Professional Documents
Culture Documents
IN CEMENT WORKS
D:\Documents and Settings\dpauling\My Documents\Temp\Ny-14340.DOC The copyrights for this document and all appendices are reserved
by HMC. The right to reproduce them entirely or in part in any
form and to make them available to third parties is subject to the
authorization of HMC.
-2-
SUMMARY
The measurement of individual process gas components in connection with emissions and
for the purpose of process optimisation is increasingly becoming one of the most important
procedures in cement works. Measuring techniques have advanced to the extent that there
is now virtually no such thing as a problem that cannot be solved technically with the
necessary specialised knowledge and means.
The point at which measurements are carried out is mainly a question of the measuring
problem involved. The most important sampling point for the control of preheater / calciner
kilns is normally after preheater, since there is the highest availability of the gas sampler.
The problems of sampling increase with increasing process temperature at the sampling
point. For gas sampling at the kiln inlet and up to gas temperatures in excess of 500C,
cooling and an emergency cooling or withdrawal system is required. Withdrawal devices
with compressed-air drives are more reliable for emergency use, so that preference is
given to electric drives.
The separation of dust from the sample gas preferably takes place as close as possible to
the entrance of the sampling probe. Sintered metal filters (alloyed steel) clog less quickly
than the usual ceramic filters. Compared to the filters outside the sampling probe, internal
filters require less maintenance and are smaller in volume. Wet sampling probes are
considered out-dated.
Three systems for gas sampling at the kiln inlet are described here in brief: Siemens type
FLK, Hartmann & Braun type 60 and type H (Holderbank probe). The latter has proved
reliable in use if daily maintenance is provided. For gas sampling below 500C, in addition
to standard products, proposals are submitted for the specific construction of sampling
probes.
The purpose of sample gas preparation is to supply the analysers with qualitatively and
quantitatively specified sample gas. The apparatus used for this purpose must be
generously dimensioned and equipped. This requirement applies in particular to the
sample gas pump, sample gas cooler and sample gas distribution. The temperature in the
measuring cabinet should be between 15 and 25C. A monitoring system must inform the
control system of any irregularities such as insufficient sample gas, system being
adjusted, etc.
The analyser room must provide the ambient conditions for the perfect functioning of the
instruments, i.e. free from vibrations, as close to the sampling point as possible, free from
dust and between 15 and 25C. As it also serves as a workroom for maintenance
personnel, it must offer sufficient space, good ventilation and a low noise level as possible.
The analysers used for the determination of individual components in the sample gas are
based on different physical effects such as light absorption. Paramagnetism, flame
ionisation and solid-state electrolysis. The instrument functions and theoretical
relationships will be dealt with only insofar as they are of significance for practical
application.
-3-
When measuring SO2 with NDIR analysers by the classic method, undefinable quantities of
SO2 are absorbed in the sample gas cooler. Corrections are possible by acidifying the
condensate in the sample gas cooler or using hot NDIR analysers. Preference is given to
the use of in-situ analysers.
Oxygen cannot be measured with light-sensitive analysers. The most reliable analysers
are still considered to be those that exploit the magnetic properties of oxygen. The life of
solid-state electrolytic systems such as zirconium dioxide probes is limited and depends on
the gas composition. Improved electrochemical measuring cells have been more widely for
the past few years and have a useful life of three years. They are small, lightweight,
require a minimum degree of evaluation and are therefore used more and more in
extractive multi-component analysers.
The flame ionisation detector is used for measurement of the total hydrocarbons (VOC =
Volatile Organic Compounds). The increasing requirements for continuous measurement
of the VOC - in connection with alternative fuels - are maintenance-intensive.
The GM 31 in-situ analyser of Messrs Sick enables three gas components to be measured
simultaneously. Compared with the earlier GM 21 (SO2 and dust) and GM 30 (additionally
NO) series, the GM 31 does not offer dust measurement.
The AR 620 and AR 650 series of Opsis are modern emission measuring instruments.
Depending on the type and number of components to be measured, infrared or ultra-violet
light used to the two together. The analyser or analysers are connected to the receiver at
the measuring channel via glass-fibre cable (light conductor). Control, monitoring and
evaluation is carried out by a standard PC. The analysers have been used several times
with good experience.
The new gas analysers with lasers are expected to have a promising future. These are in-
situ analysers that offer the benefits of high sensitivity, short response times and high
stability. They could provide the ideal preconditions for CO monitoring, but are (still)
relatively expensive.
Of the extractive measuring instruments, good experiences have been made with the
CEMAS FTIR analyser of Hartmann and Braun. The analyser can be programmed with all
IR selectable components. In order not to alter the originality of the sample gas, all gas
conducting parts are heated.
-4-
Recording the measured values with strip chart recorders should still suffice for a general
overview. However, in the majority of countries, digital evaluation of emissions is required
according to specific criteria. In the field of process optimisation, the measured values are
also only useful when available in the required form, e.g. as mean values, spreadsheet for
further evaluation, etc. At the kiln inlet, averaging or damping for lump fuel (tyres) feeding
in precalcining is absolutely necessary. In modern plants, the measured values can be
partly or wholly processed by the process control system.
The reliable functioning of each measuring instrument depends largely on the maintenance
provided. Visual checks and function tests must be carried out according to the age of the
measuring instruments and the specifications of the supplier.
Personnel must be suitably trained. Critical spare parts such as the sample gas pump,
filters valves, seals, etc., must be carried in stock.
The gas analysers must be checked according to the manufacturers instructions and
emission measurements carried out in compliance with the local requirements and
corrections made as necessary. If test gas should (still) be required, this must be obtained
in good time. Any fault signalling devices must be checked for proper functioning.
The accuracy of gas analysis as a whole is often overestimated, whereby the analysers are
least to blame in this respect. Significant errors occur when sampling (representative) due
to sorptions and reactions in sample gas preparation.
Gas analysis also has special monitoring functions for protection against explosions in the
electrostatic filter and coal pulverising mills. Simple monitoring of the CO content is usually
not sufficient for the protection of the electrostatic filter. The consequences are
unnecessary shutdowns of the filter. In addition, reference is made to a new analyser of
Messrs Sick, which has been designed specially for this task.
Finally, a table is provided showing the various conversion factors from ppm to mg/m3 for
the relevant gas components.
-5-
Table of contents
1. INTRODUCTION........................................................................................................... 9
2. PURPOSE ..................................................................................................................... 10
3. DEFINITIONS.............................................................................................................. 11
4. MEASURING METHODS.......................................................................................... 13
4.1 In-situ measurement.................................................................................................. 13
4.2 Extractive measurement............................................................................................ 14
4.3 Measuring points....................................................................................................... 15
5. GAS SAMPLING ........................................................................................................ 19
5.1 Kiln inlet (measuring point A) .................................................................................. 19
5.1.1 Preheater kilns (cyclone preheaters and grate preheaters kilns) ....................... 19
5.1.2 Wet kilns and long dry kilns ............................................................................. 22
5.1.3 Special instructions for gas sampling at the kiln inlet ...................................... 22
5.2 Other measuring points (B to G)............................................................................... 26
5.2.1 Kiln inlet chamber and riser duct (measuring point B)..................................... 26
5.2.2 Lower cyclone stages (measuring point C)....................................................... 26
5.2.3 Upper cyclone stages (measuring point D) ....................................................... 26
5.2.4 Downstream of cyclone or grate preheaters (measuring point E)..................... 27
5.2.5 Downstream of electrostatic filter (measuring points F, G).............................. 27
5.2.6 Coal pulverising mill......................................................................................... 28
6. GAS SAMPLING PROBES ......................................................................................... 31
6.1 Sampling probes in the temperature range 900 to 1500C ....................................... 31
6.1.1 Siemens type FLK............................................................................................. 31
6.1.2 Harman & Braun type 60S (formerly type 13) ................................................. 32
6.1.3 Hartmann & Braun type H (Holderbank probe)............................................ 33
6.2 Sampling probes in the temperature range 500 to 900C ......................................... 34
6.3 Sampling probes in the temperature range below 500C.......................................... 34
7. SAMPLE GAS PREPARATION ................................................................................. 37
7.1 Sample gas pipe ........................................................................................................ 37
7.2 Sample gas coolers.................................................................................................... 38
7.2.1 Specifications of a suitable sample gas cooler.................................................. 39
7.3 Sample gas pump ...................................................................................................... 39
7.4 Flow control and distribution.................................................................................... 40
7.4.1 Simple gas control............................................................................................. 41
-6-
1. INTRODUCTION
Gas analysis is among the most important, but also the most sophisticated methods of
measurement in cement works. This is due to the constantly growing demands in the
field of gaseous emissions and related requirements. Gas analysis fulfils an equally
important function in the field of automation, where the reliable and exact measurement
of certain gas components plays an essential role.
Advances in the field of gas analysis have taken place parallel to the trend of demands,
so that one can hardly speak of technically unsolvable problems any longer. If
problems should nevertheless occur, they are mainly due to the following
The aim of this report is to prevent problems of this kind occurring when renewing
and adding gas analysing apparatus.
-10-
2. PURPOSE
The continuous measurement of the constituents of certain components in the gas flows
(gas analysis) in cement works consists of three areas of activity:
c: Process measuring systems (e.g. O2, CO, NO) for optimisation of the combustion
process.
Important for all these activities is a high availability of the measuring equipment and
sufficient accuracy of the measured values. The report therefore attaches particular
importance to gas sampling and preparation, maintenance and quality assurance as
well as possible measuring errors. Finally, hints and practical advice are given, based
on many years of experience.
-11-
3. DEFINITIONS
Indication delay (90% time) of an analyser or a measuring instrument: The time that
elapses from a sudden change in the gas concentration at the entrance of the analyser
or probe up to indication of 90% of this change.
Downtime: The time during which faulty conditions of the measuring instrument and
their correction unforeseeably make the generation of measuring or adjusting signals
impossible.
Extractive sampling: The sample gas is extracted as a partial gas flow from the
process gas prior to analysis and conditioned.
Gradient monitoring: Monitoring the gas concentration change gradient (e.g. control of
CO cut-off).
In-situ: The measuring instrument is located in or on the process gas duct; analysis is
carried out in the process gas (e.g. optical dust content measuring instrument or
zirconium oxygen probe).
Adjustment: Setting or adjustment of the gas analyser with test or zero gas with the
aim of adjusting the indication of the gas analyser as close as possible to the gas
concentration to be measured in the sample gas. With newer analysers, this can also
take place with built-in gas-filled glass cuvettes.
Life zero: A specific quantity (normally mA) prevails in the electric output signal when
the analyser physically measures zero. The advantage of this system is that the
output signal is electrically set to zero only if the electrical measuring circuit fails.
Sample gas: That part of the process gas extracted from the process gas and passed
to the analyser for measurement.
Sample gas preparation: The entire equipment used for the qualitative and
quantitative preparation of the sample gas to ensure that the same conforms to the
specifications of the analysers.
Measuring equipment: All devices and instruments required for the measurement of
gas component concentrations.
Measured component: The gas component in sample gas, the concentration of which
is to be measured.
Zero gas: Test gas used for the adjustment or readjustment of the zero of a gas
analyser.
-12-
Test gas: Gas with a known composition for adjustment or readjustment of gas
analysers. It consists of the measured component and one or several accompanying
gases (e.g. nitrogen/ carbonic acid + measuring component).
Representativity: The correspondence of the measured signal with the definition of the
measured quantity, e.g. the composition of the gas at the kiln inlet measured at a point
in relation to the average gas composition over the entire cross-section.
Disturbing component: Component in the sample gas that falsifies the measured
value, e.g. if the gas analyser has a cross-sensitivity or chemical reactions distort
measurements.
Dead time of an analyser or measuring instrument: The time that elapses from a
sudden change in the gas concentration at the entrance of the analyser or probe up to
the start of the indication of this change.
Availability: The time during which the measuring instrument generates usable signals.
Supply pressure: The sample gas pressure necessary for supplying the analysers.
4. MEASURING METHODS
No gas sampling
Undelayed indication
Less maintenance
Optical systems
Optical instruments are flanged directly on to the gas channel to be measured (e.g.
chimney). These consist of a transmitter (normally light) and a receiver or reflector
arranged opposite. While passing through the channel, the light beam emitted from
the transmitter (infrared, ultraviolet or laser) is changed by certain gas components.
After appropriate processing, this change results in the measured quantity (see
chapter 8.6).
Optical measuring instruments (8.6.1, 8.6.2, 8.6.3) are used mainly for the
measurement of pollutant emissions, e.g. dust content, SO2, NO, etc. The range of
instruments available for measuring further gas components is increasing in line with
technological advances. The application of optical measuring in-situ instruments is
restricted by two factors; the dust content and the given length of the measuring path,
which is in direct relation to the sensitivity. For use in the cement works, these
restrictions mean that they only function in pure gas and their application is limited
solely to the measurement of emissions, although good experience has been made
in this respect. Thanks to improved representativity, in-situ measuring methods are
technically superior to extractive methods.
-14-
Analysers
With the exception of the dust content, the majority of components can be measured
by the extractive method. In-situ measuring instruments offer the following
advantages:
Adjustment and calibration is simpler (with built-in reference cuvettes or test gas).
Drying the gases during gas preparation by means of cooling to a dew point of about
3C can be disturbing for certain components. These are logically water vapour
(H2O) sulphur dioxide (SO2) and certain hydrocarbon compounds (VOC) (see chapter
8.4).
To exclude disturbing influences during sample gas drying, there are instruments
with operating temperatures that are far above the dew point (up to 250C). The
components of these instruments are subject to increased load when exposed to
these temperatures, which inevitably leads to an increased susceptibility to faults.
Experiences made with an instrument of this type are described in chapter 8.6.4.
-15-
Figures 1 to 3 show the typical measuring points at various kiln systems (cyclone
preheaters, grate preheaters and wet kilns). The associated tables show the usually
measured components as well as the purpose of their measurement. The choice of
measuring point at the kiln system at which measurements are to be made, depends
primarily on the purpose of measurement.
-16-
G
E
F
CO-monitoring O2 CO NO
CO-monitoring O2 CO NO
Emissions O2 CO NO
Emissions O2 CO NO CO2
Unsuitable
CO-monitoring O2 CO NO
Emissions O2 CO NO CO2
Unsuitable
5. GAS SAMPLING
At the rotary kiln itself, only the measuring point at the feed end of the kiln is
realisable with reasonable effort. The components normally measured there are
needed as reference quantities for an optimal combustion process (usually O2, CO,
NO).
To prevent distortion of the measured signals due to false air from the kiln inlet seal,
the probe tip must be located distinctly within the rotating part. The sampling
conditions at the kiln inlet vary considerably, depending on the type of kiln (preheater
kilns, long dry kilns and wet kilns).
The process gas contains large proportions of alkalis, chlorides and sulphates that
are subject to high local and time variations. Depending on the process and raw
material situation, the alkali compounds combine with the dust to form strongly
adherent deposits. To counteract their effects to make gas sampling possible at all,
elaborate measures are necessary from a measuring point of view.
Owing to the high temperatures, the provision of external cooling is necessary for all
types of sampling probes in the area of the kiln inlet.
The principle mounting arrangement is shown in Figure 4. Owing to the wide range
of kiln systems available, it is impossible to define an exact mounting point for all
types of kilns in advance. The following criteria play a dominant role:
Available space
Accessibility
According to experience, the gas composition at the kiln inlet is not homogeneous.
In consequence, the sample is only representative of the average gas composition to
a limited degree, irrespective of where the probe is positioned. The measuring point
or sampling point in the rotating part of the kiln, as will be shown later, cannot be
determined primarily on the basis of measuring criteria. An individually adapted
interpretation of the measured values (averaging), correlation with disturbance
variables, etc.) is far more important here than the position of the probe (see chapter
9).
The mounting position shown in Figure 4 is for reference only. Positions and
dimensions can vary considerably from plant to plant. The undisturbed, reliable
function of the probe and the warranty that no false air will be drawn in from the kiln
inlet seal always has the highest priority.
As the probe consists of relatively long sections, the access way must not be too
narrow. The mounting position should be accessible via steps not ladders.
Pokeholes, which must be operated regularly, are located in the immediate vicinity of
the probe. It must be ensured that the sampling probe does not interfere with the
working space provided for maintenance personnel.
In principle, the probe must be mounted on the side where the direction of rotation of
the kiln shell is directed downwards. If conditions do not permit, the other side can
also be selected, provided the probe can be positioned at a sufficient distance to the
material bed.
Meal inlet pipes, pokeholes, measuring connections, etc., mostly restrict free
selection of the mounting position, the main aim is to minimise hindrance of
maintenance personnel.
The inevitably cooled probe surfaces encourage the formation of deposits in the kiln
inlet chamber. These deposits not only hinder subsequent removal of a probe, they
can negatively influence the geometry of the kiln inlet chamber and also cause
pressure loss. To prevent this happening, it is recommended to regularly blow-out
the gap between the pipe and probe by means of a mounted air cannon. The pipe
must be dimensioned, so that a gap of about 25mm is produced between the probe
and inner pipe wall.
-21-
Retraction device
E
E
ca L+1.5m L
A = 0....300 mm = 30....45
B = Min. 500 mm L = Nominal length of the probe
C = B + wall thickness N = Inclination from horizontal axis
E = Min 200 mm X = As short as possible
Under these conditions, gas sampling is much simpler than for kilns with preheaters.
No cooling of the probe is necessary and the measured signal has improved
representativity, as better intermixture of the gases takes place due to the internals
(chains, crosses). With regard to positioning, mounting conditions, etc. roughly the
same conditions apply as for kilns with preheaters.
The above process conditions show that gas sampling from this difficult environment
requires particular measures:
A filter system must ensure that the large amounts of diversely structured dust is
removed prior to sample gas preparation
Very often, deposits build up at the sampling system and clog the probe entrance,
so that measures must be taken for their prevention.
Probe cooling
Liquid cooling is used for all known probes. The cooling medium, which is usually
water, transfers the heat in a closed system to water-air or water-water heat
exchanger to prevent calcareous deposits. The energy produced varies
considerably. Depending on the condition of the probe surface, a large amount
(clean surface) or small amount (dust or deposits on the probe) must be removed.
Based on experience, an energy flow of maximum 30kW per probe meter must be
removed within the kiln (probe with 10cm outside diameter). In extreme cases, the
energy flow varies between 1kW and 30kW per probe meter.
According to experience, the risk of deposit formation increases with reducing probe
or cooling circuit temperature. As a countermeasure, some probes are operated at
higher temperatures. This takes place by controlling the cooling circuit or by
internally heating the probe.
If a water-air heat exchanger is used, it must be taken into account in dimensioning
that the maximum cooling capacity is still sufficient even with heavy clogging of the
heat exchanger surfaces. It is also recommended to add an anti-freeze to the water
in cold regions.
-23-
In order to better control incrustations and deposit build-up in and on the probes,
cooling systems were developed which enable the probes to be operated at higher
temperatures. An example is the steam jacket probe. This probe functions as a heat
conducting tube, whereby the heat of evaporation of water is used as a heat carrier
to the directly flanged-on heat exchanger (condenser).
The probe temperature is 120 to 150C, the internal pressure is 1.5 to 5 bar. The
probe only functions properly if it can be mounted almost vertically (less than 30
from the vertical axis). This requirement virtually excludes the mounting method
shown in Figure 4 in an almost horizontal position, so that its possible applications
are very restricted.
If a synthetic heat carrying liquid is used instead of water, the probes can be
operated at even higher temperatures (up to 200C) (see chapter 6.1.1).
In the event of failure of the cooling system, the probe would be damaged within a
few minutes. Countermeasures are essential. For emergency cooling purposes,
water supplied from storage tanks can be used and feed directly into the probe circuit
if required. Emergency cooling systems of this kind must also function in the event of
a total power failure. Pneumatically operated withdrawal devices have proved more
reliable than emergency cooling systems (see Figure 5). Since compressed air can
be stored without problems, pneumatic operation has the advantage that the probe
can be withdrawn from the danger, zone even in the event of a total power failure,
within a reasonable time. The automatic insertion and withdrawal device also has
further significant advantages:
Regular insertion and withdrawal prevents the formation of deposits on the probe and
their build-up.
With dry filtration, the sample gas is withdrawn from the process by suction in an
unpurified condition. Dust separation takes place inside the probe or at its outer end.
The produced dust must be removed from the filter from time to time. Whether this
takes place manually by replacing the filter element or automatically by blowing back
into the sampling space, the sample gas flow is always interrupted. This interruption
has a negative effect on the availability of the measuring instrument. To prevent
condensation of any kind, the filter must have a minimum operating temperature of
150C.
Rail with hose guide device
Suspension point
-24-
Drive Chain
Pneumatical drive
Gas Probe
Shutter flap
Mounting flange
With wet filtration, the sample gas flows through a water curtain at the probe tip. This
produces a slurry of water and dust, which together with the sample gas discharges
through the wet probe, which in contrast to the dry probe is inclined outwards. The
slurry discharges, the gas is dried and passed to the analysers.
Wet filtration has the disadvantage that various gas components are also dissolved
in water. SO2 is completely washed out and the CO2 concentration reduces to an
uncontrollable degree. As a result of this change in the total composition, the relative
contents of the other gas components increase, so that the composition of the
sample gas is no longer correct. If the injector water circulates in a closed system,
the circulated water becomes acidic and behaves virtually inert towards CO2.
However, susceptibility to corrosion of all parts coming into contact with the water
must be observed.
The level of maintenance necessary with wet filtration is slightly less and above all
technically less demanding than with dry filtration. However, in view of the distortion
of measured values and the reliable dry systems that are currently available, its
application is no longer recommended.
Filter cleaning
The dust separated in the dry filter must be removed from time to time. The degree
of clogging of the filter, including the gas sampling probe, can be determined by
means of a vacuum gauge arranged between the filter and sample gas pump. The
maximum permissible value for cleaning differs from plant to plant and must be
individually determined.
Compressed air is normally used for cleaning and removing the dust from the filter. If
the filter is arranged at the outer end of the probe, cleaning should take place in two
stages.
The compressed air must be free from oil and water. Oil in the purging air produces
incorrect CO in the filter and excessive moisture encourages encrustation of the filter.
With regard to gas sampling, the conditions for the wet kiln and long dry kiln are
considerably simpler than for preheater kilns. The lower temperatures usually
obviate the need of probe cooling. The lower dust load enables the use of simpler
filter systems without special cleaning, as used following the heat exchanger in the
case of dry kilns (see chapter 6.3).
-26-
In the kiln inlet chamber and riser duct, as at the rotary kiln inlet, local and time-
related concentration differences can be expected. Above the meal inlet of the
lowest cyclone, there is no longer any danger of a deposit build-up through circulated
materials. However, the gas composition no longer corresponds at this point with the
gas composition in the rotary kiln inlet. It is distorted by the reduction of CO in the
riser duct and through dilution with false air and CO2 from deacidification of the
burned product.
When burning secondary fuel, the measuring point must be arranged below the
secondary fuel inlet.
With regard to gas sampling, virtually the same conditions apply as at the kiln inlet.
This applies similarly to a large extent to sampling from the calcining chambers of
grate preheater kilns.
From a measurement point of view, gas analysis in this area is virtually pointless,
particularly in plants with pre-calcining. However, if measurements are still carried
on a frequent basis, these are not for measurement reasons, but the more
favourable conditions with regard to sampling (deposit formation and temperature)
than further below, e.g. in the rotary kiln inlet. A cooling system is necessary in any
event.
At the point, similar conditions prevail for gas sampling as in the riser duct
(measuring point D). In the case of grate preheaters, this point approximately
corresponds with the conditions downstream of the grate, but less dust loading. To
be taken into account for CO monitoring in grate preheater kilns is the fact that no
cooling tower is usually available to delay the residence time of the gases in a
positive sense. For this, the gas quantity can be increased on account of the low
dust loading.
In the majority of cases, the components relevant for emission are measured.
Preference must be given to the point downstream of the fan, as the gas composition
is no longer subject to change before the exhaust gas discharges from the stack.
Owing to the turbulence of the process gas in the plant sections and fans, it can
generally be assumed that the process gas is well intermixed.
Owing to the probability of filter shutdowns, gas sampling requires a dust filter as
provided at the measuring points in the raw gas flow. The high dew point makes it
necessary for the probe to be heated at the critical points. No differences exist with
regard to the type of kiln (cyclones, grate preheaters or wet kilns).
Gas analysis in the area of the coal pulverising mill serves solely for the purpose of
quickly detecting smouldering fires and/or potentially explosive gas concentrations.
The possible measuring points are shown in Figure 6.
-29-
The measuring point in the filter serves especially for monitoring glow nests when the
coal pulverising mill is stationary. Owing to the large filter volume, glow nests can
only be detected quickly enough with sufficient circulation of the process gas. For
this reason, 200-300m3/h process gas should be drawn out by suction via a separate
gas pipe and fed back again at the filter inlet when the coal pulverising mill is
stationary. The gas sampling probe is then arranged in the gas pipe. When the coal
pulverising mill is in operation, the gas pipe is closed via a valve.
In principle, gas sampling probes can be divided into three categories according to
temperature range:
This category includes gas sampling at the inlet of preheater kilns with and without
pre-calcining.
Brief description
Compact gas sampling system with special cooling fluid permitting probe operating
temperatures up to 230C. The condensation of alkali vapours and in turn the
danger of incrustations is drastically reduced. The suction opening is arranged on
the side of the probe tube and should be located in the flow slipstream. This
arrangement offers selective dust separation. An electrically or pneumatically
operated insertion / withdrawal device is an integral part of the probe.
Dust removal filter: External, heated, blow-back with compressed air, large
volume
Comment
Despite effective measures, stubborn blockages can occur, particularly when kiln
operation is loaded by high alkali circuits. Owing to the lateral arrangement of the
suction opening, the gases within the probe tube flow through an elbow. The
removal of incrusted deposits from this elbow is far more difficult due to poor access
than if the gases were drawn out by suction at the end of the stem and through a
straight tube. In view of the turbulent flow conditions at the sampling point, it is
doubtful whether lateral suction results in selective dust separation.
The large overall volume of the dust filter permits a relatively large gas throughput.
The resulting delay and damping of the indication is of little significance from a
measurement point of view.
Available for the drive of the withdrawal device is either an electric motor or
compressed-air drive. As compressed air can be stored without problems, the
compressed-air drive offers more reliability than an electric motor in the event of a
power failure. Emergency operation with crank handle is also provided.
With difficult kiln operation, alkali condensation can occur and block the entrance of
the probe, as already mentioned, despite increased operating temperature.
Brief description
Gas sampling probe with closed cooling system (cooling medium water) and
temperature control. Integrated emergency cooling system with fresh water. Lateral
arrangement of two suction openings (Figure 7).
Comment
Widely used. Functions well in non-extreme alkali conditions. The same remarks
apply here with regard to the lateral arrangement of the suction openings as for the
Siemens system.
With large amounts of dust, the filter must be blown-out at brief intervals, The
resulting loss of signal can have a critical effect on the availability of the measuring
signal.
The operating temperature is too low to completely prevent alkali condensation. An
automatic insertion and withdrawal device is optionally available.
-33-
1 2 10 11 3 7 5 8
9
12
12
6 5
1 Gas inlet holes 7 Ceramic fliter
2 Gas tube 8 Protection case
3 Filtering unit 9 Connecting for air blaster
4 Gas outlet 10 Built in tube
5 Compressed air inlet 11 Mounting flange
6 Ring heating element 12 Cooling water in/outlet
Fig. 7
Brief description
The main features of this probe are the automatically functioning mechanical removal
of any incrustations at the probe inlet as well as a new type of filter system (see
Figure 8). Cooling takes place in a closed system (cooling medium water) with
temperature controller.
About 20 probes of this type are in use worldwide and are functioning efficiently.
Marketing takes place via Messrs ELSAG BAILEY Hartmann & Braun in Frankfurt
under the designation type H.
Dust removal filter: Internal, heated filter of sintered metal, cleaning with
compressed air
Thanks to its modular concept, the probe can be provided with the same insertion
and withdrawal device as type 60.
-34-
1 2 3 4 7 10 11
9
12
12
5 8 6
Fig. 8
Comment
This sampling system has proved highly reliable in practical use, even under difficult
conditions. The availability satisfies the high requirements of high-level control
systems. The maintenance costs are low thanks to the simple and maintenance-
friendly design of the system.
With regard to the sampling probes, similar conditions prevail in this temperature as
at higher temperatures (e.g. at the kiln inlet of preheater kilns).
Temperatures between 500 and 900C are typical for cyclone preheater kilns in the
area of the lower cyclone stages (measuring point C). As already noted,
measurements there are almost pointless, particularly in plants with precalcining. If
measurements are still to be carried out despite this, steam jack probes are
considered ideal for this purpose when mounted vertically in the cyclone roof.
Typical for the application of such probes are the measuring points downstream of
the heat exchanger or grate preheater, upstream or downstream of the electrostatic
-35-
filter as well as at the kiln inlet of wet or long dry kilns. Cooling is unnecessary. With
regard to the place of use or purpose of measurement, there are certain differences
in the design of the probes (e.g. mounting length). Decisive for the quality of these
probes is the dust filtration. The discovery of fine-pore filter tubes of sintered
stainless steel brought about great advances. The level of maintenance is low, a
block-back device is unnecessary. The sample gas is clean and available for weeks
to months without interruption. Probes with such filters are not available from
established suppliers. They still use fine ceramic filters that enable a comparatively
long useful life. However, it must be ensured that the gas conducting tube is heated
at the penetration point (Figure 9).
The optimal porosity of sintered metal filters is only 0.5m. As the main proportion of
the dust particles is larger, the pores clog at a correspondingly slow rate. According
to experience, they have a useful life of four weeks to several months, depending on
the properties of the dust. If the pressure drop becomes excessive due to clogging,
the filter tubes can be cleaned with acid.
With wet kilns and long dry kilns, the probes must be lengthened, depending on the
size of the smoke chamber. In order to ensure the necessary mechanical strength,
supporting tube of a larger diameter must be mounted, into which the probe is
inserted.
A further important element is the heating in the penetration zone from the measuring
channel to the outside. In order to prevent any kind of condensation, the
temperature must not fall below 100C at any point.
For all gas conducting parts, stainless steel, e.g. No. 1.4541 or 1.4571 must be used.
Thread e.g. 25 x 1.5
430 Length as required max 1500
10
ca 40
25 (1")
1 2 3 4 5 6 7 8
Manufacturer:
Mott Metallurgical Corp.
Farmington Industrial Park
Farmington CT 06032
Fax No (USA) 203 674 1489
3 Gas conducting tube. Outside diam. 10 mm, inner diam. 6 mm, stainless steel
4 Centering discs approx. every 500 mm
5 Carrier tube. Outside diam. 40 mm, wall thickness 2 mm, stainless steeel
6 Heating coil Heated lenght approx. 300 mm ca. 200 W
7 Mounting flange
8 Sample gas connection
The purpose of sample gas preparation is to supply the downstream gas analysers with
sample gas, so that it conforms qualitatively and quantitatively with the specifications of
the analysers. Such an arrangement normally consists of the following elements:
The probe itself as well as the sample gas pipe between the sampling probe and
sample gas cooler should be heated to between 100 and 150C (for measurements
with FID devices up to 200C). This heating has the following purpose:
> Condensation in the sample gas pipe can distort certain measured values.
> Incrustations and blockage of the sample gas pipe are accelerated by the
formation of condensate.
For selection purposes, it should be noted that with heating band controlled pipes,
the temperature is limited to a maximum of 105C. This temperature does not
normally suffice for process measurements but for emission measurements. The
following specifications must be observed:
Heated pipes are relatively expensive. When planning a measuring system, the
sample gas pipes must therefore be as short as possible. The following points must
be observed:
b) Possible installation of the sample gas cooler in the vicinity of the sampling point.
For automatic compressed-air cleaning via the sample gas pipe, it must be ensured
that this has sufficient pressure resistance.
For connections, high-quality stainless steel clamping ring fittings, e.g. Swagelok,
Serto or similar, must generally be used.
Sample gas coolers serve to remove most of the water vapour from the sample gas
at temperatures of about 3C and keep the residual content at a constant value. The
latter is important particularly for emission measurement. The temperature indication
of the sample gas cooler normally shows the operating temperature of the cooling
medium. If the sample gas cooler is subject to excessive load or ambient
temperature, the sample gas temperature at the cooler outlet can be considerably
higher than the indication. The dew point temperature then no longer corresponds
with the cooling medium indication.
The sample gas cooler is usually accommodated in a cabinet together with the
analysers and the remaining sample gas preparation apparatus. Instead of using
long heated sample gas pipes, it may be advantageous to install the sample gas
cooler immediately downstream of the sample gas probe in a separate equipment
cabinet.
-39-
Condensate collecting tanks have the disadvantage that indication of the measured
values is greatly delayed due to the additional volume. The produced condensate
should therefore be pumped out using automatic hose pumps or diaphragm liquid
pumps.
Also available for the majority of sample gas coolers are moisture monitoring devices
at the sample gas outlet. Such devices are urgently recommended.
Mainly diaphragm and compressor pumps are used. The sample gas volume
necessary for the gas analysers is between 30 and 100l/h. In order to keep
indication delays to a minimum, a larger sample gas volume is often extracted from
the process gas. The excess sample gas can be diverted via a bypass shortly
upstream of the analyser. If possible, a large part of the measuring instrument
should be operated under pressure, as false air can infiltrate during suction phase.
Sample gas pumps installed upstream of the sample gas cooler must be heated.
This arrangement has the advantage that no false air is sucked in the event of leaks
in the sample gas cooler. However, according to experience, heated pumps are
more susceptible to faults than cold operated pumps, so that arrangement
downstream of the sample gas cooler is still more advantageous.
-40-
With throttling on the pressure side, a pressure control valve must ensure that the
pressure downstream of the pump does not increase to an unnecessary degree.
Significant pressure drops in the sample gas encourages the formation of powdery
salts (white powder). The following diagram shows the characteristics of a suitable
sample gas pump.
Vacuum
Excess pressure
Gas flow [l/min] at 20C
Vacuum Overpressure
20
Gas flow [l/min] at 20C
15
10
0
0 0,2 0,6 0 0,5 1 1,5 2 2,5 [bar]
The arrangement for control and distribution of the sample gas depends on the
number and specifications of the analysers to be supplied. Multi-component
analysers with only connection solve the problem of gas distribution internally.
However, as soon as two or more analysers have to be supplied with gas, the inlet
pressure must be taken into account in distribution. At a low supply pressure, the
analysers can be connected in series. If analysers have to be connected with
varying supply pressure, they must be connected in parallel via appropriate gas
distribution.
-41-
1 3
7 13
8 9
12
2
6
5
4 11
10
A B C
The flowmeter monitors and controls the set sample and test gas volume.
Rotameters fitted with variable limit transmitters are normally used for setting the flow
rate. The signal transmitters should signal especially when the necessary flow is not
reached. This is particularly important for gas analysers with safety functions.
As already mentioned, not all gas analysers have the same conditions with respect to
supply pressures. Optical analysers normally require less pressure than oxygen
analysers, for example, which use the paramagnetic measuring principle. However,
special attention must be given to the supply pressure of the analysers when
dimensioning the sample gas preparation apparatus. As soon as several analysers
have to be operated, parallel connection is recommended, despite the additional
work involved. For this purpose, the sample gas flow must be divided appropriate to
the number of gas analysers to be supplied. The following flow diagram shows how
this division can take place. The sample gas pump supplies a common supply pipe,
which is under increased sample gas pressure. A spring-loaded pressure control
valve controls the pressure in this pipe. The excess sample gas is diverted. A
further, slightly more elaborate option, is to use a separate sample gas pump for
each analyser.
Parallel connection with pressure control valve
A B C
The available electronic flowmeters are reliable and should be used to an increasing
extent for automatic gas volume control purposes.
D:\Documents and Settings\dpauling\My Documents\Temp\Ny-14340.DOC The copyrights for this document and all appendices are reserved
by HMC. The right to reproduce them entirely or in part in any
form and to make them available to third parties is subject to the
authorization of HMC.
-43-
Setpoint
S Power
Supply
Flow Controller
Tests with automatic control have confirmed that the reliability of the gas supply can
be considerably improved with this system. It offers decisive functional advantages
compared with rotameters and is less susceptible to faults. A constant gas flow
positively affects overall gas analysis in every respect. Since the controllers have a
throttling effect on the gas flow, they are preferably arranged on the suction side of
the pump. For flow indication purposes, a volume-proportional signal is available.
This enables ideal monitoring via the control system. In addition, a rotameter
(without needle valve) can still be used for monitoring locally.
-44-
Coarse filters are normally installed inside or downstream of the probe, so that a
large proportion of the dust is already separated before the sample gas enters the
sample gas preparation apparatus. If sintered metal filters with a porosity of <0.5m
are already used in the probe, the sample would be sufficiently clean. Nonetheless,
fine afterfiltration directly upstream of the analysers is recommended, as impurities
can also infiltrate the system downstream of the probe. Ceramic filters of silicon
carbide, mineral fibres or filter casings of borosilicate glass fibres are also frequently
used. Silicone carbide composites have a high separation efficiency of about 99.9%
at an average grain diameter of 1.2m and are temperature resistant up to 660C.
The disadvantage of all these filters compared to sintered metal filters, apart from
their lower separation efficiency is that they have to be cleaned and blown out or
replaced at shorter intervals
For separation of the dust particles and sublimated salts left behind in the sample
gas, membrane filter, e.g. of glass fibres or PTFE should be provided.
7.5.3 Valves
Changeover valves
If automatic blowing out of the filter and/or probes is required, the gas path must be
able to be diverted automatically. For this purpose, stainless steel solenoid valves or
mechanically/ pneumatically operated ball sliding valves of Teflon (housing) and
stainless steel (ball) are used.
Valves installed in moist sample gas (upstream of the sample gas cooler) must be
mounted on a temperature-controlled valve plate in order to prevent corrosion as a
consequence of the dew point not being reached.
Valves, particularly solenoid valves, have the tendency to leak, even when slightly
fouled. They enable a high degree of automation of the measuring instrument, e.g.
automatic calibration, however, experience has shown that their susceptibility to
faults almost cancels out the convenience they are expected to provide (see chapter
11.7).
-45-
The purpose of the pressure control valve is to ensure that the supply pressure of the
analysers remains fairly constant. The excess gas delivered by the pump discharges
via a spring-loaded body. The nominal pressure of the valve must be slightly higher
than the highest specified supply pressure of all parallel connected analysers.
7.6 Adjustment
The necessity of regular adjustments depends on the purpose of the analysis and the
stability of the analysers. For emission measurements, a rotation of one to two
weeks is sufficient, provided the regulations in the country concerned do not contain
any tighter requirements. During adjustment, a preferably manually-operated two-
way directional control valve switches the gas supply to a bottle containing test gas.
Newer gas analysers are normally provided with an automatic adjusting device. By
means of an internal or external command a glass cuvette filled with the respective
gas is inserted into the sensor and the analyser is automatically readjusted. During
this time, the last measured value remains indicated, so that the subsequent signal
processing system is not disturbed by the procedure. With older analysers that do
not have this type of automatic device, operating personnel or the process control
system must be informed of the adjustment. With CO monitoring systems, activation
of the automatic filter shutdown system must be prevented.
Analysers with so-called calibrating cuvettes have the big advantage that no more
test gas is necessary. Manufacturers and most nationally customary regulations
accept this type of adjustment over a period of 2...3 years. After this time, the
analysers must be overhauled and inspected and readjusted by an appropriately
licensed institute.
No sample gas must discharge into the analyser room. Every gas outlet must be
connected to a pipe and discharged to the atmosphere or returned to the main gas
flow. Pipes with large cross-sections must be used in order to prevent any back
pressure.
An option should be provided to easily connect the sample gas pipe gas-tight to a
nitrogen bottle directly downstream of the sampling probe.
-46-
The complete measuring apparatus is purged with nitrogen with the gas preparation
apparatus operating normally. The oxygen analyser shows zero indication soon
afterwards. After several minutes, the nitrogen supply is interrupted, so that a
vacuum is produced in the system. Caution! The regulating valves on the test
bottles are often not vacuum-tight. If the measured value of the oxygen analyser
should increase again within a space of about 20 seconds, this means that there is a
leak in the system. An increase of several per cent within several minutes is
considered normal.
The suction pressure is a good indicator of the condition of the filter or filters between
the probe and sample gas pump. By measuring this pressure with a pressure
transmitter or contact pressure gauge, the automatic cleaning process can be
initiated when a certain level is reached. As a result of this optimising measure, the
measured signal is not interrupted unnecessarily and the filters are cleaned or blown
back before being irreversibly clogged.
13 9
A
10
14
1 18
8
Calibration Operation
16
3 4 12
2
7
6 17
15
Blow-out Operation
For the analysers and sample gas preparation apparatus, a double 19 equipment
cabinet with one field respectively for the gas preparation apparatus and analysers is
normally sufficient. However, other equipment is often installed in this clean, air-
conditioned analyser room. In such cases, it must be ensured that sufficient room is
provided to enable maintenance personnel to move freely and that there is adequate
room for material (e.g. test gas bottles, tools, etc.) and additional space.
7.9 Location
If space is available at a suitable location that is not exposed to excessive dust, heat
radiation and noise, a switchgear cabinet is sufficient for accommodating the
measuring equipment. A separate, closed room is unnecessary under such
conditions. The volume to be air-conditioned can be considerably reduced as a
result.
8. ANALYSERS
Light absorption
Paramagnetism
Flame ionisation and
Solid-state electrolytic systems
All newer analysers process the analogue signals output by the sensors digitally.
Digitalisation has provided the following advantages:
The internal functions of the analysers will not be dealt with here. Sufficient information
in this respect is provided in the manufacturers documents and technical literature.
Important for practical application is knowing how to use the various measuring methods
effectively.
The manufacturer plays a secondary role in analyser selection from a qualitative point of
view. Measuring equipment faults are rarely attributed to faulty analysers. It goes
without saying that only types designed for industrial purposes are suitable for use in
cement works and certainly no laboratory equipment. Suitability for industrial purposes
means, for example:
With the exception of oxygen, the majority of interesting gas components can be
measured by infrared absorption. Available are analysers for extractive and in-situ
application. In the main, infrared is used for the extractive method. The following
advantages are offered by these NDIR (non-dispersive infrared) analysers:
The not particularly high sensitivity of the NDIR analysers does not normally pose a
problem in the cement industry. The smallest measuring range, e.g. for CO is about
100ppm, for NO about 500ppm and for CO2 about 20 %. Cross-sensitivities to water
vapour and CO2 are present. Water vapour as a disturbing component plays a
secondary role when the gases in the sample gas cooler are dried with sufficient
stability. The cross-sensitivity of CO2 is usually within the tolerance when
compensated in the analyser. The influence can be further reduced when an amount
of CO2 corresponding to the average value of the sample gas is mixed with the test
gas. As the CO2 content in the cement process is extremely high, unestablished
analysers should be tested in this respect prior to their use.
The most well known manufacturers today produce multi-channel NDIR analysers,
enabling the prices per measured component to be distinctly reduced. The majority
of manufacturers also offer analysers in different quality classes (low-cost analysers).
As a rule, the low-cost analysers are not temperature-stabilised, but temperature-
compensated and are therefore not as stable as the analysers at the top end of the
price scale. However, if external fluctuations are kept to a minimum, the
temperature-compensated analysers meet the requirements in the majority of cases.
In principle, NDIR analysers are suitable for the measurement of SO2. However,
tests on kiln systems with cyclone preheaters have shown that insufficient SO2 is
measured with extractive sample gas sampling and preparation (in accordance with
DIN 2462, page 4). The magnitude of the deficiency can vary considerably and can
therefore not be calibrated. These measuring errors are attributed to chemical
reactions of the SO2 with NH3 (ammonia) in the sample gas cooler (see chapter
11.2). The following options are available to eliminate these errors:
With the addition of about 8 to 12ml phosphoric acid (5%) in the sample gas cooler,
the pH of the condensate reduces to a value below 1.5. The reaction of the SO2 with
ammonia is accordingly prevented.
It is obvious that hot analysers, whose entire measuring system must function at
temperatures between 150C and 250C, are more susceptible than cold analysers.
According to experience, the heated sample gas pump causes the greatest
problems.
The difference of the NDUV analysers compared with NDIR is mainly that they are
less cross-sensitive to water vapour and CO2 and higher sensitivities can be
achieved. In the cement industry, NDUV analysers were initially used for the
measurement of NO components. Meanwhile, the less expensive NDIR analysers
are being more widely used for NO and have proved reliable.
8.3 Paramagnetism
Oxygen does not have a usable spectrum either in IR or UV light. In future, laser
analysers /chapter 8.6.3) will be the first to be able to measure oxygen with light in
addition to other components. The paramagnetic properties of oxygen (oxygen
molecules are strongly attracted in a magnetic field) are still used for oxygen
measurement, however with a declining trend. In practice, two methods are widely
used:
In recent years, analysers operating on the torsion balance or also dumb bell
principle have become increasingly popular.
The measuring effect is based solely on the magnetic forces of the oxygen molecule
and is therefore comparatively less cross-sensitive to other gas components.
However, the measuring chamber reacts sensitively to dirt and condensate.
-52-
This measuring method is used for measurement of the volatile content of unburnt
hydrocarbons (VOC) (Volatile Organic Compounds). Organic carbon compounds
contained in the sample gas are ionised in a hydrogen flame. The ion quantity is
almost proportional to the number of carbon atoms involved in combustion. The ionic
voltage is detected with the aid of an electrode and brought to indication via a high-
resistance amplifier. Propane gas in nitrogen is normally used for calibration
purposes.
Continuous measurements with the flame ionisation detector (FID) are increasingly
prescribed, as soon as alternative fuels are burned. FID analysers are unfortunately
susceptible to faults and require considerable maintenance by suitably trained
personnel.
The entire sample gas preparation system must not fall at any point below the
sample gas temperature, i.e. the sampler, sample gas pipe and pump must be
heated. The temperature in the ionisation chamber of the FID analyser is normally
about 200C.
Without going into details about the individual techniques, the following points are of
importance:
> Multi-component analysers are less expensive for simple measuring tasks with
more than one component than individual analysers in all respects.
> When using alternative fuels, the necessity may arise that new, particularly critical
pollutant components have to be measured. This aspect justifies the use of
flexible, programmable measuring instruments.
8.6.1 Sick GM 31
-54-
Sick, Waldkirch (D), is a pioneer of the design and construction of in-situ measuring
systems. The instruments GM 21 for dust, and later, the GM 30 for NO SO2 and
dust, have been used successfully for more than ten years. The successor
instrument, GM 31, differs considerably from its predecessors in two ways:
The flanged design as a so-called lance instrument does not measure the whole
channel cross-section.
The GM 31 is designed for the components SO2, NO, NH3 and NO. A maximum of
three of these components can also be measured simultaneously. Despite the lack
of the benefit of dust measurement, the GM 31 is a low-cost emission measuring
instrument, which complemented with a dust measuring instrument for systems not
subject to particular requirements, meets all specific requirements.
Within the Opsis measuring system, a light beam is projected on to a receiver and
then passed through a glass fibre cable to the Opsis analyser.
In emission measurement, the light beam or measuring path continues through the
interior of a stack or exhaust gas duct. Every gas absorbs light in known spectral
regions of the total light spectrum, in a manner that is characteristic of the respective
gas. This enables the analyser to detect and measure gases defined by the user.
The results can be indicated in real-time or used for statistical and further processing
purposes.
As every gas has its own unmistakable absorption spectrum, the so-called finger
print, the concentrations of several different gases can be determined
simultaneously in the light path. With the DOAS method, a light beam is sent from a
special light source - a Xenon high-pressure lamp - over a specific path; by means of
complex, computer-aided calculations, the light losses through molecular absorption
along the path are evaluated and analysed. The light of the Xenon lamp contains
both the visible spectrum as well as the ultra-violet and infrared wavelengths. The
light is detected by a receiver and passed on via a fibre optic cable to the analyser.
This fibre optic cable makes it possible to install the analyser at a sufficient distance
from any harmful environmental influences at the measuring point.
The analyser consists, among other things, of a high-performance spectrometer, a
computer and the associated control unit. The spectrometer breaks down the light
with the aid of an optical grid in narrow wavelength bands. This optical grid can be
adjusted for examining an optimal wavelength region.
The light is converted into electrical signals. A narrow slot moves at high speed
across the detector; this combines a large number of instantaneous values that
provide an image of the spectrum in the relevant wavelength region This scanning
procedure is repeated a hundred times per second; the recorded spectrums are
added up in the multi-channel memory of the evaluation unit until their evaluation.
Evaluation is carried out individually for each wavelength region and is based on the
comparison of absorption curves. The respectively last recorded absorption
spectrum is compared with a computer calculated spectrum. The calculated
spectrum consists of a summation of the reference spectrums for the respective
evaluation. The computer alters the size factors for each reference spectrum until
optimal correspondence is achieved, so that the various gas concentrations can be
calculated with high accuracy.
For calibration: In principle, each component must be calibrated at last once (as part
of the commissioning procedure). This takes place either via a convention method or
by inserting cuvettes circulated by test gas in the light path. In this simple way, the
work involved in the normally usual convention method is avoided. In two works of
the ChB, these measuring instruments have proved highly reliable and may be
referred to as standard-setting technology.
The LDS 300 can be used for measurements in atmospheres with a varying dust
content up to 50g/m3, depending on the size of the dust particles.
The light spectrum of the semiconductor laser can be adjusted to the absorption lines
of the gas component to be measured via temperature and current. This ensures
that measurement can only take place on a selected absorption line of the respective
gas. The absorption lines of the remaining gas components are on other
wavelengths and do not influence measurement.
In the three years in which the LDS has been used, continuous tests were carried out
in order to determine the stability of the system. It emerged that no recalibration of
either the zero or measuring range were necessary.
The diode laser is located in the main unit, from where the laser beam is passed via
an optical fibre to the sensors at the measuring points. The distance between the
instrument and the measuring point can be up to 1000m. In the main unit, the laser
beam is scattered in an optical distributor and so enables simultaneous
measurements on three different process measuring levels.
The short measuring time of the LDS 3000 provides direct information (t90<1 sec).
This is made possible by in-situ measurement, the high efficiency of the diode laser
and appropriate signal evaluation. The short resonance time depends on the
number of measuring points, as the laser light is scattered in a fibre-optic distributor
and passed on to all measuring points simultaneously. Each measuring point can
therefore function independently of the other measuring channels.
Thanks to its unusually short reaction time, this measuring system was to be able to
considerably improve the CO monitoring of electrostatic filters. Measurement directly
following WT would perhaps be possible, since the manufacturer mentions that up to
50g/m3 dust has no influence. Unfortunately, the measuring distance this value
applies to is unavailable.
The LDS 300 is one of the first laser analysers to appear on the market. At present,
the price is (still) relatively high. Development is absolutely in line with the trend and
could also be used advantageously in the cement industry in the near future.
-57-
A German cement works has had good experience over a two-year operating period.
The instrument is overhauled twice a year as part of a maintenance contract. The
weak point, the hot gas pump, is changed each time or the diaphragms replaced as a
precautionary measure.
-58-
Following the measured values on the basis of recording strip charts is advantageous
for the general assessment of a process sequence. In the area of emissions, the
majority of countries require that measured values are prepared according to certain,
adapted criteria; hourly and daily average value, criteria relating to the observance or
non-observance of limiting values, etc. In such cases, digital processing of the
measured values is unavoidable.
For the purpose of clarity, not too many components should be recorded
simultaneously. Strongly varying signals must be appropriately dampened. Monitors
or line recorders are better suited for this purpose than dotted-line recorders. Scaling
should correspond with the actual physical values. This applies particularly to life-
zero signals. For strip chart recorders, a paper feed rate of 20mm/h is standard.
When lump fuel (e.g. tyres) is fed into the precalcining zone, the measured values at
the kiln inlet are subject to extreme fluctuations. Without damping of averaging, the
signals are difficult to interpret. It has emerged that sliding, linear averaging is easier
to interpret than logarithmic damping, as individual values are less important in linear
averaging. The more favourable intervals for optimal averaging are between 10 and
15 minutes. Figures 13 and 14 show the behaviour of the signals of a 1500-7/d kiln
with tyre feed without averaging and with various averaging times.
-59-
8
7
6
5
4
3
2
1
0
5
4.5
4
3.5
3
2.5
2
1.5
1
0.5
0
5
4.5
4
3.5
3
2.5
2
1.5
1
0.5
0
-60-
7000
6000
5000
4000
3000
2000
1000
5000
4500
4000
3500
3000
2500
2000
1500
1000
500
0
5000
4500
4000
3500
3000
2500
2000
1500
1000
500
0
-61-
Condition check
Adjustment check, cleaning
Replacement of wearing parts
Function test
-63-
Function test
Visual checks
Probes:
Tightness check QY
Cooling D QY
Filter change HY
Cleaning HY
Analyzers:
Indication plausibility D
Adjustment W
Filter change QY
Total overhaul 3Y
The operating conditions and type of construction of the equipment used varies
considerably. Instructions for maintenance and repairs must be adapted to local
conditions and defined according to individual experience.
The term visual check includes visual inspection of the equipment, indicating
instruments, signal states and checking heated parts for a hot or cold condition by
trained and experienced personnel.
Vacuum check
The hose connections at the probe are sealed and the sample gas pump is
started. The maximum vacuum which the pump can produce must be noted when
commissioning the measuring equipment. At maximum vacuum, the part of the
pump is closed and the pump stopped. The sample gas preparation system up to
the sample gas pump (suction area) is considered tight when the vacuum does
not change significantly for a period of 5 minutes following temperature
equalisation.
Nitrogen check
The sample gas pumps are usually diaphragm pumps, whose diaphragm and
valve plates should be replaced at least once a year.
The ball bearings of the motor and eccentric should be replaced every two years
in continuous operation. Standby sample gas pumps should be available.
-65-
Ceramic filters can be cleaned with diluted hydrochloric acid (1 part concentrated
hydrochloric acid to 10 parts water). The filters must be placed in the acid for
about 2 hours.
The parts must subsequently be flushed with water pressure from the inside to the
outside, blown out with compressed air and dried in a drying cabinet at about
100C.
The filter element should be replaced at regular intervals, e.g. every three months.
Fine filters with condensate monitor are tested by moistening the filter element.
The instrument must then respond.
The gas analysers are tested with test gases or by means of the built-in calibrating
cuvettes and readjusted if necessary. For testing with gases, it must be ensured
that the flow rate roughly corresponds with the operating values. When selecting
the test gases, the specifications of the instrument manufacturer must be
observed. The gas analysers should be adjusted at regular intervals, at least
every two weeks. Adjustments must be noted in the maintenance report. If
significant adjustments are required, or the adjustment range limits (greater than
80% or less than 20%) are reached, the instrument must be tested and inspected
if necessary.
For zero readjustment, nitrogen is normally used. With the exception of oxygen
measurement, air can also be used as a zero gas if the analysers are not too
sensitive.
For sensitivity adjustment, a test gas is used that normally consists of the
measuring components and nitrogen. For compensation of the residual moisture,
it is advantageous to feed the test gas upstream of the sample gas cooler. The
test gas concentration for sensitivity adjustment should be 80 to 90% of the
respective measuring range end value.
This term means that certain functions gas analysis are extensively tested, e.g. to
ensure that the detector in the fine filter responds if moisture is present and the
sample gas pump is stopped.
-66-
As initially mentioned, new gas analysers no longer require any test gases. The
internal calibrating and control options ensure that the accuracy is maintained over a
long period (normally two years). After this time, the analysers should be tested by
the manufacturers service department and in particular the internal control devices
(calibrating cuvettes) checked. Test gases are available in various accuracy classes
according to particular directives, e.g. VDI 3490, sheet 2. Depending on the
manufacturing and analysis accuracy, distinction is made between three classes.
The test gases are delivered with analysis certificates that contain all essential data,
such as:
Minimum and maximum storage temperature and maximum test gas storage time
(limit date)
The concentration specified on the analysis certificate should be written in large and
easily readable letters on the test gas bottle.
The test gases are available in bottles of varying size. A 10 litre bottle (about
1100mm high, preferably of aluminium) is normally sufficient for a two-year supply.
The bottle pressure is between 100 and 200 bar, depending on the gas composition.
The main valve must be closed after each use. Test gas deliveries may take time in
certain regions; this should be taken into account when the bottle pressure starts to
get low.
With one exception, all standard components can be mixed together. This
exception applies to oxygen O2.
With oxygen analysers, the measuring range should be such that air can be used
for calibration. Nitrogen must always be used as a residual gas.
The accuracy of the test gases reduces the more components are mixed. The
number of mixing components is thus limited.
Insufficient pressure in the suction pipe (blockage of the probe or gas path)
Excessive pressure in the suction pipe (possible failure or leakage of a solenoid
valve)
Status signals of the analysers (failure of the analyser or electrical fault)
Operating temperature of the sample gas cooler or failure of the sample gas
cooler
Operation of the condensate monitor (can indicate failure of the sample gas cooler
or faulty water-cooled probe
Adjustment/operation changeover valve should automatically inform the control
room that the measuring equipment is being serviced and is therefore not ready
for operation
Minimum contact of flowmeter (insufficient sample gas)
Status signals for measuring equipment faults are checked for their plausibility by
simulated operating deviations.
They have a corrosive effect, contain dust, must not be adulterated, can be very
moist, etc., to mention but a few of the unpleasant characteristics that make
automation difficult.
Practice shows time and again that solenoid valves are the weak point within
measuring equipment, so that only valves of the best quality must be used for
evaluation. The number of necessary solenoid valves increases with an increasing
degree of automation; due to their susceptibility to faults, the failure rate also
increases, so that the savings expected from automation are accordingly lost.
The same applies when a group of analysers is switched to two or more measuring
points with the intention of saving on expensive analysers. Apart from the operating
difficulties, experience has shown time and again that the saved investment costs
are quickly absorbed by the corresponding maintenance costs.
-68-
The accuracy of gas analysers is often overestimated. The reason for gas analyses
being relatively inaccurate is not because of the analysers, as is often assumed. The
following examples explain the most important influences responsible for this.
In the probe and sample gas preparation system, sample gas components can
react with each other. These reactions are reinforced by catalytic effects.
Furthermore, reactions with dust or gas conducting components are possible.
11.2.2 CO reduction
11.2.3 NO2 formation in the probe and sample gas preparation system
At lower temperatures and high gas residence times, this formation takes place
more rapidly. The indication delay of the NO measuring equipment should
therefore be as short as possible.
Rust, dust deposits, non-ferrous metal parts (especially copper) in the gas path of
the sample gas can reduce SO2 up to 100%. For the sample gas pipes and all
other gas conducting parts, Teflon, high-alloyed steel or steel with special coating
(e.g. PTFE) must therefore be used.
-69-
SO2 reacts with the CaCO3 and CaO contained in the filter dust of the sampling
probe with the formation of CaSO4. With high dust contents in the process gas,
measuring errors of up to 100% can occur. SO2 can therefore only be measured
fairly accurately process gas with a low dust content. This applies in particular to
SO2 measurements in the rotary kiln inlet. Tests have shown that SO2 is initially
indicated only a short time after cleaning the filter. Within several minutes, the
indication falls to zero again, because the SO2 is absorbed in the dust building up
in the filter.
Condensation of the water vapour which is separated as water in the sample gas
cooler by the extractive method, can also distort gas analysis. This is of particular
significance for SO2 emission measurement when low concentrations are present.
The theoretical solubility of SO2 in water is low and could be compensated by
appropriate calibration. However, tests on cement kiln systems in a direct
operating mode have shown that measuring errors for SO2 are far above the
theoretical distribution equilibrium when SO2 reacts with other gas components
such as NH3. An SO2 measuring device can be tested for these errors by the
sample gas being transferred from the sampling probe directly into concentrated
sulphuric acid (H2SO4) as the drying medium and not via the sample gas cooler. If
distinct differences in the measured value indication occur, a chemical reaction is
most probably taking place in the sample gas cooler.
Apart from SO2, NO2 also dissolves in the condensate of the sample gas cooler.
However, since the proportion of NO2 in the total NO concentration can be
disregarded with proper sampling, the resulting measuring error can normally be
disregarded.
The accuracy of the gas analysers, provided they are in a perfect condition, is not an
issue. If a gas analyser is viewed as a whole, the analyser is the most accurate link
in the measuring chain, with a few exceptions. The analyser manufacturers
-70-
If, for any reason, combustible gases should develop (ring fractures, faults in fuel
dosage, etc.), in the kiln system, there is a risk of a potentially explosive mixture
forming in the electrostatic filter. High-voltage discharges inside the filter cause
sparks, which can lead to explosions in a gas atmosphere of appropriate
composition.
As a safety measure, the CO component of the gas discharging from the kiln system
is measured. Once a certain level is reached, the high-voltage should be switched
off before the gases reach the electrostatic filter. Measurement can initially take
place in precalcining plants following complete combustion in the auxiliary firing
equipment, i.e. in the uppermost cyclone stage of the heat exchanger at the earliest.
The residence time of the gases from this point to the electrostatic filter is relatively
short (several seconds) and varies considerably from plant to plant (with or without
cooling tower, combined/direct operation, etc.).
Compared with the usual method, filter shutdown is optimised in two essential
steps. The first step consists of shutting down the filter chambers step by step.
The second step consists of taking into account a reducing or increasing CO
trend.
The residence time of the gases in the electrostatic filter itself is used in order to
further delay shutdown. The filter chambers are shutdown at approximately five
second intervals, depending on the size and gas flow rate. However, as soon as
CO measurement indicates a reducing trend, the shutdown cycle is stopped.
Large sample gas volume (up to 300l/h setting; short delay time)
12.1.3 Specifications
With regard to the CO analyser, it should be noted that the same must have a
high measuring range for CO monitoring. It cannot be used simultaneously for
continuous CO observation, as the normal CO content is below the detection limit
of this analyser. If CO is to be continuously measured, a second analyser with a
much lower measuring range of 5000ppm (0.5%) must be used.
The procedure for shutdown of the individual filter chambers is shown on the basis
of two examples in Tables 5 and 6. The specified numerical values of the
shutdown sequences and the CO level are only guide values and must be
adjusted to the respective plants.
For CO values below 10%, a risk of explosion is virtually excluded. However, the
threshold values must be set far lower, as the measurement reacts with delays
despite optimal preconditions (dead time + indication delay).
-73-
X 0 0 0 +- 0 1 1 1 1
X 1 0 0 +- 1 1 1 1 1
5 1 1 0 + 2 0 1 1 1
10 1 1 0 - 1 1 1 1 1
10 1 1 0 + 3 0 0 1 1
10 1 1 1 + 5 0 0 0 0
15 1 1 0 - 1 1 1 1 1*
15 1 1 0 + 4 0 0 0 1
15 1 1 1 + 5 0 0 0 0
Table 5 Switching procedure, for a simple heat exchanger kiln with 3-chamber filter
X 0 0 0 0 0 0 +- +- 0 1 1 1 1
X 1 0 0 0 0 0 +- +- 1 1 1 1 1
X 1 0 0 1 0 0 +- +- 1 1 1 1 1
X 1 1 0 1 0 0 +- +- 1 1 1 1 1
5 1 1 0 1 1 0 + + 2 0 1 1 1
5 1 1 0 1 1 0 - - 1 1 1 1 1
5 1 1 1 1 1 1 + - 2 0 1 1 1
5 1 1 1 1 1 0 + + 3 0 0 1 1
10 1 1 0 1 1 0 - - 1 1 1 1 1
10 1 1 0 1 1 0 + + 3 0 0 1 1
10 1 1 1 1 1 1 + + 5 0 0 0 0
10 1 1 0 1 1 0 - - 1 1 1 1 1*
15 1 1 0 1 1 0 + + 4 0 0 0 1
15 1 1 0 1 1 0 - - 1 1 1 1 1
15 1 1 1 1 1 1 + + 5 0 0 0 0
Table 6 Switch-off pocedure, for a two strings heat exchanger kiln with 3-chamber filter
-74-
Sick GM 950
Via an in-situ sensor (zirconium dioxide) CO is measured at a high response rate
and processed to a shutdown command if required. As zirconium is fairly instable
as a CO indicator, extractive supplied semiconductor sensors correct any
deviations.
The device is new and still in the testing phase. The currently available test
results have been positive up to now. The measuring method is relatively
elaborate and the device gives the impression of being somewhat complex.
During the test, no functional disturbance occurred.
In the safety concept of coal pulverising mills, the monitoring of critical gas
components covers only a small part of the necessary equipment and measures. In
addition, coal pulverising mills vary considerably in their design, so that the
application and function of gas analysis is always determined as part of the entire
plant.
In 1984, the following limiting values were listed by a working group of the
Association of German Cement Works (VDZ):
O2 content in moist exhaust gas downstream of the filter 14%
Temperature at mill inlet (coal moisture 12%) 400C
Gas temperature downstream of mill upstream of filter 120C
Gas temperature downstream of filter 120C
Gas temperature upstream of filter minimum 30C above dew point about 75C
CO concentrate during filter shutdown 50ppm
Operation with hot gas generator 70ppm
Operation with kiln exhaust gase not practical
Pressure upstream of mill -1.5bar
Coal dust temperature 110C
Air temperature of pneumatic transport 80C
-75-
The limiting value for the oxygen concentration applies only to plants with inert
gas operation. All other limiting values are the same for the safe operation of inert
and non-inert gas operated plants.
For coal pulverising mills heated with kiln exhaust gas, CO monitoring is not
practical during operation, because the gases from kiln operation contain up to 10
times more CO than the limiting value of 70ppm permits.
CO shutdowns of the electrostatic filter for the kiln, must be coupled with the filter
of the coal pulverising mill.
This information on the safety equipment of coal pulverising mills only contains the
most important information and is intended to draw attention to the complexity and
importance of such plants.
Conversion of the gas components from volume to units of weight and vice versa, is
often the cause of errors in adjustment or interpretation. Such errors occur most
frequently in connection with NO and NO2. The problem with these NO components
is that the instruments measure NO, but the emission must be converted to NO2.
Furthermore, emissions are considered in weight-related units (mg/m3), but the
instruments and the test gases are often specified in volume-related units (e.g. ppm
or %)
Table 7
Example
Given:
Required:
Solution:
Process Technology
H. Nyffenegger
-77-
14. LITERATURE
B. Thier: Safe Operation of Coal Pulverising Mills. ZKG 4/1984, page 163
H.G. Loos, Erlangen: A New Operational Compact Gas Sampling Device for Cement
Rotary Kilns. ZKG NO. 6/1987
M. Birrer, H. Nyffenegger: Gas Sampling at the Kiln Inlet with Improved Probe.
Holderbank NEWS 6/90
Bronkhurst High-Tech B.V., Ruurlo, Holland: Operating Manual for Mass Flowmeters
and Controllers for Gases
KNF Neuberger (Switzerland) AG, CH 8362 Balterswil: Prospectus and Data Sheet
for Diaphragm Compressors and Vacuum Pumps