Nemko ref. no.

: 200043138

TEST REPORT
MEDICAL ELECTRICAL EQUIPMENT
Equipment / Product LCD type Computer
Name and address of the applicant Advantech Co., Ltd.
4F, No. 108-3, Ming-Chuan Road.
Shing-Tien City, Taipei
TAIWAN, R.O.C.
Name and address of the Advantech Co., Ltd.
manufacturer 4F, No. 108-3, Ming-Chuan Road.,
Shing-Tien City, Taipei
TAIWAN, R.O.C.
Name and address of the factory Advantech Co., Ltd.
Fl. 7, NO. 1, Lane 169, Kang-Ning Street
Xi-Zhi Town, Taipei Hsien
TAIWAN, R.O.C.
Trade mark ADVANTECH
Model/type PPC-xxyM (Explanation for xxy on page 3 in this report)
Rating and principal 100-250Vac, 3A, 50/60Hz, Class I equipment
characteristics
Serial no
Test sample(s) received, date 24. oct. - 2000
Tested in the period, dates Oct. -Nov. - 2000
Tested according to IEC 60601-1 (2 ed 1988) + Amend. 1 (1991) + Amend. 2 (1995)
+ Corrigendum (June 1995)
MEDICAL ELECTRICAL EQUIPMENT - Part 1: General requirements for safety
Result of testing The equipment complies with the above mentioned standards.
The test results relate only to the sample(s) tested.
Name and address of the testing Telephone
laboratory (+47) 22 96 03 30
P.O. BOX 73 BLINDERN, Fax
N - 0314 OSLO, NORWAY (+47) 22 96 05 50

Tested by
Signature date
Johny Giang

Verified by
Signature date
Frank Skarpsno
Nemko AS

Verdicts are placed in the column to the right.: P = Pass, F = Fail, N = Not applicable, = Considered/Information.
Due to Nemkos computerised handling of test reports the layout of this form is modified compared to the original TRF published by EMEDCA; 1992-12-01.
The content fully covers the original TRF.
This form is only for use by Nemko, or by others according to special agreement with Nemko. The completed test report is not valid for external use, unless
issued by Nemko or attached to a Nemko certification document. The report shall not be reproduced without written permission from Nemko and may then only
be copied in full.
Issue Nemko 97-09-05

Page 1 of 73
Nemko ref. no.: 200043138

DESCRIPTION OF EQUIPMENT UNDER TEST:


The PPC-150 panel PC is a multimedia Pentium MMX processor-based computer that is
designed to serve as a human machine interface (HMI) and as a desktop computer. It is a PC-
based system with 15 color TFT LCD display, on-board PCI Ethernet controller, multi-COM
port interfaces and a 16-bit audio controller. With built-in CD-ROM drive, floppy drive and
PCMCIA ezpansion sockets.

NAME AND ADDRESS OF PRODUCTION-SITES (FACTORIES):


See front page.

INFORMATION ABOUT THE STANDARDS / DOCUMENTS CONSIDERED:


EN 60601-1 (1990) + A1 (1993) + A2 (1995) +A13 (1996) +
Corrigenda (July 1994).
EMC standard : EN 60601-1-2 (1993)

TESTED ACCORDING TO NATIONAL REQUIREMENTS FOR THE


FOLLOWING COUNTRIES:

LIST OF APPENDIXES / ENCLOSURES TO THE TEST REPORT:


See page 4.

Page 2 of 73
Nemko ref. no.: 200043138

SUMMARY OF TESTING
Evaluation of Results
If not explicitly stated otherwise in the standard, the test is passed if the measurement value is equal to or below the limit line,
regardless of the uncertainty of the measurement. If the measurement value is above the limit line, the test is not passed - ref.
IEC 60EE/CTL (sec) 056/94 (CLT = Committee of Testing Laboratories).
The instrument accuracy is within limits agreed by this committee (ref. Nemko proc P227):

Models PPC-xxyM
where xx can be 12 or 15 for 12 LCD or 15 LCD
where y can be 0-9

This computer uses adaptor (Skynet, Model SNP -8086-M which has CB)

Clause Remarks Information/Comments

Page 3 of 73
Nemko ref. no.: 200043138

LIST OF ATTACHMENTS PROVIDING FURTHER INFORMATION


ON THE EQUIPMENT TESTED AND THE TEST METHODS
The following attachments are added to this TRF or kept in file at the Testing Station
mentioned at the cover page:
Statements on calibrations and measurement uncertainties (where relevant) of the
measurement and test equipment as identified throughout this TRF:
Calibration All instruments used in the tests given in this report are calibrated and
traceable to international standards.
Further information about tractability will be given on request.
Measurement Measurements uncertainties are calculated for all instruments and
uncertainties instruments set-ups given in this report. Calculations are based on the
principles given in the standard EA -4/02 (Dec. 1999).
Further information about measurements uncertainties will be given on
request
Document title/identification: Kept in file at Nemko; P31000 ATT. No.: Not attached
Lis t of worksheets which describe measurement procedures or test methods where
practicable:
Document title/identification: Kept in file at Nemko; P31000 ATT. No.: Not attached
List of equipment or units tested and/or accompanying units and accessories (in case not
all individual units can be mentioned on the front-page):
Document title/identification: See summary of testing in this report. ATT. No.: Not attached
Technical design documents of components having basic, supplementary of reinforced
insulation (e.g. transformer specification mentioning the insulation class of the insulation
materials applied - A to H):
Document title/identification: Kept in file at Nemko ATT. No.: Not attached
Description of safety systems (specifying for example safety circuits, redundant circuits,
double circuits, hardware/software safety measures, temperature controls, etc.):
Document title/identification: Kept in file at Nemko ATT. No.: Not attached

PERSONNEL - INITIALS USED THROUGHOUT THIS TRF


Initials: Full name: Signatures:

This Part 1 TRF contains 73 pages, numbered from 1 to 73.


This TRF contains photographs. Attachment no. 1
Attachment no. 2. EMC report no. 200045135

Page 4 of 73
Nemko ref. no.: 200043138

3 GENERAL REQUIREMENTS
3.1 EQUIPMENT when transported, The equipment causes no hazards when used according P
stored, installed, operated in NORMAL the manufacturers instructions.
USE and maintained according to the
instructions of the manufacturer, causes
no SAFETY HAZARD which could
reasonably be fore- seen and which is
not connected with its intended
application in NORMAL CONDITION
and in S.F.C.
3.4 An alternative means of construction is No alternative construction. N
used to that detailed in this standard
and it can be demonstrated that an
equivalent degree of safety is obtained.

5 CLASSIFICATION
5.1 Type of protection against electric Class I equipment.
shock.
5.2 Degree of protection against electric No applied part. This is a Personal Computer. As the
shock. equipment is not intended to be connected to the patient
and does not have any patient applied parts, it is not
marked with the type B applied part symbol.
5.3 Classification of the equipment against Ordinary equipment.
ingress of liquids.
5.5 Degree of safety of application in the The equipment is not an AP or APG category
presence of a FLAMMABLE equipment.
ANAESTHETIC MIXTURE WITH
AIR or WITH OXYGEN OR
NITROUS OXIDE
5.6 Mode of operation. Continuous operation.

Page 5 of 73
Nemko ref. no.: 200043138

INSULATION DIAGRAM
Protection against electric shock - Refer to table on next page. P
Block diagram of system
Drawing

A-a2

Prim. Sec.
A-f

A-e
A-k

B-a

Comments

Page 6 of 73
Nemko ref. no.: 200043138

TABLE TO INSULATION DIAGRAM ON PREVIOUS PAGE


Distance Insulation type Maximum Required distances (mm) Dielectric strength
(Test Refer to 20.1 + 20.2) Basic/Supplement./ circuit test voltage
Double/Reinforced voltage Clearance Creepage Refer to 20.3
Insulation

A-a2 Double/Reinforce. 250 V a.c. 5,0 mm 8,0 mm 4 000 V a.c.

A-k Double/Reinforce. 250 V a.c. 5,0 mm 8,0 mm 4 000 V a.c.


A-e Double/Reinforce. 250 V a.c. 5,0 mm 8,0 mm 4 000 V a.c.

A-f Basic 250 V a.c. 1,6 mm 3,0 mm 1 500 V a.c.

B-a Double/Reinforce. 250 V a.c. 5,0 mm 8,0 mm 4 000 V a.c.

Comments

Page 7 of 73
Nemko ref. no.: 200043138

6 IDENTIFICATION, MARKING AND DOCUMENTS


6.1 Marking on the outside Example of marking label:

6.1 a Markings of Mains operated P


EQUIPMENT

6.1 b Markings of Internally Powered Not applicable. Not internally powered equipment N
EQUIPMENT
6.1 c Markings of EQUIPMENT supplied Not applicable. N
from a specified power supply
6.1 d Minimum requirements if limited space Not applicable. N
for marking
6.1 e Name and/or trademark of the See front page P
manufacturer or supplier
6.1 f Model or type reference See front page P
6.1 g Rated supply voltage(s) or voltage 100-250 V Single phase a.c. P
range(s)
Number of phases
Type of current 3A

6.1 h Rated frequency or rated frequency 50/60 Hz. P


range(s) in Hz

Page 8 of 73
Nemko ref. no.: 200043138

6.1 j Rated power input (VA, W or A) 3A P


6.1 k Power output of auxiliary mains socket No N
outlets
6.1 l Class II symbol Not applicable. Not Class II. N
Symbol for degree of protection with Ordinary equipment. IP X0. N
respect to harmful ingress of water
according to EN 60529 (IPX0 not
required to be marked)
Symbol for protection against electric Personal Computer. As the equipment is not intended to N
shock: be connected to the patient and does not have any
patient applied parts, it is not marked with the type B
applied part symbol.
6.1 m Mode of operation (if no marking, Continuous operation. P
suitable for continuous operation)
6.1 n Types and rating of external accessible No external fuses. N
fuses
6.1 p Rated output voltage and current or No auxiliary mains socket outlets. N
power, output frequency (where
applicable)
6.1 q Symbol for physiological effect(s): Not applicable. No need for marking. N
6.1 r Anaesthetic-proof symbol: Not of category AP or APG. N
6.1 s High voltage symbol: N
Symbol 6 of table DII, Dangerous voltage is used.

6.1 t Special cooling requirements Fan used P


6.1 u) Limited mechanical stability 10 P
6.1 v Protective packing requirement(s). P
Marking(s) for unpacking safety
hazard(s)
6.1 y Earth terminals: Appliance inlet used P
6.1 z Removable protective means P
Durability of markings Tested with a cloth of rack soaked with, in turn, water, P
methylated spirit and isopropyl alcohol.
The sample label withstood the test.
6.2 Marking on the inside
6.2 a Marking clearly legible P

Nominal supply voltage of Not permanently installed equipment N


permanently installed equipment
marked inside or outside of equipment
6.2 b Maximum power loading of heating No heating elements used N
elements or lamp holders for heating
lamps clearly and indelibly marked
near or in the heater
Heating elements or lamp holders for N
heating lamps not intended to be
changed by operator and which can be
changed only with the use of a tool,
have at least an identifying marking
referring to information stated in the
accompanying documents

Page 9 of 73
Nemko ref. no.: 200043138

6.2 c High voltage parts: P


Symbol 6 of table DII, Dangerous voltage is used.

6.2 d Type of battery and mode of insertion Back up battery. Type M4T28-BR12SH1, P
(if applicable) marked (see Sub-clause rating 5V, 50mAh
56.7 b)
Batteries not intended be changed by P
operator and which can be changed
only with use of tool have at least
marking referring to information in
accompanying documents
6.2 e Fuse type and rating or reference stated GBP, 12A, 250Vac P

6.2 f P
Protective earth terminal marked
6.2 g Functional earth terminal marked Not applicable. N

6.2 h Terminals for supply neutral conductor Not applicable. N


in permanently installed equipment
marked (N)
6.2 j Markings required in Sub-clause 6.2 f, P
h, k and l
6.2 k The supply connections are clearly P
marked adjacent to the terminals or in
accompanying documents (for small
equipment)
6.2 l Statement for suitable wiring materials Not applicable. N
(at temperatures over 75C) locates at
or near the point of the supply
connections and is clearly discernible
after connection
6.2 n Capacitors and/or circuit parts are Not applicable. N
marked as required in Sub-clause 15 c
6.3 Marking of controls and instruments
6.3 a Mains switch clearly identified. On P
and off positions marked or
otherwise indicated
6.3 b Adequate indications of different Not applicable. N
positions of controls and other switches
6.3 c Adequate indication of the direction of N
setting devices if change of setting of a
control could cause a safety hazard
6.3 f Functions of operator controls and N
indicators identified
6.3 g Numeric indications of parameters are N
in SI units according to ISO 1000.
Units outside the International System
as specified
6.4 Symbols
6.4 a Marking symbols compliance with P
appendix D, where applicable

Page 10 of 73
Nemko ref. no.: 200043138

6.4 b Symbols for controls and performance Not applicable. N


conform to IEC 60 878, where
applicable
Durability of marking symbols Tested in Clause 6.1. P
6.5 Colours of insulation of conductors
6.5 a Protective earth conductor has P
green/yellow insulation
6.5 b All insulations of internal protective P
earth conductors are green/yellow, at
least at the terminations of the
conductors
6.5 c Only protective earth-, functional Not used N
earth-, potential equalisation and inside
earthing conductors (cf. 6.5 b) are
green/yellow
6.5 d Colour of neutral conductor: light blue P
according to IEC 60 227 or 245
6.5 e Colours of phase conductors in power P
supply cord according to IEC 60 227 or
245
6.5 f Additional protective earthing in multi- P
conductor cords are marked
green/yellow at the ends of the
additional conductors
6.6 Identification of medical gas
cylinders and connections
6.6 a Identification of content in accordance No medical gas used N
with ISO/R32
6.6 b Identification of connection point that N
errors are avoided when a replacement
is made

Page 11 of 73
Nemko ref. no.: 200043138

6.7 Indicator lights and push-buttons


6.7 a Colour red only used for warning of No colour red used N
danger or/and need for urgent action.
Dot-matrix/alphanumeric displays not
considered to be indicator lights
6.7 b Colours of unilluminated push-buttons. N
Colour red only used in case of
emergency
Location Meaning of indicator lights Illuminated push-button (see IEC 60 Colour
73)



6.8 Accompanying documents
6.8.1 Equipment is accompanied at least by
- instructions for use User Manual P
- technical description User Manual P
- an address to which the user can refer User Manual P
All applicable classifications specified User Manual P
in Clause 5 are included in instructions
for use and technical description
markings in Sub-clause 6.1 included in P
accompanying documents if not
permanently affixed to equipment
Warning statements and explanations Not used N
of warning symbols are provided in
accompanying documents
Language of accompanying documents English. Will be in the language required for the market P
where the equipment is intended to be sold.

Page 12 of 73
Nemko ref. no.: 200043138

6.8.2 Instructions for use


6.8.2 a General information
Necessary information to operate the P
equipment
Explanation of the function of controls, P
displays and signals
Sequence of operation P
Connection and disconnection of Detachable power supply cord P
detachable parts and accessories
Replacement of material which is Not applicable. N
consumed during operation
Indications of recognised accessories, Not applicable. N
detachable parts and materials, if the
use of other parts or materials can
degrade minimum safety
Cleaning, preventive inspection and User Manual P
maintenance to be performed including
the frequency of such maintenance
Information about safe performance of P
routine maintenance
Information about preventive P
inspection and maintenance to be
performed by other persons
Meanings of figures, symbols, warning P
statements and abbreviations on
equipment explained in the instructions
for use
Instruction for use states the function P
and intended application of equipment
Instructions for use provide user with Not applicable. N
information regarding potential
electromagnetic or interference and
advice how to avoid such interference
6.8.2 c Signal output and signal input parts
Signal output or signal input parts Not applicable. N
intended only for connection to
specified equipment stated in
instruction for use
6.8.2 d Cleaning, disinfection and sterilisation
of parts in contact with the patient
Details about cleaning or disinfection Properly described. P
or sterilisation methods that may be
used for equipment parts which come
into contact with the patient during
normal use given in instruction for use

Page 13 of 73
Nemko ref. no.: 200043138

6.8.2 e Mains operated equipment with


additional power source
A warning statement referring to the Not applicable. No used of additional power source N
necessity for periodical checking or
replacement of an additional power
source
If Class I equipment a statement saying Not applicable. N
that where the integrity of the external
protective conductor in the installation
or its arrangement is in doubt,
equipment shall be operated from its
internal electrical power source
6.8.2 f Removal of primary batteries (i.e. not No primary batteries used
rechargeable batteries):
Instruction for use contains a warning N
to remove batteries if equipment is not
likely to be used for some time ,unless
no risk of safety hazard
6.8.2 g Rechargeable batteries: Not rechargeable batteries
Instructions to ensure safe use and N
adequate maintenance
6.8.2 h Equipment with a specified power
supply or battery charger:
Instructions for use identify power Not applicable. N
supplies or battery chargers necessary
to ensure compliance with the
requirements of IEC 601-1
6.8.2 j Instructions for use identify risks Not applicable. N
associated with disposal of waste,
residues etc. and of
equipment/accessories at end of their
lives. Further it provides advice on
minimising these risks
6.8.3 Technical description
6.8.3 a General
Technical description provides all data The Technical and Service Manual is well written and P
essential for safe operation including contains all necessary information required.
data in Sub-clause 6.1 and all
characteristics of the equipment
A statement whether particular Not applicable. N
measures or particular conditions are to
be observed for installing equipment
and bringing equipment into use
6.8.3 b Replacement of fuses and other parts :
Required type and rating of fuses Not applicable. Notpermanently installed equipment N
utilised in the mains supply circuit
external to permanently installed
equipment
Instructions for replacement of Not applicable. N
interchangeable and/or detachable parts
which are subject to deterioration
during normal use

Page 14 of 73
Nemko ref. no.: 200043138

6.8.3 c Circuit diagrams, component part lists,


etc.:
The technical description contains a Not applicable. N
statement that the supplier will make
available on request circuit diagrams,
component part lists, descriptions,
calibration instructions, etc. in order to
assist user in case of necessary
reparations
6.8.3 d Environmental conditions for transport
and storage:
The technical description contains a P
specification of the permissible
environmental conditions for transport
and storage
The above said specification also P
repeated on the outside of the
packaging

7 POWER INPUT
7.1 Steady state current or power input Input power: 100-250Vac, 3A, 50/60Hz P
does not exceed the marked rating by
Deviation: (measured-rated)*100/rated = %
more than specified
Equipment: N3635
Initials Tester / Date: JohnyG / 10-2000
Power input Function UNlower = 90 V UN = 170 V UN upper = 275V
Input current Normal operation (A) 1,4/1,4 0,98/0,98 0,6/0,6
Standby (A)
Input power Normal operation (W) 69/69 68/67 68/68
Input power Normal operation (VA)
Power factor cos 0,54/0,54 0,48/0,47 0,40/0,41

Comments Power input rated is measured at both 50 and 60Hz. The 50Hz values are showed before /
and the 60Hz is back. (Remark: #PPC-15yM-zT)

10 ENVIRONMENTAL CONDITIONS
10.1 Equipment capable, while packed for Considered OK. P
transport and storage, of being exposed
to environmental conditions as stated
by manufacturer (see 6.8.3 d)
10.2 Operation of equipment according to Considered OK. P
specified environment and power
supply

Page 15 of 73
Nemko ref. no.: 200043138

13 GENERAL
Equipment so designed that risk of Considered OK. P
electric shock obviated as far as
practicable (requirements as given in
section three)

14 REQUIREMENTS RELATED TO CLASSIFICATION


14.1 Class I equipment
14.1 a Parts with Double Insulation Transformer P
Parts with Reinforced Insulation Transformer P
Parts with SELV-Voltage N
Parts with Protective Impedance N
14.1 b If the mains part of equipment No external d.c. power source used N
specified for an external d.c. power
source is isolated from accessible
conductive parts by basic insulation
only, a separate protective earth
conductor is provided
14.2 Class II equipment
14.2 a Class II equipment is of one of the Not applicable. Class I equipment N
following types:
14.2 b If equipment is fitted with a device for Not applicable. N
changing over from class I to class II
protection, the following requirements
are fulfilled:
The change-over device indicates the N
selected class clearly
A tool for change-over is necessary N
The equipment complies with all N
requirements for the selected class at
any given time
Class II position: N
The device interrupts the connection of
protective earth conductor to
equipment or changes it into functional
earth conductor (compliance with 18.2)
14.2 c Class II equipment provided with N
functional earth connections (cf. Sub-
clause 18 k and l)
14.4 Class I and II equipment
14.4 a Equipment is provided with an P
additional protection according to the
requirements of Class I or Class II
equipment

Page 16 of 73
Nemko ref. no.: 200043138

14.4 b No safety hazard develop, when a Not applicable. N


connection with the wrong polarity is
made in equipment specified for power
supply from an external d.c. power
source (e.g. for use in ambulances)
14.5 Internally powered equipment
14.5 b Equipment also having means of Not applicable. N
connection to supply mains complies
also with requirements for Class I or II
while so connected
14.6 Types B, BF and CF applied parts
14.6 c Applied parts suitable for direct cardiac Not applicable. N
application are of type CF

Page 17 of 73
Nemko ref. no.: 200043138

15 LIMITATION OF VOLTAGE AND/OR ENERGY


15 b Equipment with mains plug so P
designed that the voltage 1 sec. after
disconnection does not exceed 60 V
Interference suppression capacitors P
Capacitance between each line and 3300pF P
protective earth. Measured value
Equipment: N2843
Initials Tester / Date: JohnyG / 10-2000
Maximum residual voltage in the mains plug 1 second after disconnection (measured in the least
favourable position of the on-off switch, if any):
Remark: pins 1 and 2 are Line and Neutral supply pins. Note! PE is equal to enclosure.
Measurement no. 1 2 2 4 5 6 7 8 9 10
Voltage between pins 1 and 14 14 14 14 14 14 14 14 14 14
2 (Limit 60 V)
Voltage between pin 1 and 14 14 14 14 14 14 14 14 14 14
PE (Limit 60 V)
Voltage between pin 2 and 14 14 14 14 14 14 14 14 14 14
PE (Limit 60 V)
15 c Accessible live parts of capacitors or Not applicable. N
related circuits having a residual
voltage above 60 V does not have a
residual energy above 2 mJ (see 15 c
below)
A non-automatic discharging device is N
included and marked
Equipment:
Initials Tester / Date: /
Capacitor and its location Working voltage Capacitance value Residual voltage Residual energy
(V) (mF) (V) (mJ)

Page 18 of 73
Nemko ref. no.: 200043138

16 ENCLOSURES AND PROTECTIVE COVERS


16 a Protection against contact with live P
parts.
- and with parts which can become live P
in single fault conditions
Instruction for use instructs operator Not applicable. No such parts N
not to touch such parts and the patient
simultaneously
Equipment: Standard test finger : Our ref. N 2489
Initials Tester / Date: JohnyG / 10-2000
Location of opening Test result finger
Back of equipment OK

16 b Check of openings in top covers with No openings in the top covers N


the vertical suspended test rod
16 c Conductive parts of actuating Not applicable. N
mechanism of electrical controls after
the removal of handles, knobs, levers
etc. have either resistance to the
protective earth terminal max. 0.2
ohms or the separation from live parts
complies with 17 g
16 d Protection of contact with internal parts Not applicable. N
of the equipment with a circuit voltage
exceeding 25 V a.c. or 60 V d.c. which
cannot be disconnected from the supply
by external mains switch or plug device
Equipment:
Initials Tester / Date:
Location of part Covering Warning notice Remarks

16 e Protective enclosures are removable Only with the aid of tool P


only with the aid of tool or an
automatic device makes these parts not
live, when the enclosure is opened or
removed
16 f Live parts are inaccessible to the test Not applicable. N
rod through openings for adjustment of
pre-set controls

Page 19 of 73
Nemko ref. no.: 200043138

17 SEPARATION
17 a Separation method of the applied part No applied part
from live parts so that allowable
leakage currents are not exceeded
17 a 1 Basic insulation - applied part earthed N
17 a 2 By protectively earthed conductive part N
(e.g. screen)
17 a 3 By separate earthed intermediate circuit N
17 a 4 By double or reinforced insulation N
17 a 5 By protective impedances limiting . N
current to applied part
17 a 6 Other method, e.g. specified in N
particular standard
Equipment:
Initials Tester / Date:
Additional leakage current test in single fault conditions
Fault condition Earth leakage current Enclosure leakage Patient leakage Patient auxiliary
(description) current current current

17 c There is no conductive connection P


between applied parts and accessible
conductive parts, which are not
protectively earthed
17 d Supplementary insulation between Not used N
hand-held flexible shafts and motor
shafts (Class I equipment)
Adequate isolation of accessible metal N
parts from motor shaft
Rated motor voltage N
Test voltage N
Air clearances/creepage distances N
(mm)/(mm)
17 g Separation method of accessible parts
other than applied parts from live parts
so that allowable leakage currents are
not exceeded
17 g 1 Basic insulation - accessible part P
earthed
17 g 2 By protectively earthed conductive part P
(e.g. screen)
17 g 3 By separate earthed intermediate circuit Not applicable. N
17 g 4 By double or reinforced insulation Transformer N

Page 20 of 73
Nemko ref. no.: 200043138

17 g 5 By protective impedances limiting N


current to accessible part
Leakage currents measurements, when N
required
17 h Adequate arrangements used to isolate N
defibrillation-proof applied parts from
other parts
Impulse voltage tests N
Peak voltage between Y1 and Y2 does N
not exceed 1 V
After recovery time equipment N
continues to perform its intended
function

18 PROTECTIVE EARTHING, FUNCTIONAL EARTHING AND


POTENTIAL EQUALIZATION
18 a Sufficiently low impedance to the P
protective earth terminal(18 f and 18 g)
18 b Suitable connection between protective P
earth terminal and protective conductor
in the installation (18 f)
18 e Means for connection of potentially Not used N
equalisation conductor complies with
specified requirements
This connection complies with N
following requirements:
- readily accessible N
- no possibility to accidental N
disconnection in normal use
- conductor is detachable without the N
use of a tool
- power supply cord does not include N
potential equalisation conductor
- connection is marked with symbol N

Page 21 of 73
Nemko ref. no.: 200043138

18 f Impedance of protective earthing U = 0,57V P


system: Impedance between any
I = 30A
accessible metal part and:
1) protective earth terminal (PE) T = 2min.
(requirement: R max. 0.1 Ohm),
2) protective earth contact in appliance
inlet (requirement: R max.
0.1 Ohm),
3) protective earth contact in the mains
plug Test equipment: See attached
list. (requirement: R max. 0,2 Ohm),
are tabled below:
Equipment: N1514, N2332, N1585
Initials Tester / Date: JohnyG / 11-2000
Accessible part and its location R (Ohm) measured against (numbers see above)
1 PE 2 inlet 3 plug
Inlet- Ground 0,019

18 g If the impedance of protective earth No N


connections other than in 18 f exceeds
0.1 Ohm, the allowable value of the
enclosure leakage current is not
exceeded in single fault condition
18 k Functional earth terminals are not used P
to provide protective earthing
18 l For Class II equipment with isolated Class I equipment N
internal screens and with a power
supply cord:
- the third conductor is used only as N
functional earth of these screens and is
coloured green/yellow
- insulation of such screens and all N
internal wiring connected to them is
double insulation or reinforced
insulation (see also 20.3)
- marking of functional earth terminal N
is distinguished from protective earth
terminal and is explained in
accompanying documents
Test of insulation (see Clause 20) N

Page 22 of 73
Nemko ref. no.: 200043138

19 CONTINUOUS LEAKAGE CURRENTS AND PATIENTS


AUXILIARY CURRENTS
19.1 General requirements (tests see 19.4)
19.1 g Equipment with multiple patient Not applicable. N
connections: Patient leakage current
and patient auxiliary current do not
exceed allowable values while one or
more patient connections are
disconnected from the patient or
disconnected from the patient and
earthed
19.2 Single fault conditions (tests see 19.4)
19.3 Allowable values (see 19.4)
19.4 Tests before humidity preconditioning See comment below.
treatment
Equipment: N2239, N2718, N2266,N3171
Initials Tester / Date: JohnyG / 11-2000

In normal condition In single fault condition


Type of leakage current Allowed value Measured max. Allowed value Measured max.
(A) value (A) (A) value (A)
Earth leakage current 500 184 1000 381
Enclosure leakage current 100 <5 500 <5
Patient leakage current (p.l.c.), a.c. : B, BF: 100 B, BF: 500
CF: 10 CF: 50
Patient leakage current (p.l.c.), d.c. : B, BF: 10 B, BF: 50
CF: 10 CF: 50
P.l.c. (110 % of the mains voltage on the B: 5000 113
signal input part or signal output part) BF, CF:
P.l.c. (110 % of mains voltage on the B:
applied part) BF: 5000
CF: 50
Patient auxiliary current, a.c. : B, BF: 100 B, BF: 500
CF: 10 CF: 50
Patient auxiliary current, d.c. : B, BF: 10 B, BF: 50
CF: 10 CF: 50
Comments
The values listed are the maximum values.

Page 23 of 73
Nemko ref. no.: 200043138

19.4 Tests after humidity preconditioning See comments below.


treatment
Equipment: N2239, N2718, N2266,N3171
Initials Tester / Date: JohnyG / 11-2000

In normal condition In single fault condition


Type of leakage current Allowed value Measured max. Allowed value Measured max.
(A) value (A) (A) value (A)
Earth leakage current 500 185 1000 382
Enclosure leakage current 100 <5 500 <5
Patient leakage current (p.l.c.), a.c. : B, BF: 100 B, BF: 500
CF: 10 CF: 50
Patient leakage current (p.l.c.), d.c. : B, BF: 10 B, BF: 50
CF: 10 CF: 50
P.l.c. (110 % of the mains voltage on the B: 5000 113
signal input part or signal output part) BF, CF:
P.l.c. (110 % of ma ins voltage on the B:
applied part) BF: 5000
CF: 50
Patient auxiliary current, a.c. : B, BF: 100 B, BF: 500
CF: 10 CF: 50
Patient auxiliary current, d.c. : B, BF: 10 B, BF: 50
CF: 10 CF: 50
Comments
The values listed are the maximum values.

Page 24 of 73
Nemko ref. no.: 200043138

20 DIELECTRIC STRENGTH (at operating temperature)


Overall compliance with Clause 20 P
Equipment: High voltage supply : N1979
Initials Tester / Date: JohnyG / 11-2000
Equipment Insulation Insulation Reference Test voltage Remarks, observations
under test resistance *) voltage
All A-a1 250 V AC 1 500 V AC OK
equipment A-a2
types A-b
A-c

A-e 250 V AC 4 000 V AC OK


A-f 250 V AC 1 500 V AC OK
A-g

A-j
A-k 250 V AC 4 000 V AC OK
Equipment B-a 250 V AC 4 000 V AC OK
types with B-b
applied part B-c
B-d
B-e

*) IEC 601-1 does not require to measure this

Page 25 of 73
Nemko ref. no.: 200043138

20 DIELECTRIC STRENGTH (after humidity preconditioning


treatment)
Overall compliance with Clause 20 P
Equipment: Humidity room : N3205
High voltage supply : N601
Initials Tester / Date: JohnyG / 11-2000
Equipment Insulation Insulation Reference Test voltage Remarks, observations
under test resistance *) voltage
All A-a1 250 V AC 1 500 V AC OK
equipment A-a2
types A-b
A-c

A-e 250 V AC 4 000 V AC OK


A-f 250 V AC 1 500 V AC OK
A-g

A-j
A-k 250 V AC 4 000 V AC OK
Equipment B-a 250 V AC 4 000 V AC OK
types with B-b
applied part B-c
B-d
B-e

*) IEC 601-1 does not require to measure this

21 MECHANICAL STRENGTH
21 a Rigidity of enclosure (45 N test force) Tested with 45 N. No damage to enclosure parts. P
21 b Strength of enclosure part and any No damage resulting in a safety hazard. P
component thereon (Impact hammer
test, 0.5 J).
21 c On portable equipment carrying No handles or grips used N
handles or grips withstand the
requirements of the loading test. (Test
force four times the weight of
equipment)

Page 26 of 73
Nemko ref. no.: 200043138

21.3 No damage to parts of patient support No patien support parts used N


and/or immobilisation system after the
loading test (1350 N and 2700 N tests)
The test force for foot rests and chairs N
shall be twice the specified maximum
load or, if not specified, the test force
shall be 2,7 kN. The test force shall be
distributed over an area of 0,1 m2
surface for 1 minute.
21.5 Hand-held equipment or equipment Not hand-held equipment N
parts are safe after drop test (dropping
height 1 m)
21.6 a Portable and mobile equipment is able P
to withstand rough handling.
21.6 b Propel test of mobile equipment Not applicable. N
(performed 20 times). Equipment
complies with requirements of this
standard.

22 MOVING PARTS
22.2 a Moving parts of transportable No moving parts used N
equipment are provided with guards
which form an integral part of the
equipment
22.2 b Moving parts of stationary equipment N
are provided with similar guards as
above unless it is evident that
equivalent protection is separately
provided during installation
22.3 Cords (ropes), chains and bands are N
confined so they cannot run off or jump
out of their guiding devices
Other means used to prevent a safety N
hazard
Mechanical safeguard means are N
removable only with a tool
22.4 Dangerous movements of equipment N
parts, which may cause physical injury
to the patient, are possible only by the
continuous activation of the control
22.6 Parts of equipment subject to N
mechanical wear are accessible for
inspection

Page 27 of 73
Nemko ref. no.: 200043138

22.7 To remove an unexpected safety hazard N


caused by an electrically produced
mechanical movement, there are means
for emergency switching of a relevant
part
The means for emergency switching N
are readily identifiable and accessible
and do not introduce a further safety
hazard
Current breaking capability N
Means for stopping of movements N
operate as a result of one single action

23 SURFACES, CORNERS AND EDGES


There are no rough surfaces, sharp The edges are well rounded. P
corners, flange or frame edges and
burrs which may cause injury or
damage

24 STABILITY IN NORMAL USE


24.1 Equipment does not overbalance, when Tested. P
tilted through an angle of 10
24.3 If equipment overbalances when tilted See above N
10, it does not overbalance when
tilted:
- 5 in any position of normal use, N
excluding transport
- 10 in the condition specified for N
transport
The equipment carries a warning notice N
for transport
24.6 a Equipment or its parts with a mass of N
more than 20 kg:
- is provided with handling devices N
(grips etc.)
- is provided with handling instructions N
for lifting and assembling
4.6 b Portable equipment with a mass of N
more than 20 kg carrying handle(s)
suitably placed that equipment may be
carried by 2 or more persons

25 EXPELLED PARTS
25.1 Protective means are provided where Not applicable. N
expelled parts of the equipment could
constitute a safety hazard

Page 28 of 73
Nemko ref. no.: 200043138

25.2 Display vacuum tubes with a face Not applicable. N


dimension of 16 cm or larger are
intrinsically safe with respect to effects
of implosion of tubes and to
mechanical impact, or the enclosure of
the equipment is provided with
adequate protection against implosion
Certificate of the test provided N

26 VIBRATION AND NOISE


No general requirement.

27 PNEUMATIC AND HYDRAULIC POWER


No general requirement.

28 SUSPENDED MASSES
28.3 Suspension systems with safety No suspended masses N
devices:
Suspension systems include a safety N
device with adequate safety factors to
protect user or patient from hazards
If after activation of a safety device the N
equipment can still be used, the
activation of the device, e.g. a
secondary rope, becomes obvious to
the operator
28.4 Suspension systems of metal without N
safety devices:
The construction of the suspension N
complies with:
28.4 1 The TOTAL LOAD does not exceed N
the SAFE WORKING LOAD
28.4 2 Where it is unlikely that supporting N
characteristics will be impaired by
wear, corrosion, material fatigue or
ageing, the SAFETY FACTOR of all
supporting parts is not less than 4
28.4 3 Where impairment by wear, corrosion, N
material fatigue is expected, the
SAFETY FACTOR is not less than 8
28.4 4 Where metal having a specific N
elongation at break of less than 5% is
used in supporting components, the
SAFETY FACTOR is not less than 1.5
times those given in 2) and 3) above

Page 29 of 73
Nemko ref. no.: 200043138

28.4 5 Sheaves, sprockets, band wheels and N


guides are so designed that the
SAFETY FACTORS of this Sub-
clause shall be maintained for a
specified minimum life till replacement
of the ropes, chains and bands
28.5 Dynamic loads: No general requirement.

29 X-RADIATION
29.1 Diagnostic X-ray equipment, see IEC No X-radiation N
601-1-3. Radiotherapy equipment, see
relevant Particular standard
29.2 For equipment not intended to produce N
X-radiation for diagnostic and
therapeutic purposes, ionising radiation
emitted by vacuum tubes excited by
voltages exceeding 5 kV does not
produce an exposure exceeding 130
nC/kg (0.5 mR)

30 ALPHA, BETA, GAMMA, NEUTRON RADIATION AND OTHER


PARTICLE RADIATION
No general requirement.

31 MICROWAVE RADIATION
No general requirement.

32 LIGHT RADIATION (INCLUDING LASERS)


No general requirement.

33 INFRA-RED RADIATION
No general requirement.

34 ULTRAVIOLET RADIATION
No general requirement.

35 ACOUSTICAL ENERGY (INCLUDING ULTRA-SONICS)


No general requirement.

Page 30 of 73
Nemko ref. no.: 200043138

36 ELECTROMAGNETIC COMPATIBILITY
IEC 60601-1-2 ( 1993 ) Compliance documented by the manufacturer. P

37 - 41 REQUIREMENTS FOR CATEGORY AP AND APG EQUIPMENT


See additional test report form. Not applicable. N

Page 31 of 73
Nemko ref. no.: 200043138

42 EXCESSIVE TEMPERATURES
Ambient temperature during See tables below.
measurements:
Humidity: 40%rh
Atmospheric pressure: 995hPa
Equipment: Humidity : N2531
Pressure : N2842
Initials Tester / Date: JohnyG / 11-2000
42.1 - 42.2 Determination of the temperature with P
thermocouples:
Equipment: N3008, N3635, N2859, N3132
Initials Tester / Date: JohnyG / 11-2000
Quantity Rated Used in tests Remarks
Supply voltage, V 100-250V AC 90, 275 V AC
Supply power, VA
Output power, VA
Measuring point Measured Calculated Measured Calculated Allowed Remarks
temp. temp. temp. temp. max.
t (K) T (C) t (K) T (C) temp. T
(C)

A: IP 90Vac, 60Hz A B
B: IP 275Vac, 60Hz
1. Ambient, T = 25 C
2. T1 coil 40 80 43 68 Cl.B Table Xa
130C
3. T1 core 36 76 39 64 Table Xa

4. L1 coil 35 60 30 55 Table Xb
5. BT1 body 26 51 26 51 Table Xb
6. V1 body 30 55 31 56 Table Xb
7. L6 coil 29 54 29 54 Table Xb
8. T1 coil for inverter 65 105 65 105 Cl.B Table Xa
130C
9. T1 core for inverter 70 110 70 110 Table Xa
10. F1 coil for inverter 58 83 58 83 Table Xb
11. Surface of PC 13 53 13 53 75 Table Xa
12. SMP for building-in:
Skynet SNP-8086-M. See
Nemko ref. 200039133

Comments According to IEC 601-1 clause 42.1, table Xa and Xb, 400 C or 250 C is added to the temperature rise for
determination of the final temperature T.
Model PPC-150M-zT

Page 32 of 73
Nemko ref. no.: 200043138

Measuring point Measured Calculated Measured Calculated Allowed Remarks


temp. temp. temp. temp. max.
t (K) T (C) t (K) T (C) temp. T
(C)

A: IP 90Vac, 60Hz A B
B: IP 275Vac, 60Hz
1. Ambient, T = 25 C
2. L1 coil 60 85 50 75 Table Xb
3. BD1 body 69 94 57 83 Table Xb

4. C8 body 46 71 42 67 Table Xb
5. T1 coil 70 110 76 116 Cl.B Table Xa
6. T1 core 67 107 73 113 Table Xa
7. L4 coil 76 101 79 104 Table Xb
8. L3 coil 61 86 63 88 Table Xb
9. HS body of Q1 41 66 42 67 Table Xb
10. V36 body 29 54 29 54 Table Xb
11.cpu body 37 62 36 61 Table Xb
12.H.DD body 25 50 25 50 Table Xb
13. F.DD body 19 44 19 44 Table Xb
14. CD rom body 27 52 27 52 Table Xb
15. V1 body 29 54 29 54 Table Xb
16. L48 body 26 51 26 51 Table Xb
17. T1 coil for inverter 62 102 62 102 Cl.B Table Xa
18. T1 core for inverter 48 88 48 88 Table Xa
19. L1 coil for inverter 56 81 56 81 Table Xb
20. Surface of PC 13 53 13 53 75 Table Xa
12. SMP for building-in:
Skynet SNP-8086-M. See
Nemko ref. 200039133

Comments According to IEC 601-1 clause 42.1, table Xa and Xb, 400 C or 250 C is added to the temperature rise for
determination of the final temperature T.
Model PPC-120M-zT

Page 33 of 73
Nemko ref. no.: 200043138

Measuring point Measured Calculated Measured Calculated Allowed Remarks


temp. temp. temp. temp. max.
t (K) T (C) t (K) T (C) temp. T
(C)

A: IP 90Vac, 60Hz A B
B: IP 275Vac, 60Hz
1. Ambient, T = 25 C
2. L1 coil 52 77 43 68 Table Xb
3. BD1 body 65 90 50 75 Table Xb

4. C8 body 40 65 34 59 Table Xb
5. T1 coil 70 110 71 111 Cl.B Table Xa
6. T1 core 62 102 66 106 Cl.B Table Xa
7. L4 coil 70 95 71 96 Table Xb
8. L3 coil 56 81 56 81 Table Xb
9. HS body of Q1 37 62 37 62 Table Xb
10. BT1 body 19 44 18 43 Table Xb
11.V2 body 22 47 20 45 Table Xb
12. upc body 24 49 22 47 Table Xb
13. H.DD body 17 42 14 39 Table Xb
14. F.DD body 13 38 11 36 Table Xb
15. CD rom body 17 42 15 40 Table Xb
16. V3 body 25 50 24 49 Table Xb
17. T2 coil for inverter 57 97 57 97 Cl.B Table Xa
18. T2 core for inverter 54 94 55 95 Cl.B Table Xa
19. F3 coil for inverter 50 75 49 74 Table Xb
20. Surface of PC 10 50 8 48 75 Table Xa
12. SMP for building-in:
Skynet SNP-8086-M. See
Nemko ref. 200039133

Comments According to IEC 601-1 clause 42.1, table Xa and Xb, 400 C or 250 C is added to the temperature rise for
determination of the final temperature T.
Model PPC-153M-zT

Page 34 of 73
Nemko ref. no.: 200043138

42.1 - 42.2 Determination of the temperature with Block vent test P


thermocouples:
Measuring point Measured Calculated Allowed Remarks
temp. temp. max. temp.
t (K) T (C) T (C)

Power input 250Vac, 60Hz


Ambient, T = 25 C
1. L1 coil 60 85 Table Xb
2. BD1 body 67 92 Table Xb
3. C8 body 54 79 Table Xb
4. T1 coil 86 126 Cl.B Table Xa
5. T1 core 82 122 Table Xa
6. L4 coil 87 112 Table Xb
7. L3 coil 72 97 Table Xb
8. H.S body & Q1 53 87 Table Xb
9. V36 body 46 71 Table Xb
10.cpu body 55 80 Table Xb
11. H.DD body 34 59 Table Xb
12.F.DD body 24 49 Table Xb
13.CD rom body 38 63 Table Xb
14.V1 body 43 68 Table Xb
15.L48 body 44 69 Table Xb
16.T1 coil for inverter 68 108 Cl.B Table Xa
17.T1 core for inverter 55 95 Table Xa
18.L1 coil for inverter 53 78 Table Xb
19. Surface of enclosure 21 61 75 Table Xa

Comments According to IEC 601-1 clause 42.1, table Xa and Xb, 400 C or 250 C is added to the temperature rise for
determination of the final temperature T.
Model PPC-120M

Page 35 of 73
Nemko ref. no.: 200043138

Measuring point Measured Calculated Allowed Remarks


temp. temp. max. temp. Block vent test
t (K) T (C) T (C)

Power input 250Vac, 60Hz


Ambient, T = 25 C
1. L1 coil 60 85 Table Xb
2. BD1 body 67 92 Table Xb
3. C8 body 51 76 Table Xb
4. T1 coil 90 130 Cl.B Table Xa
5. T1 core 83 123 Table Xa
6. L4 coil 88 113 Table Xb
7. L3 coil 74 99 Table Xb
8. H.S body & Q1 55 80 Table Xb
9. BT1 body 39 64 Table Xb
10. V2 body 40 65 Table Xb
11. CPU body 43 68 Table Xb
12. HDD body 31 56 Table Xb
13. FDD body 23 48 Table Xb
14. CD rom body 29 54 Table Xb
15. U3 body 40 65 Table Xb
16.T2 coil for inverter 63 103 Cl.B Table Xa
17. T2 core for inverter 60 100 Table Xa
18. F3 coil for inverter 58 83 Table Xb
19. Surface of enclosure 25 65 75 Table Xb

Comments According to IEC 601-1 clause 42.1, table Xa and Xb, 400 C or 250 C is added to the temperature rise for
determination of the final temperature T.
Model PPC-153M

Page 36 of 73
Nemko ref. no.: 200043138

Measuring point Measured Calculated Allowed Remarks


temp. temp. max. temp. Block vent test
t (K) T (C) T (C)

Power input 250Vac, 60Hz


Ambient, T = 25 C
1. T1 coil 65 105 Cl.B Table Xa
2. T1 core 62 102 Cl.B Table Xa
3. L1 coil 51 76 Table Xb
4. BT1 body 51 76 Table Xb
5. U1 body 58 83 Table Xb
6. L6 coil 56 81 Table Xb
7. T1 coil for inverter 70 110 Cl.B Table Xa
8. T1 core for inverter 77 117 Cl.B Table Xa
9. F1 coil for inverter 68 93 Table Xb
10. Surface of PC 34 73 75 Table Xa

Comments According to IEC 601-1 clause 42.1, table Xa and Xb, 400 C or 250 C is added to the temperature rise for
determination of the final temperature T.
Model PPC-150M

Page 37 of 73
Nemko ref. no.: 200043138

42.1 - 42.2 Determination of the temperature with Installed fan test P


thermocouples:
Measuring point Measured Calculated Allowed Remarks
temp. temp. max. temp.
t (K) T (C) T (C)

Power input 250Vac, 60Hz


Ambient, T = 24 C
1. L1 coil 50 75 Table Xb
2. BD1 body 57 82 Table Xb
3. C8 body 45 70 Table Xb
4. T1 coil 78 118 Cl.B Table Xa
5. T1 core 74 114 Table Xa
6. L4 coil 81 106 Table Xb
7. L3 coil 66 91 Table Xb
8. H.S body & Q1 44 69 Table Xb
9. V36 body 40 65 Table Xb
10.cpu body 73 98 Table Xb
11. H.DD body 32 57 Table Xb
12.F.DD body 18 43 Table Xb
13.CD rom body 28 53 Table Xb
14.V1 body 42 67 Table Xb
15.L48 body 54 79 Table Xb
16.T1 coil for inverter 61 101 Cl.B Table Xa
17.T1 core for inverter 47 87 Table Xa
18.L1 coil for inverter 57 82 Table Xb
19. Surface of enclosure 34 74 75 Table Xa

Comments According to IEC 601-1 clause 42.1, table Xa and Xb, 400 C or 250 C is added to the temperature rise for
determination of the final temperature T.
Model PPC-120M

Page 38 of 73
Nemko ref. no.: 200043138

Measuring point Measured Calculated Allowed Remarks


temp. temp. max. temp. Installed fan test
t (K) T (C) T (C)

Power input 250Vac, 60Hz


Ambient, T = 25 C
1. L1 coil 49 74 Table Xb
2. BD1 body 56 81 Table Xb
3. C8 body 42 67 Table Xb
4. T1 coil 78 118 Cl.B Table Xa
5. T1 core 73 113 Table Xa
6. L4 coil 81 106 Table Xb
7. L3 coil 66 91 Table Xb
8. H.S body & Q1 47 72 Table Xb
9. BT1 body 39 64 Table Xb
10. U2 body 68 93 Table Xb
11. CPU body 72 97 Table Xb
12. HDD body 23 48 Table Xb
13. FDD body 13 38 Table Xb
14. CD rom body 20 45 Table Xb
15. U3 body 38 63 Table Xb
16.T2 coil for inverter 55 95 Cl.B Table Xa
17. T2 core for inverter 52 92 Table Xa
18. F3 coil for inverter 50 75 Table Xb
19. Surface of enclosure 31 71 75 Table Xb

Comments According to IEC 601-1 clause 42.1, table Xa and Xb, 400 C or 250 C is added to the temperature rise for
determination of the final temperature T.
Model PPC-153M

Page 39 of 73
Nemko ref. no.: 200043138

42.1 - 42.2 Determination of the temperature rise Not applicable. SMP used N
of copper windings by the resistance
method:
R 2 R1
t = (234 .5 + t 1 ) ( t 2 t 1 )
R1

Equipment:
Initials Tester / Date: /

Quantity Rated Used in tests Remarks


Supply voltage, V
Supply power, VA
Output power, VA
Winding R1 () R2 () Room temperat. Final temperat. Allowed Remarks
t1 (C) t2 (C) t (K) T (C) T (C)

Comments According to IEC 60601-1 clause 42.1, table Xa and Xb, 400 C or 250 C is added to the temperature rise
for determination of the final temperature T.

42.3 Surface temperatures of applied parts No applied parts used N


not intended to supply heat to a patient
do not exceed 41C
42.5 Guards for preventing contact with hot Not applicable. No hot accessible surfces N
accessible surfaces are removable only
with the aid of a tool

43 FIRE PREVENTION
43.1 Strength and rigidity of the equipment The equipment is well constructed with regard to avoid P
are sufficient to avoid fire hazards (see fire hazard.
21)
43.2 Oxygen enriched atmospheres: No general requirement.

Page 40 of 73
Nemko ref. no.: 200043138

44 OVERFLOW, SPILLAGE, LEAKAGE, HUMIDITY, INGRESS OF


LIQUIDS, CLEANING, STERILIZATION, DISINFECTION AND
COMPATIBILITY
44.2 Overflow
Equipment containing a liquid No containing a liquid reservoir used N
reservoir:
- The equipment is electrically safe N
after 15% overfill steadily over a
period of 1 min
- Transportable equipment is N
electrically safe after additionally
having been tilted through 15 in least
favourable direction(s) (if necessary
with refilling)
No signs of wetting of uninsulated live N
parts
Dielectric strength test (if deemed N
necessary)
44.3 Spillage
Safety of the equipment does not Not applicable. N
change in consequence of spillage test
performed by the method mentioned in
the standard (200 ml of water for
approximately 15 s from a height of
max. 5 cm)
44.4 Leakage
Safety of the equipment does not Not applicable. N
change in consequence of leakage test
performed by the method mentioned in
the standard (drops of water).
Equipment shall be so constructed that
liquid which might escape in a single
fault condition does not cause a safety
hazard(see also Sub-clause *52.4.1)
44.5 Humidity
Checked by the pre-conditioning 48H P
treatment and tests (see 4.10)
44.6 Ingress of liquids
Equipment checked by relevant tests of Not applicable. IPX0 N
IEC 60 529
Equipment withstands dielectric N
strength test in Clause 20.
Water, if entered equipment, has no N
harmful effect, in particular there is no
trace of water on insulation for which
creepage distances are specified

Page 41 of 73
Nemko ref. no.: 200043138

44.7 Cleaning, sterilisation and


disinfection
Equipment/equipment parts capable of See Sub-clause 6.8.2 d. P
withstanding cleaning, sterilisation or
disinfection likely to be encountered in
normal use or specified by the
manufacturer
Test with saturated steam at 134 Not applicable. N
degrees C+-4 for 20 cycles, each of 20
min duration
Method specified by the manufacturer See Sub-clause 6.8.2 d. P
in instructions for use
Safety of the equipment not impaired Not applicable. N
by the test. No appreciable signs of
deterioration. Dielectric strength test
specified in Clause 20. withstood.
44.8 Compatibility with substances used No general requirement
with the equipment:

45 PRESSURE VESSELS AND PARTS SUBJECT TO PRESSURE


45.2 Pressure vessel with a pressure volume No pressure vessels used N
greater than 200 kPa x l, and pressure
greater than 50 kPa, withstands the
hydraulic test pressure.
Hydraulic test pressure: (The test N
pressure shall be maximum permissible
working pressure multiplied by a factor
obtained from Fig. 38).
45.3 The maximum pressure to which a part N
can be subjected in normal condition
and single fault condition does not
exceed the maximum permissible
working pressure for the part
The used maximum pressure is the N
highest of the following:
45.3 a Rated maximum supply pressure from N
an external source
45.3 b Pressure setting of a pressure relief N
device provided as part of the assembly
45.3 c Maximum pressure that can be N
developed by an air compressor that is
part of the assembly, unless the
pressure is limited by a pressure-relief
device
45.7 Equipment incorporates pressure-relief N
device(s) where excessive pressure
could occur
If yes, it complies with all the N
following requirements:
45.7 a Connected as close as possible to the N
pressure vessel or parts of system it is
intended to protect

Page 42 of 73
Nemko ref. no.: 200043138

45.7 b Readily accessible for inspection, N


maintenance and repair
45.7 c Not capable of being adjusted or N
rendered inoperative without a tool
45.7 d Its discharge opening located and N
directed so, that released materials is
not directed towards any person
45.7 e Operation of device does not deposit N
material on parts causing possible
safety hazard
45.7 f Ensures that the pressure does not N
exceed the maximum permissible
working pressure by more than 10%, if
a failure occurs in the control of the
supply pressure
45.7 g No shut-off valve between the N
pressure-relief device and the parts it is
intended to protect
45.7 h Minimum number of cycles of N
operation is 100 000, except for
bursting disks

46 HUMAN ERRORS
No general requirement.

47 ELECTROSTATIC DISCHARGES
No general requirement.

48 BIOCOMPATIBILITY
Equipment parts and accessories No parts of the equipment or accessories are intended N
intended to come into contact with to come into contact with biological tissues, cells or body
biological tissues, cells or body fluids fluids.
assessed and documented as given in
ISO 10993-1.
Inspection of information provided by
manufacturer.

49 INTERRUPTION OF POWER SUPPLY


49.1 If automatically resetting thermal cut- Not applicable. N
outs and overcurrent releases are used,
they give a safe condition by such
resetting
49.2 Interruption and restoration of the P
power supply do not result in hazards
49.3 Means are provided for removal of Not applicable. N
mechanical constrains on patients in
case of a supply mains failure

Page 43 of 73
Nemko ref. no.: 200043138

50 MARKING OF CONTROLS AND INSTRUMENTS


50.1 See Sub-clause 6.3.

51 PROTECTION AGAINST HAZARDOUS OUTPUT


51.1 Intentional exceeding of safety limits. No general requirement (cf. Appendix A, Sub-clause A2)
51.2 Indication of parameters relevant to No general requirement (cf. Appendix A, Sub-clause A2)
safety.
51.3 Reliability of components. See Sub-clause 3.6 f.
51.4 Equipment providing both low- and Not applicable. N
high-intensity outputs. Minimised
possibility of high intensity output
being selected accidentally
51.5 Incorrect output. No general requirement

52 ABNORMAL OPERATION AND FAULT CONDITIONS


52.1 Equipment shall be so designed and P
manufactured that even in single fault
condition no safety hazard as described
under Sub-clause 52.4 exists (see 3.1
and 13)
52.1 Safety of equipment incorporating The equipment does not include programmable
programmable electronic systems electronic subsystems as specified in IEC 60601-1-4,
(PES) conforms to the rules of future 1996
IEC 601-1-4
52.5.1 Overloading of mains supply Tests see Sub-clause 57.9. P
transformers See Nemko ref.: 200039133
52.5.2 Failure of thermostats Not used N
52.5.3 Short-circuiting one part of a double Not applicable. N
insulation
52.5.4 Interruption of the protective earth P
conductor (tests as described in Sub-
clause 19.4)
52.5.5 Impairment of cooling. Temperatures Fan used P
do not exceed 1.7 times values of
Clause 42, Tables Xa and Xb, minus
17.5 degrees C.
52.5.6 Locking of moving parts (see also No moving parts N
52.5.8)

Page 44 of 73
Nemko ref. no.: 200043138

52.5.7 Interruption and short-circuiting of Not applicable. N


motor capacitors (see also 52.5.8) Not used
Equipment:

Initials Tester / Date: /


Capacitor and Short circuit Open circuit R1 () R2 () t1 (C) t2 (C) t (K) Final t (C)
winding

Comments

52.5.8 Additional tests for motor operated Not used N


equipment:
52.5.9 Failure of components (refer to Table See the table below P
in Clause 56)
X1 and X2 capacitors between parts of
opposite polarity in the mains part and
complying with IEC 60 384-14
exempted from this requirement
52.5.10 Overload
52.5.10 a Equipment with heating elements: Not used N
52.5.10 a 1 Thermostatically controlled equipment N
(see 52.5.10.c and d)
52.5.10 a 2 Equipment with heating elements with N
short-time rating (see 52.5.10 c and e)
52.5.10 a 3 Other equipment with heating elements N
(see 52.5.10 c)
52.5.10 b Equipment having motors, which are a N
part of equipment (see 52.5.5 - 52.5.8
and 52.5.10 f - 52.5.10 h)
52.5.10 b 1 Mention circuit/component concerned N
(e.g. regulator, capacitor, wiring, etc.)
and mention the fault condition (e.g.
disconnected, short-circuited, etc.).
Mention creepage-distances separately
if applicable (refer to the text of 52.5
above 52.5.1 in the 601-1 standard)
52.5.10 b 2 Mention for example safety hazards N
according to 3.1 and 13 and 52.4

Page 45 of 73
Nemko ref. no.: 200043138

Test No Applied single fault condition (refer to 52.5) 1) Observed result (e.g. hazards which arise or not) 2)
from 52.5
See Nemko ref. No. 200039133

53 ENVIRONMENTAL TESTS
See Sub-clause 4.10 and Clause 10.

54 CONSTRUCTIONAL REQUIREMENTS - GENERAL


See Clause 55 to 59.

55 ENCLOSURES AND COVERS


See Clause 16, 21 and 24.

Page 46 of 73
Nemko ref. no.: 200043138

56 COMPONENTS AND GENERAL ASSEMBLY - LIST OF CRITICAL


COMPONENT PARTS
Code Approved by Manufacturer, Type number, Nominal ratings Operating- Location
ratings

Mains cord <HAR> R&L, WS-010, H03W-F 3G 0.75mm2 250Vac, 16A


Enclosure UL HB or better,
min. 2.0 mm
thick
PCB UL 94-1 or better
CD-rom TUV, CSA, XM-7004Bxx 5Vdc, 0.9A
(Optional) UL XM-1902Bxx
X=0-9
Appliance VDE Supercom, SC-9 10A, 250Vac
inlet Inalways, 0711
Rong Feng, SS-130
Rong Feng, SS-7B
Mains fuse UL, CSA Conquer, GBP 12A 250Vac
System General, PDC60240.A0 I/P 24Vdc/5.0A
Conquer, GFP O/P
7A, 250Vac
Mains switch VDE, CSA Fima, 1852 5A, 250Vac
Lithium UL SGS-Thomson, M4T28-BR12SH1 5V, 50mAh
Battery
Ply switch UL, TUV Raychem, SMD150-2018 1.10A, 5V
for Keyboard
mouse
DC fan for VDE, CSA, Delta, AFB0512MA 12Vdc, 0.15A
CPU UL 8.47CFM
Inverter Accept Sen Huei, TRN-0067 and TRN-0064 Cl.B
transformers tested in TDK, SRW15/20EM-T11H004
equipment NIA15/20EM-T11H004
Mains See Nemko Adaptor: Skynet SNP-8086-M
transformer ref. No.
200039133
DC fan for UL, TUV Adda, AD0412MB-G70 12Vdc, 0.08A,
DC power 4.2CGM
supplies
LCD module Accept Samsung, LT150Z-051 TFT type, 6V
for model tested in max. for iput,
PPC-xxM-zT equipment 11.2mA, 690V
for lamp.
Alt. Toshiba , LTM15C151A TFT type, 5V
max for iput,
6mA, 710V for
lamp.
Inverter for Accept Laruel Carry&Leap, TAD282 12V, 1.14A
Lc tested in
equipment

Page 47 of 73
Nemko ref. no.: 200043138

56.1 General Marking of components (refer to Table in Clause 56).


56.1 b Component markings are according The components are used according to their ratings. All P
with conditions of use components in the mains part and applied part are
marked or otherwise identified regarding their ratings.
56.1 d Fixing of components. The movement of components is prevented. P
56.1 f Fixing of wiring. Conductors and connectors are adequately secured and P
insulated. Accidental detachment does not result in a
safety hazard.
56.3 Connections - general
56.3 a Incorrect interconnection of accessible No such interconnection N
connectors is prevented where a hazard
may be caused
Accessible parts are separated from P
live parts (see 17 g)
Plugs for connection of patient circuit Not applicable. No such connection of patient N
cannot be connected to other outlets
Medical gas connections are not Not used N
interchangeable (see Sub-clause 6.6
and ISO R407)
56.3 b Accessible conductive parts are P
prevented from becoming live when
connection between different part of
the equipment is broken
56.3 c Connectors having conductive Not applicable. N
connection to patient are constructed so
that no conductive connection of the
connector remote from patient can
contact earth or hazardous voltages
Test with flat conductive surface N
Test with straight unjointed test finger N
Test if able to be plugged into a mains N
socket
56.4 Connections of capacitors
Capacitors not connected between live Not applicable. N
parts and non- protectively earthed
accessible conductive parts
Capacitors between the mains part and P
earthed accessible parts, comply with
IEC 60384-14 or equivalent
The enclosure of capacitors are not P
secured directly to non-protectively
earthed accessible metal parts
Capacitors or other spark-suppression P
devices are not connected between
contacts of thermal cut-outs

Page 48 of 73
Nemko ref. no.: 200043138

56.5 Protective devices


Protective devices which operate by P
producing a short circuit which results
in operation of an overcurrent
protection device in supply mains
system not used (see also 59.3)
56.6 Temperature and overload control
devices
56.6 a Thermal cut-outs which require Not used N
soldering to reset are not used
Thermal safety devices are provided to N
prevent temperature limits being
exceeded
An independent non-self resetting N
thermal cut-out is provided where the
failure of a thermostat may cause a
hazard
The operating temperature of the above N
thermal cut-out is between the upper
limit of the first thermostat and the safe
limit for the function
There is an audible alarm where loss of N
function could present a safety hazard
Test of thermal cut-outs and N
overcurrent releases
Heated liquid containers are protected N
against dangerous overheating when
container is empty
56.6 b Means provided for varying the N
temperature setting of thermostats, the
temperature setting is clearly indicated
The operating temperature of thermal N
cut-outs is clearly indicated
56.7 Batteries
56.7 a 1 Adequately ventilation P
56.7 a 2 Accidental short-circuiting is prevented P
(Note! Lithium batteries)
56.7 b Incorrect polarity of connection Not applicable. A back up Lithium battery used. Not N
prevented (Note! Lithium batteries) internally powered equipment
1) Establish whether there is a
possibility of making an incorrect
battery connection.
2) Where such a possibility exists,
establishing the effect of an incorrect
battery connection.
Type of battery is clearly marked (see Battery type: M4T28-BR12SH1 P
also 6.2 d)
56.7 c Battery state. No general requirement. N

Page 49 of 73
Nemko ref. no.: 200043138

56.8 Indicators
Unless indication is otherwise apparent Colours, see Sub-clause 6.7. P
to the operator from normal operating
position , indicator lights are provided:
- to indicate that equipment is See Sub-clause 6.3 a. P
energised
- to indicate the operation of non- Not applicable. N
luminous heaters
- to indicate when output exists if a Not applicable. . N
safety hazard could result
Charging mode is visibly indicated to Not applicable. N
the operator
56.10 Actuating parts of controls
56.10 a Accessible parts of electrical controls No actuating parts N
comply with requirements in Sub-
clause 16 c
56.10 b Fixing, prevention of maladjustments: N
Actuating parts are adequately secured N
to prevent them working loose during
normal use
Controls are secured to prevent N
movement relative to scale marking
Detachable indicating devices are N
prevented from incorrect connection
without the use of a tool
Equipment:
Initia ls Tester / Date: /
Rotating controls Gripping diameter of the Test torque Remarks
knob (mm) (Nm)

Pull controls Axial pull, test force (Nm) Remarks

56.10 c Stops of adequate mechanical strength Not applicable. N


are provided on rotating or movable
parts of controls, where necessary to
prevent an unexpected change from
maximum to minimum, or vice-versa,
of the controlled parameter where this
could produce a safety hazard
Manual torque test N
Axial pull test N

Page 50 of 73
Nemko ref. no.: 200043138

56.11 Cord connected hand-held and foot-


operated control devices
56.11 a Voltages on cord connected control P
devices do not exceed 25 V a.c. or 60
V d.c. or peak value
Circuits isolated from mains part (see P
17 g)
56.11 b Hand-held control devices: No safety Not hand-held equipment N
hazard as a result of the free fall test
(Sub-clause 21.5)
Foot-operated control devices able to N
support the weight of an adult human
being. Test with an actuating force of
1350 N over an area of 625 mm2 . No
damage to the device resulting in a
safety hazard.
56.11 c Hand-held and foot-operated control N
devices do not change their control
setting when inadvertently placed in
abnormal position
56.11 d IP-classification of foot-operated N
control device at least IPX1
Electrical switching parts specified for N
use in operating rooms at least IPX8
56.11 e Provided with adequate means of N
anchoring as for power supply cords
(see test of 57.4).

57 MAINS PARTS, COMPONENTS AND LAY-OUT


57.1 Isolation from supply mains
57.1 a Means of isolation: P
The equipment has means for simultaneous P
disconnection of all supply poles
Means for disconnection incorporated in P
equipment
External means for disconnection are Not applicable. N
specified in accompanying documents
57.1 d Switches for mains disconnection comply P
with creepage distance and air clearances
(IEC 60328 , replaced by IEC 60 61058-1)
57.1 f Mains switches are not incorporated in power P
supply cord or flexible leads
57.1 g The directions of movement of actuators of Two -pole switch disconnects the voltage. Mounted N
mains switches comply with IEC 60447 according to IEC 60447.
57.1 h Suitable plug device used to isolate the non- P
permanently installed equipment from the
supply mains
57.1 m Fuses and semiconductor devices are not used P
as isolating devices

Page 51 of 73
Nemko ref. no.: 200043138

57.2 Mains connectors , appliance inlets and the


like
57.2 e Auxiliary mains sockets outlets: Not used N
- cannot accept a mains plug (does not apply N
to emergency trolleys)
Number of socket-outlets on emergency N
trolleys are limited to maximum 4.
- and they are properly marked N
57.3 Power supply cords
57.3 a Equipment not provided with more than one P
connection to supply mains
Alternative connections to a different supply P
system do not cause safety hazards when
more than one connection is made
simultaneously
Mains plug not fitted with more than one P
power supply cord
Equipment not intended to be permanently Appliance inlet. P
connected to fixed wiring is provided with
either a power supply cord or an appliance
inlet
57.3 b Power supply cords not less robust than The cord is of type HO3VV-F P
ordinary flexible cord, designation 53 of IEC
60245 or IEC 60227
PVC insulated power supply cords not used Not applicable. No external metal parts excedding N
for equipment having external metal parts 75C
exceeding 75C , unless cord rated for the
temperature measured
57.3 c Cross-sectional area of power supply cords 0,75mm Cu. See list of components. P
(see Table XV) (Data sheet)
Equipment:
Initials Tester / Date:
57.3 d Stranded conductors fixed by any clamping Not applicable. N
means or screws are not soldered

Page 52 of 73
Nemko ref. no.: 200043138

57.4 Connection of Power Supply Cords


57.4 a Cords anchorages: Detachable power supply cord used N
The equipment or its mains connectors N
incorporate a cable anchoring device to
relieve strain on conductors and prevent
abrasion of covering (no knots etc.)
The construction and materials of the cord N
anchorages are made of insulating material or
insulated from unearthed accessible metal
parts by supplementary insulation,
or having an insulating lining complying with N
the requirements for basic insulation fixed to
the cord anchorage if an insulation fault on
power supply cord could make accessible
parts live
Clamping screws do not bear directly on the N
cord insulation
Screws associated with cable replacement are N
not used to fix other components
Conductors so arranged that if the cord N
anchorage fails no strain on protective earth
conductor occurs as long as phase conductors
are in contact with their terminals
Pull test 25 times N
Torque test for 1 min N
After the tests, the cord sheath has not been N
longitudinally displaced by more than 2 mm
and the conductor ends have not moved over
a distance of more than 1 mm from their
normally connected position
Creepage distances and air clearances are not N
reduced below the values specified in Sub-
clause 57.10.
Not possible to push the cord into equipment N
to such an extent that the cord, or internal
parts of the equipment, could be damaged
57.4 b For other than stationary equipment, the N
power supply cord is adequately protected
against excessive bending by means of a cord
guard of insulating material
Alternatively shaped opening for the power N
supply cord
Bending test: N
Flexing test (if guards fail the dimensional N
test) (see IEC 60335-1 AM 6:1988, Sub-
clause 25.10)
Results of flexing test: N
Percent of broken conductor strands: N

Page 53 of 73
Nemko ref. no.: 200043138

57.4 c There is sufficient space inside the equipment N


to allow the supply cable conductors to be
introduced and connected
Any covers can be fitted without risk of N
damage to the conductors or their insulation
It is possible to check that conductors are N
correctly connected before the cover is fitted
57.5 Mains Terminal Devices and wiring of
mains part
57.5 a Mains connected equipment other than those Not applicable. Detachable supply cord used N
with a detachable supply cord is provided
with mains terminals, where connections are
made with screws, nuts , soldering , clamping
, crimping of conductors or equally effective
methods
Reliance not placed upon terminals alone to N
maintain conductors in position, unless
barriers are provided
Terminals of components used for external N
conductors comply with specified
requirements and are properly marked (see
6.2 h, j and k)
Screws and nuts which clamp external N
conductors are not used to fix any other
component (other than internal conductors
unlikely to be displaced)
Barriers provided N
57.5 b Terminals including any protective earth N
terminal are closely grouped to allow easy
connection
Protective earth conductor, see 58 N
Marking of mains terminals, see 6.2 N
Mains terminal devices are inaccessible N
without the use of a tool
Mains terminal devices are so located or N
shielded that there is no risk of accidental
contact between live parts and accessible
conductive parts
- and for class II equipment between live N
parts and conductive parts separated from
accessible parts by supplementary insulation
only
Single strand of wire (8 mm). Test result: N
57.5 c Tightening or loosening of clamping means N
of conductors does not subject internal wiring
to stress
- does not reduce creepage distances and air N
clearances below allowable values
Cross-sectional area of the conductor used in N
the fastening and loosening test:

Page 54 of 73
Nemko ref. no.: 200043138

57.5 d Effective connection to mains terminals does N


not require special preparation of cables
Conductors are not damaged or dis placed N
when tightening or loosening of clamping
screws or nuts
57.5 e Fixing of wiring See Sub-clause 56.1 f. P
57.6 Mains Fuses and Over-current Releases
List of fuses and over-current releases: See P
table at the beginning of Clause 56 in this
TRF
There are fuses or over-current releases in P
each supply lead for class I equipment and
class II equipment having a functional earth
according to Sub-clause 18.1
There is at least one mains fuse for other Class I equipment N
single phase class II equipment
A protective earth conductor is not fused P
No fuse is fitted in neutral conductor of P
permanently installed equipment
57.8 Wiring of Mains Part
57.8 a The insulation of an individual conductor is The internal mains wiring has at least the same P
at least equivalent to that required by IEC quality and equivalency to that required in IEC
60227 or IEC 60245 (otherwise the conductor 60227/245.
is considered bare)
- dielectric strength test of 2000 V for 1 min P
57.8 b The cross-sectional area of internal wiring Not applicable. Not used N
between the mains terminal and protective
devices is not less than minimum required for
the power supply cord. Measured area:
(see Sub-clause 57.3 c)
Equipment:
Initials Tester / Date: /
The cross-sectional area of other wiring in the N
mains part and sizes of tracks on printed
wiring circuits are sufficient to prevent any
fire hazard in case of possible fault currents
In case of doubt concerning adequacy of N
incorporated overcurrent protection
connecting test performed. No safety hazard
arises
57.9 Mains supply transformers (refer to table See list of components.
below)
57.9.1 Insulation of mains supply transformers P
protected against overheating in event of
short-circuit or overload (see 57.9.1 a and b)
External protective devices connected such P
that failure of components cannot render the
protective devices inoperative

Page 55 of 73
Nemko ref. no.: 200043138

57.9.1 a Short-circuit test(s) at 90 to 110 percent of See Nemko ref. No. 200039133. P
rated supply voltage / voltage range SMP mdoel SNP -8086-M
Supply voltage:
Equipment:
Initials Tester / Date:
Winding under test*) Comments
T1:
pin 8-13,14 Shutdown after 1 sec., cycle protection
pin FL2-10,11,12 Shutdown after 1 sec., cycle protection
pin 10,11,12-FL1 Shutdown after 1 sec., cycle protection

*) Test under the conditions specified in


Clause 42.
57.9.1 b Overload
Tests under the conditions specified in Clause Temperatures during tests do not exceed values P
42 until steady thermal conditions are given in Table XIX.
obtained. Results: See table below.
Loading of a section or winding of the See Nemko ref. No. 200039133. P
transformer: SMP mdel SNP-8086-M
Supply voltage: 250Vac, 60Hz
Equipment: N3133, N3126, N3636, N3747,N382
Initials Tester / Date: JohnyG / 09-2000
Winding under Test Duration t Final Ins. Notes
test current (C) t (C) Class
(min.)
max. B
(A) (C)
T1: after D4 25 124 78 118 175 Unit shutdown, No danger, No P
(for +5V) damaged
T1: after D3 5,4 130 63 103 175 Unit shutdown, No danger, No P
(for +12V) damaged

OUTPUT OVERLOAD

For +5V 23 270 66 106 175 Unit shutdown, No danger, No P


damaged
For +12V 5,3 230 60 100 175 Unit shutdown, No danger, No P
damaged

Page 56 of 73
Nemko ref. no.: 200043138

57.9.2 Dielectric strength of electrical insulation The electrical insulation between primary winding
between turns and layers and other windings, screens and the core of the
transformer(s) have been investigated by the tests
performed as described in Clause 20.
Insulation between turns and layers, test Not applicable. SMP used N
performed according to 57.9.2:
Voltage used in test:
Frequency used in test:
57.9.4 Construction
Identification, Manufacturer and type Total Insula- Nominal rated General Protection
reference, rated tion voltages and construction*) devices
marking (e.g. T1, power material currents of the (manufacturer,
T2, etc.) (VA) class windings type and rating,
location)
Mains SKYNET (H.K.) Class B 100-250Vac,
transforme. T1 Electronic Co. Ltd. 3A
360-8086-M
Inverter Sen Huei, TRN-0067 Class B
transformers:
T1, T2
TDK Class B
SRW15/20E M-
T11H004
NIA15/20EM-T11H004 Class B

*) A construction diagram has to be added to this test report for each transformer mentioned above.

Drawing
See Nemko ref. No 200039133
Comments The transformer consists of windings wounded on a bobbin with separate integrated forms. The
barrier between the primary and secondary windings has a split which increase the distance
between them.
Creepage distances and air clearances of See Nemko ref. No 200039133
transformers
Insulation between Ref. Air clearance (mm) Creepage distance (mm) Notes
voltage (V) Required Measured Required Measured
Primary - Primary 250 1,6 2,1 3,0 3,4 Parts other
than windings
Primary - Secondary 250 5,0 11,5 8,0 11,5 Windings

57.9.4 a One of the following methods is used to One bobbin with concentric windings separated by P
separate primary from secondary and patient double or reinforced insulation
connected windings:
57.9.4 c Means are provided to prevent displacement P
of end turns

Page 57 of 73
Nemko ref. no.: 200043138

57.9.4 d An earthed screen has a minimum overlap of Not used N


3 mm and a width at least equal to the axial
length of the primary winding
57.9.4 e The insulation between the primary and P
secondary winding in reinforced or double
insulated transformers consists of:
number of insulation layer(s): 5 layers P
total thickness 0,5mm P
combination of two layers withstands Test voltage 4000Vac P
dielectric strength test for reinforced
insulation
57.9.4 f The creepage distances between the primary See Sub-clause 57.9.4 e. P
and secondary windings comply with
requirements for reinforced insulation, with
specified allowances
mm at location: P
Notes: P
57.9.4 g For toroidal transformers: Not applicable. SMP used
the exit of wires from internal windings is N
provided with double sleeving complying
with requirements for double insulation
its total thickness is at least 0.3 mm extending N
at least 20 mm outside the winding. Measured
thickness:

Page 58 of 73
Nemko ref. no.: 200043138

57.10 Creepage distances and air clearance


57.10 a Values (refer to table )
Creepage distances and air clearances comply P
at least with the values of Table XVI
Creepage distances for slot insulation of No motor used N
motors are at least 50% of the values in the
table with a minimum of 2 mm at 250 V
Creepage distances and air clearances No defibrillation proof applied parts N
between defibrillation proof applied parts and
other parts are not less than 4 mm
Insulation Ref. Air clearance (mm) Creepage distance (mm) Notes
voltage
between (V) Required Measured Required Measured
Equivalent to basic A -f 250 1,6 4,6 3,0 4,6
insulation between
parts of opposite
polarity
Basic or A -a1

supplementary A -b

insulation A -c

A -j

B-d

B-c

Double insulation or A -a2 250 5,0 12,3 8,0 25,0

reinforced insulation A -e 250 5,0 25,0 8,0 35,0

A -k 250 5,0 7,0 8,0 12,4

B-a 250 5,0 7,0 8,0 8,6


B-e

Comments
57.10 b Statements , test specifications etc. observed
57.10 d Statements regarding measurement of
creepage distances and clearances observed

Page 59 of 73
Nemko ref. no.: 200043138

58 PROTECTIVE EARTHING -TERMINALS AND CONNECTIONS


58.1 Clamping means of the protective earth P
terminal for fixed supply conductors or
power supply cords comply with
requirements of Sub-clause 57.5 c:
Not possible to loosen the clamping P
means without use of a tool
Screws for internal protective earth P
connections are covered or protected
against loosening from outside
58.2 Internal protective earthing connections P
by means of screw, soldering,
crimping, wrapping, welding or a
reliable pressure contact
58.3 Not used (see 57.5 b)
58.7 Where an appliance inlet is used for the P
supply connection, its earth pin is
regarded as the protective earth
terminal
58.8 The protective earth terminal is not P
used for connection between different
parts or fixing of any component not
related to protective or functional
earthing
58.9 Where the protective earth connections P
are made via a plug or socket device
the protective earth connection is made
before and interrupted after the supply
connections during connection and
interrupting

59 CONSTRUCTION AND LAYOUT


59.1 Internal wiring
Fixing of wiring in the applied part and Not applicable. No applied part N
the mains part, see Sub-clause 56.1 f
59.1 a Internal cables and wiring:
- are protected against contact with No moving parts. The wiring is adequately protected N
moving parts and friction with sharp against sharp edges etc.
corners and edges
- are protected by additional fixed N
sleeving or similar means , if there is
movement relative to metal parts where
it is in direct contact with metal parts
- and wiring , cord forms or P
components are not likely to be
damaged by opening or assembling the
equipment or opening or closing of
inspection doors

Page 60 of 73
Nemko ref. no.: 200043138

59.1 b The bending radius of cables and cable Not applicable. N


forms is at least 5 times the outer
diameter of the lead
59.1 c Insulation:
- insulating sleeving is adequately P
secured. It can only be removed by
breaking or cutting or is secured at both
ends
Sheath of a fle xible cord used as Not applicable. N
supplementary insulation inside
equipment is not subject to undue
mechanical or thermal stresses and its
insulation properties are at least as
specified in IEC 60227 or IEC 60245
- insulated conductors, which are Not applicable. No temp over 70 deg.C N
subject to temperatures greater than
70C, have an insulation of heat-
resistant material
Dielectric strength test of insulation of N
2000 V for 1 min
59.1 d Aluminium wires below 16 mm cross- Not used N
section are not used
59.1 e Separation of circuits (see Clause 17)
59.1 f Connecting cords between equipment Not applicable. No such connection N
parts (e.g. parts of an x-ray or patient
monitoring installation or a data-
processing installation or
combinations) are considered
belonging to equipment and subject to
requirements of this standard
59.2 Insulation
59.2 b Insulation characteristics, mechanical Test of enclosure parts. Temperature of ball 90 C. P
strength and resistance to heat and fire d=1.3mm
is retained. Result of the ball-pressure
test:
59.2 c Insulation is not likely to be impaired P
by deposition of dirt or dust resulting
from wear of parts within the
equipment such that creepage distances
and clearances are reduced below
specified values
Ceramic materials and the like Not used N
specified in this sub-clause are not used
as supplementary or reinforced
insulation
Rubber materials used as Class I equipmet N
supplementary insulation in class II
equipment are resistant to ageing
(oxygen test) and arranged and
dimensioned adequately
Creepage distances are not reduced P
below values specified in 57.10 despite
any cracks in such insulation

Page 61 of 73
Nemko ref. no.: 200043138

Insulating materials in which heating No heating elements used N


elements are embedded are not used as
reinforced insulation
59.3 Excessive current and voltage
protection
See Sub-clause 57.6. An internal No internal electrical power source N
electrical power source in equipment is
provided with an appropriately rated
device to protect against fire hazards.
Protective means used. Inspection of
design data
Fuse elements replaceable without Not applicable. Must open the enclosure N
opening the enclosure are fully
enclosed in a fuseholder and fuse
replacement does not cause safety
hazard. Test with standard test finger.
Live parts of fuseholders are shielded See above N
to prevent electric shock when
replacing fuses (if replaceable without
the use of a tool)
Protective devices between an F-type Not applicable. No F-type applied part N
applied part and the enclosure do not
operate below 500 V r.m.s.
59.4 Oil containers
Oil containers in portable equipment Not applicable. No oil containers used N
are adequately sealed to prevent oil loss
in any position and the oil container
design allows for the expansion of oil
Oil containers in mobile equipment are N
sealed to prevent loss of oil during
transport
An oil level indicator is provided on N
partially sealed oil-filled equipment or
parts. Inspection of equipment and
technical description , and by manual
test

Page 62 of 73
Nemko ref. no.: 200043138

19.4 f APPENDIX 1
Fig. 16 in Measurement of the earth leakage P
IEC 60601-1 current

L(N)
T1 S5 P1

Rseau P3
Mains V 5
N(L) P1
S1

S10 P1
FE

MD PE

S12 P1

(Measuring supply circuit with one side of the supply mains at approximately earth potential (fig. 10)).
Switch positions NC (S1 = 1) SFC (S1 = 0) Additional SFC*
(see Clause 17.a)* (S1 = 1)
Measured (A) Measured (A) Measured (A)
1 2 1 2 1 2
S5 S10 S12 Before After Before After Before After
1 1 1 180 180
1 1 0 380 381
1 0 1 182 182
1 0 0 378 379
0 1 1 184 185
0 1 0 381 382
0 0 1 181 181
0 0 0 381 382
1)Insulation-enclosed
2)Metal enclosed
3)Combination of Insulation- and metal-enclosed 2
and = Before and after humidity preconditioning treatment.
Note: NC = Normal Conditions 1 = Switch Closed * = describe additional SFC according to
SFC = Single Fault Condition 0 = Switch Open Sub-clause 17 a in notes above

Page 63 of 73
Nemko ref. no.: 200043138

19.4 g APPENDIX 2
Fig. 18 in Measurement of the enclosure FOR MAINS OPERATED EQUIPMENT P
IEC 60601-1 leakage current

L(N)
T2 S9

Rseau
Mains
V V2
N(L)
P1

L(N) 1
T1 S5 P1

Rseau P3
Mains V V1 5
N(L) MD2
P1
S1

S10 P1
FE

S7 P1
PE

MD1
P1
S12

Switch position NC SFC SFC Additional SFC*


S1 = 1, S7 = 1 S1 = 0, S7 = 1 S1 = 1, S7 = 0 S1 = 1, S7 = 1
Measured (A) Measured (A) Measured (A) Measured (A)
1 2 1 2 1 2 1 2
S5 S10 S12 Before After Before After Before After Before After
1 1 1 0,4 0,4 0,5 0,5 0,4 0,4
1 1 0 0,4 0,4 0,5 0,5 0,4 0,4
1 0 1 0,4 0,4 1,7 1,7 2,5 2,6
1 0 0 0,4 0,4 1,7 1,7 2,5 2,6
0 1 1 0,4 0,4 0,5 0,5 0,4 0,4
0 1 0 0,4 0,4 0,5 0,5 0,4 0,4
0 0 1 0,3 0,3 1,5 1,5 2,4 2,4
0 0 0 0,3 0,3 1,5 1,5 2,4 2,4
1 2
and = Before and after humidity preconditioning treatment.
Note: S7 not used for Class II NC = Normal Conditions 1 = Switch Closed * = describe additional SFC
equipment SFC = Single Fault Condition 0 = Switch Open according to Sub-clause 17 a in
notes above

Page 64 of 73
Nemko ref. no.: 200043138

19.4 g APPENDIX 2
Fig. 18 in Measurement of the enclosure FOR INTERNALLY POWERED EQUIPMENT N
IEC 60601-1 leakage current

See fig. 18 on previous page.


Using MD2 between different parts of the enclosure.

Normal Condition
Measured (A)
1 2
Before After

Comments :
1 2
and = Before and after humidity preconditioning treatment.

Page 65 of 73
Nemko ref. no.: 200043138

19.4 h APPENDIX 3
Fig. 20 in Measurement of the patient leakage FOR MAINS OPERATED EQUIPMENT N
IEC 60601-1 current

1
L(N)
T1 P3
S5 P1

Rseau
Mains V 5
N(L) P1
S1

S10 P1
FE

S7 P1
PE

6
P1
MD

S13

Switch position NC SFC SFC Additional SFC*


S1 = 1, S7 = 1 S1 = 0, S7 = 1 S1 = 1, S7 = 0 S1 = 1, S7 = 1
Measured (A) Measured (A) Measured (A) Measured (A)
1 2 1 2 1 2 1 2
S5 S10 Before After Before After Before After Before After
1 1
1 0
0 1
0 0
1 2
and = Before and after humidity preconditioning treatment.
Note: S7 not used for Class II NC = Normal Conditions 1 = Switch Closed * = describe additional SFC
equipment SFC = Single Fault Condition 0 = Switch Open according to Sub-clause 17 a in
notes above

Page 66 of 73
Nemko ref. no.: 200043138

19.4 h APPENDIX 3
Fig. 23 in Measurement of the patient leakage FOR INTERNALLY POWERED EQUIPMENT N
IEC 60601-1 current

Normal Condition
Measured (A)
1 2
Before After

Comments :
1 2
and = Before and after humidity preconditioning treatment.

Page 67 of 73
Nemko ref. no.: 200043138

19.4 h2 APPENDIX 4
Fig. 21 in Measurement of the patient leakage FOR MAINS OPERATED EQUIPMENT N
IEC 60601-1 current with mains voltage on the f-
type isolated (floating) applied part

1 6
L(N)
T1 S5 P1

Rseau P3
Mains V V1 5
N(L) P1
S1

3
S10 P1
FE

S7 P1 3
PE

P1

L(N) MD
T2 S9 R

Rseau
Mains V V2
N(L)

S13

Switch position Measured A


1 2
S5 S10 S13 Before After

1 1 1
1 1 0
1 0 1
1 0 0
0 1 1
0 1 0
0 0 1
0 0 0
1 2
and = Before and after humidity preconditioning treatment.
Note: S7 not used for Class II equipment NC = Normal Conditions 1 = Switch Closed
SFC = Single Fault Condition 0 = Switch Open

Page 68 of 73
Nemko ref. no.: 200043138

19.4 h2 APPENDIX 4
Fig. 24 in Measurement of the patient leakage FOR INTERNALLY POWERED EQUIPMENT N
IEC 60601-1 current with mains voltage on the f-
type isolated (floating) applied part

Single Fault Condition


Measured (A)
1 2
Before After

Comments :
1 2
and = Before and after humidity preconditioning treatment.

Page 69 of 73
Nemko ref. no.: 200043138

19.4 h3 APPENDIX 5
Fig. 22 in Measurement of the patient leakage FOR MAINS OPERATED EQUIPMENT P
IEC 60601-1 current with mains voltage on a
signal input or signal output part

Switch position Measured A


1 2
S5 S10 S13 Before After

1 1 1 113 113
1 1 0 113 113
1 0 1
1 0 0
0 1 1 113 113
0 1 0 113 113
0 0 1
0 0 0
1 2
and = Before and after humidity preconditioning treatment.
Note: S7 not used for Class II equipment NC = Normal Conditions 1 = Switch Closed
SFC = Single Fault Condition 0 = Switch Open

Page 70 of 73
Nemko ref. no.: 200043138

19.4 h3 APPENDIX 5
Fig. 25 in Measurement of the patient leakage FOR INTERNALLY POWERED EQUIPMENT N
IEC 60601-1 current with mains voltage on a
signal input or signal output part

Single Fault Condition


Measured (A)
1 2
Before After

Comments :
1 2
and = Before and after humidity preconditioning treatment.

Page 71 of 73
Nemko ref. no.: 200043138

19.4 j APPENDIX 6
Fig. 26 in Measurement of the patient FOR MAINS OPERATED EQUIPMENT N
IEC 60601-1 auxiliary current

1
L(N)
T1 P3
S5 P1

Rseau
Mains V 5 MD
N(L) P1
S1

S10 P1
FE

S7 P1
PE

Switch position NC SFC SFC Additional SFC*


S1 = 1, S7 = 1 S1 = 0, S7 = 1 S1 = 1, S7 = 0 S1 = 1, S7 = 1
Measured (A) Measured (A) Measured (A) Measured (A)
1 2 1 2 1 2 1 2
S5 S10 Before After Before After Before After Before After
1 1
1 0
0 1
0 0
1 2
and = Before and after humidity preconditioning treatment.
Note: S7 not used for Class II NC = Normal Conditions 1 = Switch Closed * = describe additional SFC
equipment SFC = Single Fault Condition 0 = Switch Open according to Sub-clause 17 a in
notes above

Page 72 of 73
Nemko ref. no.: 200043138

19.4 j APPENDIX 6
Fig. 27 in Measurement of the patient FOR INTERNALLY POWERED EQUIPMENT N
IEC 60601-1 auxiliary current

Normal Condition
Measured (A)
1 2
Before After

Comments :
1 2
and = Before and after humidity preconditioning treatment.

Page 73 of 73