Professional Documents
Culture Documents
1Page
Lab report1
Subject :
Physical chemistry
Submitted to:
Mam Afia
Submitted by:
Shaherbano
Roll no:
(58)
Apparatus:
Beaker
Volumetric flask
Pipette
Burette
Measuring cylinder
Procedure:
1. Burette:
Using temperature-equilibrated water, fill the burette past the 0 mL mark (be sure to record the
3
temperature), insuring that no air bubbles are trapped in the tip. Give the water about a minute to
Page
drain, and then allow the water to drain to the 0.00 mL mark in your waste beaker. Touch the side
of the tip to the beaker to remove any water sticking to the tip. Replace the waste beaker with the
weighed 125 mL Erlenmeyer flask, (analytical balance) and slowly transfer 10 mL of water to
the Erlenmeyer flask (this should take about 1 minute.) Reweigh the flask (again using the
analytical balance) and calculate the volume delivered as you did for the pipette. Repeat this
procedure at least 2 more times, or until your delivered volumes are 0.02 mL. Once you get an
appropriate agreement between the 10 mL values, perform the same procedure for 20, 30, 40,
and 50 mL. Remember to allow the burette to drain at about 10 mL/min for all of these
determinations!
2. Flask:
Weigh empty, dry flask. Fill it with distilled water to mark. Weigh again. Mass difference is the
mass of water. Repeat 3-4 times.
3. Pipette:
Weigh empty, dry, closed weighing bottle. Pipette water into the bottle, close, weigh again. Mass
difference is the mass of water. Repeat 3-4 times.
Quite often single volume pipette and volumetric flask are used together - diluted sample is
pipetted from the flask to carry on separate titrations. To calculate sample size we have to
multiply titration result by the ratio of the flask volume and pipette volume. Instead of using each
time volumes of the pipette and flask it is easier and faster to use so called commensurability of
the flask and pipette - just divide flask volume by the pipette volume and use this number to
calculate original sample size.
For Flask;
4 Page
=23.81+24.25+24.31/3
=23.90ml
Error=+1.1ml
For Beaker;
=24.03+24.11+23.96/3
=24.03
Error =+0.97ml
For Cylinder;
=23.81+23.57+23.56/3
=24.60
Error=+0.40ml
Discussion:
According to result it can be noted that actual volume calculated were not equal to the volume
5 Page
predicted . the actual volume were less then the predicted volume this could be due to errors, the
result of instruments can be reported as: 23.90 +1.1 for flask and 24.03+0.97 for
beaker,24.60+0.40 for measuring cylinder. All the instruments were not precise in experiment
conducted, for precision is the repeatability of measured values of the same quantity. According
to our result the minimum error were found in case of measuring cylinder so it is more accurate
measuring glass ware then beaker and flask.
Result:
It was observed that the volume of various standardized was not accurate ,this to tell us that glass
ware in various lab were not accurately contain the volume prescribe on them unlike the
parameters like density and temperature of the substance are noted.
Page 6