Professional Documents
Culture Documents
24 (2006) 717735
The same old Watson! You never learn that the gravest issues may depend
upon the smallest things. d Sherlock Holmes in The Adventure of the
Creeping Man by Sir Arthur Conan Doyle
* Corresponding author.
E-mail address: mark0003@mc.duke.edu (J.B. Mark).
0889-8537/06/$ - see front matter 2006 Elsevier Inc. All rights reserved.
doi:10.1016/j.atc.2006.08.008 anesthesiology.theclinics.com
718 BARBEITO & MARK
Catheter transducer systems with low natural frequency can still be useful
clinically because another mechanical property, damping, has additional in-
uence on the monitored waveforms. Damping describes the absorption of
oscillatory energy by frictional forces. A monitoring system is optimally
damped if it dissipates the energy produced by the components of the mea-
suring system, conducting only the oscillations that correspond to the mon-
itored pressure waveform. Optimal damping is dicult to achieve. Most
monitoring systems are underdamped, but have a natural frequency high
enough so that the eect on the monitored waveform is limited [4]. Fig. 1
shows the relationship between damping and natural frequency of pressure
monitoring systems, and describes the characteristics of typical transducer
systems in clinical use.
When systolic pressure overshoot is observed, it is common practice to
incorporate a small air bubble into the system to increase damping. How-
ever, while air bubbles increase damping, they simultaneously lower the nat-
ural frequency of the system and, paradoxically, can cause artifactual
increases in systolic pressure and artifactual decreases (but much less so)
in diastolic pressure.
Fig. 1. Inuence of natural frequency and damping coecient on the dynamic response of pres-
sure monitoring systems. Cathetertubingtransducer systems in the optimal dynamic response
area will record accurately a wide range of arterial pressure waveforms. Those falling within the
adequate dynamic response range will faithfully reproduce most pressure waveforms, while
those in the overdamped or underdamped regions will have waveform distortions that are char-
acteristic of those technical limitations. Monitoring systems with natural frequencies in the un-
acceptable range should be adjusted to improve performance. Typical clinical monitoring
systems reside in the hatched area, some showing optimal dynamic response, but many being
underdamped. (Adapted from Mark JB. Atlas of cardiovascular monitoring. New York:
Churchill Livingstone; 1998. p. 111; with permission.)
720 BARBEITO & MARK
For the reasons stated above, it is important to know the natural fre-
quency and damping coecients of the monitoring system in use. Although
not completely accurate in vivo [5], the fast ush test is a clinically useful
method that allows the determination of these values through examination
of the artifact that follows a pressurized ush of the monitoring catheter
[6,7]. From this ush artifact, one can calculate with reasonable accuracy
the natural frequency and damping coecient of the monitoring system.
considerably lower than the pressure at the level of the heart in the sitting pa-
tient and will be exactly equal to the pressure measured at heart level minus the
hydrostatic pressure dierence between the heart and brain [9].
Fig. 2. Normal arterial pressure (ART) waveform. (Adapted from Mark JB. Atlas of cardiovas-
cular monitoring. New York: Churchill Livingstone; 1998. p. 94; with permission.)
PRESSURE MONITORING 723
Fig. 3. Arterial pressure (ART) waveform showing systolic pressure variation during positive-
pressure mechanical ventilation. End-expiratory systolic blood pressure (1) serves as a baseline
from which an early inspiratory increase (2, Dup) can be measured, followed by a delayed de-
crease (3, Ddown). The large Ddown and total systolic pressure variation of nearly 30 mm Hg
suggests the diagnosis of hypovolemia, even though tachycardia and hypotension are not pres-
ent. (Adapted from Mark JB. Atlas of cardiovascular monitoring. New York: Churchill Living-
stone; 1998. p. 282; with permission.)
PRESSURE MONITORING 725
Fig. 4. Abnormal arterial pressure (ART) waveforms in various disease states. (A) Pulsus alter-
nans (congestive heart failure). (B) Pulsus bisferiens (aortic insuciency). (C) Pulsus parvus
and tardus (aortic stenosis). (D) Spike-and-dome pattern (hypertrophic cardiomyopathy).
(E) Pulsus paradoxus (cardiac tamponade). The arrows indicate the onset of spontaneous inspi-
ration. (Adapted from Mark JB. Atlas of cardiovascular monitoring. New York: Churchill Liv-
ingstone; 1998. p. 311, 325; with permission.)
PRESSURE MONITORING 727
Fig. 6. CVP changes during positive-pressure mechanical ventilation. Mean CVP increases at
onset of each positive-pressure breath (arrows), but venous return decreases when the measured
CVP increases. (Adapted from Mark JB. Atlas of cardiovascular monitoring. New York:
Churchill Livingstone; 1998. p. 269; with permission.)
730 BARBEITO & MARK
Fig. 7. Normal CVP and its temporal relationship to the ECG. a, a wave; c, c wave; P, P wave;
R, R wave; T, T wave; v, v wave; x, x descent; y, y descent. (Adapted from Mark JB. Atlas of
cardiovascular monitoring. New York: Churchill Livingstone; 1998. p. 20; with permission.)
annulus moves toward the cardiac apex. The former produces the c wave,
which is small and inconsistent, while the latter generates a drop in pressure
in the right atrium, producing the x descent. Ventricular ejection drives ve-
nous return through the vena cavae, which increases right atrial volume and
pressure in mid- to late systole, thereby inscribing the CVP v wave. Ventric-
ular diastole then begins with the opening of the tricuspid valve, and blood
lls the right ventricle, producing the y descent in CVP. Finally, the last part
of diastole is marked by atrial contraction, which raises atrial pressure to its
highest value and produces the a wave.
Fig. 8. Junctional rhythm (right panels) causes retrograde atrial depolarization during ventric-
ular systole. Atrial contraction against the closed tricuspid valve inscribes a tall early systolic
cannon a wave in the CVP trace (asterisk). Loss of atrioventricular synchrony reduces ventric-
ular preload, leading to arterial hypotension. Note that the ECG does not clearly reveal the
rhythm change because the P-wave amplitude is very small. Abbreviations: ART, arterial pres-
sure. R, ECG R wave. (Adapted from Mark JB. Atlas of cardiovascular monitoring. New York:
Churchill Livingstone; 1998. p. 228; with permission.)
Fig. 9. Tricuspid regurgitation inscribes a tall regurgitant c-v wave in the CVP trace with oblit-
eration of the x descent. Mean CVP is markedly increased in TR, and right ventricular end-
diastolic pressure is lower than mean CVP and best estimated before onset of the c-v wave
(arrows). Abbreviations: R, ECG R wave. (Adapted from Mark JB. Atlas of cardiovascular mon-
itoring. New York: Churchill Livingstone; 1998. p. 294; with permission.)
the y descent is characteristically attenuated. Note that the increased CVP seen
in both constriction and tamponade reects the extramural compressive eect
of the restrictive pericardium or eusion, and cardiac lling is reduced in both
conditions.
Summary
Pressure monitoring systems inuence the contour of the displayed wave-
forms and, on occasion, can introduce signicant artifact in the pressure
traces. It is important to understand the technical details of invasive pres-
sure monitoring to interpret better the information presented.
Careful observation of the arterial pressure waveform can provide infor-
mation about ventricular function, the arterial system, and ventricular pre-
load. In particular, systolic pressure variation during the respiratory cycle in
mechanically ventilated patients is a clinically useful indicator of volume
status.
CVP monitoring is also used to assess intravascular volume, but this mea-
surement is signicantly inuenced by ventricular compliance and intratho-
racic pressure. Under most clinical circumstances, a trend in CVP values or
its change with therapeutic maneuvers is more reliable than a single
PRESSURE MONITORING 733
Fig. 10. Pericardial constriction increases right- and left-sided cardiac lling pressures. The
CVP shows an M or W pattern, with tall a and v waves and steep x and y descents. A mid-
diastolic plateau wave or h wave (asterisk) may also be seen when the heart rate is slow. Abbre-
viations: ART, arterial pressure; R, ECG R wave. (Adapted from Mark JB. Atlas of cardiovas-
cular monitoring. New York: Churchill Livingstone; 1998. p. 318; with permission.)
References
[1] Kleinman B. Understanding natural frequency and damping and how they relate to the mea-
surement of blood pressure. J Clin Monit 1989;5(2):13747.
[2] Gardner RM. Direct blood pressure measurementddynamic response requirements. Anes-
thesiology 1981;54(3):22736.
[3] Hunziker P. Accuracy and dynamic response of disposable pressure transducer-tubing sys-
tems. Can J Anaesth 1987;34(4):40914.
[4] Pittman JA, Sum Ping J, Mark JB. Arterial and central venous pressure monitoring. Int
Anesthesiol Clin 2004;42(1):1330.
[5] Hipkins SF, Rutten AJ, Runciman WB. Experimental analysis of catheter-manometer sys-
tems in vitro and in vivo. Anesthesiology 1989;71(6):893906.
734 BARBEITO & MARK
[6] Kleinman B, Powell S, Gardner RM. Equivalence of fast ush and square wave testing of
blood pressure monitoring systems. J Clin Monit 1996;12(2):14954.
[7] Kleinman B, Powell S, Kumar P, et al. The fast ush test measures the dynamic response
of the entire blood pressure monitoring system. Anesthesiology 1992;77(6):121520.
[8] Courtois M, Fattal PG, Kovacs SJ Jr, et al. Anatomically and physiologically based
reference level for measurement of intracardiac pressures. Circulation 1995;92(7):
19942000.
[9] Mark JB. Technical requirements for direct blood pressure measurement. In: Mark JB,
editor. Atlas of cardiovascular monitoring. New York: Churchill Livingstone; 1998.
p. 99126.
[10] Tegtmeyer K, Brady G, Lai S, et al. Videos in clinical medicine. Placement of an arterial line.
N Engl J Med 2006;354(15):e13.
[11] Singleton RJ, Webb RK, Ludbrook GL, et al. The Australian Incident Monitoring Study.
Problems associated with vascular access: an analysis of 2000 incident reports. Anaesth In-
tensive Care 1993;21(5):6649.
[12] Mandel MA, Dauchot PJ. Radial artery cannulation in 1,000 patients: precautions and com-
plications. J Hand Surg [Am] 1977;2(6):4825.
[13] Wilkins RG. Radial artery cannulation and ischaemic damage: a review. Anaesthesia 1985;
40(9):8969.
[14] Mangano DT, Hickey RF. Ischemic injury following uncomplicated radial artery catheter-
ization. Anesth Analg 1979;58(1):557.
[15] Slogo S, Keats AS, Arlund C. On the safety of radial artery cannulation. Anesthesiology
1983;59(1):427.
[16] OGrady NP, Alexander M, Dellinger EP, et al. Guidelines for the prevention of intravascu-
lar catheter-related infections. Infect Control Hosp Epidemiol 2002;23(12):75969.
[17] Chowet AL, Lopez JR, Brock-Utne JG, et al. Wrist hyperextension leads to median nerve
conduction block: implications for intra-arterial catheter placement. Anesthesiology 2004;
100(2):28791.
[18] Goedje O, Hoeke K, Lichtwarck-Ascho M, et al. Continuous cardiac output by femoral
arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arte-
rial thermodilution. Crit Care Med 1999;27(11):240712.
[19] Linton NW, Linton RA. Estimation of changes in cardiac output from the arterial blood
pressure waveform in the upper limb. Br J Anaesth 2001;86(4):48696.
[20] Zollner C, Haller M, Weis M, et al. Beat-to-beat measurement of cardiac output by intravas-
cular pulse contour analysis: a prospective criterion standard study in patients after cardiac
surgery. J Cardiothorac Vasc Anesth 2000;14(2):1259.
[21] Pittman J, Bar-Yosef S, SumPing J, et al. Continuous cardiac output monitoring with pulse
contour analysis: a comparison with lithium indicator dilution cardiac output measurement.
Crit Care Med 2005;33(9):201521.
[22] Karamanoglu M. A system for analysis of arterial blood pressure waveforms in humans.
Comput Biomed Res 1997;30(3):24455.
[23] Benetos A, Laurent S, Hoeks AP, et al. Arterial alterations with aging and high blood pres-
sure. A noninvasive study of carotid and femoral arteries. Arterioscler Thromb 1993;13(1):
907.
[24] London GM, Guerin AP. Inuence of arterial pulse and reected waves on blood pressure
and cardiac function. Am Heart J 1999;138(3 Pt 2):2204.
[25] Vlachopoulos C, Hirata K, ORourke MF. Pressure-altering agents aect central aortic pres-
sures more than is apparent from upper limb measurements in hypertensive patients: the role
of arterial wave reections. Hypertension 2001;38(6):145660.
[26] Dorman T, Breslow MJ, Lipsett PA, et al. Radial artery pressure monitoring underestimates
central arterial pressure during vasopressor therapy in critically ill surgical patients. Crit
Care Med 1998;26(10):16469.
PRESSURE MONITORING 735
[27] Chauhan S, Saxena N, Mehrotra S, et al. Femoral artery pressures are more reliable than
radial artery pressures on initiation of cardiopulmonary bypass. J Cardiothorac Vasc Anesth
2000;14(3):2746.
[28] ORourke MF, Yaginuma T. Wave reections and the arterial pulse. Arch Intern Med 1984;
144(2):36671.
[29] Perel A, Pizov R, Cotev S. Systolic blood pressure variation is a sensitive indicator of hypo-
volemia in ventilated dogs subjected to graded hemorrhage. Anesthesiology 1987;67(4):
498502.
[30] Rooke GA, Schwid HA, Shapira Y. The eect of graded hemorrhage and intravascular vol-
ume replacement on systolic pressure variation in humans during mechanical and spontane-
ous ventilation. Anesth Analg 1995;80(5):92532.
[31] Michard F, Teboul JL. Predicting uid responsiveness in ICU patients: a critical analysis of
the evidence. Chest 2002;121(6):20008.
[32] Pizov R, Cohen M, Weiss Y, et al. Positive end-expiratory pressure-induced hemodynamic
changes are reected in the arterial pressure waveform. Crit Care Med 1996;24(8):13817.
[33] Mark JB, Slaughter TF. Cardiovascular monitoring. In: Miller R, editor. Millers anesthesia.
6th edition. Philadelphia: Elsevier Churchill Livingstone; 2005. p. 1265362.
[34] McGee DC, Gould MK. Preventing complications of central venous catheterization. N Engl
J Med 2003;348(12):112333.
[35] Randolph AG, Cook DJ, Gonzales CA, et al. Ultrasound guidance for placement of central
venous catheters: a meta-analysis of the literature. Crit Care Med 1996;24(12):20538.
[36] Manseld PF, Hohn DC, Fornage BD, et al. Complications and failures of subclavian-vein
catheterization. N Engl J Med 1994;331(26):17358.
[37] Rothschild J. Chapter 21. Ultrasound guidance of central vein catheterization. In: Making
health care safer: a critical analysis of patient safety practices. Rockville (MD): Agency
for Healthcare Research and Quality; 2001. p. 25361.
[38] Pittet D, Tarara D, Wenzel RP. Nosocomial bloodstream infection in critically ill patients.
Excess length of stay, extra costs, and attributable mortality. JAMA 1994;271(20):1598601.
[39] Saint S. Chapter 16. Prevention of intravascular catheter-associated infections. In: Making
health care safer: a critical analysis of patient safety practices. Rockville (MD): Agency for
Healthcare Research and Quality; 2001. p. 16989.
[40] Maki DG, Stolz SM, Wheeler S, et al. Prevention of central venous catheter-related blood-
stream infection by use of an antiseptic-impregnated catheter. A randomized, controlled
trial. Ann Intern Med 1997;127(4):25766.
[41] Hirsch DR, Ingenito EP, Goldhaber SZ. Prevalence of deep venous thrombosis among
patients in medical intensive care. JAMA 1995;274(4):3357.
[42] Timsit JF, Farkas JC, Boyer JM, et al. Central vein catheter-related thrombosis in intensive
care patients: incidence, risks factors, and relationship with catheter-related sepsis. Chest
1998;114(1):20713.