P/N: 01.54.20094-16
Statement
This manual will help you understand the operation and maintenance of the product better. It is
reminded that the product shall be used strictly complying with this manual. Users operation
failing to comply with this manual may result in malfunction or accident for which EDAN
INSTRUMENTS, INC. (hereinafter called EDAN) can not be held liable.
EDAN owns the copyrights of this manual. Without prior written consent of EDAN, any
materials contained in this manual shall not be photocopied, reproduced or translated into other
languages.
Materials protected by the copyright law, including but not limited to confidential information
such as technical information and patent information are contained in this manual, the user shall
not disclose such information to any irrelevant third party.
The user shall understand that nothing in this manual grants him, expressly or implicitly, any
right or license to use any of the intellectual properties of EDAN.
EDAN holds the rights to modify, update, and ultimately explain this manual.
The electrical installation of the relevant room complies with national standards, and
Upon request, EDAN may provide, with compensation, necessary circuit diagrams, and other
information to help qualified technician to maintain and repair some parts, which EDAN may
define as user serviceable.
I
Terms Used in this Manual
This guide is designed to give key concepts on safety precautions.
WARNING
A WARNING label advises against certain actions or situations that could result in personal
injury or death.
CAUTION
A CAUTION label advises against actions or situations that could damage equipment, produce
inaccurate data, or invalidate a procedure.
NOTE
II
Table of Contents
Chapter 1 Warranty and Service.................................................................................................. 1
7.1 Cleaning................................................................................................................................ 46
7.1.1 Cleaning the Main Unit and the Patient Cable............................................................... 46
7.1.2 Cleaning the Electrodes.................................................................................................. 46
7.1.3 Cleaning the Print Head ................................................................................................. 46
III
7.2 Disinfection .......................................................................................................................... 47
7.3 Care and Maintenance .......................................................................................................... 47
7.3.1 Recharge and Replacement of Battery ........................................................................... 47
7.3.2 Recorder Paper ............................................................................................................... 48
7.3.3 Maintenance of Main Unit, Patient Cable and Electrodes ............................................. 49
IV
SE-3 Electrocardiograph Service Manual Warranty and Service
EDAN provides a one-year-warranty for the warranted products (accessories are included). The
warranty period begins on the date the products are shipped to customers. If a customer promptly
notifies EDAN of customers warranty claim hereunder, EDAN will either repair, adjust or
replace (with new or exchange replacement parts) EDANs products. EDAN warrants that any
service it provides to customers will be performed by trained individuals in a workmanlike
manner.
Limitation of Warranty
Direct, indirect or final damage and delay caused by the following situations for which EDAN is
not responsible may void the warranty:
Service Procedure
Fill in the SCF with detailed information including: Model Name, Serial Number (SN) and
Problem Phenomena.
EDAN should not have any obligation to take over the case without this information. The
form can be downloaded at: http://www.edan.com.cn or obtained from EDANs service
department.
Once the service department receives the fully filled SCF, EDANs engineer will offer a
solution in three working days. EDAN will follow out the case based on the two conditions
below:
-1-
SE-3 Electrocardiograph Service Manual Warranty and Service
Within Warranty:
ii) The customer signs the Declaration Form and sends it back by email or fax. This form is
legally certificated to make sure the customer or the end-user will return the defective parts to
EDAN on time. We will, at this option, dispatch the replacement one(s) with the confirmed
shipping invoice.
NOTE:
(1) Both Return Material Authorization Form and Declaration Form are offered
by EDAN service department once the SCF is confirmed by service
engineers.
(2) The customer is responsible for freight & insurance charges when the
equipment is shipped to EDAN for service, including custom charges. EDAN
is responsible for the freight, insurance & custom charges from EDAN to the
customer.
Out of Warranty:
After receiving the RMA form from the service department, the customer sends defective
parts to EDAN in advance. We will analyze the problems and discuss with the customer about
either repairing or replacing the part(s). Once the maintenance fee is invoiced and paid, we
will make sure to dispatch good part(s) to the confirmed address.
NOTE: The customer is responsible for any freight & insurance charge for the
returned product.
-2-
SE-3 Electrocardiograph Service Manual Warranty and Service
NOTE: EDAN should not have any obligation to the end-user or the customer
who returns the goods without the notification by EDANs service
department. The sender takes full responsibility for the accounted fee.
Please disassemble the parts with anti-static facility, do not touch the parts with naked hands.
Please describe the returned parts as sample of ***** and put the total value on the invoice,
and note on the invoice as sample, no commercial value.
Contact Information
If you have any question about maintenance, technical specifications or malfunctions of devices,
do not hesitate to contact us.
E-mail: support@edan.com.cn
-3-
SE-3 Electrocardiograph Service Manual Safety Guidance
External output
External input
Caution
Potential equalization
Mains supply
On (mains supply)
Battery indicator
-4-
SE-3 Electrocardiograph Service Manual Safety Guidance
Recall key
Print/Stop key
ON/OFF key
Menu key
Recycle
Part Number
-5-
SE-3 Electrocardiograph Service Manual Safety Guidance
Serial Number
Date of Manufacture
Manufacturer
Classification:
Degree of safety of application in the Equipment not suitable for use in the presence
-6-
SE-3 Electrocardiograph Service Manual Safety Guidance
Environment:
-7-
SE-3 Electrocardiograph Service Manual Safety Guidance
WARNING
1. The service provider must read and understand all the information presented in the
user manual and this service manual before installing or servicing a unit.
2. Only qualified service engineers can install this equipment, and only service engineers
authorized by the manufacturer can open the shell.
3. Only qualified installation or service engineers can shift the mains supply shift switch
(100V-115V/220V-240V) according to local mains supply specifications.
6. SHOCK HAZARD - The power receptacle must be a hospital grade grounded outlet.
Never try to adapt the three-prong plug to fit a two-slot outlet.
7. If the integrity of the external protective conductor is in doubt, the equipment should be
operated by using the built-in rechargeable battery.
8. Do not use this equipment in the presence of high static electricity or high voltage
equipment which may generate sparks.
9. Only the patient cable and other accessories supplied by the manufacturer can be
used. Or else, the performance and electric shock protection can not be guaranteed.
10.Ensure that the conductive parts of electrodes and associated connectors, including
neutral electrodes, do not come in contact with earth or any other conducting objects.
12.Do not touch the patient, bed, table or the equipment while using the ECG together
with a defibrillator.
-8-
SE-3 Electrocardiograph Service Manual Safety Guidance
WARNING
13.In order to avoid being burned, please keep the electrodes far away from the radio
knife while using electrosurgical equipment.
14.If reusable electrodes with electrode gel are used during defibrillation, ECG recovery
will take more than 10 seconds. The manufacturer recommends the use of disposable
electrodes at all times.
15.Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing
equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configuration
shall comply with the valid version of the standard IEC/EN 60601-1-1. Therefore
anybody, who connects additional equipment to the signal input or output connector to
configure a medical system, must make sure that it complies with the requirements of
the valid version of the system standard IEC/EN 60601-1-1. If in doubt, consult our
technical service department or your local distributor.
16.The summation of leakage current should never exceed leakage current limits while
several other units are used at the same time; otherwise, electric shock may happen.
WARNING
1. Improper operation may cause lithium battery (hereinafter called battery) to be hot,
ignited or exploded, and it may lead to the declination of the battery capacity. It is
necessary to read the user manual carefully and pay more attention to warning
messages.
2. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery. The battery of the same model and specification
provided by the manufacturer should be used.
-9-
SE-3 Electrocardiograph Service Manual Safety Guidance
WARNING
3. Danger of explosion -- Do not reverse the anode and the cathode when installing the
battery.
5. When leakage or foul smell is found, stop using the battery immediately. If your skin or
cloth comes into contact with the leakage liquid, cleanse it with clean water at once. If
the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with clean
water first and go to see a doctor immediately.
6. Remove the battery from the electrocardiograph when the electrocardiograph is not
used for a long time.
7. If the battery is stored alone and not used for a long time, we recommend that the
battery should be charged at least once every 6 months to prevent overdischarge.
CAUTION
1. Avoid liquid splash and excessive temperature. The temperature must be kept
between 5 C and 40 C during operation, and it should be kept between -20 C and
55 C during transportation and storage.
2. Do not use the equipment in a dusty environment with bad ventilation or in the
presence of corrosive.
3. Make sure that there is no intense electromagnetic interference source around the
equipment, such as radio transmitters, mobile phones etc. Attention: large medical
electrical equipment such as electrosurgical equipment, radiological equipment and
magnetic resonance imaging equipment is likely to bring electromagnetic interference.
4. Before use, the equipment, the patient cable and electrodes should be checked.
Replacement should be taken if there is any evident defectiveness or aging symptom
which may impair the safety or the performance.
5. The following safety checks should be performed at least every 24 months by a
qualified person who has adequate training, knowledge, and practical experience to
perform these tests.
- 10 -
SE-3 Electrocardiograph Service Manual Safety Guidance
CAUTION
a) Inspect the equipment and accessories for mechanical and functional damage.
c) Inspect the fuse to verify compliance with the rated current and breaking
characteristics.
d) Verify that the device functions properly as described in the instructions for use.
e) Test the protection earth resistance according to IEC/EN 60601-1: Limit 0.1 ohm.
f) Test the earth leakage current according to IEC/EN 60601-1: Limit: NC 500 A,
SFC 1000A.
- 11 -
SE-3 Electrocardiograph Service Manual Safety Guidance
CAUTION
1. Turn off the power before cleaning and disinfection. If the mains supply is used, the
power cord should be dragged out of the outlet. Prevent the detergent from seeping
into the equipment during cleaning.
2. Do not immerse the unit or the patient cable into liquid under any circumstances.
3. Do not clean the unit and accessories with abrasive fabric and avoid scratching the
electrodes.
4. Any remainder of detergent should be removed from the unit and the patient cable
after cleaning.
- 12 -
SE-3 Electrocardiograph Service Manual Operating Principle
The heart is a power organ of the blood circulation. Before the systole or the diastole, a cardiac
impulse happens in the heart muscle, and a faint bioelectric signal is thus generated. The
bioelectric signal is transmitted through the whole body, and the potential difference is generated
on the different skin surfaces because of the different distances from the heart.
The cardiogram is a record of the amplification of the potential distribution on the body skin
surface. The potential difference is sampled by the electrodes, and amplified and processed by the
electrocardiograph. Then it is recorded on the paper. The cardiogram recorded by the
electrocardiograph can help doctors to analyze and diagnose heart disease. The intended use of
the electrocardiograph is to acquire ECG signals from adult and pediatric patients through body
surface ECG electrodes. The electrocardiograph only records the heart's electrical activity, and
does not produce any electricity of its own. The test does not hurt and has no known side effects.
It does not require any preparations except possibly shaving chest hair to get a better recording.
The recording itself takes only a few seconds.
It is more than one hundred years since the electrocardiograph was applied in clinical diagnosis.
The electrocardiograph is an important measurement in clinical diagnosis of heart disease, and is
equipped in almost every hospital and clinic.
- 13 -
SE-3 Electrocardiograph Service Manual Operating Principle
3.3 Lead
The 12-lead ECG provides spatial information about the heart's electrical activity in three
approximately orthogonal directions.
Right Left
Superior Inferior
Anterior Posterior
- 14 -
SE-3 Electrocardiograph Service Manual Operating Principle
Each lead represents a particular orientation in space, as indicated below (RA = Right Arm, LA =
Left Arm, LL = Left Leg):
Bipolar limb leads (frontal plane):
Lead I: RA (-) to LA (+) (right left, or lateral)
Lead II: RA (-) to LL (+) (superior inferior)
Lead III: LA (-) to LL (+) (superior inferior)
Augmented unipolar limb leads (frontal plane):
Lead aVR: RA (+) to [LA & LL] (-) (rightward)
Lead aVL: LA (+) to [RA & LL] (-) (leftward)
Lead aVF: LL (+) to [RA & LA] (-) (inferior)
Unipolar (+) chest leads (horizontal plane):
Leads V1, V2, V3: (posterior anterior)
Leads V4, V5, V6: (right left, or lateral)
- 15 -
SE-3 Electrocardiograph Service Manual Operating Principle
- 16 -
SE-3 Electrocardiograph Service Manual Operating Principle
- 17 -
SE-3 Electrocardiograph Service Manual Operating Principle
This module is a sampling and processing module of 12-channel ECG signals. It can amplify and
preprocess ECG signals in every channel, detect lead situations, and send data to ARM9 MCU.
This module is connected to ARM9 MCU through the UART port.
The electrocardiograph has 10 lead cables. Nine lead cables input ECG signals and one right foot
lead cable provides the float ground. The ECG signals pass through the defibrillator protection
circuit, the buffering circuit (U1~U3), and enter the difference amplifier circuit. Then they
continue to pass through the pacemaker pulse restraining, high-pass filter, low-pass filter, channel
switch, second level amplification, and enter the A/D sampling port.
Because the ECG system must meet the safety requirement of the photo-electricity isolation, the
serial signals will be sent to the photo-electricity component.
- 18 -
SE-3 Electrocardiograph Service Manual Operating Principle
- 19 -
SE-3 Electrocardiograph Service Manual Operating Principle
The AC voltage passes the linear transformer and is transformed into 20V AC voltage. One part is
provided to charge the battery, and the other part passes DC/DC to provide +8V for the thermal
print head, +5V for the electric machine, and +5V for the main board and the ECG board.
- 20 -
SE-3 Electrocardiograph Service Manual Functional Verification
WARNING
Only qualified service personnel should perform a full functional checkout procedure.
1. Equipment Requirement
Equipment Model
ECG simulator FLUKE MPS450
ECG detector NMI EGC-1C
- 21 -
SE-3 Electrocardiograph Service Manual Functional Verification
4. Key Tests
Press every key on the key panel to test whether they work as described in the user manual.
2) Make the ECG detector output sine waves with the amplitude of 2mV and the frequency of
10Hz. Press the PRINT/STOP key to begin to print ECG reports. Examine whether the printed
contents are integrated and consistent with the information displayed on the LCD screen. The
amplitude of the ECG waveform of lead I should be 20mm.
3) The ECG report includes date and time, ID, name, sex, age, sensitivity, paper speed, filter, lead,
1mV calibration mark, ECG waveform, heart rate etc. 1mV calibration mark and lead names are
printed before ECG waves in the ECG reports. The printed characters and waveforms are clear.
4) Open the recorder casing and remove the recorder paper, and then the hint information Paper?
should be displayed. Install the recorder paper again and close the recorder casing, and then the
hint information Paper? should disappear.
- 22 -
SE-3 Electrocardiograph Service Manual Functional Verification
1) Test the protection earth resistance according to IEC/EN 60601-1: Limit: 0.1 ohm.
2) Test the earth leakage current according to IEC/EN 60601-1: Limit: NC 500 A, SFC 1000A.
3) Test the enclosure leakage current according to IEC/EN 60601-1: Limit: NC 100A, SFC
500A.
4) Test the patient leakage current according to IEC/EN 60601-1: Limit: NC a.c. 10A, d.c. 10A;
SFC a.c. 50A, d.c. 50A.
5) Test the patient auxiliary current according to IEC/EN 60601-1: Limit: NC a.c. 10A, d.c.
10A; SFC a.c. 50A, d.c. 50A.
6) Test the patient leakage current under single fault condition with mains voltage on the applied
part according to IEC/EN 60601-1: Limit: 50A (CF).
- 23 -
SE-3 Electrocardiograph Service Manual Modules Malfunction Verification
1) Open the main unit following the procedures described in Section 6.1, Disassembly Steps.
3) Measure the voltage to earth of the test points listed below by using a multimeter.
T3 18V~24V
T5 +5V0.25V
T7 +5V0.25V
T2 18V~24V
T6 +5V0.25V
T4 +8V0.5V
- 24 -
SE-3 Electrocardiograph Service Manual Modules Malfunction Verification
T3 T5 T7
T2 T6 T4
4) Compare the measurement results with the reference results in the list. If any one of the
results exceeds the reference range, the main board defection is confirmed. Replacement of
the main board is recommended.
1) Make sure that the malfunction of ECG measurement is not caused by other defects, such as
defective connection, inoperative electrodes or main board, etc.
2) Open the main unit following the procedures described in Section 6.1, Disassembly Steps.
- 25 -
SE-3 Electrocardiograph Service Manual Modules Malfunction Verification
4) Measure the voltage to earth of the test points listed below by using a multimeter.
-VCC -7.5V
+VCC 7.5V
+5V 5.0V0.15V
+3.3V 3.3V0.1V
GND 0V
+2.5V 2.5V0.05V
+5V
+3.3V
GND
5) Compare the measurement results with the reference results in the list. If any one of the results
exceeds the reference range, the ECG board defection is confirmed. Replacement of the ECG
board is recommended.
- 26 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
1. Main Unit
Remove the crosshead screws (M310) on the bottom of the electrocardiograph. Open the
upper unit and remove the cables between the upper and lower units, and then separate the
upper unit from the lower unit. In the following pictures, is the lower unit, is the recorder
casing.
- 27 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
There are two fuses of the same specifications installed on the bottom of the main unit. The
specifications are shown on the fuse label: AC220V-240V: T200mA; AC100V-115V: T400mA;
520.
WARNING
Ruptured fuses must only be replaced with those of the same type and rating as the
original.
- 28 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
3. Disassembly Steps of the Upper Unit (19264 Dot Single Color LCD Screen)
Note: You should take care not to damage the LCD screen when disassembling it.
The upper unit of the electrocardiograph (19264 dot single color LCD screen) includes the
lower casing of the displaying box, filmy switch panel, LCD screen, rotation base etc.
The disassembly steps:
Remove the cables among the LCD screen , key panel and transfer board . Remove
the five crosshead panhead screws (M36), and dismantle the transfer board Remove
the four crosshead panhead screws (M38), and dismantle the upper casing of the displaying
box to show the LCD screen and the rotation base fixed on the lower casing of the
displaying box. Remove the screws used to fix the LCD screen and dismantle the LCD screen.
Remove the two crosshead sunk screws (M320) to dismantle the lower casing of the
displaying box Remove the four crosshead sunk screws (M36) to dismantle the
rotation base .
The label and the key panel are stuck to the upper unit , and can be torn off.
- 29 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
4. Disassembly Steps of the Upper Unit (320240 Dot Single Color LCD Screen)
Note: You should take care not to damage the LCD screen when disassembling it.
The LCD screen (320240 dot single color LCD screen) is embedded in the upper casing of the
displaying box, while the LCD screen (19264 dot single color LCD screen) is fixed in the lower
casing of the displaying box by using screws.
The following picture shows: crosshead panhead screws M36, the transfer board, the
upper unit, the rotation base, crosshead panhead screws M38, the lower casing of the
displaying box, crosshead sunk screws M36, crosshead sunk screws M320, the LCD
screen, the filmy switch panel, the upper casing of the displaying box, the label,
the protection screen.
- 30 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
Remove the cables between two parts before disassembly. Remove the screws on the ECG board
and dismantle the ECG board. Remove the main board which is embedded in the groove of
the lower unit. Remove the two crosshead panhead screws (M26) and dismantle the print
head . Remove the four crosshead sunk screws (M36) and dismantle the print bracket .
Remove crosshead panhead screws (M38) and dismantle the transformer . is the power
transition board.
- 31 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
- 32 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
- 33 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
Net Chip
(Optional)
- 34 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
5 +5V
- 35 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
2. SE-3 Main Board -- Integrating Power Supply, Input/Output Interfaces and ARM9
Circuit
- 36 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
Pin Description
1, 2 16.8V AC INPUT
Pin Description
6, 7, 25, 26 +8V
2 +5V
15 +3.3V
24 N.C
J9 -- connected to BP3 of the connecting board (Only for the ECG with narrow screen)
Pin Description
4, 40 +5V
Others N.C
J10 -- connected to BP3 of the connecting board (Only for the ECG with wide screen)
Pin Description
- 37 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
10, 40 +3.3V
20 LCD backlight
Others N.C
Pin Description
1 Enable
2 GND
3, 4 UART Signal
5 +5V
- 38 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
Pin Description
1 Battery supply
2 Battery detecting
3 GND
- 39 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
4. Connecting Board
P1 BP2 BP3
Pin Description
2 +5V
1, 20 GND
- 40 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
Pin Description
1 ON/OFF
2 GND
BP3-- connected to J9 of the main board (Only for the ECG with narrow screen)
Pin Description
4, 40 +5V
Others N.C
BP3-- connected to J10 of the main board (Only for the ECG with wide screen)
Pin Description
10, 40 +3.3V
20 LCD Backlight
Others N.C
- 41 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
6.3 Troubleshooting
WARNING
Replace parts, components, or accessories only with parts supplied or approved by the
manufacturer. The use of any other parts can lead to inferior device performance and will
void the product warranty.
This section describes some of the most common symptoms and provides possible causes and
solutions to help you solve potential problems.
When you encounter a problem, address the issue in the following order:
1. Review the troubleshooting section to see whether your issue is addressed.
2. Contact your local distributor.
3. Contact EDAN service department.
EDAN service department can assist with product troubleshooting and provide technical
expertise to help with any issues with SE-3.
- 42 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
2. Display Troubles
- 43 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
3. Transmission Troubles
4. Operation Troubles
- 44 -
SE-3 Electrocardiograph Service Manual Electrocardiograph Servicing
Some characters or
The print head is dirty. Clean the print head
waveforms recorded are
Print head failure Change the print head
missing.
There is a stain on the
detecting position of the Clean the print head
Fail to detect paper print head. Change the print head
Print head failure Change the main board
Main board failure
5. Parameter Troubles
- 45 -
SE-3 Electrocardiograph Service Manual Cleaning, Care and Maintenance
7.1 Cleaning
CAUTION
Turn off the power before cleaning and disinfection. The mains supply must be switched
off if it is in use.
CAUTION
1. Prevent the detergent from seeping into the main unit while cleaning. Do not immerse
the unit or the patient cable into liquid under any circumstances.
2. Do not clean the unit and accessories with abrasive fabric and avoid scratching the
electrodes.
- 46 -
SE-3 Electrocardiograph Service Manual Cleaning, Care and Maintenance
7.2 Disinfection
To avoid permanent damage to the equipment, disinfection can be performed only when it is
considered as necessary according to your hospitals regulations.
Before disinfection, clean the equipment first. Then wipe the surfaces of the unit and the patient
cable with hospital standard disinfectant.
CAUTION
1) Capacity Identification
The battery capacity of the rechargeable battery can be identified according to the battery
symbol in the top right corner of the LCD screen.
: Full capacity
flashes in the top right corner of the LCD screen. When the battery capacity is full, the
symbol stops flashing, and the battery recharging indicator lamp ( ) is black. The
3-channel electrocardiograph can not be recharged when it is printing reports, and the
battery recharging indicator is black; when the 3-channel electrocardiograph is switched off,
the battery recharging indicator lamp ( ) will be black if the battery is fully recharged.
- 47 -
SE-3 Electrocardiograph Service Manual Cleaning, Care and Maintenance
Because of the capacity consumption during the storage and transport course, the battery
capacity is not full when it is used for the first time. Battery recharge should be considered
before the first use.
Note: If the battery has not been used for more than two months, it should be
recharged before use.
3) Replacement
When the useful life of the battery is over, or foul smell and leakage are found, please
contact the manufacturer or the local distributor for replacement.
WARNING
1. Only qualified service engineers authorized by the manufacturer can open the
battery compartment and replace the battery. The battery of the same model and
specification provided by the manufacturer must be used.
2. Danger of explosion -- Do not reverse the anode and the cathode when installing the
battery.
3. Remove the battery from the electrocardiograph when the electrocardiograph is not
used for a long time.
4. If the battery is stored alone and not used for a long time, we recommend that the
battery should be charged at least once every 6 months to prevent overdischarge.
Note: Recorder paper provided by the manufacturer should be used. Other paper may
shorten the life of the thermal print head. The deteriorated print head may lead
to illegible ECG reports and block the advance of the paper.
Storage Requirements:
Recorder paper should be stored in a dry, dark and cool area, avoiding excessive
temperature, humidity and sunshine.
Do not put the recorder paper under fluorescence for a long time.
Make sure that there is no polyvinyl chloride or other chemicals in the storage
environment, which will lead to color change of the paper.
Do not overlap the recorded paper for a long time, or else the ECG reports may
trans-print each other.
- 48 -
SE-3 Electrocardiograph Service Manual Cleaning, Care and Maintenance
The following safety checks should be performed at least every 24 months by a qualified
person who has adequate training, knowledge, and practical experience to perform these
tests.
a) Inspect the equipment and accessories for mechanical and functional damage.
b) Inspect the safety related labels for legibility.
c) Inspect the fuse to verify compliance with the rated current and breaking characteristics.
d) Verify that the device functions properly as described in the instructions for use.
e) Test the protection earth resistance according to IEC/EN 60601-1: Limit 0.1 ohm.
f) Test the earth leakage current according to IEC/EN 60601-1: Limit: NC 500A, SFC
1000A.
g) Test the enclosure leakage current according to IEC/EN 60601-1: Limit: NC 100A,
SFC 500A.
h) Test the patient leakage current according to IEC/EN 60601-1: Limit: NC a.c. 10A, d.c.
10A; SFC a.c. 50A, d.c. 50A.
i) Test the patient auxiliary current according to IEC/EN 60601-1: Limit: NC a.c. 10A,
d.c. 10A; SFC a.c. 50A, d.c. 50A.
j) Test the patient leakage current under single fault condition with mains voltage on the
applied part according to IEC/EN 60601-1: Limit: 50A (CF).
The leakage current should never exceed the limit. The data should be recorded in an
equipment log. If the device is not functioning properly or fails any of the above tests, the
device has to be repaired.
WARNING
Failure on the part of the responsible individual hospital or institution employing this
equipment to implement a satisfactory maintenance schedule may cause undue
equipment failures and possible health hazards.
1) Main Unit
Avoid excessive temperature, sunshine, humidity or dirt.
Put the dustproof coat on the main unit after use and prevent shaking it violently when
moving it to another place.
Prevent any liquid from seeping into the equipment, otherwise the safety and the
performance of the electrocardiograph can not be guaranteed.
- 49 -
SE-3 Electrocardiograph Service Manual Cleaning, Care and Maintenance
2) Patient Cable
Integrity of the patient cable, including the main cable and lead wires, should be
checked regularly. Make sure that it is conductible.
Do not drag or twist the patient cable with excessive stress while using it. Hold the
connector plug instead of the cable when connecting or disconnecting the patient cable.
Align the patient cable to avoid twisting, knotting or crooking in a closed angle while
using it.
Store the lead wires in a big wheel to prevent any people from stumbling.
Once damage or aging of the patient cable is found, replace it with a new one
immediately.
3) Electrodes
Electrodes must be cleansed after use and make sure there is no remainder gel on them.
Keep suction bulbs of chest electrodes away from sunshine and excessive temperature.
After long-term use, the surfaces of electrodes will be oxidized because of erosion and
other causes. By this time, electrodes should be replaced to achieve high-quality ECG
records.
CAUTION
The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer for
recycling or proper disposal.
- 50 -
SE-3 Electrocardiograph Service Manual Renewal Parts
- 51 -