Professional Documents
Culture Documents
442M
6200
9
Installation 2-1
Filling the Jacket Reservoir ......................................................................... 2-1
Power Connection....................................................................................... 2-1
Jacket Cooling Water Connection............................................................... 2-1
Oxygen Filling Connection .......................................................................... 2-2
Printer and Balance Connections................................................................ 2-2
Swagelock Tube Fittings ............................................................................. 2-3
Retightening Swagelock Tube Fittings ........................................................ 2-3
6200 External Plumbing .............................................................................. 2-5
Standardizing the Calorimeter..................................................................... 2-7
3
6200 Calorimeter Operating Instruction Manual
TABLE OF CONTENTS
Operating Instructions 5-1
Operating the 1108 Oxygen Bomb.............................................................. 5-1
Operating the Filling Connection................................................................. 5-1
Operating the Calorimeter........................................................................... 5-2
Samples and Sample Holders..................................................................... 5-5
Combustion Aids ......................................................................................... 5-5
Oxygen Charging Pressure......................................................................... 5-5
Combustion Capsules ................................................................................. 5-6
Foodstuffs and Cellulosic Materials ............................................................ 5-6
Coarse Samples.......................................................................................... 5-6
Corrosive Samples...................................................................................... 5-7
Explosives and High Energy Fuels ............................................................. 5-7
Volatile Sample Holders .............................................................................. 5-7
Poor Combustion ........................................................................................ 5-8
4
6200 Calorimeter Operating Instruction Manual
TABLE OF CONTENTS
Appendix B Calculations B-1
Calculating the Heat of Combustion............................................................ B-1
General Calculations................................................................................... B-1
Thermochemical Corrections ...................................................................... B-2
ASTM and ISO Methods Differ.................................................................... B-3
Acid and Sulfur Corrections ........................................................................ B-4
ASTM Treatment for Acid and Sulfur .......................................................... B-6
ISO Calculations ......................................................................................... B-6
Spiking Samples ......................................................................................... B-7
Conversion to Other Moisture Bases .......................................................... B-7
Conversion to Net Heat of Combustion....................................................... B-7
Tables
B-1 Settings for ISO & BSI Methods ......................................................... B-5
C-1 Calorimeter Control Limits.................................................................. C-2
D-1 6200 Calorimeter Serial Ports Pin-Out ............................................... D-3
D-2 Calorimeter Run Data Template......................................................... D-4
6
6200 Calorimeter Operating Instruction Manual
PREFACE
This manual contains instructions for component parts and peripheral items Scope
installing and operating the Parr 6200 used with the 6200 Calorimeter have
Calorimeter. For ease of use, the manual is been included and made a part of these
divided into nine chapters. instructions.
No. Description
Concept of Operation
201M Limited Warranty
Installation 483M Introduction to Bomb
Calorimetry
Instrument Description
205M 1108 Oxygen Combustion
Program Installation & Control Bomb
207M Analytical Methods for
Operating Instructions
Oxygen Bombs
Corrections & Final Reports 230M Safety Precautions
Reporting Instructions
Additional instructions for the printer,
File Management cooler, and water handling systems are
found in the respective package and should
Maintenance & Troubleshooting
be made a part of this book.
I On position
O Off position
8II
6200 Calorimeter Operating Instruction Manual
PREFACE
Safety Information General Specifications
To avoid electrical shock, always:
1. Use a properly grounded electrical Electrical Ratings
outlet of correct voltage and current 120VAC, 6.0 Amps. 50/60 Hz
handling capability. 240VAC, 3.0 Amps, 50/60 Hz
2. Ensure that the equipment is
connected to electrical service Before connecting the calorimeter to an
according to local national electrical electrical outlet, the user must be certain
codes. Failure to properly connect that the electrical outlet has an earth ground
may create a fire or shock hazard. connection and that the line, load and other
3. For continued protection against characteristics of the installation do not
possible hazard, replace fuses with exceed the following limits:
same type and rating of fuse.
4. Disconnect from the power supply Voltage: Fluctuations in the line voltage
before maintenance or servicing. should not exceed 10% of the rated nominal
voltage shown on the data plate.
To avoid personal injury:
1. Do not use in the presence of Frequency: Calorimeters can be operated
flammable or combustible materials; from either a 50 or 60 Hertz power supply
fire or explosion may result. This without affecting their operation or
device contains components which calibration.
may ignite such material.
2. Refer servicing to qualified Current: The total current drawn should not
personnel. exceed the rating shown on the data plate
on the calorimeter by more than 10 percent.
Intended Usage
If the instrument is used in a manner not Environmental Conditions
specified by Parr Instrument Company, the
protection provided by the equipment may Operating: 15C to 30C; maximum relative
be impaired. humidity of 80% non-condensing.
Installation Category II (overvoltage)
in accordance with IEC 664.
Cleaning & Maintenance Pollution degree 2 in accordance
Periodic cleaning may be performed on with IEC 664.
the exterior surfaces of the instrument with Altitude Limit: 2,000 meters.
a lightly dampened cloth containing mild Storage: -25C and 65C; 10% to 85%
soap solution. All power should be relative humidity.
disconnected when cleaning the instrument.
Provisions for Lifting and
There are no user serviceable parts
inside the product other than what is Carrying
specifically called out and discussed in this Before moving the instrument,
manual. Advanced troubleshooting disconnect all connections from the rear of
instructions beyond the scope of this the apparatus. Lift the instrument by
manual can be obtained by calling Parr grabbing underneath each corner.
Instrument Company in order to determine
which part(s) may be replaced or serviced.
III
9
6200 Calorimeter Operating Instruction Manual
PREFACE
Getting
Started These steps are offered to help the user functions used to modify the program
become familiar with, install, operate and contained in the 6200 Calorimeter. A
develop the full capabilities of the Parr 6200 review of the menus will provide a good
Calorimeter. idea of the capabilities and flexibility
designed into this instrument.
1. Review the Concept of Operations,
Chapter 1, to get an understanding of 7. Review the Calculations, Appendix B.
the overall capabilities of the calorimeter This provides information about
and microprocessor control. calculations performed by the 6200
Calorimeter.
2. Unpack and install the calorimeter in
accordance with the Installation 8. Review Standardization, Appendix C. This
Instructions, Chapter 2. This simple, will serve two important functions. First, it
step-wise procedure will acquaint the provides instructions on generating the
user with the various parts of the energy equivalent factor required to
calorimeter and make it easier to calculate the heat of combustion of
understand the operating instructions unknown samples. Secondly, it will give the
which follow. user the opportunity to run tests on a
material with a known heat of combustion to
3. Turn the power switch ON (located on become familiar with the instrument and
the back). Turn to the Instrument confirm that the instrument and operating
Description, Chapter 3, to review the procedures are producing results with
touch screen controls. acceptable precision. Most 6200
Calorimeters will have an energy equivalent
4. Review the Program Installation and of approximately 2400 calories per C. The
Control, Chapter 4, to match the factory runs for standardization and determinations
settings to the intended mode of are identical, except for the setting of the
operation. Any required changes can be instrument to the standardization or
made to the program parameters located determination mode.
in the Main Menu.
9. Review the Communication Interfacing,
5. Review the Reporting Instructions, Appendix D, for the correct installation of
Chapter 7, to become familiar with the any peripherals connected to the 6200
manner in which calorimetry corrections Calorimeter.
are entered. Also discussed are
generating final reports, editing and 10. After successful standardization, the
clearing memory. 6200 Calorimeter should be ready for
testing samples.
6. Turn to the Menu Operating Instructions,
Appendix A, to review the menu
IV10
6200 Calorimeter Operating Instruction Manual
1 CONCEPT OF OPERATION
The 6200 Calorimeter has been inlet check valve and an adjustable Overview
designed to provide the user with: needle valve for controlled release
of residual gasses following
A traditional design calorimeter with combustion. They are intended for
removable oxygen bomb and bucket. samples ranging from 0.6 to 1.2
grams with a maximum energy
A moderately priced calorimeter which release of 8000 calories per charge.
uses real time temperature
measurements to determine heat leaks The 1108 oxygen bomb is made of high-
using a controlled calorimeter jacket. strength, high nickel stainless steel
designed to resist the corrosive acids
A high precision calorimeter capable of produced in routine fuel testing. An
exceeding the repeatability and alternative 1108CL bomb is available,
reproducibility requirements of all constructed of an alloy containing additional
international standard test methods. cobalt and molybdenum to resist the
corrosive conditions produced when burning
A compact calorimeter requiring samples containing chlorinated compounds.
minimum laboratory bench space.
The Model 6200 can also be equipped
A modern intuitive graphical user with a variety of special purpose oxygen
interface for ease of operation and bombs for unusual samples and/or
training. applications. The 1104 high strength
oxygen bomb is designed for testing
A calorimeter with up to date digital explosives and other potentially hazardous
hardware, software and materials. The 1109/1109A semimicro
communications capabilities. oxygen bombs can be fitted along with its
unique bucket to test samples ranging from
A calorimeter that is cost effective and 25 to 200 mg.
which can incorporate a users current
bombs, buckets, and accessories. Removable Bucket
11
1-1
6200 Calorimeter Operating Instruction Manual
1
Dynamic In its Dynamic Operating Mode, Repeated testing, and over 20 years of
Operation the calorimeter uses a sophisticated routine use in fuel laboratories, has
curve matching technique to compare the demonstrated that this technique can cut
temperature rise with a known thermal the time required for a test by one-half
curve to extrapolate the final temperature without significantly affecting the precision
rise without actually waiting for it to develop. of the calorimeter.
Full Micro- In addition to its process A. Collect and store all required
control functions, the test data.
processor
microprocessor in the B. Apply all required corrections for
Based Data calorimeter has been combustion characteristics.
Acquisition preprogrammed to: C. Compute and report the heat of
and Handling combustion for the sample.
1-2
12
6200 Calorimeter Operating Instruction Manual
2 INSTALLATION
The 6200 Calorimeter is completely This apparatus is to be used indoors. It
assembled and given a thorough test before it requires at least 4 square feet of workspace
is shipped from the factory. If the user follows on a sturdy bench or table in a well-ventilated
these instructions, installation of the area with convenient access to an electric
calorimeter should be completed outlet, running water and a drain. The supply
with little or no difficulty. If the factory settings voltage must be within 10% of marked
are not disturbed, only minor adjustments will nominal voltage on the apparatus. The supply
be needed to adapt the calorimeter to voltage receptacle must have an earth ground
operating conditions in the users laboratory. connection.
The water reservoir of the calorimeter must the tank fill elbow on the back of the Filling the
be filled with approximately 1.4 liters of water calorimeter. The tank is full once water Jacket
(distilled or de-ionized preferred). This must be stands in the horizontal run of the filling Reservoir
done prior to turning on the heater and the elbow.
pump. The reservoir is filled through
Plug the power line into any grounded Turn the power switch to the on Power
outlet providing proper voltage that matches position. After a short time, the Parr Connection
the specification on the nameplate of the logo will appear on the LCD display
calorimeter. The calorimeter will draw followed by a running description of the
approximately 300 watts of power. Grounding instrument boot sequence. When the boot
is very important not only as a safety measure, sequence is complete, the calorimeter Main
but also to ensure satisfactory controller Menu is displayed. Go to the Calorimeter
performance. If there is any question about Operation page and turn the heater and pump
the reliability of the ground connection through on. This begins circulating and heating the
the power cord, run a separate earth ground calorimeter jacket water. Add water to the
wire to the controller chassis. filling elbow at the rear of the instrument as
required in order to keep it full.
It becomes necessary to use the jacket Tap Water Cooling Jacket
cooling water connection only if the calorimeter Connect the tap water supply Cooling
operating room temperature exceeds 24 C to the cold water inlet on the back Water
(75F). of the calorimeter using either Connection
copper or nylon tubing
When required, an external water source is (HJ0025TB035). A 196VB metering valve is
used to cool the jacket of the 6200 Calorimeter. provided with the calorimeter. This valve
This is done in either of the following ways: should be installed in this inlet line near the
1. Tap water is used for cooling and then run to a calorimeter. This valve is used to adjust the
drain. flow of water to the heat exchanger to
2. Cooling water is re-circulated to the compensate for differences in tap water
calorimeter from a Parr water handling system. temperatures and water line pressures. Once
the calorimeter is operating at equilibrium,
The water that provides the cooling goes check the jacket temperature that is displayed
through a heat exchanger and does not mix on the operating page. If this temperature is
with the water in the jacket and its reservoir. cycling significantly, close down on the
There is a very low cooling load and tap water metering valve to reduce the flow of cooling
up to a temperature of 27C should be water. If the jacket rises above its 30 C set
adequate. point, open this valve to increase the cooling.
A flow rate of 100 ml / minute is generally all
that is required.
132-1
6200 Calorimeter Operating Instruction Manual
2
Cooling with the Water Handling
Jacket Connect the cooling water outlet System
Cooling on the back of the calorimeter to a
Water drain using either nylon If the calorimeter is to be operated with
Connection (HJ0025TB035) or copper tubing. a Parr Water Handling System, connect the
A shut off valve in tap water supply pump output to the cooling water inlet and
Continued
line is also a good idea if the connect the cooling water outlet to the
calorimeter will not be used for an extended return connection on the water handling
period. system. With this installation it is neither
necessary nor desirable to install the 196VB
metering valve in the inlet line. It is a good
idea to keep all water line runs as short as
practical to avoid unwanted temperature
changes in the water between the source
and the calorimeter.
Oxygen
The 6200 Calorimeter is All connections should be checked
Filling equipped with an automatic bomb for leaks. Any leaks detected must be
Connection oxygen filling system. This system corrected before proceeding.
consists of an oxygen pressure regulator Instructions for operating the filling
with a relief valve that mounts on an oxygen connection are in the Operating
tank and a controlled solenoid inside the Instructions chapter.
calorimeter. To install the regulator on the
oxygen supply tank, unscrew the protecting Adjust the pressure regulator to deliver
cap from the oxygen tank and inspect the 450 psi of O2. Assemble the oxygen bomb
threads on the tank outlet to be sure they without a charge and attach the filling hose
are clean and in good condition. Place the to the bomb inlet valve. Press the O2 fill key
ball end of the regulator in the outlet and on the Calorimeter Operation page and
draw up the union nut tightly, keeping the observe the delivery pressure on the 0
gages tilted slightly back from an upright 600 psi gage while the oxygen is flowing
position. Connect the regulator to the into the bomb. Adjust the regulator, if
oxygen inlet fitting on the back of the needed, to bring the pressure to 450 psi. If
calorimeter case. This hose should be there is any doubt about the setting, release
routed so that it will not kink or come in the gas from the bomb and run a second
contact with any hot surface. Connect the check.
high-pressure nylon hose with the push on
connector to the back of the calorimeter.
Connect the printer to the directly to the RS232C printer port on the
Printer and
calorimeter at this time. The Parr back of the calorimeter.
Balance 1757 Printer is configured and
Connections furnished with a cord to connect The balance port connection, if used,
should be made at this.
2-2
14
6200 Calorimeter Operating Instruction Manual
2
Swagelok
When Swagelok Tube Fittings are Tube
used, the instructions for installation are: Fittings
1. Simply insert the tubing into the
Swagelok Tube Fitting. Make sure
that the tubing rests firmly on the
shoulder of the fitting and that the
nut is finger-tight.
2. Before tightening the Swagelok nut,
scribe the nut at the 6 oclock
position.
3. While holding the fitting body
steady with a back-up wrench,
tighten the nut 1-1/4 turns. Watch
the scribe mark, make one
complete revolution and continue to
the 9 oclock position.
4. For 3/16" and 4mm or smaller tube
fittings, tighten the Swagelok nut
3/4 turns from finger-tight.
Swagelok tubing connections can be thickness also has an effect on the Retightening
disconnected and retightened many times. amount of tightening required. Plastic Swagelok
The same reliable leak-proof seal can be tubing requires a minimal amount of Tube Fittings
obtained every time the connection is additional tightening while heavy wall
remade using the simple two-step metal tubing may require somewhat
procedure. more tightening. In general, the nut only needs
1. Insert the tubing with pre-swaged ferrules to be tightened about 1/8 turn beyond finger
into the fitting body until the front ferrule tight where the ferrule seats in order to obtain a
seats. tight seal.
2. Tighten the nut by hand. Rotate the nut to
the original position with a wrench. An Over tightening the nut should be avoided.
increase in resistance will be encountered Over tightening the nut causes distortion
at the original position. Then tighten slightly (flaring) of the lip of the tube fitting where the
with a wrench. Smaller tube sizes (up to ferrule seats. This in turn causes the threaded
3/16 or 4mm) take less tightening to reach portion of the body to deform. It becomes
the original position than larger tube sizes. difficult to tighten the nut by hand during a
The type of tubing and the wall subsequent re-tightening when the fitting body
becomes distorted in this manner.
2-315
6200 Calorimeter Operating Instruction Manual
2
Figure 2-1
6200 Calorimeter Back Panel
2-4
16
6200 Calorimeter Operating Instruction Manual
2
Figure 2-3
2-5
17
6200 Calorimeter Operating Instruction Manual
2
Open Loop Configurations
Figure 2-4
Figure 2-5
2-6
18
6200 Calorimeter Operating Instruction Manual
2
The calorimeter must be accurately Calorimeter Operations Page, Standardizing
standardized prior to actually performing using the Bomb key. All bomb the Calorimeter
calorimetric tests on sample materials. and bucket combinations will need to be
Review Appendix C - Standardization, in standardized separately. The end result of a
order to become familiar with the general standardization test is an energy equivalent
procedure and calculations. The user value, or the amount of energy required to
should configure the calorimeter at this time raise the temperature of the calorimeter one
to accommodate the desired sample weight degree. Repeated standardization with any
entry mode. The calorimeter can be placed given bomb and bucket combination should
into standardization mode on the yield an energy equivalent value with a
Calorimeter Operation Page, with the range of 14 calories per degree, centered
operating mode key. If two bombs and around the mean value for all tests using
buckets are being used with the calorimeter that bomb and bucket combination. The
to maximize sample throughput, the calorimeter is ready for testing samples
calorimeter can be configured to prompt for after an energy equivalent value has been
a Bomb ID at the start of each test. The obtained.
Bomb ID can also be selected on the
2-7
19
6200 Calorimeter Operating Instruction Manual
2
This page intentionally left blank.
2-8
20
6200 Calorimeter Operating Instruction Manual
3 INSTRUMENT DESCRIPTION
All calorimeter configurations and Types of
operations are handled by a menu-driven Controls
system operated from the bright touch
screen display. The settings and controls
are organized into nine main sections or
pages which comprise the MAIN MENU.
Note:
Keys with a double box in the upper
left hand corner lead to sub-menus.
The controls that change the data field Some keys lead to multiple choices. Menu Keys
information in the menus will be one of the Always clear the current value before
following: entering a new value. Once entered the
1. Toggles. These data fields contain screen will revert to the previous menu
ON/OFF or YES/NO choices. Simply and the new value will be displayed in
touching the key on the screen toggles the lower right corner of the key.
the choice to the other option. The 4. Data Displays. Most of these keys
current setting is displayed in the lower display values that have been calculated
right corner of the key. by the calorimeter and are informational
2. Option Selection. These data fields only. Certain ones can be overridden by
contain a list of options. Touching the the user entering a desired value
key on the screen steps the user through a sub-menu. The value is
through the available choices. The displayed in the lower right corner of the
current setting is displayed in the lower key.
right corner of the key. Note:
3. Value Entry Fields. These data fields Some keys will respond with an
are used to enter data into the opportunity for the user to confirm
calorimeter. Touching the key on the the specified action to minimize
screen brings up a sub menu with a accidental disruptions to the
key pad or similar screen for entering program and/or stored data.
the required value.
There are five control keys which 3. Start. This key is used to start a Control Keys
always appear in the right column of the calorimeter test.
primary displays. These keys are 4. Report. This key is used to access the
unavailable when they are gray instead test results stored in the calorimeter, to
of white. enter thermochemical corrections and to
1. Escape. This key is used to go up one initiate report on the display, printer or
level in the menu structure. attached computer
2. Main Menu. This key is used to return 5. Help. This key is used to access help
to the main menu touch screen from screens related to the menu currently
anywhere in the menu structure. displayed on the touch screen.
3-1
21
6200 Calorimeter Operating Instruction Manual
3
This page intentionally left blank.
3-2
22
6200 Calorimeter Operating Instruction Manual
4 PROGRAM INSTALLATION & CONTROL
Software The program in the 6200 Numerous provisions are included to
Installation Calorimeter can be extensively permit the use of other reagent
modified to tailor the unit to a wide concentrations, techniques, combustion
variety of operating conditions, aids and short cuts appropriate for the
reporting units, laboratory techniques, users work.
available accessories and
communication modes. Note:
Changes to the program are made by
In addition, the calculations, use of the menu structure described in
thermochemical corrections and Appendix A of this manual. Any of these
reporting modes can be modified to items can be individually entered at any
conform to a number of standard test time to revise the operating program.
methods and procedures.
4-1
23
6200 Calorimeter Operating Instruction Manual
4
The default parameters of the 6200 Go to the Program Info and Control Revising
Calorimeter can be changed to guarantee Menu, User/ Factory Settings, User Default
that the 6200 Calorimeter, when cold Setup ID, and enter the desired User Settings
restarted, will always be in the desired Setup ID.
configuration before beginning a series of
tests. Select Save User Default Settings
4-2
24
6200 Calorimeter Operating Instruction Manual
4
Factory Default Figure 4- 1
Settings 6200 Factory Default Settings
254-3
6200 Calorimeter Operating Instruction Manual
4
Data Entry Controls Communication Controls
Prompt for Bomb ID ON Printer Port (RS232)
Weight Entry Mode Touch Screen Balance Port (RS232)
Acid Entry Mode Touch Screen Network Interface
Hydrogen Entry Mode Touch Screen
Auto Sample ID Controls ON Printer Destination Local (RS232)
Sample Weight Warning above 2.0 Bar Code Port (RS232)
Spike Weight Entry Mode Touch Screen Network Data Devices
Sulfur Entry Mode Touch Screen
Moisture Entry Mode Touch Screen
Auto Preweigh Controls ON Printer Port Communications
Number of Data Bits 8
Parity None
Auto Sample ID Controls Number of Stop Bits 1
Automatic Sample ID ON Handshaking Xon/Xoff
Automatic Sample ID Increment 1 Baud Rate 9600
Automatic Sample ID Number 1 Printer Type Parr 1757
Data Logger
Data Logger OFF
Data Log Interval 12s
Data Log DestinationLog File and Printer
Select Data Log Items
Data Log Format Text Format
4-4
26
6200 Calorimeter Operating Instruction Manual
5 OPERATING INSTRUCTIONS
Operating the Detailed Instructions for 205M. Follow these instructions carefully,
1108 Oxygen preparing the sample and charging giving particular attention to the precautions
Bomb the 1108 Oxygen Bomb are given to be observed in charging and handling the
in Operating Instructions No. bomb.
Operating To fill the bomb, connect the overnight or longer, close the tank valve to
the Filling hose to the bomb inlet valve and prevent leakage. When changing oxygen
Connection push the O2 button on the tanks, close the tank valve and push the O2
calorimeter control panel. The calorimeter FILL key to exhaust the system. Do not use
will then fill the bomb to the preset pressure oil or combustible lubricants on this filling
and release the residual pressure in the system or on any devices handling oxygen
connecting hose at the end of the filling under pressure. Keep all threads, fittings,
cycle. It will take approximately 60 seconds and gaskets clean and in good condition.
to fill the bomb. During this time a Replace the two 394HCJE O-rings in the
countdown timer on the O2 fill button will slip connector if the connector fails to
display the remaining fill time. Pushing the maintain a tight seal on the bomb inlet
O2 key a second time will stop the flow of valve.
oxygen at any time. Once the display
returns to its normal reading, the user can The recommended filling pressure is 450
disconnect the coupling and proceed with psig (3 MPa or 30 bar). This pressure is
the combustion test. prescribed by most of the standard bomb
calorimetric test methods. Higher or lower
If the charging cycle should be started filling pressures can be used, but the bomb
inadvertently, it can be stopped immediately must never be filled to more than 600 psig
by pushing the O2 fill key a second time. (40 atm).
During extended periods of inactivity,
5-1
27
6200 Calorimeter Operating Instruction Manual
5
All operations required to standardize Each bomb and bucket Operating the
the 6200 Calorimeter, or test an unknown combination will have to be Calorimeter
sample, should proceed step-wise in the standardized separately and the
following manner: proper energy equivalent for each
set must be used when calculating
1. Turn on the calorimeter and activate the heat of combustion.
the pump and heater using
3. Fill the calorimeter bucket by first
Calorimeter Operations. Allow at
taring the dry bucket on a solution or
least 20 minutes for the calorimeter
trip balance; then add 2000 (+/- 0.5)
to warm up and the jacket
grams of water. Distilled water is
temperature to stabilize. Once the
preferred, but demineralized or tap
jacket temperature comes within
water containing less than 250 ppm
0.5C of 30C and stays there for
of dissolved solids is satisfactory.
approximately 10 minutes, the
The bucket water temperature
calorimeter is ready to begin testing.
should be approximately 3 to 5 C
The bomb parts should be wetted
below the jacket temperature. It is
and then dried in the manner used at
not necessary to use exactly 2000
the conclusion of a test. This serves
grams, but the amount selected
to wet all sealing parts as well as
must be duplicated within +/- 0.5
leaving the bomb with the same
gram for each run. Instead of
amount of residual water which will
weighing the bucket, it can be filled
exist in all subsequent testing.
from an automatic pipet, or from any
2. Prepare the sample weighing the other volumetric device if the
material to 0.1 mg and charge the repeatability of the filling system is
oxygen bomb as described in the within +/- 0.5 ml.
section entitled, Operating the Filling
To speed and simplify the bucket
Connection. Using an additional
filling process, and to conserve
bomb and bucket can increase the
water and energy, Parr offers a
throughput of the 6200 Calorimeter.
closed circuit Water Handling
With this arrangement, the
System (No. 6510). This provides a
calorimeter can operate almost
water supply, cooled to the starting
continuously since the operator will
temperature and held in an
be able to empty a bomb and
automatic pipet ready for delivery in
recharge it while a run is in progress.
the exact amount needed to fill the
A bomb and bucket for the next run
bucket. Instructions for this
will be ready to go into the
automatic system are given in
calorimeter as soon as it is opened.
Operating Instruction No. 454M.
5-2
28
6200 Calorimeter Operating Instruction Manual
5
Operating the 4. Set the bucket in the a series of short beeps to warn the
Calorimeter calorimeter. Attach the lifting user to move away from the
Continued handle (421A) to the two calorimeter. Once the bomb has
holes in the side of the been fired, the status bar will display
screw cap and partially lower the POSTPERIOD. The calorimeter will
bomb in the water. Handle the check to make certain that a
bomb carefully during this operation temperature rise occurs and will then
so that the sample will not be look for the final equilibrium
disturbed. Push the two ignition lead conditions to be met. If it fails to
wires into the terminal sockets on meet either the initial or final
the bomb head. Orient the wires equilibrium conditions, or if it fails to
away from the stirrer shaft so they detect a temperature rise within the
do not become tangled in the stirring allotted time, the calorimeter will
mechanism. Lower the bomb terminate the test and advise the
completely into water with its feet user of the error.
spanning the circular boss in the
bottom of the bucket. Remove the 8. At the conclusion of the test, the
lifting handle and shake any drops of calorimeter will signal the user.
water back into the bucket and
check for gas bubbles. 9. Open the cover and remove the
bomb and bucket. Remove the
5. Close the calorimeter cover. This bomb from the bucket and open the
lowers the stirrer and thermistor knurled valve knob on the bomb
probe into the bucket. head to release the residual gas
pressure before attempting to
6. Select determination or remove the cap. This release
standardization as appropriate on should proceed slowly over a period
the Calorimeter Operation menu by of not less than one minute to avoid
toggling the operating mode key. entrainment losses. After all
After pressing the start key, the pressure has been released,
calorimeter will now prompt the unscrew the cap; lift the head out of
operator for Bomb ID number, the cylinder and examine the interior
sample ID number, sample weight of the bomb for soot or other
and spike weight in accordance with evidence of incomplete combustion.
the instructions set into the operating If such evidence is found, the test
controls page. will have to be discarded.
Otherwise, wash all interior surfaces
7. The calorimeter will now take over of the bomb, including the head, with
and conduct the test. During the a jet of distilled water and collect the
time it is establishing the initial washings in a beaker.
equilibrium, it will display
PREPERIOD on the status bar. Just
before it fires the bomb, it will sound
5-3
29
6200 Calorimeter Operating Instruction Manual
5
10. Remove all unburned pieces of fuse dissolving 3.76 grams of Na2CO3 in
wire from the bomb electrodes; the water and diluting to one liter.
straighten them and measure their NaOH or KOH solutions of the same
combined length in centimeters. normality may be used.
Subtract this length from the initial
length of 10 centimeters and multiply 12. Analyze the bomb washings to
this burned length by 2.3 calories determine the sulfur content of the
per cm (for Parr 45C10 Fuse Wire) sample if it exceeds 0.1%. Methods
to obtain the fuse correction. The for determining sulfur are discussed
scale on the fuse wire card can be in Operating Instructions No. 207M.
used to obtain this value directly.
13. At the end of the testing period, turn
11. Titrate the bomb washings with a OFF the calorimeter at the power
standard sodium carbonate solution switch.
using methyl orange, red or purple
indicator. A 0.0709N sodium
carbonate solution is recommended
for this titration to simplify the
calculation. This is prepared by
5-4
30
6200 Calorimeter Operating Instruction Manual
5
Samples
and Sample Particle Size and Moisture addition in giving controlled burn rates, the
Holders Content. Solid samples burn best in an pelletizing of samples keeps the sample in
oxygen bomb when reduced to 60 the fuel capsule during combustion.
mesh, or smaller, and compressed into
a pellet with a 2811 Parr Pellet Press. Materials, such as coal, burn well in
the as-received or air-dry condition, but do
Large particles may not burn not burn completely dry samples. A
completely and small particles are certain amount of moisture is desirable in
easily swept out of the capsule by order to control the burning rate. Moisture
turbulent gases during rapid content up to 20% can be tolerated in many
combustion. cases, but the optimum moisture is best
determined by trial combustions.
Note:
Particle size is important because If moisture is to be added to retard
it influences the reaction rate. the combustion rate, drop water directly into
Compression into a pellet is a loose sample or onto a pellet after the
recommended because the sample has been weighed. Then let the
pressure developed during sample stand for awhile to obtain uniform
combustion can be reduced as distribution.
much as 40% when compared to
the combustion of the material in
the powder form. In
Combustion Note:
Aids Some samples may be difficult to It must be remembered, however, that a
ignite or they may burn so slowly that combustion aid adds to the total energy
the particles become chilled below the released in the bomb and the amount of
ignition point before complete sample may have to be reduced to
combustion is obtained. In such cases compensate for the added charge.
powdered benzoic acid, white oil or any
other combustible material of known Also, when benzoic acid is combusted
purity can be mixed with the sample. for standardization runs or for combustion
Ethylene glycol, butyl alcohol or decalin aid purposes, it should be in the form of a
may also be used for this purpose. pellet to avoid possible damage to the bomb
which might result from rapid combustion of
the loose powder.
Oxygen The 6200 Calorimeter has been changes from this value are not
designed to operate with an oxygen recommended.
Charging
filling pressure of 30 atm. Significant
Pressure
5-5
31
6200 Calorimeter Operating Instruction Manual
5
Non-volatile samples to be tested in Combustion
Capsules should be monitored for Capsules
Parr oxygen bombs are weighed and
wear. Do not use the capsule if the
burned in shallow capsules measuring
wall or base thickness is less than 0.025.
approximately 1" diameter and 7/16" deep.
These are available in stainless steel, fused
New capsules are heated in a muffle
silica and platinum alloyed with 3-1/2%
furnace at 500C for 24 hours to develop
rhodium.
this protective coating uniformly on all
surfaces. This treatment should be repeated
Stainless steel capsules (43AS) are
after a capsule has been polished with an
furnished with each calorimeter. When
abrasive to remove any ash or other surface
combusting samples that contain metal
deposits. Heating in a muffle is also a good
particles such as aluminum or magnesium,
way to destroy any traces of carbon or
the non-metallic (fused silica) 43A3 Capsule
combustible matter which might remain in
is required. When superior corrosion
the capsule from a previous test.
resistance is needed, the Platinum Rhodium
43A5 Capsule is required.
Note:
After heating, place the capsules in a
The stainless steel capsules will acquire
clean container and handle them only
a dull gray finish after repeated use in an
with forceps when they are removed to
oxygen bomb due to the formation of a
be weighed on an analytical balance.
hard, protective oxide film. This dull finish
not only protects the capsule, but it also
promotes combustion and makes it easier to
burn the last traces of the sample.
In most cases it may be necessary to completely. Whole wheat grains and Coarse
burn coarse samples without size reduction coarse charcoal chunks are typical of Samples
since grinding or drying may introduce materials which will burn satisfactorily
unwanted changes. There is no objection to without grinding and without additives or a
this if the coarse sample will ignite and burn special procedure.
5-6
32
6200 Calorimeter Operating Instruction Manual
5
Corrosive The 1108 Oxygen Bomb is made constructed of an alloy selected to
Samples of a corrosion resistant alloy specifically resist the corrosive effects of
designed to withstand the corrosive samples with high chlorine or chloride.
mixture of sulfuric and nitric acids
produced in normal fuel testing While no material will offer complete
operations. Samples containing chlorine corrosion resistance to these samples, the
and particular samples containing more 1108CL Bomb offers significantly enhanced
corrosion resistance for this service.
than 20 mg of chlorine samples with
high sulfur contents will greatly
accelerate corrosion of the bomb. An
alternate 1108CL Bomb is available
Explosives
The 1108 and 1108CL Bombs used unusually high energy levels, should not be
and High
in the 6200 Calorimeter have been tested with these bombs.
Energy designed to provide highly automated Rather, they should be tested in a model
Fuels testing of routine samples. Materials 1104 High Pressure Oxygen Bomb
which release large volumes of gas which designed specifically for these types of
detonate with explosive force or burn with samples.
Volatile Volatile samples can be handled determination. This can be done by running
Sample in a Parr 43A6 Platinum Capsule with a blank test with tape alone using a sample
Holders a spun rim, or in a Parr 43AS Alloy weighing 1.0 gram. The compensation for
Capsule which has a sturdy wall with heat of tape may be done through the spike
a flat top rim. These holders can be sealed option; see Spike Controls, Heat of
with a disc of plastic adhesive tape Combustion of Spike. The heat of
prepared by stretching tape across the top combustion of the 517A tape is
of the cup and trimming the excess with a approximately 6300 cal/gram.
sharp knife. The seal obtained after
pressing this disc firmly against the rim of Note:
the cup with a flat blade will be adequate for Tape should always be stored in a
most volatile samples. The tape used for sealed container to minimize changes in
this purpose should be free of chlorine and its moisture and solvent content
as low in sulfur as possible. Borden Mystic
Tape, No. M-169-C or 3M Transparent
Tape, No. 610, are recommended for this
purpose. The 3M Transparent Tape can be
ordered through Parr, Part No. 517A.
. Figure 5-7
Combustion Capsule
with Adhesive Tape Seal
5-7
33
6200 Calorimeter Operating Instruction Manual
5
Use the following procedure when filling small opening which is needed to Volatile
and handling any of these tape-sealed prevent collapse of the disc when Sample
sample holders: pressure is applied. Holders
Weigh the empty cup or capsule; then Fill the bomb with the usual oxygen Continued
cover the top with tape, trim with a knife charging pressure.
and press the trimmed edge firmly The calorimeter will fire the bomb and
against the metal rim. Also cut and complete the test in the usual manner.
attach a small flag to the disc (see
Figure 5-7). Volatile samples are defined as one with
Puncture the tape at a point below the an initial boiling point below 180C per
flag, then re-weigh the empty cup with ASTM D-2.
its tape cover.
Add the sample with a hypodermic Low volatile samples with a high water
syringe; close the opening with the flag content, such as urine or blood, can be
and re-weigh the filled cup. burned in an open capsule by absorbing the
Set the cup in the capsule holder and liquid on filter paper pulp or by adding a
arrange the auxiliary fuse so that it combustion aid, such as ethylene glycol.
touches the center of the tape disc.
Just before starting the test, prick the
disc with a sharp needle to make a
Poor
Because of the difference in combustion pass a 60 mesh screen may Combustion
characteristics of the many different materials not burn completely.
which may be burned in an oxygen bomb, it is The use of a sample pellet which has
difficult to give specific directions which will been made too hard or too soft. Either
assure complete combustions for all samples. condition can cause spalling and the
ejection of unburned fragments.
The following fundamental conditions Insufficient space between the
should be considered when burning samples: combustion cup and the bottom of the
Some part of the sample must be heated bomb. The bottom of the cup should
to its ignition temperature to start the always be at least one-half inch above
combustion and, in burning, it must the bottom of the bomb or above the
liberate sufficient heat to support its own liquid level in the bomb to prevent
combustion regardless of the chilling thermal quenching.
effect of the adjacent metal parts. Excessive moisture or non-
The combustion must produce sufficient combustible material in the sample. If
turbulence within the bomb to bring the moisture, ash and other non
oxygen into the fuel cup for burning the combustible material in the sample
last traces of the sample. amounts to approximately 20% or
Loose or powdery condition of the sample more of the charge, it may be difficult
which will permit unburned particles to be to obtain complete combustion. This
ejected during a violent combustion. condition can be remedied by adding a
The use of a sample containing coarse small amount of benzoic acid or other
particles which will not burn readily. combustion aid.
Coal particles which are too large to
5-8
34
6200 Calorimeter Operating Instruction Manual
6 CORRECTIONS & FINAL REPORTS
Entering Final reports for each test can be These are:
Corrections obtained whenever the operator is Manual Entry
and Obtaining prepared to enter any required Fixed Corrections
the Final corrections for fuse, acid and
sulfur. Program Installation and Control,
Report Chapter 4, provides the default settings
When entering corrections, the user can used to setup the method preferred by the
choose either of two methods. user.
Refer to the Reporting Instructions,
Chapter 7, for the steps necessary to initiate
a report from the controller
Manual During the reporting process, the If fixed values for fuse, acid and sulfur
Entry controller will prompt the user to enter are turned OFF on the Thermochemical
the following values: Corrections Page, then the user must
manually enter the values at the prompt.
Fuse Key in the Fuse
Correction Correction and press the If the Spiking Correction is used, a
ENTER key. The default setting spiking correction must be entered before
for this value is to be entered in obtaining a Final Report.
calories. This is explained in
Appendix B, Page B-2. If values for these corrections are
not available, the operator can use the SKIP
Acid Key in the Acid Correction key to bypass any of the corrections,
Correction and press the ENTER key. however, a Final Report will not be printed
The default setting for this value until an entry is made for fuse, acid and
is to be entered in milliliters of sulfur or spike if used.
standard alkali required to titrate
total acid or calories. After the last entry has been made,
the calorimeter will automatically produce a
Sulfur Key in the Sulfur Correction Final Report.
Correction and press the ENTER key.
The default setting for this value
is to be entered as percent
sulfur in the sample.
Fixed In many cases, fixed values for Details for applying fixed corrections are
Corrections fuse and acid can be used without found in Appendix B, Thermochemical
introducing a significant error since Calculations.
the corrections are both relatively
small and constant. Any value set-up as a fixed correction
will be automatically applied and the
Fixed sulfur corrections can also be controller will not prompt the user for this
used whenever a series of samples will be value.
tested with a reasonably constant sulfur
content.
6-1
35
6200 Calorimeter Operating Instruction Manual
6
This page intentionally left blank.
36
6-2
6200 Calorimeter Operating Instruction Manual
7 REPORTING INSTRUCTIONS
The 6200 Calorimeter can transmit its optional printer connected to the RS232C Report
stored test data in either of two ways. The port of the calorimeter. This page also Option
REPORT DESTINATION key on the selects the type of reports that are Selection
Reporting Controls Page toggles the report generated automatically by the calorimeter.
destination between the display and an
7-1
37
6200 Calorimeter Operating Instruction Manual
7
Report The displayed files can be sorted The DESEL ALL key is used to cancel
Generation by sample ID number, by type, by the current selection of files.
Continued status or by date of test by simply To bring the selected report or series of
touching the appropriate column. report to the display, press the DISPLAY
key. To send the reports to the printer
Individual files can be chosen by press the PRINT key.
highlighting them using the up and down
arrow keys to move the cursor. Press the The EDIT key brings up a sub-menu
SELECT key to actually enter the selection. which enables the operator to edit any of
Once selected the highlight will turn from the data in the report or add
dark blue to light blue. A series of tests can thermochemical corrections to convert
be selected by scrolling through the list and preliminary reports to final reports. Final
selecting individual files. reports can only be edited if EDIT FINAL
REPORTS on the reporting control page is
The double up and down keys will jump turned on.
the cursor to the top or bottom of the current
display. Net Heat of Combustion
To have the Net Heat of Combustion
If a range of tests is to be selected, print as part of preliminary and final reports,
select the first test in the series, scroll the go to the Thermochemical Corrections Page
selection bar to the last test in the series and turn ON Calculate Net Heat of
and press EXTEND SEL to select the Combustion. During the reporting process,
series. the controller will prompt for the hydrogen
(H) value.
7-2
38
6200 Calorimeter Operating Instruction Manual
8 FILE MANAGEMENT
The 6200 Calorimeter will hold data for
1000 tests in its memory. These tests may
be pre weights, preliminary or final reports
for either standardization or determination
runs. Once the memory of the controller is
filled, the controller will not start a new
analysis until the user clears some of the
memory.
The FILE MANAGEMENT key on the will then ask the user to confirm Clearing
main menu leads to the file management that this file is to be deleted. Memory
sub-menu. The RUN DATA FILE A series of files can be deleted by
MANAGER key leads to a listing of the files. selecting the first file in the series
Single files can be deleted by and then the last file in the series
highlighting the file and pressing using the EXTEND SEL key and
the DELETE key. The controller then pressing the DELETE key.
The controller of the 6200 calorimeter Compact flash memory cards Removable
can accept compact flash memory cards. are inserted into the slot on the Compact
These cards can be used to: back of the control section of the Flash
Copy test file data for transfer to a calorimeter. Keys are provided on Memory
computer the FILE MANAGEMENT sub-
Copy user settings for back up menu to initiate each of the above
Reload user settings to the three actions.
controller to restore or update the
controllers operating system.
8-1
39
6200 Calorimeter Operating Instruction Manual
8
This page intentionally left blank.
8-2
40
6200 Calorimeter Operating Instruction Manual
9 MAINTENANCE & TROUBLESHOOTING
Oxygen Under normal usage the 1108 Parr recommended the 1108 Oxygen
Bomb Oxygen Bomb will give long service if Combustion Bomb have O-rings
handled with reasonable care. and valve seats replaced after 6 months or
However, the user must remember 500 firings which ever comes first or at more
these bombs are continually subjected frequent intervals if the bomb has been
to high temperatures and pressures subject to heavy usage or if it shows any
that apply heavy stresses to the evidence of damage. Detailed information
sealing mechanism. The mechanical can be found in Bulletin 205M supplied as a
condition of the bomb must therefore part of this manual. This 1108 Oxygen
be watched carefully and any signs of Combustion Bomb is the only part of the
weakness or deterioration should be calorimeter system that requires routine
replaced before they fail. It is maintenance. All other problems will
require diagnosis and parts replacement.
9-1
41
6200 Calorimeter Operating Instruction Manual
9
Fuses The replacement of protective Note: Check the labels on the instrument
fuses for the 6200 Calorimeter should for correct fuse rating.
be performed by qualified personnel.
Parr Description Type Ratings
All fuses except Parr part # 139E23 are Part #
located on the A1794E I/O board located 139E23 Lines Fast- 15 Amps,
inside the instrument. To gain access to the Protective Acting 250Vac
inside, first disconnect the power cord from Fuses
the rear of the unit. Remove the two screws 1641E2 Heater Fuse Fast- 2.5
at the upper rear of the unit that secure the (F2) Acting Amps,
front cover. Pry up on the back lip of the 250Vac
cover and carefully hinge it forward. Make 1641E Pump Fuse Fast- 1 Amp,
sure that the cables attached to the front (F1) Acting 250VAC
portion of the cover do not become
stretched or disconnected. Follow this
procedure in reverse order to re-secure the
front cover.
6200 Calorimeter The calorimeter will run a The Current Run Has Aborted Due To
Error List number of diagnostic checks Timeout. The calorimeter has failed to
upon itself and will advise the operator if it establish an acceptable final temperature
detects any error conditions. Most of these within the time allowed. Possible causes for
errors and reports will be self-explanatory. The this error are listed below:
following list contains errors that are not
necessarily self-evident and suggestions for Poor jacket water circulation due to a
correcting the error condition. kinked hose or insufficient water in the tank.
A bomb leak.
A Misfire Condition Has Been Detected This Poor bucket stirring.
error will be generated in the event the total
temperature rise fails to exceed 0.5 C after There Is A Problem With The Bucket
the first minute of the post-period. Thermistor. Possible electrical open.
9-2
42
6200 Calorimeter Operating Instruction Manual
9
A / D Initialization Failed. Bomb EE Standard Deviation 6200 Calorimeter
Shortly after power is applied to the calorimeter Warning. The relative Error List
controller and the operating system has started, standard deviation for the Continued
the CPU attempts to read the unique IO board calibration runs in memory for
calibration information from the IO board. the indicated bomb exceeds
If the IO board is not connected to the CPU, or the preset limit.
the information on the board is not valid, this error
will be issued. The heater loop break limit has been detected.
The heater will now be shutdown. This error
Bomb ID Has Been Fired Times Which means that the calorimeter is trying to heat the
Exceeds The Bomb Service Interval. The water in the unit for an extended period of time.
calorimeter controller keeps track of how many The calorimeter suspects that there is a short
times the bomb has been fired. When this count and shuts the system down in order to save
exceeds a preset limit (usually 500) this message itself. This is a fairly normal occurrence if the
will be issued each time the bomb is used for a lab temperature is very cool at night. It is
test. Perform bomb maintenance and reset acceptable practice to ignore the warning and
counter on Calibration Data and Control page for re-start the unit. However, if this error occurs
appropriate bomb number. more than three times in a row, then it may be a
true thermistor problem and the user should
You Have Exceeded The Run Data File Limit contact Parr Technical Service.
(1000 Files) - The memory set aside for test runs
has been filled. Use the memory management Sample Weight Warning.
techniques to clear out non-current tests. The entered sample mass exceeds the value
entered via the Sample Weight Warning Above
key on the Data Entry Controls page. This
warning threshold is normally 2 grams.
9-3
43
6200 Calorimeter Operating Instruction Manual
9
This page intentionally left blank.
9-4
44
6200 Calorimeter Operating Instruction Manual
APPENDIX A MENU OPERATING INSTRUCTIONS
The settings and controls are organized This appendix describes all pages of the
into nine main sections or pages which menu-based operating system of the 6200
comprise the Main Menu. Calorimeter.
The calorimeter will normally be must only be turned on after the calorimeter
Calorimeter operated from the Calorimeter water tank is filled with water.
Operation Operation Page, although tests can
Menu always be started from any menu page. Note:
The heater and pump must be turned
ON to bring jacket to the correct starting
temperature before testing can
commence.
A-1
45
6200 Calorimeter Operating Instruction Manual
APPENDIX A
Setup. Press this key to access Temperature
the Temperature Plot Setup Menu, vs. Time Plot
which has many keys that permit Screen
the user to fully customize both the
x (time) axis and the scaling of the
y axis.
A-2
46
6200 Calorimeter Operating Instruction Manual
APPENDIX A
Temperature Time Units. Toggles between Bucket Max Value. Press this key to
Plot Setup minutes and seconds. access its numeric dialog box to set a
Menu maximum bucket value.
Continued Bucket Plot Color. Toggles
between: Jacket Plot Color: (same as Bucket Plot
Red Color, above).
Green
Yellow Jacket Max Value. Press this key to
Blue access its numeric dialog box to set a
Magenta maximum jacket value.
Cyan
White Time Minimum. Press this key to access
its numeric dialog box to set the least
Black
amount of time for the run.
A-3
47
6200 Calorimeter Operating Instruction Manual
APPENDIX A
Spiking Correction. Accesses sub-menu,
Operating
Spiking Controls. Spiking is the material
addition, such as benzoic acid or mineral oil, Controls
to samples which are difficult to burn in order Menu
to drive the combustion to completion.
A-4
48
6200 Calorimeter Operating Instruction Manual
APPENDIX A
Operating Calibrate Touchscreen. This keep the backlight ON at all times.
Controls key prompts the user to touch the
Menu screen at predefined points in LCD Contrast. This key accesses a
Continued order to facilitate touchscreen sub-menu with a slide control which
calibration. adjusts the contrast on the LCD display
for optimum viewing.
LCD Backlight Time-out. The
unit is equipped with an automatic Print Error Messages. When turned
circuit to shut-off the backlight ON, all error messages will be printed
when it is not being used. The on the printer as well as displayed on
back light will shut-off if there is the screen.
no keyboard activity for the
number of seconds entered. Language. Steps the calorimeter
Pressing any key will through the installed operating
automatically turn the back languages.
lighting ON. A setting of 0 will
Program
Information Settings Protect. Provides protection
and Control for the program options and settings on
the menus. If this is turned ON, the user
Menu
will be warned that enumeration keys
are locked when a key is pressed.
Enumeration Keys either toggle a value
(ON / OFF) or select from a predefined
list. This feature is used primarily to
protect the instrument settings from
accidental changes if one were to
inadvertently touch or bump up against
the touchscreen.
Date. Displays current date and User/Factory Settings. This key leads
accesses sub-menu on which to a sub-menu that allows the user to
date is set in (YY/MM/DD) format. save or recall user defined instrument
settings. Additionally, factory
Time. Displays current time and preinstalled settings supporting different
accesses sub-menu on which bombs or special operating modes can
time is set in (HH:MM) format. also be recalled.
A-5
49
6200 Calorimeter Operating Instruction Manual
APPENDIX A
Save User Default Settings. Used to record available for the calorimeter. When the Program
the setup to the memory once the user has user chooses a bomb, the instrument Information
configured the instrument to their operating must be re-booted to load the correct and Control
requirements. version of the software. Menu
Continued
Note: User Function Setup This key leads to
Keys which make global changes to the sub menus that support the configuration
setup of the calorimeter contain a YES or of five factory / user definable function
NO response to make certain that the user keys. The function keys are accessible
wishes to proceed. This two step entry is from the Diagnostics page.
intended to prevent inadvertent global
program changes. Cold Restart. Returns the instrument to its
last known state. Otherwise the instrument
Bomb Type Select This key toggles will return to the factory default settings.
through the different bomb models
A-6
50
6200 Calorimeter Operating Instruction Manual
APPENDIX A
A-7
51
6200 Calorimeter Operating Instruction Manual
APPENDIX A
Print Standardization Runs. Will print all of Bomb 2. Accesses sub-menu, Bomb 2. Calibration
the tests that have been incorporated into Provides the same controls as described and Data
the calculated EE value. This will be helpful for Bomb 1. Controls
in evaluating a series of tests which fail to Menu
produce a satisfactory EE value and relative Bomb 3. Accesses sub-menu, Bomb 3. Continued
standard deviation. Provides the same controls as described for
Bomb 1.
Reset Bomb Fire Count. After bomb
service, press this button to reset the fire Bomb 4. Accesses sub-menu, Bomb 4.
count to zero Provides the same controls as described for
Bomb 1.
A-9
53
6200 Calorimeter Operating Instruction Manual
APPENDIX A
Auto Sample ID Increment. Establishes Data Entry
the increment between sample numbers. Controls
Menu
Next Auto Sample ID Number. Establishes the
initial sample number for a series of tests and then
shows the next sample ID which will be assigned.
A-10
54
6200 Calorimeter Operating Instruction Manual
APPENDIX A
Data Entry Auto Preweigh ID Controls. Accesses Automatic Preweigh ID. ON/OFF toggle
Controls sub-menu, used to automatically assign for this feature.
Menu Sample ID numbers when a series of
Continued samples are preweighed ahead of the Automatic Preweigh ID Prefix. An entry here will
time they are actually tested. be used as a prefix for all pre-weigh sample IDs.
Reporting Report Width. The column width of the Automatic Report Destination. Directs the
Controls printer being used can be set to 40 or 80 reports to the Printer Port or the display.
Menu columns. Select 40 when the 1757 Printer is
used. Individual Reports. When set to ON, will
generate header information for each report
printed. In the OFF position, only one header
will be printed for a series of tests.
A-11
55
6200 Calorimeter Operating Instruction Manual
APPENDIX A
Number of Data Bits. Standard Communication
options for data bits. Toggles Controls Menu
between 7 and 8.
A-12
56
6200 Calorimeter Operating Instruction Manual
APPENDIX A
Communication Balance Port (RS232). Baud Rate. Standard options for baud rate.
Controls Menu Accesses sub-menu, Balance Choose from 19.2K , 9600, 4800, 2400, 1800,
Port Communications. 1200, 600, 300, 150, 134.5, 110, and 75.
Continued
Balance Type. This key toggles through the Data Characters from Balance. This setting is only
available balance templates. used when the generic balance format is selected.
This value determines the number of numeric data
Customize Balance Settings. This key leads characters (0-9 . + -) to accept. Any additional
to a submenu that allows the balance template characters after this value and before the string
and related port frame parameters to be altered terminating <CR> are discarded.
as required.
Data Precision. This key allows the user to
establish the number of digits to the right of the
decimal point that are passed from the balance
handler.
A-13
57
6200 Calorimeter Operating Instruction Manual
APPENDIX A
File Management
Run Data File Manager - This key activates as a means of either Menu
the File Manager. The File Manager is used archiving data or
to delete or rename test report files. It is also transferring it to a PC.
used to convert file types.
A-14
58
6200 Calorimeter Operating Instruction Manual
APPENDIX A
Diagnostics
Menu Allows the user to test many of View System Log - This key is used to
the components and subsystems of the display the contents of /flash/log/messages.
calorimeter. These capabilities must be This file is used primarily to log application
used in conjunction with the Maintenance program debug messages.
Instructions to obtain the maximum benefits
User Defined Functions This key leads to
from these capabilities.
a sub-menu that offers five special purpose
user / factory definable function keys.
A-15
59
6200 Calorimeter Operating Instruction Manual
APPENDIX A
This page intentionally left blank.
8-2
60
6200 Calorimeter Operating Instruction Manual
APPENDIX B CALCULATIONS
The 6200 Calorimeter will automatically instrument is set up so the Calculating the
make all of the calculations necessary to calculations match the Heat of
produce a gross heat of combustion for the procedures and the units are Combustion
sample. However, it is important that the user consistent throughout the
understand these calculations to ensure the process.
The calculation for the gross heat of These calculations are made in General
combustion is done by: calories, grams, and degrees Calculations
Celsius, and then converted to
WT- e1 e2 e3 other units if required.
Hc =
m Temperature Rise.
Where: The 6200 Calorimeter produces a
Hc = Gross heat of combustion. corrected temperature rise reading
T = Observed temperature rise. automatically. Corrections for heat leaks
W = Energy equivalent of the during the test are applied. (For a complete
calorimeter being used. discussion of this process see Introduction
e1 = Heat produced by burning to Bomb Calorimetry, Manual No. 483M).
the nitrogen portion of the air
trapped in the bomb to form Energy Equivalent.
nitric The energy equivalent (represented by
acid. W in the above formula, or abbreviated as
e2 = The heat produced by the EE) is determined by standardizing the
formation of sulfuric acid calorimeter as described in Appendix C -
from the reaction of sulfur Standardization. It is an expression of the
dioxide, water and oxygen. amount of energy required to raise the
e3 = Heat produced by the temperature of the calorimeter one degree.
heating wire and cotton It is commonly expressed in calories per
thread. degree Celsius. Since it is directly related to
m = Mass of the sample. the mass of the calorimeter, it will change
whenever any of the components of the
calorimeter (i.e. the bomb, bucket or amount
of water) is changed.
B-1
61
6200 Calorimeter Operating Instruction Manual
APPENDIX B
Thermochemical Nitric Acid Correction. correction and the fuse wire supplied by
Corrections In the high pressure oxygen Parr.
environment within the oxygen
bomb, nitrogen that was present as part of the When using the 1108P bomb, there are two
air trapped in the bomb is burned to nitric components to the fuse correction:
oxide which combines with water vapor to
form nitric acid. All of this heat is artificial The heat introduced by heating the wire
since it is not a result of the sample burning. used to ignite the cotton thread.
The nitric acid correction removes this excess
heat from the calculation. The heat of combustion of the cotton
thread used to ignite the sample.
Sulfur Correction.
In the oxygen rich atmosphere within the The semi-permanent heating wire is
bomb, sulfur in the sample is oxidized to heated by dissipating an electrical charge
sulfur trioxide which combines with water from a capacitor. Since this charge is
vapor to form sulfuric acid. This liberates controlled by the size of the capacitor and
additional heat over the normal combustion the charging voltage, and because the
process which converts sulfur to sulfur capacitor is fully discharged for each test,
dioxide. The sulfur correction removes this the energy released can be calculated. In
excess heat from the calculation. the 6200 Calorimeter this is a fixed
correction of 10 calories per test.
Fuse Correction. Cotton has a heat of combustion of
The fuse correction applied by the 4000 calories per gram. The actual thread
calorimeter is calculated as: being used should be weighed to see how
much is being burned. Ten centimeters of a
e3 = (fuse value)(fuse multiplier fine thread will weigh approximately 0.003
from grams which would release 12 calories as it
calculation factors page) burns. Heavier threads weigh up to 0.010
= (entered value)(fuse multiplier grams per 10 centimeters and increase this
from thermochemicals page) correction to 40 calories per test. The finer
the thread, the smaller errors will be if the
Fuse Value is the number entered by thread is not exactly ten centimeters in
the user and the value which appears in the length. Polyester thread is not
test report. recommended for use in the bomb because
Note: it has a tendency to melt and fall away from
Calculation Factors, - Fuse Multiplier is the heating wire before it ignites.
normally set to 1.0 so the entered value Using the fine thread mentioned above,
is in calories the fuse correction for the calorimeter would
Users may find it convenient to enter a be the 10 calories from electrical heating
fixed value for the fuse correction and avoid plus 12 calories from the burning thread for
the need to determine this correction for a total of 22 calories per test. The thread
each test. Fixed fuse corrections can be supplied by Parr has a mass of
entered when Thermochemical Corrections, approximately 1 milligram per centimeter.
is set to ON. By default a fixed fuse This results in a total fuse correction of 50
correction of 15 calories is applied to all calories. This value should be changed
tests. Total errors of more than 5 calories from the default value of 15 calories in the
will seldom occur when using a fixed fuse Thermochemical Corrections menu.
B-2
62
6200 Calorimeter Operating Instruction Manual
APPENDIX B
Current ASTM, ISO, and British proper corrections are applied, and the ASTM
Standard Methods differ on their treatment calculations made are consistent with and ISO
of the nitric and sulfuric acid the procedure used. Methods
thermochemical corrections. ASTM Differ
Methods call for titrating the bomb washings Note:
to determine the total acid present. This is Please review the following section on
assumed to be all nitric acid with a heat of Acid and Sulfur Corrections. Different
combustion of -14.1 Kcal per mole. The standard test methods use different
amount of sulfur is then determined and values for the heat of formation of
converted to equivalents of sulfuric acid. sulfuric acid. These differences are
The difference between the heat of generally insignificant. The 6200
formation of sulfuric acid (-72.2 Kcal per Calorimeter uses the most recent,
mole or -36.1 calories per milliequivalent) published values for all thermochemical
and nitric acid is then subtracted as the data.
sulfur correction.
Thermochemical Calculation
Most other test methods treat nitric Details
and sulfuric acid corrections as entirely
separate values instead of combined Traditionally, standard solutions and
values. This eliminates the requirement for procedures have been established to
a total acid determination and permits the simplify the calculations related to the
nitric acid correction to be handled in a thermochemical corrections. The 6200
variety of ways, including the assumption of Calorimeter has been programmed to
a fixed nitric acid correction. permit the user to use standard solutions
and units which are most convenient, since
The 6200 Calorimeter can be set up the microprocessor can easily apply any
to apply the acid correction by either the conversion factors required.
ASTM or ISO convention, as the user
prefers. Care must be used to ensure the
B-363
6200 Calorimeter Operating Instruction Manual
APPENDIX B
Acid and Total acid is the amount of base Example:
Sulfur required to titrate the bomb
Corrections washings (milliliters). (Percent Sulfur) x (Sample Mass) x
Nitric acid is that portion of the total 10 = milligrams of sulfur
acid in the bomb washings that result
when the nitrogen in the air that is (milligrams of sulfur) / (equivalent wt. of S
trapped in the bomb is burned at high in H2SO4) = milliequivalents of H2SO4
pressure. Since this nitric acid does not
result from the sample, and the Sulfur Multiplier is then 10 / equivalent wt.
combustion conditions are reasonably of S in H2SO4 or 10/16.03 = 0.6238
constant from test to test, the amount of
nitric acid formed is also constant. e1 is the nitric acid portion of the total acid
Acid multiplier is multiplied by the user correction.
entered acid value to arrive at the
number of milliequivalents of acid. This e2 is the sulfuric acid portion of the total
value is normally the concentration acid correction.
(normality) of the base in equivalents
per liter (N). e1 + e2 is the total bomb acid correction.
Percent sulfur is the concentration of
sulfur in the sample (weight %). Calculation for e1 when entered acid value
Molecular weight of sulfur is 32.06. is total acid:
Equivalent weight of sulfur in H2SO4
is 16.03 (one half of the molecular [((total acid) (acid mult))-(percent sulfur)
weight). (sample mass) (sulfur mult) ] (heat of
Heat of formation of nitric acid is 14.1 formation of HNO3)
calories/milliequivalent.
Total acid milliequivalents - milliequivalents
Heat of formation of sulfuric acid
of H2SO4 = milliequivalents of HNO3.
(from SO2) is 36.1 calories /
(milliequivalents of HNO3) (heat of formation
milliequivalent.
of HNO3) = heat (in calories) due to HNO3
Sample mass is the mass of sample
produced.
burned in the bomb (grams).
Sulfur multiplier is multiplied by the Calculation for e1 when entered acid value
product of the user entered sulfur value is nitric acid only:
and the sample mass to arrive at the
number of milliequivalents of sulfuric (nitric acid) (acid multiplier) (heat of
acid in the bomb washings. formation of HNO3).
B-4
64
6200 Calorimeter Operating Instruction Manual
APPENDIX B
Acid and
Users may find it convenient to enter a to the difference between the
fixed value for the acid correction and avoid actual and assumed value for Sulfur
the need to determine this correction for sulfur. Corrections
each test. Use of a fixed value for the acid Continued
correction is highly recommended. Fixed For ordinary work where benzoic acid is
acid corrections can be entered when Fixed used, for standardizing the calorimeter, the
Acid - Thermochemical Corrections, is set to Fixed Sulfur Correction, for
ON. A correction of 10 calories is a good Standardizations should be ON applying a
number for the fixed nitric acid value. For fixed value of 0.0 to all standardization
most work, it is recommended to set Acid tests. Benzoic acid contains no sulfur.
Value is Nitric Acid Only, in Calculation
Factors to ON. Total errors of more than 3 Please note that the values entered into
calories will seldom occur when using fixed the test report appear as entered in the
nitric acid corrections. report. Values for e1, e2 and e3 are
calculated and used as energy corrections
Fixed sulfur corrections can be entered in accordance with the formulas and
if a series of samples contain a constant settings given above. The formulas used
amount of sulfur. Fixed sulfur corrections above to arrive at e1 or e2 are not the same
can be entered when Fixed Sulfur - as the formulas used for e1 and e2 which
Thermochemical Corrections, is set to ON appear in most ASTM bomb calorimetric
and then enter percent sulfur as indicated procedures. However, the sum of e1 and e2,
on this line. Any errors will be proportional above, is equal to the sum of the ASTM
treatment of e1 and e2.
Table B-1
Settings for ISO & BSI Methods
65
B-5
6200 Calorimeter Operating Instruction Manual
APPENDIX B
ASTM In the ASTM treatment, the dioxide, under bomb conditions, which is -
correction for acid formation 72.2 Kcal per mole or -36.1 calories per
Treatment for
assumes that all the acid titrated milliequivalent. But remember, a correction
Acid and is nitric acid. Obviously, if sulfur of 14.1 calories per milliequivalent of sulfuric
Sulfur is present in the sample, which acid is already included in the ASTM nitric
in turn produces sulfuric acid, part of the acid correction (e1). Therefore the
correction for the sulfuric acid formed is additional correction which must be applied
already included in the ASTM nitric acid for sulfur will be the difference between 36.1
correction (e1). This is adjusted by a and 14.1 or 22.0 calories per milliequivalent
separate computation based upon the sulfur (44.0 Kcal per mole). For convenience, this
content of the sample. An additional is expressed, in the ASTM e2 formula, as
correction of 1.37 Kcal must be applied for 13.7 calories (44.0/32.06) for each
each gram of sulfur converted to sulfuric percentage point of sulfur per gram of
from sulfur dioxide. This is based upon the sample.
heat of formation of sulfuric acid, from sulfur
ISO Both the ISO 1928 and BSI The settings in Table B-1 assume that
Calculations 1016: Part 5 methods for testing the same procedure is carried out for both
the calorific value of coal and standardization and determination.
coke, deal with acid and sulfur corrections in
a manner which is somewhat different than The offset value is the product of -1, the
ASTM procedures. Provision has been Heat of Formation of Nitric Acid, the acid
made in the 6200 Controller for dealing with multiplier, and the 20 ml of 0.1 N sodium
these different procedures. carbonate used in the analysis.
The analysis of bomb washings in The formula used to get the total
these methods call for a titration, first using correction in calories is as follows:
0.1N barium hydroxide (V2) followed by V1(Acid Multiplier)(Heat of Formation of
filtering, and a second titration using 0.1N Nitric Acid)V2(Sulfur Multiplier)(Heat of
HCL(V1) after 20 ml of a 0.1N sodium Formation of Sulfuric Acid)+offset value.
carbonate has been added to the filtrate.
Table B-1 gives the settings which allows The values for fixed acid and sulfur,
the results of the two titrations, V1 and V2, which are used in preliminary reports, will
to be entered into the controller directly for reflect a sulfur correction of 0, and a nitric
the calculation of the total acid correction. acid correction of 10 calories.
V1 should be entered at the prompt for acid
and V2 is entered at the prompt for sulfur.
B-666
6200 Calorimeter Operating Instruction Manual
APPENDIX B
It is sometimes necessary to add a Ms = Mass of spiking material Spiking
spiking material to samples which are very Samples
small, have a low heat of combustion, or This factor is added to the
have a high moisture content to add calculations when Spike Controls, Use
sufficient heat to drive the combustion to Spiking is set to ON. Heat of Combustion of
completion. Benzoic acid is an excellent Spike is entered as calories per gram. The
material for spiking for all of the same controller will prompt the user to enter the
reasons it is a good standard material. weight of spiking material. Fixed spikes can
White oil is also an excellent material, be used when, Use Fixed Spike is set to ON
particularly for liquid samples. The 6200 and entering the mass of the spike on -
Calorimeter can automatically compensate Weight of Fixed Spike.
for the addition of spiking materials to these
samples. The calculations are modified in
these cases as follows:
The calculations described above give moisture free or other basis by Conversion
the calorific value of the sample with determining the moisture content to Other
moisture as it existed when the sample was of the air-dry sample and using Moisture
weighed. For example, if an air-dried coal conversion formulae published in Bases
sample was tested, the results will be in ASTM Method D3180 and in other
terms of heat units per weight of air-dry references on fuel technology.
sample. This can be converted to a
The calorific value obtained in a bomb latent heat from the gross calorific Conversion
calorimeter test represents the gross heat of value is therefore an important to Net
combustion for the sample. This is the heat figure in power plant calculations. Heat of
produced when the sample burns, plus the If the percentage of hydrogen H, in Combustion
heat given up when the newly formed water the sample is known, the net heat
vapor condenses and cools to the of combustion, Hnet Btu per pound
temperature of the bomb. In nearly all can be calculated as follows:
industrial operations, this water vapor
escapes as steam in the flue gases and the Hnet = 1.8Hc - 92.7H
latent heat of vaporization, which it contains, (Solid fuels, ASTM D2015)
is not available for useful work. The net heat Hnet = 1.8Hc - 91.23H
of combustion obtained by subtracting the (Liquid fuels, ASTM D240)
67
B-7
6200 Calorimeter Operating Instruction Manual
APPENDIX B
This page intentionally left blank.
B-8
68
6200 Calorimeter Operating Instruction Manual
STANDARDIZATION APPENDIX C
Standardizing bomb washings are titrated to determine the
the The Energy Equivalent Factor. nitric acid correction. The energy equivalent
Calorimeter The term standardization, as is computed by substituting the following
used here, denotes the operation of the equation:
calorimeter on a standard sample from Hm + e1 + e2 + e3
which the energy equivalent or effective W=
heat capacity of the system can be T
determined. The energy equivalent, W or Where:
EE of the calorimeter, is the energy required
to raise the temperature one degree, usually W = Energy equivalent of the
expressed as calories per degree Celsius. calorimeter in calories per degree
Standardization tests should be repeated Celsius.
after changing any parts of the calorimeter, H = Heat of combustion of the
and occasionally as a check on both the standard benzoic acid sample in
calorimeter and operating technique. calories per gram.
m = Mass of the standard benzoic
Standardization Procedure. acid sample in grams.
The procedure for a standardization test T = Temperature rise in C.
is exactly the same as for testing a fuel e1 = Correction for heat of
sample. Use a pellet of calorific grade formation of nitric acid in calories.
benzoic acid weighing not less than 0.9 nor e2 = Correction for sulfur which is
more than 1.1 grams. The corrected usually 0.
temperature rise, T, is determined from the e3 = Correction for heating wire and
observed test data and the combustion of cotton thread.
C-1
69
6200 Calorimeter Operating Instruction Manual
APPENDIX C
The 6200 Calorimeter includes a provision Any outliers or other tests which Automatic
for calculating and using a mean energy should not be included in the average Statistical
equivalent for each of up to 4 separate bomb EE Value must be deleted from the
Calculations
and bucket combinations. ASTM procedures memory using the memory
recommend that the energy equivalent be management procedures (see Chapter 8).
determined by averaging ten tests. The 6200 A list of all tests associated with any Cal ID
Calorimeter automatically determines and can be printed from the Calibration Data
uses up to 10 tests in its memory and will Page using Print Standardization Runs.
update the EE Value as additional
standardizations are run. Only Final Tests will The user can elect to have any number
be used in determining and updating EE of stored standardization runs used in
values. These values, the number of tests, determining the EE value by entering this
and the relative standard deviation for the number on Calibration Data & Controls
tests used in determining the EE value are Page - Calibration Run Limit.
stored in the Calibration Data Page under the
EE Value for each bomb. EE Max Std Deviation on this same
page establishes the maximum allowable
The user can chose to turn off the standard deviation for the EE Value before
automatic averaging and updating procedure an error condition is reported. The default
and protect the EE Values by turning ON the value is zero which turns off this limit. But
protection feature for the appropriate bomb on the user should enter a value appropriate
the Calibration Data and Control Page using for the test being made.
Protected EE Value.
Calorimeter control limits when benzoic acid is used as a test sample
Accepted heat of combustion taken as 26454 J/g.
Instrument precision 0.10% RSD.
Control limits based on 99% confidence (3 sigma) values. Values are in J/g.
NUMBER OF UCL FOR THE RANGE UCL FOR THE RSD MAXIMUM PERMISSIBLE DEVIATION
OBSERVATIONS IN A (HIGH LOW) WITHIN THE WITHIN THE GROUP OF THE GROUP MEAN FROM THE
GROUP GROUP ACCEPTED VALUE OR GRAND MEAN
1 79.4
2 97.5 0.2606 56.1
3 115.3 0.2276 45.8
4 124.3 0.2088 39.7
5 130.1 0.1964 35.5
6 134.3 0.1874 32.4
7 137.6 0.1806 30.0
8 140.4 0.1751 28.1
9 142.7 0.1707 26.5
10 144.7 0.1669 25.1
11 146.4 0.1637 23.9
12 147.9 0.1610 22.9
13 149.4 0.1585 22.0
14 150.7 0.1563 21.2
15 151.8 0.1544 20.5
16 153.0 0.1526 19.8
17 154.0 0.1511 19.2
18 154.9 0.1496 18.7
19 155.8 0.1483 18.2
20 156.7 0.1470 17.7
21 157.4 0.1459 17.3
22 158.2 0.1488 16.9
23 158.9 0.1438 16.5
24 159.5 0.1429 16.2
25 160.2 0.1420 15.9
70
6200 Calorimeter Operating Instruction Manual
APPENDIX D COMMUNICATION INTERFACES
Printer RS232C Connections be changed to achieve this match. Printer
Port Port Communications for the 6200
The 6200 Calorimeter is also equipped
Controller are on the Printer Port
with an RS232C port for connection to
Communications Page.
either a 40 or 80 column printer and/or a
balance. Before making either of these The default parameters for the 6200
connections the data transmission rate of Calorimeter are set up for use with the Parr
the calorimeter and the printer or balance 1757 Printer. Table D-1 on page D-3
must be matched. Generally the baud rates identifies and describes the pin-out for the
on either device can RS232C port.
71
D-1
6200 Calorimeter Operating Instruction Manual
APPENDIX D
Sartorious
Field Length [ESC] P [CR] [LF] Balance
polarity 1 Interface
space 1 Note:
data 8 The automatic data output option
space 1 should not be used.
stability 2
CR 1 The calorimeter will send this command
LF 1 string once every few seconds after the
ENTER key has been pressed during a
mass entry sequence. The ENTER key
The polarity field must contain either a
should only be pressed when the mass
+ or a space. Leading zeros in the data
reading is stable. However, unstable
field are blanked, except for the one to the
readings will be rejected and a warning will
left of the decimal point. The stability field
be issued. Acknowledging the warning will
must contain g_ for the calorimeter to
re-issue the command string to the balance
accept a mass. The balance should be
on a periodic basis.
configured to transmit data upon receipt of
the following command string:
72
D-2
6200 Calorimeter Operating Instruction Manual
APPENDIX D
Generic Field Length (CR). Leading zeros may be blanked as
Interface data 9 spaces and are counted. Non-numeric
CR 1 characters are ignored and will reset the
input buffer if the data field has not been
The data field should consist of 9 filled. Any characters received after filling
numeric characters (0 through 9, +, - and the data field and before the carriage return
space) terminated with a carriage return are ignored.
Table D-1
6200 Calorimeter Serial Ports Pin-Out
There are three RS-232 serial ports at the rear of the calorimeter. These ports are designated
Terminal, Printer and Balance. The pin-out of these three ports are identical. The pin-out is
illustrated in the following table.
Both the terminal and balance port are female. The printer port uses a male connector.
D-3
73
6200 Calorimeter Operating Instruction Manual
APPENDIX D
Table D-2
6200 Calorimeter Run Data Template
Field Description
SampleID Field will hold ID up to sixteen characters in length
Timestamp MM/DD/YY HH:mm:ss
Mode 0 = determination, 1 = standardization
Method 0 = equilibrium, 1 = dynamic
State 0 = preweigh, 1 = preliminary, 2 = final
Units 0 = MJ/kg, 1 = Btu/lb, 2 = cal/g, 3 = J/kg, 4 = other
UnitMultIfOther unit multiplier in effect at time of report
BombID [1-4]
BombEE bomb energy equivalent
SampleWt sample weight
SpikeWt spike weight
Fuse fuse value
FuseFinal fuse value is final
Acid acid value
AcidFinal acid value is final
Sulfur sulfur value
SulfurFinal sulfur value is final
Hydrogen hydrogen value (net calc option)
HydrogenFinal hydrogen value is final (net calc option)
Moisture moisture value (dry calc option)
MoistureFinal moisture value is final (dry calc option)
JacketTemp jacket temperature
InitTemp initial temperature
DeltaT temperature rise
HOC gross heat of combustion
NetHOC dry net HOC (if both dry and net calc options enabled)
DryHOC dry gross HOC (if dry calc option enabled)
DryNetHOC dry net HOC (if both dry and net calc options enabled)
D-4
74
6200 Calorimeter Operating Instruction Manual
APPENDIX D
Ethernet Calorimeter test data can be by connecting a router with DHCP server
Interface transferred to an Ethernet network capability to the calorimeter using an
connected computer using the FTP File ordinary CAT 5 network cable. The
Transfer Protocol. First, you must know calorimeter should be connected to LAN
the IP address of the network-connected side of the router. The PC in turn is also
calorimeter. The network DHCP connected to the LAN side of the router
(Dynamic Host Configuration Protocol) using a similar CAT 5 cable. A D-Link 614+
server provides this address shortly after router is recommended for this purpose. For
the calorimeter is turned on. The address this router, operated without a WAN
can be seen on the software and connection, the primary DNS address of the
hardware info page, under program router (WAN setup) must be set to the IP
information and control. See the address of the router found on the LAN
example screenshot. setup page. Other routers behave differently
in the absence of a WAN connection.
Providing an active upstream connection to
the WAN port of most routers generally
minimizes the use of any obscure setup
configurations.
ftp://root:rootroot@192.168.0.125/../flash/data/
D-5
75
6200 Calorimeter Operating Instruction Manual
APPENDIX D
The following screenshot illustrates the contents of the calorimeter data directory as presented
by a web browser.
You can drag and drop or copy and paste test data files (with the csv suffix) from the web
browser window to any convenient folder or directory on the PC.
D-6
76
6200 Calorimeter Operating Instruction Manual
APPENDIX D
The calorimeter offers a web server service. Test reports can be viewed with a web browser
using a URL of the following form.
http://192.168.0.125
Where 192.168.0.125 is the IP address of the calorimeter. The following screenshot illustrates
the calorimeter home page.
6200 Calorimeter
D-7
77
6200 Calorimeter Operating Instruction Manual
APPENDIX D
Clicking on the Sample Data tab displays a list of reports currently in the instrument memory.
6200 Calorimeter
D-8
78
6200 Calorimeter Operating Instruction Manual
APPENDIX D
Clicking on any given report will provide a display as follows.
6200 Calorimeter
79D-9
6200 Calorimeter Operating Instruction Manual
APPENDIX D
Bar Code
The use of barcodes in the laboratory Key and enter the feature key Port
has become a highly accurate, rapid and purchased from Parr Instrument
inexpensive way to identify samples. When Company into the instrument by using the
purchasing this feature, the user must touchpad. Pressing the key labeled ABC
supply Parr with the MAC address of the allows the user to switch from upper case
calorimeter (found in the Software & letters, to lower case letters and finally to
Hardware Info menu screen). This allows numerals.. A CD containing all the
Parr to activate the feature key. necessary documentation and setup
information for using both the scanner and
In order to enable the calorimeter to use
the printer is provided at the time of
the bar code feature, the feature key needs
purchase. A PC based program used for
to be entered into the instrument. Select
printing bar coded labels is also provided on
the Program Information and Control key
from the Main Menu. Next, select Feature this CD.
These keys allow the user to specify the Interface devices are polled from device Network
IP addresses of one or more Balance 1 to 15 for sample and / or spike Data
Interface devices on the network. Balance weights when the weight entry mode is Devices
set to Network.
D-10
80
6200 Calorimeter Operating Instruction Manual
APPENDIX E TECHNICAL SERVICE
Should you need assistance in the Return for Repair
operation or service of your instrument, To return the instrument for repair,
please contact the Technical Service please call the Technical Service
Department. Department for shipping instructions and a
RETURN AUTHORIZATION NUMBER.
Telephone: (309) 762-7716 This number must be clearly shown on the
Toll Free: 1-800-872-7720 outside of the shipping carton in order to
Fax: (309) 762-9453 expedite the repair process.
Email: parr@parrinst.com
If you have not saved the original carton
Any correspondence must include the and traps, please request an A1341DD
following basic information: packaging return kit.
E-1
81
6200 Calorimeter Operating Instruction Manual
APPENDIX E
This page intentionally left blank.
E-2
82
6200 Calorimeter Operating Instruction Manual
APPENDIX F 6200 CALORIMETER PARTS LIST
F-183
6200 Calorimeter Operating Instruction Manual
APPENDIX F
Parts List for Temperature Control Assembly
Item Description
1281DD Manifold, Temperature Control
1417E Thermistor, Jacket
538VB Male Connector, 1/8 NPTM-T-BT Nylon
252HWHJ Elbow, Hose Barb, 1/2 x 3/8 M
280HWHJ Elbow, Hose Barb, 1/4 T X3/8 M
A1275DDEB Cartridge, Heater Assembly 120V
A1275DDEE Cartridge, Heater Assembly 240V
535VB Male Connector, 1/4T x 3/8 NPTM
337VB Male Connector, 1/4T x 1/8 NPTM
283VB Adapter, Male 1/4T x 1/8 NPTM
A1276DD Cold Water Solenoid Assembly w/ connector
117HW3 Elbow, Male, 1/4T x 1/8 NPTM
F-2
84
6200 Calorimeter Operating Instruction Manual
APPENDIX F
F-385
6200 Calorimeter Operating Instruction Manual
APPENDIX F
Figure F-1
6200 Oxygen Bomb Calorimeter Cutaway Front
F-4
86
6200 Calorimeter Operating Instruction Manual
APPENDIX F
Figure F-2
6200 Oxygen Bomb Calorimeter Cutaway Rear
F-587
6200 Calorimeter Operating Instruction Manual
APPENDIX F
Figure F-3
A1279DD Control Schematic
F-6
88
6200 Calorimeter Operating Instruction Manual
APPENDIX F
Figure F-4
Oxygen Solenoid Assembly & Fittings
F-789
6200 Calorimeter Operating Instruction Manual
APPENDIX F
Figure F-5
Water Tank Assembly
F-8
90
6200 Calorimeter Operating Instruction Manual
APPENDIX F
Figure F-6
A1311DD Circulating Pump Assembly with Fittings
F-991
6200 Calorimeter Operating Instruction Manual
APPENDIX F
Figure F-7
Temperature Control Assembly with Fittings
F-10
92
6200 Calorimeter Operating Instruction Manual
APPENDIX F
Figure F-8
A1284DD2 Stirrer Hub Assembly
F-11
93
6200 Calorimeter Operating Instruction Manual
APPENDIX F
Figure F-9
Stirrer Motor Assembly
F-12
F-12
94
PARR INSTRUMENT COMPANY
http://www.parrinst.com parr@parrinst.com
Revision 01/29/10