 41cm long arms are used.
The excess cable must be housed in the housing provided.
The 110cm long arms are equipped with a special extension cable.

OPERATING INSTRUCTIONS (Fig. 10)


1. Remove the terminal board cover

 by removing the screw 




Fig. 10
xgenus ac/dc

21

2. Proceed to the electric connection as shown in Fig. 11


PLEASE NOTE
The xgenus dc unit, differently from xgenus ac, is equipped with 2 communication cables.

Fig. 11
TERMINAL BOARD CONNECTION DIAGRAM









BROWN

L (line)

YELLOW GREEN

T (ground)

BLUE

N (neutral)

BLACK

COMMUNICATION

RED

COMMUNICATION

YELLOW GREEN

T (ground)

YELLOW GREEN

T (ground)

3. Connect the positioner arm shielded cable to the grounding potential


4. Connect the wall plate to the grounding potential

5. Clamp the cables with the stress relief clamps provided


6. Reassemble the terminal board insulating cover.

xgenus dc only

22

xgenus ac/dc

5.6 ASSEMBLING THE RADIOGRAPHIC UNIT


ASSEMBLY INSTRUCTIONS
1. Take out the radiographic unit from the packaging (refer to Fig. 5)
2. Check that all the rating data match the power supply voltage
3. Remove both guards from the positioner arm by loosening the relevant screws (Fig. 12)

Fig. 12
4. Using a screw driver tip, separate the front coupling device (Fig. 13A)
5. Remove both guards (Fig. 13B)

Fig. 13A

xgenus ac/dc

Fig. 13B

23

6. Insert the arm unit pin into the x-ray head (Fig. 14A) and insert the supporting retainer C clip
(Fig. 14B)

Fig. 14A

7.

Fig. 14B

Check that during insertion, the pin of the rotation prevention device correctly fits the seat located
on the x-ray head (Fig. 15)

Fig. 15

24

xgenus ac/dc

8. Couple the arm and x-ray head connectors and make sure that they are seated (Fig. 16)

Fig. 16

xgenus ac/dc

25

5.7 BALANCING THE POSITIONING ARM

CAUTION
The arm must be adjusted only when the radiographic unit is fully assembled and the xray head is mounted.

WARNING
To prevent damage to the internal mechanism while performing adjustments and the balancing
tests, the adjustment wrench/key must not be left in place.

WARNING
The wrench/key provided must remain with the unit for future adjustments.
PLEASE NOTE
To reach the adjustment screw X the arm A must be put in vertical position.
To reach the adjustment screw Y the arm B must be put in horizontal position.
The adjustment key provided can be inserted only under the above conditions (Fig. 17)

X
arm B

Fig. 17

arm A

Fig. 17
26

xgenus ac/dc

Procedure to be followed to correctly balance the positioner arm (refer to Fig. 17):
1.

BALANCING THE ARM A


PLEASE NOTE
The positioner type arm is supplied with arm A already factory adjusted.
The Arm B is supplied unloaded (no tension) for safety reasons.

2.

BALANCING THE ARM B


For correct balancing, proceed as follows:

3.

a.

ARM A VERTICAL

b.

ARM B HORIZONTAL

c.

INSERT THE ADJUSTMENT WRENCH/KEY IN Y

d.

LOAD THE SPRING BY APPLYING 22 TURNS WITH THE WRENCH IN A CLOCKWISE


DIRECTION.

e.

WITHDRAW THE WRENCH/KEY

CHECKING THE BALANCE


To check the balance proceed as follows:
BRING THE ARM B TO SEVERAL DIFFERENT POSITIONS
if the head and arm do not hold the position you have selected:
a.

BRING THE ARM B BACK TO THE HORIZONTAL POSITION

b.

INSERT THE ADJUSTMENTWRENCH IN Y

c.

ROTATE THE ADJUSTMENT WRENCH BY HALF TURN:


CLOCKWISE IF IT TENDS TO DRIFT DOWN
COUNTERCLOCKWISE IF IT DRIFTS UP

d.

WITHDRAW THE WRENCH

PLEASE NOTE
Repeat the test and adjustment until the arm A is steady and stable in all positions,
even when the arm B is completely extended.
4.

READJUSTMENT OF ARM A
If the head and arm do not hold the position you have selected:
a.

BRING THE ARM A TO THE VERTICAL POSITION AGAIN

b.

INSERT THE ADJUSTMENT WRENCH IN X

c.

TURN THE ADJUSTMENT WRENCH BY HALF TURN:


CLOCKWISE IF IT TENDS TO COME DOWN
COUNTERCLOCKWISE IF IT DRIFTS UP

d.

WITHDRAW THE WRENCH

xgenus ac/dc

27

PLEASE NOTE
Repeat the test and adjustment until the arm A is steady and stable in all positions,
even when the arm B is completely extended.
5. UPON COMPLETION OF BALANCING:
a. Insert the movable finish cover (A) between the arm guard and metal frame (Fig. 18)

Fig. 18
b.

Insert the pins  of one guard  into the relevant seats; then position it and check that the
finish piece  is coupled to the guards (Fig. 19)







28

Fig.19

xgenus ac/dc

5.8 ASSEMBLING xgenus ac/dc X-RAY CONTROL

CAUTION
Check that the cable runs are arranged in the x-ray control installation wall; check the
compliance
of
the
power
supply
with
the
installation
specifications
(refer to 4 INSTALLATION SPECIFICATIONS).

CAUTION
Check that the rating data match the power supply voltage.
ASSEMBLY INSTRUCTIONS (Fig. 20)
1.

Take the x-ray control out of the packaging (refer to Fig. 5)

2.

Take out the drilling template

3.

Confirm the vertical and horizontal alignment. Mark the mounting holes on the wall using
the drilling template.

4.

Drill using a 3mm tip, then drill again with a 6mm tip to prevent any flaking of the white
coat

5.

Insert the three screw anchors provided into the holes

6.

Open the x-ray control by unscrewing and removing the three screws




Fig. 20

xgenus ac/dc

29

7.

Remove the 26-pole connector from its seat to release both control guards (Fig. 21)

CONNECTOR 26 pole

Fig. 21
8.

Insert the electric feed cables into the hole (Fig. 22)

POWER SUPPLY CABLE


INLET HOLE

INLET HOLE
FOR THE RADIOGRAPHIC UNIT(S)
AND WARNING LAMP(S) CABLE

Fig. 22
9.

Insert the connection cables coming from the radiographic units into the rectangular slot

10.

Insert the cables of the external signal lamps (IF INSTALLED) (Fig. 22)

11.

Insert the cables of the remote control button (IF INSTALLED) (Fig. 22)

12.

Attach the x-ray control base to the wall, matching the three anchor screws with the holes

13.

Tighten the screws (refer to Fig. 20


the screw anchors

) with the relevant washers (refer to Fig. 20 ) into

PLEASE NOTE
If the wall is uneven, use appropriate shims to prevent the x-ray control from warping.
PLEASE NOTE
Keep the x-ray control clean. Protect the board from dust, dirt or wire particles.

30

xgenus ac/dc

5.9 X-RAY CONTROL ELECTRICAL CONNECTION

CAUTION
Before proceeding to connections, the power supply must be turned off.

CAUTION
If the equipment is mounted on metal wall or studs, these must be connected to the
grounding circuit of the system.

WARNING
While carrying out electrical connection procedures, always respect the polarity PHASE NEUTRAL.

WARNING
While stripping the cable, pay attention to the small copper wires that may fall on the printed circuit
board and cause short circuits or malfunctions.
OPERATING INSTRUCTIONS (Fig. 23) (refer to 19 ATTACHMENTS)
1.

Connect the power supply cable to the terminal board

2.

Insert the three mains cables into the rack

3.

Fix them with the cable clamp

4.

Connect the cables coming from the operatory unit# 1 to the terminals XRAY1

4a.

Connect the communication cables from the tube head 1 to the terminals C11 and C12

5.

Connect the YELLOW-GREEN grounding cable to the metal "Grounding plate 

6.

Connect the cables coming from the operatory unit# 2 to the terminals XRAY2

6a.

Connect the communication cables from the tube head 2 to the terminals C21 and C22

7.

Connect the YELLOW-GREEN grounding cable to the metal Grounding plate 

8.

Fasten the cables in the cable clamp

9.

Connect the remote control button (if installed) (refer to 19 ATTACHMENTS)

10.

Check the configuration on the dip-switches (refer to section 7 SYSTEM CONFIGURATION)

11.

Reconnect the 26-pole connector

12.

Close the x-ray control with the three screws

13.

Mount the plate guard (refer to Fig. 6

14.

Reconnect the power supply

)

xgenus dc only

xgenus ac/dc

31




3A-3B

4A-4B

Fig. 23
The electronic PC card terminal boards include the follow:

1
2
3A
3B
4A
4B
5
6
7

KEY1
KEY2
S12/S11
S22/S21
C12/C11
C22/C21
XRAY1
XRAY2
LINE

RADIOGRAPHIC
RADIOGRAPHIC
RADIOGRAPHIC
RADIOGRAPHIC

UNIT N
UNIT N
UNIT N
UNIT N

1
2
1
2

CONTROL BUTTON WITH LED


CONTROL BUTTON (OPTIONAL) WITH LED
SIGNAL LAMP OUTSIDE THE OPERATORY
SIGNAL LAMP OUTSIDE THE OPERATORY

RS232 COMMUNICATION FOR DC TUBE HEAD 1


RS232 COMMUNICATION FOR DC TUBE HEAD 2
RADIOGRAPHIC UNIT N 1 POWER SUPPLY
RADIOGRAPHIC UNIT N 2 POWER SUPPLY
X-RAY CONTROL POWER SUPPLY

WARNING
Check the value of the fuses F1-F2 (see schemes on 19) which must be in compliance with the type
of tube head connected to the XRAY1 and XRAY2 terminal boards:
FUSE
F1 (XRAY1)
F2 (XRAY2)

TUBE HEAD TYPE


xgenus ac
xgenus dc
8AF - 250V
12.5A - 250V
8AF - 250V
12.5A - 250V

xgenus dc only

32

xgenus ac/dc

6. X-RAY CONTROL PANEL

The control panel is the user's interface, where the operator can:
with the keys





select the radiographic unit operating parameters


change the exposure times displayed
store any new customized exposure times

with the light indicators green Led





xgenus ac/dc

check the x-ray control operating conditions


check the active function key
check the selected radiographic unit parameter

33

control panel
MAIN ON/OFF SWITCH

LED INDICATOR
DISPLAY
X-RAY EXPOSURE
BUTTON
RADIOGRAPHIC
DISTANCE INDICATOR
(LONG CONE
RECOMMENDEDMANDATORY OF DC
OPERATION)
KEY TO DECREASE
EXPOSURE TIME

KEY TO INCREASE
EXPOSURE TIME

TUBEHEAD TYPE
INDICATOR

TUBEHEAD
SELECTION

SELECTION OF PATIENTS
SIZE

RADIOGRAPHIC
KV/VOLTAGE SELECTOR
MEMORY

MAXILLARY
UPPER TEETH

RADIOGRAPHIC CURRENT
SELECTOR

MANDIBULARY
LOWER TEETH

OCCLUSAL EXAM

X-RAY EMISSION
SIGNAL
BITE-WING EXAM
DIGITAL RADIOGRAPHIC
TECHNIQUE (HOLD FOR 3
SEC.-ALWAYS USE LONG
CONE TECHNIQUE)

DUTY CICLE PAUSE


INDICATOR
MALFUNCTIONING
INDICATOR

FILM TYPE SELECTION


KEY SWITCH LOCK

34

xgenus ac/dc

7. SYSTEM CONFIGURATION
The xgenus ac/dc radiographic systems is factory configured for the following operative standard
mode configuration
on the control panel by pressing the RX button
Installation of 2 xgenus tube heads

A radiographic distance
SSD = 31cm
with long 12 cone

A radiographic distance
SSD = 20cm
with short 8 cone

Led 1 lights up,


by pressing again the button, Led 2 lights up

IF THE LONG 12 (31cm) CONE IS USED

on the control panel Led 12

is lit

IF THE SHORT 8 (20cm) CONE IS USED




with film type D


CONTROL BUTTON
to perform the exposure

on the control panel Led 8

is lit

on the control panel Led D

is lit

The control includes a control button with long


flexible coil cord.

The installed radiographic units include the following features:


tube head type

Led AC lit: xgenus ac


Led DC lit: xgenus dc

radiographic voltage

the Led 70 kV or 60 kV
on the control panel must be lit

radiographic current

the Led 4mA or 8mA


on the control panel must be lit
(ONLY AVAILABLE IN DC MODE)

) on the

The above configuration depends on the position of n 8 mini-switches (dip-switch


x-ray control board.
DIP
PARAMETER
SWITCH
1
RADIOGRAPHIC UNIT N 1
2
TYPE OF RADIOGRAPHIC UNIT N 1
3
RADIOGRAPHIC UNIT N 2
4
TYPE OF RADIOGRAPHIC UNIT N 2
5
CONTROL BUTTON N 2

CONE

7
8

NOT AVAILABLE
NOT AVAILABLE

ON

OFF

INSTALLED
NOT INSTALLED
DC
AC
INSTALLED
NOT INSTALLED
DC
AC
INSTALLED
NOT INSTALLED
LONG (12)
SHORT (8)
INSTALLED
INSTALLED
NOT AVAILABLE
NOT AVAILABLE

xgenus dc only

xgenus ac/dc

35

IF THE LONG 12 (31cm) CONE IS USED

ON

2
3

OFF
radiographic unit n 1
AC (=off) / DC (=on)

radiographic unit n 2

4
5
6

AC (=off) / DC (=on)

7
8

control button n 2
cone 12 (31cm)
not available
not available

IF THE SHORT 8 (20cm) CONE IS USED

ON
1

OFF

2
3

4
5
6
7
8

AC (=off) / DC (=on)

control button n 2
cone 8 (20cm)
not available
not available

radiographic unit n 1
AC (=off) / DC (=on)
radiographic unit n 2

xgenus dc only

36

xgenus ac/dc

The configuration may be changed if:

POSSIBLE MODIFICATIONS

HOW TO CARRY OUT THE MODIFICATIONS

type E and F films are used


a digital system is used

radiographic current (4mA /8mA)

radiographic voltage (60kV /70kV)


the short 8 (20cm) cone is used
the long 12 (31cm) cone is used
one single radiographic unit is used
n 2 CONTROL BUTTONS are used




By pressing the keys of the control panel


refer to USERS MANUAL
sect. 6 USE INSTRUCTIONS

by changing the dip-switch position

THIS OPERATION MUST BE CARRIED OUT BY THE AUTHORIZED


SERVICE PERSONAL ONLY

PLEASE NOTE
xgenus dc tube head allows the following combinations for the radiographic voltage and current:
60kV/4mA 60kV/8mA 70kV/4mA 70kV/8mA
xgenus ac tube head allows only:

70kV/8mA

xgenus dc only

xgenus ac/dc

37

8. CHANGING THE CONFIGURATION

CAUTION
This operation must be carried out by a qualified installer/technician.

WARNING
To install the changes you have made, turn off the x-ray control and then turn it on again.
To change the configuration, the dip-switch positions in the x-ray control must be changed:
DIP
PARAMETER
SWITCH
1
RADIOGRAPHIC UNIT N 1
2
TYPE OF RADIOGRAPHIC UNIT N 1
3
RADIOGRAPHIC UNIT N 2
4
TYPE OF RADIOGRAPHIC UNIT N 2
5
CONTROL BUTTON N 2

1.

CONE

7
8

NOT AVAILABLE
NOT AVAILABLE

ON

OFF

INSTALLED
NOT INSTALLED
DC
AC
INSTALLED
NOT INSTALLED
DC
AC
INSTALLED
NOT INSTALLED
LONG (12)
SHORT (8)
INSTALLED
INSTALLED
NOT AVAILABLE
NOT AVAILABLE

CHANGING THE NUMBER OF RADIOGRAPHIC UNITS INSTALLED


To change the number of radiographic units installed, move the dip-switch n 1 or n 3
a. If the radiographic unit is connected to the RX1 terminal board, put the dip-switch n 1
to the ON position, otherwise to the OFF position.
b. If the radiographic unit is connected to the RX2 terminal board, put the dip-switch n 3
to the ON position, otherwise to the OFF position.

2.

CHANGE THE TYPE OF TUBE HEAD INSTALLED


a. if the tube head connected to the XRAY1 is ac type, move the dip-switch n 2 to the

OFF position; otherwise to the ON position.


b. if the tube head connected to the XRAY2 is ac type, move the dip-switch n 4 to the

OFF position; otherwise to the ON position.

3.

CHANGING THE CONE


To replace the 12 cone with the 8 cone, move the dip-switch n 6 to the ON position.
Check that the led 8 lights up on the control panel.
PLEASE NOTE
After modification, the set exposure times are automatically changed.

xgenus dc only

38

xgenus ac/dc

To replace the 8 cone with the 12 cone, move the dip-switch n 6 to the ON position.
Check that the led 12 lights up on the control panel.
PLEASE NOTE
After modification, the set exposure times are automatically changed.
4.

CHANGING THE AMOUNT OF THE REMOTE BUTTONS


To change the number of control buttons, move the dip-switch n 5 to the ON position.
PLEASE NOTE
After modification, each radiographic unit is controlled by the relevant button.
PLEASE NOTE
Its possible to connect to the timer:
n 2 xgenus ac
n 2 xgenus dc
n 1 xgenus ac + n1 xgenus dc

xgenus ac/dc

39

9. START UP

CAUTION
For further information, see the USER'S MANUAL.
TURN ON THE X-RAY CONTROL to provide power to the system
a. Turn THE KEY SWITCH to position I (ON).

b. Set the MAIN SWITCH located on the upper part of the control to the
position I (ON).





The green light turns on indicating that the system is powered


The LEDs of the set radiographic parameters automatically light up
The exposure time is shown on the display

c. THE RADIOGRAPHIC SYSTEM IS NOW READY FOR USE

ATTENTION
If an error is detected when the system is turned on, the anomaly is indicated as
follows:

an intermittent beep sounds off

The MALFUNCTIONING INDICATOR Led

turns on intermittently

The error code (E .) appears on the display


(refer to 14 ERROR MESSAGES)

All CONTROL PANEL functions are inhibited

In this case, turn off the timer and then turn it back on again.
If the error should repeat itself, call for Technical Assistance.
PLEASE NOTE
The exposure time and radiographic parameters which appear on the display are the last that
were set before the x-ray control was turned off.
PLEASE NOTE
If installed outside the operatory, the RX signaling lamp corresponding to the selected tube head
turns on.
PLEASE NOTE
If the x-ray control remains unused and inactive for a few minutes, it switches to the stand-by
mode. Press any key of the CONTROL PANEL to restore it to the operative mode.
The previous settings will be retained.
40

xgenus ac/dc

10. CHECKING THE INSTALLATION

CAUTION
When assembly and all connections are completed, the installer must check the electrical
safety and function of the system for compliance.
OPERATING INSTRUCTIONS:
1.

Checking the configuration


Check on the control panel that all LEDs corresponding to the required configuration are lit;
otherwise, change them.
(Refer to section. 7 SYSTEM CONFIGURATION and section.8 CHANGING THE
CONFIGURATION.)

2.

3.

Checking the x-ray control operation


a.

Verify the correct operation of the control panel by selecting different exposure times

b.

Verify the time on the display

c.

When changing the selected radiographic unit, check that the corresponding external
RX signal lamps (if installed) turn on.

Checking the exposure: CAUTION! Radiation will be emitted.


a.

Set an exposure time of 1 sec

b.

Lift the "CONTROL BUTTON" on the x-ray control

c.

Using the extensible cable of the CONTROL BUTTON, keep a safe distance (length
of the cable) or at least 2 meters (7 feet) from the radiographic unit, to check the
radiographic exposure.

d.

Press the X-RAY KEY

and keep it pressed until the acoustic signal (beep)

stops and the yellow X-RAY OUTPUT SIGNAL

Led turns off.

PLEASE NOTE
If the EXPOSURE BUTTON is released early, the exposure is immediately
interrupted and the E12 error message appears on the display.

e.

f.

xgenus ac/dc

At the end of the exposure, verify the operation of the Led PAUSE INDICATOR

Verify on the display the indication of the actual exposure time


(Refer to section 11.4 EXPOSURE TIME (sec))

41

4.

Checking the operation of the radiographic unit:


emitted

CAUTION! Radiation will be

Carry out several exposures on the installed radiographic units installed and check for the
following:

5.

a.

There are no errors.

b.

The LED (AC or DC) of the selected radiographic unit is lit.

c.

The "CONTROL BUTTON" LED is lit over the duration of the acoustic signal.

Checking the power drawn by the radiographic system: CAUTION! Radiation will
be emitted.
To check the power drawn by the radiographic system, a meter must be used, with the
Ammeter in AC
a.

Connect the instrument to the power supply line


(refer to section 5.9 X-RAY CONTROL ELECTRICAL CONNECTION)

b.

Set an exposure time of approx. 3 sec. on the x-ray control.

b. If using an xgenus dc, set 70kV-8mA-3.2s on the x-ray control.

c.

Carry out an exposure and read the current value on the instrument
PLEASE NOTE
The radiographic system complies with the requirements when:
the current drawn is 8A for line voltage of 115V

the current drawn is 12.5A for line voltage of 115V


Otherwise, check the electrical system or call for Technical Assistance.
6.

Checking the electrical system


To check the electrical system, a meter must be used, with the Voltmeter in AC mode.
a.

Connect the instrument to the terminals L and N on the x-ray control

b.

Measure the line voltage (load voltage V0)

c.

Connect the instrument to the terminals L and N of the wall plate terminal board
(refer to section 5.5 fig.11 - CONNECTION TO THE TERMINAL BOARD)

d. Set an exposure time of approx. 3 sec on the x-ray control

d. If using an xgenus dc, set 70kV-8mA-3.2s on the x-ray control.


e. Release the exposure
f.

Measure the line voltage during exposure (load voltage V1)


PLEASE NOTE
The electric system complies with the requirements when the result of the formula is
as follows:

(V0 V1 ) / V0 0.03 or 0.012


Otherwise, the electric system must be adjusted
(refer to section 4: INSTALLATION SPECIFICATIONS)

xgenus dc only

42

xgenus ac/dc

11. CHECKING THE EXPOSURE FACTORS

1. Radiographic voltage (kVp)


The radiographic high voltage is measured using a calibrated "non invasive" instrument with an initial

delay set of 50ms . The kVp value is factory checked with a PMX-I D.

SET
TECHNICAL
FACTORS

xgenus ac
xgenus dc
Line Voltage
Vn -5% / +10%
Vn 15%
Max Voltage drop
1.2%
3%
Nominal High Voltage
70 kV
60-70 kV
Nominal current
8mA
4mA 8 mA
Set exposure time
3.2 s
3.2 s
The radiographic VOLTAGE is: 70kVp 15% for xgenus ac
60kVp 70kVp 10% for xgenus dc

2. Radiographic current (mA)


The radiographic current is measured by connecting a milliampmeter. The mA value is factory
checked with a digital multimeter.
xgenus dc
xgenus ac
Line Voltage
Vn -5% / +10%
Vn 15%
Max Voltage drop
1.2%
3%
Nominal
High
Voltage
70
kV
60-70
kV
SET
TECHNICAL
Nominal current
8mA
4mA 8 mA
FACTORS
Set exposure time
3.2 s
3.2 s

The radiographic CURRENT is: 8mA 15% for xgenus ac


4mA 8mA 10% for xgenus dc
3. Dose (mGy)
The dose in air is measured with a "non invasive" instrument, by positioning the detector at a spotskin distance = 310mm (12). The mGy are factory checked with the instrument SOLIDOSE.

SET
TECHNICAL
FACTORS

Line Voltage
Max Voltage drop
Nominal High Voltage
Nominal current
Set exposure time

xgenus ac
Vn -5% / +10%
1.2%
70 kV
8mA
1s

xgenus dc
Vn 15%
3%
60-70 kV
4mA 8 mA
1s

 3.3 mGy/s 30% (with 12 cone)


6.5 mGy/s 30% (with 8 cone)

Dose in air  (60kVp 4mA)=2.2mGy/s30%


(70kVp 4mA)=3mGy/s30%
(60kVp 8mA)=4.5mGy/s30%
(70kVp 8mA)=5.5mGy/s30%
Dose in air

(with 12 cone)
(with 12 cone)
(with 12 cone)
(with 12 cone)

xgenus ac only
xgenus ac only

xgenus dc only equipped with a 12 cone. See 13 of the USERS MANUAL for the dose in air with a 8 cone.

xgenus ac/dc

43

4. Exposure time (s)


The exposure time is measured with a non invasive instrument.

PLEASE NOTE
The xgenus x-ray control features the "self-compensating" technology.
The automatic correction or compensation of the exposure time offsets the effects of the
line voltage fluctuations on x-ray film quality.
Radiographic over/under exposure is eliminated.
The automatic correction system works as follows:

A voltmeter located inside the timer constantly reads the mains voltage
while the operator selects the required exposure time.

The last voltage measured (before the exposure) is sent to the


microprocessor. The selected exposure time is corrected to voltage by
applying a special algorithm.

At the end of the exposure, during the pause time, the time-corrected value
is shown on the display of the x-ray control.
THE PAUSE TIME IS APPROX. 32 TIMES THE SET TIME.

The correct time is the actual duration of the exposure


The accuracy of the actual time is factory checked with instrument PMX-I.

SET
TECHNICAL
FACTORS

xgenus dc
xgenus ac
Line Voltage
Vn -5% / +10%
Vn 15%
Max Voltage drop
1.2%
3%
Nominal High Voltage
70 kV
60-70 kV
Nominal current
8mA
4mA 8 mA
Set exposure time
0.2 s
1.6 s

The EXPOSURE TIME measured is: 0.2s 10% for xgenus ac


1.6s 10% for xgenus dc

5. Checking the exposure time

(continued on p. 45: Operating instructions)

WARNING
While performing periodic quality checks, some differences between the set exposure time and
the measured exposure time may be detected.
The differences are due to the automatic time compensation that intervenes when the power
voltage differs from the nominal one.
The x-ray control is provided with two types of accuracy measurements for the exposure time:
A. Set exposure (is the exposure time set by the operator on the x-ray control).

The max error is +100% / - 100% when the voltage fluctuation is -5% / +10%.
B. After exposure: its the exposure duration (automatically adjusted). Its calculated by the
microprocessor according to the line voltage, and its shown on the display.

xgenus ac only

44

xgenus ac/dc

OPERATING INSTRUCTIONS:

CAUTION
Radiations will be emitted.
To check the exposure times (sec) it is advisable to use calibrated non invasive instruments;
i.e.:
NERO by Victoreen
PMX by RTI
TRIAD QA by Keithley
a. Put the X-ray detector at the cone outlet; center the primary beam.
b. Turn on the x-ray control
c. Set one exposure time on the control.
d. Release the exposure.
e. At the end of the exposure, when the functional pause period starts, read the actual time of
exposure on the display.
f. To check the accuracy of the exposure time, use the chart below.

CONFORMITY CHECK CHART


At the end of the exposure
1

read the time value


on the display

= Actual exposure time

read the measure


on the instrument

= Emission time measure

subtract 1 from 2

divide 3 by 2

= Absolute error

multiply 4 by 100

= Percentage error (%)

The accuracy is checked when the


% of error is

10% o 20msec

by adopting the most favorable


condition.

PLEASE NOTE
The exposure time complies with the requirements when:
((Measured time Actual time) / Measured time)) x 100 10% or 20 ms
Otherwise, please contact Technical Support.

xgenus ac/dc

45

12. DIAGNOSTIC
The xgenus ac/dc radiographic systems allows the installer to set or display some functional
parameters of the x-ray control.
1.

To set the minimum exposure time:


a.
b.

Select the tube head


Turn off the x-ray control

c.
d.
e.

key pressed
Turn on the x-ray control, by keeping the
The message SEC is displayed for approx 1sec
The present value of the lower set limit is displayed

f.

To change the value, press the keys

g.
h.

key
To confirm, press the
To exit this mode, turn the x-ray control off and then on again

and

WARNING
Both in digital and in conventional mode, the minimum exposure time MUST NOT BE LESS than

0,08s
2.

To visualize the parameters, the installer must:

a.

Press simultaneously and keep pressed the following keys


(2) UPPER JAW MOLAR + (31) LOWER JAW MOLAR

b.

Press the key associated to the parameter one wishes to see


KEY

DISPLAYED PARAMETER

Example

U.M.

BITE-WING
ANTERIOR

RADIOGRAPHIC SYSTEM
VOLTAGE

110

Volt

BITE-WING
POSTERIOR

LINE VOLTAGE

105

Volt

UPPER
INCISOR

MAXIMUM LINE VOLTAGE

VALUE DETECTED

115

Volt

LOWER
INCISOR

MINIMUM LINE VOLTAGE

VALUE DETECTED

118

Volt

OCCLUSAL

SOFTWARE VERSION

1.5

----

xgenus ac only

46

xgenus ac/dc

13. TUBE HEAD CALIBRATION

CAUTION
During this operation radiations will be emitted.
It is mandatory to carry out all the safety measures relevant to radioprotection.

INSTRUCTIONS
1.

turn off the timer

2.

turn on the timer by keeping the key

3.

the message TUBE is displayed.

4.

select the

5.

take an exposure by pressing the EXPOSURE BUTTON key

pressed.

appropriate tube head to calibrate.

pressed until the acoustic signal (beeping) stops and the yellow
6.

and keep it
led turns off.

Once the exposure has been taken, if the display does not show errors, the calibration has been
successfully done.

xgenus dc only

xgenus ac/dc

47

14. ERROR MESSAGES


The following chart gives a list of error messages that may appear while the xgenus ac/dc
radiographic systems is working.
The chart also includes the reason for the error messages and what to do to solve them.
ERROR
MESSAGES

CAUSE
RX1 TUBEHEAD
IS NOT CONNECTED OR IS OUT OF ORDER
RX2 TUBEHEAD
IS NOT CONNECTED OR IS OUT OF ORDER

CALL TECHNICAL SUPPORT


CALL TECHNICAL SUPPORT

CORRUPTED EEPROM DATA

CALL TECHNICAL SUPPORT

EEPROM DATA NOT SAVED PROPERLY

CALL TECHNICAL SUPPORT

LINE VOLTAGE VALUE NOT


WITHIN RANGE
THE X-RAY KEY
ALWAYS SEEMS TO BE PRESSED

CALL TECHNICAL SUPPORT


MAKE SURE IT IS NOT JAMMED OR DEFECTIVE

ANOMALY IN THE CONTROL PANEL

CALL TECHNICAL SUPPORT

THE EXPOSURE
HAS BEEN PREMATURELY INTERRUPTED
ANOMALY IN THE TRIAC/RELAY

KEEP THE X-RAY KEY


PRESSED TILL THE END OF THE EXPOSURE
CALL TECHNICAL SUPPORT

ANOMALY IN THE ELECTRONIC CIRCUIT

CALL TECHNICAL SUPPORT

ANOMALY IN THE CONTROL CIRCUIT

CALL TECHNICAL SUPPORT

INCORRECT DIP-SWITCH CONFIGURATION

CALL TECHNICAL SUPPORT

THE CONTROL BUTTON DOES NOT CORRESPOND


TO THE SELECTED TUBEHEAD
THE TUBEHEAD
DOES NOT WORK PROPERLY
THE TUBEHEAD
IS NOT IN THE CORRECT MODE
THE TUBEHEAD
HAS NOT COMPLETED THE EXPOSURE
THE TUBEHEAD
IS NOT CALIBRATED

CALL TECHNICAL SUPPORT

EEPROM DATA NOT SAVED PROPERLY

CALL TECHNICAL SUPPORT

CORRUPTED EEPROM DATA

CALL TECHNICAL SUPPORT

OVERVOLTAGE ERROR

CALL TECHNICAL SUPPORT

ANODE VOLTAGE OUT OF TOLERANCE

CALL TECHNICAL SUPPORT

ANODE CURRENT OUT OF TOLERANCE

CALL TECHNICAL SUPPORT

CONTROL CONNECTOR

CALL TECHNICAL SUPPORT

PROBLEM IN THE REFERENCE VOLTAGE

CALL TECHNICAL SUPPORT

MAJOR ERROR

ALL FUNCTIONS ARE DISABLED


CALL TECHNICAL SUPPORT

SOLUTION

CALL TECHNICAL SUPPORT


CALL TECHNICAL SUPPORT
REPEAT THE EXPOSURE
CALL TECHNICAL SUPPORT
CALL TECHNICAL SUPPORT

xgenus dc only

48

xgenus ac/dc

15. REPLACEMENT OF FUSES


The x-ray control electronic equipment is protected by 4 fuses located on the electronic P.C.B.
To replace them, proceed as follows (Fig. 25):
1.

Turn off the power supply

2.

Remove the front cover from the x-ray control

3.

Identify the fuse to be replaced (Fig. 26)

4.

Remove the plastic protection (Fig. 27)

5.

Remove the fuse

6.

Replace it with one of same type

TYPE OF TUBE HEAD CONNECTED

xgenus ac

xgenus dc

POWER VOLTAGE
OF THE RADIOGRAPHIC SYSTEM

115V (-5% /+10%)

115V (15%)

Internal protection fuses


(5x20) quick

F1 (XRAY1)
F2 (XRAY2)

8AF 250V
8AF 250V

12.5A 250V
12.5A 250V

(see schemes on 19)

F3 F4

12.5A 250V

12.5A 250V

7.

Put the fuse protection back in place.

8.

Fig. 26
Replace the x-ray control cover.

9.

Restore the power supply.

 by unscrewing the fixing screws 

Fig. 27

n 2 spare fuses are supplied in a plastic bag inside the XGT control.

xgenus ac/dc

49

16. MAINTENANCE
To guarantee the safety of the radiographic system, the following checks must be carried out:

CAUTION
The owner is responsible for arranging the maintenance schedule.
The radiographic system must undergo maintenance to check correct operation every
12 months.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.

20.
21.

22.
23.
24.
25.
26.
27.
28.
29.
30.
50

Turn off the power


Release the spring tension of the arm B of the positioner using the wrench provided.
(Please see page 27)
Remove the radiographic generator.
(Refer to section 5.6)
Remove the wall plate cover.
(Refer to section 5.2)
Remove the terminal board cover and disconnect the cable
Remove the bracket plug and withdraw the guard slat.
(Refer to section 5.4)
Remove the positioner and the relevant cable from the bracket
Remove the bracket from the wall plate
(Refer to section 5.3)
Check the vertical alignment of the wall plate adjust if required
Check the six screws on the wall plate tighten if required
Clean off the old lubricating grease from the bracket shaft - should the bracket shaft be
damaged, install a new bracket
Clean the old lubricating grease off the bracket bushing - should the bracket bushing be
damaged, install a new bracket
Grease the bracket shaft (use grease MOLIKOTE D)
Lubricate the bracket bushing with lubrication grease (use grease MOLIKOTE D)
Install the bracket in the wall plate
(Refer to section 5.3)
Check the arm cable should it be damaged, send the arm to the manufacturer for repairs
Check the arm covers
Replace if the guards are damaged
Clean the old grease off the shaft - should the shaft be damaged, send the arm to the
manufacturer/dealer for repairs.
(Refer to section 5.4)
Lubricate the arm shaft with lubricating grease (use grease MOLIKOTE D) and reposition it in
the bracket
Put again the arm cable in the bracket and the wall plate. Connect it to the terminal board
and re-attach the terminal board cover
(Refer to section 5.5)
Position the guard slat in the bracket
Position the bracket plug
Position the plate guard
Check the electric contact of the radiographic unit. If damaged, send the radiographic unit to
the manufacturer/dealer for repairs
Clean the old grease from the radiographic unit assembly shaft
Grease the assembly shaft of the radiographic unit with a thin layer of lubricating grease
(use grease MOLIKOTE D)
Position the radiographic unit again
Load the spring of the positioner arm B using the wrench provided
(Refer to 5.7)
Turn on the power and check the correct operation of the radiographic system.
xgenus ac/dc

17. CLEANING THE OUTER SURFACES


Use a soft cloth dampened with water and mild soap to clean the outer surfaces of the x-ray.
DO NOT permit water or any liquid to penetrate the machine or severe electrical component damage
may occur. Damage of this nature is not covered under warranty.
For protection against cross-infection, it is recommended that barrier protection be used and replaced
for each patient.
Germicides must be selected carefully, realizing that many are not suitable for use with such plastics
that are used in manufacturing. Damage of this nature is not covered under warranty.

18. IF A REPAIR SHOULD BECOME NECESSARY

In case of a malfunction, contact your local Technical Service Representative.


They will call TG Group for an RA Number (Return Authorization). This will authorize one the return of
the defective part to TG Group for replacement or repair.

USING THE ORIGINAL PACKAGING


Send all materials to:
TG Group, Inc.
3505 Laird Road, Unit # 7
Mississauga, Ontario L5L 5Y7
Canada
Tel. 1-877-557-4888
Fax 1-905-832-0732

Manufacturer:
de Gtzen S.r.l.
Via Roma, 45
21057 OLGIATE OLONA VA - ITALY
Tel. +39 0331 376760 r.a.
Fax +39 0331 376763
E-mail: degotzen@degotzen.com

xgenus ac/dc

51

19. ATTACHMENTS
The manufacturer undertakes to supply, upon request, drawings, circuit diagrams, component parts
lists, instructions or other information needed by qualified technical personnel to perform repairs on
those parts of the xgenus ac/dc radiographic systems which may be repaired.

52

xgenus ac/dc

xgenus ac/dc

53

REF
1
2
3
4
5
A
B
C

54

SPARE PARTS TUBEHEAD xgenus ac


AC XG (1) FORK PLASTIC COVER
AC/DC XG (2) RING NUT SET FOR THE
FORK
AC/DC XG (3) FORK
AC/DC XG (4) ROTATION LIMITING
DEVICE
AC XG (5) CABLE + 3 POLES CONNECTOR
LONG CONE 31cm (12) 60
SHORT CONE 20cm (8) 60
RECTANGULAR CONE 31cm (12) 44x35

CODE
29700983
29700980

Q.TY PCS
7
5

29700777
29700750

3
1

39700302
29700397
29700396
29700615

1
1
1
1

xgenus ac/dc

REF
1
2
3
4
5
6
7
8

xgenus ac/dc

SPARE PARTS TUBEHEAD xgenus dc


DC XG (1) FORK PLASTIC COVER
AC/DC XG(2) RING NUT SET FOR THE
FORK
AC/DC XG (3) FORK
AC/DC XG (4) ROTATION LIMITING
DEVICE
DC XG (5) WASHER
DC XG (6) AIR CARD
DC XG (7) AIR CARD COVER
DC XG (8) HEAD CABLE

CODE
29700979
29700980

Q.TY PCS
8
5

29700777
29700750

3
1

27500959
29700981
29700982
39700385

1
9
11
1

55

REF
1
2
3

56

SPARE PARTS xgenus ECB


(OPTIONAL)
XG ECB (1) SUP & INF SHELL
MAIN BOARD WITH SWITCH AND LED
XG ECB (3) KEYLOCK SWITCH

CODE

Q.TY PCS

29700978
39700303
35100298

4
1
1

xgenus ac/dc

REF
1
2
3
3
4
4

XG
XG
XG
XG
XG
XG

xgenus ac/dc

SPARE PARTS xgenus LIGHT


(OPTIONAL)
LIGHT (1) SUP & INF SHELL
LIGHT (2) LAMP HOLDER
LIGHT (3) LAMP 230V 15W
LIGHT (3) LAMP 115V 15W
LIGHT (4) 230V BOARD
LIGHT (4) 115V BOARD

CODE

Q.TY PCS

29700977
33400372
33300352
33300390
39200354
39200353

4
1
1
1
3
3

57

Ref
1
2
2
3
4
5
6
7
8
8
9
10
10
11
12
13
14
58

SPARE PARTS TIMER xgenus ac/dc


AC/DC XG TIMER (1) TEMPLATE
AC/DC XG TIMER (2) CONTROL BOARD
AC/DC XG TIMER (2) CONTROL BOARD US
AC/DC XG TIMER (3) XM COILED CABLE
AC/DC XG TIMER (4) XM DOUBLE COILED
CABLE
AC/DC XG TIMER (5) EXPOSURE BUTTON
HOUSING
AC/DC XG TIMER (6) EXPOSURE BUTTON
BOARD
AC/DC XG TIMER (7) LED BOARD
AC/DC XG TIMER (8) DEG KEYBOARD
AC/DC XG TIMER (8) SAT KEYBOARD
AC/DC XG TIMER (9) REAR COVER
AC/DC XG TIMER (10) DEG FRONT COVER
AC/DC XG TIMER (10) SAT FRONT COVER
AC/DC XG TIMER (11) KEYSWITCH LOCK
AC/DC XG TIMER (12) FILTER
GREEN BIPOLAR SWITCH
BIPOLAR SWITCH PROTECTION

CODE
24200883
39200306
39200307
32000325
32000359

Q.TY PCS
1
7
7
1
1

29700974

39700303

39200313
39200308
39200309
29700975
29700304
29700976
39700305
36200382
35000044
33500228

7
1
1
1
4
4
2
3
1
1
xgenus ac/dc

REF
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15

SPARE PARTS POSITINER ARM


xgenus ac/dc
AC XG B (1) CABLE + 3 POLES CONNECTOR
DC XG B (2) CABLE + 5 POLES CONNECTOR
AC/DC XG B (3) HEAD SIDE SHELLS
POSITIONER ARM CONNECTING ROD ANT
POSITIONER ARM TIE ROD HOLDER
AC/DC XG B (6) CAPS (set)
AC/DC XG B (7) EXT COVER
AC/DC XG B (8) INT COVER
AC/DC XG B (9) CENTRAL SHELLS
POSITIONER ARM LOADING WRENCH
AC/DC XG B (11) INT COVER
AC/DC B (12) EXT COVER
POSITIONER ARM CONNECTING ROD POST
AC/DC XG B SAFETY KEY
AC/DC B (15) BRACKET SIDE SHELLS

xgenus ac/dc

CODE

QTY PCS

39700384
39700381
29700990
29700460
29700458
29700991
27500757
27500758
29700993
22000543
27500758
27500757
29700992
22000176
29700994

1
1
9
4
6
7
1
1
8
1
1
1
7
1
8

59

REF
1
1
2
3
4

60

SPARE PARTS WALL PLATE 350


WALL BOX COVER (1) DEG
AC/ XG WALL BOX (1) COVER SAT
AC/DC XG B & WALL PLATE SAFETY KEY
WALL PLATE (3) TEMPLATE
WALL PLATE (4) INSERT + CAPS

CODE
29700985
29700986
22000176
24200984
29701036

Q.TY PCS
5
5
1
1
3

WALL PLATE 350 DEG


AC/ XG WALL PLATE 350 SAT

29700383
29700987

COMPLETO
COMPLETO

xgenus ac/dc

REF
1
1
2
3
4
5

SPARE PARTS WALL PLATE 350 TOP


WALL PLATE (1) COVER DEG
AC/ XG WALL PLATE (1) COVER SAT
AC/DC XG B & WALL PLATE SAFETY KEY
WALL PLATE (3) TEMPLATE TOP MOUNT
WALL PLATE (4) INSERT + CAPS
WALL PLATE (5) WASHER TOP MOUNT

CODE
29700985
29700986
22000176
24201002
29701036
22000597

WALL PLATE 350 TOP DEG


AC/ XG WALL PLATE 350 TOP SAT

29700988
29700989

xgenus ac/dc

Q.TY PCS
5
5
1
1
3
1
COMPLETO
COMPLETO

61

Ref

ARM SPARE PARTS


400
800
1100
1100 DC XG

CODE
29700380
29700378
29700379
29701019

Q.TY PCS
COMPLETO
COMPLETO
COMPLETO
CPMPLETO

Ref

ARM SPARE PARTS


CEILING BRACKET

CODE
29700382

Q.TY PCS
COMPLETO

Ref

ARM SPARE PARTS


UNIT BRACKET

CODE
29700381

Q.TY PCS
COMPLETO

CODE
29700620
29700621
29700622
29701035

Q.TY PCS
COMPLETO
COMPLETO
COMPLETO
COMPLETO

BRACKET
BRACKET
BRACKET
BRACKET

Ref
BRACKET
BRACKET
BRACKET
BRACKET

62

ARM SPARE PARTS


400 TOP MOUNT
800 TOP MOUNT
1100 TOP MOUNT
1100 TOP MOUNT DC XG

xgenus ac/dc

Dichiarazione

di Conformit

Declaration of Conformity Statement -

Dclaration
Konformitts-Erklrung - Dclaration de Conformidad de la

de Conformit

Impresa
Company
Societ
Firma
Empresa
dichiara che il dispositivo medico
declares that the medical device
declare que le dispositif medical
erklrt, da das medizinisches Gert
declara que el dispositivo medico

sistema radiografico per diagnosi intra-orale


intraoral X-ray system
appareil pour radioagnostic dentaire
Rntgendiagnose-Einheit
rayos-X infraorales

modello
type of machine
modle de machine
bauart der maschine
modelo de la mquina

xgenus

conforme ai requisiti delle direttive


is in compliance with the directives
est conforme aux directives
ist den Richtlinien gem
comple las directives
dispositivi medici
medical devices
dispositives medicaux
medizinische gerte
dispositivos medicos

93/42/CEE

organismo notificato
notified body
organisme notizie
verantwortliche gestaltung
organismo responsible

0434

certificato n
certificate No.
certificat n
zertifikat Nr.
certificato n

2002 OSL MDD 0044

compatibilit elettromagnetica
electromagnetic compatibility
compatibilit electromagntique
elektromagnetische verordnung
contabilidad electromagnetica

89/336/CEE

bassa tensione
low voltage
basse tension
niederspannung
baja tension

73/23/CEE

e successive modifiche
and subsequent amendments
et modifications suivantes
und folgenden nderungen
y modificaciones singuentes

Il dispositivo definito di classe IIb conforme ai requisiti dellallegato II della direttiva 93/42/CEE
This product, definited as a class IIb device, is conformed to the requirements of annex II of the council directive 93/42/ECC
Le produit ci-mentionn, defini comme dispositif de classe IIb, est conforme aux requises de lannex II de la directive 93/42/EEC
Dieses Gert, bestimmt als Klasse IIb, ist den richtilinien 93/42/EWG, annex II gem
El producto aqui mencionado esta definido como dispositivo de clase IIb y esta conformado a los requisitos de lannex II de la directiva 93/42/CEE

data di emissione
issuing date
date de mission
ausstelldatum
fecha de emisio

xgenus ac/dc

18.03.2002

63

Dichiarazione

di Conformit

Declaration of Conformity Statement -

Dclaration
Konformitts-Erklrung - Dclaration de Conformidad de la

de Conformit

Impresa
Company
Societ
Firma
Empresa
dichiara che il dispositivo medico
declares that the medical device
declare que le dispositif medical
erklrt, da das medizinisches Gert
declara que el dispositivo medico

sistema radiografico per diagnosi intra-orale


intraoral X-ray system
appareil pour radioagnostic dentaire
Rntgendiagnose-Einheit
rayos-X infraorales

modello
type of machine
modle de machine
bauart der maschine
modelo de la mquina

xgenus dc

conforme ai requisiti delle direttive


is in compliance with the directives
est conforme aux directives
ist den Richtlinien gem
comple las directives
dispositivi medici
medical devices
dispositives medicaux
medizinische gerte
dispositivos medicos

93/42/CEE

organismo notificato
notified body
organisme notizie
verantwortliche gestaltung
organismo responsible

0434

certificato n
certificate No.
certificat n
zertifikat Nr.
certificato n

2002 OSL MDD 0044

compatibilit elettromagnetica
electromagnetic compatibility
compatibilit electromagntique
elektromagnetische verordnung
contabilidad electromagnetica

89/336/CEE

bassa tensione
low voltage
basse tension
Niederspannung
baja tension

73/23/CEE

e successive modifiche
and subsequent amendments
et modifications suivantes
und folgenden nderungen
y modificaciones singuentes

Il dispositivo definito di classe IIb conforme ai requisiti dellallegato II della direttiva 93/42/CEE
This product, definited as a class IIb device, is conformed to the requirements of annex II of the council directive 93/42/ECC
Le produit ci-mentionn, defini comme dispositif de classe IIb, est conforme aux requises de lannex II de la directive 93/42/EEC
Dieses Gert, bestimmt als Klasse IIb, ist den richtilinien 93/42/EWG, annex II gem
El producto aqui mencionado esta definido como dispositivo de clase IIb y esta conformado a los requisitos de lannex II de la directiva 93/42/CEE

data di emissione
issuing date
date de mission
ausstelldatum
fecha de emisio

64

25.06.2002

xgenus ac/dc

NOTES