Professional Documents
Culture Documents
User Manual
Bruker Elemental Suite OES User Manual
Content Overview
0.01 Release Notes .............................................................................................................................. 2
0.02 General Information .................................................................................................................... 6
0.03 Login ............................................................................................................................................ 7
0.04 Software structure....................................................................................................................... 8
1. Analysis Plugin ..................................................................................................................................... 9
1.01 Analysis table ............................................................................................................................... 9
1.02 Hotkeys ...................................................................................................................................... 10
1.03 Activate a method ..................................................................................................................... 11
1.04 Enter ID Information.................................................................................................................. 12
1.05 Performing a measurement ...................................................................................................... 13
1.06 Deleting a measurement ........................................................................................................... 14
1.07 Complete an analysis ................................................................................................................. 15
1.08 Workflow: Standardization........................................................................................................ 16
1.09 Workflow: Type standardization ............................................................................................... 19
1.10 Workflow: Control ..................................................................................................................... 22
1.11 Workflow: PMI........................................................................................................................... 24
1.12 Workflow: Maintenance............................................................................................................ 25
1.13 Workflow: SORT ........................................................................................................................ 27
1.14 Workflow: Stdz. Replacement ................................................................................................... 29
1.15 Settings: Analysis completion settings ...................................................................................... 32
1.16 Settings: Publishing Settings...................................................................................................... 33
1.17 Settings: Edit Limit Cards for Qualities ...................................................................................... 36
1.17.1 Add new Limitcard.............................................................................................................. 37
1.17.2 Edit Limitcards .................................................................................................................... 38
1.17.3 Duplicate Limitcards ........................................................................................................... 38
1.17.4 Local Standards................................................................................................................... 39
1.18 Settings: Edit Sorting Settings ................................................................................................... 40
1.19 Settings: Standardization Status................................................................................................ 42
1.20 Reset Daily Counter ................................................................................................................... 43
1.21 Switch User ................................................................................................................................ 43
2. Development Plugin .......................................................................................................................... 44
2.01 Methods editor overview .......................................................................................................... 44
2.02 Change Language....................................................................................................................... 45
2.02 Change skin................................................................................................................................ 46
DOC-ES-CCD-UM V1.0 02.2016 Seite 3 von 80
Bruker Elemental Suite OES User Manual
Thank you for choosing a product for metal analysis from Bruker Elemental GmbH.
The software Bruker Elemental Suite OES is part of your new OES analytical system.
For the configuration of the data like alloys, control samples, type standards or report you can use
the Development Plugin, which is topic of the chapter 2. For instruments with a touch screen, you
will need an external mouse and keyboard to perform the configuration.
If you see strange icons on the screen in most of the cases a mouse over (bringing the mouse pointer
over the strange icon) will open a box with detailed information.
0.03 Login
First of all in each menu you will find the button to save in the top left corner
The magnifying glass allows you to increase the size of the window to full screen.
The page with the magnifying glass reduces the size of the window.
Every button with a blue hook has a sub menu (might be not available for all users).
With a click on the right mouse button on the left side of a table or on the top row you can open a
submenu if available.
With important changes in the database you will be asked to store the changes. A new MDS will be
created. Please fill in the comment box to find old restore points later.
1. Analysis Plugin
2: Sample identification
5: Analysis table
1.02 Hotkeys
Please be aware that the following hotkeys work on all instruments that are supplied with an
external keyboard. This includes also those instruments with a touchscreen that have an additional
external keyboard in addition to the screen-keyboard.
Hotkey Function
F2 Start measurement
F3 Stop measurement
F4 Finish analysis
F5 Clear screen
Ctrl-M Start Multiple measurements
Ctrl-F Toggle argon flush on/off
Ctrl-I Enter sample ID
Alt-F Switch off all Argon flows
In your OES System we use different methods to analyze different materials or alloys. To give you the
best results please always use the method that fits your alloy the best.
To activate a method please press the button on the bottom left corner marked as.
A Menu will open that shows you all of the existing methods.
On the top of the box you can select the Basic element of your
material by clicking on the chemical expression.
With the result of the sample you might want to store some details about the sample (e.g. furnace,
material number, date, customer name, etc.). You can find that information in the top row above the
analyses grid.
If you filled in all necessary information you can click on the disc-symbol
Depending on the type of instrument you have you will find a button or trigger to start and stop a
measurement. Please check your instrument manual to find out which button will be used for this
purpose.
In addition you can start the measurement with the button in the software which can be found in the
bottom on the right.
1 Element symbol
6 Single measurements
If you want to delete single measurements that do not match the others due to bad burn spot or bad
sample you can delete it.
Just click with the right mouse button on the number of the measurement in the header and a
dialogue will open that allows deleting a single measurement.
On the bottom underneath the analyses grid you will find the buttons for the completion of an
analysis.
With the button Complete Analysis defined actions like storing, printing and
deleting will be performed. (See 1.14 Settings Analysis Completion Settings)
By using the button Report Analysis you will be entering a new Menu.
Here you can print the analysis table like it is seen on the
screen, copy the results to the clipboard to use it in other
programs or save it in .txt or .xlsx Format.
This allows you to store each result in an own file for further
use.
With the last button Reject Analysis all measurements on the screen will be
deleted.
First open the Workflow menu by using the button in the top right
corner of the analyses plugin and select the Standardization.
You can also open a menu for the Standardization by using the
button which can be found in the lower right corner.
Click on Yes.
Place the requested sample on the spark stand and start the
measurement.
After the measurement you will see a box with a green hook and the
information, that the X-Shift is okay.
If there is a problem you will see a red cross and the information that the X-
Shift limits are exceeded. In this case Measure again. If X-Shift fails again
please read chapter 1.12 Workflow: Maintenance.
Automatically the name of the next sample that should be sparked appears in the Sample ID
information.
First open the Workflow menu by using the button in the top right
corner of the analyses plugin and select the Type-Standardization.
In the information row the display changes from Element Concentration to Type Stdz Element
Concentration and an additional box with the name of the active type standard appears.
First open the Workflow menu by using the button in the top right
corner of the analyses plugin and select Control.
A box with a green hook is displayed to show that the measurement was
successful.
In the box after completing analysis there will be a red cross to indicate
it was not successful. In this case you will not see a new timestamp in
the list.
First open the Workflow menu by using the button in the top right
corner of the analyses plugin and select PMI.
You can also open it by using the button in the bottom right
corner.
On the TOP you will get the name of the grade that matches the measurement and a Matchcode
that represents the reliability of the matching grade.
Violations of the recommended quality will be marked in yellow (for a violation of the minimum) or
in red (for a violation of the maximum).
To check the correct function of the optic you will have to perform a drift
correction before standardization.
You can choose to perform this drift correction also manually in the
Workflow: Maintenance.
Depending on the Matrix you are currently using, you will be asked
by the system to measure the pure sample. The name of the
sample will also appear in the Sample Identification field.
After the measurement you will see a box with a green hook and the
information, that the X-Shift is okay.
If there is a problem you will see a red cross and the information that the X-
Shift limits are exceeded.
- Pollution of the spark stand/probe head with other material (e.g. Cu material when measuring pure
Al-sample; mainly in Multi-Matrix-Instruments)
- Bad spark spot (e.g. Argon switched off, oxidized sample surface, humidity in the spark stand/ probe
head or on the sample surface)
- Bad argon atmosphere (in instruments with UV-Optic, the atmosphere is surveyed during the X-
Shift as well)
In those cases solve the problem and repeat the X-Shift measurement again.
Sometimes it is enough to spark 4-5 times on the pure sample before performing the X-Shift.
First open the Workflow menu by using the button in the top right
corner of the analyses plugin and select the SORT.
Read chapter 1.18 Settings: Edit Sorting Settings for further details.
On the bottom of the screen you will find a counter of how many samples were checked and how
many failed the sorting.
A window opens, that allows you to select the sample which should
be replaced.
From a drop down menu you can select the sample to be replaced
from the currently used samples. You can only replace one sample
at a time.
The name of the selected sample will be used for the new sample. It
can be changed to the name on the sample you want to use in the
future. If the names are identical you cant go on.
Now change the name of the new sample. Click on Save & Close.
As the old sample might be required for future adjustments or as the standardization replacement
might have failed, please do not remove the sample from the database.
6: Type of file that should be created (one table for all results or single file for each result)
12: After publishing an external program can be started automatically to transfer the data from the
.csv to your SPC.
With the Insert Meta Data Columns you can select additional
information such as sample identification as separate columns.
At the end you can change the order of the columns by moving
columns up and down in the content table.
If you want to change the limit cards or add new limit cards to
the system you can use the Settings menu.
2: Alarm limit low and high (used as external limits; e.g. for tolerances)
4: Target value
With a right mouse click on the front bar you can open a menu
that allows you to add, edit or duplicate limit cards.
The export function allows you to export the limit cards to a file
(e.g. a .csv-file or .txt-file)
From a drop down list you can select the elements to be used in
the limit card.
After entering the low and high values you get a new row for the next element.
Here you can change the values of the element settings or add new elements to the limit card.
The Duplicate Limitcard is used if you want to add limitcards similar to existing ones or if you want
to use different copies of a limitcard for incoming material, production and final product.
The menu Set this as used local Standard activates this name to be used in the sample
identification information in the analysis plugin. For those limitcards which do not have a separate
local name, the limit card name will be used. The activated column with local names will be displayed
in bold letters.
The SORT program allows you to compare samples with a reference sample in a quick way. Here only
a few elements will be observed and deviations are tolerated. This is used for example for incoming
material check to verify that all items have the same composition.
Initially there is no sorting program preinstalled, as it highly refers to the requirements of the
analyzed material.
After entering the properties of the program it will be shown in the list.
A click on the + in front of the program name will open the details.
Enter the allowed Relative Deviation between reference sample and sorted samples for each
chemical element. This is not the deviation on the concentration but on the intensity ratios.
In the new window you can see the standardization table for the active method. That means only
those analytical channels are displayed, that are activated in the method. You can check the Factor
and the offset to find irregularities. The Standardization Status of all used lines for the matrix can be
found in the Methods Editor (See chapter 2.04 START: Standardization Status).
If you see odd factors for a channel you can see details
by clicking on the channels name and then on
You will be asked if you want to reset the daily counter first.
Measurements (Counter)
should show 0. The
Measurements (Total) cant
be reset.
2. Development Plugin
1: Program Menu (change language and skin; load old restore points)
3: Configuration tab
4: Active matrix element (some changes are only valid for one base element)
6: Configuration menu
The Skin Bruker Light Blue might cause some problems with some resolution settings.
Every time you change something on the methods a new MDS (restore point) is created. This helps to
undo changes you did by accident.
To activate and older MDS file just click on the MDS you want to restore
and use the button Activate.
As there is only a limited number of MDS files allowed, the oldest MDS will be deleted if the
maximum is reached. To save MDS from being deleted you can protect important restore points.
1: Sample ID Key (String which is used in the database for the ID field)
5: Enable/disable visibility in the Sample ID form (automatic filled ID fields can be disabled here)
By default all new ID fields allow are configured for Manual Input.
(This Macro will not be executed if you click into the ID information
bar)
The Lookup List (fix) allow you to create a list of different values that
have to be entered regular. (e.g. furnace number, supplier, customer,
etc.)
Type in the Values you want to use for the ID field. Each
row gives you a new selection for the Sample ID form.
Type in the Values you want to use for the ID field. Each
row gives you a new selection for the Sample ID form.
The Lookup List from file (fix) allow you to use a list of different
values that have to be entered regular from a file on the computer.
(e.g. furnace number, supplier, customer, etc.) The file should have a
separate row for each new value and should be in a Text-format.
The Lookup List from file (editable) allow you to use a list of different
values that have to be entered regular from a file on the computer.
(e.g. furnace number, supplier, customer, etc.) The file should have a
separate row for each new value and should be in a Text-format.
Values that are not used that often can still be entered manually.
With a right mouse click on the front of each row you can open a menu
that allows you to create new reports, duplicate existing reports or open
the report designer.
7: ID Fields
The design element bar (1) gives you the option to include new text fields or pictures in the
report design.
You can either click into the field on the report design that you wish to
change or chose it from the BEReportA4Portrait menu in the content
element box (3)
In the Content elements list (3) you can select from the
tblReport List the values that should be filled into the
report (8) from the analysis grid.
The content elements will be moved into the report design by drag and drop (hold the left mouse
button while moving an element onto the report field (2). Violations in the syntax due to overlapping
contents will be indicated with red color.
If you click on the BEReportA4Portrait in the content element menu (3) you will find the box
Behavior in the design configuration (4). Here you can change the name of the Report Design.
With a click on Edit Limit Cards you will see the full list of Limit
Cards for the active Base.
For further information see chapter 1.17 Settings: Edit Limit Cards
for Qualities.
The function Export Limit Cards gives you the option to export
limitcards for the use on another Bruker Elemental OES System or
as a backup.
If you want to import data from a different file type you change
the view from Quality Export Format to All Files you will be
able to find your own quality files as well.
Select the file that contains your quality details and click on
Open.
With a right mouse click on the column header you can assign the
columns to a header from the list.
A control sample allows you to observe the accuracy of the instrument. It is recommended to use a
sample matching the produced material.
Enter the Low and high limits for all chemical elements that should be observed.
In the overview of the samples you can click on the + to see the details for the control sample.
Select Methods in the Methods Editor. You will see a list of the methods for the active base
element.
You can select the control sample from the drop down menu.
The control sample has been assigned to the method. The Workflow: Control is now available in the
Workflow Menu for this method.
A type-sample allows you to put a correction on the accuracy for a special alloy. It is recommended
to use a sample matching the produced material.
Enter the low limits, high limits and target values for all chemical elements that should be observed.
In the overview of the samples you can click on the + to see the details for the control sample.
In the Type-Standardization
Settings you can set an expire
point of the validation for the
control samples by time or by
number of measurements.
Select Methods in the Methods Editor. You will see a list of the methods for the active base
element.
You can select the type-sample from the drop down menu.
The control sample has been assigned to the method. The Workflow: Type standardization is now
available in the Workflow Menu for this method.
A detailed instruction can be found in chapter 1.18 Settings: Edit Sorting Settings
To add new Users to the system you need to open the Manage
Usergroups and Users menu.
With Add new User you can create a new user for the login screen.
With Edit User you can change the password of the existing users.
The Login Name has to be filled in. The information for First
Name, Last Name and Password are optional.
With a right mouse click on the field User Names you can
Add new Group.
Enter the Name for the Usergroup and click on Save & Close
In the tab Methods You can see the list of the available methods for the active base element.
With Add new Method you can create a new method that is
completely empty.
Clicking on the + in front of the methods name will open the details.
In the tab Sequences You will find the measurement parameters. The prespark is the sequence
starting with P. For wire or metal sheet programs the prespark might be too long.
Changing the prespark time here will also change the Expected Total Duration. Please talk to your
responsible service contact to find the perfect duration for the prespark for your application.
2: Included in Matrix Sum (bold elements have to be activated due to calculation of influences)
4: Unit (% or ppm)
5: Fix the Numeric Format (by default the numeric format is set to dynamic which means the systems
sets the number of decimals depending on the value)
7: Restriction to the maximum absolute standard deviation for the calculation of the average
8: Restriction to the maximum relative standard deviation for the calculation of the average
If you use the right mouse click on the front bar you can add new
chemical elements manual or from the channel list. Also you can
change the order of the elements in the analysis grid or reset to
the default order.
With a right mouse click on the front bar of the details for a
method you can find the option Add new Expression
Enter the name for the expression. In the example the carbon
equivalent is used as CEQ.
Chemical elements will be entered into the expression box (2) with a double mouse click on the
element in the element box (3). Arithmetic operators and numbers will be entered with the
keyboard.
At the end the new expression can be tested. Enter any Test Value in
the top row and click on Test. The Test Value will be used in the
expression for each element concentration.
In the View menu you will find the option Toggle Grid
Layout.
In the View menu you will find the option Chose visible
Items.
Now you can select the columns that should be shown in the analysis grid. A
+ adds the column to the view and a - removes the column from the
view.
3: Alarm limits from the active limit card (low and high)
4: Warning limits from the active limit card (low and high)
9: Single measurements
3. External Programs
In the Windows Start Menu you can find the Bruker Elemental
OES Folder. In Tools you can find the Backup and Restore
With the program you can create a full backup of the data.
You can store the created 3 files on external device for safety.
Do the same with the CCD-Backup and the Results-Backup. There no additional window will be
opened.
3.02 CCDServer
If the Elemental Suite is closed the CCDServer to communicate with the CCD instruments such as Q2,
Q4 or Q4m is not closed automatically.
In the task menu from windows you can find a mechanical wheel as a
symbol for the CCDServer.
A right mouse click on the wheel will open a menu. Here you can
Shutdown CCD-Server.
You have to confirm that you want to shut down the CCD Server. This
should only be done if the Elemental Suite was shut down before.