You are on page 1of 8

Chapter 26

Chemical Analysis: Moisture Measurement


D. B. Meadowcroft; edited by I. Verhappen

26.1 Introduction the waste of overdrying, since the moisture would be picked
up again during storage. Finally, many granulated foods
The measurement and control of the moisture content of must have a defined moisture content.
gases, liquids, and solids is an integral part of many indus- The purpose of this chapter is to introduce the reader to
tries. Numerous techniques exist, none being universally the major measurement techniques that are available. The
applicable, and the instrument technologist must be able three statesgas, liquid, and solidwill be treated sepa-
to choose the appropriate measurement technique for the rately. In addition, many commercial instruments measure
application. It is particularly important to measure moisture some parameter that changes reproducibly with moisture
because of its presence in the atmosphere, but it is awkward concentration, and these instruments must be regularly cali-
because it is a condensable vapor that will combine with brated by the user. The chapter therefore ends with a discus-
many substances by either physical adsorption or chemical sion of the major calibration techniques that the average user
reaction. Moisture measurement may be needed to ensure must be willing to employ when using such instruments.
that the level remains below a prescribed value or within a First, it is necessary to clarify a further aspect of mois-
specified band, and the range of concentrations involved can ture measurement that can confuse the newcomer, which is
be from less than one part per million to percentage values. to define the large number of units that are used, particularly
A few examples will illustrate the range of applications. for gases, and show how they are interrelated.

GasesIn gas-cooled nuclear reactors, the moisture level of


the coolant has to be within a prescribed band (e.g., 250500
volume parts per million) or below a certain value (e.g., 10
26.2Definitions
vppm) depending on the type of reactor. Rapid detection of
26.2.1 Gases
small increases due to leaks from the steam generators is
also essential. Moisture must be excluded from semicon- Although water vapor is not an ideal gas, for most hygrom-
ductor device manufacture, and glove boxes are fitted with etry purposesand to gain an understanding of the units
moisture meters to give an alarm at, say, 40 vppm. Environ- involvedit is sufficient to assume that water vapor does
mental control systems need moisture measurement to con- behave ideally. The basic unit of moisture in a gas against
trol the humidity, and even tumble dryers can be fitted with which other units can readily be referred is vapor pressure,
sensors to automatically end the clothes drying cycle. and Daltons law of partial pressures can be assumed to hold
if the saturated vapor pressure is not exceeded.
Liquids The requirement is usually to ensure that the water In environmental applications, the unit thats often
contamination level is low enough. Examples are the preven- used is relative humidity, which is the ratio in percent of
tion of corrosion in machinery, breakdown of transformer the actual vapor pressure in a gas to the saturation vapor
oil, and loss of efficiency of refrigerants or solvents. pressure of water at that temperature. It is therefore tem-
perature dependent but is independent of the pressure of
Solids Specified moisture levels are often necessary for the carrier gas.
commercial reasons. Products sold by weight (e.g., coal, ore, For chemical measurements, the concentration of mois-
tobacco, textiles) can most profitably have moisture contents ture is usually required. The volume concentration is given
just below the maximum acceptable limit. Some textiles and by the vapor pressure of moisture divided by the total pres-
papers must be dried to standard storage conditions to pre- sure, often multiplied by 106 to give volume parts per mil-
vent deterioration caused by excessive wetness and to avoid lion (vppm). The concentration by weight in wppm is given
2010 Elsevier Inc. All rights reserved.
doi: 10.1016/B978-0-7506-8308-1.00026-7 429
430 PART | III Measurement of Temperature and Chemical Composition

by the volume concentration multiplied by the molecular It shouldbe noted that below 0C the dewpoint and frost-
weight of water and divided by that of the carrier gas. Mete- point differ. It is possible for supercooled water to exist
orologists often call the weight concentration the mixing below 0C, which can give some ambiguity, but this is
ratio and express it in g/kg. unlikely very much below 0C (certainly not below -40C).
When the prime aim is to avoid condensation, the appropri- In addition, it can be seen that the saturated vapor pressure
ate unit is the dewpoint, which is the temperature at which the increases by an order of magnitude every 1520 degrees so
vapor pressure of the moisture would become saturated with that in the range -80C to 50C dewpoint, there is a vapor
respect to a plane surface. Similarly, the frostpoint refers to pressure change of five orders of magnitude. Table 26.1 lists
the formation of ice. The relationship among dewpoints, frost- the vapor pressure for dew- or frostpoint between -90C
points, and saturated vapor pressure is derived from thermo- and +50C.
dynamic and experimental work and is shown in Figure 26.1. Table 26.2 gives the interrelationships between these
various units for some typical values.

26.2.2 Liquids and Solids


Generally, measurements are made in terms of concentra-
tion, either as a percentage of the total wet weight of the
sample (e.g., in the ceramics industry for clay) or of the
dry weight (e.g., in the textile industry, where the moisture
concentration is called regain). In addition, if a liquid or
solid is in equilibrium with the gas surrounding it, the equi-
librium relative humidity of the gas can be related to the
moisture content of the solid or liquid by experimentally
derived isotherms (e.g., Figure 26.2) or by Henrys law
for appropriate unsaturated liquids. For liquids that obey
Henrys law, the partial vapor pressure of the moisture P is
related to the concentration of water dissolved in the liquid
by W = KP, where K is Henrys law constant. K can be
derived from the known saturation values of the particular
liquid, that is, K = WsPs, where Ws and Ps are, respectively,
Figure 26.1 The relationship between saturation vapor pressure and saturation concentration and saturation vapor pressure at a
dewpoint and frostpoint temperatures. given temperature.

Table 26.1 The relationship between dew/frostpoint and vapor pressure (n bar, which is equivalent to vppm at
1 bar total pressure)
Frostpoint (C) Saturated vapor Frostpoint (C) Saturated vapor Dewpoint (C) Saturated vapor
pressure (n bar) pressure (n bar) pressure (n bar)
90 0.10 40 128 0 6,110
80 0.55 36 200 4 8,120
75 1.22 32 308 8 10,700
70 2.62 28 467 12 14,000
65 5.41 24 700 16 19,200
60 10.8 20 1,030 20 23,400
56 18.4 16 1,510 25 31,700
52 30.7 12 2,170 30 41,800
48 50.2 8 3,100 40 73,000
44 81.0 4 4,370 50 120,000
Chapter | 26 Chemical Analysis: Moisture Measurement 431

Figure 26.3) using a small mirror for which the tempera-


Table 26.2 Some examples of the relationships ture can be controlled by a built-in heater and thermoelec-
between the various units for moisture in gases tric cooler. The temperature is measured by a thermocouple
Dew/frost Vapor pressure RH at 20C Mixing ratio or platinum resistance thermometer just behind the mirror
point (C) (n bar or vppm ambient (%) in air (g/kg) surface, and the onset of dew is detected by the change of
at 1 bar) reflectivity measured by a lamp and photocell. A feedback
70 2.5 0.01 1.5 103 circuit between the cell output and the heater/cooler circuit
45 72 0.3 0.045 enables the dewpoint temperature to be followed automati-
20 1,030 4.4 0.64 cally. Systematic errors can be very small, and such instru-
ments are used as secondary standards, yet with little loss of
0 6,110 26 3.8
sophistication they can be priced competitively for labora-
10 12,300 53 7.6
tory and plant use. Mirror contamination can be a problem
20 23,400 100 14.5 in dirty gases, and in some instruments the mirror is periodi-
cally heated to reduce the effect of contamination. Condens-
able carrier gases, which condense at similar temperatures to
the moisture, invalidate the technique. It is an ideal method
if the dewpoint itself is required, but if another unit is to
be derived from it, accurate temperature measurements are
essential because of the rapid change in vapor pressure with
dewpoint temperature (see Section 26.2.1).

26.3.1.2 Coulometric Instruments


The gas is passed at a constant rate through a sampling tube
in which the moisture is absorbed onto a film of partially
hydrated phosphoric anhydride (P2O5) coated on two plati-
num electrodes (see Figure 26.4). A dc voltage is applied
Figure 26.2 The relationship between the moisture content of a sub-
across the electrodes to decompose the water, the charge
stance and the equilibrium relative humidity of the surrounding gas, for the produced by the electrolysis being directly proportional to
example of wool. the mass of water absorbed (Faradays law). Thus the cur-
rent depends on the flow rate, which must be set and con-
trolled accurately at a predetermined rate (usually 100 ml
26.3 Measurement techniques min-1) so that the current meter can be calibrated directly in
ppm. Several points are worth making:
Techniques that allow automatic operation have the impor-
tant advantage that they can be used for process control. 1. The maximum moisture concentration measurable by
Here we therefore concentrate our attention on such tech- this technique is in the range 1,0003,000 vppm, but care
niques. Again, those available for gases, liquids, and solids must be taken to ensure that surges of moisture level do
are discussed separately. not wash off the P2O5.
2. There is generally a zero leakage current equivalent to a
few ppm. To allow for this error, when necessary, the cur-
26.3.1 Gases rent should be measured at two flow rates and the differ-
ence normalized to the flow for 100 ml min-1.
There is a huge choice of techniques for the measurement
of moisture in gases, reflecting the large number of ways in
which its presence is manifested. The techniques range from
measuring the extension of hair in simple wall-mounted
room monitors to sophisticated electronic instruments. To
some extent, the choice of technique depends on the prop-
erty required: dewpoint, concentration, or relative humidity.
Only the major techniques are discussed here. More exten-
sive treatments are given in the bibliography.

26.3.1.1 Dewpoint Instruments


The determination of the temperature at which moisture Figure 26.3 A schematic diagram of a sensor of a dewpoint mirror
condenses on a plane mirror can be readily estimated (see instrument.
432 PART | III Measurement of Temperature and Chemical Composition

are less affected by deposits on the cell windows and give


better calibration stability in polluted gases.

26.3.1.4 Electrical Sensor Instruments


There are many substances for which electrical impedance
changes with the surrounding moisture level. If this absorp-
tion process is sufficiently reproducible on a thin film, the
impedance, measured at either an audio frequency or a radio
frequency, can be calibrated in terms of moisture concen-
tration or relative humidity. Materials used in commer-
cial instruments include polymers, tantalum oxide, silicon
oxide, chromium oxide, aluminum oxide, lithium chloride
mixed with plastic, and carbon-loaded plastics that change
length and hence resistance with moisture level. Many such
instruments are available commercially, particularly using
an anodized aluminum oxide layer that has a very narrow
columnar pore structure (see Figure 26.5), but aging and
other deterioration processes can occur so that regular cali-
bration is essential. A major advantage of such sensors is
that as no imposed gas flow is necessary, they can simply be
placed in the gas to be measuredfor example, in an envi-
ronmental chamber. In addition, they can be used at high
Figure 26.4 A schematic diagram of a sensor of a coulometric
instrument.
pressure, they have a wide response range (typically 50C to
-80C dewpoint for a single aluminum oxide sensor), have
a rapid response, and are generally not expensive. These
advantages often outweigh any problems of drift and stabil-
3. Platinum electrodes are not suitable for use in gases con- ity and the requirement for regular calibration, but the sen-
taining significant amounts of hydrogen. The platinum sors must be used with care.
can catalyze the recombination of the electrolyzed oxy-
gen, and this water is also electrolyzed, giving inaccurate
measurements. Gold or rhodium elements reduce this 26.3.1.5 Quartz Crystal Oscillator Instrument
effect. The oscillation frequency of a quartz crystal coated with
4. In the absence of recombination and gas leaks, the hygroscopic material is a very sensitive detector of the
response of a coulometric instrument can be regarded as weight of absorbed water because very small changes in
absolute for many purposes. frequency can be measured. In practice, as shown in Fig-
5. Cells, which work at pressure, can be obtained. This can ure 26.6, two quartz crystal oscillators are used, and the
increase the sensitivity at low moisture levels because wet and a dry gasare passed across them alternately, usu-
it is possible to use a flow rate of 100 ml min-1 at the ally for 30seconds at a time. The frequency of crystal oscil-
measuring pressure, which does not increase the velocity lation is about 9.106 Hz and that of the crystal exposed to
of gas along the element and hence does not impair the the wet gas will be lowered and that of the crystal exposed
absorption efficiency of the P2O5. to the dry gas willrise. The resultant audio frequency dif-
ference is extracted, amplified, and converted to voltage
26.3.1.3 Infrared Instruments
Water vapor absorbs in the 12 nm infrared range, and infra-
red analyzers (see Chapter 29) can be successfully used as
moisture meters. For concentrations in the vppm range, the
path length has to be very long, and high sample flow rates
of several liters per minute can be necessary to reduce the
consequent slow response time. Both single-beam instru-
ments, in which the zero baseline is determined by measur-
Figure 26.5 An idealized representation of the pore structure of anod-
ing the absorption at a nearby nonabsorbing wavelength, ized alumina. The pores are typically less than 20 nm in diameter and more
and double-beam instruments, in which a sealed parallel cell than 100 nm deep. A porous gold layer is deposited on the alumina for
is used as reference, can be used. Single-beam instruments electrical contact when used as a hygrometer sensor.
Chapter | 26 Chemical Analysis: Moisture Measurement 433

by a brown color showing the presence of free iodine. This


is the basic standard technique and is incorporated into many
commercial instruments with varying levels of automation.
In process instruments the endpoint is determined electro-
metrically by amperometric, potentiometric, or coulometric
methods (see Chapter 23). In the amperometric method two
platinum electrodes are polarized; when free iodine appears
they are depolarized and the resultant current is measured
to define the endpoint. Potentiometrically, the potential of
an indicator electrode is monitored against a calomel elec-
trode and the endpoint is characterized by a sudden change in
potential. Coulometrically, iodine is generated by a constant
electrolyzing current from a modified reagent, and the time
taken to reach the endpoint gives the mass of water in the
sample. This last technique lends itself to automatic opera-
Figure 26.6 A block diagram of the arrangement of a piezoelectric tion, with samples injected sequentially or, in one instrument,
humidity instrument. Courtesy of Du Pont Instruments (UK) Ltd. the moisture in a sample flow is measured continuously by
mixing with standardized reagent, and the electrolysis cur-
to give a meter response of which the maximum value on rent is a measure of the mass flow of water.
each 30-second cycle is a measure of the moisture level. The
range of applicable concentrations is 13,000 vppm, and at
26.3.2.2 Infrared Instruments
lower levels the fact that the value after a certain time is
measured rather than an equilibrium value means that the The same comments apply as for gases (Section 26.3.1.3),
instrument can have a more rapid response than alternative but sample cell lengths are usually shorter, in the range of
methods (sample lines, however, often determine response 1100mm. It is an attractive method for online analysis,
time). Because the crystals see the sample gas for equal but care must be taken that other components in the liquid
times, contamination of the two crystals should be similar do not interfere with the measurement. Single-beam instru-
and the frequency difference little affected, resulting in sta- ments are most often used.
bility. However, regular calibration is still necessary, and the
complexity of the instrument makes it expensive. 26.3.2.3 Vapor Pressure Methods
26.3.1.6 Automatic Psychrometers As discussed in Section 26.2.2, the equilibrium relative
humidity above a liquid can be used to determine the
The measurement of the temperature difference between a moisture content in the liquid. Either the relative humid-
dry thermometer bulb and one surrounded by a wet muslin ity in a closed volume above the liquid can be measured
bag fed by a wick is the classical meteorological humidity or a sensor that responds to the moisture vapor pressure in
measurement. This is called psychometry, and automated the liquid can be immersed in the liquid. The aluminum
instruments are available. The rate of evaporation depends oxide sensor (Section 26.3.1.4) can be used either above the
on the gas flow as well as on the relative humidity, but gener- liquid because it does not require a gas flow rate or within
ally a flow rate greater than 3ms-1 gives a constant tempera- the liquid because, although the aluminum oxide pores will
ture depression. It is most useful at high relative humidities adsorb water molecules, they will not adsorb the liquid mol-
with accurate temperature measurements. ecules. These techniques are not appropriate if suspended
free water is present in the liquid.
One manufacturer has developed a system in which the
26.3.2 Liquids
sensor is a moisture-permeable plastic tube that is immersed
26.3.2.1 Karl Fischer Titration in the liquid. A fixed quantity of initially dry gas is circulated
through the tube, and the moisture in the gas is measured by
The Karl Fischer reagent contains iodine, sulfur dioxide,
an optical dewpoint meter. When equilibrium is reached, the
and pyridine (C5H5N) in methanol; the iodine reacts quanti-
dewpoint measured equals that of the moisture in the liquid.
tatively with water as follows:
7 3C 5 H 5 N + I 2 + SO 2 A + H 2 O " 2C 5 H 5 NHI + C 5 H 5 NSO 3
26.3.2.4 Microwave Instruments
C 5 H 5 NSO 3 + CH 3OH " C 5 H 5 NHSO 4 CH 3
The water molecule has a dipole moment with rotational
If a sample containing water is titrated with this reagent, vibration frequencies, which give absorption in the micro-
the endpoint at which all the H2O has been reacted is indicated wave, S-band (2.63.95 GHz), and X-band (8.212.4 GHz)
434 PART | III Measurement of Temperature and Chemical Composition

are pushed into the material, the meter on the instrument


being calibrated directly in percentage moisture. For
online measurements of granular materials, electrodes can
be rollers, plates (see Figure 26.8(b)), or skids, but uni-
form density is essential. A difficulty with this and other
online methods that require contact between the sensor
Figure 26.7 The basic concept for measuring moisture by microwave and the material is that hard materials will cause rapid ero-
absorption. sion of the sensor.

suitable for moisture measurement (see Figure 26.7). The


26.3.3.3 Microwave Instruments
S-band needs path lengths four times longer than the X-band Most comments appropriate to liquids also apply to solids,
for a given attenuation, and therefore the microwave band but, as mentioned, constant packing density is necessary.
as well as cell dimensions can be chosen to give a suitable For sheet materials such as paper or cloth, measurement is
attenuation. Electronic developments are causing increased simple, with the sheet passing through a slot in the wave-
interest in this technique. guide. For granular materials, uniform material density is
achieved by design of the flow path; alternatively, extruders
26.3.2.5 Turbidity-nephelometer or compactors can be useful.

Undissolved water must be detected in aviation fuel during


transfer. After thorough mixing the fuel is divided into two
26.3.3.4 Infrared Instruments
flows; one is heated to dissolve all the water before it passes The basic difference from measurements in gases and liquids
into a reference cell, and the other passes directly into the is that, for solids, reflectance methods (see Figure 26.9) are
working cell. Light beams split from a single source pass usually used rather than transmission methods. Single-beam
through the cells, suspended water droplets in the cell scat-
ter the light, and a differential output is obtained from the
matched photoelectric detectors on the two cells. Moisture
of 0 to 40 ppm can be detected at fuel temperatures of 30
to 40C.

26.3.3Solids
The range of solids in which moisture must be measured
commercially is wide, and many techniques are limited to
specific materials and industries. In this book just some of
the major methods are discussed.

26.3.3.1 Equilibrium Relative Humidity Figure 26.8 Two techniques for electrical measurements of moisture
in solids: (a) pointed probes for insertion in wood, plaster, etc. to measure
The moisture level of the air immediately above a solid can resistance, (b) capacitance plates to measure moisture in flowing powder
be used to measure its moisture content. Electrical probes, or granules.
as discussed in Section 26.3.1.4, are generally used, and if
appropriate they can be placed above a moving conveyor.
If a material is being dried, its temperature is related to its
equilibrium relative humidity, and a temperature measure-
ment can be used to assess the extent of drying.

26.3.3.2 Electrical Impedance


Moisture can produce a marked increase in the electrical
conductivity of a material and, because of waters high
dielectric constant, capacitance measurements can also be
valuable. Electrical resistance measurements of moisture
in timber and plaster are generally made using a pair of Figure 26.9 The principle of infrared reflectance used to measure
sharp pointed probes, as shown in Figure 26.8(a), which moisture in a solid on a conveyor.
Chapter | 26 Chemical Analysis: Moisture Measurement 435

operation is used with a rotating absorption and reference 26.4.1 Gases


frequency filter to give regular zero readings. The calibra-
First, the difficulties of making accurate moisture
tion of a reflectance method can be substantially indepen-
measurements must be stressed. This is particularly so
dent of the packing density as it measures only the surface
at low levels, say, less than 100 vppm, because, since all
concentration. For material on a conveyor belt, a plough is
materials absorb moisture to some extent, sample lines
often used in front of the sensing position to ensure a mea-
must come to equilibrium as well as the detector. At low
surement more typical of the bulk. The method is not suit-
moisture levels this can take hours, particularly at low-
able for poorly reflecting materials such as carbon and some
flow rates. A rapid-flow bypass line can be valuable.
metal powders.
Patience is mandatory, and if possible the outputs of the
instruments should be recorded to establish when stable
26.3.3.5 Neutron Moderation conditions are achieved. Many plastics are permeable to
Hydrogen nuclei slow (moderate) fast neutrons; therefore, moisture and must never be used. At high moisture levels
if a fast neutron source is placed over a moist material with copper, Teflon, Viton, glass, or quartz can be satisfactorily
a slow neutron detector adjacent, the detector output can be used, but at low levels stainless steel is essential. Finally,
used to indicate the moisture concentration. The concen- at high moisture levels it must be remembered that the
tration of any other hydrogen atoms in the material and its sample lines and detectors must be at least 10 Kelvins
packing density must be known. This technique is described hotter than the dewpoint of the gas.
in Chapter 30. Nuclear magnetic resonance can also be used There are two basic calibration methods that can, with
to detect hydrogen nuclei as a means of measuring moisture advantage, be combined. Either a sample gas is passed
content. through a reference hygrometer and the instrument under
test or a gas of known humidity is generated and passed
through the instrument under test. Obviously, it is ideal to
double-check the calibration using a known humidity and a
26.4Calibration reference hygrometer.
It will be seen from the preceding sections that many mois- The most suitable reference hygrometer is the dewpoint
ture measurement techniques are not absolute and must be meter, which can be readily obtained with certified calibra-
calibrated, generally at very regular intervals. It must first be tion traceable to a standard instrument. For many applica-
emphasized that the absolute accuracy of moisture measure- tions, less sophisticated dewpoint instruments would be
ment, particularly in gases, is not usually high. Though it is adequate, and coulometric analyzers are possible for low
possible to calibrate moisture detectors for liquids or solids moisture levels. At high levels gravimetric methods can be
to 0.1 to 1.0 percent, such accuracies are the exception rather used, but they are slow and tedious and difficult to make
than the rule for gases. Figure 26.10 shows the accuracies of accurate.
some of the techniques discussed in this chapter compared There are a range of possible humidity sources, some of
with the absolute gravimetric standard of the U.S. National which are available commercially; the choice depends on
Bureau of Standards. the facilities available and the application:
1. A plastic tube permeable to moisture and held in a ther-
mostatically controlled water bath will give a constant
humidity for a given flow rate. Some manufacturers sell
such tubes precalibrated for use as humidity sources, but
obviously, the method is not absolute, and the perme-
ation characteristics of the tubes may change with time.
2. Gas cylinders can be purchased with a predetermined
moisture level that does not significantly drift because of
the internal surface treatment of the cylinder. However,
to prevent condensation in the cylinder, the maximum
moisture level is limited to about 500 vppm, even with a
cylinder pressure of only 10 bar. They are most suitable
for spot checks of instruments on site.
3. If an inert gas containing a known concentration of
hydrogen is passed through a bed of copper oxide heated
to ~350C, the hydrogen is converted to water vapor.
This method relies on the measurement and stability of
Figure 26.10 The accuracy of some of the major techniques for mea- the hydrogen content, which is better than for moisture.
suring moisture in gases, after Wexler (1970). The generated humidity is also independent of flow rate.
436 PART | III Measurement of Temperature and Chemical Composition

4. Water can be continuously injected into a gas stream Balances can be obtained with a built-in heater, which
using either an electrically driven syringe pump or a gives a direct reading of moisture content for a fixed initial
peristaltic pump. The injection point should be heated to sample weight. Other favored techniques include measuring
ensure rapid evaporation. The method can be used very the water vapor given off by absorbing it in a desiccant to
successfully, syringes in particular allowing a very wide avoid the effects of volatiles; the Karl Fischer method again;
range to be covered. or mixing the substance with calcium carbide in a closed
5. If a single humidity level can be generated, a range can bomb and measuring the pressure of acetylene produced.
be obtained using a flow mixing system, but to achieve The method must be carefully chosen to suit the substance
sufficient accuracy mass-flow meters will probably be and process technique being used. Finally, it is worth noting
necessary. that rather than an absolute calibration, calibration directly
in terms of the desired quality of the substance in the manu-
facturing process may be the most appropriate.
26.4.2 Liquids
The basic absolute method is that of the Karl Fischer titra-
tion, which was described in Section 26.3.2.1. References
Mitchell, J., and Smith, D., Aquametry. Part I, A Treatise on Methods for
the Determination of Water, Chemical Analysis Series No. 5, Wiley,
26.4.3Solids New York (1977).
Mitchell, J., and Smith, D., Aquametry. Part 2, The Karl Fischer Reagent,
There are several methods that allow the absolute moisture
Wiley, New York (1980).
level of a solid to be determined, but for all of them, samples
Verdin, A., Gas Analysis Instrumentation, Macmillan, London (1973).
of the specific substance being measured by the process tech- Wexler, A., Electric hygrometers, National Bureau of Standards Circular
niquemust be used. The most common technique is, of course, No 586 (1957).
to weigh a sample, dry it, and then weigh again. Drying temper Wexler, A. (ed.), Humidity and Moisture (3 volumes), papers presented at
ature and time depend on the material; if necessary, the temper- a conference, Reinhold, New York (1965).
ature must be limited to avoid decomposition, loss of volatile Wexler, A., Measurement of humidity in the free atmosphere near the sur-
components, or absorption of gases from the atmosphere. face of the Earth, Meteorological Monographs, 11, 262282 (1970).

You might also like