Professional Documents
Culture Documents
LOT 5
Mobile Multi-functioning Power Monitoring Meter
a) Voltage inputs:
Class A performance over all of the input range inclusive of 63.5 V (if LN
measurements are done in a 110 V line-line VT circuit)
6 terminals for 4 voltage input channels with maximum input ratings of
600 V.
2. Current inputs:
4 X 1/5 A CT inputs
4 X 1 Vrms differential voltage transducer inputs with a 5 V power supply
(to be used to power Rogowski coils).
c) Power supply
90-300 V AC/DC power factor corrected power supply.
1/2 hour power failure ride-through via an internal LiFePO4 battery.
d) Data storage
4 GB on-board storage providing data storage of all IEC 61000-4-30
voltage and current 10-minute values up to the 63rd harmonic component.
e) Time stamping
GPS time synchronization resulting in a time-stamping certainty better
than 100 ns.
PTP time synchronization resulting in a time-stamping certainty better
than 1 ms.
NTP time synchronization as an option.
f) Communication
Communication shall be via Ethernet.
Modbus over Ethernet.
DNP3 over Ethernet.
IEC 61850
g) Advanced triggering
User-configurable triggering between different instruments
5.1.2.1 Voltages
The voltage inputs to the instrument shall interface directly to 110 V VT circuits or
directly to a mains connection, which has to be less than 600 V line to neutral.
Voltage measurements shall be differential.
6 input terminals for voltages shall be used.
The first 4 terminals shall be used for 3 voltage inputs V1, V2 and V3 and
are neutral (Vcom) referenced (V1-N, V2-N, V3-N).
5.1.2.2 Currents
Permanently installed 1 A or 5 A CTs, or clamp-on CTs interface directly to the
galvanically isolated current inputs.
The instrument shall calculate the missing current by assuming the sum of all
currents is zero.
It shall also be powered from an IEEE 802.3-2008 compliant PoE Plus (< 12 W) source.
The internal LiFePO4 battery shall power the instrument in the absence of power. The
battery shall be capable of 2,000 charge/discharge cycles.
5.3 Communication
Two (2) Gigabyte Ethernet ports shall facilitate local and networked external
communication. One port shall be used to interface with corporate networks (DHCP or
fixed IP support). The second Ethernet port shall be used to create a local network by
means of being a local DHCP server.
The remote support software shall support automatic service discovery protocols.
Laptops, when connected to the instrument, shall receive an IP address from the
instrument. The remote support software should be in a position to automatically
discover which instruments are connected to the local network.
Waveform and other events shall be triggered from the 500 kHz digitised waveforms.
These waveforms shall be locked to the fundamental frequency to prevent multiple zero
crosses.
Captured waveform data shall be further subsampled to reduce stored data size. The user
should be able to configure the waveform-sampling rate from 1 kHz to 50 kHz.
Upper and/or lower thresholds shall define an event and can be set by the user based on
waveform
Single-phase and multiple-phase triggers shall be in such away that they can be defined.
When a threshold is exceeded, the instrument shall retain pre- and post-diagnostic
information of the event.
VOLTAGE INPUTS
Number of channels 4 x differential inputs (3/4 Wire + 4th Diff)
Measurement input range 0 600V rms
Voltage measurement Single Phase, 3 phase (Star, Delta), DC
Input impedance/cannel 1 M
CURRENT INPUT
Number of channels 4 x galvanically isolated inputs
Measurement input range 0 5 Arms
Maximum continuous cureent 6.0 Arms
3 sec over-current withstand 50
VA Budern at 5 1 VA
Galvanic isolation 1,000V
CURRENT TRANSDUCER
INPUT
Number of channels 4 diffrential inputs
Measurement input range 0 1.0 Vrms (1.414 V peak)
Input impedance/cannel 1 M
DIGITAL INPUTS
Number of channels 2 sets of 2 galvanically isolated digital outputs
Switch rating 0.1 Arms, 300 Vrms (Max)
ACCURACY & BANDWIDTH
Power frequency range DC, 50Hz (40 60 Hz), 60 Hz (50 70 Hz)
Harmonic bandwith 1-64th (harmonic and inter-harmonic)
Measurement sampling rate 0.5 MHz (simultaneously sampled)
Waveform storage rate 1 KHz (user configured)
ADC resolution 16-Bt
Overall accuracy class 0.1%
Fast Transient capturing 20s
COMMUNICATION
Ethernet 2 x Gigabit Ethernet IEEE 802.3.2008
compliant
USB USB 2.0 480 Mbit (powered)
CLOCKS
POWER
Maximum power consmption 15W
AC/DC supply voltage input 90 300 Vrms 42 69Hz (power factor
corrected)
Power over Ethernet IEEE 802.3 compliant (35 60 Vdc)
On-board battery LifePO4 (2,000 charge/discharge cycles)
PHYSICAL
Construction (mnimum) Aluminium, 250 x 135 x 65 (L x W x H)
Mounting options 3U-19 Rack, Panel mount, Portable case
Electrical connections Plugged screw type (screw lock-down)
Electrical isolation class 600 V Class II
Operating temperatura 0c to 75c
COMPLIANCE
PQ Measurement IEC 61000-4-30, Class A
Harmonics IEC 61000-4-7
Flicker IEC 61000-4-15
SCADA IEC 61850
LOT 6
VARIOUS TEST EQUIPMENT
The battery discharge test equipment shall be designed for the test of battery benches
using the discharge method. The testing device shall be capable of discharging the
battery in different modes, measuring the capacity of new and used battery.
The test equipment shall allow the connection to auxiliary modules to enhance the
discharging capability. With the optional current clamp, or with the external shunt, it
should also be possible to use the real burden as the additional burden, and to control
the total discharging current in the selected mode.
The test set shall be microprocessor based in order to allow the control of all the
discharging parameters in real time. It shall also allow communication with a PC for the
control and the data representation of the discharging parameters.
The test set shall have the following features:
Two wheels for ease movement.
The number of discharging steps programmed from 0 to 20 for ease of
drawing any discharge profile.
In addition to the external clamp, the external burden current can now be
measured also with a shunt.
If a test is interrupted, for instance because of a faulty contact or cell, after the
repair it is possible to continue the test: this was not possible with the former
device.
The test set firmware can be upgraded with a supplied software.
The discharge actual parameters shall be displayed on the graphic screen and all the
measured parameters saved into the selected memory.
The device shall be equipped with an emergency pushbutton, located on the front.
When pressed, it should stop all the test activities. The pushbutton shall be
mechanically self-locking: the operator has to rotate it in order to reset the normal
operation.
The device shall be powered by a power supply plug for a wide range AC supply or
by the battery under test itself, for batteries with a nominal voltage of 110 V 125V
Table below provides specifications for the hardware, firmware, TDMS software and
optional accessories required.
A) Hardware
24 V range limits 20 to 34 V DC
48 V range limits 34 to 60 V DC
72 V range limits 60 to 84 V DC
External clip-on CT conversion factor range 1 to 500 mV. NOTE: the 1 mV range is to
(programmable) be used for currents greater than 400 A.
PARAMETER VALUE
B) Firmware
Number of memories 8
Test set control mode Via encoder + push-button and menu on the
display
C) TDMS Software
Results and settings data base Integrated in the TDMS Substation data base
D) Options
Transport case, with wheels Dimensions: 330 x 900 x 600 mm; weight 16 kg.
The tester is used to be used to check the efficiency of the metal oxide surge arrester
in service that are installed on the high voltage transmission and distribution
networks.
The equipqment shall have to perform the tests according to the IEC standard 60099-5
A1 ED.
The equipment shall be capable to measure leakage current while the surge arrester is
in service. The leakage current shall be analyzed by means of a special current clip-on
transformer in the surge arrester ground connection. The values of the current may
normally range from fractions of milliampere to a few milliampere.
6.2.5 The accuracy of the tester shall be within specified limits below:
50 Hz measure, rms and peak value: total maximum of 5% from 0.1 to 10
mA.
150 Hz rms measure: total maximum of 10% from 10 to 1000 A.
6.2.9 The equipment shall have the test probe for measuring the third harmonic of the system
voltage, 2 m long, with banana plugs for the connection to the PT secondary and
connector for the connection to the tester.
Nominal voltage: 100 V phase to phase (57.8 V phase to neutral).
Maximum voltage: 200 V phase to phase.
The cable shall incorporate a resistor divider made of two resistors,
respectively 56.8 k and 1 k maximum, so that the output to the tester is 1 V
when the input voltage is 57.8 V.
Carry case:
A plastic carry case holds: the clip-on transformer, the measuring instrument,
the instruction manual, the short-circuit cable.
Carry Case dimensions: Enough to hold above items.
Weight with all the components: Not more than 3 kg.
6.2.11 the tester shall be supplied together with the following accessories:
Instruction manual.
Measuring probe for the third harmonic of the system voltage.
Short-circuit cable to carry out the calibration test.
Plastic carry case.
Electromagnetic compatibility
Directive 2004/108/EC. Applicable Standard: EN61326:2006.
Low voltage:
Directive 2006/95/EC (CE conform). Applicable standard, for a class I instrument,
pollution degree 2, installation category II: CEI EN 61010-1.
6.3.1 GENERAL
The portable, high accuracy test set shall perform all tests foreseen by international
Standards on
CTs, measurement VTs, Power Transformers, and measure the Tan (), dissipation
factor and capacitance of any device.
6.3.3 CHARACTERISTICS
The test set shall incorporate a generator with six outputs. When an output is
generated, there shall be voltage also on all other outputs, unless for the high AC
voltage.
6.3.7 Timer
The test set shall allow to measure a device timing, using a special input. At the test
start with voltage or current generation, a timer shall be started and should stop when
the relay trips. It shall also be possible to stop the timer when the injected current is
stopped by the intervention of the device under test.
6.3.9 Display
The display shall have the following main features:
Pixel Light LCD type View area
640 x 480, colors Backlight TFT 132 x 99 mm
The menu shall be entered by pressing the knob and selecting the item by moving the
knob.
Note: Total cost on practical training should be summed up under this lot.
Where Factory Acceptance Tests (FAT) have been specified the following will be the arrangement. In
the event of a contract being signed, three representatives consisting of two engineers and a
procurement specialist will witness the factory tests. The cost of these representatives in terms of
travel allowances, accommodation and airfare will be met by the supplier. The rate of allowances
received by ESCOMs Officers is USD300.00 per day per officer from date of departure to date of
return.
Lot 1 and Lot 4 shall have Factory Acceptance Tests.