You are on page 1of 14

Article

An Intelligent Optical Dissolved Oxygen


Measurement Method Based on a Fluorescent
Quenching Mechanism
Fengmei Li, Yaoguang Wei *, Yingyi Chen, Daoliang Li and Xu Zhang
Received: 1 October 2015; Accepted: 1 December 2015; Published: 9 December 2015
Academic Editor: Frances S. Ligler
College of Information and Electrical Engineering, China Agricultural University, 17 Tsinghua East Road,
Beijing 100083, China; lifm@cau.edu.cn (F.L.); chyingyi@126.com (Y.C.); dliangl@cau.edu.cn (D.L.);
zhangxu_zx888@sina.com (X.Z.)
* Correspondence: weiyaoguang@gmail.com; Tel.: +86-10-6273-6764; Fax: +86-10-6273-7741

Abstract: Dissolved oxygen (DO) is a key factor that influences the healthy growth of fishes
in aquaculture. The DO content changes with the aquatic environment and should therefore
be monitored online. However, traditional measurement methods, such as iodometry and
other chemical analysis methods, are not suitable for online monitoring. The Clark method
is not stable enough for extended periods of monitoring. To solve these problems, this paper
proposes an intelligent DO measurement method based on the fluorescence quenching mechanism.
The measurement system is composed of fluorescent quenching detection, signal conditioning,
intelligent processing, and power supply modules. The optical probe adopts the fluorescent
quenching mechanism to detect the DO content and solves the problem, whereas traditional
chemical methods are easily influenced by the environment. The optical probe contains a
thermistor and dual excitation sources to isolate visible parasitic light and execute a compensation
strategy. The intelligent processing module adopts the IEEE 1451.2 standard and realizes intelligent
compensation. Experimental results show that the optical measurement method is stable, accurate,
and suitable for online DO monitoring in aquaculture applications.
Keywords: dissolved oxygen; fluorescent quenching; intelligent compensation

1. Introduction
Dissolved oxygen (DO) refers to the oxygen molecules dissolved in water and is essential to
maintain human and animal life. Oxygen is an important analyte because of its key role in the
life science, biotechnology, medicine, and aquaculture industries. The DO content in water is an
indication of water quality, and careful control of oxygen levels is important in the self-purification
processes of wastewater [1,2]. Water quality is closely related to the contaminants present in water,
such as H2 S, NO2 , NH4 + , and organic matter. Wastewater characteristics, including color, chemical
oxygen demand (COD), and biological oxygen demand (BOD), specifically indicate the level of
pollutants in industrial wastewater [3]. At the same time, DO plays a very important role in the
health and growth of aquatic organisms [4,5]. A DO content of less than 2 mg/L for a certain
number of hours causes the suffocation and death of aquatic organisms [6]. For humans, the DO
content of drinking water should not be less than 6 mg/L. Consequently, the determination of
oxygen concentrations is of high importance in the aquaculture industry and in daily life. However,
monitoring the DO content with all its external influencing factors, such as temperature, pressure,
and salinity, is difficult. To obtain an accurate DO content, the detection method should implement

Sensors 2015, 15, 3091330926; doi:10.3390/s151229837

www.mdpi.com/journal/sensors

Sensors 2015, 15, 3091330926

intelligent compensation. In general, three methods can be used to detect DO content: iodometric,
electrochemical, and optical methods [7,8].
The iodometric method [9,10] is a popular and precise method of detecting the DO content in
water. It is a fiducial method but has a complex detection process and cannot be used to detect the
water quality online. This method is mainly used as a benchmark in the laboratory environment
(off-line). The electrochemical method [1114] uses electrodes to detect the current produced by
redox reactions at the electrodes. This method can be classified as polarographic type or galvanic
cell type based on the detection principle. The electrochemical method has a long history in the
detection of DO content; the first so-called Clark polarographic method was designed by Clark of
the YSI Company in 1956 [12]. In contrast to iodometry, the electrochemical method monitors DO
content by the oxidation-reduction reaction that occurs between the electrode and DO molecules
and consumes oxygen in the detection process. Given that instrumental drift is inevitable with the
large number of factors that are involved in determining the detection result, electrochemical sensors
require regular calibration and replacement. Optical DO sensors [15,16] are more attractive than the
iodometry and electrochemical methods because they have a fast response time, do not consume
oxygen, have a small drift over time, have the capability to withstand external disturbances, and
require marginal calibration. The detection principle of optical DO sensors is based on fluorescent
quenching, including fluorescent lifetime detection and fluorescent intensity detection. Intensity
detection can be achieved through the photodiode, unlike lifetime, which should be detected based
on the phase shift [17]. This study develops an intelligent optical measurement method based on the
fluorescent quenching mechanism.
The aforementioned methods have some advantages and disadvantages that make them
unsuitable to the aquaculture industry in China. First, the detection of DO content is difficult for
the aquarist who must deal with numerous factors which influence the aquaculture industry. A
comparison of the three methods shows that the electrochemical method is not a good choice because
of its weak anti-interference properties. Second, the DO content is not constant, and insufficient
concentration in natural water leads to the death of fishes. Thus, detecting DO content in real time
is very important. However, iodometry method water samples must be tested in the laboratory,
rendering this method unsuitable for monitoring organisms in actual production for this reason.
Finally, traditional optical sensors have several disadvantages, including susceptibility to changes
in external temperature, pressure, and salinity and attenuation of light source and drift because
of the degradation or leaching of the dye. The influence off all these factors can be decreased by
adding intelligent processing modules. The conventional optical DO sensor introduced from abroad
is expensive and does not have high accuracy when used in the aquaculture industry. Thus, designing
and developing an inexpensive and intelligent dedicated optical DO sensor is necessary.
This study proposes and develops an intelligent DO measurement method based on the
fluorescent quenching mechanism. The sensor contains four modules: fluorescent quenching
detection, signal conditioning, intelligent processing, and power supply modules. The sensor based
on fluorescent quenching has several advantageous aspects: lower power consumption, smaller size,
higher accuracy, and stronger anti-interference properties than iodometry or electrochemical sensors.
2. Materials and Methods
2.1. The Overall Design of Optical Dissolved Oxygen Sensor
Given the presence of unstable influencing factors, the sensor adopts an optical probe based
on the quenching of fluorescence. Compared with the traditional DO sensor, the intelligent optical
DO sensor proposed in this study has an enhanced probe structure and an additional intelligent
processing module. These calibration parameters are stored in the transducer electronic data sheet
(TEDS) memory. Figure 1 shows that the fluorescent quenching detection, signal conditioning,
intelligent processing, and power supply modules are included in the intelligent sensor.

30914

Sensors 2015, 15, 3091330926

The fluorescent quenching detection module contains a temperature probe and a DO probe. The
temperature probe is responsible for collecting the water temperature signals, and the DO probe is
responsible for collecting the DO signals. The original input signal can be converted into 02.5 V
Sensors
2015, 15,
voltage
signal
bypagepage
the signal conditioning circuits. The MSP430 microcontroller, which is the core of
the intelligent
processing
module, is connected to the signal conditioning circuits, TEDS memory,
Sensors
2015, 15, pagepage
the
intelligent
processing module, is connected to the signal conditioning circuits, TEDS memory,
and serial
interface [18]. The collected data are fused through multi-probe data fusion technology,
and serial interface [18]. The collected data are fused through multi-probe data fusion technology,
the
intelligent
processing
is connected
tocompatible
the signal conditioning
circuits,
TEDS
memory,
and the
DO
value
obtained
ismodule,
transferred
through
compatible
RS485
interface
after
being
processed
and the DO
value
obtained
is
transferred
through
RS485
interface
after
being
processed
and
serial
interface
[18].
The
collected
data
are
fused
through
multi-probe
data
fusion
technology,
and analyzed
by the
microcontroller.
allowsthe
the
microcontroller
to communicate
and analyzed
by the
microcontroller.The
TheRS485
RS485 interface
interface allows
microcontroller
to communicate
and upper
the
value
obtained
is
through
compatible
RS485
interface
after
being
processed
with with
the
PC.
The
sensor
is transferred
powered
an
on-off
powersupply,
supply,
which
is also
controlled
by the
the DO
upper
PC.
The
sensor
is
poweredby
by
an on-off
power
which
is also
controlled
by the
and
analyzed
by
the
microcontroller.
The
RS485
interface
allows
the
microcontroller
to
communicate
MSP430
microcontroller.
MSP430
microcontroller.
with the upper PC. The sensor is powered by an on-off power supply, which is also controlled by the
MSP430 microcontroller.

Figure 1. Overall design of the optical DO sensor.

Figure 1. Overall design of the optical DO sensor.


Figure 1. Overall
design
of the optical DO sensor.
2.2. Design of the Fluorescent Quenching
Detection
Module

2.2. Design of the Fluorescent Quenching Detection Module

The schematic
of the fluorescent
detection module is shown in Figure 2. The probe
2.2. Design
of the Fluorescent
Quenchingquenching
Detection Module

The
schematic
of the
fluorescent
quenching
detection
module
is shown
in Figure 2. isThe
has an
approximate
length
of 16 cm and
a diameter
of 4 cm. The
compact
probe configuration
usedprobe
The
schematic
of
the
fluorescent
quenching
detection
module
is
shown
in
Figure
2.
The
probe
forapproximate
compatibility with
theofrequirements
aquaculture
industry.
illustrated
in Figure
2, the
has an
length
16 cm and of
a the
diameter
of 4 cm.
The As
compact
probe
configuration
is
has an
approximate
length of 16 cm and a blue
diameter ofsol-gel
4 cm. The
compact
probered
configuration
is used
contains dual
film,
glass
slide,
optical filter,
used DO
for probe
compatibility
with high-brightness
the requirements ofLEDs,
the aquaculture
industry.
As illustrated
in blue
Figure 2,
for compatibility
with
the requirements
of theThis
aquaculture
industry.
As illustrated
in resistance
Figure 2, the
optical
filter
paper,
and
photodiode.
modulesol-gel
also includes
a platinum
the DO
probe
contains
dualsilicon
high-brightness
blue LEDs,
film, glass
slide, red
opticaltofilter,
DO
probe
contains
dual
high-brightness
blue
LEDs,
sol-gel
film,
glass
slide,
red
optical
filter, blue
monitor
ambient
temperature
during
measurements.
The
fluorescent
intensity
and
temperature
are to
blue optical
filter paper,
paper, and
andsilicon
siliconphotodiode.
photodiode.
This
module
includes
a platinum
resistance
optical filter
This
module
alsoalso
includes
a platinum
resistance
to
processed
in the software
for temperature
calibration.
monitor
ambient
temperature
during
Thefluorescent
fluorescent
intensity
temperature
monitor
ambient
temperature
duringmeasurements.
measurements. The
intensity
andand
temperature
are are
processed
in the
forfor
temperature
processed
in software
the software
temperaturecalibration.
calibration.
PCB

OF1

PCB
Temperature Probe

OF1 Reference
Blue
LED
Blue Reference
LED
Blue
Excitation
LED
Blue Excitation
LED
OF2

Silicon
Photodiode
Temperature
Probe
Silicon Photodiode
OF3
OF3
Metal Case

OF2
Sol-Gel Film

Metal Case
Glass Slide

Glass Slide
Sol-Gel Film
Figure 2. Schematic of the fluorescence quenching detection module.

Figure
2. Schematic
fluorescence
quenching
detection
module.
Figure
2. Schematic
ofthe
the(LA470-02)
fluorescenceare
quenching
detection
module.
The dual high-brightness
blue of
LEDs
modulated
at the
same frequency, so that
the reference LED can be used to compensate the excitation LED because the light intensity loss of
The dual high-brightness
blue LEDs
(LA470-02)
are modulated
at the same
frequency,
so that
the blue
are exactly theblue
same.
At the
same frequency,
the photodiode
cansame
reduce
background
The
dualLEDs
high-brightness
LEDs
(LA470-02)
are modulated
at the
frequency,
so that
the
reference
LED
can
be
used
to
compensate
the
excitation
LED
because
the
light
intensity
loss
of
radiation
due
to
the
ambient
light
in
the
measurement
environment
and
avoid
the
excitation
of
any
the reference LED can be used to compensate the excitation LED because the light intensity loss of
the
blue
LEDs
are
exactly
the
same.
At
the
same
frequency,
the
photodiode
can
reduce
background
fluorescent
material.
In the
addition,
the
intensity
of frequency,
the LEDs are
to a low
level
at which
the
the blue
LEDsdue
are
exactly
same.
Atthe
themeasurement
same
thereduced
photodiode
reduce
background
radiation
to
the ambient
light
in
avoid can
the[19].
excitation
of any
photobleaching
phenomenon
of the
dye
has a smallenvironment
probability and
of occurring
The central
radiation
due tomaterial.
the ambient
light inthe
the
measurement
environment
and aavoid
the excitation
of any
fluorescent
In excitation
addition,
intensity
thenm,
LEDs
are reduced
low
level
at whichfilter
the
wavelength
of the blue
LED
is aboutof465
which
is filteredtoby
a blue
bandpass
fluorescent
material.phenomenon
In addition,
the
intensity
the probability
LEDs are reduced
to a[19].
low The
levelcentral
at which
photobleaching
thewavelengths.
dye has aofsmall
of occurring
paper
to screen out
the light of of
other
The experimental
results
show that
the blue
light
wavelength of the blue excitation LED is about 465 nm, which is filtered by a blue bandpass filter
paper to screen out the light of other wavelengths. The experimental results show that the blue light
30915
3

Sensors 2015, 15, 3091330926

the photobleaching phenomenon of the dye has a small probability of occurring [19]. The central
wavelength of the blue excitation LED is about 465 nm, which is filtered by a blue bandpass filter
paper to screen out the light of other wavelengths. The experimental results show that the blue light
can induce the sensitive membrane to emit fluorescence at 650 nm. To reduce the influence of parasitic
light, the probe is equipped with blue bandpass filter papers (OF1 and OF2) in front of the LEDs and
red high-pass filter (OF3) in front of the silicon photodiode. A silicon photodiode (OPT 301) is used
to receive the fluorescence emitted from the sol-gel film and blue light from the reference LED. The
blue excitation LED and blue reference LED are separated on different sides of the red high-pass
filter, which is beneficial in cutting off parasitic light and guaranteeing the accuracy of the optical
signal detection.
The fluorescent sensing film is the most important part of the DO sensor and its performance
significantly influences the accuracy, efficiency, and stability of the sensor. Researchers have
conducted several studies on fluorescence indicators [2022] and found that the most common
fluorescent indicators contain metal porphyrin complexes, organic polycyclic aromatic hydrocarbons,
and transition metal complexes [23]. Ru(bpy)3 Cl2 has been chosen as the fluorescence indicator in this
study because of its highly emissive metal-to-ligand charge-transfer state, long lifetime, and strong
absorption in the blue-green region of the spectrum, which is compatible with the high-brightness
blue LED [20]. The dye is entrapped in a porous and hydrophobic sol-gel film that is approximately
0.04 mm. The sol-gel film is mounted on the glass slide surface, which should be transparent for the
excitation and luminescence to penetrate. The film should also be in the shape of an arc and maintain
a stable size; the arc surface is designed to increase the area of contact and avoid surface bubble. The
operating principle of the sensor is based on the fluorescent quenching mechanism. The fluorescent
quenching process is described by the Stern-Volmer equation [2426]:
F0
1 ` Ksv rQs
F

(1)

where:
F0 denotes the fluorescence signal intensity of anaerobic water;
F denotes the fluorescence signal intensity of the water samples;
KSV denotes the Stern-Volmer constant; and
[Q] denotes the concentration of quencher.
As the equation shows, the degree of fluorescent quenching F0 /F has a straight-line relationship
with the concentration of the quencher [Q]. In this system, the quencher is the DO, so the DO content
can be determined by using Equation (1).
2.3. Design of the Signal Conditioning Module
The block diagram of the intensity measurement system is shown in Figure 3. The signal
conditioning module is responsible for collecting and processing the information. The dual LEDs
are modulated (MOD) into on-off at a frequency of 20 kHz. LED1 induces the sol-gel film to emit red
fluorescence. At the same time, LED2 emits blue light as a contrast. The luminescence signal from
the sensor dyes arrives time-delayed at the single photodiode (because of the luminescence lifetime
of the dye), so that the reference and luminescence signals are detected in different time windows
without interference. The fluorescence signal from the photodiode is converted to a voltage signal
via an I/V switching circuit. The signal is then amplified (AMP) to obtain a voltage signal of 02.5 V
and eliminate the higher harmonics of the signal. The voltage signal is filtered by a low pass (LP)
filter and provides a voltage signal proportional to the measured intensity shift. Finally, the output
voltage is divided by the voltage value based on the intensity in the anaerobic water to obtain F0 /F,
which has a linear correlation with the DO content. As presented in the Stern-Volmer equation, the
DO content can be determined.
30916

Sensors 2015, 15, 3091330926


Sensors 2015, 15, pagepage

Pt

LED2

Sensors 2015, 15, pagepage


RS485

OF1

MOD
MCU
Pt

Sensitive
membrane

LED2
LED1

PD

TEDS
OF2
OF1

RS485
MOD

TEDS

MCU
LP

amp

OF3
Sensitive
membrane

LED1 I/V
OF2

Figure
3. Schematic
the
signalconditioning
conditioning module.
Figure
3. Schematic
of of
the
signal
module.
LP
2.4. Design of the Intelligent Processing
Module

PD

OF3

I/V

amp

2.4. Design of the Intelligent Processing Module

With the rapid development of computers, communications, and networks, the DO sensor

Figure 3.of
Schematic
of the signal
conditioning module.
With
theberapid
computers,
communications,
and
thetraditional
DO sensor
should
more development
intelligent, networked,
and integrated
[20]. To solve
thenetworks,
problems of
shouldoptical
be more
intelligent,
networked,
andsensor
integrated
[20].
Tostudy
solveisthe
problems
traditional
sensors,
the intelligent
optical DO
proposed
in this
developed
withof
automatic
2.4. Design of the Intelligent Processing Module
which are optical
suitable DO
for the
online
detectionin
of this
aquaculture
long periods.
opticalcompensations,
sensors, the intelligent
sensor
proposed
study isover
developed
with automatic
With the
rapidare
development
computers,
communications,
and networks,
theperiods.
DO sensor
compensations,
which
suitable forofthe
online detection
of aquaculture
over long

IEEE1451.2
Standard
should
be more
intelligent, networked, and integrated [20]. To solve the problems of traditional
optical
sensors,
the
intelligent optical DO sensor proposed in this study is developed with automatic
IEEE1451.2The
Standard
data transfer is implemented using the protocols described in IEEE 1451.2 [27,28]. The
compensations, which are suitable for the online detection of aquaculture over long periods.
intelligent
sensor is
includes
a smart using
transducer
interface module
(STIM)
and 1451.2
network
capableThe
The
data transfer
implemented
the protocols
described
in IEEE
[27,28].
application processor (NCAP), and the sensor provides a general interface standard and a TEDS
intelligent
sensor
includes a smart transducer interface module (STIM) and network capable
IEEE1451.2
Standard
standard format for the intelligent sensor and the field bus. NCAP is connected to the STIM through
application
processor
(NCAP),
and the sensor
provides a general interface standard and a TEDS
The data
transfer
is implemented
using the protocols described in IEEE 1451.2 [27,28]. The
an RS485
interface,
as shown
in Figure 4.
standard
format
for theincludes
intelligent
sensor
and the
bus. NCAP
is(STIM)
connected
STIM
through
intelligent
sensor
a smart
interface
module
and to
network
capable
The calibration
parameters
aretransducer
stored
in field
the
TEDS
as shown
in Figure
4. the
Through
these
an RS485
interface,
as
shown
in
Figure
4.
application
processor
(NCAP),
and
the
sensor
provides
a
general
interface
standard
and
a
TEDS
parameters, the original signals can been converted into corrected signals, including DO,
The
calibration
are
stored
in
the
TEDS
as isshown
in toFigure
4. through
Through
standard
format
forparameters
the intelligent
sensor
and the
field
bus.
NCAP
connected
the STIM
temperature,
pressure,
salinity, and
luminous
flux
signals.
RS485 interface,
shown in
Figure
4.(2) been
these an
parameters,
the as
original
signals
converted into corrected signals, including DO,
D (1) can
D
D( n)
p
The pressure,
calibration
are
stored
in C
the
Figure
f (salinity,
Xparameters
X n ) luminous
...flux
[ X1 Has1 ]i [shown
X 2 H 2in
] j ...[
X n H4.
(2)
temperature,
signals.

1 , X 2 ,...and
i , j ... pTEDS
n ] Through these
i 0
j 0
p 0
parameters, the original signals can been converted into corrected signals, including DO,
D
pand
1qthe
D
pluminous
2sensor
q D
pnoutput
q flux signals.
temperature,
pressure,Xsalinity,
In the equation,
variables; Hn denotes the revised output variables;
n denotes
i

f pX
X2 , ...X
q variables;
... and
s rX
D (1) D (2)
D ( n )C
n Hn s
Dn denotes
the
ofnoutput
Ci,j...p
i,j,,prX
denotes
polynomial
coefficient.
2 H2 s ...rX
1 , order
1 H1the
i
j
p
f ( X1 , X 2 ,... X n i)0
0 Ci , j ... p [ X 1 H1 ] [ X 2 H 2 ] ...[ X n H n ]
j0 p...
i 0

j 0

p 0

(2)

(2)

In theInequation,
Xn denotes
the
denotes
revised
output
variables;
the equation,
Xn denotes
thesensor
sensoroutput
output variables;
variables; HHn ndenotes
thethe
revised
output
variables;
Dn denotes
the order
of output
variables;
denotes
polynomial
coefficient.
Dn denotes
the order
of output
variables;and
andC
Ci,j,...
i,j,,p,p
denotes
thethe
polynomial
coefficient.

Figure 4. Model of data correction.

The calibration standard equation is shown in Figure 4. The relationship between the output
variables and revised output variables, such as DO, temperature, pressure, salinity, and luminous
Figure4.4.Model
Model of
of5data
data correction.
Figure
correction.

The calibration standard equation is shown in Figure 4. The relationship between the output
variables and revised output variables, such as DO, temperature, pressure, salinity, and luminous
5

30917

Sensors 2015, 15, 3091330926

The
calibration
standard equation is shown in Figure 4. The relationship between the output
Sensors
2015,
15, pagepage
variables and revised output variables, such as DO, temperature, pressure, salinity, and luminous flux
flux signals
are converted
a standard
The sensor
can directly
determine
accurate
signals
are converted
into ainto
standard
form. form.
The sensor
can directly
determine
accurate
valuesvalues
when
when
the calibration
parameters
in the
TEDS,
not onlyautomatic
facilitatesrecognition
automatic
it
uses it
theuses
calibration
parameters
stored instored
the TEDS,
which
not which
only facilitates
recognition
but the
alsosensor
solvesautomatic
the sensorcorrection
automaticproblem.
correctionInproblem.
In theoptical
intelligent
but
also solves
the intelligent
DO optical
sensor, DO
the
sensor,
the total,
channel,
and TEDS
calibration
must In
bethis
defined.
this
study, the
of total
total,
channel,
and
calibration
must TEDS
be defined.
study,Inthe
definition
of definition
total and channel
and channel
TEDS
makes the
intelligent
optical
sensor moreand
standardized,
and that of
the
TEDS
makes the
intelligent
optical
DO sensor
moreDO
standardized,
that of the calibration
TEDS
calibration
may correct
original
at any
time. These
havefor
significance
for the
may
correctTEDS
the original
data the
at any
time.data
These
functions
havefunctions
significance
the maintenance
maintenance
diagnosis
the sensors
[29,30].
Thus, theoptical
intelligent
DO
sensor
is called
and
diagnosisand
of the
sensorsof[29,30].
Thus,
the intelligent
DO optical
sensor is
called
a plug
anda
plug device.
and play device.
play
3.
Experimentsand
andDiscussions
Discussions
3. Experiments
3.1.
Calibration and
3.1. Calibration
and Validation
Validation
3.1.1. Temperature Compensation of the Optical DO Sensor
3.1.1. Temperature Compensation of the Optical DO Sensor
The temperature dependence of the oxygen sensor has a number of contributions, including the
The temperature dependence of the oxygen sensor has a number of contributions, including the
temperature dependence of the fluorescence sensing film, light source, and the diffusion coefficient
temperature dependence of the fluorescence sensing film, light source, and the diffusion coefficient
in the water. As discussed in Section 2.2, the light intensity related to temperature is measured by
in the water. As discussed in Section 2.2, the light intensity related to temperature is measured by the
the probe, and the temperature effect is measured by the sensor based on the thermistor [31]. This
probe, and the temperature effect is measured by the sensor based on the thermistor [31]. This study
study designs a correction model based on the temperature applicable to online DO monitoring. This
designs a correction model based on the temperature applicable to online DO monitoring. This
model can perform temperature compensation to reduce the influence of temperature through the
model can perform temperature compensation to reduce the influence of temperature through the
curve fitting algorithm.
curve fitting algorithm.
To eliminate the influence of temperature on sensitive film and light, anaerobic water and
To eliminate the influence of temperature on sensitive film and light, anaerobic water and
oxygen saturated solutions are prepared as standard solutions, which are heated with an integrated
oxygen saturated solutions are prepared as standard
solutions, which are heated with an integrated
thermostatic magnetic blender (HWCL-1) from 0 C to 40 C in an off-light environment. The values
thermostatic magnetic blender (HWCL-1) from 0 C to 40 C in an off-light environment. The values
are measured by a commercial DO meter (YSI 6150) in steps of 5 C (040 C), and the data between
are measured by a commercial DO meter (YSI 6150) in steps of 5 C (040 C), and the data between
output signal and temperature obtained are displayed in the plot shown in Figure 5.
output signal and temperature obtained are displayed in the plot shown in Figure 5.

Figure
Line chart
chart of
of the
Figure 5.
5. Line
the DO
DO content
content and
and output
output voltage
voltage signal.
signal.

The relationship can


can be
be described
described by
by the
the following
following equations:
equations:
DOt1 1 t1U
t1 1 ` t1 U
DO
DOt 2 2 t 2U
DO
(3)
t2 2 ` t2 U
(3)
..

DO . U

% DO tn n ` tn U
tn
n
tn
The slope is an important parameter to achieve the DO content, which changes with
temperature. The relationship between the slope and temperature can be obtained based on the
309186, the slope of the output voltage signal can be
curve fit method in this study. As shown in Figure
fitted by Equation (4):

&

Sensors 2015, 15, 3091330926

The slope is an important parameter to achieve the DO content, which changes with temperature.
The relationship between the slope and temperature can be obtained based on the curve fit method in
this study. As shown in Figure 6, the slope of the output voltage signal can be fitted by Equation (4):
Sensors 2015, 15, pagepage

t 0 ` 1 T ` 2 T 2 ` 3 T 3 ` 4 T 4 ` 5 T 5
(4)
t 0 1T 2T 2 3T 3 4T 4 5T 5
(4)
In the equation, 0 = 52.488, 1 = 4.3436, 2 = 0.2157, 3 = 0.0061, 4 = 0.00008,
In the equation,
20 == 52.488,
1 seen
= 4.3436,
2 =6,0.2157,
3 = 0.0061, 4 = 0.00008, 5 = 0.0000004,
5 = 0.0000004,
and R
0.9998, as
in Figure
which denotes
a very good linear fitting. The DO
2 = 0.9998, as seen in Figure 6, which denotes a very good linear fitting. The DO content can be
and
R
content can be obtained at any temperature, taking Equation (4) into Equation (3), and Equation (5)
obtained
atachieved.
any temperature, taking Equation (4) into Equation (3), and Equation (5) can then
can
then be
be achieved.
2
3
4
DOt DO
`0 U
` 4 UT
` 55 UT 5
2 ` 3 UT
3
4
`0U1UT
1UT`2 UT
t
2UT 3UT 4UT 5UT

(5)
(5)

Figure 6.
6. Fitting
Fitting curve
curve of
of the
the slope.
slope.
Figure

The calibration standard Equation (2) in the TEDS calibration can be adapted as follows:
The calibration standard Equation (2) in the TEDS calibration can be adapted as follows:
1

i
j
DOt 1
5 Cij [U H1 ] [T H 2 ]

i
0 j 0
DOt
Cij rU H1 si rT H2 s j
0 j
0 be presented by Equation (7):
If H1 = 0 and H2 = 0, then Equationi
(6)
can

(6)
(6)

DOt Equation
C00 C01T
C02Tbepresented
C03T C04by
T Equation
C05T
If H1 = 0 and H2 = 0, then
(6)can
(7):
2

(7)
C10U C11UT C12UT 2 2 C13UT 3 3 C14UT 4 4 C15UT 5 5
DOt C00 ` C01 T ` C02 T ` C03 T ` C04 T ` C05 T
(7)
3 ` C UT 4 ` C
5
The calibration values
with
and
output
voltage. The DO
`C10 Uare
` Cconsistent
UT 2the
` Cactual
11 UT ` C12
13 UT temperature
14
15 UT
value must be corrected according to the real-time temperature. Taking C00 = , C00 C05 = 0,C10 = 0,
calibration
temperature
and output
The and
DO
C11 =The
1, C
12 = 2, C13 values
= 3, Care
14 = consistent
4, and C15with
= 5 the
intoactual
Equation
(7) and detecting
the voltage.
temperature
value
be corrected
according
to can
the be
real-time
temperature.
Taking
C00 =
, the
C00range
C05of=the
0,
outputmust
voltage,
the accurate
DO value
determined
in the 040
C range
(i.e.,
C
=

,
C
=

,
C
=

,
C
=

,
C
=

,
and
C
=

into
Equation
(7)
and
detecting
the
10
0The11temperature
1
12
2
13
14
4 track 15
5
sensor).
probe
in this3 sensor
can
the temperature
response of the DO sensor
temperature
outputachieving
voltage, the
accurate DO
value can beindetermined
in thewhere
040 frequent
C range
in real time,and
thereby
temperature
compensation
an application
(i.e.,
the
range
of
the
sensor).
The
temperature
probe
in
this
sensor
can
track
the
temperature
short-term changes in temperature occur. The calibration curve based on intensity measurement
response
of the DO
in real time,
thereby
temperature
inindicator,
an application
changes because
ofsensor
the impactsof
typical
driftachieving
sources, including
thecompensation
leaching of the
LED
where
short-term
in temperature
occur.
Therequires
calibration
curve based
on aintensity
output,frequent
and ageing
of the changes
sol-gel matrix.
Thus, the
sensor
recalibration
after
certain
measurement
because
of the
impactsof
typical than
drift the
sources,
including the
leaching
of the
period, but thechanges
frequency
achieves
much
more progress
electrochemical
sensor.
In addition,
indicator,
LED
output,
and
ageing
of
the
sol-gel
matrix.
Thus,
the
sensor
requires
recalibration
after
the temperature compensation should be processed in the shade to avoid the influence of daylight
aand
certain
period, but
the frequency achieves much more progress than the electrochemical sensor. In
background
fluorescence.
addition, the temperature compensation should be processed in the shade to avoid the influence of
daylight
and background
fluorescence.
3.1.2. Pressure
Compensation
of the Optical DO Sensor
Given that the DO value increases when the sol-gel film is moderately affected by pressure, the
pressure affects the accuracy of the measurement. Pressure compensation is needed in the
30919 [32] is as follows:
measurement of oxygen, and the correction formula

DO CS *

P
101.3

(8)

Sensors 2015, 15, 3091330926

3.1.2. Pressure Compensation of the Optical DO Sensor


Given that the DO value increases when the sol-gel film is moderately affected by pressure,
the pressure affects the accuracy of the measurement. Pressure compensation is needed in the
measurement of oxygen, and the correction formula [32] is as follows:
Sensors 2015, 15, pagepage

DO1 CS

P
101.3

(8)

In the equation, DO denotes


the DO in the pressure of P; CS denotes the normal atmospheric
In the equation, DO1 denotes the DO in the pressure of P; CS denotes the normal atmospheric
pressure value; and P denotes the actual atmospheric pressure value.
pressure value; and P denotes the actual atmospheric pressure value.
CS can
through
the
tableatatdifferent
different
temperatures.
At different
CS be
canobtained
be obtained
through
thenational
national standard
standard table
temperatures.
At different
temperature,
the the
aqueous
solutions
to different
differentpressures.
pressures.
mapped
in Figure
temperature,
aqueous
solutionsare
aresubjected
subjected to
AsAs
mapped
in Figure
7, the7, the
experiment
shows
thatthat
thethe
optical
inthis
thisstudy
study
can
effectively
eliminate
experiment
shows
opticalDO
DOsensor
sensor introduced
introduced in
can
effectively
eliminate
the the
effecteffect
of pressure
andand
that
the
pressure
hasa agood
good
effect.
of pressure
that
the
pressurecompensation
compensation has
effect.

Figure
test.
Figure7.7.Pressure
Pressure compensation
compensation test.

3.1.3.3.1.3.
Salinity
Compensation
ofof
the
Salinity
Compensation
theOptical
OpticalDO
DO Sensor
Sensor
Given
solubility of
of oxygen
decreases
as the
increases,
the DO must
Given
thatthat
thethe
solubility
oxygenininwater
water
decreases
assalt
thecontent
salt content
increases,
the DO
be
compensated
according
to
the
salinity.
When
the
salinity
(i.e.,
expressed
as
the
total
salinity)
is
must be compensated according to the salinity. When the salinity (i.e., expressed as the total salinity)
below
ppt,the
thecorrection
correction formula
given
by:by:
is below
3535ppt,
formula[32]
[32]is is
given

DO
nC
DO2
C
CSS nC
S S

(9) (9)

In the
DODO
denotes
the the
DODO
value
at aatsalinity
of n;
thethe
DODO
value
in in
pure
water;
Inequation,
the equation,
denotes
value
a salinity
of C
n;S denotes
CS denotes
value
pure
n denotes
salinitythe
value;
andvalue;
CS denotes
reduction
of DO according
to salinity
(1 ppt).(1 ppt).
water;the
n denotes
salinity
and CSthe
denotes
the reduction
of DO according
to salinity
CS and
be obtained
throughthe
thenational
national standard
different
temperatures.
The The
CS and
CSC
can
be obtained
through
standardtable
tableat at
different
temperatures.
S can
++, and CO 2
2
aqueous
solution
is
measured
using
KCl,
NH
as
ions
in
the
experiment.
The
ion
density
aqueous solution is measured using KCl, NH44 , and CO3 3 as ions in the experiment. The ion density
of the
measurementsolutions
solutions is
0 and
40 ppt.
values
separatelyseparately
are shown are
of the
measurement
isbetween
between
0 and
40 The
ppt.DO
The
DO measured
values measured
on
the
basis
of
the
experimental
data
in
Figure
8.
The
analysis
of
the
data
shows
that
the
DO
values
shown on the basis of the experimental data in Figure 8. The analysis of the data shows that
the DO
have certain effects when ionic strength regulators are added. The experiment shows that the optical
values have certain effects when ionic strength regulators are added. The experiment shows that the
sensor satisfies the salinity compensation in the aquatic environment. The salinity compensation
optical sensor satisfies the salinity compensation in the aquatic environment. The salinity
parameters are also stored in the calibration TEDS.
compensation parameters are also stored in the calibration TEDS.

30920

aqueous solution is measured using KCl, NH4+, and CO32 as ions in the experiment. The ion density
of the measurement solutions is between 0 and 40 ppt. The DO values measured separately are
shown on the basis of the experimental data in Figure 8. The analysis of the data shows that the DO
values have certain effects when ionic strength regulators are added. The experiment shows that the
optical2015,
sensor
satisfies the salinity compensation in the aquatic environment. The salinity
Sensors
15, 3091330926
compensation parameters are also stored in the calibration TEDS.

Figure 8.
8. Salinity
test.
Figure
Salinity compensation
compensation test.

3.1.4. Excitation
Compensation
3.1.4.
Excitation LED
LED Compensation
The experimental
experimental results
results show
show that
of the
the excitation
excitation LED
The
that different
different intensities
intensities of
LED have
have aa large
large
influence
on
the
fluorescence
intensity
of
the
sensitive
film.
The
intelligent
processing
module
also
influence on the fluorescence intensity of the sensitive film. The intelligent processing module also
allows
for
the
adjustment
of
the
calibration
parameters
to
correct
the
excitation
blue
LED
at
the
time
allows for the adjustment of the calibration parameters to correct the excitation blue LED at the time
of measurement (i.e., if this differs from the excitation at the time of calibration). The luminous flux
of measurement (i.e., if this differs from the excitation at the time of calibration). The luminous flux
8 exponential form at 100 C. When the service
attenuation curve of the blue LED shows a similar
time reaches about 10,000 h, the luminous flux of the blue LED (L470-02) drops to about 70% of
the original luminous flux. Thus, when the LEDs work beyond 10,000 h, the sensor probe must be
replaced. Before replacement, illuminant compensation is required to guarantee the accuracy of the
DO sensor. The experimental results show that this sensor can effectively compensate the effect of
the light source caused by the instability or aging of the light source. The compensation formula is
shown in the following equation, and the calibration parameters are also stored in the TEDS. When
the intensity of the reference blue LED changes, Equation (8) is used to correct the intensity value of
fluorescence quenched by oxygen:
E E

e
e
p0.7
1q
e
e

(10)

In the equation, e denotes the original intensity value of the reference LED; e denotes the
measured intensity value of the reference LED; E denotes the measured intensity value of the
quenched fluorescence; and E denotes the corrected intensity value of the quenched fluorescence.
The long-term stability of the sensor in aquaculture tank water has been continuously studied
for over 12 months. With the passage of time, the sol-gel matrix is prone to losing its original features,
which may cause drifts in calibration, leaching, and bleaching of the dye. As described above, if the
intensity of the fluorescence declines (i.e., after about 12 months), then the sensor probe should be
replaced rather than the whole sensor device with the separate structure.
3.2. Analysis of the Performance of the Sensor
To verify the performance of the sensor, the optical DO sensor is tested in terms of our
aspects: accuracy, stability, precision, repeatability, rapidity. Its superior performance proves that
the intelligent optical DO sensor can be used to monitor the DO content in the aquaculture industry.
The experimental processes are shown below.
3.2.1. Accuracy Test
In the laboratory, the room temperature is about 25 C, and three groups of standard aqueous
solutions are measured: 5.02, 9.98, and 18.03 mg/L. The defined oxygen contents are adjusted by
mixing the oxygen and nitrogen with mass flow controllers. The sample contents are checked using
a commercial DO meter. Compared with reference values, the experimental data in Table 1 show that
30921

Sensors 2015, 15, 3091330926

the relative error is less than 2%, the measurement error is less than 0.2 mg/L, and the resolution of
the optical DO sensor is 0.01 mg/L within the range of 020 mg/L, which conforms to the accuracy
requirements. At the same time, the temperature error is less than 0.5 C, which shows that the
temperature probe is accurate.
Table 1. Accuracy test.
Samples
5.02
9.98
18.03

Measurements
4
5

4.86
10.12
17.85

4.79
10.09
17.83

4.89
10.21
17.91

4.93
9.89
17.87

4.89
10.06
17.96

Absolute Relative
Error
Error

AVG

4.91
9.91
17.87

5.09
9.89
17.93

4.98
10.13
18.03

4.92
10.04
17.91

0.1
0.06
0.12

1.99%
0.60%
0.67%

3.2.2. Stability Test


The response of the optical sensor appears to be stable within the measurement error, which
meets the specified stability requirement of 0.2 mg/L. In the laboratory, the DO content is measured
Sensors 2015,
within
12 h15,inpagepage
steps of 1 h. The standard aqueous solution is adjusted by mixing the oxygen and
nitrogen with mass flow controllers. The experimental results are 2.51, 5.58, 7.22, 10.01, 14.56,
18.1218.12
mg/L.
As shown
in Figure
9, the
largest
error
the
and
mg/L.
As shown
in Figure
9, the
largest
errorisisless
lessthan
than 0.02
0.02 mg/L,
mg/L, which
which meets
meets the
requirements
of
the
optical
dissolved
oxygen
sensor.
requirements of the optical dissolved oxygen sensor.

9. Stability test.
Figure 9.

3.2.3. Precision
Precision Test
Test
3.2.3.
Precision refers
refers to
to the
the discrete
discrete degree
degree of
of the
the values
values that
that are
are repeatedly
repeatedly measured
measured by
by the
the same
same
Precision
method
under
the
same
experimental
conditions
for
the
same
sample.
Generally,
the
relative
method under the same experimental conditions for the same sample. Generally, the relative standard
standard deviation
(RSD)
used tothe
explain
sensor precision.
RSD indicates
that the
deviation
(RSD) is used
to isexplain
sensorthe
precision.
The RSD The
indicates
that the results
areresults
more
are
more
concentrated.
The
standard
solution
contains
Na
2
SO
3
,
and
this
sample
is
continuously
concentrated. The standard solution contains Na2 SO3 , and this sample is continuously measured ten
measured
times. The results
measurement
results
are shown
Table 2. The micro-controller
calculates
times.
The ten
measurement
are shown
in Table
2. Thein
micro-controller
calculates the RSD
every
the measurements,
RSD every ten indicating
measurements,
indicating
that
the precision
of this
group
of data
is the
veryRSD
good
ten
that the
precision
of this
group of data
is very
good
when
is
when
the
RSD
is
less
than
2%.
The
micro-controller
inputs
these
data
into
the
internal
memory.
As
less than 2%. The micro-controller inputs these data into the internal memory. As shown in Figure 10,
shown
in Figure
10, the
sensor has a high precision.
the
results
show that
theresults
sensorshow
has a that
highthe
precision.
Table 2. Precision test.
Sample

10

AVG
Average
average

RSD

5.56

5.62

5.60

5.44

5.49

5.51

5.50

5.82

5.71

5.60

5.59

1.97%

9.96

9.91

9.89

10.22

10.15

10.04

10.00

9.98

10.02

10.15

10.03

1.10%

18.90

18.85

18.86

18.74

18.69

30922
18.56

18.74

18.63

18.60

18.59

18.73

0.64%

measured ten times. The measurement results are shown in Table 2. The micro-controller calculates
the RSD every ten measurements, indicating that the precision of this group of data is very good
when the RSD is less than 2%. The micro-controller inputs these data into the internal memory. As
shown in Figure 10, the results show that the sensor has a high precision.
Sensors 2015, 15, 3091330926

Table 2. Precision test.


Table 2. Precision test.
Sample
A
B
C

Sample
1 5.62 2
5.56

3
5.60

45.44

5 5.49 6

7
5.51

85.50

9 5.8210

10

AVG Average
5.71 average
5.60

A
5.60 5.44
5.51 10.04
5.50 5.82
5.60 10.02 5.59
9.96 5.56
9.91 5.62 9.89
10.22 5.4910.15
10.00 5.71 9.98
10.15
B
9.96 9.91 9.89 10.22 10.15 10.04 10.00 9.98 10.02 10.15
10.03
18.90
18.8518.8518.86
18.74 18.69
18.69
18.74 18.60
18.63
18.59
C
18.90
18.86 18.74
18.56 18.56
18.74 18.63
18.59 18.60 18.73

AVG
Average
average

RSD

RSD
5.59

1.97%

1.97%
10.03
1.10%
18.73
0.64%

1.10%
0.64%

Figure 10.
10. Precision
Precision test.
test.
Figure

3.2.4. Repeatability Test


Na2 SO3 is added into the aqueous solutions 10
to achieve the sample solutions (i.e., 2.51, 9.24, and
12.57 mg/L). The sensor is used to measure the DO values of the sample solutions. The data are
recorded every hour and continuously measured ten times. Table 3 shows that the relative error is
less than 2%. The result of the data analysis indicate that the repeatability meets the requirements of
the optical DO sensor.
Table 3. Stability test.
Samples
2.51
9.24
12.51

2.54
9.22
12.49

2.5
9.19
12.44

2.44
9.16
12.45

Measurement Results
4
5
6
7
2.56
9.13
12.44

2.62
9.33
12.67

2.48
9.25
12.71

2.63
9.23
12.58

10

2.55
9.22
12.55

2.59
9.16
12.49

2.63
9.13
12.51

Relative Error
1.59%
0.43%
0.16%

3.2.5. Speed Test


Two cups of water are prepared in laboratory. 5% Na2 SO3 and a suitable amount of CoCl2 are
added to one of them. They are placed in the air for two days to ensure that both temperatures
are same, which can eliminate the influence of temperature on the measurement results. Series A
indicates the change of DO content from tap water to anaerobic water, and the Series B line indicates
the change from anaerobic water to tap water. From Figure 11, we can see that the response time is less
than 3 min. After 1 min, the Series A is nearly 0 mg/L. However, the response time of Series B is about
3 min. The reason of this phenomenon is that the process of fluorescence quenching must achieve
dynamic balance. Series B should take some time to react, but it doesnt need time in anaerobic water.
The data indicates that detection response speed meets the requirements of the optical DO sensor.

30923

indicates the change from anaerobic water to tap water. From Figure 11, we can see that the
response time is less than 3 min. After 1 min, the Series A is nearly 0 mg/L. However, the response
time of Series B is about 3 min. The reason of this phenomenon is that the process of fluorescence
quenching must achieve dynamic balance. Series B should take some time to react, but it doesnt
need
time
anaerobic water. The data indicates that detection response speed meets the
Sensors
2015,in
15, 3091330926
requirements of the optical DO sensor.

Figure 11. Speed Test.


Figure 11. Speed Test.

3.2.6. On-Site Verification Test


3.2.6. On-Site Verification Test
Figure 12 is the real-time graph of dissolved oxygen changes in aquaculture ponds. After
Figure 12 is the real-time graph of dissolved oxygen changes in aquaculture ponds. After sunrise,
sunrise, phytoplankton produces photosynthesis, which releases a large amount of oxygen. The
phytoplankton produces photosynthesis, which releases a large amount of oxygen. The increase of
increase of oxygen is more than the consumption, so the DO content gradually increases. Over a
oxygen is more than the consumption, so the DO content gradually increases. Over a period of time,
period of time, at the peak of photosynthesis, DO content is at a maximum. After sunset, the
at the peak of photosynthesis, DO content is at a maximum. After sunset, the phytoplankton undergo
phytoplankton undergo respiration which consumes oxygen instead of photosynthesis. In addition,
respiration which consumes oxygen instead of photosynthesis. In addition, with the respiration
with the respiration of aquatic animals, DO content decreases rapidly. After all night, the
of aquatic
animals, DO content decreases rapidly. After all night, the accumulation is reduced to
Sensors
2015, 15, pagepage
accumulation is reduced to a minimum at dawn. These changes perfectly accord with the test data,
a minimum at dawn. These changes perfectly accord with the test data, which proved that the
which proved that the developed DO sensor has a reliable performance. In contrast to change rates
of developed
oxygen consumed
byhas
organisms,
temperature
littletoinfluence
on DO
contentconsumed
in the
DO sensor
a reliablethe
performance.
In has
contrast
change rates
of oxygen
11
on-site
test.
by organisms,
the temperature has little influence on DO content in the on-site test.

Figure
12. 12.
On-site
Test.
Figure
On-site
Test.

4. Conclusions
4. Conclusions
To detect the dissolved oxygen content exactly in aquaculture industry applications, an
To detect the dissolved oxygen content exactly in aquaculture industry applications, an
intelligent optical dissolved oxygen measurement method based on a fluorescent quenching
intelligent optical dissolved oxygen measurement method based on a fluorescent quenching
mechanism has been designed in this paper. The advantages of the unique sensor head based on
mechanism has been designed in this paper. The advantages of the unique sensor head based
fluorescent quenching have been highlighted and intensity measuring circuits and intelligent
on fluorescent quenching have been highlighted and intensity measuring circuits and intelligent
processing have been described. With the optical dissolved oxygen sensor probe in the paper, the
processing have been described. With the optical dissolved oxygen sensor probe in the paper, the
detection of dissolved oxygen became more accurate and interference-free. In view of the influence
detection of dissolved oxygen became more accurate and interference-free. In view of the influence
of temperature, pressure, salinity and blue excitation LED on measurement, this paper has designed
of temperature, pressure, salinity and blue excitation LED on measurement, this paper has designed
an intelligent processing module. The accurate DO content can be achieved by the intelligent optical
DO sensor with the correction coefficients stored in the calibration TEDS. The experiments have
shown that the sensor performance can meet the30924
specifications for the monitoring DO content in
aquaculture industry, such as low-power consumption, fast response, ability to detect online, high
accuracy and strong stability, which will enable reliable operation for periods of months at the very

Sensors 2015, 15, 3091330926

an intelligent processing module. The accurate DO content can be achieved by the intelligent optical
DO sensor with the correction coefficients stored in the calibration TEDS. The experiments have
shown that the sensor performance can meet the specifications for the monitoring DO content in
aquaculture industry, such as low-power consumption, fast response, ability to detect online, high
accuracy and strong stability, which will enable reliable operation for periods of months at the very
least. In addition, the use of an optical measurement method and the intelligent calibration technique
could be applicable in other medical and environmental monitoring areas.
Acknowledgments: Funds for this research was supported by Special Fund for Agro-scientific Research in the
Public Interest, Modern Fishery Digitization and Integration & Demonstration of Internet of Things Technology
(201203017).
Author Contributions: Study concept and design: Yaoguang Wei, Daoliang Li and Yingyi Chen; data acquisition:
Fengmei Li and Xu Zhang; statistical analysis: Xu Zhang and Fengmei Li; writing of the manuscript: Fengmei Li
and Xu Zhang; study supervision: Daoliang Li, Yaoguang Wei and Yingyi Chen.
Conflicts of Interest: The authors declare no conflict of interest.

References
1.

2.

3.
4.

5.

6.

7.
8.
9.

10.

11.
12.
13.
14.

Gao, M.; Zhang, F.; Tian, J. Dissolved oxygen online acquisition terminal based on wireless sensor network.
In Proceedings of the 2008 4th International Conference on Wireless Communications, Networking and
Mobile Computing (WiCOM), Dalian, China, 1214 October 2008; pp. 14.
Niu, C.; Zhang, Y.; Zhou, Y.; Shi, K.; Liu, X.; Qin, B. The Potential Applications of Real-Time Monitoring
of Water Quality in a Large Shallow Lake (Lake Taihu, China) Using a Chromophoric Dissolved Organic
Matter Fluorescence Sensor. Sensors 2014, 14, 1158011594. [CrossRef] [PubMed]
Chong, S.; Aziz, A.; Harun, S. Fibre Optic Sensors for Selected Wastewater Characteristics. Sensors 2013, 13,
86408668. [CrossRef] [PubMed]
Wilhelm Filho, D.; Torres, M.A.; Zaniboni-Filho, E.; Pedrosa, R.C. Effect of different oxygen tensions on
weight gain, feed conversion, and antioxidant status in piapara, Leporinus elongatus (Valenciennes, 1847).
Aquaculture 2005, 244, 349357. [CrossRef]
Foss, A.; Vollen, T.; iestad, V. Growth and oxygen consumption in normal and O2 supersaturated water,
and interactive effects of O2 saturation and ammonia on growth in spotted wolffish (Anarhichas minor
Olafsen). Aquaculture 2003, 224, 105116. [CrossRef]
Wang, J.; Bian, C.; Tong, J.; Sun, J.; Li, Y.; Hong, W.; Xia, S. Modification of Graphene on Ultramicroelectrode
Array and Its Application in Detection of Dissolved Oxygen. Sensors 2015, 15, 382393. [CrossRef]
[PubMed]
Ramamoorthy, R.; Dutta, P.K.; Akbar, S.A. Oxygen sensors: Materials, methods, designs and applications.
J. Mater. Sci. 2003, 38, 42714282. [CrossRef]
Dai, W.Y.; Sun, L. The Measurement Methods of Dissolved Oxygen in Water. Anhui Agri. Sci. Bull. 2007, 13,
7779.
Novic, M.; Pihlar, B.; Dular, M. Use of flow injection analysis based on iodometry for automation of
dissolved oxygen (Winkler method) and chemical oxygen demand (dichromate method) determinations.
Fresenius Z. Anal. Chem. 1988, 332, 750755. [CrossRef]
Zhao, Y.; Sun, L.; Li, F.M. The research about detection of dissolved oxygen in water based on C8051F040.
In Proceedings of the 2009 International Conference on Information Engineering and Computer Science,
ICIECS 2009, Wuhan, China, 1920 December 2009; pp. 14.
Landman, M.J.; van den Heuvel, M.R. An improved system for the control of dissolved oxygen in
freshwater aquaria. Water Res. 2003, 37, 43374342. [CrossRef]
Jalukse, L.; Leito, I.; Mashirin, A.; Tenno, T. Estimation of uncertainty in electrochemical amperometric
measurement of dissolved oxygen concentration. Accred. Qual. Assur. 2004, 9, 340348. [CrossRef]
Hasumoto, H.; Imazu, T.; Miura, T.; Kogure, K. Use of an optical oxygen sensor to measure dissolved
oxygen in seawater. J. Oceanogr. 2006, 62, 99103. [CrossRef]
Helm, I.; Jalukse, L.; Vilbaste, M.; Leito, I. Micro-Winkler titration method for dissolved oxygen
concentration measurement. Anal. Chim Acta 2009, 648, 167173. [CrossRef] [PubMed]

30925

Sensors 2015, 15, 3091330926

15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.

29.

30.
31.
32.

Li, A.; Wang, Y.; Hu, Q.; Shieh, W. Few-mode fiber based optical sensors. Opt. Express 2015, 23, 11391150.
[CrossRef] [PubMed]
Deepa, N.; Balaji Ganesh, A. Sol-gel based portable optical sensor for simultaneous and minimal invasive
measurement of pH and dissolved oxygen. Measurement 2015, 59, 337343. [CrossRef]
Eich, S.; Schmlzlin, E.; Lhmannsrben, H. Distributed Fiber Optical Sensing of Oxygen with Optical Time
Domain Reflectometry. Sensors 2013, 13, 71707183. [CrossRef] [PubMed]
Zheng, G.L.; Xu, Z.W. A new design of high-precision dissolved oxygen sensor. Transducer Microsyst.
Technol. 2012, 112114.
McDonagh, C.; Kolle, C.; McEvoy, A.K.; Dowling, D.L.; Cafolla, A.A.; Cullen, S.J.M. Phase fluorometric
dissolved oxygen sensor. Sens. Actuators B Chem. 2001, 74, 124130. [CrossRef]
Zhao, S.Y.; Harrison, B.S. Morphology impact on oxygen sensing ability of Ru(dpp)3 Cl2 containing
biocompatible polymers. Mater. Sci. Eng. C 2015, 53, 280285. [CrossRef] [PubMed]
Lee, S.; Ibey, B.L.; Cot, G.L.; Pishko, M.V. Measurement of pH and dissolved oxygen within cell culture
media using a hydrogel microarray sensor. Sens. Actuators B Chem. 2008, 128, 388398. [CrossRef]
Chu, C.; Chuang, C. Optical fiber sensor for dual sensing of dissolved oxygen and Cu2+ ions based on
PdTFPP/CdSe embedded in sol-gel matrix. Sens. Actuators B Chem. 2015, 209, 9499. [CrossRef]
Hartmann, P. Photochemically Induced Energy-Transfer Effects on the Decay Times of Ruthenium
Complexes in Polymers. Anal. Chem. 2000, 72, 28282834. [CrossRef] [PubMed]
Feng, W.W.; Zhou, N.; Chen, L.X. An optical sensor for monitoring of dissolved oxygen based on phase
detection. J. Opt. 2013, 15, 5550255507. [CrossRef]
Ding, Q.S.; Ma, D.K.; Li, D.L. Research and application on compensation and calibration methods for smart
sensor of dissolved oxygen. J. Shandong Agric. Univ. 2012, 42, 567571.
Sakaue, H.; Ozaki, T.; Ishikawa, H. Global Oxygen Detection in Water Using Luminescent Probe on
Anodized Aluminum. Sensors 2009, 9, 41514163. [CrossRef] [PubMed]
Woods, S.P.; Lee, K.; Bryzek, J. An overview of the IEEE-P1451.2 smart transducer interface module. Analog
Integr. Circuits Signal Process. 1997, 14, 165177. [CrossRef]
Shao, H.; Zhuang, J. Research into the Technology of Intelligent Sensors Based on IEEE1451 Standard.
In Proceedings of the 2012 International Conference on Industrial Control and Electronics Engineering
(ICICEE), Xian, China, 2325 August 2012; pp. 17971799.
Petru, P.; Mircea, B. Hardware implementation of a PIC18F448 based TIM for IEEE1451.2 compliant actuator
control. In Proceedings of the 2010 9th International Symposium on Electronics and Telecommunications
(ISETC 2010), Timisoara, Romania, 1112 November 2010; pp. 119122.
Kumar, A.; Singh, I.P.; Sud, S.K. Energy Efficient and Low-Cost Indoor Environment Monitoring System
Based on the IEEE 1451 Standard. IEEE Sens. J. 2011, 11, 25982610. [CrossRef]
Ding, Q.S.; Tai, H.J.; Ma, D.K. Development of a Smart Dissolved Oxygen Sensor Based on IEEE1451.2.
Sens. Lett. 2011, 9, 10491054. [CrossRef]
Lv, B.; Lei, Z.; Liu, J.; Qu, J.L. Design of a PIC18 F2520 based polarographic dissolved oxygen sensor.
Shandong Sci. 2012, 25, 7377.
2015 by the authors; licensee MDPI, Basel, Switzerland. This article is an open
access article distributed under the terms and conditions of the Creative Commons by
Attribution (CC-BY) license (http://creativecommons.org/licenses/by/4.0/).

30926

You might also like