Professional Documents
Culture Documents
Toray Engineering | Measurement & Control Technologies | Principle of Measurement of Zirconia Oxygen Analyzers
Fig. 1
In Fig. 1, a gas with fixed oxygen concentration (norm ally air) is placed on the A side as the reference
gas (PR), while the gas whose oxygen concentration is to be m easured (PM) is placed on the B side.
When this is done, ion conduction occurs from the side with higher oxygen concentration to the side
with lower oxygen concentration.
The electrochem ical reaction on both sides follows Form ula 1, and the EMF accom panying ion
conduction is given by the Nernst equation (Form ula 2).
Formula 1
Formula 2
Since the EMF varies depending on the tem perature of the zirconia sensor and the oxygen
http://www.toray-eng.com/measuring/tec/zirconia.html
1/2
12/16/2016
Toray Engineering | Measurement & Control Technologies | Principle of Measurement of Zirconia Oxygen Analyzers
concentration of the reference gas (PR), in the actual device, the zirconia sensor is placed in a
constant tem perature oven, and air is generally used as the reference gas.
With Toray's zirconia sensor, a stable EMF conform ing to the Nernst equation is obtained from about
500C, as shown in the Table.
http://www.toray-eng.com/measuring/tec/zirconia.html
2/2