You are on page 1of 9

INFANT INCUBATOR SERIES

MODEL:B-800

MODEL:B-2000

MODEL:B-1000

MODEL:B-3000

Optional Parts

SC:(double Wall Canopy)

GL(Phototherapy

Unit)

GB(top shelf )

HL( only KGW-800 is optional ,other is standard configuration )

YEC( infant scale )

SGL(Phototherapy

Unit with up and below)

Model

B-3000

B-2000

B-1000

B-800

Mattress to hood

450mm

450mm

450mm

410mm

Observation side of
hood

Two incline

Two incline

Two incline

Two incline

Operate port

6(2
ports)

6(2 Iris ports)

6(2 Iris ports)

4(2 Iris ports)

Tubing grommets

Front access panel

826*310mm

826*310mm

826*310mm

826*280mm

Mattress tray size

630*360mm

630*360mm

630*360mm

630*360mm

Power requirement

400VA

400VA

400VA

400VA

Temperature
mode

Air and skin


servo control

Air and skin


servo control

Air servo control

Air servo control

Air
temperature
control range

20-37

20-37

20-37

20-37

Temperature override
mode to

20-39

20-39

20-39

None

Skin temp
accuracy

sensor

30-37

30-37

None

None

Temperature override
mode to

30-38

30-38

None

None

Humidity
range

5-99%

None

None

None

Temperature
fluctuation

0.5

0.5

0.5

0.5

Mattress temperature
uniformity

0.8

0.8

0.8

0.8

Temperature rise time

30min

30min

30min

30min

Noise level within


hood environment

55db

55db

55db

55db

control

display

Iris

Power failure alarm

Yes

Yes

Yes

Yes

Fan failure alarm

Yes

Yes

Yes

Yes

Sensor failure alarm

Yes

Yes

Yes

Yes

High air temperature


alarm

1.5

1.5

1.5

1.5

Low air temperature


alarm

2.5

2.5

2.5

2.5

High
skin
temperature alarm

1.0

1.0

None

None

Low skin temperature


alarm

1.0

1.0

None

None

Over air temperature


alarm

38
40

38
40

38
40

38

Over
skin
temperature alarm

38
39

38
39

None

None

System failure alarm

Yes

Yes

Yes

none

Alarm
function

Yes

Yes

Yes

none

Environment
temperature

20-30

20-30

20-30

20-30

Environment relative
humidity

5-99%

5-99%

5-99%

5-99%

Environment
velocity of flow

0.3m/s

0.3m/s

0.3m/s

0.3m/s

RS-232 connector

Yes

Yes

Yes

None

Rail

Yes

Yes

Yes

None

10
stepless

10stepless

10stepless

None

10kg

10kg

10kg

10kg

Transfusion.pole max
load

2kg

2kg

2kg

2kg

Cabinet Max.load

6kg

10kg

10kg

None

silence

Infant
bed
adjusting
Mattress
max.load

air

tilt
tray

INFANT WARMER SERIES

MODEL:BNT-1000

MODEL: BNT-2000

MODEL:BNT-2000B

OPTIONAL PARTS

MODEL

BNT-1000

BNT-2000

BNT-2000B

Environment
temperature

18-30

18-30

18-30

Environment relative
humidity

30-75%

30-75%

30-75%

Environment
velocity of flow

0.3m/s

0.3m/s

0.3m/s

Power requirement

1000VA

1000VA

1000VA

Temperature
mode

control

Auto

Pre-heat,auto
manual

Temperature
range

control

25-37

25-37

25-37

Temperature override
model to

None

38

38

Temperature
accuracy

sensor

0.3

0.3

0.3

Mattress temperature
uniformity

2.0

2.0

2.0

Temperature rise time

45min

45min

45min

Heat output indicate

0-100%

0-100%

0-100%

Radiant box horizon


angle adjust range

090stepless

090stepless

090stepless

air

and

Pre-heat,auto
manual

and

Mattress tray size

68*56cm

68*56cm

68*56cm

Mattress
range

tilt

010stepless

010stepless

010stepless

Mattress
load

tray

10kg

10kg

10kg

Transfusion pole max


load

2kg

2kg

2kg

Shelf max load

2kg

2kg

2kg

x-ray shelf

Yes

Yes

Yes

RS-232 connector

None

Yes

Yes

Mattress
box

radiant

81cm

81cm

81cm

infrared

1-3um

1-3um

1-3um

APGAR TIMER

None

Yes

Yes

Power failure alarm

Yes

Yes

Yes

Sensor failure alarm

Yes

Yes

yes

Temperature deviation
alarm

Over
alarm

38

38, 39

38, 39

System failure alarm

None

Yes

yes

Low pressure suction

None

None

Yes

Pressure adjust range

None

None

0-22Kpa(0-165mmHG)

Operation noise

None

none

55db

With
unit

phototherapy

None

None

1 set

7L oxygen Cylinder
(
With
oxygen
manometer&Oxygen
Flowmeter

None

None

2sets

Neonate resuscitation
bag

None

None

1pc

to

Heater
wavelength

adjust
max

temperature

Infant Phototherapy Unit

B-100( With 5 blue lights)

B-200( With 5 blue lights)

B-100( With LED)

B-200(With LED )

B-400(LED)

Model

B-100

B-100 (LED)

B-200

B-200 (LED)

B-400

Input power

150VA

150VA

150VA

150VA

150VA

Light source

Blue light
tube

LED

Blue light
tube

LED

LED

Light quantity

435

435

435

Light source
effective life

2000hours

50000hours

2000hours

50000hours

50000hours

Wavelength

425-475nm

460-470nm

425-475nm

460-470nm

460-470nm

Ebi in
effective area
surface

500W/cm

500W/cm

500W/cm

500W/cm

500W/cm2

Time display
range

9999H
59MIN

9999H
59MIN

9999H
59MIN

9999H
59MIN

9999H
59MIN

Display time
accuracy

2H1MIN

2H1MIN

2H1MIN

2H1MIN

2H1MIN

Adjust angle

30,60
90

30,60
90

NONE

NONE

NONE

Adjust height

110-150MM

110-150MM

NONE

NONE

NONE

With castor

YES

YES

NONE

NONE

NONE

Working
environment
requirement

10-30

10-30

10-30

10-30

10-30