Professional Documents
Culture Documents
Aluminum
DOC316.53.01002
Aluminon Method1
Method 8012
Powder Pillows
Adapted from Standard Methods for the Examination of Water and Wastewater.
Test preparation
How to use instrument-specific information
The Instrument-specific information table displays requirements that may vary between
instruments. To use this table, select an instrument then read across to find the corresponding
information required to perform this test.
Sample cell
Cell orientation
DR 6000
2495402
DR 5000
2495402
DR 3900
2495402
2495402
Quantity
Aluminum
Page 1 of 8
Aluminum
Aluminon method with powder pillows
Programs
10 Aluminum Alumin.
Start
1.
An orange to orange-red
color will develop if
aluminum is present.
6. Blank Preparation:
Pour 10 mL of the mixture
into a square sample cell.
Insert an adapter if
required (Instrumentspecific information).
Aluminum
Page 2 of 8
A 15-minute reaction
period will begin.
Aluminum
Aluminon method with powder pillows (continued)
Zero
Read
Interferences
Table 39 Interfering substances
Interfering substance
Interference level
Greater than 300 mg/L as CaCO3. Samples with greater than
300 mg/L acidity as CaCO3 must be treated as follows:
1.
2.
Acidity
3.
Alkalinity
Fluoride
Iron
Phosphate
Polyphosphate
Aluminum
Page 3 of 8
Aluminum
Fluoride interferes at all levels by complexing with aluminum. The actual aluminum
concentration can be determined using the Fluoride interference graph when the fluoride
concentration is known.
To use the fluoride interference graph:
1. Select the vertical grid line along the top of the graph that represents the aluminum reading
obtained in step 15.
2. Locate the point on the line where it intersects with the horizontal grid line that indicates how
much fluoride is present in the sample.
3. Extrapolate the true aluminum concentration by following the curved lines on either side of the
intersect point down to the true aluminum concentration.
For example, if the aluminum test result was 0.7 mg/L Al and the fluoride present in the sample
was 1 mg/L F, the point where the 0.7 grid line intersects with the 1 mg/L F grid line falls
between the 1.2 and 1.3 mg/L Al curves. In this case, the true aluminum content would be 1.27
mg/L.
mg/L F
Aluminum
Page 4 of 8
Aluminum
Accuracy check
Standard additions method (sample spike)
Required for accuracy check:
TenSette Pipet
1. After reading test results, leave the sample cell (unspiked sample) in the instrument. Verify the
chemical form.
2. Select Options>More>Standard additions from the instrument menu.
3. Press OK to accept the default values for standard concentration, sample volume, and spike
volumes. Press EDIT to change these values. After values are accepted, the unspiked sample
reading will appear in the top row.
4. Open an Aluminum Voluette Ampule Standard, 50-mg/L Al.
5. Prepare three sample spikes. Fill three mixing cylinders* with 50 mL of sample. Use the
TenSette Pipet to add 0.1 mL, 0.2 mL, and 0.3 mL of standard, respectively, to each sample
and mix thoroughly.
6. Analyze each sample spike as described in the procedure above, starting with the 0.1 mL
sample spike. Accept each standard additions reading by pressing READ. Each addition
should reflect approximately 100% recovery.
7. After completing the sequence, press GRAPH to view the best-fit line through the standard
additions data points, accounting for matrix interferences. Press IDEAL LINE to view
relationships between the sample spikes and the Ideal Line of 100% recovery.
Standard solution method
1. Prepare a 0.4-mg/L aluminum standard solution as follows: Pipet 1.00 mL of Aluminum
Standard Solution, 100-mg/L as Al3+, into a 250-mL volumetric flask.
2. Dilute to the mark with deionized water. Prepare this solution daily. Follow the Aluminon
method with powder pillows test.
Or, use the following alternative procedure:
1. Using the TenSette Pipet, add 0.8 mL of solution from an Aluminum Voluette Ampule Standard
Solution (50-mg/L as Al) into a 100-mL volumetric flask.
2. Dilute to volume with deionized water. Follow the Aluminon method with powder pillows test.
3. To adjust the calibration curve using the reading obtained with the 0.4-mg/L aluminum
standard solution, select Options>More>Standard Adjust from the instrument menu.
4. Turn on the Standard Adjust feature and accept the shown concentration. If an alternate
concentration is used, enter the concentration and adjust the curve to that value.
Aluminum
Page 5 of 8
Aluminum
Method performance
Program
Standard
Precision
95% Confidence Limits of
Distribution
Sensitivity
Concentration change
per 0.010 Abs change
10
Summary of method
Aluminon indicator combines with aluminum in the sample to form a red-orange color. The
intensity of color is proportional to the aluminum concentration. Ascorbic acid is added before the
AluVer 3 reagent to remove iron interference. To establish a reagent blank, the sample is split after
the addition of the AluVer 3. Bleaching 3 Reagent is then added to one-half of the split sample to
bleach out the color of the aluminum aluminon complex. The AluVer 3 Aluminum reagent,
packaged in powder form, shows exceptional stability and is applicable for fresh water
applications. Test results are measured at 522 nm.
Aluminum
Page 6 of 8
Aluminum
Consumables and replacement items
Required reagents
Description
Quantity/Test
Unit
Catalog number
2242000
100/pkg
1429099
100/pkg
1457799
1429449
100/pkg
varied
500 mL
88449
Water, deionized
varied
4L
27256
Quantity
Unit
Catalog number
each
189641
Unit
Catalog number
100 mL
1417442
Required apparatus
Description
Cylinder, graduated mixing, 50-mL, with glass stopper
Recommended standards
Description
Aluminum Standard Solution, 100-mg/L as
Al3+
100 mL
2305842
16/pkg
1479210
each
2196800
Unit
Catalog number
946 mL
2088153
100 mL
247632
500 mL
254049
100/pkg
2601300
each
1970001
50/pkg
2185696
50 mL
245026
100 mL
244932
each
69000
each
1457442
each
1457446
Aluminum
Page 7 of 8
Hach Company, 2007, 2010, 2012. All rights reserved. Printed in the U.S.A.
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
Edition 7