Professional Documents
Culture Documents
CALCULATED HEMOGLOBIN/HB
Hematocrit is determined conductometrically. The measured conductivity, after correction for electrolyte
concentration, is inversely related to the hematocrit.
See below for information on factors affecting results. Certain substances, such as drugs, may affect analyte
levels in vivo.1
If results appear inconsistent with the clinical assessment, the patient sample should be retested using
another cartridge.
Intended Use
The test for hematocrit, as part of the i-STAT System, is intended for use in the in vitro quantification of
packed red blood cell volume fraction in arterial, venous, or capillary whole blood.
Hematocrit measurements can aid in the determination and monitoring of normal or abnormal total red
cell volume status including, but not limited to, conditions such as anemia, erythrocytosis, and blood loss
related to trauma and surgery.
Contents
Each i-STAT cartridge contains one reference electrode (when potentiometric sensors are included in the
cartridge configuration), sensors for the measurement of specific analytes, and a buffered aqueous calibrant
solution of known conductance that contains known concentrations of analytes and preservatives.
Metrological Traceability
The iSTAT System test for hematocrit measures packed red blood cell volume fraction in arterial, venous,
or capillary whole blood (expressed as the % packed cell volume) for in vitro diagnostic use. Hematocrit
values assigned to iSTATs working calibrators are traceable to the Clinical and Laboratory Standards
Institute (CLSI) H7-A3 procedure for determining packed cell volume by the microhematocrit method.2
Further information regarding metrological traceability is available from Abbott Point of Care Inc.
Expected Values
Reportable Reference
Test/Abbreviation Units* Range Range3
Hematocrit/Hct %PCV 10 75 38 51**
Fraction 0.10 0.75 0.38 0.51
Hemoglobin/Hb g/dL 3.4 25.5 12 17
g/L 34 255 120 170
mmol/L 2.1 15.8 7 11
* The i-STAT System can be configured with the preferred units.
**The reference ranges for hematocrit and hemoglobin span both female and male populations.
Clinical Significance
Hematocrit is a measurement of the fractional volume of red blood cells. This is a key indicator of the
bodys state of hydration, anemia or severe blood loss, as well as the bloods ability to transport oxygen.
A decreased hematocrit can be due to either overhydration, which increases the plasma volume, or a
decrease in the number of red blood cells caused by anemias or blood loss. An increased hematocrit can be
due to loss of fluids, such as in dehydration, diuretic therapy, and burns, or an increase in red blood cells,
such as in cardiovascular and renal disorders, polycythemia vera, and impaired ventilation.
Performance Characteristics
The typical performance data summarized below were collected in health care facilities by health care
professionals trained in the use of the i-STAT System and comparative methods.
Precision data were collected in multiple sites as follows: Duplicates of each control fluid were tested in the
morning and in the afternoon on five days for a total of 20 replicates. The averaged statistics are presented
below.
Method comparison data were collected using CLSI guideline EP9-A4. Venous blood samples, collected in
lithium heparin Vacutainer tubes, were analyzed in duplicate on the i-STAT System and on the comparative
methods for hematocrit within 20 minutes of collection.
Deming regression analysis5 was performed on the first replicate of each sample. In the method comparison
table, n is the number of specimens in the data set, Sxx and Syy refer to estimates of imprecision based
on the duplicates of the comparative and the i-STAT methods respectively, Sy.x is the standard error of the
estimate, and r is the correlation coefficient.*
Method comparisons will vary from site to site due to differences in sample handling, comparative method
calibration and other site specific variables.
Interference studies were based on CLSI guideline EP7-P.6
*The usual warning relating to the use of regression analysis is summarized here as a reminder: For any analyte, if the data is collected over a narrow
range, the estimate of the regression parameters are relatively imprecise and may be biased. Therefore, predictions made from these estimates may be
invalid.5 The correlation coefficient, r, can be used as a guide to assess the adequacy of the comparative method range in overcoming this problem.
As a guide, the range of data can be considered adequate if r>0.975.
Interferent Effect
WBC Grossly elevated white blood cell counts may increase results.
Total Protein Hematocrit results are affected by the level of total protein as follows:
Displayed Result TP < 6.5 g/dL TP > 8.0 g/dL
HCT < 40 %PCV Hct decreased by ~1% PCV Hct increased by ~1% PCV
for each decrease of for each increase
1 g/dL TP 1 g/dL TP
HCT > 40 % PCV Hct decreased by ~0.75 % PCV Hct increased by ~0.75 %PCV
for each decrease of for each increase
1 g/dL TP 1 g/dL TP
Total protein levels may be low in neonatal and burn patient populations, as well
as in additional clinical populations listed in Statland.3 Total protein levels may also
be decreased in patients undergoing cardiopulmonary bypass (CPB) or ECMO,
and with patients receiving large volumes of saline-based IV fluids. Care should be
taken when using hematocrit results from patients with total protein levels below
the adult reference range (6.5 to 8 g/dL).
The CPB sample type can be used to correct the hematocrit result for the dilutional
affect of the pump prime in cardiovascular surgery. The CPB algorithm assumes
that cells and plasma are diluted equally and that the pump priming solution has
no added albumin or other colloid or packed red blood cells. Since perfusion
practices vary, it is recommended that each practice verify the use of the CPB
sample type and the length of time in which the CPB sample type should be used
during the recovery period. Note that for hematocrit values above 30 %PCV, the
CPB correction is 1.5 %PCV; the size of the correction at this level should not
impact transfusion decisions.
Lipids Abnormally high lipids may increase results. Interference from lipids will be about
two-thirds the size of the interference from protein.
Cartridge Comparison
The performance characteristics of the sensors are equivalent in all cartridge configurations. System
difference analysis was performed on 40 patient samples using the i-STAT 6+ and i-STAT E3+ cartridges.
In the 1530 %PCV range the average difference was 0.462. In the 3050 %PCV range the average
difference was 0.097.
i-STAT is a registered trademark of the Abbott Group of Companies in various jurisdictions. Vacutainer is a registered trademark of Becton Dickinson
and Company, Franklin Lakes, NJ USA. Coulter S Plus is a registered trademark of Beckman Coulter Incorporated, Fullerton, CA USA. Cell-Dyn is a
registered trademark of Abbott Laboratories, Abbott Park, IL USA. SE9500 is a trademark of Sysmex America Inc., Mundelein, IL USA. STAT Profile
is a registered trademark of Nova Biomedical, Waltham, MA USA.
2014 Abbott Point of Care Inc. All rights reserved. Printed in USA.