Volume Conversion Device EK220

Short-Form Instructions (73020051)


Volume Conversion Device Type EK220

The display is activated on pressing any key and the


momentary counter reading Vb (standard volume) is
EK220 displayed in the menu User. All further data
Status Menu
associated with the standard volume can be
displayed by pressing the key (see first column
"Standard volume" on the inside page of these
Volume Conversion Device
Year: 2007
M 07 0102 instructions).
MPE at ref. conditions: 0,5%
IP 65
S/N: XXXXXXX More EN12405-1 DE-07-MI 002-PTB xyz
info: Press key.
The key is pressed to display a value in the
column, standard volume. Now, the menu Std. V
(Standard volume) is displayed. Using the keys
Made in Germany
and - , you can view all the values associated with
the standard volume.
To change to a different menu (e.g.: Press.), the keys
Instrom et
or are pressed until the desired menu name
appears in the display (see inside page of the operating
instructions).
Entering values
Values in the volume conversion device which are not subject to the calibration lock or only computed
(e.g. flow) or measured (e.g. pressure or temperature) can be changed. even without a PC or readout
device.
In these short-form instructions all values which are subject to the calibration lock are identified with "C".
All values which are determined or measured and therefore can only be read are identified with a "".
Example of changing a value
( adjustable counter in the menu actual volume (Act.V.)
- The display is activated by pressing any key.
The momentary counter reading Vb (standard volume) is indicated in the menu User in the display.
Prefix Arc. Status Menu

o k . U s e r Submenu

V b 0 0 0 0 0 1 2 3 4 m 3
- Changing to the column Actual volume occurs by pressing the key
The momentary counter reading V (actual volume) is indicated in the menu Act.V. in the display.
Prefix Arc. Status Menu

o k . A c t . V . Submenu

V m 0 0 0 0 0 1 2 3 4 m 3
- Within the menu, Actual volume, repeated pressing of the key changes to the value VmA
(adjustable counter).
Prefix Arc. Status Menu

o k . A c t . V . Submenu

V m A 0 0 0 0 0 2 3 4 5 m 3
- The entry-mode is activated by pressing and - keys (ENTER).
The modifiable display location flashes.
The keys and enable skipping to the other places of the displayed value.
These can be changed with the keys and - and refreshed by pressing the ENTER keys.
Prefix Arc. Status Menu

V m A Submenu

0 0 0 0 0 2 3 4 5 . 0 0 0 0

Pressing and keys (before pressing ENTER) causes the entry to be cancelled.

Elster GmbH August 2008 Technical modifications reserved SW version from V1.20 Page: 1 of 4
Volume Conversion Device EK220

User list Stand. volume Actual volume Pressure Temperature


to to
Energ.
User Std.V. Act.V. Press Temp. Conv.

Vb Vb Vm p T
Vol. at base PL Vol. at base PL PL - -
Actual volume Pressure Temperature
cond. (pre-dec.) cond. (post-dec.)
VmA S Qb - Qm - pMin C Tmin C
Act. vol. adjust. Flow at base cond. Actual flow Lower alarm limit Lower alarm limit
p - VbD S VmD S pMax Tmax
C C
Pressure Dist. quant. Disturbance quant. Upper alarm limit Upper alarm limit
T - VbT - VmT - MRL.p MRL.T
C C
Temperature Total quantity Total quantity Meas. range bottom Meas. range bottom
Z - VbA S VmA S MRU.p MRU.T
C C
Compres. factor Adjustable count. Adjustable counter Meas. range top Meas. range top
Zb VbME VmME p.F T.F
Compres. factor C - - S S
Month end value Month end value Substitute value Substitute value
at base cond.
C - Time Time pb Tb
- - C C
Conversion fact. Time of VnME Time of VbME Press. at base cond. Tem. at base cond.

K.F S Md.p Md.T


K subst. value
C C
Access rights Pressure mode Temp. mode
VbME The EK220 differentiates between the following Typ.p Typ.T
- access parties. C C
Month end value Press. sens. type Temp. sens. type
C = Calibration lock; PL = Certification data log;
Time - M = Manufactures lock; S = Suplieres lock; SNp C SNT C
Time of VbME K = Customers lock Serial no. of sen. Serial no. sensor
C/S Calibration or supplier's lock, depending on SMenu p coeff. SMenu T coeff.
VmME - S S
national regulations. In case of verification (C) (C)
Month end value of device under calibration regulations (e.g.
Submenu p coeff. Submenu p coeff.
acc. MID) the write access C must be
Time - used. p1Adj C/S TAdj1 C/S
Time of VmME - Values which are measured or computed Adjustment val. 1 Adjustment val. 1
by the volume corrector, but can only be
Menu p2Adj TAdj2
S layed and are identified with a dash. C/S C/S
Dislpay menu () Letters in brackets: Values can only be Adjustment val. 2 Adjustment val. 2
changed via interface and not via keypad. Prog Prog
C/S C/S
Accept adjust Accept adjust.

Entry errors p.atm T.Mes


C -
Entry errors are displayed if incorrect entries are Amb. press. fixed
Temp. meas.
value
made by the operator via the keypad. After the entry
key is released, the display skips back to the original p.Mes
-
state. Pressure meas
----x---- the possible error codes correspond to the p.Abs
following table. -
Absolute press.
Example: SMenu Press.2
S (C)
o k I n p u t s Submen p sen. 2
. (acc. to Selp2)
C P . I 1 - - - - 5 - - - -
p2Mes
Pres. meas p -
sens. 2
(acc. to Selp2)

Code Description Code Description


The entry of the calorific value Ho.b in the energy list is not
1 The archive is empty 11
permitted. Please change Ho.b in the Volume corrector list
2 The archive value cannot be read. 12 The entry of this source (address) is not permitted.
4 Parameter cannot be changed (constant) 13 Clock has to be set to its starting value
5 No authorisation for changing the value 14 Gas analysis parameters for AGA-NX-19 do not match
6 Invalid value. 20 Value for the application-specific display is not defined
7 Incorrect combination Because the certification data log is full the calibration lock
21
8 Entry not possible due to special setting must be opened for changing the parameter.

Elster GmbH August 2008 Technical modifications reserved SW version from V1.20 Page: 2 of 4
Volume Conversion Device EK220

Conversion Archive Status + Logbook System


zu
to
Temp.
Conv. Archiv Status System Serv.

C - ArMo1 Arc (S) S.Reg (S) Time S


Conversion factor Month archive 1 Status register Date and time
K ArMo2 Arc Stat MdTim
Compressibility - (S) - Daylight sav.: yes/no S
Month archive 2 Momentary status
ratio factor
pbX ArDay Arc Clr MCyc
pb for gasanalysis S (S) S Measurement cycle C
Day archive Clear S.Reg
(acc. to Md.K) time
TbX ArMP Arc Logb. MCPul
S (S) (S) Measurement cycle C
Tb for gasanalysis Meas. per. archive Logbook triggered by input
pulse
Ho.b MPer AudTr OCyc
Calorific value S S (S) S
Meas. period Audit Trail Operating cycle time
(acc. to Md.K)
CO2 MP.Re PLogB Arc Disp
S - Certification data log - S
Carbon diox. cont. Rem. time mea. p. Disp. switch-off time
(acc. to PL)
H2 FrMP ClrPL Aut.V
Hydrogen content S S Clear certification C C
Freeze ArMP Disp. changover time
(acc. to Md.K) data log (acc. to PL)
Rhob SNo
Density gas at base S -
Serial no. device
cond. (acc. to Md.K)
N2 Ta.Rg
Nitrogen content S C
Ambient temp.
(acc. to Md.K)
Clr
dv Vers
Density ratio S Clear S.Reg -
Software version
(acc. to Md.K)
K.F S Chk -
Substitute value Software checksum
Md.K C/S
K mode Clear status register
Warnings (W) and / or alarms (A) which are no longer prevailing, i.e. only
K mode changeover displayed for information, but no longer flashig, are cleared in the menu
Md.K = 0 "Status" - "Clr" by pressing the - + keys (ENTER) . To the right in the
display a "0" flashes. By pressing the - key the value is set to "1". Pressing
the - + keys (ENTER) again clears the status register and ok appears in
Conv. the display. Alarm or warning statuses still prevailing are then again indicated
C - with the letter A and/or W flashing in the display.
Conversion factor Device status
Ho.n S Output in the first line of the display:
Calorific value
K.F A = Alarm; W = Warning; B = Batteries discharged; P = Calibration lock open;
S L = Certification data log ( Calibration logbook) full, o = online
Substitute value
Md.K A flashing device status signals a currently prevailing message, a steady device
C/S status shows a message no longer prevailing, but which is still located in the
K mode
status register. The status register "S.Reg" documents all alarms and warnings
since the last "clear".

Example of the display of a temperature sensor


- Change to the column Status by repeated pressing of the key .
- The "1" on the right in the display, indicates a temperature alarm.
- Press ENTER to recall Status information as short text

Status Menu Status Menu


ENTER
A S t a t u s A S R . 6
S . R e g 1 # 1 T A l a r m L i m .

Elster GmbH August 2008 Technical modifications reserved SW version from V1.20 Page: 3 of 4
Volume Conversion Device EK220
Service Inputs Outputs Interfaces Energy
to
Serv. Inputs Outp. Ser.IO Energy to
System User

Bat.R - cp.I1 C/S Md.O1 S Md.S2 S W S


Remaining bat. life cp-value Input 1 Mode for outp. 1 Mode interface 2 Energy
Bat.C cp.I2 SC.O1 DF.S2 P
S S S S -
Battery capacity cp-value Input 2 Source for out. 1 Data format interf. 2 Power
St.SL Md.I2 cp.O1 Bd.S2 WD
K S S S S
Supplier lock Mode for Input 2 cp value outp. 1 Baud rate interface 2 W disturb.
Cod.S St.I2 SpO1 TypS2 W.T
S - Status pointer S Type interface 2 S -
Supplier code Status on Inp. 2 W total
for output 1 (acc. to Md.S2)
St.CL MdMI2 Md.O2 BusS2 W.A
K Mode S S Bus mode RS485 on / S S
Customer lock Mode for outp. 2 W adjust.
monitoring I2 off (acc. to Md.S2)
Cod.C SC.I2 SC.O2 Num.T Ho.b
K S S No. of ringing tones S S
Source Source for Calorific
Customer code before accept. call
monitoring I2 output 2 value for W
(acc. to Md.S2)
St.PL L1.I2 cp.O2 M.INI WME
K S S Initialise modem S Month end -
Calibration lock Limit 1 for I2 cp value outp. 2
(acc. to Md.S2) value
Contr L2.I2 SpO2 SMenu S Time
S S Status pointer S Submenu GSM&SMS (C) Time of -
Display contrast Limit 2 for I2
for output 2 para. (acc. to Md.S2) WME
Adj.T SpI2 Md.O3 DProt S
C Stat. pointer S S Submenu IDOM prot. (C)
Clock adjust. factor Mode for outp. 3
mon. I2 (acc. to Md.S2)
Save St.I3 SC.O3 SMenu S
S - Source for S Submenu Modbus (C)
Savbe all data Status on Inp. 3
output 3 para. (acc. to Md.S2)
Clr.A C MdMI3 S cp.O3 S Bd.S1 S
Clear achives Mode moni. I3 cp value outp. 3 Baudrate interface 1
Clr.V SC.I3 SpO3 CW1.S
C Source S Status pointer S S
Clear counter Call window 1 start
monitoring I3 for output 3
Clr.X C L1.I3 S Md.O4 S CW1.E S
Initialise device Limit 1 for I3 Mode for outp. 4 Call window 1 end
Bin.T SpI3 SC.O4 CW2.S
- Stat. pointer S S S
Temp. binary value Source for out. 4 Call window 2 start
mon. I3
Bin.p - SNM S cp.O4 S CW2.E S
Press. binary value SerNo gasmeter cp value outp. 4 Call window 2 end
Bin2p SpO4 CW3.S
Press. 2 binary - Status pointer S S
Call window 3 start
value for out. 4
Sel.T C CW3.E S
Select T sensor Call window 3 end
Sel.p C CW4.S S
Select p sensor Call window 4 start
Selp2 C CW4.E S
Select p sensor 2 Call window 4 end
SMenu S CWTst
Submenu ambient (C) Test call window S
temperature (acc. to Md.S2)
Addr S
Addr. user display
...
User display
SMenu S (C)
Submenu Revisal
ArCal (S)
Frozen data
Frz S
Freeze
- -
Display test

Elster GmbH August 2008 Technical modifications reserved SW version from V1.20 Page: 4 of 4