Technical specification

FLUXUS F608**-F2

Portable ultrasonic flow measurement of liquids in hazardous areas


Portable instrument for non-invasive, quick ultrasonic flow measurement with clamp-on technology for all
types of piping

Features
Precise bi-directional and highly dynamic flow measurement with the non-invasive clamp-on technology
High precision at fast and slow flow rates, high temperature and zero point stability
Portable, easy-to-use flow transmitter with 2 flow channels, multiple inputs, an integrated data logger with a
serial interface
Extremely resistant carbon fiber housing
Covered by FM Class I Div. 2 certification
Compact and very lightweight, allowing the measuring
system to be easily carried as personal luggage, e.g.
for offshore visits
Water tight; resistant against oil, many liquids and dirt
Li-Ion battery provides up to 14 hours of measurement
operation
Automatic loading of calibration data and transducer
detection for a fast and easy set-up (less than 5 min),
providing precise and long-term stable results
User-friendly design
Transducers available for a wide range of inner pipe diameters and fluid temperatures
( -274 to +1112 F)
Rugged transducers (FM Class I Div. 2, resistant to
rough environments and humidity)
Robust, water-tight (NEMA 4) transport case with comprehensive accessories
HybridTrek automatically switches between transit time
and NoiseTrek mode of measurement when high particulate flows are encountered
QuickFix for fast mounting of the flow transmitter in difficult conditions
Measurement is unaffected by fluid density, viscosity
and solid content (max. 10 % of volume)

FLU
XUS

F 608

FLUXUS F608 supported by handle

Measurement with transducers mounted with the portable


Variofix VP

FLUXUS F608

Applications
Designed for the following industries:
Upstream (on- and offshore)
Midstream and downstream (pipelines and refineries)
Chemical industry
Energy sector (e.g. HVAC, geothermal, power plants)

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

Measurement with the flow transmitter fixed to the pipe with


the QuickFix pipe mounting fixture

FLUXUS F608**-F2

Technical specification

Table of contents

Function ........................................................................................................................................................... 3
Measurement principle...................................................................................................................................... 3
Calculation of volumetric flow rate .................................................................................................................... 3
Number of sound paths..................................................................................................................................... 4
Typical measurement setup .............................................................................................................................. 4
Flow transmitter .............................................................................................................................................. 5
Technical data................................................................................................................................................... 5
Dimensions ....................................................................................................................................................... 6
Standard scope of supply.................................................................................................................................. 7
Adapters (optional)............................................................................................................................................ 7
Transducers..................................................................................................................................................... 8
Transducer selection......................................................................................................................................... 8
Transducer order code...................................................................................................................................... 9
Technical data................................................................................................................................................. 10
Transducer mounting fixture ....................................................................................................................... 13
Coupling materials for transducers ............................................................................................................ 15
Connection systems ..................................................................................................................................... 16
Transducer cable ............................................................................................................................................ 16
Clamp-on temperature probe (optional) ..................................................................................................... 17
Wall thickness measurement (optional) ..................................................................................................... 18

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

Technical specification

FLUXUS F608**-F2

Function
Measurement principle
Transit time difference principle
In order to measure the flow of a fluid in a pipe, ultrasonic signals are used, employing the transit time difference principle. Ultrasonic signals are emitted by a transducer installed on the pipe and received by a second
transducer. These signals are emitted alternately in the flow direction and against it.
As the fluid in which the signals propagate is flowing, the transit time of the ultrasonic signals in the flow direction is shorter than against the flow direction.
The transit time difference, t, is measured and allows the flowmeter to determine the average flow velocity
along the propagation path of the ultrasonic signals. A flow profile correction is then performed in order to obtain the area averaged flow velocity, which is proportional to the volumetric flow rate.
Two integrated microprocessors control the entire measuring process. This allows the flowmeter to remove
disturbance signals, and to check each received ultrasonic wave for its validity which reduces noise.
HybridTrek
If the gaseous or solid content in the fluid increases occasionally during measurement, a measurement with
the transit time difference principle is no longer possible. NoiseTrek mode will then be selected by the flowmeter. This measurement method allows the flowmeter to achieve a stable measurement even with high gaseous or solid content.
The transmitter can switch automatically between transit time and NoiseTrek mode without any changes to
the measurement setup.

t
Path of the ultrasonic signal

t
1

t
2

Transit time difference t

Calculation of volumetric flow rate

V = kRe . A . ka . t/(2 . tfl)


where

V
kRe
A
ka
t
tfl

=
=
=
=
=
=

volumetric flow rate


fluid mechanics calibration factor
cross-sectional pipe area
acoustical calibration factor
transit time difference
transit time in the fluid

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

FLUXUS F608**-F2

Technical specification

Number of sound paths


The number of sound paths is the number of transits of the ultrasonic signal through the fluid in the pipe. Depending on the number of sound paths, the following methods of installation exist:
reflect arrangement
The number of sound paths is even. Both of the transducers are mounted on the same side of the pipe.
Correct positioning of the transducers is easier.
diagonal arrangement
The number of sound paths is odd. Both of the transducers are mounted on opposite sides of the pipe.
direct mode
Diagonal arrangement with 1 sound path. This should be used in the case of a high signal attenuation by
the fluid, pipe or coatings.
The preferred method of installation depends on the application. While increasing the number of sound paths
increases the accuracy of the measurement, signal attenuation increases as well. The optimum number of
sound paths for the parameters of the application will be determined automatically by the transmitter.
As the transducers can be mounted with the transducer mounting fixture in reflect arrangement or diagonal arrangement, the number of sound paths can be adjusted optimally for the application.
.

a = transducer distance

Reflect arrangement, number of sound paths: 2


a

Diagonal arrangement, number of sound paths: 3


a>0

a<0

Direct mode, number of sound paths: 1

Direct mode, number of sound paths: 1,


negative transducer distance

Typical measurement setup


transducers

U L T R A S O N IC

F L O W M E T E R

C H A N N E L A

C H A N N E L B
Q

N E X T

O N

7
4

Q -

E N T E R

M U X

1
0

5
Q

6
2

O F F

L IG H T

3 x O F F

B R K

D IS P

B A T T E R Y

Q +

E N T E R

D IS P

M O D E

O N

transmitter
temperature inputs
Example of a reflect arrangement
4

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

Technical specification

FLUXUS F608**-F2

Flow transmitter
Technical data
FLUXUS
design

measurement
measurement principle
flow velocity
repeatability
fluid
temperature compensation
accuracy1
with standard calibration
with advanced calibration
(optional)
with field calibration2
flow transmitter
power supply
integrated battery
power consumption
number of flow measuring
channels
damping
measuring cycle (1 channel)
response time
housing material
degree of protection
dimensions
weight
fixation
ambient temperature
display
menu language
explosion protection
marking
F
M

F608**-F2
portable, FM class I Div. 2

transit time difference correlation principle,


automatic NoiseTrek selection for measurements with high gaseous or solid content
0.03 to 82 ft/s
0.15 % of reading 0.03 ft/s
all acoustically conductive liquids with < 10 % gaseous or solid content in volume (transit time difference
principle)
corresponding to the recommendations in ANSI/ASME MFC-5.1-2011
1.6 % of reading 0.03 ft/s
1.2 % of reading 0.03 ft/s
0.5 % of reading 0.03 ft/s
100 to 230 V/50 to 60 Hz (power supply unit, outside of explosive atmosphere)
10.5 to 15 V DC (socket at transmitter, with power adapter (optional))
integrated battery
Li-Ion, 7.2 V/4.5 Ah
operating time (without inputs and backlight): > 14 h
<6W
2
0 to 100 s, adjustable
100 to 1000 Hz
1 s (1 channel), option: 70 ms
PA, TPS, PC, Polyester, stainless steel
NEMA 4
see dimensional drawing
4.9 lb
QuickFix pipe mounting fixture
14 to 140 F
2 x 16 characters, dot matrix, backlight
English, German, French, Dutch, Spanish
NI/Cl. I /Div. 2/
GP. A,B,C,D /
T5 Ta = 60 C

for transit time difference principle, reference conditions and v > 0.49 ft/s

reference uncertainty < 0.2 %

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

FLUXUS F608**-F2

FLUXUS
measuring functions
physical quantities
totalizer
calculation functions
diagnostic functions
data logger
loggable values
capacity
communication
interface
serial data kit
software (all Windows
versions)

Technical specification

F608**-F2
volumetric flow rate, mass flow rate, flow velocity,
heat flow (if temperature inputs are installed)
volume, mass, optional: heat quantity
average, difference, sum
sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit times
all physical quantities, totalized values and diagnostic values
> 100 000 measured values
RS232/USB
- FluxData: download of measurement data, graphical presentation,
conversion to other formats (e.g. for Excel)
- FluxDiag (optional): online diagnostics and report generation

able
adapter
transport case
dimensions
inputs
number
accessories
type
connection
range
resolution
accuracy

- FluxSubstanceLoader: upload of fluid data sets


RS232
RS232 - USB
19.7 x 15.7 x 7.5 in
The inputs are galvanically isolated from the transmitter.
max. 4
input adapter (if number of inputs > 2)
temperature input
Pt100/Pt1000
4-wire
-238 to +1040 F
0.01 K
0.01 % of reading 0.03 K

2.32

Dimensions

C H A N N E L B
Q

N E X T

O N

Q -

E N T E R

M U X

5
Q

Q +

E N T E R

D IS P

O F F

L IG H T

3 x O F F

B R K

D IS P

8.39

U L T R A S O N IC F L O W M E T E R

C H A N N E L A

B A T T E R Y

M O D E

O N

8.9
in inch

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

Technical specification

FLUXUS F608**-F2

Standard scope of supply


F608 Standard
flow measurement on liquids
2 independent measuring channels

application

F608 Energy

temperature-compensated calculation of mass


flow rate
integrated heat flow computer for monitoring of
energy flows
including energy calculator for BTU and heat
measurements
inputs
temperature input
accessories
transport case
power supply unit, mains cable
battery
QuickFix pipe mounting fixture
for transmitter
serial data kit
measuring tape
user manual,
safety instructions,
Quick start guide
connector board at the upper
side of the transmitter

x
x
x
x

x
x
x
x

x
x
x

x
x
x

C H

C H
C O M M

D C

C H
B

T 1 /T 3

C O M M

- IN

In p u t

C H

T 2 /T 4

D C

- IN

Adapters (optional)

input adapter

transducers
measuring channel A

C H

T 1 /T 3

C O M M

RS232

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

In p u t

C H

T 2 /T 4

D C

inputs

transducers
measuring channel B

- IN

power supply unit/battery charging unit

FLUXUS F608**-F2

Technical specification

Transducers
Transducer selection
transducer order code
FSG

15.7

FSK

3.9

FSM

FSP

0.98

FSQ

0.39

FSS

0.24
0.2

19.7

7.9

78.7

3.9
2
5.9

0.39

255.9

94.5

39.4 47.2
15.7

0.98

157.5

18.9

9.4

2.8

0.4

20

40

200
inner pipe diameter [in]

transducer order code


0.43

FSG
FSK

0.2

FSM
FSP
FSQ
FSS

0.08
0.04
0.02
0.01
0.2

recommended

0.4

0.6

0.8

1.2
pipe wall thickness [in]

possible

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

Technical specification

FLUXUS F608**-F2

Transducer order code


1, 2

5, 6

7, 8

9 to 11

12, 13

no. of character

option

extension cable

connection
system

explosion
protection

ambient
temperature

transducer
frequency

transducer

description

FS

set of ultrasonic flow transducers for liquids measurement, shear


wave
G

0.2 MHz

0.5 MHz

1 MHz

2 MHz

4 MHz

8 MHz
N

normal temperature range

extended temperature range (FSM, FSP, FSQ)


F2

FM Class I Div. 2
NL

with Lemo connector


XXX

cable length in m, for max. length of extension cable see page 16


LC

long transducer cable

example
FS

F2

NL

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

010

shear wave transducer 1 MHz, normal temperature range,


FM Class I Div. 2, connection system NL with Lemo connector
and extension cable 10 m
/

FLUXUS F608**-F2

Technical specification

Technical data
Shear wave transducers

in
in
in
in

15.7
19.7
157.5
255.9

15.7
19.7
157.5
255.9

3.9
7.9
78.7
94.5

3.9
7.9
78.7
94.5

in

0.43

0.43

0.2

0.2

PEEK with stainless steel


cap 304
PEEK
NEMA 6

PEEK with stainless steel


cap 304
PEEK
NEMA 6

PEEK with stainless steel


cap 304
PEEK
NEMA 6

PEEK with stainless steel


cap 304
PEEK
NEMA 6

ft

1699
16

1699
29

1699
16

1699
29

in
in
in

5.1
2.01
2.64

5.1
2.01
2.64

4.98
2.01
2.66

4.98
2.01
2.66

ambient temperature
min.
F
-40
max.
F
+266
temperature compenx
sation
explosion protection
order code
FSG-NF2NL
explosion protection temperature
min.
F
-40
F max.
F
+257
M marking
NI/Cl. I,II,III/Div. 2 /
GP A,B,C,D,E,F,G/
Temp. Codes dwg 3860
type of protection
non incendive
remark
on request

10

CLK1N51
FSK-NF2NL/LC
0.5

CDK1N51
FSK-NF2NL
0.5

contact surface
degree of protection
transducer cable
type
length
dimensions
length l
width b
height h
dimensional drawing

CLG1N51
FSG-NF2NL/LC
0.2

inner pipe diameter d


min. extended
min. recommended
max. recommended
max. extended
pipe wall thickness
min.
material
housing

CDG1N51
FSG-NF2NL
MHz 0.2

technical type
order code
transducer frequency

-40
+266
x

-40
+266
x

-40
+266
x

FSG-NF2NL/LC

FSK-NF2NL

FSK-NF2NL/LC

-40
+257

-40
+257

-40
+257

NI/Cl. I,II,III/Div. 2 /
NI/Cl. I,II,III/Div. 2 /
NI/Cl. I,II,III/Div. 2 /
GP A,B,C,D,E,F,G/
GP A,B,C,D,E,F,G/
GP A,B,C,D,E,F,G/
Temp. Codes dwg 3860 Temp. Codes dwg 3860 Temp. Codes dwg 3860
non incendive
non incendive
non incendive
on request

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

Technical specification

CDP1N51
FSP-NF2NL
2

CDQ1N51
FSQ-NF2NL
4

CDS1N51
FSS-NF2NL
8

in
in
in
in

2
3.9
78.7
133.9

0.98
2
7.9
23.6

0.39
0.98
5.9
15.7

0.24
0.39
2.8
2.8

in

0.1

0.06

0.04

stainless steel 304


PEEK
NEMA 6

stainless steel 304


PEEK
NEMA 6

stainless steel 304


PEEK
NEMA 6

stainless steel 304


PEI
NEMA 6

ft

1699
13

1699
13

1699
9

1699
6

in
in
in

2.36
1.18
1.32

2.36
1.18
1.32

1.67
0.71
0.85

0.98
0.51
0.67

CDM1N51
FSM-NF2NL
MHz 1

ambient temperature
min.
F
-40
max.
F
+266
temperature compenx
sation
explosion protection
order code
FSM-NF2NL
explosion protection temperature
min.
F
-40
F max.
F
+257
M marking
NI/Cl. I,II,III/Div. 2 /
GP A,B,C,D,E,F,G/
Temp. Codes dwg 3860
type of protection
non incendive

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

technical type
order code
transducer frequency
inner pipe diameter d
min. extended
min. recommended
max. recommended
max. extended
pipe wall thickness
min.
material
housing
contact surface
degree of protection
transducer cable
type
length
dimensions
length l
width b
height h
dimensional drawing

FLUXUS F608**-F2

-40
+266
x

-40
+266
x

-22
+266
-

FSP-NF2NL

FSQ-NF2NL

FSS-NF2NL

-40
+257

-40
+257

-40
+257

NI/Cl. I,II,III/Div. 2 /
NI/Cl. I,II,III/Div. 2 /
NI/Cl. I,II,III/Div. 2 /
GP A,B,C,D,E,F,G/
GP A,B,C,D,E,F,G/
GP A,B,C,D,E,F,G/
Temp. Codes dwg 3860 Temp. Codes dwg 3860 Temp. Codes dwg 3860
non incendive
non incendive
non incendive

11

FLUXUS F608**-F2

Technical specification

Shear wave transducers (extended temperature range)


CDP1E51
FSP-EF2NL
2

CDQ1E51
FSQ-EF2NL
4

in
in
in
in

2
3.9
39.4
47.2

0.98
2
15.7
18.9

0.39
0.98
5.9
9.4

in

0.08

0.04

0.02

stainless steel 304


Sintimid
NEMA 4

stainless steel 304


Sintimid
NEMA 4

stainless steel 304


Sintimid
NEMA 4

ft

1699
13

1699
13

1699
9

in
in
in

2.36
1.18
1.32

2.36
1.18
1.32

1.67
0.71
0.85

ambient temperature
min.
F
-22
max.
F
+392
temperature compenx
sation
explosion protection
order code
FSM-EF2NL
explosion protection temperature
min.
F
-40
F max.
F
+374
M marking
NI/Cl. I,II,III/Div. 2 /
GP A,B,C,D,E,F,G/
Temp. Codes dwg 3860
type of protection
non incendive

12

inner pipe diameter d


min. extended
min. recommended
max. recommended
max. extended
pipe wall thickness
min.
material
housing
contact surface
degree of protection
transducer cable
type
length
dimensions
length l
width b
height h
dimensional drawing

CDM1E51
FSM-EF2NL
MHz 1

technical type
order code
transducer frequency

-22
+392
x

-22
+392
x

FSP-EF2NL

FSQ-EF2NL

-40
+374

-40
+374

NI/Cl. I,II,III/Div. 2 /
NI/Cl. I,II,III/Div. 2 /
GP A,B,C,D,E,F,G/
GP A,B,C,D,E,F,G/
Temp. Codes dwg 3860 Temp. Codes dwg 3860
non incendive
non incendive

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

Technical specification

FLUXUS F608**-F2

Transducer mounting fixture


6

fixation

size

measurement
arrangement

transducer

transducer
mounting fixture

1, 2

7 to 9

no. of character

outer pipe diameter

Order code
description

FS

mounting frames (transducers with transducer frequency S)

LM

ladder chain mounting accessory

VP

portable Variofix

WL

transducer box for WaveInjector


A

all transducers

transducers with transducer frequency G, K

transducers with transducer frequency M, P

transducers with transducer frequency Q

transducers with transducer frequency S


D

reflect arrangement or diagonal arrangement/direct mode

reflect arrangement
S

small

medium
C

chains

without fixation
L08

0.5 to 8 in

L22

0.5 to 22 in

010

0.39 to 3.9 in

025

0.39 to 9.8 in

055

0.39 to 21.7 in

055

portable Variofix and chains

example
VP

mounting frames FS and chains

transducer frequency: S
material: stainless steel 304,
301, 303
dimensions:
8.27 x 1.26 x 1.73 in
chain length: 1 ft
outer pipe diameter:
max. 5.9 in

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

13

FLUXUS F608**-F2

Technical specification

ladder chain mounting accessory LM

transducer frequency: M, P, Q
chain length: 30/78 in
outer pipe diameter:
max. 24 in

portable Variofix VP and chains (optional)

material: stainless steel 304,


301, 303
dimensions:
16.3 x 3.7 x 2.99 in
chain length: 6 ft

transducer box WL for WaveInjector


transducer box and
adapter

see Technical specification


TSWaveInjectorVx-x

14

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

Technical specification

FLUXUS F608**-F2

Coupling materials for transducers


normal temperature range
(4th character of transducer order
code = N)
< 212 F
< 338 F
coupling compound coupling compound
type N
type E

extended temperature range


(4th character of transducer order
code = E)
< 302 F
< 392 F
coupling compound coupling compound
type E
type E or H

WaveInjector WI-400
< 536 F
coupling pad type A
and coupling pad
type VT

536 to 752 F
coupling pad type B
and coupling pad
type VT

Technical data
type

ambient temperature
F
-22 to +266

material
mineral grease paste

-22 to +392

silicone paste

coupling compound
type N
coupling compound
type E
coupling compound
type H
coupling pad type A

-22 to +482

fluoropolymer paste

max. 536

lead

coupling pad type B

> 536 to 752

silver

coupling pad type VT

14 to +392

fluoroelastomer

remark

for transducers with transducer frequency F


for transducers with transducer frequency G, H, K
for shear wave transducers with transducer frequency
M, P
for shear wave transducers IP68 and Lamb wave transducers with transducer frequency M, P
for transducers with transducer frequency Q

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

15

FLUXUS F608**-F2

Technical specification

Connection systems
connection system NL
x

transmitter

transducer frequency
(3d character of transducer
order code)
N
L

cable length ft
cable length (option LC) ft

F, G, H, K
x
6
6

y
9
22

M, P
l
32
32

x
6
22

Q
y
6
6

l
32
32

x
6
26

S
y
3
3

l
32
32

x
3
-

y
3
-

l
32
-

x, y = transducer cable length


l = max. length of extension cable

Transducer cable
Technical data
type
standard length
max. length
ambient temperature
cable jacket
material
outer diameter
thickness
color
shield
sheath
material
outer diameter

16

ft

transducer cable
1699
see table above

ft
F

-67 to +392

extension cable
1750
16
32
32
< 144

PTFE
0.11
0.01
brown
x

PE
0.24
0.02
black
x

stainless steel 304


0.31

stainless steel 304


0.35

in
in

in

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

Technical specification

FLUXUS F608**-F2

Clamp-on temperature probe (optional)


Technical data
technical type
design
type
connection
measuring range

PT13N

Pt1000
4-wire
-40 to +392
(0.27 F + 2 . 10-3 . (|T [F]| - 32 F))
class A
0.03 F (at 50 F)

accuracy T
accuracy T
(2x Pt matched according
to EN 1434-1)
response time
housing
degree of protection
weight
fixation
accessories
thermal conductivity
paste 392 F
thermal conductivity foil
482 F
plastic protection plate,
insulation foam
dimensions
length l
width b
height h

PT13F
short response time
Pt1000
4-wire
-58 to +482
(0.27 F + 2 . 10-3 . (|T [F]| - 32 F))
class A
0.1 K (3 K < T < 6 K), more corresponding to
EN 1434-1

0.437
clamp-on

8
PEEK, stainless steel 304, copper
NEMA 4
0.7
clamp-on

0.59
0.49
0.79

0.55
1.18
1.06

360 brass alloy


lb

in
in
in

PT13N
l

extension cable

PT13F
h

Connector
pin
1
2
3, 4, 5
6
7
8

cable of temperature probe


PT13F
white/blue
red/blue
not connected
red
white
not connected

cable of temperature probe


PT13N
white
red
red
white

extension cable
white
black

2
4

green
red

5
8

1
6

3
7

Cable
temperature probe
type
standard length
max. length
cable jacket

ft
ft

cable of temperature probe


PT13F
PT13N
4 x 0.25 mm black
4 x 24 AWG
9
20
PTFE
PTFE

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

extension cable
4 x 18 AWG
328
LS PVC

17

FLUXUS F608**-F2

Technical specification

Wall thickness measurement (optional)


The pipe wall thickness is an important pipe parameter which has to be determined exactly for a good measurement. However, the pipe wall thickness often is unknown.
The wall thickness probe can be connected to the transmitter instead of the flow transducers and the wall
thickness measurement mode is activated automatically.
Acoustic coupling compound is applied to the wall thickness probe which then is placed firmly on the pipe.
The wall thickness is displayed and can be stored directly in the transmitter.

Technical data
technical type
measuring range1
resolution
accuracy
fluid temperature
explosion protection
able
type
length
1

DWR1NZ7
in
in
F

ft

0.04 to 9.8
0.0004
1 % 0.004 in
-4 to +392,
short-time peak max. 932
2616
4

The measuring range depends on the attenuation of the ultrasonic signal in the
pipe. For strongly attenuating plastics (e.g. PFA, PTFE, PP) the measuring range is smaller.

Cable
type
ambient temperature
cable jacket
material
outer diameter
color
shield

in

2616
<392
FEP
0.2
black
x

DWR1NZ7

18

TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01

FLUXUS F608**-F2

FLEXIM AMERICAS Corporation


Edgewood, NY 11717
USA

20

Technical specification

internet: www.flexim.com
e-mail: usinfo@flexim.com
1-888-852-7473

Tel.: (631) 492-2300

Subject to change without notification. Errors excepted.

Fax: (631) 492-2117

FLUXUS is a registered trademark of FLEXIM GmbH.


TSFLUXUS_F608xx-F2V2-1US_Lus, 2016-08-01