Foreword
Thank you for using MV-600IBinfusion pump!
Please read this manual carefully before using it so as to operate the pump correctly and safely.
Contents
Brief Introduction....3
Intended Use..3
Safety Precaution......3
Specification...4
System Structure...5
Packing ...5
Main Key Features and Functions......6
Basic Operation Steps..7
Bolus Function..10
Fat Emulsion Infusion .10
Method to Change Different Brand of Infusion Set.10
Load Infusion Pipe Parameters 12
Battery Charge 13
Cleaning, Maintenance, Transport and Storage 13
Alarms Indicator and Remove. .14
1. Brief Introduction
1.1 Product Introduction
MV-600IBInfusion Pump is a combination of microelectronic technology and modern nursing,
an important outcome of microcomputer applied to clinical nursing. It can control the speed
and general quantity of infusion precisely and continuously over long hours on a large scale,
which can completely meet the various requirements of modern clinical treatment on different
occasions and is now widely used in hospitals.
MV-600IBInfusion Pump is an economical product of intellectualization, collectivization and
high technology oriented to oversea customers. It can be used in all clinics or wards, thus
promoting the previously single pump intravenous infusion control to mass control. The
patients' infusion situation and alarm information (e.g.: infusion completion, pipe occlusion)
can all be seen at the Nurse's Station. This greatly improves the medical nursing quality,
reduces the nurse's workload, and ensures the patient's nursing safety.
1.2 Intended Use
It is used in hospitals where patients need to be given steady and continuous intravenous
infusion or precise medication.
1.3 Safety Precaution
1. The user must be thoroughly familiar with the infusion pump describe in this manual prior to
use.
2. Attention for setting up the pump:
Do not exceed the time the doctor diagnoses for the treatment.
Observe and confirm the condition of the patient and the pump operation are both good.
Once any abnormal situation is found on the patient or machine , first assure the patient
is safe , and then turn off the machine and take proper action.
5. Attention after finishing using the pump
3
Turn off the power supply according to the requirement, and pull out the AC Power cord.
In order to use the pump conveniently next time, please clean it and then keep it in
storage.
1.4 Product Functions
MV-600IB
Bolus
600ml/h
1-2000ml/h
Volume limit
1-9999 ml
Accuracy
Power supply
Battery
Battery recharge
Displayed information
Alarm function
Max. weight
2.5 kg
Classification
Type IIB.
ABS plastic
Operating condition
condition
Work mode
2. List of Accessories:
infusion pump
AC power cord
instruction manual
certificate of approval
maintenance card
1 unit
1 set
1 set
1 set
1 set
Notice: If any accessory is missing in the box, please contact us or our agents in your local
place.
13
1
7
8
11
12
3
4
9
10
Serial
No.
Description
Functions
100 key
100 key
10 key
10 key
key
key
"STOP" key
"CLEAR" key
"SET" key
10
"SELECT" key
11
START/BOLUS key
2.BOLUS: Press this key in the infusion without release, the pump
starts to fast infusion at the maximum rate; release the key, the
pump comes to infusion at the initial rate.
Turn on/off the pumpPress the key for about 3 to 5 seconds until
12
"POWER" key
13
LCD
Diagram 1
Diagram 2
Diagram 3
1. Rotate the knob of the pole clamp anticlockwisely, and leave enough room to put IV
stand in as Diagram 1 shows.
2. Fix the infusion pipe on the IV stand, and rotate the knob of pole clamp clockwisely to
get the pump fixed on the IV stand.
Diagram4
Diagram 5
Pull
straight
the pipe
Diagram 6
Diagram 8
Wrong
Diagram 7
No..
Diagram 9
Diagram 10
Diagram 11
Diagram 12
Diagram 13
Diagram 14
Diagram 15
displays "OVER" and sends an audible and visible alarm to remind the user that the infusion
is finished. Press "STOP" key to stop infusion as Diagram 17 shows.
Diagram 16
Diagram 17
Diagram 18
During infusion, if the on-going rate is lower than 600ml/h and yet you need a moments
fast-rate infusion, you may press the "BOLUS" key by keeping nger on i t for a wh i le. The
pump shall infuse at 600ml/h as long as you keep your nger on BOL US key. A er remov i ng
your nger from the key, i t wi ll run at the or i gi nal i nf us i on rat e that you have set . If the
on-going rate is higher than 600ml/h, there is no Bolus func on.
Indicator
light is not
on
a er
installa on
the pipe
Diagram 19
Diagram 20
Diagram 21
1. Install a new brand of infusion pipe filled with liquid. Press "UP 100" and "DOWN 100" keys
10
together as Diagram 22 shows. The LCD displays 3 values as Diagram 23 shows. Record the
upper one: pressure value.
2. Press "SET" and "STOP" keys together as Diagram 24 shows. The LCD will display another 3
values as Diagram 25 shows, and the upper one is the occlusion value for occlusion alarm.
When the occlusion pressure is about 50 to 150 higher than pressure value, the occlusion alarm
shall sound.
For instance, the pressure value we recorded is 50, so we should set the occlusion value within
the range of 100 and 200.The higher the value, the less sensitive the pressure sensor shall be
and thus the slower the occlusion alarm shall sound. The lower the value, the more sensitive the
pressure sensor shall be and thus the sooner the occlusion alarm shall sound. Adjust a suitable
occlusion value by pressing number keys as Diagrams 26 shows. At last press "SET" key to
save the setting.
Diagram 22
Diagram 25
Diagram 23
Diagram 26
Diagram 24
Diagram 27
Example 2The actual liquid flowed into the measuring cup is 8ml, and the initial accuracy value
is 50, we should adjust the accuracy to 48.
3. Repeat step 1 and step 2 until the accuracy value is accurate. (That means the actual liquid
volume flowed into the measuring cup is the same as volume limit.) The accuracy of infusion
pump is 5.
2.5 Load the Infusion Pipe Parameter
1. MV-600IBinfusion pump can save three kinds of infusion pipe parameters. It displays A, B or
C on the top right corner of the LCD as Diagram 28 shows, and the "A","B" or "C" stands for the
type of infusion pipe in use.
2. Press "STOP" and "SET" key together as Diagram 29 shows. The LCD shows the infusion
pipe's parameters in use as Diagram 30 shows.
If you want to load the parameter of Type A infusion pipe, press "CLEAR" and "DOWN 100"
keys at the same time as Diagram 31 shows, and the LCD will display A on the top right corner.
If you want to load the parameters of type B infusion pipe, press " CLEAR" and " DOWN 10"
keys at the same time as Diagram 32 shows, and the LCD will display B on the top right corner.
If you want to load the parameters of type C infusion pipe, press " CLEAR" and " DOWN 1"
keys together as Diagram 33 shows, and the LCD will display C on the top right corner.
3. Load the type of infusion pipe you need, and then press "SET" key to save it as Diagram 34
shows.
Diagram28
Type A
Type B
Type C
Occlusion
Value
Accuracy
Diagram 29
Diagram 30
Diagram 31
12
Diagram 32
Diagram 33
Diagram 34
Diagram 35
Low ba er y capaci ty
Diagram 36
Diagram 37
ba er y
supply
symbol
Diagram38
13
3.4 Maintenance
The warranty year is 1 year beginning from the purchasing date. We can offer free repair
service within the warranty time on condition that the product is operated properly. The
following situation is not within the range of free maintenance and repair:
1. Malfunction caused by false use, repair or reconstruction by any unprofessional, disqualified
or untrained people.
2. Malfunction or damage caused during the transportation.
3. Malfunction or damage caused by fire, salt, poisonous gas, earthquake, hurricane, flood,
abnormal voltage and other nature factors.
Display
Reason
Solution
Press "STOP" key to stop the
Alarm for
Preset infusion
infusion
volume
completion
completion.
2.Infusion pipe
installation error
3.Malfunction
with the infrared
sensor
1. Infusion loop
occlusion
2.The occlusion
Occlusion
value is much
alarm
too sensitive.
3. Something
sensor.
2.Battery aging
or something
wrong with the
battery charge
circuit.
1. No AC power
Alarm for no
AC power
supply
infusion pump.
2. Something
power supply
circuit of infusion
repair.
pump.
Solution for No.1
Alarm for
installation
error
1.Air bubble in
2.Infusion pipe
installation error
in the pipe.
3. Something
sensor of
infusion pump.
Alarm for
Something
abnormal
control
motor.
15