Professional Documents
Culture Documents
88
1 drop = 2 or 10 ppm as CaCO3 / 25 mL blue caps
420-C0385P.88
KIT COMPONENTS:
001-H09048.88 Hardness Titrant Low, 60 mL
460-S0889.71 Hardness Titrant High, 60 mL
460-S0850.70 Hardness Buffer Solution, 30 mL
6024987 Hardness Indicator Powder, 10g
6040993 Vial, 10-50 mL
INTERFERENCES: Metals may cause difficulty in ATTENTION: As necessary, calibrate this kit against
seeing the endpoint. If metal interference is presumed, a known standard made with plant / make-up water. Be
add one drop of Hardness Titrant to the sample before sure to collect a representative sample.
adding buffer or indicator. Include this drop of titrant when Video Procedure
calculating your results. Additional Hardness Buffer may
be necessary to view a clean endpoint.
KIT COMPONENTS:
001-H09048.88 Hardness Titrant Low, 60 mL
460-S0889.71 Hardness Titrant High, 60 mL
460-S0850.70 Hardness Buffer Solution, 30 mL
6024987 Hardness Indicator Powder, 10g
6040993 Vial, 10-50 mL
INTERFERENCES: Metals may cause difficulty in ATTENTION: As necessary, calibrate this kit against
seeing the endpoint. If metal interference is presumed, a known standard made with plant / make-up water. Be
add one drop of Hardness Titrant to the sample before sure to collect a representative sample.
adding buffer or indicator. Include this drop of titrant when Video Procedure
calculating your results. Additional Hardness Buffer may
be necessary to view a clean endpoint.
STEP 1 STEP 3
Add Hardness Titrant one
2 Add 5 drops of Hardness
Buffer (460-S0850) and swirl to
4 drop at a time while swirling. Count
the number of drops until the sample
mix.
color changes from red to blue.
STEP 1 STEP 3
Add Hardness Titrant one
2 Add 5 drops of Hardness
Buffer (460-S0850) and swirl to
4 drop at a time while swirling. Count
the number of drops until the sample
mix.
color changes from red to blue.