You are on page 1of 6

Health Devices IPM System

INSPECTION AND PREVENTIVE MAINTENANCE


__________________________________________________________________________________________

Procedure No. 461-20010301 (Acceptance)


__________________________________________________________________________________________

Anesthesia Unit Ventilators


Used For:
Anesthesia Unit Ventilators [10-145]

______________________________________________________________________________________

Commonly Used In: Delivery rooms and operating rooms

Scope: Applies to ventilators used to deliver inhalation anesthetic agents during surgical procedures
that require general anesthesia

Risk Level: High

Type Interval Time Required

Major 6 months _____ hours


Minor Not Applicable

Notes:
Interval Note: Inspection and preventive maintenance intervals should be scheduled according to the
manufacturer's recommendations. However, units should have a major inspection at least every six
months. Pre-use checks should be performed before each case by the anesthetist who will be
operating the equipment.
_______________________________________________________________________________________

Overview
The ventilator can be built into the anesthesia
Patients undergoing surgery under general machine or can be a stand-alone unit connected
anesthesia are routinely paralyzed with muscle to the machine by gas tubing and, perhaps,
relaxants to stabilize the surgical field. sensor cables. Some anesthesia ventilators have
Consequently, they are unable to breathe on built-in displays and alarms; others rely on the
their own and must be mechanically ventilated sensors, displays, and alarms of the anesthesia
either manually by the anesthetist, who machine to monitor their performance.
squeezes a reservoir bag in the breathing circuit,
or automatically by an anesthesia ventilator. A In general, an anesthesia ventilator is less
switch valve allows the choice of the method by sophisticated than a critical care ventilator,
which ventilation is to be supported. The having only a control mode of operation, with
anesthesia ventilator is typically turned on and time cycling. (However, there is at least one ICU-
off independently of the switching between type ventilator that can be used to administer
manual and automatic ventilation. inhalation anesthetics.) A pressure limit prevents
exposure of the lungs to excessive pressure.
Anesthesia ventilators use positive pressure to Several other breathing waveshape parameters
inflate a patient's lungs and deliver a prescribed (e.g., inspiratory:expiratory [I:E] ratio, tidal
mixture of gases and vapors to them. This volume, minute volume, flow) are adjustable by
mixture is produced by the anesthesia machine. the operator and controlled by the ventilator.
Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 1 of 6
Procedure No. 461-20010301 Anesthesia Unit Ventilators (Acceptance)

Ventilators designed solely for anesthetic


administration typically do not have
compressors.
Procedure
During extended surgical procedures and
Before beginning an inspection, carefully read
procedures involving open breathing circuit
this procedure and the manufacturer's
configurations, a humidifier may be included in
instruction and service manuals; be sure that
the breathing circuit. Otherwise, a circle system
you understand how to operate the equipment,
with an absorber, along with one-way inspiratory
the significance of each control and indicator, and
and expiratory valves, is used, typically without a
the alarm capabilities. Also review special
humidifier. The ventilator's pressure-relief and
inspection or preventive maintenance procedures
limit valve(s) should be connected to a waste gas
or frequencies recommended by the
scavenging system.
manufacturer.

Citations Manufacturers' recommended procedures for


(from Health Devices if not otherwise inspection and preventive maintenance of
noted) mechanical anesthesia ventilators vary in both
methods and required accuracy. In addition,
Anesthesia systems [Evaluation], 1988 Jan;17:3. ventilator controls can vary greatly among
Who should service anesthesia equipment [User manufacturers and models. This procedure
Experience Network(TM)], 1988 Feb;17:70. provides the basic framework for complete
Barotrauma from anesthesia ventilators ventilator inspection and preventive
[Hazard], 1988 Nov;17:354. maintenance. Manufacturers' recommended
Damage to plastic components from Loctite procedures should be added where appropriate.
[Hazard], 1989 Jul-Aug; 18:288.
Risk of barotrauma and/or lack of ventilation
with ventilatorless anesthesia machines
[Hazard], 1994 Jan-Feb;23:54.
Qualitative Tasks
Anesthesia Systems [Evaluation], 1996 May-
Jun;25:158+. Chassis/Housing. Check for shipping
Anesthesia Systems [Evaluation update], 1998 damage; report any damage to the
Jan;27:4+. manufacturer, shipper, or service organization,
Carbon monoxide exposures during inhalation and arrange for repair or replacement.
anesthesia: the interaction between halogenated
anesthetic agents and carbon dioxide absorbents Check that the ventilator is suitably constructed
[Hazard] 1998 Nov;27:402-4. to withstand normal hospital use and abuse. For
instance, a unit with venting on the top of the
housing or poorly protected or sealed controls
Test Apparatus, Supplies, and indicators may be prone to fluid entry. (Such
Parts design deficiencies should usually be recognized
during prepurchase evaluation. However, if any
Ground resistance ohmmeter are evident, discuss corrective action with the
manufacturer. If not correctable, warn users or
Leakage current meter or electrical safety take other preventive measures.)
analyzer
Examine the exterior of the ventilator for
Lung simulator with adjustable compliance or cleanliness and general physical condition.
ventilator tester Ensure that plastic housings are intact, that all
assembly hardware (e.g., screws, fasteners) is
Pressure gauge or meter with 2 cm H2O present and tight.
resolution from -20 to +120 cm H2O
Mount/Fasteners. Ensure that the
assembly and weight distribution is stable and
Oxygen analyzer
that the ventilator will not tip over when pushed
or when a caster is jammed on an obstacle (e.g.,
Various breathing circuit adapters
line cord, threshold), as may occur during
transport. If the ventilator is designed to rest on
Additional items as required for specific
a shelf, ensure that it has nonslip legs or
manufacturers' procedures
supports.
Fresh CO2 absorbent (if procedure or
Casters/Brakes. Verify that the correct
investigation results in excessive gas flows
casters have been supplied with the ventilator
through the carbon dioxide absorber)

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 2 of 6
Procedure No. 461-20010301 Anesthesia Unit Ventilators (Acceptance)

(e.g., size, correct swivel). (ECRI recommends 5 Strain Reliefs. Examine the strain reliefs at
in [12.7 cm] diameter casters for mobile both ends of the line cord, if so equipped. Be
ventilators to reduce shock to the unit and to sure that they hold the cord securely.
minimize the effort required to roll the unit
across elevator thresholds and other uneven Circuit Breaker/Fuse. If the ventilator
surfaces.) Verify caster and brake operation.
has a switch-type circuit breaker, check that it
moves freely. If the ventilator is protected by an
AC Plug/Receptacles. A solidly external fuse, verify that the fuse type is labeled
constructed, good quality plug with adequate and that all fuses and spares are the proper
strain relief is acceptable, but the use of a current rating and type. If the value and type are
Hospital Grade plug (identifiable by a green dot not labeled, check the manual for the proper
and/or labeling) will eliminate guesswork and current rating and type and permanently mark
ensure a plug of acceptable construction quality. this information on the ventilator housing near
Right-angle plugs are unacceptable for devices the fuse holder. If no spare fuse is provided,
that are moved frequently. A good quality two- consider attaching a fuse clip and spare fuse,
prong plug is acceptable for double-insulated particularly for high-risk devices.
devices. Replace the plug or have the supplier
replace it if it is not Hospital Grade or otherwise Especially for critical or life-support devices,
suitable. Hospital Grade molded plugs are verify that accessory outlets have independent
acceptable. overcurrent protection (fuse or circuit breaker)
so that a short in a device plugged into the
Examine the AC power plug for damage. Attempt accessory outlet or an accessory overload will
to wiggle the blades to determine if they are not disable the primary device. If this is not
secure. Shake nonmolded plugs and listen for available, then consider labeling the primary
rattles that could indicate loose screws. device to clearly indicate where the ventilator's
fuse or circuit breaker is located, and/or install a
AC Plug. Examine the AC power plug for fused Hospital Grade (or similar quality) plug on
damage, if so equipped. Attempt to wiggle the any commonly used accessories that are not
blades to check that they are secure. Shake the already provided with suitable overcurrent
plug and listen for rattles that could indicate protection.
loose screws. If any damage is suspected, open
the plug and inspect it. Tubes/Hoses. Check the condition of all
tubing and hoses. Be sure that they are not
Line Cord. Ensure that the line cord is long cracked, kinked, or dirty. Check that they are
enough for the ventilator's intended application; connected to the correct locations.
an extension cord should not be required. (A
length of 10 ft [3 m] is suitable for most Cables. Inspect any cables (e.g., for sensors)
applications, although 18 ft [5.5 m] has been and their strain reliefs for general condition.
suggested for operating room equipment.) Carefully examine cables to detect breaks in the
insulation and to ensure that they are securely
The cord should be of suitable quality and gripped in the connectors at each end, which will
current-carrying capacity. Hard Service (SO, ST, prevent rotation or other strain. Where
or STO), Junior Hard Service (SJO, SJT, or SJTO), appropriate, verify that there are no intermittent
or an equivalent-quality cord should be used. faults by flexing cables near each end and
looking for erratic operation or by using an
Verify that the ventilator has adequate protection ohmmeter.
against power loss (e.g., from accidental
disconnection of a detachable power cord, Fittings/Connectors. Verify appropriate
disconnection of the power cord from the wall, or
connectors are supplied if connection to other
depleted battery if a battery-powered device is
hospital equipment or systems is required.
not plugged in). Equipment having a detachable
Ventilators that connect to the central piped
power cord should also have adequate capture
medical gas system should have the matching
devices, cleats, or channels to hold the cord in
DISS or quick-connect fitting for the appropriate
place. If these are absent, request that the
gas. Verify that suitable connectors are supplied
supplier provide suitable means of securing the
with the ventilator so that adapters are not
cord. Verify that the ventilator has adequate
required.
alarms or indicators for line-power loss and
battery depletion and an adequate battery-
Examine all gas and liquid fittings and
charging indicator.
connectors, as well as all electrical cable
connectors, for general condition. Electrical
contacts should be straight, clean, and bright.
Gas and liquid fittings should be tight and should

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 3 of 6
Procedure No. 461-20010301 Anesthesia Unit Ventilators (Acceptance)

not leak. If keyed connectors are used (e.g., pin- Pressure-Relief Mechanism. Check the
indexed gas connectors), ensure that no pins are proper operation of the pressure-relief
missing and that the keying is correct. Keying mechanism, if so equipped, by occluding the
pins should be securely seated in "blind" holes so breathing circuit and observing the resulting
that they cannot be forced in farther. peak pressure on the unit's pressure indicator.
Verify that pressure is vented in the breathing
Filters. Check the condition of gas filters, if circuit.
included in the ventilator.
Absorber. See Anesthesia Units Procedure
Controls/Switches. Verify that software 400.
setup parameters accessible through hidden or
service menus are correctly set for the Fan. Check physical condition and proper
appropriate application and are consistent for all operation, if so equipped.
ventilators. Instruction and service manuals may
contain instructions regarding such modes. If Battery/Charger. Determine the
they do not, contact the manufacturer. Discuss
replacement interval for all batteries and
appropriate settings with the department head
document the interval(s) in an equipment control
and users. If alarm capabilities are included, see
system or a hard copy file for the ventilator. Be
the Acceptance test for Alarms.
sure to include batteries/cells for clocks and/or
memory logs. For critical care monitors and
Examine all controls and switches for physical
therapeutic devices, it may be desirable to
condition, secure mounting, and correct motion.
disconnect the battery and determine if the
If a control has fixed-limit stops, check for
device still operates on line power.
proper alignment, as well as positive stopping.
Operate the ventilator on battery power for
Bellows. Check the physical condition and several minutes to verify that the battery is
proper operation of the bellows. Perform a leak charged and can hold a charge. Activate the
test as follows: Fill the bellows and set the battery test function, if so equipped. Check the
Vent/Bag valve to Bag to trap gas in the bellows. condition of the battery charger, and verify that
There should be no visible deflation after one battery charge indicators function. Provide users
minute. with instructions and/or checklist procedure to
ensure adequate battery charging and
Gas cylinders (and gauges and performance.
regulators, if so equipped). Verify that
these are present and securely mounted. Verify Indicators/Displays. During the course
that one and only one washer is used to seal the of the inspection, confirm the operation of all
tank to its yoke. Verify that all index pins are lights, indicators, meters, gauges, and visual
present and protruding to the proper length to displays on the ventilator and charger (if so
engage the hole in the tank valve stem and in equipped). Be sure that all segments of a digital
the correct positions for the gas to be supplied display function. Record the reading of an hour
through the yoke. meter, if present.

Breathing circuit (including filters). Alarms/Interlocks. Verify that


Verify that these components are compatible configurable alarm features are appropriately set
with the ventilator according to the and consistent among all ventilators. It should
manufacturer's recommendations. Check for not be possible for critical alarms to be turned
leaks, the absence of obstructions, and proper off, silenced, or defeated without adequate
flow direction in the breathing circuit, ensuring warning to the operator or automatic alarm
the proper assembly and function of fittings, reactivation after a short delay. Such deficiencies
adapters, the CO2 absorber, inspiratory and should usually be recognized during prepurchase
expiratory valves and PEEP valves, the APL evaluation. However, if any are found, review the
valve, the scavenger, and other components. justification for purchasing this ventilator and
With the ventilator connected to the anesthesia discuss corrective action with the manufacturer.
system, check for leaks in the entire system, (Alarm features may be optional or
including the breathing circuit. This task does not programmable.) If no remedy is available, a user
have to be duplicated if done as part of the training program should be instituted to reduce
Anesthesia Units Procedure (Procedure 400). the risk of incorrect use. A warning label on the
ventilator or a poster in the area of use may be
Humidifier. See Heated Humidifiers appropriate.
Procedure 431.
Verify that alarms are loud, distinctive, and/or
bright enough to be noticed in the environment

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 4 of 6
Procedure No. 461-20010301 Anesthesia Unit Ventilators (Acceptance)

in which the ventilator will normally be used. If a so that each operates while readings are taken.
remote alarm-indicator is required, verify that it Record the maximum leakage current; it should
is available and functioning. Audible alarm- not exceed 300 A for patient-care equipment
volume controls should not allow the alarm to be used in a patient-care vicinity [NFPA 99 7-
turned off or lowered to an indiscernible volume. 5.1.3.5(a)].

Induce alarm conditions to activate audible and The measurements should be made with all
visual alarms. Check that all associated accessories that are normally powered from the
interlocks or features function (e.g., an infusion same line cord connected and turned on. This
pump initiates KVO rate upon alarm). If the applies to devices that are plugged into
ventilator has an alarm-silence feature, check accessory outlets on the ventilator and to
the method of reset (i.e., manual or automatic) devices that are plugged into a multiple-outlet
against the manufacturer's specifications. strip so that the devices are grounded through a
single line or extension cord.
Labeling. Check that all necessary placards,
labels, and instruction cards are present and Leakage current must be measured with the
legible. ventilator powered by a conventional (grounded)
power system, even if it is normally used in an
Accessories. Verify that all necessary area with isolated power. If the ventilator has a
features and accessories (e.g., transducers) have special plug (e.g., explosion proof), a
been supplied with the ventilator. At least one corresponding adapter is required.
copy each (two are generally preferred) of the
instruction and service manuals, including Though confirmation of grounding integrity
schematics, should be shipped with the unit and provides reasonable assurance of safety for
filed in the central equipment file. A copy of the ventilators with permanent redundant grounding
instruction manual should be kept with the (e.g., a bedside monitor grounded through the
ventilator and read by all operators before the line cord and its central station connection),
ventilator is put in use. NFPA 99 calls for measurement of chassis
leakage current with the redundant ground intact
[NFPA 99 7-5.1.3.5]. Where practical, verify that
these ventilators meet appropriate chassis
Quantitative Tasks leakage current requirements before installation
or connection to ground during acceptance
Grounding Resistance. Using an testing.
ohmmeter, electrical safety analyzer, or
multimeter with good resolution of fractional Be alert for leakage current of the ventilator in
ohms, measure and record the resistance the off mode that is greater than about 30 A
between the grounding pin of the power cord and is greater than or equal to the leakage
and exposed (unpainted and not anodized) metal current in the on mode. Although this may be
on the chassis of the ventilator or of the system normal and proper for the ventilator, it may
in which the ventilator is mounted. We indicate that the on/off switch is incorrectly
recommend a maximum of 0.5 . If the wired in the neutral (instead of the hot) line.
ventilator is a component within an anesthesia Incorrect switch wiring poses a risk to service
unit, grounding and leakage current personnel who believe that the power is
measurements can be referenced to that unit. disconnected when the switch is off. Check the
wiring, or contact the manufacturer.
Chassis Leakage Current. Note: Some
devices (especially devices incorporating a Modes and Settings. Anesthesia
microprocessor, motor, or compressor) may be ventilators are usually equipped only with a
damaged by switching polarity while the device control mode. However, specialized units may
is on. If you perform reverse polarity testing, have additional modes such as assist/control and
turn off the unit until the motor stops or for at pressure support. Adjustable positive end-
least 10 sec before switching polarity. Although expiratory pressure (PEEP) may also be
reversed polarity testing is not required, it may available. The function of these modes should be
be advisable on a ventilator of questionable inspected and verified for proper operation.
quality. Check the operation and accuracy of ventilation
controls, which may include tidal volume, breath
With the polarity of the power line normal and rate, inspiratory time, expiratory time, I:E ratio,
the equipment ground wire disconnected, pressure limit, or flow. Typically, these tests are
measure chassis leakage current with the performed by attaching the ventilator to a lung
ventilator operating in all normal modes, simulator or ventilator tester and comparing
including on, standby, and off. If the unit has measured values to settings on the ventilator.
heating and cooling modes, set the thermostats The manufacturer should recommend the

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 5 of 6
Procedure No. 461-20010301 Anesthesia Unit Ventilators (Acceptance)

appropriate ventilator settings (e.g., tidal


volume, rate, inspiratory time, compliance) to
verify proper operation and accuracy (generally
within 10%).
Before Returning to Use
Ensure that all controls are set properly. Set
Monitors and Alarms. The following alarms loud enough to alert personnel in the
breathing circuit parameters may be monitored area in which the ventilator will be used. Other
by the ventilator or by the system in which the controls should be in their normal pre-use
ventilator is mounted. They should be inspected positions.
for accuracy (generally within 10%) according to
the manufacturer's specifications: Attach a caution tag in a prominent position so
that the user will be aware that control settings
Breathing rate may have been changed.
Inspiratory time
Airway pressure (e.g., PIP, PEEP, MAP, apnea) Recharge battery-powered ventilators, or equip
Volume (e.g., tidal volume, minute volume, them with fresh batteries, if needed.
apnea)
Fraction of inspired oxygen (FIO2; see Oxygen Change absorbent in CO2 absorber (if procedure
Analyzers Procedure 417) or investigation results in excessive gas flows
through the carbon dioxide absorber).
Alarm settings (e.g., high PIP, low MAP, low
pressure, low FIO2) should be tested for proper
and accurate activation.

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 6 of 6

You might also like