MC-2

PRECAUTIONS ON INSPECTIONS, REPLACEMENTS,


AND ADJUSTMENTS
This chapter does not describe parts which require no special note or adjustments in
removal and installation.
For details on the removal of these parts, refer to the Parts List.
Also follow the precautions below when performing inspections, replacements, and
adjustments of parts.

WARNING
To prevent electrical shocks, be sure to turn OFF the power of this equipment
before starting work.

WARNING/CAUTION
Observe the warnings and cautions described in SAFETY PRECAUTION.

CAUTION
Be sure to wear a wristband for grounding when performing maintenance of boards and
optical units. Otherwise static electricity in the body may cause damage to electronic
parts on the boards and optical units.

CAUTION
Never remove the screws painted in red.

<INSTRUCTIONS>
The screws painted in yellow must be adjusted after the installation of parts.
When installing parts, follow the check/adjustment procedure provided.
<NOTE>
When removing the covers, insert shutter into the film tray to prevent exposure of film to light.

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-2

MC-3
The diagrams in this manual displays the CHECK and FIT ON icons where necessary.
When removing/reinstalling, following these icons.
CHECK icon:

CHECK
FIT ON icon:

006-259-01
11.10.2002 FM3601

When reinstalling removed parts, this is indicated if reinstalling


positions must be checked or adjusted.
This symbol is shown inside illustrations of parts removal
procedures. Whenever this is shown, be sure to refer to " CHECK/
Adjusting procedure".
When installing parts, this is indicated if alignment to the boss is
required.

DRYPIX 7000 Service Manual

MC-3

MC-4

HOW TO READ REMOVAL AND ADJUSTMENT VOLUME


The REMOVAL AND ADJUSTMENT (MC) Volume is constructed in such a way that it
can be referred together with the SERVICE PARTS LIST (SP) Volume.
The following describes how to refer to this REMOVAL AND ADJUSTMENT (MC)
Volume.
Example (REMOVAL AND ADJUSTMENT Volume)
Chapter No.

(1)

Upper Conveyor Unit Exit Motor (MD2)

Section. No.

(2)

FP7C0523.EPS

Removing Procedure
(1)

(2)
(3)

Remove the following cover.


Rear cover
Remove the conveyor unit exit clutch (CLD1)
Remove the exit conveyance motor (MD2).
(3) #3
[Disconnect]
Connector MD2

(3) #3
[Disconnect]
Connector CLD1

(3) #3
[Release] Clamp

(3) #4
[Disconnect]
Clutch connector

(3) #1
[Remove] 2-DT3x6

(3) #2
[Remove]
Upper conveyor unit exit motor (MD2)

(2) #1
[Remove] E5

(2) #2
[Pull out]
Upper conveyor unit exit clutch (CLD1)
FP7C0504.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

MC-58

FP7C0005.EPS

(1) Chapter No. .. The Chapter No. in the REMOVAL AND ADJUSTMENT Volume is related to
the INDEX No. of the SERVICE PARTS LIST Volume.
(2) Section No. ... The Section No. in the REMOVAL AND ADJUSTMENT Volume indicates
enclosed REF. No. (example 3 ) in the SERVICE PARTS LIST Volume
(exploded view).
For the above example, the item to be referred for MC
Conveyor Unit Exit Motor (MD2) is SP 05C 3 .
006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

Upper

MC-4

MC-5
Example (SERVICE PARTS LIST Volume)
(1)

INDEX No.

05C
E6
FP7Z0513.EPS

15

(2)

REF. No.

3
2

16

2-BR3X8

E6

6
E4
E3
DT3X6

14

E6

13
10
E6

17
6

E6

9
DT3X6

10

3-BR3X10

11

18

12
FP7Z0503.EPS

006-259-01
11.10.2002 FM3600

SP-30

FP7C0006.EPS

To refer to the service parts information on MC


(MD2), refer to SP 05C 3 .

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

Upper Conveyor Unit Exit Motor

MC-5

MC-6

CONTENTS
1. COVERS

1.1 Covers

MC-10

2. FRAME

2.1 BCR Assembly

MC-12

3. TRAY LOADING UNIT

3.1 Tray Loading Unit


3.2 Tray
3.3 Film Pack/Shutter Detection Sensor
(SA11, SA21, SA31)
3.4 Tray Detection Switch (SA12, SA22,
SA32)
3.5 Tray Lock Assembly
3.6 Tray Loading Unit Rail
3.7 Tray Rail
3.8 Squeezing Roller
3.9 Special Screw

MC-14
MC-18
MC-22
MC-24
MC-25
MC-26
MC-30
MC-32
MC-34
FP7C0001.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-6

MC-7

4. REMOVAL UNIT

4.1
4.2
4.3
4.4
4.5

Removal Unit
Removal Drive Cam
Suction Cup Arm
Suction Cup
Bellows

MC-36
MC-40
MC-44
MC-48
MC-50

5. CONVEYOR UNIT

5.1
5.2
5.3
5.4

Upper Conveyor Unit


Film Positioning Motor (Lower) (MD6)
Upper Conveyor Unit Exit Motor (MD2)
Cleaning Roller Grip Release Motor
(MD4)
Film Positioning Assembly (Lower)
Film Positioning Motor (Upper) (MD5)
Film Positioning Assembly (Upper)
Conveyor Unit Exit Guide
Conveyance Driving Roller (Upper)
Conveyance Driving Roller (Lower)
Conveyor Unit Guide
Grip Idle Gear
Lower Conveyor Unit

MC-52
MC-58
MC-60

Sub-Scanning Unit
Sub-Scanning Motor (ME1)
Sub-Scanning Roller
Film Leading Edge Detection Sensor
(SE1)

MC-82
MC-92
MC-94
MC-98

5.5
5.6
5.7
5.8
5.9
5.10
5.11
5.12
5.13

6. SUB-SCANNING UNIT

6.1
6.2
6.3
6.4

MC-62
MC-64
MC-66
MC-68
MC-72
MC-74
MC-76
MC-78
MC-79
MC-80

FP7C0002.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-7

MC-8

7. SCANNING OPTICS UNIT

7.1 Scanner Unit

MC-104

8. HEAT DEVELOPMENT
UNIT

8.1
8.2
8.3
8.4
8.5
8.6

MC-110
MC-112
MC-114
MC-116
MC-118
MC-122

9. FILM RELEASE UNIT

9.1 Film Release Unit


9.2 Density Measurement Light-Emitting
Board (LED15A)

MC-124
MC-126

10. CONTROLLER

10.1
10.2
10.3
10.4
10.5

MC-128
MC-130
MC-132
MC-134
MC-135

Cooling Fan Assembly


Filter Section Cooling Fan
Film Cooling Section
Heat Development Unit
Heat Development Rack Assembly
Heat Development Roller

Operation Panel
Controller Assembly
HDD
PC Board
MCT Board

FP7C0003.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-8

MC-9

11. SORTER

11.1 Film Conveyance Timing Belt


11.2 Jam Removing Guide Plate Magnet

MC-136
MC-139

12. VERSION UPGRADE AND


FULL INSTALLATION OF
MAIN UNIT SOFTWARE

12.1 Preparations (Same for Version Upgrades


and Full Installation)
12.2 Version Upgrade
12.3 Full Installation

MC-140

13. BOARD SETTING AND


LED LIGHTNG
INFORMATION

MC-141
MC-144

MC-

FP7C0004.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-9

COVERS

FP7Z0111.EPS

Covers

FP7C0105.EPS

Removing Procedure
(1)

Remove the following covers.


Jam clear cover
Upper left side cover
Upper right side cover
Side ejection cover
Upper left side cover

2-T4x12

Jam clear cover


Operation panel
fixing screw
2-T4x12

Upper right side cover


Side ejection cover

2-T4x8

3-BR3x8

BR3x12

FP7C0100.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-10

MC-11
(2)

Remove the following covers.


Film release unit cover
Right side cover
Left side cover
Rear cover
Lower front cover
8-T4x25
Film release unit cover
Left side cover

BR3x8

Rear cover
8-T4x16

Front cover

8-T4x16

8-BR3x8
Lower front cover

Right side cover


FP7C0101.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-11

FRAME

FP7Z0211.EPS

BCR Assembly

FP7C0201.EPS

Removing Procedure
(1)
(2)

Remove the following cover.


Rear cover
Remove the trays
3.2 Tray

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-12

MC-13
(3)

Remove the BCR assembly.

BCR assembly

#3
[Remove] 2-TP3x8

#2
[Release] Clamp

#1
[Disconnect] Connector
FP7B0541.EPS

<REMARKS>
Attaching position of the BCR

35 x 43 cm size
BCR

26 x 36/20 x 25 cm size
Shield plate

Shield plate

BCR

FP7B0542.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-13

TRAY LOADING UNIT

FP7Z0311.EPS

Tray Loading Unit

FP7C0322.EPS

<NOTE>
The following describes the procedure for the upper tray loading unit. Perform the same
procedure for the middle and lower tray loading units. Steps that differ between the three units
are indicated separately.

Removing Procedure
(1)
(2)

Remove the following covers.


Rear cover
Remove the trays.
3.2 Tray

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-14

MC-15
(3)

Remove the tray inside cover.


Tray inside cover

[Remove] 4-TP4x8
FP7C0301.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-15

MC-16
(4)

Remove the bracket.

(5)

Remove the tray loading unit.

(5) #2

(5) #2

[Disconnect]
Connector
(5) #2
[Disconnect] SA12-COM
Connector
SA12-NC

[Disconnect]
Connector
SOLA11

(5) #1

(5) #2

[Release] Clamp x 5

[Disconnect]
Connector
SA13

(4) #2
[Release] Clamp
Bracket

(4) #1
[Remove] 2-BR4x8
FP7C0302.EPS

(5) #4
[Disconnect]
Connector SA11

(5) #3
[Release] Clamp x 2

(5) #5
[Remove] 3-BR4x8
Tray loading unit
Film pack/shutter detection sensor
FP7C0303.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-16

MC-17
Reinstalling Procedure
Reinstall in the reverse order of removal.

<INSTRUCTIONS>
Bind the extra length of the cable so that it does not come out of the tray insertion
portion.

Clamp

#1
[Bind] Cable

FP7B0552.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-17

Tray

FP7C0323.EPS

Removing Procedure
(1)

Remove the following cover.


Rear cover

<NOTE>
When pulling out the tray from the equipment, insert shutter in the tray to prevent exposure of
film to light.

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-18

MC-19
(2)

Remove the tray.

#2
[Push slightly] Tray

#1
[Release manually] Tray lock

FP7B0506.EPS

#3
[Push] Stopper

#4
[Pull out] Tray

#3
Tray

[Push] Stopper
FP7C0304.EPS

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-19

MC-20
CAUTION
After removing the tray, return the rail back into the equipment as it will come in the way
of the job and is also dangerous.

#2

[Push] Rail (Left)

#1

#3

#4

[Release] Latch

[Release] Latch

[Push] Rail (Right)


FP7C0305.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.
However, there is no need to pull out the rail pushed into the equipment.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-20

MC-21
BLANK PAGE

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-21

Film Pack/Shutter Detection Sensor


(SA11, SA21, SA31)

FP7C0324.EPS

Removing Procedure
(1)

Remove the tray.


3.2 Tray

(2)

Remove the tray inside cover.


Tray inside cover

[Remove] 4-TP4x8
FP7C0301.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-22

MC-23
(3)

Remove the film pack/shutter detection sensor (SA11, SA21, SA31).

#2
[Disconnect]
Connector SA11

#1
[Release] Clamp x 2

#3
[Remove] BR4x8

Film pack/shutter detection sensor


FP7C0306.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

<INSTRUCTIONS>
Bind the extra length of the cable so that it does not come out of the tray insertion
portion.

Clamp

#1
[Bind] Cable

FP7B0552.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-23

3.4

Tray Detection Switch (SA12, SA22,


SA32)

FP7C0325.EPS

Reinstalling Procedure

<INSTRUCTIONS>
Check that the connectors are connected at the correct positions.

Tray lock switch


SA12-NC

SA12-COM
FP7C0307.EPS

Reinstall in the reverse order of removal.

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-24

Tray Lock Assembly

FP7C0326.EPS

Removing Procedure
(1)

Remove the following cover.


Rear cover

(2)

Pull out the tray.

(3)

Remove the tray lock assembly.


#3

#2
[Disconnect]
Connector
SOLA11

#2

[Remove] 2-BR4x8

[Disconnect]
Connector
SA13

#2
[Disconnect] Connector
SA12-COM

#1
[Release] Clamp

#2
[Disconnect] Connector
SA12-NC
Tray lock assembly
FP7C0308.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.
Tray lock assembly

Back side
FP7C0309.EPS

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-25

Tray Loading Unit Rail

FP7C0327.EPS

Removing Procedure
(1)

Remove the tray loading unit.

(2)

Remove the tray lock assembly.


3.5 Tray Lock Assembly

<NOTE>
When removing the tray loading unit rail, while sliding the inner rail and inner slide rail, remove
the screws.
Outer rail

Inner slide rail

Inner rail

Inner slide rail

FP7C0320.EPS

(3)

Remove the tray loading unit rail.

#1

#2

[Push to rear] Inner slide rail

[Remove] 4-T4x4

#3
[Remove] 4-T4x4

Tray loading unit rail


Auxiliary plate
Tray loading unit rail

#4
[Remove] 2-T4x4
FP7C0310.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-26

MC-27
Reinstalling Procedure
(1)

While pressing the tip in front and behind the left rail of the tray loading unit
against the aperture of the tray loading unit, secure the rail with the screws.

#2
[Press] Rail
Tray loading unit rail

#4

#1

[Press] Rail

[Temporarily tighten]
2-T4x4

#3
[Tighten] 2-T4x4

#1
[Temporarily tighten]
2-T4x4

aperture

#5
[Tighten] 2-T4x4
aperture
FP7C0311.EPS

(2)

Secure the auxiliary plate to the tray loading unit right rail using the screws.

(3)

Place the tip in front and behind the right rail of the tray loading unit on the
aperture of the tray loading unit, tighten the rail temporarily with the screws.
Auxiliary plate

(3)
[Temporarily tighten]
2-T4x4

Tray loading
unit rail (right)

(2)
[Tighten] 2-T4x4
Aperture

(3)
[Temporarily tighten]
2-T4x4

Aperture
FP7C0312.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-27

MC-28
(4)

Push the protrusion in front and behind the auxiliary plate secured to the tray
loading unit right rail against the part for attaching on the tray loading unit, and
while pressing the rail against the aperture, tighten the temporarily tightened
screws firmly.

(4) #2
(4) #1

[Press] Rail

[Push] Rail

(4) #4

(4) #5
[Press] Rail

[Push] Rail

Auxiliary
plate

(4)#3
[Tighten] 2-T4x4

Tray loading
unit rail (right)

(4)#6
[Tighten] 2-T4x4

FP7C0334.EPS

CHECK/Adjusting Procedure
After reinstalling the tray loading unit rail, insert the tray into the tray loading unit, and
check that it moves smoothly.
Apply bonding to the ten screws securing the tray loading unit rail.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-28

MC-29
BLANK PAGE

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-29

Tray Rail

FP7C0328.EPS

Removing Procedure
(1)

Remove the tray.


3.2 Tray

(2)

Remove the tray rail.


#1
[Remove] 5-T4x4
Tray rail

Tray rail

CHECK

#2
CHECK

[Remove] 5-T4x4
FP7C0313.EPS

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-30

MC-31
Reinstalling Procedure
(1)

Temporarily secure the tray rail to the tray.

(2)

While pressing the tray rail downwards, tighten the screws.


(1)
[Temporarily tighten] 5-T4x4

(2) #2

(2) #3

[Push against]
Tray rail

[Tighten] 5-T4x4

Tray rail

(2) #1
[Press downwards] Tray rail
FP7C0314.EPS

CHECK/Adjusting Procedure
After reinstalling the tray rail, insert the tray into the tray loading unit, and check that
it moves smoothly.
Apply bonding to the ten screws securing the tray rail.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-31

Squeezing Roller

FP7C0329.EPS

Removing Procedure
(1)

Remove the tray.


3.2 Tray

<NOTE>
When removing the squeezing roller, be careful not to damage the bearing of the tray,

(2)

Remove the squeezing roller.


#2
[Remove] Squeezing roller

#1
[Pull] Bearing

FP7C0315.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-32

MC-33
BLANK PAGE

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-33

Special Screw

FP7C0333.EPS

Removing Procedure
(1)

Remove the tray.


3.2 Tray

(2)

Remove the upper rear cover of the tray.

(3)

Remove the special screws.


Upper rear cover

(3)
[Remove] Special Screw

(2) #1
[Remove] 3-Q4x6
FP7C0318.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-34

MC-35
Reinstalling Procedure

<INSTRUCTIONS>
Attach the special screws so that only the ball bearing at the tip of the special screw
protrudes from the tray surface.
REMARKS>
These special screws are producing the click sound when inserting shutters.
Shutter

Ball bearing
Tray surface

Special screw
FP7C0319.EPS

Reinstall in the reverse order of removal.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-35

REMOVAL UNIT

FP7Z0411.EPS

Removal Unit

FP7C0412.EPS

Removing Procedure
(1)

Remove the tray loading unit.


3.1 Tray Loading Unit

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-36

MC-37
(2)

Remove the shield cover.

(3)

Remove the removal unit.

<NOTE>
Remove the screws (4-BR4x8) securing the removal unit while supporting the removal unit
with the hand.
To remove the removal unit, slide it to the left until the pins dislocate from the frame, then tilt
the removal unit to remove. Make sure that the removal unit does not touch the BCR
assembly at this time.

(3) #1
[Disconnect]
Connecter
CNB1

(3) #2

(3) #1

[Remove] 4-BR4x8

[Disconnect]
Connecter
MB11

(2) #2
[Remove]
Shield cover
Removal unit

CHECK

(2) #1
Pin
Shield cover

[Remove]
BR3x8
FP7C0401.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-37

MC-38
Reinstalling Procedure
Reinstall in the reverse order of removal.
CHECK/Adjusting Procedure
When reinstalling the removal unit, insert the pins on the removal unit into the holes
on the equipment.
Perform the following adjustment if the idle gear of the removal unit has been
removed or if the idle gear has been loosened.
(1)

Secure the gear while pressing the flange of the idle gear against the gears in the
arrow direction shown below.
Upper removal unit

Idle gear
Upper conveyor unit

Middle removal unit

Middle conveyor unit

Idle gear

Lower removal unit

Lower conveyor unit

Idle gear

FP7C0402.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-38

MC-39
BLANK PAGE

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-39

Removal Drive Cam

FP7C0413.EPS

Removing Procedure
(1)

Remove the removal unit.


4.1 Removal Unit

<NOTE>
When performing the following procedure, move the suction cup arm assembly to the home
position first.

(2)

Remove the gear fixing cover.

(3)

Remove the gear 1 and gear 2.

(3)
[Remove] Gear 1
Gear fixing cover

(2) #1
[Release] Clamp

(3)
[Remove] Gear 2
Bearing

(2) #1
[Release] Clamp

(2) #3
[Remove] 4-DT3x6

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

(2) #2
[Remove] E6
FP7C0403.EPS

MC-40

MC-41
(4)

Remove the removal drive cam together with the crank.


Crank
Crank

#1
[Remove] E3

#1
[Remove] E3
Removal drive cam
Removal drive cam
FP7C0404.EPS

Reinstalling Procedure
<NOTE>
When performing the following procedure, move the suction cup arm assembly to the home
position first.

(1)

Reinstall the removal drive cam to the removal unit together with the crank.

(2)

Insert pins such as hexagonal wrench into the two holes on gear 1, and adjust the
phase of the removal drive cam.

(3)

Reinstall gear 2.

Gear 1

(3)
[Reinstall]
Gear 2

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

(2) #1
[Insert] Pin
FP7C0405.EPS

MC-41

MC-42
(4)

Align the gear fixing cover to the attaching position, and pull out the pin.

(5)

Align the rib positions of gears 1 and 2.


(5) #1
[Align] Rib
Gear 2

Gear 1

Gear 1
Gear 2

(4)
[Pull out] Pin

Gear fixing cover

(5) #2
[Reinstall] Bearing
FP7C0406.EPS

(6)

Secure the gear fixing cover with the screws.

(7)

Secure the cable with the clamp.

CHECK/Adjusting Procedure
After reinstalling the removal drive cam, check that the suction cup arm assembly
moves smoothly.

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-42

MC-43
BLANK PAGE

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-43

Suction Cup Arm

FP7C0414.EPS

4.3.1

Removal/Reinstallation

Removing Procedure
(1)

Remove the tray.


3.2 tray

(2)

Remove the tray inside cover.

Tray inside cover

[Remove] 4-TP4x8
FP7C0301.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-44

MC-45
(3)

Remove the suction cup arm.


(3) #2
[Disconnect] Connector
SB2

Suction cup arm

(3) #1
[Release] Clamp

(3) #3
[Remove] 4-BR3x8
FP7C0407.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-00
10.15.2002 FM3343

DRYPIX 7000 Service Manual

MC-45

MC-46

4.3.2

Changing Film Size to 20 x 25 cm

Perform the following procedure to change the film size of the current film tray (35 x 43
cm, 26 x 36 cm) to 20 x 25 cm size.
(1)

Remove the removal unit.


4.1 Removal Unit

(2)

Remove the suction cup arm, and replace with the 20 x 25 cm size suction cup
arm.
#3
[Remove] suction cup arm

#4
[Replace] 20 x 25 cm size
suction cup arm

#2
[Remove] 4-BR3x8

#5
[Secure] 4-BR3x8

Removal unit

#1
[Disconnect] Connector

#6
[Connect] Connector

(3)

FP7C0542.EPS

Change the attaching position of the film release plate.


#1
[Remove] 4-DT3x6

#3
[Secure] 4-DT3x6
For 20 x 25 cm size

Mark

#2
[Move] Film release plate x 2
FP7C0543.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-46

MC-47
(4)

To change the film size from 35 x 43 cm to 20 x 25 cm, change the attaching


position of the barcode reader.
2.1 BCR Assembly

35 x 43 cm size
BCR

26 x 36/20 x 25 cm size
Shield plate

Shield plate

BCR

FP7B0542.EPS

(5)

Reinstall the removed parts in the reversed order of the removing procedure.

<REMARKS>
The attaching position of the film release plate of the removal unit is marked with the
respective film sizes. To return the film size from 20 x 25 cm to other sizes, refer to the
following figure and change the attaching position of the film release plate.
Not used

For 20x25 cm size

For 35x43, 26x36


cm sizes

FP7C0544.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-47

Suction Cup

FP7C0415.EPS

Removing Procedure
(1)

Remove the suction cup arm.


4.3 Suction Cup Arm

(2)

Remove the suction cup.

#2
[Remove] Suction cup

#1
[Remove] Nozzle
FP7C0408.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-48

MC-49
BLANK PAGE

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-49

Bellows

FP7C0416.EPS

Removing Procedure
(1)

Remove the suction cup arm.


4.3 Suction Cup Arm

<INSTRUCTIONS>
Rotate the bellows in the tightened state to remove.
(2)

Remove the bellows.


CHECK

#3
#2

[Remove] Bellows

[Pull out] Bellows

#1
[Lower] Plate

O ring
Stopper
FP7C0409.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-50

MC-51
Reinstalling Procedure

<INSTRUCTIONS>
If the bellows has been removed, be sure to replace the O ring.
To reinstall, insert the second groove on the top of the bellows into the plate.
To reinstall, leave a space of 1 mm between the bellows and stopper.

2nd groove

Bellows

Space of approx. 1 mm

Stopper
FP7C0410.EPS

Reinstall in the reverse order of removal.


CHECK/Adjusting Procedure
Check for air leakage as follows.
(1)

Push down the plate, and with an approximately 50 mm space between the plate
and stopper, push the suction cup against the film.

(2)

Two minutes later, check that the space is still 50 mm.

[Push down] Plate

50mm
Stopper
Film
FP7C0411.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-51

CONVEYOR UNIT

FP7Z0511.EPS

Upper Conveyor Unit

FP7C0521.EPS

Removing Procedure
(1)

Insert the shutter into the film tray.

(2)

Remove the following covers.


Upper right side cover
Rear cover

(3)

Remove the cleaning roller.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-52

MC-53
(4)

If the equipment mounts film tray 2, loosen the screws securing the idle gear of
the middle conveyor unit.
Upper removal unit

Upper conveyor unit

Middle removal unit

#1
[Loosen] 2-BR3x6
Middle conveyor unit

Lower removal unit

Lower conveyor unit

FP7C0540.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-53

MC-54
(5)

Remove the sub-scanning unit fixing bracket.


Sub-scanning unit fixing bracket

[Remove] 4-BR4x8

FP7C0613.EPS

(6)

Reinstall the sub-scanning unit fixing bracket at the position shown below.
[Tighten] 2-BR4x8
Sub-scanning unit fixing bracket

FP7C0546.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-54

MC-55
(7)

Remove the screws securing the sub-scanning unit.

#1
[Remove] 2-BR4x8
Sub-scanning unit
FP7C0535.EPS

(8)

Remove the upper conveyor unit removing jig.

[Remove] BR4x8

Upper conveyor unit removing jig


FP7C0533.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-55

MC-56
(9)

Insert the upper conveyor unit removing jig below the sub-scanning unit.
Turn the upper conveyor unit removing jig clockwise direction to lift the subscanning unit.
Sub-scanning unit

#2
[Lift] Sub-scanning unit

#1
[Turn]
Upper conveyor unit
removing jig

#3
[Insert] Pin
FP7C0534.EPS

(10) Remove the upper conveyor unit.


#1 [Disconnect]
Connector CND2

#1 [Disconnect]
Connector CND1

#1 [Disconnect]
Connector CND3
#1 [Disconnect]
Connector CND4

#1 [Disconnect]
Connector CND4

#1 [Disconnect]
Connector MD4

#1 [Disconnect]
Connector MD2
#1 [Disconnect]
Connector CLD1

#3
[Remove] BR3x8

Upper conveyor unit

#4
#2

#1 [Disconnect]

[Release] Clamp Connector CND6


006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

Shield cover

#3
[Remove] BR3x8

[Remove] 5-BR4x8
FP7C0501.EPS

MC-56

MC-57
Reinstalling Procedure

<INSTRUCTIONS>
When reinstalling the upper conveyor unit, check that the pin on the upper conveyor
unit is inserted into the groove of the sub-scanning unit rail.
Sub-scanning unit rail

Upper conveyor unit

Pin

FP7C0545.EPS

Reinstall in the reverse order of removal.


CHECK/Adjusting Procedure
Perform the following adjustment if the idle gear of the middle conveyor unit has been
removed or if the idle gear has been loosened.
(1)

Secure the gear while pressing the flange of the idle gear against the gears in the
arrow direction shown below.
Upper removal unit

Upper conveyor unit

Middle removal unit


Idle gear

Middle conveyor unit

FP7C0541.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-57

Film Positioning Motor (Lower) (MD6)

FP7C0522.EPS

Removing Procedure
(1)
(2)

Remove the following cover.


Right side cover
Remove the film positioning motor (lower) (MD6).

#2
[Disconnect] Connector MD6

#1
[Release] Clamp

#3
[Remove] 2-DT3x6
Film positioning motor (lower) MD6
FP7C0502.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-58

MC-59
Reinstalling Procedure

<INSTRUCTIONS>
Open the left and right positioning guides of the positioning guide assembly until they
touch the sides when installing the film positioning motor (MD6).
[Push] Positioning guide

Positioning guide assembly

Positioning guide

[Push] Positioning guide


FP7C0503.EPS

Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-59

Upper Conveyor Unit Exit Motor (MD2)

FP7C0523.EPS

Removing Procedure
(1)

Remove the following cover.


Rear cover

(2)

Remove the conveyor unit exit clutch (CLD1)

(3)

Remove the exit conveyance motor (MD2).


(3) #3
[Disconnect]
Connector MD2

(2) #3
[Disconnect]
Connector CLD1

(3) #3
[Release] Clamp

(3) #4
[Remove]
Clutch connector

(3) #1
[Remove] 2-DT3x6

(3) #2
[Remove]
Upper conveyor unit exit motor (MD2)

(2) #1
[Remove] E5

(2) #2
[Pull out]
Upper conveyor unit exit clutch (CLD1)
FP7C0504.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-60

MC-61
BLANK PAGE

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-61

Cleaning Roller Grip Release Motor


(MD4)

FP7C0524.EPS

Removing Procedure
(1)
(2)

Remove the following cover.


Rear cover
Rotate the worm gear of the cleaning roller grip release motor (MD4) in the arrow
direction shown in the figure with the hand so that it grips the cleaning roller.

#1
[Rotate] Worm gear

FP7C0539.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-62

MC-63
(3)

Remove the cleaning roller grip release motor (MD4).

<NOTE>
Remove the one-way gear assembly making sure not to drop the bearing.
Bearing

One-way gear
assembly
Bearing

CHECK

#1
[Disconnect]
Connector MD4

#2
[Remove] 2-DT3x6
Cleaning roller grip
release motor (MD4)
FP7C0505.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.
CHECK/Adjusting Procedure
Reinstall the motor with the D cut side of the one-way gear assembly shaft facing up.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-63

Film Positioning Assembly (Lower)

FP7C0525.EPS

Removing Procedure
(1)

Remove the upper conveyor unit.


5.1 Upper Conveyor Unit

(2)

Remove the grip position detection sensor bracket.

(3)

Remove the connector bracket.

(2) #2
[Disconnect] Connector SD3

(2) #1
[Remove] DT3x6

(2) #2

Grip position detection sensor bracket


[Disconnect]
Connector SD4

(3) #1

Connector bracket

[Remove] BR3x8

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

FP7C0506.EPS

MC-64

MC-65
(4)

Remove the film positioning assembly (lower).

Flange gear

#1
[Remove] E6

#2
[Release] Clamp

#3
[Disconnect] Connector MD6
Film positioning assembly (lower)

#4

#3

[Remove] 3-DT3x6

[Disconnect] Connector SD7


FP7C0507.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-65

Film Positioning Motor (Upper) (MD5)

FP7C0526.EPS

Removing Procedure
(1)

Remove the upper conveyor unit.


5.1 Upper Conveyor Unit

(2)

Remove the film positioning motor (upper) (MD5).

#2
[Disconnect]
Connector MD5

#1
[Release] Clamp

Film positioning assembly


(upper)
#3
[Remove]
2-DT3x6

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

Film positioning motor


(upper) (MD5)
FP7C0508.EPS

MC-66

MC-67
Reinstalling Procedure

<INSTRUCTIONS>
Open the left and right positioning guides of the positioning guide assembly until they
touch the sides when installing the film positioning motor (MD5).
[Push] Positioning guide

Positioning guide assembly

Positioning guide

[Push] Positioning guide


FP7C0503.EPS

Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-67

Film Positioning Assembly (Upper)

FP7C0527.EPS

(1)

Remove the upper conveyor unit.


5.1 Upper Conveyor Unit

(2)

Remove the film positioning motor (upper) (MD5).


5.6 Film Positioning Motor (Upper) (MD5)

(3)

Remove the conveyance guide.

#1
[Remove] Guide

Conveyor guide

#3
[Remove] 2-DT3x6

#4
[Release]
Clamp

#2
[Disconnect]
Connector SD1
FP7C0509.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-68

MC-69
(4)

Remove the turn guide.

(5)

Remove the connector bracket.

(4) #2
[Remove] Bearing

Turn guide

(4) #1
[Remove] E4

(5) #1
[Remove] DT3x6
Connector bracket

(4) #2
[Remove]
Bearing

(4) #1

(5) #1

[Remove] E4

[Remove]
DT3x6
Connector bracket
FP7C0510.EPS

(6)

Remove the stay.

(7)

Remove the screws of the conveyor unit exit guide.

(8)

Remove the film positioning assembly (upper).

(6) #1
[Remove] 4-DT3x6
Stay
Conveyor unit exit guide

(7) #1
[Remove] 2-DT3x6

(8) #1
[Release] Clampx6

(7) #2
[Remove] 2-BR3x8

(8) #3
[Remove] 3-DT3x6

(8) #2

Film positioning assembly (upper)


[Disconnect] Connector
FP7C0511.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-69

MC-70
Reinstalling Procedure

<INSTRUCTIONS>
Open the left and right positioning guides until they touch the sides when installing the
film positioning motor (MD5).
[Push] Positioning guide

Positioning guide assembly

Positioning guide

[Push] Positioning guide


FP7C0503.EPS

Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-70

MC-71
BLANK PAGE

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-71

Conveyor Unit Exit Guide

FP7C0528.EPS

Removing Procedure
(1)

Remove the upper conveyor unit.


5.1 Upper Conveyor Unit

(2)

Remove the sensor bracket.

(3)

Remove the stay.

(4)

Remove the conveyor unit exit guide.

#1
[Release] Clamp
Sensor bracket

#4

#2

[Loosen]

[Remove] DT3x6
Stay

#5
[Remove] 2-DT3x6

#7
[Release]
Conveyor unit exit guide

#3
[Remove]
4-DT3x6

#6
[Remove] 2-BR3x8
Conveyor unit exit guide
FP7C0512.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-72

MC-73
BLANK PAGE

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-73

Conveyance Driving Roller (Upper)

FP7C0529.EPS

(1)

Remove the upper conveyor unit.


5.1 Upper Conveyor Unit

(2)

Remove the knob.

(3)

Remove the handle.

Knob

(3) #1
[Remove]
BR3x10

(2) #1
[Remove] BR3x8

Handle
FP7C0513.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-74

MC-75
(4)

Remove the grip idler gear.

(5)

Remove the grip arm.

(6)

Remove the spring.

(6) [Remove] Spring


Grip arm

(6) [Remove] Spring


(5) #3
[Remove] E6

(5) #1
[Remove]
Conveyor drive gear

(5) #3
[Remove] E4
Grip arm

(4) #1

(5) #2
(5) #2

[Remove] E6

[Remove] E6

Grip idler gear

[Remove] E4

FP7C0514.EPS

(7)

Remove the conveyance grip roller (upper).

(8)

Remove the conveyance driving roller (upper).


Conveyance grip roller (upper)

(8) #2
[Remove] Bearing

(8) #1
[Remove] E6

(7) #1
[Remove] Bearing

(8) #2
[Remove] Bearing

(8) #1
[Remove] E6

(7) #1
[Remove] Bearing
Conveyance driving roller (upper)
FP7C0515.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.
006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-75

10

Conveyance Driving Roller (Lower)

FP7C0530.EPS

Removing Procedure
(1)

Remove the upper conveyor unit.


5.1 Upper Conveyor Unit

(2)

Remove the upper conveyor unit lower bracket.

(3)

Remove the drive transmission gear.

(3) #2
[Remove] E6
Bearing

(2) #3

(3) #1

[Remove] 4-DT3x6

[Remove] E6
Upper conveyor unit lower bracket

(2) #1
[Disconnect]
Connector

(2) #2
[Release] Clamp

Shield cover

Drive transmission
gear
FP7C0517.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-76

MC-77
(4)

Remove the grip idler gear.

(5)

Remove the grip arm.

(6)

Remove the spring.

Grip idler gear

(4) #1
[Remove] E6

(6) [Remove] Spring


Grip arm

(6) [Remove] Spring

(5) #3
[Remove]
Conveyor drive gear

(5) #5
[Remove] E6

(5) #4

(5) #5

[Remove] E6

[Remove] E4

(5) #2
Grip arm

[Remove]
Idler gear

(5) #1

(5) #4

[Remove] E6 [Remove] E4
FP7C0518.EPS

(7)

Remove the conveyance grip roller (lower).

(8)

Remove the conveyance driving roller (lower).

(8) #2
[Remove] Bearing

(8) #1
[Remove] E6

(7) #1
[Remove] Bearing

Conveyance driving roller


(lower)

(8) #2
[Remove] Bearing

(8) #1

Conveyance grip roller (lower)

(7) #1

[Remove] E6 [Remove] Bearing


FP7C0519.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-77

11

Conveyor Unit Guide

FP7C0531.EPS

Removing Procedure
(1)

Remove the upper conveyor unit.


5.1 Upper Conveyor Unit

(2)

Remove the film positioning assembly (lower).


5.5 Film Positioning Assembly (Lower)

(3)

Remove the conveyance driving roller (upper).


5.9 Conveyance Driving Roller (Upper)

(4)

Remove the conveyance driving roller (lower).


5.10 Conveyance Driving Roller (Lower)

(5)

Remove the conveyor unit guide.


Boss
Conveyor unit guide

Boss

#1

[Release] Conveyor unit guide

#2
[Remove] 2-BR3x8
FP7C0520.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.
006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-78

12

Grip Idle Gear

If the grip idle gear of the upper conveyor unit has been removed, reinstall the gear as
following.
Adjusting Procedure
(1)

Align the holes on the upper and lower grip release gears with the holes on the
side frame, insert the pins of a hexagonal wrench into the holes to adjust the
upper and lower arm phases, and then reintall the grip idle gear.

Pin

#1
[Reinstall] Grip idle gear
Grip release gear

#2
[Attach] E6
Pin

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

FP7C0516.EPS

MC-79

13

Lower Conveyor Unit

FP7C0536.EPS

Removing Procedure
(1)

(2)

Remove the following covers.


Right side cover
Rear cover
Remove the lower conveyor unit.

Idle gear
Middle
conveyor unit

#1
[Loosen] 2-BR3x6

#4
[Remove]
Lower conveyor unit

#3
[Remove] 4-BR4x8

#2

Shield cover

[Remove] 2-BR3x8
FP7C0537.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-80

MC-81
CHECK/Adjusting Procedure
Perform the following adjustment if the idle gear of the middle conveyor unit has been
removed or if the idle gear has been loosened.
(1)

Secure the gear while pressing the flange of the idle gear against the gears in the
arrow direction shown below.

Upper removal unit

Upper conveyor unit

Middle removal unit

Middle conveyor unit

Idle gear

Lower removal unit

Lower conveyor unit

FP7C0538.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-81

SUB-SCANNING UNIT

FP7Z0611.EPS

Sub-Scanning Unit

FP7C0609.EPS

6.1.1

Removal/Reinstallation

Removing Procedure
(1)
(2)

Insert the shutter into the film tray.


Remove the following covers.
Right side cover
Rear cover

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-82

MC-83
(3)

Remove the sub-scanning unit fixing bracket.


Sub-scanning unit fixing bracket

[Remove] 4-BR4x8

FP7C0613.EPS

(4)

Reinstall the sub-scanning unit fixing bracket at the position shown below.

#2
[Tighten] 2-BR4x8
Sub-scanning unit fixing bracket

#1
[Tighten] 2-BR4x8

Sub-scanning unit fixing bracket

FP7C0614.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-83

MC-84
(5)

Remove the shield plate.

(6)

Remove the connector cover.

(7)

Remove the sub-scanning unit.

(7) #1
[Disconnect] Connector
LDD16A-CN1

(7) #1
[Disconnect] Connector
LDD16A-CN3

(7) #1
[Disconnect] Connector
LDD16A-CN7
Shield plate

(5) #1
[Remove] 2-BR3x8

(7) #1
[Disconnect] Connector
LDD16A-CN9

(6) #1
[Loosen] 2-BR4x12
Connector cover

FP7C0601.EPS

(7) #2
[Disconnect]
Connector ME1

(7) #2
[Disconnect]
Connector CNS2

(7) #3
[Remove]
BR4x12

(7) #3
[Remove] 2-BR4x8
Sub-scanning unit
FP7C0602.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-84

MC-85
Reinstalling Procedure
(1)

Insert the sub-scanning unit into the equipment, and temporarily secure with the
screws.

#1

#1

[Temporarily tighten]
2-BR4x8

[Temporarily tighten]
BR4x12

#1
[Temporarily tighten]
2-BR4x8

FP7C0616.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-85

MC-86
(2)

Press the positioning screw against the rail, and secure the sub-scanning unit.

#2

#2

[Tighten] 2-BR4x8

[Tighten] BR4x12

#2
[Tighten] 2-BR4x8

#1
[Press against] Positioning screw

Positioning screw

Rail of equipment
FP7C0603.EPS

(3)

Remove the sub-scanning unit fixing bracket, and reinstall it to its original
position.

CAUTION
When removing the sub-scanning unit fixing bracket, be careful never remove the
screws painted in red.

CHECK/Adjusting Procedure
Check that connectors are connected properly.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-86

MC-87

6.1.2

Adjusting the Sub-Scanning Unit Positioning Screw

When adjusting the sub-scanning unit positioning screw, or if the positioning screws
has been turned accidentally, perform the following procedure.
Adjusting Procedure
(1)

Print the grid pattern.


Service Mode, [2-5] Grid.

(2)

Measure the D and E on the printed film and check that they meet the specified
values.
For 35 x 43 cm size

20

20
D

Film conveyance
direction

(Unit: mm)
Position

35 x 43cm size

26 x 36 cm size

20 x 25 cm size

D, E

14.5 1.0

27.5 1.0

26.0 1.0

1.2

1.2

1.2

|D-E|

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

FP7C0622.EPS

MC-87

MC-88
(3)

If the measured values are not the specified values, loosen the screws securing
the sub-scanning unit.

#1
[Loosen] 2-BR4x8

#1
[Loosen] 2-BR4x8

FP7C0617.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-88

MC-89
(4)

Remove the sub-scanning unit fixing bracket.


Sub-scanning unit fixing bracket

[Remove] 4-BR4x8

FP7C0613.EPS

(5)

Reinstall the sub-scanning unit fixing bracket at the position shown below.
#2
[Tighten] 2-BR4x8
Sub-scanning unit fixing bracket

#1
[Tighten] 2-BR4x8

Sub-scanning unit fixing bracket

FP7C0614.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-89

MC-90
(6)

Rotate the positioning screw in the clockwise or counterclockwise direction and


adjust.

<NOTE>
Rotating once in the clockwise or counterclockwise direction changes the D and E on the
printed film by about 0.5 mm.

[Rotate] Positioning screw


FP7C0615.EPS

Film conveyance direction

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

FP7C0624.EPS

MC-90

MC-91
(7)

Press the positioning screw against the rail, and secure the sub-scanning unit.

#2
[Tighten] 2-BR4x8

#2
[Tighten] 2-BR4x8

#1
[Press against] Positioning screw

Positioning screw

Rail of equipment
FP7C0618.EPS

(8)

Print the grid pattern.


Service Mode, [2-5] Grid.

(9)

Measure the D and E on the printed film and check that they meet the specified
values.

(10) If the measured values are not the specified values, repeat steps (3) to (9).

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-91

Sub-Scanning Motor (ME1)

FP7C0610.EPS

6.2.1

Removal/Reinstallation

Removing Procedure
(1)

Remove the sub-scanning unit.


6.1 Sub-Scanning Unit

(2)

Remove the sub-scanning motor (ME1).


#1
[Remove] Q4x14

Sub-scanning motor (ME1)

#2
[Remove] 4-TP4x8
FP7C0604.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-92

MC-93

6.2.2

Adjustment When Replacing Sub-scanning Motor

If the sub-scanning motor (ME1) has been replaced, check/adjust the film conveyance
amount as follows.
(1)

Print the grid pattern.


Service Mode, [2-5] Grid.

(2)

Measure F on the printed film and check that it meets the specified value.
For 35 x 43 cm size
20

Film conveyance
direction

(Unit: mm)
Position
F

(3)

35 x 43 cm size
413.0 1.2

26 x 36 cm size

20 x 25 cm size

226.5 1.0

225.0 1.0

FP7C0623.EPS

If the measured values are not the specified values, adjust the film conveyance
amount.
Service Mode, [5-2-3] Adjusting subscanner.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-93

Sub-Scanning Roller

FP7C0611.EPS

Removing Procedure
(1)

Remove the sub-scanning unit.


6.1 Sub-Scanning Unit

(2)

Remove the scanner unit.


7.1 Scanner Unit

(3)

Remove the sub-scanning motor (ME1).


6.2 Sub-Scanning Motor (ME1)

(4)

Remove the handle.

(5)

Remove the belt cover.

(6)

Remove the SUS belt.

(6) #1
[Remove] Spring

(5) #1
[Remove] BR4x8
Belt cover

(4) #1
[Remove] BR3x10

SUS belt

(6) #2
[Lower] SUS belt tensioner

Handle
FP7C0605.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-94

MC-95
(7)

Remove the sub-scanning roller (left).

#2
[Pull out]
Sub-scanning roller (Left)
Waveform washer

#1
[Remove] BR4x8
Sub-scanning roller (left)
W5
FP7C0606.EPS

(8)

Remove the sub-scanning unit entrance guide.

(9)

Remove the sub-scanning roller (right).


Sub-scanning unit entrance guide

(9) #3
[Pull out]
Sub-scanning unit roller (Right)

(8) #1

Waveform washer

[Remove] BR4x8
W5

Sub-scanning roller (right)

(9) #1
[Remove] BR4x8

(9) #2
[Lower] SUS belt tensioner
FP7C0607.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-95

MC-96
Reinstalling Procedure
<NOTE>
When installing the sub-scanning roller, be sure to reinstall the waveform washer.
After reinstalling the sub-scanning roller, rotate the roller four or five times with the hand, and
check that it turns smoothly.

Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-96

MC-97
BLANK PAGE

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-97

Film Leading Edge Detection Sensor


(SE1)

FP7C0612.EPS

Removing Procedure
(1)
(2)

(3)

Insert the shutter into the film tray.


Remove the following covers.
Right side cover
Rear cover
Remove the sub-scanning unit fixing bracket.
Sub-scanning unit fixing bracket

[Remove] 4-BR4x8

FP7C0613.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-98

MC-99
(4)

Reinstall the sub-scanning unit fixing bracket at the position shown below.
#2
[Tighten] 2-BR4x8
Sub-scanning unit fixing bracket

#1
[Tighten] 2-BR4x8

Sub-scanning unit fixing bracket

FP7C0614.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-99

MC-100
(5)

Remove the shield plate.

(6)

Remove the connector cover.

(7)

Draw the sub-scanning unit.

(7) #1
[Disconnect] Connector
LDD16A-CN1

(7) #1
[Disconnect] Connector
LDD16A-CN3

(7) #1
[Disconnect] Connector
LDD16A-CN7
Shield plate

(5) #1
[Remove] 2-BR3x8

(7) #1
(6) #1

[Disconnect] Connector
LDD16A-CN9

[Loosen] 2-BR4x12
Connector cover
FP7C0601.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-100

MC-101
(7)#2
[Disconnect] Connector ME1

(7)#2
[Disconnect] Connector CNS2

(7)#3
[Remove] 2-BR4x8
Sub-scanning unit
FP7C0619.EPS

(8)

Remove the film leading edge detection sensor (SE1).

#1
[Disconnect] Connector
SED
Film leading edge detection sensor (SE1)

#2
[Remove] 2-BR3x8
FP7C0608.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-101

MC-102
Reinstalling Procedure
Reinstall in the reverse order of removal. However, perform the following when
securing the sub-scanning unit into the equipment.
(1)

Insert the sub-scanning unit into the equipment, and temporarily secure with the
screws.

#1
[Temporarily tighten]
2-BR4x8

#1
[Temporarily tighten]
2-BR4x8

FP7C0620.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-102

MC-103
(2)

Press the positioning screw against the rail, and secure the sub-scanning unit.

#2
[Tighten] 2-BR4x8

#2
[Tighten] 2-BR4x8

#1
[Press against] Positioning screw

Positioning screw

Rail of equipment
FP7C0621.EPS

(3)

Remove the sub-scanning unit fixing bracket, and reinstall it to its original
position.

CAUTION
When removing the sub-scanning unit fixing bracket, be careful never remove the
screws painted in red.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-103

SCANNING OPTICS UNIT

FP7Z0711.EPS

Scanner Unit

FP7C0703.EPS

CAUTION
The LD/IC may damage due to static electricity. When removing/reinstalling it, take
anti-static measures by wearing a wrist band, etc.
Do not remove the upper cover of the scanner unit.
Do not place your finger inside the laser beam window (at the bottom of the scanner
unit).
Handle the optical unit gently, making sure not to subject it to shock.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-104

MC-105

7.1.1

Removal/Reinstallation

Removing Procedure
(1)

Remove the sub-scanning unit.


6.1 Sub-scanning Unit

(2)

Remove the scanner unit.

#1
[Remove] 4-BR4x12
Scanner unit

CHECK
CHECK

CHECK

FP7C0701.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-105

MC-106
Reinstalling Procedure
Reinstall in the reverse order of the removing procedure. However perform the
following when installing the scanner unit to the sub-scanning unit.

<INSTRUCTIONS>
When pushing the scanner unit against the positioning plate, push gently without
subjecting it to shock.
(1)

Align the scanner unit to the reinstalling position of the sub-scanning unit, and
secure temporarily with the screws.

(2)

Push the two positioning brackets and protruded part on the scanner unit against
the sub-scanning unit at the positions shown in the following figure.

(3)

Secure the scanner unit with the screws.


Scanner unit

(2)

(1)
[Secure temporarily] 4-BR4x12

[Push] Scanner unit

(3)
[Tighten] 4-BR4x12

(1)
[Secure temporarily] 4-BR4x12

(3)
[Tighten] 4-BR4x12

(2)
[Push] Scanner unit
FP7C0702.EPS

CHECK/Adjusting Procedure
Make sure that there is no gap between the sub-scanning unit positioning plate and
scanner unit positioning parts.
Make sure that the connectors are connected properly.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-106

MC-107

7.1.2

Adjustments in Replacement of Scanner Unit

If the scanner unit has been replaced, the data in the scanner unit data FD provided
with the scanner unit needs to be sent to the equipment.
(1)

Turn ON the power of the equipment.

(2)

Prepare the PC for servicing.


INSTALLATION, 10.1 Connecting the PC for Servicing to the Network
INSTALLATION, 10.2 Starting the DPX7 PC Tool/Connecting DPX7 PC
Tool to the Equipment

(3)

Import the individual data.


Printer name importing data [Control] [File Save/Load] [Save from the
printer] [Execute]
#1

#3

#4

#2

#5

FP7C0F09.EPS

The individual data is sent to the PC for servicing. When the data transfer is
completed, the [File transfer is completed.] screen appears.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-107

MC-108
(4)

Insert the scanner unit data FD provided with the scanner unit into the FDD of the
PC for servicing.

(5)

Copy scanner unit data to the PC for sevicing.


Printer name [Data Transfer] [FROM FD]
Scanner unit data is copied to the PC for servicing.
#1

#2

#3

FP7C0F11.EPS

When completed, the [FD File transfer is completed.] screen appears.


(6)

Send the individual data containing the scanner unit data copied to the
equipment.
Printer name [Control] [File Save/Load] [Load to the printer]
[Execute]
The individual data is sent to the equipment, and only the scanner unit data is updated.
When the data transfer is completed, the [File transfer is completed.] screen appears.

(7)

End the DPX7 PC Tool.

(8)

Reboot the equipment, and start the M-Utility.

(9)

Output the grid pattern using the M-Utility.


SERVICE MODE, [2-5] Grid

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-108

MC-109
(10) Measure C of the film output, and check that the values satisfy the specification.
For 35 x 43 cm size

Film conveyance
direction

(Unit: mm)
Position

35 x 43 cm size

|A-B|
C

26 x 36 cm size

20 x 25 cm size

1.2

1.2

2.0

300 1.2

300 1.2

200 1.2

FP7C0705.EPS

(11) If they do not, change the value of [4-3-1] Scanning Width of the M-Utility.
SERVICE MODE, [4-3-1] Scanning Width
(12) Measure A and B of the film output, and check that the values satisfy the
specification.
(13) If they do not, change the value of [4-3-2] Scanning Position of the M-Utility.
SERVICE MODE, [4-3-2] Scanning Position
(14) Perform automatic density correction.
SERVICE MODE, [3-1] AUTO F.D.C.
(15) Check the film density.
SERVICE MODE, [3-2] Check Density

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-109

HEAT DEVELOPMENT UNIT

FP7Z0811.EPS

Cooling Fan Assembly

FP7C0814.EPS

Removing Procedure
(1)

(2)

Remove the following covers.


Jam clear cover
Upper left cover
Remove the fan fixing bracket.
Cooling fan assembly

#1
[Remove] 4-BR3x8

Fan fixing bracket

FP7C0801.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-110

MC-111
(3)

Remove the cooling fan assembly.

#2
[Remove] 4-BR4x8
Cooling fan assembly

#1
[Disconnect] Connector CNG3

FP7C0802.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-111

Filter Section Cooling Fan

FP7C0815.EPS

Removing Procedure
(1)

Remove the cooling fan assembly.


8.1 Cooling Fan Assembly

(2)

Remove the fan duct.

(3)

Remove the filter section cooling fan.


Fan duct

#1
[Remove] 4-BR3x8

Auxilary plate

#2
[Remove] 2-BR4x40

Filter section cooling fan


FP7C0803.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-112

MC-113
Reinstalling Procedure

<INSTRUCTIONS>
Reinstall the filter section cooling fan as shown in the figure.
Filter section cooling fan
Label

Cable

Cooling fan assembly


FP7C0804.EPS

Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-113

Film Cooling Section

FP7C0816.EPS

Removing Procedure
(1)

Remove the following covers.


Jam clear cover
Upper left cover
Upper right cover

(2)

Remove the operation panel.


10.1 Operation Panel

(3)

Remove the film release unit.


9.1 Film Release Unit

(4)

Remove the film cooling section.


Round hole

#1
[Release]
Clamp

#2

CHECK

[Disconnect]
Connector THG10
Film cooling section

CHECK

#3

#4

[Loosen] 2-BR4x8

[Remove] 5-BR4x8
FP7C0805.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-114

MC-115
Reinstalling Procedure

<INSTRUCTIONS>
When reinstalling the film cooling section, press the shaft against the bracket as shown
in the figure, and secure with the screws.

[Press] Film cooling section shaft


FP7C0806.EPS

Reinstall in the reverse order of removal.


CHECK/Adjusting Procedure
After reinstalling the film cooling section, check from the round hole on the back of
the film cooling section that the gears of the film cooling section and heat
development unit are engaged.
After reinstalling the film cooling section, rotate the jam handle, and check that it
moves smoothly.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-115

Heat Development Unit

FP7C0817.EPS

Removing Procedure
(1)

(2)

Remove the following covers.


Jam clear cover
Upper left side cover
Upper right side cover
Rear cover
Right side cover
Left side cover
Remove the sub-scanning unit.
6.1 Sub-Scanning Unit

(3)

Remove the film release unit.


9.1 Film Release Unit

(4)

Remove the cooling fan assembly.


8.1 Cooling Fan Assembly

(5)

Remove the operation panel.


10.1 Operation Panel

(6)

Remove the film cooling section.


8.3 Film Cooling Section

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-116

MC-117
(7)

Remove the left inside cover.

(8)

Remove the top panel.


Left inside cover

(7) #1

Top panel

[Remove] 2-BR3x8

(8) #1
[Remove] 4-BR4x8

(8) #2
[Release] Clamp
FP7C0807.EPS

(9)

Remove the heat development unit.

#1
[Disconnect]
Connector CNG2

Heat development unit

#1
[Disconnect]
Connector CNG4

#3
[Remove] 3-BR4x8

#1
[Disconnect]
Connector CNG1

#1
[Disconnect]
Connector MG1

#2
[Release] Clamp
FP7C0808.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-117

Heat Development Rack Assembly

FP7C0818.EPS

Removing Procedure
(1)

Insert the shutter into the film tray.

(2)

Remove the following covers.


Jam clear cover
Upper left side cover
Rear cover
Left side cover

(3)

Remove the film release unit.


9.1 Film Release Unit

(4)

Remove the film cooling section.


8.3 Film Cooling Section

(5)

Remove the cooling fan assembly together with the bracket.


Cooling fan assembly

#2
[Remove] 2-BR3x8
Bracket

Bracket

#3
[Remove] 4-BR4x8

#1
[Disconnect] Connector CNG3
FP7C0823.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-118

MC-119
(6)

Remove the left inside cover.


Left inside cover

[Remove] 2-BR3x8

FP7M0320.EPS

(7)

Disconnect the connectors and earth cables.

(8)

Remove the rack fixing brackets (small).

(9)

Remove the rack fixing bracket (large).

<NOTE>
Do not remove the left and right fixing pins on the heat development rack assembly.

(10) Remove the heat development rack assembly.

(7) #2
[Disconnect] Connector THG7-9

(8) #2

(7) #2

[Remove] BR4x8

[Disconnect]
Connector HEAT3

CHECK

Rack fixing bracket (small)

(8) #1
CHECK

[Remove]
BR4x8

Rack fixing
bracket (small)

(10)

(7) #1
[Remove] TP4x8

[Remove]
Heat development rack assembly

(9)
[Remove] 2-BR4x8 Rack fixing bracket (large)
FP7C0809.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-119

MC-120
Reinstalling Procedure
<NOTE>
When reinstalling the adiabatic cover, do not sandwich the thermistor cable.

Reinstall in the reverse order of removal.


CHECK/Adjustment Procedure
Check that connectors are connected properly.

Third rack assembly

THG7-9
HEAT3
Second rack assembly
THG4-6
HEAT2
THG1-3
HEAT1
First rack assembly
FP7C0810.EPS

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-120

MC-121
BLANK PAGE

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-121

Heat Development Roller

FP7C0819.EPS

Removing Procedure
(1)

Remove the heat development rack assembly, and remove adiabatic cover.
8.5 Heat Development Rack Assembly

(2)

Remove the fixing pins.

(3)

Remove the heater assembly.

(4)

Remove the heat development roller.


(3)
[Remove] Heater assembly

(2) #1
[Remove] Fixing pin (long)

(4) #1
[Remove] WP3x10
Bearing

(2) #2
[Remove]
Fixing pin (short)

Bearing

Heat development roller


FP7C0811.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-122

MC-123
BLANK PAGE

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-123

FILM RELEASE UNIT

FP7Z0911.EPS

Film Release Unit

FP7C0904.EPS

Removing Procedure
(1)

Remove the following covers.


Jam clear cover
Upper left side cover
Upper right side cover
Side ejection cover

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-124

MC-125
(2)

Remove the film release unit.

#1
[Release] Clamp

#2
[Disconnect]
Connector CNJ2

#2
[Disconnect]
Connector CNJ1

#3
[Remove] 4-BR4x8

Film release unit

FP7C0901.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-125

Density Measurement Light-Emitting


Board (LED15A)

FP7C0905.EPS

Removing Procedure
(1)

Remove the following covers.


Jam clear cover
Upper left side cover
Upper right side cover
Film release unit cover

(2)

Remove the LED15A installing bracket.

(3)

Remove the density measurement light-emitting board.


(2) #3
[Disconnect]
Connector LED-CN85

(2) #1
[Remove] 2-BR3x8

(3) #1
[Remove] BR3x12

Shield
plate

(3) #1
[Remove] BR3x8

Spacer

(2) #2
[Remove] Clamp
Density measurement
light-emitting board (LED15A)

(2) #3
[Disconnect] Connector SJ1
FP7C0902.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-126

MC-127
Reinstalling Procedure

<INSTRUCTIONS>
Press the LED cover of the density measurement light-emitting board against the
position shown in the figure below, and secure.
[Press]
Density measurement light-emitting board (LED15A)

FP7C0903.EPS

Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-127

10

CONTROLLER

FP7Z0A11.EPS

10 -

Operation Panel

FP7C0106.EPS

Removing Procedure
(1)

Remove the following covers.


Jam clear cover
Upper left side cover

(2)

Remove the operation panel.

#2

#2

[Disconnect]
Connector
PANEL-CN2

[Disconnect]
Connector
PANEL-CN1

#4
[Disconnect]
Connector
PANEL-CN5

#4
[Disconnect] Connector
PANEL-CN7

#4

Operation panel

[Disconnect]
Connector
PANEL-CN8

#4
[Disconnect]
Connector
PANEL-CN6

#4

#3

[Disconnect]
Connector
PANEL-CN4

[Remove] BR3x8

#1
[Remove] 2-BR3x8
006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

FP7C0104.EPS

MC-128

MC-129
Reinstalling Procedure
<NOTE>
When reinstalling the upper left side cover, make sure that it does not interfere with the
operation panel.

Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-129

10 -

Controller Assembly

FP7C0A04.EPS

Removing Procedure
(1)

(2)

Remove the following covers.


Left side cover
Rear cover
Lower front cover
Remove the controller assembly.
#2
[Disconnect]
Shaded connector

#2
[Disconnect]
Connector PC-CN4

#2
[Disconnect]
Connector
PRINT16A-CN2

#2
[Disconnect]
Connector

[Disconnect]
Connector
PRINT16A-CN4

#1
[Release]
Clamp

#4
[Remove]
3-BR3x8

#2

Controller
assembly

#2
[Disconnect]
Connector
PC-CN26

#2
[Disconnect]
Connector CNK1

#1
[Release]
Clamp

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

#2
[Disconnect]
Connector
PRINT16A-CN16

FP7C0A01.EPS

MC-130

MC-131

#5
[Pull out] Controller assembly
FP7C0A02.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-131

10 -

10.2

HDD

FP7C0A05.EPS

Removing Procedure
(1)

Remove the controller assembly.


10.2 Controller Assembly

(2)

Remove the HDD.


#2
[Remove] 4-BR3x12
HDD

#1
[Disconnect]
Connector HDD

#1
[Disconnect]
Connector HDD
FP7C0A03.EPS

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-132

MC-133
Replacing Procedure
The service parts HDD is installed with the OS and main unit software.
However, because the main unit software may not be of the latest version, if the HDD
has been replaced, perform full installation of the main unit software and re-installation
of individual data.
Prepare the latest main unit software CD-R when replacing the HDD.
(1)

Turn OFF the power of the equipment.

(2)

Replace the HDD.


Removing Procedure/ Reinstalling Procedure

(3)

Turn ON the power of the equipment, and check the version of the main unit
software displayed at startup.

(4)

Start the M-Utility, and set the network.


INSTALLATION, 7.2.2 System Management/Set Network

(5)

End the M-Utility, and turn OFF and then ON the power of the equipment.

(6)

Prepare the PC for servicing.


INSTALLATION, 10.1 Connecting the PC for Servicing to the Network
INSTALLATION, 10.2 Starting the DPX7 PC Tool/Connecting DPX7 PC
Tool to the Equipment

<INSTRUCTIONS>
If the version of the main unit software in the HDD is old: Perform steps (7) and
onwards.
If the version of the main unit software in the HDD is latest: Perform steps (8) and
onwards.
(7)

Perform full installation of the latest main unit software and installation of the
application program of the printer in the MCT board.
12.3 Full Installation Steps (1) to (3)

(8)

Re-install the individual data.


12.3 Full Installation Step (4)

(9)

Set the system date, film counter, and remaining number of films.
12.3 Full Installation Steps (5) to (7)

(10) Reboot the equipment.


12.3 Full Installation Step (8)
(11) Click [Exit] and shutdown the DPX7 PC Tool.
(12) Remove the CD-R from the PC for servicing.
(13) Shutdown the PC for servicing.
(14) Disconnect the network cable of the PC for servicing from the network to which
the equipment is connected.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-133

10 -

PC Board

FP7C0A06.EPS

CAUTION
Replace the PC board if the lithium battery has worn out.
Return the old PC board to the Parts Center (factory).
Improper replacement of the battery may result in explosion. When replacing the
lithium battery, use the battery designated by Fuji Photo Film Co., Ltd. or use an
equivalent battery. Dispose the old battery as specified by Fuji Photo Film Co., Ltd.
Removing Procedure
(1)

Remove the controller assembly.


10.2 Controller Assembly

(2)

Remove the PRN board, MCT board, and image memory board.
SERVICE PARTS LIST, 10. Controller

(3)

Disconnect all connectors on the PC board.

(4)

Remove the PC board.

Reinstalling Procedure
Reinstall in the reverse order of removal.

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-134

10 -

MCT Board

FP7C0A07.EPS

Removing Procedure
(1)

Remove the controller assembly.


10.2 Controller Assembly

(2)

Remove the MCT board.


SERVICE PARTS LIST 10. Controller

Reinstalling Procedure
Reinstall in the reverse order of removal.
Replacing Procedure
If the MCT board has been replaced, the software on the HD (printer installer, printer
OS, printer application) needs to be installed with a new MCT board.
(1)

Turn OFF the power of the equipment.

(2)

Replace the MCT board.


Removing Procedure/ Reinstalling Procedure

(3)

Turn ON the power of the equipment, and start the M-Utility.

(4)

Install the software on the HD in the MCT board in the following order; PRT INST
PRT OS PRT APL.
SERVICE MODE, [1-1-9] Reload Software

(5)

Reboot the equipment.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-135

11

SORTER

FP7C0B02.EPS

11 -

Film Conveyance Timing Belt

FP7C0B03.EPS

Belt Structure

Top belt

Inside belt

Bottom belt

FP3F0S43.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-136

MC-137
Reinstalling Procedure
(1)

There are three timing belts. Install the inside belt first and then the top and
bottom belts.
Top belt

Top belt

Bottom belt

Inside belt
Bottom belt

Inside belt

FP3F0S42.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-137

MC-138
Adjusting Procedure
(1)

With the fixing screws of the tension pulleys loosened, secure the tension pulleys
at the position where they are in balance by the force of the springs.
Tension spring

Top belt
Fixing screw

Bottom belt

Fixing screw

Inside belt
Fixing screw

Tension spring
Tension spring
FP3F0S40.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-138

11 -

Jam Removing Guide Plate Magnet

FP7C0B04EPS

(1)

Loosen the fixing screws of the magnet.

(2)

Close the jam removing guide plate, and tighten magnet fixing screws where it
touches the stopper.
Stopper

#2 [Push] Guide plate


#1 [Loosen] Two screws
#3 [Tighten] Two screws

Magnet
FP3F0S41.EPS

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-139

MC-140

12.

VERSION UPGRADE AND FULL INSTALLATION


OF MAIN UNIT SOFTWARE
The main unit software is supplied on CD-R. The DRX7 PC Tool is required for version
upgrading and full installation of the main unit software.
Version upgrade:
The concerned files excluding individual data of the currently
installed software are overwritten.
Full installation:
The concerned files including individual data of the currently
installed software are overwritten.
<NOTE>
As performing full installation will clear the individual data in the HD, the individual data
needs to be installed after completing full installation.
Use the latest individual data for installation.
In version upgrades, individual data will be preserved.

<REMARKS>
The following shows the times required for software installation. Take note that these times
differ according to the performance and network environment of the PC for servicing used.
Version upgrade (PC for servicing Equipment): Approx. 2 minutes
Full installation (PC for servicing Equipment): Approx. 3 minutes
PRT APL installation (HDD MCT board): Approx. 1 minute

12.1

Preparations (Same for Version Upgrades and Full


Installation)

(1)

Turn ON the power of the equipment.

(2)

Prepare the PC for servicing.


INSTALLATION, 10.1 Connecting the PC for Servicing to the Network
INSTALLATION, 10.2 Starting the DPX7 PC Tool/Connecting to DPX7 PC
Tool the Equipment

006-259-01
11.10.2002 FM3601

DRYPIX 7000 Service Manual

MC-140

MC-141

12.2

Version Upgrade

Flow of Data When Version Upgrading


DRYPIX 7000

PC for servicing (DPX7 PC Tool)

HDD

(1)

(2)
Formatter
program

FTP folder

CD-R
(Main unit software)

Printer
program

(3)
MCT board
Memory

FP7C0F08.EPS

Version Upgrading Procedure


(1)

Insert the latest main unit software CD-R in the CD-ROM drive of the PC for
servicing, and copy the main unit software to the FTP folder of the PC for
servicing.
Printer name [Soft Ver.UP] CD-ROM drive [Copy]
#1

#2

#3

#4

FP7C0F01.EPS

The main unit software is copied to the FTP folder of the PC for servicing.
When completed, [File Copy Finish!!] screen appears.
006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-141

MC-142
(2)

Write the main unit software in the HD of the equipment from the PC for servicing
(FTP folder).
[Verup] [Execute]

#1

#2

FP7C0F02.EPS

The main unit software is installed (version upgraded) in the equipment.


When installation completed, the [System File Load Finish!!] screen appears.
(3)

Of the software written in the HD of the equipment, write the printer application
program in the MCT board.
[PRT APL] [Install]

#1

#2

FP7C0F03.EPS

The printer application program in the HD is installed in the MCT bard.


When installation completed, the [Software Install Finish!!] screen appears.
<REMARKS>
The version of the main unit software is upgraded after the equipment is rebooted.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-142

MC-143
(4)

Reboot the equipment from the PC for servicing to upgraded the software.
Printer name [Control] [System Config] [Reboot] [Reboot All]
[Reset]
#1

#4

#3

#2

#6

#5

FP7C0F04.EPS

The equipment is rebooted.


(5)

Click [Exit] to end DPX7 PC Tool.

(6)

Remove the CD-R from the PC for servicing.

(7)

Shutdown the PC for servicing.

(8)

Disconnect the network cable of the PC for servicing from the network to which
the equipment is connected.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-143

MC-144

12.3

Full Installation

Flow of Data When Performing Full Installation


PC for servicing (DPX7 PC Tool)

DRYPIX 7000
HDD

(1)

(2)
Formatter
program
FTP folder

CD-R
(Main unit software)

Printer
program

(4)
Indv Folder

(3)
MCT board
Memory

FP7C0F07.EPS

Full Installation Procedure


(1)

Insert the latest main unit software CD-R in the CD-ROM drive of the PC for
servicing, and copy the main unit software to the FTP folder of the PC for
servicing.
Printer name [Soft Ver.UP] CD-ROM drive [Copy]
#1

#2

#3

#4

FP7C0F01.EPS

The main unit software is copied to the FTP folder of the PC for servicing.
When completed, [File Copy Finish!!] screen appears.
006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-144

MC-145
(2)

Write the main unit software in the HD of the equipment from the PC for servicing
(FTP folder).
[Full] [Execute]

#1

#2

FP7C0F05.EPS

The main unit software is installed in the equipment.


When installation completed, the [System File Load Finish!!] screen appears.
(3)

Of the software written in the HD of the equipment, write the printer application
program in the MCT board.
[PRT APL] [Install]

#1

#2

FP7C0F03.EPS

The printer application program in the HD is installed in the MCT board.


When installation completed, the [Software Install Finish!!] screen appears.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-145

MC-146
(4)

Install the individual data.

<INSTRUCTIONS>
The individual data needs to be installed if full installation has been performed. To
install, use the latest individual data of the corresponding printer saved in the PC for
servicing.
Installing the Individual Data in the PC for Servicing
(i)

Printer name [Control] [File Save/Load] [Load to the printer]


[Execute]
#1

#3

#4

#2

#5

FP7C0F06.EPS

The individual data in the PC for servicing is sent to the equipment.


When the data transfer is completed, the [File transfer is completed.] screen appears.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-146

MC-147
When Installing the Individual Data Saved on FD
(i)

Insert the FD with the latest individual data in the FDD of the PC for servicing.

(ii)

Copy the individual data saved on FD to the PC for servicing.


Printer name [Data Transfer] [FROM FD]
#1

#2

#3

FP7C0F11.EPS

The individual data is copied to the PC for servicing.


When the data copy is completed, the [FD File transfer is completed.] screen appears.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-147

MC-148
(iii)

Install the individual data in the equipment.


Printer name [Control] [File Save/Load] [Load to the printer]
[Execute]
#1

#3

#2

#4

#5

FP7C0F06.EPS

The copied individual data in the PC for servicing is sent to the equipment.
When the data transfer is completed, [File transfer is completed.] screen appears
(5)

Set the system date and time.


Printer name [Control] [System Config] [Set Date/Time]
SERVICE MODE, [PC9-2-1] System Config

(6)

Set the film counter.


Printer name [Control] [System Config] [Set Counters]
SERVICE MODE, [PC9-2-5] Set Counters

(7)

Set the number of remaining films.


Printer name [Control] [System Config] [Set Remain. Films]
SERVICE MODE, [PC9-2-6] Set Remain. Films

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-148

MC-149
(8)

Reboot the equipment from the PC for servicing so that the changes made in the
main unit software becomes effective.
Printer name [Control] [System Config] [Reboot] [Reboot All]
[Reset]
#1

#4

#3

#2

#6

#5

FP7C0F04.EPS

The equipment reboots.


<REMARKS>
The main unit software is replaced with that fully installed after the equipment is rebooted.

(9)

Click [Exit] and shutdown the DPX7 PC Tool.

(10) Remove the CD-R and FD from the PC for servicing.


(11) Shutdown the PC for servicing.
(12) Disconnect the network cable of the PC for servicing from the network to which
the equipment is connected.

006-259-02
04.01.2003 FM3978

DRYPIX 7000 Service Manual

MC-149

You might also like