You are on page 1of 7

Anaesthesia 2012, 67, 377383

doi:10.1111/j.1365-2044.2011.07018.x

Original Article
A comparison of the Nexfin and transcardiopulmonary thermodilution
to estimate cardiac output during coronary artery surgery
O. Broch,1 J. Renner,2 M. Gruenewald,1 P. Meybohm,2 J. Schottler,3 A. Caliebe,4 M. Steinfath,5
M. Malbrain6 and B. Bein7
1 Staff member, 2 Associate Professor, 6 Professor and Chair, 7 Professor, Department of Anaesthesiology and Intensive
Care Medicine, 3 Staff member, Department of Cardiothoracic and Vascular Surgery, 4 Staff member, Institute of Medical
Informatics and Statistics, Christian-Albrechts University, Kiel, Germany
5 Associate Professor of Internal Medicine, ICU and High Care Burn Unit, ZNA Campus Stuivenberg, Antwerp, Belgium

Summary
The newly introduced Nexn device allows analysis of the blood pressure trace produced by a non-invasive nger cuff.
We compared the cardiac output derived from the Nexn and PiCCO, using transcardiopulmonary thermodilution,
during cardiac surgery. Forty patients with preserved left ventricular function undergoing elective coronary artery bypass
graft surgery were studied after induction of general anaesthesia and until discharge to the intensive care unit. There was
a signicant correlation between Nexn and PiCCO before (r2 = 0.81, p < 0.001) and after (r2 = 0.56, p < 0.001)
cardiopulmonary bypass. BlandAltman analysis demonstrated the mean bias of Nexn to be )0.1 (95% limits of
agreement )0.6 to +0.5, percentage error 23%) and )0.1 ()0.8 to +0.6, 26%) l.min)1.m)2, before and after
cardiopulmonary bypass, respectively. After a passive leg-raise was performed, there was also good correlation between
the two methods, both before (r2 = 0.72, p < 0.001) and after (r2 = 0.76, p < 0.001) cardiopulmonary bypass. We
conclude that the Nexn is a reliable method of measuring cardiac output during and after cardiac surgery.
. ..............................................................................................................................................................

Correspondence to: Dr O. Broch


Email: broch@anaesthesie.uni-kiel.de
Accepted: 19 November 2011

Monitoring of haemodynamic variables, such as cardiac


output and left ventricular stroke volume, is of
increasing interest, as it allows treatment to be tailored
to the individual patient. Recent studies have shown
that optimisation of left ventricular stroke volume or
cardiac output is associated with benecial effects on
both morbidity and length of stay on the intensive care
unit (ICU) [13]. The gold standard is the pulmonary
artery catheter; however, this is associated with a
number of complications and clinical limitations [4
6]. Therefore, investigators have examined the utility of
less invasive techniques such as transcardiopulmonary

Anaesthesia 2012 The Association of Anaesthetists of Great Britain and Ireland

thermodilution or continuous arterial waveform analysis [79].


The Nexn (BMEYE, Amsterdam, The Netherlands) has recently been introduced into clinical
practice. It consists of a model-based method that
provides beat-to-beat measurement of cardiac output by
analysis of the non-invasive nger arterial blood
pressure trace, which is measured continuously by the
use of an inatable nger cuff. Stroke volume is
determined by dividing the pulsatile systolic area of
each beat by impedance, which is estimated by the
device based on patient characteristics.
377

Anaesthesia 2012, 67, 377383

We designed a study to compare the Nexn with


transcardiopulmonary thermodilution using the PiCCO
(Pulsion Medical Systems, Munich, Germany), looking
at the accuracy of cardiac output estimation and the
ability to track haemodynamic changes and trends.

Methods
After approval from our institutional ethics committee,
patients with preserved left ventricular function undergoing elective coronary artery bypass surgery were
approached and consent for participation in the study
was sought. Exclusion criteria included age < 18 years,
left ventricular ejection fraction 0.5, emergency
surgery, haemodynamic instability, intracardiac shunt,
severe aortic, tricuspid or mitral stenosis or insufciency, and mechanical circulatory support.
Patients were prescribed midazolam 0.1 mg.kg)1
orally, 30 min before induction of anaesthesia. Routine
monitoring was established including peripheral oxygen
saturation and heart rate (S 5 monitor; GE Healthcare,
Helsinki, Finland).
The Nexn monitoring system was set up by
entering patient-specic data followed by attaching the
pneumatic nger cuff to the middle phalanx of the
third nger (on the opposite hand to the arterial line),
as recommended by the manufacturer. Finger arterial
pressure measurement is based on the volume clamp
method [1012] in combination with Physiocal calibration [13]. Continuous nger pressure is reconstructed to brachial artery levels using a generalised
waveform lter and a level correction formula, correcting for differences in wave shape and pressure values
between these sites. Once the brachial artery waveform
has been developed, the pulsatile systolic area (determined by the time integral of the pressure curve above
the diastolic pressure and between the upstroke and the
dicrotic notch) is calculated for each beat. Stroke
volume is then calculated using the estimated arterial
impedance. To that end, a patient-specic threeelement Windkessel model is produced using patient
characteristics (age, sex, height and weight). Moreover,
for each beat, mean arterial pressure is used to correct
the elements that are pressure-dependent in a nonlinear way [14]. Dividing the pulsatile systolic area by
the impedance gives beat-to-beat stroke volume. Heart
rate was calculated from the pulse interval, followed by
378

Broch et al. | Nexfin vs PiCCO

cardiac output (stroke volume multiplied by heart rate);


variables were automatically indexed to body surface
area.
After administration of sufentanil (0.5 lg.kg)1),
propofol (1.5 mg.kg)1) and rocuronium (0.6 mg.kg)1),
the trachea was intubated. Anaesthesia was maintained
by administering a continuous infusion of both sufentanil (1 lg.kg)1.h)1) and propofol (3 mg.kg)1.h)1). The
patients lungs were ventilated with an oxygen air
mixture using a tidal volume of 8 ml.kg)1 ideal body
weight, and positive endexpiratory pressure was set at
5 cmH2O. Subsequently, a central venous catheter and a
transpulmonary thermodilution PiCCO catheter were
inserted into the right internal jugular vein and the
femoral artery, respectively. The thermodilution catheter
was connected to the PiCCO monitor (Software version
1.3.0.8).
After induction of anaesthesia, Nexn and PiCCO
measurements as well as other haemodynamic values
were recorded every 10 min, both before and after
cardiopulmonary bypass (CPB). These were only measured when the following conditions were met: stable
haemodynamics; an undamped arterial pressure trace;
and (with respect to the Nexn) a frequency of
physiological autocalibration > 30 heartbeats. Thermodilution measurements were made by injecting 15 ml cold
saline ( 8 C) through the central venous line at least
three times, at random points in the respiratory cycle. If
a difference of 15% was detected between the PiCCO
measurements, they were repeated. At the same time,
Nexn measurements were performed by collecting ve
numerical values over a period of 3 min and determining the mean value.
A passive leg-raising manoeuvre was performed
after induction of anaesthesia and after surgery (before
transfer to the ICU); this involved elevating the legs to
45 while the patient was in the horizontal position. This
manoeuvre causes blood to be shifted from the lower
part of the body to the intrathoracic compartment,
generating an increase in right and left ventricular
preload, and leads to well recognised haemodynamic
effects. All measurements were repeated before, during,
and after raising the legs.
Statistical comparisons were performed using commercially available statistics software (GraphPad Prism
5; GraphPad Software Inc., San Diego, CA, USA and
Anaesthesia 2012 The Association of Anaesthetists of Great Britain and Ireland

Broch et al. | Nexfin vs PiCCO

MedCalc for Windows, version 11.6.1.0; MedCalc Software, Mariakerke, Belgium). To demonstrate the relationship between sample size and the width of the
condence interval of the estimated variable, we calculated the width of the 95% condence interval of the
limits of agreementqas
recommended by Bland and
Altman (as 1:96 n3  s, where s is the standard
deviation of the bias). To describe the agreement
between PiCCO and Nexn, BlandAltman plots for
repeated measures were calculated for each time period
(before and after CPB). Percentage error was calculated
as described by Critchley et al. [15], using the limits of
agreement (2 SD) of the bias divided by the mean values
of the two methods. BlandAltman plots for haemodynamic trends and paired t-tests followed by Bonferroni
correction for multiple comparisons were also performed.

Results
We enrolled 40 patients, 29 (73%) of whom were men;
mean (SD) age was 64 (4) years, body mass index was 28
(3) kg.m)2 and left ventricular ejection fraction was 0.6
(0.1). A total of 404 data pairs (213 before and 191 after

Anaesthesia 2012, 67, 377383

CPB) were obtained. There was a signicant difference


(p < 0.05) between mean arterial pressure, systemic
vascular resistance index, heart rate and cardiac index
using both measurement devices before and after CPB.
There was a signicant correlation between Nexn and
thermodilution before and after CPB (Fig. 1). The bias,
limits of agreement and percentage error were within the
acceptable range (Table 1).
The passive leg-raise manoeuvre was performed in
37 40 patients before and 32 40 patients after CPB
(Table 2). Patients who increased their stroke volume
index by > 15% were dened as responders. We
observed 19 responders both before (51%) and after
(59%) CPB. Again, the measurements obtained by the
Nexn and PiCCO were correlated, with acceptable bias
and limits of agreement.
Changes in values displayed by both devices are
illustrated in Fig. 2. Again, they were well correlated,
with limits of agreement from )31% to +23%; after CPB,
limits of agreement ranged from )12% to +13%.
There was no signicant correlation between mean
arterial pressure and Nexn before (r2 = 0.01, p = 0.08)
and after (r2 = 0.009, p = 0.72) CPB; however, there was

Figure 1 Correlation plots (left) and BlandAltman plots (right) showing correlation and agreement between the
cardiac index measured by the Nexn and by PiCCO in 40 patients undergoing coronary artery surgery, before and after
cardiopulmonary bypass. CI, cardiac index; CPB, cardiopulmonary bypass; PE, percentage error.
Anaesthesia 2012 The Association of Anaesthetists of Great Britain and Ireland

379

Broch et al. | Nexfin vs PiCCO

Anaesthesia 2012, 67, 377383

Table 1 Comparison of Nexn and PiCCO in 40 patients undergoing coronary artery surgery. Values are mean (SD) or
percentage.
Cardiac index; l.min)1.m)2
Bias; l.min)1.m)2
95% limits of agreement; l.min)1.m)2
Percentage error

Before CPB (213 measurements)

After CPB (191 measurements)

2.4 (0.6)
0.06 (0.27)
)0.60 to +0.49
23%

2.8 (0.6)
0.09 (0.37)
)0.81 to +0.63
26%

CPB, cardiopulmonary bypass.

Table 2 Comparison of Nexn and PiCCO in patients


undergoing coronary artery surgery in whom a passive
leg-raise manoeuvre was performed. Values are mean
(SD) or percentage. CPB, cardiopulmonary bypass.

r2
Bias; l.min)1.m)2
95% limits of
agreement;
l.min)1.m)2
Percentage error

Before CPB
(n = 37)

After CPB
(n = 32)

r2 = 0.72
(p < 0.001)
0.14 (0.43)
)0.98 to +0.70

r2 = 0.76
(p < 0.001)
0.07 (0.26)
)0.59 to +0.44

39%

18%

some correlation between the Nexn and systemic


vascular resistance index before (r2 = 0.31, p < 0.0001)
and after (r2 = 0.16, p < 0.0001) CPB.

Discussion
We have shown that the haemodynamic values and
indices measured by the Nexn and the PiCCO in
patients with preserved left ventricular function who do
not require inotropic support are well correlated. Such a
comparison has not been described before. The Nexn
was able to track haemodynamic changes and trends
before and after CPB however values were inuenced by
systemic vascular resistance.
The recently introduced Nexn monitoring system
is completely non-invasive in that it requires only the
use of a pneumatic nger cuff, without the insertion of
any intravascular lines. We have shown that the values
of cardiac output that it displays are well correlated with
the more invasive PiCCO system that requires the
insertion of both a central venous line and a femoral

Figure 2 Correlation plots (left) and BlandAltman plots (right) showing correlation and agreement of percentage
change in cardiac index measured by the Nexn and PiCCO in 40 patients undergoing coronary artery surgery before
and after cardiopulmonary bypass. CI, cardiac index; CPB, cardiopulmonary bypass.
380

Anaesthesia 2012 The Association of Anaesthetists of Great Britain and Ireland

Broch et al. | Nexfin vs PiCCO

arterial line. The PiCCO is, however, in our opinion, less


invasive again than the pulmonary artery catheter, to
which many more complications are attributed [16].
As we have demonstrated, the Nexn does not
appear to need calibration before use. Its accuracy
depends on correct estimation of impedance, compliance and resistance [17], which are calculated using the
patient-specic data entered before connecting to the
patient. Aortic diameter and arterial pressure are then
estimated [1820], before cardiac output index is calculated.
Apart from the non-invasive nature and good
correlation of the device, we have shown that it is
easy and rapid to set up and use, meaning that we
can recommend use in similar settings. However, we
have not yet shown that its accuracy can be relied on
during haemodynamic instability or when inotropic or
vasoactive drugs are in use, meaning that further
studies in these areas must be carried out before more
widespread use can be recommended.
Other studies have also shown non-invasive measurement of nger arterial pressure to be quite accurate
[10, 21]. However, up to now, studies looking at their
use to estimate stroke volume and cardiac index have
been limited, with equivocal results [11, 2225].
Although in the study applying the Modelow algorithm
to the nger-derived arterial pressure curve, the device
showed less accuracy compared with pulmonary thermodilution, other investigations using the Nexn found
good correlation with pulmonary thermodilution and
echocardiography-derived cardiac output, respectively.
As recommended by Critchley et al. [26], we used
30% as our cut-off for limits of agreement, and we have
shown that the Nexn and PiCCO are adequately
correlated at varying times during coronary surgery, and
also before and after rapid intravascular volume expansion, as estimated using the passive leg-raise manoeuvre.
However, at the time of the passive leg-raise the Nexn
results did differ from the PiCCO by a small amount;
this nding has also been shown in a previous similar
study carried out in the ICU [27]. A number of previous
studies have also shown that other devices that rely on
pulse contour analysis are also not as accurate during
marked intravascular volume changes [2831].
Of interest, the mean arterial pressure was not
correlated with the cardiac output measured by the
Anaesthesia 2012 The Association of Anaesthetists of Great Britain and Ireland

Anaesthesia 2012, 67, 377383

Nexn during the study period, but it was related to


the systemic vascular resistance. This has also been
shown in a number of other studies using pulse
contour analysis with other devices [7, 3235]. It must
be emphasised, however, that the relationship that we
have shown is based on only a few data points from a
small number of patients. However, from a physiological standpoint, cardiac index will be higher for a
given mean arterial pressure if systemic vascular
resistance index is low, which may explain some of
our ndings.
As suggested by a recent study [36], we have shown
that the Nexn and PiCCO are precise during stable
haemodynamic conditions, with no rapid changes
during data collection. This was also demonstrated
during the passive leg-raise test. This means both
systems are able to track such haemodynamic changes
and trends with acceptable accuracy.
Some limitations of our study should be noted. Our
data were derived from patients with preserved left
ventricular function undergoing elective coronary surgery under stable haemodynamic conditions and without ongoing pharmacologic support. Therefore, our
results cannot automatically be transferred to other
patient groups. Furthermore, we used PiCCO instead of
a pulmonary artery catheter as our gold standard. We
feel that this was justied because the PiCCO has been
shown to be interchangeable with the pulmonary artery
catheter in a number of other studies [7, 8, 37, 38].
However, the PiCCO has some limitations, particularly
after weaning from CPB, with transient thermal changes
leading to increased bias [35].
From the point of view of clinical relevance,
individualised intra-operative goal directed therapy has
been shown to lead to reduced morbidity [2]. Noninvasive and continuous estimation of stroke volume
and cardiac output is required for this to be carried out,
and we have demonstrated that the Nexn is a good
candidate for such a monitor.

Acknowledgements
The authors are indebted to Volkmar Hensel-Bringmann and Bernd Kuhr for excellent technical assistance
and logistic support. BB and MM are members of the
medical advisory board of Pulsion Medical Systems, the
manufacturer of the PiCCO device, and have received
381

Anaesthesia 2012, 67, 377383

honoraria for consulting and giving lectures. The other


authors have no competing interests to declare.

References
1. Pearse R, Dawson D, Fawcett J, Rhodes A, Grounds RM, Bennett
ED. Early goal-directed therapy after major surgery reduces
complications and duration of hospital stay. A randomised,
controlled trial. Critical Care 2005; 9: R68793.
2. Rhodes A, Cecconi M, Hamilton M, et al. Goal-directed therapy
in high-risk surgical patients: a 15-year follow-up study.
Intensive Care Medicine 2010; 36: 132732.
3. Jans O, Tollund C, Bundgaard-Nielsen M, Selmer C, Warberg J,
Secher NH. Goal-directed fluid therapy: stroke volume optimisation and cardiac dimensions in supine healthy humans. Acta
Anaesthesiologica Scandivacica 2008; 52: 53640.
4. Wheeler AP, Bernard GR, Thompson BT, et al. Pulmonary-artery
versus central venous catheter to guide treatment of acute
lung injury. New England Journal of Medicine 2006; 354:
221324.
5. Sandham JD, Hull RD, Brant RF, et al. A randomized, controlled
trial of the use of pulmonary-artery catheters in high-risk
surgical patients. New England Journal of Medicine 2003; 348:
514.
6. Richard C, Warszawski J, Anguel N, et al. Early use of the
pulmonary artery catheter and outcomes in patients with shock
and acute respiratory distress syndrome: a randomized controlled trial. Journal of the American Medical Association 2003;
290: 271320.
7. Sander M, Spies CD, Grubitzsch H, Foer A, Muller M, von
Heymann C. Comparison of uncalibrated arterial waveform
analysis in cardiac surgery patients with thermodilution cardiac
output measurements. Critical Care 2006; 10: R164.
8. Friesecke S, Heinrich A, Abel P, Felix SB. Comparison of
pulmonary artery and aortic transpulmonary thermodilution
for monitoring of cardiac output in patients with severe heart
failure: validation of a novel method. Critical Care Medicine
2009; 37: 11923.
9. Ritter S, Rudiger A, Maggiorini M. Transpulmonary thermodilution-derived cardiac function index identifies cardiac dysfunction
in acute heart failure and septic patients: an observational
study. Critical Care 2009; 13: R133.
10. Eeftinck Schattenkerk DW, van Lieshout JJ, van den Meiracker
AH, et al. Nexfin noninvasive continuous blood pressure validated against Riva-Rocci Korotkoff. American Journal of Hypertension 2009; 22: 37883.
11. Bogert LW, Wesseling KH, Schraa O, et al. Pulse contour cardiac
output derived from non-invasive arterial pressure in cardiovascular disease. Anaesthesia 2010; 65: 111925.
12. Penaz J, Voigt A, Teichmann W. [Contribution to the continuous
indirect blood pressure measurement]. Zeitschrift fur die Gesamte Innere Medizin und ihre Grenzgebiete 1976; 31: 10303.
13. Imholz BP, Wieling W, van Montfrans GA, Wesseling KH. Fifteen
years experience with finger arterial pressure monitoring:
assessment of the technology. Cardiovascular Research 1998;
38: 60516.
14. Wesseling KH, Jansen JR, Settels JJ, Schreuder JJ. Computation of
aortic flow from pressure in humans using a nonlinear, threeelement model. Journal of Applied Physiology 1993; 74: 2566
73.
15. Critchley LA, Lee A, Ho AM. A critical review of the ability of
continuous cardiac output monitors to measure trends in cardiac
output. Anesthesia and Analgesia 2010; 111: 118092.

382

Broch et al. | Nexfin vs PiCCO


16. Sakka SG, Reinhart K, Wegscheider K, Meier-Hellmann A. Is the
placement of a pulmonary artery catheter still justified solely for
the measurement of cardiac output? Journal of Cardiothoracic
and Vascular Anesthesia 2000; 14: 11924.
17. Jansen JR, Schreuder JJ, Mulier JP, Smith NT, Settels JJ, Wesseling
KH. A comparison of cardiac output derived from the arterial
pressure wave against thermodilution in cardiac surgery
patients. British Journal of Anaesthesia 2001; 87: 21222.
18. Wuyts FL, Vanhuyse VJ, Langewouters GJ, Decraemer WF, Raman
ER, Buyle S. Elastic properties of human aortas in relation to age
and atherosclerosis: a structural model. Physics in Medicine and
Biology 1995; 40: 157797.
19. Winer N, Sowers JR, Weber MA. Gender differences in vascular
compliance in young, healthy subjects assessed by pulse
contour analysis. Journal of Clinical Hypertension 2001; 3:
14552.
20. Langewouters GJ, Wesseling KH, Goedhard WJ. The pressure
dependent dynamic elasticity of 35 thoracic and 16 abdominal
human aortas in vitro described by a five component model.
Journal of Biomechanics 1985; 18: 61320.
21. Hofhuizen CM, Lemson J, Hemelaar AE, et al. Continuous noninvasive finger arterial pressure monitoring reflects intra-arterial
pressure changes in children undergoing cardiac surgery. British
Journal of Anaesthesia 2010; 105: 493500.
22. Hirschl MM, Kittler H, Woisetschlager C, et al. Simultaneous
comparison of thoracic bioimpedance and arterial pulse waveform-derived cardiac output with thermodilution measurement.
Critical Care Medicine 2000; 28: 1798802.
23. Lu Z, Mukkamala R. Continuous cardiac output monitoring in
humans by invasive and noninvasive peripheral blood pressure
waveform analysis. Journal of Applied Physiology 2006; 101:
598608.
24. Rang S, de Pablo Lapiedra B, van Montfrans GA, Bouma BJ,
Wesseling KH, Wolf H. Modelflow: a new method for noninvasive assessment of cardiac output in pregnant women.
American Journal of Obstetrics and Gynecology 2007; 196:
2318.
25. van Geldorp IE, Delhaas T, Hermans B, et al. Comparison of a
non-invasive arterial pulse contour technique and echo Doppler
aorta velocity-time integral on stroke volume changes in
optimization of cardiac resynchronization therapy. Europace
2011; 13: 8795.
26. Critchley LA, Critchley JA. A meta-analysis of studies using bias
and precision statistics to compare cardiac output measurement
techniques. Journal of Clinical Monitoring and Computing 1999;
15: 8591.
27. Stover JF, Stocker R, Lenherr R, et al. Noninvasive cardiac output
and blood pressure monitoring cannot replace an invasive
monitoring system in critically ill patients. BMC Anesthesiology
2009; 9: 6.
28. Hamzaoui O, Monnet X, Richard C, Osman D, Chemla D, Teboul JL.
Effects of changes in vascular tone on the agreement between
pulse contour and transpulmonary thermodilution cardiac output measurements within an up to 6-hour calibration-free
period. Critical Care Medicine 2008; 36: 43440.
29. Gruenewald M, Renner J, Meybohm P, Hocker J, Scholz J, Bein B.
Reliability of continuous cardiac output measurement during
intra-abdominal hypertension relies on repeated calibrations: an
experimental animal study. Critical Care 2008; 12: R132.
30. Bein B, Meybohm P, Cavus E, et al. The reliability of pulse
contour-derived cardiac output during hemorrhage and after
vasopressor administration. Anesthesia and Analgesia 2007;
105: 10713.

Anaesthesia 2012 The Association of Anaesthetists of Great Britain and Ireland

Broch et al. | Nexfin vs PiCCO


31. Eleftheriadis S, Galatoudis Z, Didilis V, et al. Variations in arterial
blood pressure are associated with parallel changes in FlowTrac Vigileo-derived cardiac output measurements: a prospective comparison study. Critical Care 2009; 13: R179.
32. Sakka SG, Kozieras J, Thuemer O, van Hout N. Measurement of
cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis. British
Journal of Anaesthesia 2007; 99: 33742.
33. Yamashita K, Nishiyama T, Yokoyama T, Abe H, Manabe M.
Effects of vasodilation on cardiac output measured by PulseCO.
Journal of Clinical Monitoring and Computing 2007; 21: 3359.
34. Yamashita K, Nishiyama T, Yokoyama T, Abe H, Manabe M. The
effects of vasodilation on cardiac output measured by PiCCO.
Journal of Cardiothoracic and Vascular Anesthesia 2008; 22:
68892.

Anaesthesia 2012 The Association of Anaesthetists of Great Britain and Ireland

Anaesthesia 2012, 67, 377383


35. Sander M, von Heymann C, Foer A, et al. Pulse contour analysis
after normothermic cardiopulmonary bypass in cardiac surgery
patients. Critical Care 2005; 9: R72934.
36. Squara P, Cecconi M, Rhodes A, Singer M, Chiche JD. Tracking
changes in cardiac output: methodological considerations for
the validation of monitoring devices. Intensive Care Medicine
2009; 35: 18018.
37. Sakka SG, Reinhart K, Meier-Hellmann A. Comparison of
pulmonary artery and arterial thermodilution cardiac output
in critically ill patients. Intensive Care Medicine 1999; 25:
8436.
38. Breukers RM, Groeneveld AB, de Wilde RB, Jansen JR. Transpulmonary versus continuous thermodilution cardiac output after
valvular and coronary artery surgery. Interactive Cardiovascular
and Thoracic Surgery 2009; 9: 48.

383

You might also like