You are on page 1of 8

AJCP / Original Article

Do the Flags Related to Immature Granulocytes


Reported by the Sysmex XE-5000 Warrant a
Microscopic Slide Review?
Heidi Eilertsen, MSc,1,2 and Tor-Arne Hagve, MD, PhD1,3

From the 1Division of Diagnostics and Technology, Akershus University Hospital, Lrenskog, Norway; 2Faculty of Health Sciences, Oslo, and Akershus
University College of Applied Sciences, Oslo, Norway; and 3Institute of Clinical Medicine, University of Oslo, Akershus University Hospital, Oslo, Norway.

Key Words: Sysmex XE-5000; IG count; Imm Gran? flag; Analytical quality

Am J Clin Pathol October 2014;142:553-560

DOI: 10.1309/AJCP4V4EXYFFOELL

ABSTRACT Modern automated hematology instruments, in addition


Objectives: The Sysmex XE-5000 instruments (Sysmex, to a fully five-part differential, also provide characterization
Kobe, Japan) count immature granulocytes (IGs) and use and counting of abnormal cells in the peripheral blood, such as
the Imm Gran? flag to signal unreliable results. This blasts, atypical lymphocytes, and immature granulocytes (IGs).
study investigated the usefulness of the Imm Gran? flag The Sysmex XE-5000 (Sysmex, Kobe, Japan) is a
and the analytical and diagnostic performance of the IG fully automated hematology instrument that counts IGs
measurements in a side-by-side evaluation. (metamyelocytes, myelocytes, and promyelocytes) in the
differential (DIFF) channel based on flow cytometry.1 In
Methods: In total, 408 samples were analyzed on three addition, information on immature cells is derived from
XE-5000 instruments. The IG count and the Imm Gran? the immature myeloid information (IMI) channel based on
flag reports from all three instruments were used for radiofrequency and direct current measuring principles.1,2 In
reproducibility studies. The diagnostic performance of the the IMI channel, not only IGs but also bands and myeloblasts
automated IGs and the Imm Gran? flag were studied by are detected. If a difference occurs between the numbers
comparing the XE-5000 results with the results of the manual of cells detected in the DIFF channel and the IMI channel,
differential. a suspect IG flag (Imm Gran?) is reported in addition to
Results: The reproducibility of the Imm Gran? flagging the immature count derived from the DIFF channel.1 A flag
between instruments was poor (, 0.75-0.80). The most report indicates the presence of abnormal cells, a finding that,
significant contributor to the report of the Imm Gran? according to the current recommendations, has to be verified
flag was bands, and the flag played a minor role in detecting by a manual evaluation of cell morphology.3
blasts. The interinstrument reproducibility of the IG counts According to our standard procedure, a blood smear
was high (intraclass correlation, 0.99). The IG count shall be examined for (1) all samples with an Imm Gran?
reported by XE-5000s was higher than the manual IG count flag and (2) all samples with a relative IG count of 5%
(36%-55%), and the difference and the variability tended to or more (presented as an IG Present flag). The purpose
increase with increasing levels of IGs. of the examination is to verify the presence of IGs and to
identify or exclude the presence of other pathologic cells
Conclusions: The Imm Gran? flag has a poor analytical in peripheral blood such as blasts. The microscopic slide
quality and gives no substantial information on the presence review is not supposed to replace the automated IG count
of blasts in the sample. We therefore suggest reporting the with a manual IG count.
automated IG count without initial microscopic slide review. To our knowledge, no studies have investigated the
quality and the usefulness of the IG-related flags. The aim of
this study was thus to evaluate the analytical and diagnostic
performance of IG measurements as well as the usefulness of

American Society for Clinical Pathology Am J Clin Pathol 2014;142:553-560 553


553 DOI: 10.1309/ AJCP4V4EXYFFOELL 553
Eilertsen and Hagve / Usefulness of Immature Granulocyte Flags on the Sysmex XE-5000

the IG flag reports on the Sysmex XE-5000, questioning the channel, which identifies and counts IG cells in addition to
need for a manual slide review or a manual count of samples the standard five-part differential populations. The number
with a flag report. Three XE-5000 instruments were evaluated of cells the instrument routinely counts in each sample to
side-by-side. The interinstrument agreement and the accuracy generate a differential cell count depends on the total number
of the IG counts, as well as the usability of the IG flag reports, of WBCs, the dilution ratio, and the counted volume. At a
were studied by comparing with a manual differential. WBC count value of 5.0 109/L, 3,922 cells are counted
in the DIFF channel. The IG counts derived from the
DIFF channel include metamyelocytes, myelocytes, and
promyelocytes. The software reports both the absolute and
Materials and Methods
the relative IG counts.4 To alert crucial levels, one can flag
the samples with a user-defined flag, termed IG Present,
Sample Selection and Study Design which occurs when the absolute or relative IG count exceeds
At Akershus University Hospital, Oslo, Norway, the the user-defined threshold. In a separate IMI channel, not
routine hematology samples are analyzed randomly on one only IGs but also bands and myeloblasts are detected. The
of three Sysmex XE-5000 instruments (hereafter, XE1, XE2, number of cells the instruments routinely counts using
and XE3) integrated onto a track-based automation system. radiofrequency in the IMI channel, at a WBC count value of
The instruments were purchased at the same time, calibrated 5.0 109/L, is 5,000 cells. The instrument software examines
and harmonized by the manufacturer, and used with identical the scattergrams from both the DIFF and the IMI channels.
software packages, including firmware upgrades. During a If there are any disagreements in the information derived
period of 5 months, 408 samples from the daily workload from the scattergrams, the IG flag algorithm comes into
were selected and reanalyzed on all three instruments. For operation and the Imm Gran? flag will be triggered. The
the purpose of this study, all samples were kept anonymous. IG count derived from the DIFF channel will, however, still
Only samples reported initially with one or more of the four be reported. A Q value provides information on the degree of
suspect flags, blasts, IGs, atypical lymphocytes, and abnormal positivity or negativity of the IG flag on a scale of 0 to 300,
lymphocytes/lymphoblasts were included. The triggering with increments of 10 arbitrary units. At a given threshold,
thresholds for the suspect flags were all factory settings. The the instrument will report the Imm Gran? flag. The factory
IG count and the flag report results from all three instruments setting threshold is preset to an arbitrary value of 100 and is
were used for the study of interinstrument agreement. The user adjustable. In this study, the factory setting threshold
diagnostic performance of the IG count and the Imm Gran? setting was 100 for trigging the Imm Gran? flag, and the
flag were determined by comparing the results reported by the user-defined IG percentage threshold for triggering the IG
instruments with the results from a blood smear examination. Present flag was 5%.

Blood Sampling and Analytical Conditions Manual Differential Leukocyte Count


Venous blood samples were collected in Vacuette tubes For each sample, two blood smears were prepared
(Greiner Bio-One, Frickenhausen, Germany) containing and stained with May-Grnwald-Giemsa using the Sysmex
dipotassium EDTA. The specimens were transported to SP-1000. A total of 200 cells were counted by two highly
the laboratory by a pneumatic tube transporting system trained technicians, each counting 100 cells. The total count of
and processed on the Sysmex XE-5000 instruments using metamyelocytes, myelocytes, and promyelocytes represented
software version 00-04 in closed mode within 4 hours the IG count. The mean IG percentage was calculated for
after collection. The instruments were continuously involved each sample based on the two manual differential counts. The
in the routine workload during the study period, and their absolute count was computed by the WBC count obtained
performance was monitored with an internal quality control from the XE-5000. To determine a positive smear finding,
(QC) system, including Sysmexs QC material e-Check and we used the criteria listed by the International Consensus
an external quality surveillance program. The QC system, Group for Hematology Review.3 The presence of a single
based on the e-Check, includes evaluation of the parameter IG myelocyte/promyelocyte (0.5%) and/or two metamyelocytes
number and IG percentage but not the instruments flagging (1%) qualified for a true-positive smear finding.
performance.
Statistical Analysis
Sysmex XE-5000 Statistical analyses were performed using the SPSS
The Sysmex XE-5000 provides WBC differential counts software version 16 (SPSS, Chicago, IL) for Windows
based on an optical principle, including forward, side (Microsoft, Redmond, WA) and Analyse-it version 2.26
scatter, and side fluorescence. The XE-5000 has a DIFF (Analyse-it Software, Leeds, England). A two-tailed P value

554 Am J Clin Pathol 2014;142:553-560 American Society for Clinical Pathology


554 DOI: 10.1309/AJCP4V4EXYFFOELL
AJCP / Original Article

Table 1
of less than .05 was considered statistically significant. To Interinstrument Variability for the Imm Gran? Flag
describe the data, the IG counts were summarized as median Reports Generated by Three Sysmex XE-5000 Instruments
and range before the analysis of concordance. The intraclass
Pair of 95% Confidence P Value
correlation coefficient (ICC [2.1], two-way random, single Instruments a Interval
measures) was calculated to quantify the interinstrument
XE1/XE2 0.79 0.71-0.87 <.001
reliability of the IG counts, and values were used to determine XE1/XE3 0.75 0.67-0.83 <.001
interinstrument reliability for the Imm Gran? flag.5 The XE2/XE3 0.80 0.73-0.88 <.001
diagnostic performance of the automated IG counts was a The extent of agreement between two instruments (n = 408).
examined by receiver operating characteristic (ROC) curve
analyses and compared for all three instruments with respect Samples With the Imm Gran? Flag
to smear findings. Passing-Bablok regression was used for
the method comparison between the Sysmex XE-5000s Based on the factory setting thresholds for flagging, the
and the manual differential counts.6 Bland-Altman analysis XE1, XE2, and XE3 instruments reported the Imm Gran?
was used to determine the degree of agreement between the flag for 81, 80, and 77 samples, respectively. The values for
Sysmex XE-5000s and the manual differential IG counts.7 agreement between the pair of instruments were 0.79 (XE1/
Pearson correlation coefficient was used to determine the XE2), 0.75 (XE1/XE3), and 0.80 (XE2/XE3) Table 1. One
association between the two variables. Logistic regression or more instruments flagged for Imm Gran? in 102 samples.
was used to assess the association between the predictors; The three instruments showed full agreement in reporting the
band, metamyelocytes, myelocytes, promyelocytes, and flag in 60 (59%) samples. Blast cells were detected by manual
blasts analyzed by manual microcopy of the smears; and counting in 17 (XE1), 17 (XE2), and 18 (XE3) of the samples
the IG flag generated by the Sysmex XE-5000 as the reported with an Imm Gran? flag. For most of these samples
dichotomous dependent outcome.8 All significant variables (16, 15, and 16, respectively), a Blast? flag was also reported.
in the univariate analysis were included in the multivariate Only one to two of the 48 samples with blasts present in the
analysis to adjust for confounders. blood smear were reported with an Imm Gran? flag as a lone
flag. For these samples, one of the technicians observed one
blast by the 100-cell manual differential count, corresponding
to the presence of 0.5% circulation blasts in peripheral blood.
Results
In addition, in one to two of the 48 samples with blasts present
The results of the leucocyte counts of the 408 samples in the blood smear, neither the Imm Gran? flag nor the
performed on the three Sysmex instruments varied from Blast? flag was reported by any of the instruments. For one
leukopenia to severe leukocytosis (0.5-207 109/L) with a of these samples, one of the technicians observed one blast by
median of 9.9 109/L and mean of 20.3 109/L. A manual the 100-cell manual differential count, corresponding to the
differential count of the samples showed metamyelocytes presence of 0.5% circulation blasts in peripheral blood. Case 2
(1% cells) in 133 samples, myelocytes (0.5% cells) in was a leukopenic sample (0.9 109/L).
280, promyelocytes (0.5% cells) in 41, and blasts (0.5% The odds ratios (ORs) for association between the
cells) in 48. No IGs were found in 31% of the samples, while myeloid cells observed in the blood film and the Imm
one, two, or three of the populations of myeloid cells were Gran? flag reported by XE1 are presented in Table 2. The
found in 34%, 38%, and 7% of the samples, respectively. most significant contributor to the report of the Imm Gran?
The agreement between the two manual differential IG cell flag is bands. The odds for a flag report increase by 8.2% per
counts in the 408 cases, presented as an ICC, was 0.903 (95% percentage increase in the number of bands. The increase in
confidence interval [CI], 0.884-0.920). the odds per percentage increase in myeloblasts was 5.4%.

Table 2
Logistic Regression Analyses of the Risk of the Imm Gran? Flag Generated by Sysmex XE-5000 (XE1) Based on the Myeloid
Cell Proportions (%) in Blood Smears

Unadjusted Effect Adjusted Effect

Variable OR 95% CI P Value OR 95% CI P Value

Bands 1.082 1.046-1.119 <.001 1.082 1.041-1.124 <.001


Immature granulocytes 1.103 1.049-1.160 <.001 1.056 0.999-1.115 .053
Myeloblasts 1.044 1.019-1.070 <.001 1.054 1.028-1.081 <.001

CI, confidence interval; OR, odds ratio.

American Society for Clinical Pathology Am J Clin Pathol 2014;142:553-560 555


555 DOI: 10.1309/ AJCP4V4EXYFFOELL 555
Eilertsen and Hagve / Usefulness of Immature Granulocyte Flags on the Sysmex XE-5000

Table 3
The results from the other two instruments were similar Mean, Median, and Range for Immature Granulocyte (IG)
(results not shown). Counts Reported by Three Sysmex XE-5000 Instrumentsa
An IG count was reported for 46% to 52% of the samples
Instrument Mean, 109/L Median, 109/L Range, 109/L
with an Imm Gran? flag by the XE-5000 instruments. On
the XE1, the IG count was 55% higher than the smear-based XE1 0.44 0.15 0-4.28
XE2 0.43 0.15 0-4.59
count (y = 1.55x [95% CI, 1.00 to 3.15] + 0.00 [95% CI, 0.09 XE3 0.46 0.18 0-4.72
to 0.01]). Similar results were found for the XE2 and the XE3
a In total, 306 samples were measured on each instrument (XE1, XE2, and XE3).
(data not shown). The difference between the two methods as
well as the variability tended to increase with the level of the
IG counts (data not shown). 0.931 (95% CI, 0.903-0.959), and 0.922 (95% CI, 0.891-
0.953) for XE1, XE2, and XE3, respectively.
Samples With No Imm Gran? Flag
For the 306 samples with no Imm Gran? flag, an IG
count result was reported for 289, 295, and 279 samples
Discussion
analyzed by XE1, XE2, and XE3, respectively. The IG counts
ranged from 0 to 4.7 109/L with a median of 0.2 109/L The main purpose of the present study was to evaluate the
and a mean of 0.4 109/L Table 3. The interinstrument usefulness of the IG flags generated by the Sysmex XE-5000.
reproducibility of the absolute IG counts between the three The Sysmex XE-5000 reports two flag types that are related
instruments, presented as an ICC, was 0.988 (95% CI, 0.985- to the presence of IGs in peripheral blood: the IG Present
0.990). The interinstrument variability of the absolute IG flag, indicating the potential presence of IGs, and the Imm
counts between XE1/XE2 and XE1/XE3 as measured by the Gran? flag, indicating the presence of suspect cells such as
coefficient of variation was 14% and 19%, respectively. blasts. Based on the established review criteria, blood smear
Of the 306 samples with no Imm Gran? flag report, examination is performed to assess the validity of the reported
the XE1, XE2, and XE3 instruments reported the flag IG flags. This practice is time-consuming and expensive. Thus,
present (IG >5%) in 95, 90, and 96 samples, respectively. further research is needed to address the clinical significance
Blast cells were detected by manual review in 14 (XE1), 12 of the flagging system and the need for follow-up.
(XE2), and 13 (XE3) of the samples reported with an IG The main finding in this study is the insufficient
Present flag. For most of these samples (13, 12, and 12, analytical performance of the Imm Gran? flag, indicating
respectively), a Blast? flag also was reported. Consequently, that the Imm Gran? flag is not a sufficient criterion
only one of the samples with blasts present in the blood smear for a microscopic review. The analysis of the same
was reported with IG present as a lone flag. In addition, samples on three instruments revealed low concordance in
in nine to 10 of the 48 samples with blasts present in the generating the Imm Gran? flag. In only 60% of cases,
smear, neither the IG Present flag nor the Blast? flag was all three instruments reported the IG flag on the same
reported by any of the instruments. samples. These findings are difficult to explain because
Figure 1 presents the regression analysis of the results all samples were analyzed at the same time and under the
of the manual IG count vs the automated count on the three same conditions on all three instruments to reduce the
XE-5000 instruments. The coefficient of correlation varied factors that usually influence comparative evaluations.9,10
between 0.82 (manual count and XE3) and 0.89 (manual In addition, the instruments were under a continuous QC
count and XE2). The automated IG count was 36% to 40% system. It is, however, a problem that the commercial
higher than the manual smear count. The difference between control programs do not include an assessment of clinically
the two methods as well as the variability tended to increase important cell types such as blasts or the instruments
with increasing levels of IGs Figure 2. The agreement flagging performance. With a common Q value cutoff
between the IG count from technician 1 and the XE1 and at 100, the observed value (0.75-0.80) for agreement
technician 2 and the XE1, presented as an ICC, was 0.815 between pairs of instruments indicates poor reproducibility
(95% CI, 0.751-0.861) and 0.738 (95% CI, 0.619-0.814), of the Imm Gran? flag, supporting the conclusion that
respectively. Similar results were found for the XE2 and the the analytical performance of the flag is low. In general,
XE3 (data not shown). the usefulness of instrument-generated flags is evaluated
Figure 3 shows the ROC curves for the overall by their sensitivity and specificity.1,11-13 However, on
performance of the IG count on the Sysmex XE-5000s to the basis of the present findings, we suggest that the
predict IGs when we used the criteria for a positive smear interinstrumental reproducibility in reporting flags also
established by the International Consensus Group.3 The areas has to be taken into account when the usefulness of flag is
under the curve (AUCs) were 0.918 (95% CI, 0.886-0.949), evaluated.

556 Am J Clin Pathol 2014;142:553-560 American Society for Clinical Pathology


556 DOI: 10.1309/AJCP4V4EXYFFOELL
AJCP / Original Article

A B

6 6
Sysmex XE-5000 IG Count (109/L) XE1

Sysmex XE-5000 IG Count (109/L) XE2


5 5

4 4

3 3

2 2

1 1
Identity Identity
Passing-Bablok Passing-Bablok
0 0
0 1 2 3 4 5 6 0 1 2 3 4 5 6
Manual IG Count (10 /L)
9
Manual IG Count (10 /L) 9

Figure 1 Passing-Bablok regression analysis of the absolute


C
immature granulocyte (IG) count determined by manual
6 differential and three Sysmex XE-5000 instruments (Sysmex,
Sysmex XE-5000 IG Count (109/L) XE3

Kobe, Japan). A, XE1. y = 1.36x (95% CI, 1.27-1.47) + 0.01


5 (95% CI, 0.00-0.01). B, XE2. y = 1.38x (95% CI, 1.29-1.50) +
0.00 (95% CI, 0.00-0.01). C, XE3. y = 1.40x (95% CI, 1.31-
4 1.49) + 0.00 (95% CI, 0.00-0.01).

1
Identity
Passing-Bablok
0
0 1 2 3 4 5 6
Manual IG Count (109/L)

An interesting observation in this study is that the most IG count and the level of the threshold. The present data
important contributor to the report of the Imm Gran? flag clearly demonstrate that the IG Present flag gives almost
is the presence of band cells. It may thus be questioned no information on the presence of blasts in addition to the
whether the Imm Gran? flag also is a measure of left- Blasts? flag. In samples with blasts detected by a manual
shift. Consequently, depending on the actual guidelines, the slide review, a Blasts? flag in 93% to 100% of the cases
presence of band cells may result in a significant delay in the accompanied the IG Present flag. It is thus questionable
report of results. whether the presence of the Imm Gran? or IG Present
The present data also show that the Imm Gran? flag flag provides sufficient information to warrant a slide review.
plays only a minor role in detecting blasts. In samples with In this study, up to two of the 48 samples with blasts detected
an Imm Gran? flag, blasts were detected by slide review in by a manual differential were reported with Imm Gran? as
approximately 20% of the samples. In 88% to 94% of these, the only flag. The consequence of omitting a slide review of
a Blasts? flag was reported together with the Imm Gran? samples with Imm Gran? is the risk of missing myeloblasts
flag. In accordance with previous results, in approximately 5% in peripheral blood, which may have an important clinical
of all samples with blasts, no Blasts? flag was reported.14 impact. However, in the two present cases, the smear review
The IG Present flag is reported when the IG count revealed only a single blast, observed by one of the two
from the DIFF channel is increased above a given threshold technicians. This discordance could be due to the limitations
and thus depends only on the analytical quality of the of the 100-cell manual differential or the fact that one of the

American Society for Clinical Pathology Am J Clin Pathol 2014;142:553-560 557


557 DOI: 10.1309/ AJCP4V4EXYFFOELL 557
Eilertsen and Hagve / Usefulness of Immature Granulocyte Flags on the Sysmex XE-5000

Identity 1.0
Bias (0.136)
95% Limits of
agreement
0.8

True-Positive Rate (Sensitivity)


2.0

0.6
Difference Sysmex XE-5000 IG Count

1.5
Manual IG Count (109/L)

1.0 0.4

0.5 XE1
0.2
XE2
XE3
0 No discrimination
0.0
0.0 0.2 0.4 0.6 0.8 1.0
0.5
False-Positive Rate (1 Specificity)

1.0 Figure 3 Receiver operator characteristic curves for overall


0.5 0.5 1.5 2.5 3.5 4.5 performance of the immature granulocyte count on the
Mean of All IG Count (109/L) Sysmex XE-5000 (Sysmex, Kobe, Japan) to predict immature
Figure 2 Bland-Altman plot for the absolute immature granulocytes in blood samples.
granulocyte (IG) counts determined by manual differential and
the counting of IGs compared with counting on the Sysmex
the Sysmex XE-5000 (XE1; Sysmex, Kobe, Japan). The 95%
XE-2100 and by flow cytometry.17-19 High imprecision for
limits of agreement are 0.525 to 0.797.
the manual method at low counts of cells is well known.15
technicians overlooked the presence of blasts. It is well known However, the present study shows a negative bias as well
that manual cell classification is subjective and has limited as a significantly higher variability for the manual method
application in low-frequency cell populations.15,16 Thus, it is at IG counts of 0.5 109/L or more. This may be explained
difficult to assess the relevance of the single blast detection. by the counting statistics as suggested by Koepke et al,16 but
In a previous study, we found that the miss rate of the it may also be a result of a nonhomogeneous distribution
Sysmex XE-5000 platform in detecting circulation blasts of leukocytes on the blood smears. Wedge techniques for
was high, with an observed sensitivity of the Blasts? flag preparing smears may cause nonhomogeneous distribution
from 0.54 to 0.75.14 In this context, the loss of clinical value of leukocytes in the films. Large cells as IGs may be pushed
by disregarding the Imm Gran? flag seems insignificant. to the edge and to the sides of the blood film and thereby
However, the disregard of Imm Gran? as the sole reason disappear from the counting area.20 Another possible reason
for manual review will result in a reduced number of smears, is that of misclassification. Inadequate classification of
simplified routines, reduced use of resources, and a shorter metamyelocytes as bands excludes the metamyelocytes from
turnaround time. Consequently, there is no obvious rationale the manual IG count and thereby produces inconsistency
for manual examination of blood smears in cases flagged between automated and manual IG counting.21 It is difficult
solely for Imm Gran?. to assess the significance of the inconsistency since the
The comparison of the Sysmex XE-5000 count with comparison is halting for many reasons. The instruments
the manual count of IGs (metamyelocytes, myelocytes, have an assumed statistical advantage over the manual
and promyelocytes) in samples with no Imm Gran? flag count by the larger number of cells counted.15 In this study,
revealed a correlation coefficient of 0.82 to 0.89 for the the agreement observed between the IG counts from the
absolute count depending on the three instruments. This is instruments (0.99) was greater than the agreement of IG
in agreement with previous studies.17,18 The comparison counts from the two manual counts (0.90). In addition, the
between the manual count and the XE-5000 method XE-5000 measurements had a smaller range of uncertainty.
shows a linear proportional bias, with XE-5000 reporting Therefore, although quantitative discordant results are
approximately 35% higher IG counts than the manual method. difficult to evaluate, the IG counts from the XE-5000 are
Similar results were found in samples with an Imm Gran? supposed to be more accurate than the manual cell counts.
flag. This is in agreement with several previous studies Consequently, there is no reason for not using the instrument
demonstrating that the manual count method underestimates counts. However, a manual review is needed, which would

558 Am J Clin Pathol 2014;142:553-560 American Society for Clinical Pathology


558 DOI: 10.1309/AJCP4V4EXYFFOELL
AJCP / Original Article

yield quantitative results of the separate nonblastic IG types References


(metamyelocytes, myelocytes, and promyelocytes). 1. Briggs CJ, Linssen J, Longair I, et al. Improved flagging rates
on the Sysmex XE-5000 compared with the XE-2100 reduce
The analytical accuracy of IG counts observed in
the number of manual film reviews and increase laboratory
samples with and without an Imm Gran? flag was productivity. Am J Clin Pathol. 2011;136:309-316.
comparable. These findings indicate that the IG count might 2. Cimenti C, Erwa W, Herkner KR, et al. The predictive
be reliable even when accompanied by an Imm Gran? flag value of immature granulocyte count and immature myeloid
and can be reported. information in the diagnosis of neonatal sepsis. Clin Chem
Lab Med. 2012;50:1429-1432.
The reproducibility of the absolute IG count obtained
3. Barnes PW, McFadden SL, Machin SJ, et al; International
from the three instruments evaluated by the ICC value was Consensus Group for Hematology. Review: suggested criteria
high (0.99). The interinstrument variability between pairs of for action following automated CBC and WBC differential
instruments measured as percent coefficient of variation was analysis. Lab Hematol. 2005;11:83-90.
also acceptable (14%-19%). These findings are comparable 4. Briggs C, Harrison P, Grant D, et al. New quantitative
parameters on a recently introduced automated blood cell
to those of Fernandes and Hamaguchi17 and Briggs et al22 for counter: the XE-2100. Clin Lab Haematol. 2000;22:345-350.
within-run reproducibility of IGs on the Sysmex XE-2100 5. Shoukri MM. Measures of Interobserver Agreement. Boca
and indicate that random use of different instruments seems Raton, FL: CRC Press; 2004:39-61.
to have a minor influence on the analytical performance with 6. Bilic-Zulle L. Comparison of methods: Passing and Bablok
respect to IG counts. regression. Biochem Med. 2011;21:49-52.
The ability of the instrument count to predict the 7. Bland JM, Altman DG. Statistical methods for assessing
agreement between two methods of clinical measurement.
presence of IGs in smears is very good, with AUCs ranging Lancet. 1986;1:307-310.
from 0.918 to 0.931. Manual examination of blood smears 8. Altman DG. Practical Statistics for Medical Research. Boca
does not necessarily count IGs with high accuracy depending Raton, FL: CRC Press; 1999.
on the number of IGs in the sample.23 In previous studies, 9. Imeri F, Herkloz R, Risch L, et al. Stability of hematological
difficulties in identifying IG populations by morphology due analytes depends on the hematology analyser used: a stability
study with Bayer Advia 120, Beckman Coulter LH 750 and
to a low number of cells have been reported.24 The manual Sysmex XE 2100. Clin Chim Acta. 2008;397:68-71.
count of IGs may thus not be the perfect gold standard in 10. de Baca ME, Gulati G, Kocher W, et al. Effects of storage
an ROC curve test. In this study, however, the relatively high of blood at room temperature on hematologic parameters
numbers of IGs in the samples support the validity of the measured on Sysmex XE-2100. Lab Med. 2006;37:28-36.
ROC curve test. 11. Kang SH, Kim HK, Ham CK, et al. Comparison of four
hematology analyzers, CELL-DYN Sapphire, ADVIA 120,
Taken together, the present data confirm a systematic Coulter LH 750 and Sysmex XE-2100, in terms of clinical
bias in IG counting between the manual slide method and the usefulness. Int J Lab Hematol. 2008;30:480-486.
Sysmex XE-5000 method and that the analytical quality and 12. Ruzicka K, Veitl M, Thalammer-Scherrer R, et al. The
the diagnostic accuracy of the instrument counting are high. new hematology analyzer Sysmex XE-2100: performance
evaluation of a novel white blood cell differential technology.
This is an important finding since it has been demonstrated Arch Pathol Lab Med. 2001;125:391-396.
that the IG count may have a central role as an inflammation 13. Stamminger G, Auch D, Diem H, et al. Performance of
marker, especially in differentiating between infection and the XE-2100 leucocyte differential. Clin Lab Haematol.
inflammation of other causes.25-28 2002;24:271-280.
The agreement between the IG counts from the 14. Eilertsen H, Vllestad NK, Hagve TA. The usefulness of
blast flags on the Sysmex XE-5000 is questionable. Am J Clin
XE-5000 instruments is good. The IG count from the Pathol. 2013;139:633-640.
XE-5000 is higher than that from the manual smear count, 15. Rumke CL. Statistical reflections on finding atypical cells.
and the difference between the two methods increases Blood Cells. 1985;11:141-144.
with an increasing number of IGs. The lack of agreement 16. Koepke JA, Dotson MA, Shifmann MA. A critical
between the manual IG count and the IG count from the evaluation of the manual/visual differential leukocyte
counting method. Blood Cells. 1985;11:173-186.
XE-5000 may be linked to errors in the manual counting.
17. Fernandes B, Hamaguchi Y. Automated enumeration of
The Imm Gran? flag has poor analytical quality and immature granulocytes. Am J Clin Pathol. 2007;128:454-463.
gives no substantial information on the presence of blasts 18. Field D, Taube E, Heumann S. Performance evaluation of
in the sample. Therefore, we suggest reporting the IG immature granulocyte parameter on the Sysmex XE-2100
count from the XE-5000 without smear confirmation. automated hematology analyzer. Lab Hematol. 2006;12:11-14.
19. Cherian S, Levin G, Wing YL, et al. Evaluation of an 8-color
flow cytometric reference method for white blood cell differential
Address reprint requests to Ms Eilertsen: Faculty of Health enumeration. Cytometry B Clin Cytom. 2010;78:319-328.
Sciences, Oslo and Akershus University College of Applied 20. Stiene-Martin EA. Causes for poor leukocyte distribution
Sciences, PO Box 4, St Olavs plass, N-0130 Oslo, Norway; heidi. in manual spreader-slide blood films. Am J Med Tech.
eilertsen@hioa.no. 1980;46:624-632.

American Society for Clinical Pathology Am J Clin Pathol 2014;142:553-560 559


559 DOI: 10.1309/ AJCP4V4EXYFFOELL 559
Eilertsen and Hagve / Usefulness of Immature Granulocyte Flags on the Sysmex XE-5000

21. Novak RW. The beleaguered band count. Clin Lab 26. Briggs C, Kunka S, Fujimoto H, et al. Evaluation of immature
Med.1993;14:895-903. granulocyte counts by the XE-IG master: upgraded software
22. Briggs C, Harrison P, Grant D, et al. Performance evaluation for the XE-2100 automated hematology analyzer. Lab
of the Sysmex XE-2100, automated haematology analyser. Hematol. 2003;9:117-124.
Sysmex J Int. 1999;9:113-119. 27. Ansari-Lari AM, Kickler T, Borowitz M. Immature
23. Valenstein PN. Evaluating diagnostic tests with imperfect granulocyte measurement using the Sysmex XE-2100:
standards. Am J Clin Pathol. 1990;93:252-258. relationship to infection and sepsis. Am J Clin Pathol.
2003;120:795-799.
24. Fujimoto H, Sakata T, Hamaguchi Y, et al. Flow cytometric
method for enumeration and classification of reactive immature 28. Nigro KG, ORiordan MA, Molloy EJ, et al. Performance of
granulocyte populations. Cytometry. 2000;42:371-378. an automated immature granulocyte count as a predictor of
neonatal sepsis. Am J Clin Pathol. 2005;123:618-624.
25. Nierhaus A, Klatte S, Linssen J, et al. Revisiting the white
blood cell count: immature granulocytes count as a diagnostic
marker to discriminate between SIRS and sepsisa
prospective, observational study. BMC Immunol. 2013;14:8.

560 Am J Clin Pathol 2014;142:553-560 American Society for Clinical Pathology


560 DOI: 10.1309/AJCP4V4EXYFFOELL

You might also like