Professional Documents
Culture Documents
discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/261996563
CITATIONS
READS
166
5 authors, including:
Pasquale Lino Calvani
Argiris Pissiotis
Tufts University
7 PUBLICATIONS 24 CITATIONS
SEE PROFILE
SEE PROFILE
Hiroshi Hirayama
56 PUBLICATIONS 877 CITATIONS
SEE PROFILE
Clinical Implications
Al-Ghafli et al
141
September 2009
Edentulous patients with severely
resorbed mandibles may experience
problems with conventional dentures
because of impaired load-bearing capacity. These problems include pain
during mastication, as well as insufficient stability and retention of the
denture.1 Perhaps the most significant
biological condition associated with
loss of stability and retention in mandibular complete dentures is physiological alveolar ridge resorption,
which also results in diminished oral
tissue volume for denture support.2
It is not always possible to achieve
optimal results using conventional
complete denture treatment. Therefore, alternatives should be considered. When satisfactory denture support is present, denture adhesives
can improve the treatment outcome.
However, when problems arise from
inadequate supporting tissue volume
for the mandibular denture, denture
adhesives can prove inadequate.3 Alveolar ridge augmentation using a
variety of natural and synthetic materials has been suggested as an alternative to increase supporting tissue
volume.4 Similarly, alveoloplasty and
tissue extension procedures have been
used to expose additional intraoral
tissues. Muscles can also be repositioned to increase denture support. It
should be noted, however, that these
treatments have occasionally introduced significant complications and
morbidity.5
Studies have shown that unsatisfactory retention and stability can
be successfully managed by fabricating a fixed prosthesis supported by 5
or 6 endosseous implants.6-9 Another
treatment modality used to provide
predictable retention and stability for a severely resorbed mandible
is mandibular implant overdenture
treatment. This treatment has a reported implant survival rate of 94.5100%.10,11
Currently, the placement of 2
implants and the fabrication of an
implant-retained overdenture is considered by some to be the standard of
care.12 Implants can be used in con-
Al-Ghafli et al
142
Al-Ghafli et al
143
September 2009
Table I. Percentage of retention decline in each cycle for all groups (n=5)
Control
Percentage
(%)
Mean
Force
ORC* (N)
0D
5D
Mean
Mean
Force
Force
ORC* (N) ORC* (N)
10D
15D
20D
Mean
Force
ORC* (N)
Mean
Force
ORC* (N)
Mean
Force
ORC* (N)
100
81.31
81.75
91.94
104.72
84.86
78.04
75
217
60.98
484
61.31
514
68.96
243
78.54
371
63.65
180
58.53
50
598
40.66
1217
40.87
1220
45.97
725
52.36
914
42.43
581
39.02
25
2313
20.33
6497
20.40
4990
22.99
2949
26.18
3210
21.22
2234
19.51
experiment.
Under normal circumstances,
a patient places/removes an overdenture prosthesis 4 times each day
(in the morning (placement), after
breakfast, after lunch, after dinner
(removal and replacement), and before bedtime (removal)).31 Based on
this assumption, time was calculated
by the equation:
Days = Number of cycles / 4
The attachments included in this
study were tested for loss of retention after being subjected to 720
(6 months), 1440 (1 year), 2880 (2
years), 7200 (5 years), and 14,400
(10 years) cycles of overdenture removal.31,32 According to the studies of
Caldwell,22 Walmsley et al,23 and Petropoulos et al,25,26 it can be assumed
that an initial retentive force of 20 N
is sufficient for overdentures in the
edentulous mandible. Consequently,
retention loss (and estimated replacement time), was defined as the number of cycles required for the dislodgement force to drop below 20 N.
The initial retentive forces among
the groups were not identical. In 4
of the 6 groups included, it was observed that the initial retentive values
were in the range of 80 N. The mean
initial retentive values of the 2 remaining groups (5D and 10D) were 91.94
N and 104.72 N, respectively (Table
I). As a result, the loss of attachment
retention was difficult to compare.
Therefore, the dislodging retentive
Al-Ghafli et al
RESULTS
The results of the 1-way ANOVA
for the number of cycles required for
144
Table II. Mean (SD) cycles, per group, for retentive force to decline from initial value to 60 N (n=5)
Group
Overdenture Removal
Mean Cycles (SD)
Time of Mean
(Months/Days)
Group
Comparison*
Control
242.40 (37.51)
2M
0D
546.40 (47.50)
4M/17D
5D
753.40 (103.59)
6M/7D
10D
571.80 (102.56)
4M/24D
15D
422.20 (139.97)
3M/15D
20D
235.40 (54.79)
2M
*Means with same uppercase letter are not significantly different (P>.05), according to Tukey
HSD test.
Table III. Mean (SD) cycles, per group, for retentive force to decline from initial value to 40 N (n=5)
Group
Overdenture Removal
Mean Cycles (SD)
Time of Mean
(Years/Months/Days)
Group
Comparison*
Control
655.00 (109.75)
5M/14D
0D
1281.40 (174.29)
10M/20D
CD
5D
1590.60 (167.84)
1Y/1M/7D
10D
1187.40 (170.78)
9M/27D
15D
1030.40 (202.06)
8M/12D
BC
20D
796.00 (118.59)
6M/19D
AB
*Means with same uppercase letter are not significantly different (P>.05), according to Tukey
HSD test.
Table IV. Mean cycles, per group, for retentive force to decline from initial value to 20 N (n=5)
Group
Overdenture Removal
Mean Cycles (SD)
Time of Mean
(Years/Months/Days)
Group
Comparison*
Control
2399.80 (396.03)
1Y/8M
0D
6974.00 (396.78)
4Y/10M/4D
5D
8300.40 (1863.19)
5Y/9M/5D
10D
4977.00 (863.14)
3Y/5M/14D
15D
3553.80 (683.06)
2Y/5M/18D
AB
20D
2272.60 (115.73)
1Y/6M/28D
*Means with same uppercase letter are not significantly different (P>.05), according to Tukey
HSD test.
Al-Ghafli et al
145
September 2009
Table V. Regression analysis of logarithmic number of cycles on angle and retentive force
Model
Sum of
Squares
df
Mean
Square
290.76
<.001
Regression
74.76
24.92
Residual
6.09
71
0.09
Total
80.85
74
Unstandardized Coefficients
Model
Standard
Error
Constant
10.12
0.155
Angle
0.06
0.013
0.41
Force
0.06
0.004
Angle x force
<0.001
<0.001
DISCUSSION
The present in vitro study investigated the effect of implant angulation and cyclic dislodgement on the
retentive properties of green ball attachments. The results of this study
indicate that implant angulations
negatively affect the attachment retention longevity. Therefore, the null
hypothesis is rejected.
Retentive forces of attachments
found in this study are in agreement
with those reported previously in the
literature. Setz et al28 reported that
retentive forces of ball attachments
range from 3 to 85 N, while Petropoulos et al4 reported that the retentive
forces for ball attachments ranged
from 27.2 4.2 to 34.6 18.8 N.
The results of the present study
Al-Ghafli et al
Standardized
Coefficients
t
65.32
<.001
4.75
<.001
0.97
17.14
<.001
0.12
1.24
.219
Beta
146
CONCLUSIONS
Within the limitations of this in
vitro study, the following conclusions
were drawn:
1. Implant angulations contribute
significantly to the rate of retention
loss of the implant attachment system
evaluated.
2. Zero- and 5-degree angulations
demonstrated the greatest longevity,
or longest period of time before the
components of the attachment system required replacement.
3. Control (0/0 degrees with standard plastic component) and 20-degree angulations presented the shortest longevity, or shortest period of
time before attachment system components required replacement.
REFERENCES
1. van Waas MA. The influence of clinical variables on patients satisfaction
with complete dentures. J Prosthet Dent
1990:63:307-10.
2. Redford M, Drury TF, Kingman A, Brown
LJ. Denture use and the technical quality
of dental prostheses among persons 18-74
years of age: United States, 1988-1991. J
Dent Res 1996;75 Spec No:714-25.
3. Slaughter A, Katz RV, Grasso JE. Professional attitudes toward denture adhesives: a Delphi technique survey of
academic prosthodontists. J Prosthet Dent
1999;82:80-9.
4. el Deeb M, Tompach PC, Morstad AT,
Kwon P. Long-term follow-up of the use of
nonporous hydroxyapatite for augmentation of the alveolar ridge. J Oral Maxillofac
Surg 1991;49:257-61.
5. Jennings DE. Treatment of the mandibular
compromised ridge: a literature review. J
Prosthet Dent 1989;61:575-9.
6. Adell R, Lekholm U, Rockler B, Brnemark
PI. A 15-year study of osseointegrated
implants in the treatment of the edentulous
jaw. Int J Oral Surg 1981;10:387-416.
7. Albrektsson T. A multicenter report on
osseointegrated oral implants. J Prosthet
Dent 1988;60:75-84.
8. Ahlqvist J, Borg K, Gunne J, Nilson H,
Olsson M, strand P. Osseointegrated
implants in edentulous jaws: a 2-year longitudinal study. Int J Oral Maxillofac Implants
1990;5:155-63.
9. Adell R, Eriksson B, Lekholm U, Brnemark
PI, Jemt T. A long-term follow-up study of
osseointegrated implants in the treatment
of totally edentulous jaws. Int J Oral Maxillofac Implants 1990;5:347-59.
10.Jemt T, Chai J, Harnett J, Heath MR, Hutton JE, Johns RB, et al. A 5-year prospective
multicenter follow-up report on overdentures supported by osseointegrated
implants. Int J Oral Maxillofac Implants
1996;11:291-8.
11.Naert I, Gizani S, Vuylsteke M, van Steenberghe D. A 5-year prospective randomized
clinical trial on the influence of splinted
and unsplinted oral implants retaining a
mandibular overdenture: prosthetic aspects
and patient satisfaction. J Oral Rehabil
1999;26:195-202.
12.Feine JS, Carlsson GE, Awad MA, Chelade
A, Duncan WJ, Gizani S, et al. The McGill
consensus statement on overdentures.
Mandibular two-implant overdentures as
the first choice standard of care for edentulous patients. Gerodontology 2002;19:3-4.
13.Allen EP, Bayne SC, Brodine AH, Cronin RJ
Jr, Donovan TE, Kois JC, et al. Annual review of selected dental literature: report of
the Committee on Scientific Investigation
of the American Academy of Restorative
Dentistry. J Prosthet Dent 2003;90:50-80.
14.Engquist B, Bergendal T, Kallus T, Linden
U. A retrospective multicenter evaluation
of osseointegrated implants supporting
overdentures. Int J Oral Maxillofac Implants
1988;3:129-34.
15.Naert I, Quirynen M, Theuniers G, van
Steenberghe D. Prosthetic aspects of osseointegrated fixtures supporting overdentures. A 4-year report. J Prosthet Dent
1991;65:671-80.
16.Gotfredsen K, Holm B. Implant-supported
mandibular overdentures retained with
ball or bar attachments: a randomized
prospective 5-year study. Int J Prosthodont
2000;13:125-30.
17.Walton JN, Huizinga SC, Peck CC. Implant
angulation: a measurement technique,
implant overdenture maintenance, and
the influence of surgical experience. Int J
Prosthodont 2001;14:523-30.
18.Trakas T, Michalakis K, Kang K, Hirayama
H. Attachment systems for implant retained
overdentures: a literature review. Implant
Dent 2006;15:24-34.
19.Naert I, Alsaadi G, Quirynen M. Prosthetic
aspects and patient satisfaction with twoimplant-retained overdentures: a 10-year
randomized clinical study. Int J Prosthodont
2004;17:401-10.
20.Burns DR, Unger JW, Elswick RK Jr, Beck
DA. Prospective clinical evaluation of
mandibular implant overdentures. Part 1:
retention, stability and tissue response. J
Prosthet Dent 1995;73:354-63.
21.Sadowsky SJ. Mandibular implant-retained
overdentures: a literature review. J Prosthet
Dent 2001;86:468-73.
22.Caldwell RC. Adhesion of foods to teeth. J
Dent Res 1962;41:821-32.
23.Walmsley AD, Frame JW. Implant supported overdenturesthe Birmingham experience. J Dent 1997; 25 Suppl 1:S43-7.
24.Walmsley AD. Magnetic retention in prosthetic dentistry. Dent Update 2002;29:42833.
25.Petropoulos VC, Smith W. Maximum dislodging forces of implant overdenture stud
attachments. Int J Oral Maxillofac Implants
2002;17:526-35.
Al-Ghafli et al
147
September 2009
26.Petropoulos VC, Smith W, Kousvelari E.
Comparison of retention and retention and
release periods for implant overdenture
attachments. Int J Oral Maxillofac Implants
1997;12:176-85.
27.Gamborena JI, Hazelton LR, NaBadalung
D, Brudvik J. Retention of ERA direct
overdenture attachments before and
after fatigue loading. Int J Prosthodont
1997;10:123-30.
28.Setz I, Lee SH, Engel E. Retention of
prefabricated attachments for implant
stabilized overdentures in the edentulous
mandible: an in vitro study. J Prosthet Dent
1998;80:323-9.
Corresponding author:
Dr Konstantinos X. Michalakis
3, Greg. Palama str.
Thessaloniki 546 22
GREECE
Fax: +30 2310 272-228
E-mail: kmichalakis@the.forthnet.gr
Copyright 2009 by the Editorial Council for
The Journal of Prosthetic Dentistry.
Al-Ghafli et al