FLOW/NO-FLOW MONITOR
INSTALLATIONANDTECHNICAL
MANUAL
6/12/2008
Maxi-Tronic, Inc.
3530 IRWIN SIMPSON ROAD
MASON, OH 45040
513.398.2500
800.659.8250 FAX: 513.398.2536
info@maxitronic.com
www.maxitronic.com
INTRODUCTION
TheFM200MultiApplicationFlowMonitorisveryeffectiveinmonitoringFLOW/
NOFLOWconditionsinpipes,chutes,spouting,etc.Itsuniqueultrasonicsensing
techniquemakestheFM200versatileforalmostanyFLOW/NOFLOWapplication.
ThefollowingaresomeoftheuniqueadvancedfeatureswhichallowtheFM200to
exceedindustrystandards.
FEATURES
Control Unit:
Monitorsmovementofmostanythingthroughapipeorchutesystem:Air,Liquids,
andSolids.
Sensesthepresenceorabsenceofmaterialsinairflow.
3Userselectablesensitivityranges:
Low
Medium
High
Calibrationfineadjustmenttosuittheapplication.
Monitorspipesystemsfrom1/2to60ormoreindiameter.
ControlOutputs:
RelayContacts:N.O.andN.C.ratedfor4Ampsat125/250VAC
TTL:05Volt
RS232:Low/HighOutput
Standalonecontrolsystem.
EasyInstallation.
NoMaintenance,nomovingparts.
L.E.D.Indicators:
Flow=Green,NoFlow=Red
Polycarbonatehousingwithconduitknockouts:5x7x3(125mmx175mmx75mm).
Flow Sensor:
Sensormountsexternallytotheoutsideofthepipeorchute.
MountingTabsforeasyinstallation.
Prewiredwith6of18AWG2conductorwire.
Lowvoltagesensor:5VDC,200MAmaximum.
3sensorstylesavailable.
FM 200 WIRING
TheFM200polycarbonatehousingcomeswith10PGconduitknockoutsinthehousing.
Highvoltageinputsabove24voltACshouldberoutedintothehousingthroughthe
(20mm)conduithole.
AllLowvoltagewiring(TTL,RS232,andsensor)shouldberoutedthroughthe
(14mm)conduithole.
Figure1
Figure2
A1Ampfuseislocatednexttothetransformer.
Figure3
SENSOR INPUTS
ConnectthelowvoltageFlowSensortotheSensorterminalsTB1(Fig.4):
PositiveRED(orWhite)sensorleadtothe+SensorterminalonTB1.
NegativeBLACKsensorleadtotheSensorterminalonTB1.
Thesensorwiringshouldbeaminimumof18AWGstranded2conductorcable.
Use18AWGforupto500(155m)and16GAupto2000(620m).
AllSensorwiresplicesmustbeSOLIDandWATERTIGHT.
Figure4
SENSOR INSTALLATION
TheFM200sensorisinveryruggedpolymerhousingwithasealed6cablepigtail(Fig.5):
Red/Black=Positive
White/Black=Negative(thisiselectricallyconnectedtothemetalsensorstud)
FM200 SENSOR
BLACK LEAD
RED/WHITE LEAD
Figure5
Refertothemountingholeslocatedateachendofsensorhousing(Fig.6).
Markmountingholes.
Drillpilotholesformountingscrews.
Screwsensorhousingtoflatsurface.
SIDE VIEW
FM200
SENSOR
FRONT VIEW
FM200
SENSOR
Figure6
Usingstrapping(suchasheavydutywireties)orclampstosecurethesensor
housingtothepipesothemetalsensingstudmakesfirmcontactwiththesurface.
FRONT VIEW
SIDE VIEW
FM200
SENSOR
NOTE: Straps or
Clamps are Used
for Mounting to
NON-FLAT
SURFACE.
FM200
SENSOR
Figure7
SENSOR LOCATION:
TOENHANCESENSITIVYINSTALLTHESENSORWHERETHEFLOWISMOST
TURBULENTORWHEREMATERIALIMPACTSTHEPIPEWALL.
Thefollowingpointsareideal(Fig8):
Pipeelbow
Pointwherematerialdropsintospoutingorchutes
Pointwherematerialdropsfromconveyorontothechute
INPUT
PUMP
Figure8
Lowleveloutputsavailable(Fig.9):
TTLlogicoutput
RS232logicoutput
Figure9
NOFLOW:RedLEDlightsand0VDCwillbepresentattheOUTterminal.
FLOW:GreenLEDwilllightand5VDCwillbepresentattheOUTterminal.
ToobtainanRS232logicaloutputdonotinstalljumperJ3asindicatedonFM200PCB.
NOFLOW:RedLEDlightsand0VDCwillbepresentattheOUTterminal.
FLOW:GreenLEDwilllightand5VDCmin.willbepresentattheOUTterminal.
NOTE:forproperlogicinputtoyourcomputer,consultyourcomputermanualforthe
correctpinconfigurationforRS232orTTL.
Relay(CTR1)ContactSpecifications:
4ampat125or250VAC
1/10HPat125or250VAC
3ampat30VDC
RelayOutputWiringonTB2(Fig.10):
COM=Commoncontact
N/C=Normallyclosed
N/O=Normallyopen
Figure10
MACHINE
SUPPLYING
MATERIAL
MOTOR
STARTER
START
SWITCH
TB2
117VAC
COM
N/O
117VAC
ALARM
N/C
AC COM
Figure11
STOP
START
STARTER
COIL
Figure12
Beforethecontrolledequipmentcanbestartedagain,thefollowingmustoccur:
MaterialisflowingpasttheFM200sensor.
TheGreenL.E.D.(FLOW)indicatorlightislitindicatingmaterialflow.
YoumayhavetoholddowntheStartbuttonuntilflowreachesthepresetrate.
AslongasmaterialcontinuestoflowpasttheFM200sensorabovethepresetrate,the
machineequipmentsupplyingmaterialwillcontinuetorun.Iftherateofmaterialflow
fallsbelowthepresetlevel,therelaywillopenandthemachineequipmentsupplyingthe
materialwillstop.
STOP
START
STARTER
COIL
TB2
COM
N/O
N/C
STARTER CIRCUIT WITH FM200 SHUTDOWN
Figure13
TheFM200has(3)threesensitivityrangeinputs.Selecttherangetogetthedesiredsensitivity
fortheapplication.
IMPORTANT:TOAVOIDELECTRICALSHOCKDISCONNECTACSUPPLY
VOLTAGEBEFOREADDINGORREMOVINGJUMPERSONTB1.
Minimum(low)range:
InstalljumperJ2(only)
Mediumrange:
Nojumpersinstalled(asshippedfromfactory)
Maximum(high)range:
InstalljumperJ1(only)
1. WithnomaterialflowingturntheSensitivityadjustmenttothefullclockwiseposition:
REDLEDwillturnon.
2. Startthematerialflowingatthefullrateyouwishtomonitor:
IftheGREENLEDturnson,thenturntheSensitivityadjustmentfullycounter
clockwise.
IfthatcausestheREDLEDtoturnonthenthecorrectsensitivityrangeis
selected.
IfthatdoesnotcausetheREDLEDtoturnon(Greenstayslit)theninstall
jumperJ2(Minimumrange)andrepeatthetestbeginningwithstep#1.
IftheGREENLEDdoesnotturnonwhenmaterialisflowing,theninstalljumper
J2(Maximumrange)andrepeatthetestbeginningwithstep#1.
Figure14
10
FM200 CALIBRATION:
Performsensitivityrangejumperselectionbeforeperformingcalibration.
1. Stopallmaterialfromflowing.
2. TurntheSensitivityadjustmentfullyclockwise.
3. Startmaterialflowingattheminimumfullflowratedesired
GREENL.E.D.willlighttoindicateFLOW.
4. TurntheSensitivityadjustmentcounterclockwiseuntiltheREDLEDlightstoindicate
NOFLOW.
5. TurntheSensitivityadjustmentclockwiseveryslowlyuntilGREENLEDlightsto
indicateFLOW.
6. Calibrationisnowcomplete.