Advance version
Developed to deliver outstanding conventional HD treatments, the Advance
version of the AK 96 machine with its selected features raises the quality and
level of treatments, operational efciency, control and ease of use. Selected
features include a complete integration with Gambro water treatment systems
and a heparin syringe pump.
The Advance version of the AK 96 machine combines technological
advancement with reliability, durability and user-friendliness.
Diascan
Controlled dialysis dose delivery
Battery back-up
For safe treatments in all
environments
HCEN4966_2_AK96_advance.indd 1
09-02-12 15.35.01
Technical specications
Disinfection log
Syringe pump
Alarm pole
Option
pH probe
Option
Operational efciency
Automatic self-test
Stand-by mode
Proling modules
Assisted priming
Battery back-up
Safety
The AK 96 machine complies with the following standards:
IEC 60601-1 General requirements for safety, Class I, type B
IEC 60601-2-16 Particular requirements for safety of hemodialysis,
hemodialtration and hemoltration equipment
IEC 60601-1-1 Safety requirements for medical electrical systems
IEC 60601-1-2 Electromagnetic compatibility, class B
IEC 60601-2-30 Particular requirements for the safety of automatic
cycling indirect blood pressure monitoring equipment
EN 1060-1 Non-invasive sphygmomanometers Part 1: General
requirements
EN 1060-3 Non-invasive sphygmomanometers Part 3: Supplementary
requirements for electromechanical blood pressure measuring systems
Gambro is a registered trademark of Gambro Lundia AB
AK 96 is a trademark of Gambro Lundia AB registered in the European
Community, China and in other countries and led in the United States
Diascan is a trademark of Gambro Lundia AB/Gambro Hospal
Switzerland ltd registered in the United States, China and in other
countries and led in the European Community
CleanCart is a trademark of Gambro Lundia AB/Gambro Hospal
Switzerland ltd, registered in China and in other countries and led in
the United States
U9000 is a trademark of Gambro Lundia AB, registered in the
European Community
Specications subject to change without notice. For further information
and operating instructions, please refer to the operator manual
Blood ow control
Flow rate, double needle: 0 and 20 to 500 ml/min
Flow rate, single needle: 0 and 20 to 500 ml/min, pressure-controlled
Blood circuit pressure supervision
Arterial pressure: -700 to +750 mmHg
Venous pressure*: -700 to +750 mmHg
Air detection
Method: Ultrasonic detector
Drip chamber size: 22 mm diameter
Heparin syringe pump
Flow rate: 0 to 10 ml/h
Programmable stop time, accumulated volume read-out
Water supply
Inlet pressure: 0.12 to 0.6 MPa (1.2 to 6 bar)
Inlet water temperature during treatment: 5 to 30 C
Inlet water quality: Fluid must comply with appropriate regulations and as
minimum ISO 13959
Dialysis uid preparation and monitoring
Flow rate: 300-700 ml/min (by step of 20 ml/min)
Acetate range: Na+ 130 to 160 mmol/l
Bicarbonate range: Na+ 130 to 160 mmol/l,
HCO3- 20 to 40 mmol/l
pH measurement: Range 1.0 to 9.9
Proling (Na+, HCO3-, UF)
Concentrate standby mode
Ultraltration control
UF volume: Adjustable, 0 to 10.00 l
Accuracy: 50 ml/h or 1%
UF rate: 0 to 4 l/h
Blood leakage detection
Method: Infrared light
Disinfection and cleaning
Automated disinfection process with Gambro water treatment systems
Heat, liquid citric acid or CleanCart disinfectant
Chemical: Peracetic acid, hypochlorite or formaldehyde
Disinfection log
Power supply
Mains voltage: 115, 230V
Frequency: 50 to 60 Hz
Power consumption: Max 2025 W at 230 V, 1575 W at 115 V
Connection of external equipment
Connector: 25 pin D-sub with RS232C or RS-422/current loop
Dimensions and weight
Width: Machine 345 mm, stand 585 mm
Depth: Machine 600 mm, stand 620 mm
Height: 1305 mm
Weight: Approx. 60 kg (without options)
Operating environment
Ambient temperature: 18 to 35 C
Relative humidity: 15 to 85%
Air pressure: Up to approx. 2500 meters above sea level (70 to 106 kpa)
*An event of venous needle disconnection is not guaranteed to be detected by most dialysis machines.
International standards recommend additional venous access monitoring is used to safeguard patient safety.