INTERNAL CALIBRATION PROCEDURE

Approved by:H. Woodyatt Date:Feb 2004


BMTS-CP-006, Issue 1, Feb 2004

INTERNAL CALIBRATION PROCEDURE AND RESULTS


Soil Resistivity Box measured against ASTM G 57-95A Reapproved 2001

ITEM DETAILS
Bodycote Serial Number Description Range
Reference
Soil Resistivity Box NA

Procedure

1. Equipment/Consumables required

REFERENCE EQUIPMENT REQUIRED/USED


Description Bodycote Reference Date of Last Calibration Stated
Uncertainty
Vernier Calipers Readable to
0.1mm min
Balance readable to 0.01 g
Ruler
Thermometer, 1 C

2 Method
Dimensions
2.1 Clean the soil box thoroughly to remove dust and contaminants.
2.2 Measure the length of the box above the electrode plate at both the ends and centre to the nearest 1 mm.(a)
2.3 Measure internal width of the box at both the ends and centre of the box to the nearest 0.01 mm.(b)
2.4 Measure the height of the box at both the ends and centre of the box to the nearest 0.01mm.(c)
2.5 Measure the clear distance between the outer electrodes and inner electrodes at three approximately equal
positions to the nearest 0.01mm.(d)
2.6 Measure the clear distance between the inner electrodes at three approximately equal positions. To the
nearest 0.01 mm.(e)
Volume
2.7 Weigh the soil box to the nearest 0.01 g.(f)
2.8 Fill the box with distilled water(20-25C) up to the brim and weight to the nearest 0.01g.(g)
2.9 Calculate the volume of the soil box.

Page 1 of 2 BMTS- CP- 006


INTERNAL CALIBRATION PROCEDURE
Approved by:H. Woodyatt Date:Feb 2004

Results of Calibration

Mass or Dimension Measured Recorded Values Specification


1 2 3 Mean
Length ,mm (1) a
Width, mm (0.01) b
Height, mm (0.01) c
Distance between O1 and I1, mm (0.01) d
Distance betweenO2 and I2, mm (0.01) d
Mean distance, mm (0.01) d
Distance between I1 and I2,mm (0.01) e
Mass of box, g(0.01) f
Mass of box + water, g(0.01) g
Water temperature, C (1)
Volume of soil box, cc (0.01) = g-f , h

Data Results Specification

Cross Sectional Area, cm2 (1) = Mean(a x b)


Inner electrode spacing, cm (0.01) = Mean e
Outer electrode spacing, cm (0.01) =Mean d
Volume of soil box, cc(0.01) = Mean h

Date of Calibration Ambient Temperature (C)

Suitability for Use/Restrictions on Use

Name of Operator Signature of Operator Checked by, Name Checked by, Signature