You are on page 1of 8

Nama Penyedia PT.

Enseval Medika Prima


Alamat Penyedia Jl. Pulo Lentut no. 10, Kawasan Industri Pulo Gadung, Jakarta Timur
Lokasi Pabrik (Provinsi) -
Merk bioMerieux, France
Periode 01 Januari 2014 s/d 31 Maret 2014

No Nama Produk Type


2 Automated Microbiology analyzer Vitek 2 Compact
for ID & AST Bacteri and yeast Kode : 27560
an Industri Pulo Gadung, Jakarta Timur

Merk/Asal negara
bioMerieux
France
, Jakarta Timur

Spesification
Instrument
Fully Automated Instrument for Identification and suscetibility test for bacteria and yeast

Test Kit / Result Integrity


Pre bar coded test kit.
No addition of reagents during sample preparation.
No addition of reagents to test kit in order to obtain final result.
No biohazard confirmed sealed unit no risk of opening and contaminating reader.
Software that verifies the consistency of the identification and susceptibility result.
Report susceptibility data for both tested and deduced antibiotics.
The system provides bacterial phenotype-based ID/AST result validation. The knowledge
The knowledge base for result validation is located on the ID/AST testing computer
The system can deliver accurate, validated identifications for 80% of routine in 6 hours.
Susceptibility test kits cover MIC ranges in excess of those recommended by CLSI
System has the ability to do separately ID and AST
Weight of disposable unit (card) and ancillaries is less than 3 oz.

Dimension / Weight
Height : 60 Cm
Width : 72 Cm
Depth : 68 Cm
Weight : Approx 75 Kg

Environment
Indoor use
Do not install in direct sunlight
Bench top leveled to + 3 degrees
Bench top clearance 11cm side and 21 cm above
Relative humidity: 20% to 80% non-condensing
Operating temperature: 15C to 30C
Altitude: -100 m to 2000 m sea-level
Shipment and storage temperature: -20 to +55C

Electrical Characteristics
Input Voltage : 200/240 VAC @ 50/60 Hz
Max Input Current : 2.5 Amps @ 240 VAC
Nominal Input Current : 1.25 Amps @ 240 VAC
Power : 300 Watts nominal, 600 watts peak
Colorimetric Optics
Emission Wavelength : 430 nm
Detection Wavelength : 568 nm for 4 GP tests
Measurement Precision : +/- 0.5 %
Lamp Life : = 80.000 cards

GENERAL CHARACTERISTICS
Cassette
Capacity, cards : 10 Cards per cassette

Dispenser
Dinspensing Volume : 3.0 ml
Diluent Capacity : 1.000 ml, user replaceable

Pipettors
Dinspensing Volume : 145 ul to 300 ul
280 ul to 300 ul

Incubator
Temperature : 35.5 C 1 C Average
Capacity : Max 60 Cards = VITEK 2 Compact 60
No. AKL
KEMKES RI AKL
20301010572