MULTITESTEROPERATION SUGGESTIONS
1 Unknown quantity.
When unknown voltage or current is measured, always start
with the highest range. After the first reading, the range
switch can be reset to a lower range for a more accurate
reading.
2 Selection of a range.
For accurate reading, use the range next higher to the
approximate value known. For instance, the voltage of a
dry cell should be checked on the DC 2.5V range; on the 10V
or 50V range, no better result is obtained. For resistance,
reading is most accurate around in the middle of the scale.
3 Protective front cover,
The metacrylic-resin front cover is treated with anti-electrifi-
cation coating. Do not wipe it hard, or the coating will come
off to cause erroneous reading. Volatile solvent should not
be used to clean it; use a soft brush to remove dust.
Metacrylic resin is softer than glass quite liable to get
bruised.
4 Maintenance cares.
Avoid placing the instrument where severe shock or vibra-
tion is encountered, or where there is high temperature or
moisture. Do not leave it in the direct sun or oily atmosphere.
_—a
5 Measurement of small-size transistors.
While testing a small-size RF germanium transistor, IcEo
gradually increases heated by the current applied. It may
cause the pointer to creep, and must be checked quickly.
6 Resistor burned.
Not seldom some internal resistor is burnt on account of AC
line voltage erroneously applied to the 250mA DC or the X1
ohm range. A resistor burnt can be replaced referring to
Fig. 9 and the table. The protection device is to safe-
guard the indicator from accidental damage, while some
resistor may burn on account of the overload, but its
replacement readily restores the instrument to normal
performance.
6 NOT ;
Noss Range Resistor |
N
GO0G 250mA DC} Nat - 0.970
100mA 7” No2 - 1.59
10mA ” No3 - 22.59
Oo imA Nod - 2250
2 x100 No5 - 3.72k2
corr ” x10 No6 - 3440
1 oN ‘N ok -340
wot Nez No3 Noa xl No7 |
Fig. 9
— 2 -SUPPLEMENTARY DATA
1 Parts list.
Part No. Description
ZXRO1 | Resistor (1.3 -1.8k 1) for mV calibration RI
ZXRO2Z | Resistor (8kQ)0.5V DC multiplier R2
ZXRO03 | Resistor (40k) 2.5V DC multiplier R3
ZXRO4 | Resistor (150k) 10V DC multiplier R4
ZXRO05 | Resistor (800k(2)50V DC multiplier RS
ZXRO06 | Resistor (4MQ) 250V DC multiplier R6
ZXRO7 | Resistor (5MQ) 500V DC multiplier R7
ZXRO08 | Resistor (10MQ) 1kV DC multiplier R8
ZXRO9 | Resistor (2.75kQ) series RO
ZXR10 | Resistor (2250) tmA DC shunt R10
ZXRI11 | Resistor (22.52) 10mA DC shunt Ril
ZXRI1Z | Resistor (1.5) 100mA DC shunt Riz
ZXR13 | Resistor (0.971) 250mA DC shunt R13
ZXRI4 | Resistor (0.0250) 10A DC shunt Ri4
2XRI15 | Resistor (11.9k1) 2.5V AC multiplier RIS
ZXRI6 | Resistor (30kQ) 10V AC multiplier RIG
ZXRIT | Resistor (160k) 50V AC multiplier RI7
ZXRI8 | Resistor (800k) 250V AC multiplier R18
2XR19 | Resistor (3MQ)1kV AC multiplier R19
ZXR20 | Resistor (31.8k) for AC sensitivity calibration R20
ZXR21_| Resistor (2k) for rectifying circuit R21
ZXR22 | Resistor (2k) for rectifying circuit R22
Resistor (340) Q X1 shunt
= BH
Resistor (344.2) Q X10 shunt
Resistor (3.72k) 2100 shunt
Resistor (220k) 2 X1k shunt
Resistor (34. 2k) Q series
Resistor (334k) 0 X10k series
Potentiometer (10kQ)0 Q, adjuster
Resistor (6.7k) shunt
Germanium diode
Germanium diode
Capacitor (0.1 F)
Capacitor (0.054F)
Meter movement (442A)
Range selector switch
Front panel (501-ZXR type)
Dry cell (UM-2/1.5V), 2 required
Dry cell (015/22.5V)
Varister
Rear case (501- ZX TR type)
Carrying handle
Pin-type jack, 6 required
Range selector knob
Meter movement case
Meter movement front cover
Test leads, pair
{———
Alligator clip cord, 2 required
Polarity switch
R.S_ Reference symbol2 Arrangement of parts.
7 Bb98
“1g
SZ
MODEL 501-ZX1R SCHEMATIC DIAGRAM
PCV (20ka/V) DO mA ACV (4kQ/V) 81
c as B—#
tkaHV 500250 50 10 25.0.5 S25" 1 10 100 250 2.5 10 50 250 Ik X1Ok Xtk x10 X10 XI -
9000000000
c =e
c 2 — a a[¢
WEIZEIP IP RUIZYIC