Overview
This procedure is for a new installation of Centralized Monitor ESRS IP Client. ESRS IP Client
software kit is installed using the InstallAnywhere installer which verifies the customers workstation
environment, downloads ESRS GW Device Client software from Powerlink, installs GW Device
Client, Unisphere monitoring agent, ConnectEMC, ACU, and UDoctor. Next the installer discovers
storage systems to monitor, configures Unisphere monitor, ACU deploys/registers monitored storage
systems, and executes a test call home.
The Centralized Monitor ESRS IP Client software is designed to be customer installable not
requiring written procedures to install and configure. All pertinent information to performing the install
is available within the installer wizard screens. These procedures a guide to assist you through the
installation process.
Done
Done
Installer Overview
ESRS IP Client uses InstallAnywhereapplication that automates the installation and configuration.
You supply the required configuration data elements and the program will install and configure all
required software components.
All installation error conditions will be clearly identified during the installation process. However, if you
need to examine the log file, the filename is esrsipclient_installer.log. It is located in C:\Documents
and Settings\Administrator\emc\ESRSIPClient directory.
.4
Done
Before installing the ESRS IP client for VNX on a Windows 7 host with the Windows
firewall enabled, ensure TCP/IP port 6389 is open. TCP/IP port 6389 must be open to allow
the Unisphere Host Agent to function properly. If TCP/IP port 6389 is blocked, the
installation will fail because the client will not be able to communicate with the target
storage systems. The HostAgent.exe is located at C:\Program
Files\EMC\HostAgent\HostAgent.exe for 32 bit Windows versions and C:\Program Files
(x86)\EMC\HostAgent\HostAgent.exe for 64 bit Windows versions.
Also, the ESRS IP Client for VNX and the Unisphere Server for Windows cannot coexist on
the same server. A pre-installation check in the ESRS IP Client for VNX will prevent
installation of the ESRS IP Client for VNX on a system that already has the Unisphere
Server installed on it. If this is the case, uninstall the Unisphere Server before installing
the ESRS IP Client for VNX. The ESRS IP Client for VNX uses the presence of the registry
key HKEY_LOCAL_MACHINE\Software\EMC\Management Server to determine if the Unisphere
Server is installed. If that key is not removed by the Unisphere Server uninstaller, it can
prevent the ESRS IP Client for VNX from being installed. You can delete the key from the
registry by using regedit and then you should be able to install the ESRS IP Client for
VNX.
.5
Done
Figure 1
Introduction
.i Read the introduction information and help contained in this panel to gain an understanding of
what actions this installer will perform and click Next.
License Agreement panel opens
.j Select I accept the terms and click Next.
Your Role panel opens
.k Select An Authorized Service Provider and click Next.
Configure Internet Connection panel opens (Fig 2)
Figure 2
Figure 3
Environment Check
.m Workstation environment check runs and displays the output in the Results box. If any checks fail
they must first be resolved before you can continue with the install. When all requirement checks
are met you can continue, click Next.
The Connect to Powerlink panel opens (Fig 4)
Figure 4
Connect to Powerlink
Figure 5
.p By default Install and use ESRS, Use email as a back up., and Notify the customer.
options are selected.
Enter the customers notification Email Address and customers SMTP Server Name or IP
Address.
If either email options are not desired clear the option selection and click Next
Systems to Monitor panel opens (Fig 6)
Figure 6
Systems to Monitor
Figure 7
Add System
10
Figure 8
Connectivity Checks
11
Figure 9
It is important that all the connectivity checks complete successfully for remote access to function
properly. Failed checks do not prevent you from continuing with the install but should be addressed
and corected. Click Next
Portal System panel opens
.s Select a qualified storage system IP address from the Portal list box or retain the installer selected
system, click Next
Customer Contact Information panel opens (Fig 10)
12
Figure 10
Figure 11
13
Figure 12
Installation Summary
.u Verify the installation summary is correct for this installation and click Install
Multiple popups will open describing what actions are being performed
Installation Complete panel opens (Fig 13)
14
Figure 13
Install Complete
.v Verify the installation was successful, all software components where installed, all storage systems
where registered, and a test event was sent by all monitored storage systems. Click Done
.w The installer automatically launches Navisphere/Unisphere Manager. Login to
Navisphere/Unisphere Manger and verify the monitoring configuration.
The Navisphere/Unisphere monitoring GUI displays the following:
Centralized monitoring agent that is running on this workstation
The latest Global Call-Home templated applied to the central monitor
All storage systems defined in step q are monitored by the central monitor
The ACU configuration GUI displays the following:
All monitored storage systems are ACU enabled
.6
15
Done
Verify that a test call-home file was sent and received by EMC
.a Record the CLARiiON Management Station Device Client serial number
Start > Programs > ESRS > Configuration Tool
The ESRS-IP Configuration Tool panel opens (Figure 1)
16
Done
Figure 1
Figure 2
.d Enter the customers CSI party number and click Search (Fig 2)
Browse for Devices page opens
17
Figure 3
Figure 4
.e Find and click the customers CLARMGMTDC gateway device serial number link recorded in
step a.
CLARMGMTDC device home page opens (Fig 4)
Figure 5
.f Scroll down to the Uploaded Files panel and click View All
18
Figure 6
Figure 7
.g Select type ConnectEMC from the filter box and click Filter
Filtered by ConnectEMC uploaded files appear (Fig 7)
Figure 8
.h Verify that an uploaded ConnectEMC file for every storage system in this monitoring configuration
is listed on this page. Each uploaded ConnectEMC file displays a storage system serial number in
the file name to identify which storage system generated the file.
Note: A ConnectEMC call home file that is encrypted can not be opened from the ServiceLink portal.
.i Access CSI and verify that every monitored storage system in this configuration opened an auto
cancelled a test event Service Request.
.j If a test call home Service Request was not generated for each storage system in the monitoring
configuration, something is wrong with the ESRS IP Client installation and configuration.
Retrace the installation and configuration steps. Refer to the errsipclient_installer.log file to
assist with debugging problems with the installation.
If you cannot diagnose the problem, call the Customer Support Center for assistance.