Purpose of the Supplement

It contains the release information for the digitizer software


versions ACP_4005 and ACP_4007.

For an overview of all valid supplements refer to the order list in MedNet.

CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

printed in Germany 2006-07-14


Agfa Company Confidential
Repair and Service
DD+DIS202.06E Software Releases and Patches

Chapter 3.8

List of Contents

1 Release Information for Software ACP_4007 .................... 1


1.1 Changed Functionality compared to Version ACP_4005 ...................1
1.2 Open Bugs..............................................................................................1

2 Release Info for Digitizer Software ACP_4005 .................. 3


2.1 Changed Functionality ..........................................................................3
2.2 Solved Bugs ...........................................................................................3
2.3 Open Bugs..............................................................................................3

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 CR 85-X Chapter 3.8 / I


2006-07-14 (Type 5148 / 100) Agfa Company Confidential
Repair and Service
DD+DIS202.06E Software Releases and Patches

1 Release Information for Software ACP_4007

1.1 Changed Functionality compared to Version ACP_4005

Internal configuration values of the photomultiplier in CR 85-X are now


checked to be within the specified range.

1.2 Open Bugs

1.2.1 User Interface

1. The language for the Operator and Key Operator menus cannot be
changed since the necessary language files are not available.
This bug does not impact the digitizer operation.

2. The digitizer type shown in the Web interface is wrong.


This bug does not impact the correct functioning of the Web interface.

1.2.2 Image Processing

1. A barely recognizable density change in black regions of images may


occur. The cause is that the photomultiplier signal (SAL) for very low
signal levels does not follow the Gaussian distribution.
This bug is not significant for medical diagnostics.

2. In a special test case the error 20443 occurred:


"Pixel FIFO full condition detected. Pixels were lost."

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 CR 85-X Chapter 3.8 / 1


2006-07-14 (Type 5148 / 100) Agfa Company Confidential
Repair and Service
Software Releases and Patches DD+DIS202.06E

1.2.3 Repair and Maintenance

1. If the installation of a hardware modification floppy disk is not


immediately followed by a digitizer reset then there are no modification
entries in the info counter.

Workaround Press the <Reset> switch at the cPCI rack after the
installation of each hardware modification floppy disk.

2. After photomultiplier exchange the CR 85-X refuses scanning with the


message "Calibration is missing".

Workaround Perform IP Center calibration and Shading calibration.

3. After installation of a new Customization Parameters File (CPF) the


application Mammo is disabled.

Workaround Enable Mammo manually in the Service menu.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Chapter 3.8 / 2 CR 85-X Edition 1, Revision 1


Agfa Company Confidential (Type 5148 / 100) 2006-07-14
Repair and Service
DD+DIS202.06E Software Releases and Patches

2 Release Info for Digitizer Software ACP_4005

NOTE:
The digitizer software for CR 85-X is based on the software for the
CR 75.0 digitizer. Changes and solved bugs refer to the
predecessor CR 75.0 software, version ACP_3403.

2.1 Changed Functionality

The digitizer CR 85-X is supported additionally.

2.2 Solved Bugs

4. An internal configuration value (Upper Limit MFB) is corrected which


had been changed erroneously in software version ACP_3403.

5. The Mammo factor has been adapted.

2.3 Open Bugs

2.3.1 User Interface

1. The language for the Operator and Key Operator menus cannot be
changed since the necessary language files are not available.
This bug does not impact the digitizer operation.

2. The digitizer type shown in the Web interface is wrong.


This bug does not impact the correct functioning of the Web interface.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 CR 85-X Chapter 3.8 / 3


2006-07-14 (Type 5148 / 100) Agfa Company Confidential
Repair and Service
Software Releases and Patches DD+DIS202.06E

2.3.2 Image Processing

1. A barely recognizable density change in black regions of images may


occur. The cause is that the photomultiplier signal (SAL) for very low
signal levels does not follow the Gaussian distribution.
This bug is not significant for medical diagnostics.

2. In a special test case the error 20443 occurred:


"Pixel FIFO full condition detected. Pixels were lost.".

2.3.3 Repair and Maintenance

1. If the installation of a hardware modification floppy disk is not


immediately followed by a digitizer reset then there are no modification
entries in the info counter.

Workaround Reset the digitizer after the installation of each


hardware modification floppy disk.

2. After photomultiplier exchange the CR 85-X refuses scanning with the


message "Calibration is missing".

Workaround Perform IP Center calibration and Shading calibration.

3. After installation of a new Customization Parameters File (CPF) the


application Mammo is disabled.

Workaround Enable Mammo manually in the Service menu.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Chapter 3.8 / 4 CR 85-X Edition 1, Revision 1


Agfa Company Confidential (Type 5148 / 100) 2006-07-14

You might also like