Table of Contents
Title
Page
Overview
3
Unpacking Inspection
4
Safety Information
5
Rules For Safe Operation
6
International Electrical Symbols
8
The Meter Structure
9
Rotary Switch
10
Functional Buttons
11
Display Symbols
14
Measurement Ranges
17
A.
Selecting a Measurement Range
17
B.
Manual Ranging and Autoranging
17
Measurement Operation
19
A.
AC & DC Voltage Measurement
19
B.
Measuring Continuity, Diodes &
Resistance
21
C.
Capacitance Measurement
24
D.
Frequency & Duty Cycle Measurement
26
E.
DC or AC Current Measurement
28
The POWER Button
30
The BLUE Button
30
The Use of Relative Value Mode
30
Operation of Hold Mode
31
General Specifications
32
Accuracy Specifications
33
A.
AC Voltage
33
B.
DC Voltage
33
C.
Continuity, Diodes & Resistance Test
34
D.
Capacitance
35
E.
Frequency & Duty Cycle
35
F.
DC Current
36
G.
AC Current
36
Title
Maintenance
A.
General Service
B.
Testing the Fuses
C.
Replacing the Battery
D.
Replacing the Fuses
RS232C Serial Port
A.
RS232C Port Cable
B.
Setting of RS232C Serial Ports
C.
System Requirements for Installing
the UT60A Interface Program
Page
37
37
37
38
39
41
41
41
42
Overview
This Operating Manual covers information on safety and
cautions. Please read the relevant information carefully
and observe all the Warnings and Notes strictly.
Warning
To avoid electric shock or personal injury, read the
"Safety Information" and "Rules for Safe Operation"
carefully before using the Meter.
Digital Multimeter Model UT60A (hereafter referred to
as "the Meter") has autorange and manual range options
with maximum reading 3999. The enclosure structure
design adopted advanced "co-injection" technique in
order to provide sufficient insulation.
In addition to the conventional measuring functions, it
is equipped with a RS232C standard serial port for easy
connection with computer to realize macro recording and
monitoring and capture of transient dynamic data,
displaying change of waveform during the measurement,
providing data and evidence to engineering technicians
for scientific research. This is also a highly applied digital
multimeter of good performance with full overload
protection.
Unpacking Inspection
Open the package case and take out the Meter. Check
the following items carefully to see any missing or
damaged part:
Description
Qty
Operating Manual
1 piece
Test Lead
1 pair
9V Battery (NEDA1604, 6F22 or 006P) 1 piece
(installed)
4 RS232C Interface Cable
1 piece
5 CD-ROM (Installation Guide & Computer 1 piece
Interface Software)
In the event you find any missing or damage, please
contact your dealer immediately.
Item
1
2
3
Safety Information
This Meter complies with the standards IEC61010: in
pollution degree 2, overvoltage category (CAT. III 1000V,
CAT. IV 600V) and double insulation.
CAT. III: Distribution level, fixed installation, with smaller
transient overvoltages than CAT. IV
CAT IV:Primary supply level, overhead lines, cable
systems etc.
Use the Meter only as specified in this operating manual,
otherwise the protection provided by the Meter may be
impaired.
In this manual, a Warning identifies conditions and
actions that pose hazards to the user, or may damage
the Meter or the equipment under test.
A Note identifies the information that user should pay
attention on.
International electrical symbols used on the Meter and
in this Operating Manual are explained on page 8.
l
l
l
l
l
l
l
l
10
11
12
13
7
8
15
1
kM
nFHz
VA
AC
16
14
10 ~ 14
15
16
If the range is too low for the input, the Meter displays
to indicate an overload.
If the range is too high, the Meter will not display
the most accurate measurement.
17
2.
Press RANGE.
The Meter enters the manual range mode and
turns off.
Each presses of RANGE increments the range.
When the highest range is reached, the Meter wraps
to the lowest range.
To exit the manual range mode, press and hold
RANGE for over 2 seconds.
The Meter returns to the autorange mode and
is displayed.
Note
l If you manually change the measurement range
after entering the REL or Hold modes, the Meter
exits these modes.
l Under frequency/duty cycle and capacitance
measurement, only autorange mode is available.
18
Red
Black
(figure 3)
Insert the red test lead into the HzV terminal and
the black test lead into the COM terminal.
Set the rotary switch to Hz V and press BLUE
button to select AC measurement mode.
Connect the test leads across with the object being
measured.
The measured value shows on the display, which
is effective value of sine wave (mean value
response).
19
Select
Hz
V
red
black
(figure 4)
3.
Insert the red test lead into the HzV terminal and
the black test lead into the COM terminal.
Set the rotary switch to Hz V; DC measurement
is default or press BLUE button to select DC
measurement mode.
Connect the test leads across with the object being
measured.
The measured value shows on the display.
20
Select
red
black
(figure 5)
21
Select
red
black
(figure 6)
Insert the red test lead into the HzV terminal and
the black test lead into the COM terminal.
Set the rotary switch to
and press BLUE
button to select
measurement mode.
For forward voltage drop readings on any
semiconductor component, place the red test lead
on the component's anode and place the black test
lead on the component's cathode.
The measured value shows on the display.
22
Select
red
black
(figure 7)
Insert the red test lead into the HzV terminal and
the black test lead into the COM terminal.
Set the rotary switch to
, resistance
measurement ( ) is defaults or press BLUE button
to select measurement mode.
23
Note
l The test leads can add 0.1 to 0.2 of error to
resistance measurement. To obtain precision
readings in low-resistance measurement, that is the
range of 400.0 , short-circuit the input terminals
beforehand, using the relative value function button
REL to automatically subtract the value measured
when the testing leads are short-circuited from the
reading.
l For high-resistance measurement (>1M ), it is
normal taking several seconds to obtain a stable
reading.
l If reading with shorted test leads is not 0.5 ,
check for loose test leads, incorrect function selection,
or enabled Data Hold function.
l The LCD displays
indicating open-circuit for the
tested resistor or the resistor value is higher than
the maximum range of the Meter.
l When resistance measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test.
red
black
(figure 8)
24
Insert the red test lead into the HzV terminal and
the black test lead into the COM terminal.
Set the rotary switch to .
Connect the test leads across with the object being
measured.
The measured value shows on the display.
Note
l For testing the capacitor with polarity, connect the
red test lead to anode & black test lead to cathode
instead.
l To increase the accuracy of capacitance
measurement especially when measuring under
400nF, use REL mode to automatically subtract the
Meter built-in equalized capacitance and residual
capacitance of test leads from the result.
l It takes a longer time when testing a high capacitor
value, the testing time is around 15 seconds in
100F range.
l When capacitance measurement has been
completed, disconnect the connection between the
testing leads and the circuit under test.
25
Red
Black
(figure 9)
Frequency Measurement
3.
Insert the red test lead into the HzV terminal and
the black test lead into the COM terminal.
Set the rotary switch to Hz%; frequency
measurement (Hz) is default or press Hz% button
to select Hz measurement mode.
Connect the test leads across with the object being
measured.
The measured value shows on the display.
Note
l To obtain a stable reading when measuring:
Input scope > 30V rms & 1 kHz frequency signal:
Set the rotary switch to Hz V, A Hz,
mA Hz or Hz A.
Then press Hz% to select Hz measurement
mode to obtain frequency value.
Input scope 30V rms frequency signal:
Follow the above step 2 carrying out the
measurement.
l When making frequency measurement at voltage
or current range, please mind the following signal
requirement table:
26
Signal
Frequency
Requirement
Range
200mV
A
200A
10Hz~1kHz
mA
20mA
A
2A
When Hz measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test.
A Hz,
mA Hz or
Hz A.
Black
Red
(figure 10)
Warning
Never attempt an in-circuit current measurement
where the open-circuit voltage between the circuit
and ground is greater than 250V.
If the fuse burns out during measurement, the Meter
may be damaged or the operator himself may be
hurt. Use proper terminals, function, and range for
the measurement. When the testing leads are
connected to the current terminals, do not parallel
them across any circuit.
The current measurement has 3 measurement positions
on the rotary switch: A Hz, mA Hz and Hz A.
The A Hz has a 400.0A and 4000A range, with
auto ranging; the mA Hz has a 40.00mA and 400.0mA
range, with auto ranging; Hz A position has a 4.000A
and 10.00A range, with auto ranging.
To measure current, do the following:
1. Turn off power to the circuit. Discharge all highvoltage capacitors.
2. Insert the red test lead into the AmA or 10A or
terminal and the black test lead into the COM
terminal.
3. Set the rotary switch to A Hz, mA Hz or Hz A.
Use the 10A terminal and Hz A measurement
position if the current value to be tested is an
unknown.
28
5.
6.
Note
l For safety sake, the measuring time for high current
should be 10 seconds for each measurement and
the interval time between 2 measurements should
be greater than 15 minutes.
l When current measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test.
29
31
General Specifications
l
Maximum Voltage
between any Terminals: 1000V rms.
and Grounding
l
l
l
: Display:3999.
Maximum Display
Measurement Speed : Updates 3 times /second.
o
o
o
o
Temperature: Operating: 0 C~40 C (32 F ~104 F).
o
o
o
o
Storage : -10 C~50 C(14 F~122 F).
o
o
: 75% @ 0 C - 30 C;
Relative Humidity
o
o
50% @ 31 C- 40 C.
: 2000 m; Storage: 10000 m.
Altitude:Operating
: One piece of 9V NEDA1604
Battery Type
or 6F22 or 006P.
: Display .
Battery Deficiency
: Display .
Data Holding
: Display .
Negative reading
: Display .
Overloading
Equipped with full icons display.
Auto and manual range selectable.
Dimensions (HxWxL) : 177 x 85 x 40 mm.
: Approximate 300g (battery
Weight
included).
Safety/Compliances : IEC61010 CAT.III 1000V,
CAT.IV 600V overvoltage
and double insulation
standard.
Certification
:
UL & CUL
l
l
l
l
l
l
l
l
l
l
l
l
32
A. AC Voltage
Range Resolution
4V
1mV
40V
10mV
400V
100mV
750V
1V
Overload
Protection
Accuracy
(1%+5)
1000V DC
750V AC rms
continuous.
(1.2%+5)
Remarks:
l Input impedance 10M .
l Displays effective value of sine wave (mean value
response).
l Frequency response 40Hz ~ 400Hz.
B. DC Voltage
Range
Resolution
400mV
0.1mV
4V
1mV
40V
10mV
400V
100mV
1000V
1V
(0.8%+3)
1000V DC
750V AC rms
continuous.
(0.8%+1)
(1%+3)
33
Overload
Protection
Accuracy
Resolution Accuracy
0.1
1mV
0.1
1
10
100
1k
10k
Approximate
100
N/A
(1.2%+2)
Overload
Protection
600Vp
(1%+2)
(1.2%+2)
(1.5%+2)
Remarks:
l Continuity Test (400.0 ) Range:
Buzzer beeps continuously.
Open circuit voltage approximate 0.45V.
l Diode Range:
Open circuit voltage approximate 1.48V.
Displays approximate forward voltage drop
reading 0.5V~0.8V.
l 400 ~ 40M Range:
Open circuit voltage approximate 0.45V.
34
Accuracy
10pF
100pF
1nF
10nF
100nF
Overload
Protection
(3%+5)
600Vp
(4%+5)
Resolution
N/A
Accuracy
Overload
Protection
(0.1%+3)
600Vp
0.01%
N/A
Remarks:
l 10Hz~10MHz Range:
Input sensitivity as follows:
1MHz: 300mV rms;
>1MHz: 600mV rms.
Maximum input amplitude: 30V rms.
l 0.1%~99.9% Range:
Reading is only for reference purpose.
35
DC Current
Overload
Protection
0.1 A
1 A
0.01mA
0.1mA
4A
0.001A
10A
0.01A
(1%+2)
(1.2%+3)
0.5A, 250V,
fast type Glass
fuse, 5x20mm
10A, 250V,
(1.5%+5) fast type Glass
fuse, 5x20mm
Remark:
l 4A & 10A Range:
For continuous measurement 10 seconds and
interval time between 2 measurements greater than
15 minutes.
G. AC Current
Range
Resolutio
400 A
4000 A
40mA
400mA
0.1 A
1 A
0.01mA
0.1mA
4A
0.001A
10A
0.01A
Accuracy
Overload
Protection
Remarks:
l Frequency response 40Hz ~ 400Hz.
l Displays effective value of sine wave (mean value response).
l 4A & 10A Range:
For continuous measurement 10 seconds and
interval time between 2 measurements greater than
15 minutes.
36
Maintenance (1)
This section provides basic maintenance information
including battery and fuse replacement instruction.
Warning
Do not attempt to repair or service your Meter unless
you are qualified to do so and have the relevant
calibration, performance test, and service information.
To avoid electrical shock or damage to the Meter, do
not get water inside the case.
A. General Service
l
l
l
37
Maintenance (2)
To test the fuse:
1. Set the rotary switch to
and select
by
pressing BLUE button.
2. Plug a test lead into the terminal HzV and touch
the probe tip to the 10A or AmA terminal.
l If the Meter beeps, the fuse is good.
l If the display shows , replace the fuse and
test again.
l If the display shows any other value, have the
Meter serviced and contact your dealer
immediately.
If the Meter does not work while the fuse is all right, send
it to your dealer for repair.
Screw
(figure 11)
Warning
To avoid false readings, which could lead to possible
electric shock or personal injury, replace the battery
as soon as the battery indicator " " appears.
To replace the battery:
1.
38
Maintenance (3)
2.
3.
4.
5.
Rubber foot
Screw
(figure 12)
Warning
To avoid electrical shock or arc blast, or personal
injury or damage to the Meter, use specified fuses
ONLY in accordance with the following procedure.
To replace the Meter's fuse:
1.
2.
3.
4.
39
Maintenance (4)
5.
6.
7.
40
Computer
D-sub
D-sub
9 Pin 25 Pin
Female Female
2
3
Pin
Name
Remark
RX
Receiving
Data
Transmitting
Data
Data Terminal
Ready
TD
20
DTR
5
6
5
6
7
6
GND
DSR
RTS
CTS
Grounding
Data Set
Ready
Request To
Send
Clear To Send
41
~ END ~
This operating manual is subject to change without notice.
42
43
44