You are on page 1of 379

BS-300

Chemistry Analyzer

Operation Manual
2007 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.

For this Operation Manual, the issued Date is 2007-07 (Version: 1.8).

Intellectual Property Statement


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called
Mindray) owns the intellectual property rights to this Mindray product and this
manual. This manual may refer to information protected by copyrights or patents and
does not convey any license under the patent rights of Mindray, nor the rights of
others. Mindray does not assume any liability arising out of any infringements of
patents or other rights of third parties.

Mindray intends to maintain the contents of this manual as confidential information.


Disclosure of the information in this manual in any manner whatsoever without the
written permission of Mindray is strictly forbidden.

Release, amendment, reproduction, distribution, rent, adaption and translation of this


manual in any manner whatsoever without the written permission of Mindray is
strictly forbidden.

, , , , are the registered


trademarks or trademarks owned by Mindray in China and other countries. All
other trademarks that appear in this manual are used only for editorial purposes
without the intention of improperly using them. They are the property of their
respective owners.

Responsibility on the Manufacturer Party


Contents of this manual are subject to changes without prior notice.

All information contained in this manual is believed to be correct. Mindray shall not
be liable for errors contained herein nor for incidental or consequential damages in
connection with the furnishing, performance, or use of this manual.

Mindray is responsible for safety, reliability and performance of this product only in
the condition that:

all installation operations, expansions, changes, modifications and repairs of this


product are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable
national and local requirements;
the product is used in accordance with the instructions for use.

I
WARNING:
It is important for the hospital or organization that employs this
equipment to carry out a reasonable service/maintenance plan.
Neglect of this may result in machine breakdown or injury of human
health.

Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY
OR FITNESS FOR ANY PARTICULAR PURPOSE.

Exemptions

Mindray's obligation or liability under this warranty does not include any
transportation or other charges or liability for direct, indirect or consequential
damages or delay resulting from the improper use or application of the product or the
use of parts or accessories not approved by Mindray or repairs by people other than
Mindray authorized personnel.

This warranty shall not extend to:

any Mindray product which has been subjected to misuse, negligence or


accident;
any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;
any product of any other manufacturer.

Return Policy

Return Procedure

In the event that it becomes necessary to return this product or part of this product to
Mindray, the following procedure should be followed:

1 Obtain return authorization: Contact the Mindray Customer Service


Department and obtain a Customer Service Authorization (Mindray)
number. The Mindray number must appear on the outside of the shipping
container. Returned shipments will not be accepted if the Mindray number
is not clearly visible. Please provide the model number, serial number,
and a brief description of the reason for return.
2 Freight policy: The customer is responsible for freight charges when this
product is shipped to Mindray for service (this includes customs charges).
3 Return address: Please send the part(s) or equipment to the address
offered by Customer Service department.

II
Company Contact

Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.


Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial
Park, Nanshan, ShenZhen518057, P.R.China
Tel: +86 755 26582479 26582888
Fax: +86 755 26582934 26582500

III
Preface

Who Should Read This Manual


This manual is written for clinical laboratory professionals to:

learn about the BS-300 hardware and software


set up and operate the BS-300
maintain and troubleshoot the BS-300

WARNING:
The BS-300 Chemistry Analyzer is to be operated only by clinical
professionals, doctors or laboratory experimenters trained and
authorized by Mindray or Mindray-authorized distributors.

What Can You Find in This Manual


The BS-300 Chemistry Analyzer should be operated and serviced strictly as
instructed by this manual.

This manual contains 7 chapters, plus 3 appendixes:

1 System Description
2 Installation
3 Basic Operations
4 Advanced Operations
5 Maintenance
6 Troubleshooting
7 Calculation Methods
Appendix A Specifications
Appendix B Supplies
Appendix C Index

Conventions Used in This Manual


This manual uses certain typographical conventions to clarify meanings in the text.

All capital, italic font indicates a key name, such as ENTER.

Bold font indicates a chapter title, such as 5 Maintenance

1
Preface

Bold and Italic font indicates:

Screen text displayed by the analyzer, such as Cant save.


A menu name, such as Setup.
A screen field or value, such as Method.

Safety Symbols

This chart explains the symbols used in this manual.

When you see Then


Read the statement following the symbol. The
WARNING: statement is alerting you to an operating hazard
that can cause personal injury.

Read the statement following the symbol. The


BIOHAZARD: statement is alerting you to a potentially
biohazardous condition.

Read the statement following the symbol. The


CAUTION: statement is alerting you to a possibility of system
damage or unreliable results.

Read the statement following the symbol. The


NOTE: statement is alerting you to information that
requires your attention.

Labels Used on the System

The labels attached to some panels use symbols with the text to clarify the meaning
of the text. If any of the labels peels off, contact Mindray customer service
department or your local distributor for replacement. The table below explains the
symbols on the labels.

Serial Number

Manufacturer

Date of Manufacture

Authorized Representative in the European Community

The device is fully in conformance with the Council Directive


Concerning In Vitro Diagnostic Medical Devices 98/79/EC.

2
Preface

The following definition of the WEEE label applies to EU


member states only: The use of this symbol indicates that
this product should not be treated as household waste. By
ensuring that this product is disposed of correctly, you will
help prevent bringing potential negative consequences to the
environment and human health. For more detailed
information with regard to returning and recycling this
product, please consult the distributor from whom you
purchased the product.

In Vitro Diagnostic equipment

Biohazard Warning: risk of potentially biohazardous infection

Warning: risk of personal injury or equipment damage

Warning: risk of electric shock

Warning: risk of burn

Warning: risk of eye hurt

ON (POWER)

OFF (POWER)

ON (ANALYZING UNIT POWER)

OFF (ANALYZING UNIT POWER)

Alternating current (AC)

Graphics

All graphics, including screens and printout, are for illustration purposes only and
must not be used for any other purpose.

EC Representative

Name: Shanghai International Holding Corp. GmbH(Europe)


Address: Eiffestrae 80, 20537 Hamburg Germany
Phone: 0049-40-2513175
Fax: 0049-40-255726

3
Preface

Safety Precautions
Observe the following safety precautions when using the BS-300 Chemistry Analyzer.
Ignoring any of these safety precautions may lead to personal injury or equipment
damage.

WARNING:
If the system is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.

Preventing Electric Shocks

WARNING:
When the POWER is on, users must not open the rear or side cover.
Spillage of reagent or sample on the analyzer may cause equipment
failure and even electric shock. Do not place sample and reagent on
the analyzer. In case of liquid ingression, shutspillage, switch off the
power suppliesimmediately, remove the spillage and contact Mindray
Customer Service Departmentcustomer service department or your
local distributor.

Preventing Personal Injury

WARNING:
Do not touch such moving parts as sample probe, reagent probe and
mixing bar, when the analyzer is in operation.
Do not put your finger or hand into any open part when the analyzer is
in operation.
If you want to replace the photometer lamp, first switch off the POWER
and then wait at least 15 minutes for the lamp to cool down before
touching it. Do not touch the lamp before it cools down, or you may get
burned.

Eye Protection

WARNING:
Light sent by the photometer lamp or the bar code scanner may hurt
your eyes. Do not stare into the lamp or the laser beam from the bar
code scanner when the analyzer is in operation.

4
Preface

Preventing Infection

BIOHAZARD:
Inappropriately handling samples may lead to biohazardous infection.
Do not touch the sample, mixture or waste with your hands. Wear
gloves and lab coat and, if necessary, goggles.
In case your skin contacts the sample, follow standard laboratory
safety procedure and consult a doctor.

Handling Reagents and Wash Solution

WARNING:
Reagents and wash solution are corrosive to human skins. Exercise
caution when using the reagents. In case your skin or clothes contact
the reagents, wash them off with soap and clean water. In case the
reagents spill into your eyes, rinse them with much water and consult
an oculist.

Treating Waste Liquids

BIOHAZARD:
Dispose of the waste in accordance with your local or national
guidelines for biohazard waste disposal and consult the manufacturer
or distributor of the reagents for details.

Treating Waste Analyzer

Please observe the following instructions to dispose of the waste analyzer.

WARNING
Materials of the analyzer are subject to contamination regulations.
Dispose of the waste analyzer in accordance with your local or
national guidelines for waste disposal.

Preventing Fire or Explosion

WARNING:
Ethanol is flammable substance around the analyzer. Please exercise
caution while using the ethanol.

5
Preface

Precautions on Use
To use the analyzer safely and effectively, pay attention to the following operation
notes.

Intended Use

WARNING:
The BS-300 Chemistry Analyzer (hereinafter referred to as the
BS-300) is a chemistry system that is designed for the in vitro
quantitative determination of clinical chemistries in serum, plasma,
urine, CSF and other samples. Please consult Mindray first if you want
to use the BS-300 for other purposes.
To draw a clinical conclusion, please also refer to the patients clinical
symptom and other test results.

Operator

WARNING:
The BS-300 is to be operated only by clinical professionals, doctors or
laboratory experimenters trained and authorized by Mindray or
Mindray-authorized distributors.

Environment

CAUTION:
Please install and operate the analyzer in an environment specified by
this manual. Installing and operating the analyzer in other environment
may lead to unreliable results and even equipment damage.
To relocate the analyzer, please contact Mindray Customer Service
Department or your local distributor.

Preventing Interference by Electromagnetic Noise

CAUTION:
Electromagnetic noise may interfere with operation of the analyzer. Do
not install devices generating excessive electromagnetic noise around
the analyzer. Do not use such devices as mobile phones or radio
transmitters in the room housing the analyzer. Do not use other CRT
displays around the analyzer.
The analyzing unit is designed and tested to EN 61326:2003 Class A.
In a domestic environment it may cause radio interference, so you
may need to take measures to mitigate the interference.
Do not use other medical instruments around the analyzer that may
generate electromagnetic noise to interfere with their operations.

6
Preface

Operating the BS-300

CAUTION:
Operate the BS-300 strictly as instructed by this manual. Inappropriate
use of the analyzer may lead to unreliable test results or even
equipment damage or personal injury.
Before using the system for the first time, run the calibration program
and the QC program to make sure it is in a proper state.
Be sure to run the QC program every time you use the system,
otherwise the result man be unreliable.
Do not open the sample disk or reagent disk cover when the analyzer
is operating.
The RS-232 port on the analyzing unit is to be used for connection
with the operation unit only. Do not use it for other purposes. Only use
the supplied cable for the connection.
The operation unit is a personal computer with the analyzer control
system installed. Installing other software or hardware on this
computer may interfere with the system operation. Do not run other
software when the system is working.
Do not use this computer for other purposes. Inappropriate use of this
computer may introduce computer virus, which may spread through
floppy disks, software or network, into the system.
Do not touch the display, mouse or keyboard with wet hands or hands
with chemicals on them.
Dont place the POWER to ON again within 10 seconds since placing
it to OFF, otherwise the system may enter the protection status. If it
does so, place the POWER to OFF and place it to ON again.

Maintaining the BS-300

CAUTION:
Maintain the BS-300 strictly as instructed by this manual. Inappropriate
maintenance may lead to equipment damage or personal injury.
To wipe off dust from the analyzer surface, use a soft, clean and wet
(not too wet) cloth, soaked with soap water if necessary, to clean the
surface. Do not use such organic solvents as ethanol for the cleaning.
After cleaning, wipe the surface dry with dry cloth.
Switch off all the powers and disconnect the power plug before
cleaning. Take necessary measures to prevent water ingression into
the system, otherwise it may lead to equipment damage or personal
injury.
Replacements of such major parts as the photometer, sample probe,
reagent probe, mixing bar and syringe plungers must be followed by a
calibration.

7
Preface

Samples

CAUTION:
Use serum samples that are completely separated from blood clots or
urine samples that are free from suspended matter. If fibrin exists in
the serum samples or suspended matter exist in the urine samples,
the sample probe may be blocked.
Medicines, anticoagulants or preservative in the samples may lead to
unreliable test results.
Hemolysis, jaundice or chylomicron in the samples may lead to
unreliable test results, so sample blanks are recommended

Store the samples properly. Improper storage may change the


compositions of the samples and lead to unreliable results.
Sample volatilization may lead to unreliable results. Do not leave the
sample open for too long.
Not all the tests the reagents claim capable of analyzing can be
analyzed on the analyzer. Consult the reagent suppliers for detailed
information.
Certain samples need to be processed before being analyzed by the
analyzer. Consult the reagent suppliers for details.
The analyzer has a specific requirement on the minimum sample
volume. Refer to this manual for the proper sample volume.
Load the sample to proper tube positions on the sample disk before
the analysis begins; otherwise you will not obtain correct results.

Reagents, Calibrators and Controls

CAUTION:
Use proper reagents, calibrators and controls in the analyzer.
Select appropriate reagents according to the performance
characteristics of the analyzer. Consult the reagent supplier or
Mindray-authorized distributor for details when you are not sure about
your reagent choice.
Store and use the reagents, calibrators and controls strictly as
instructed by the suppliers. Otherwise, you may not obtain reliable
results or the best performance of the analyzer.
Perform calibration after changing the reagents. Otherwise, you may
not obtain reliable results.
Contamination among reagents by carryover may lead to unreliable
test results. Consult the reagent supplier for details.

8
Preface

Setting up BS-300

CAUTION:
You need to set up the analyzer strictly as instructed by this manual
before using it. To define such parameters as sample volume, reagent
volume and wavelength, consult instructions of reagents.

Backing Up Data

NOTE:
The analyzer automatically backs up the data to the built-in hard disk.
However, data loss is still possible due to mis-deletion or physical
damage of the disk. Mindray recommends you regularly back up the
data to such medium as CDs.

Computer and Printer

NOTE:
Refer to their users manuals for details.

External Equipment

WARNING:
External equipment connected to the system, such as PC and printer,
shall be consistent with IEC 60950 or EN 60950.

9
Contents

Preface.......................................................................................................................................... 1
Who Should Read This Manual ............................................................................................. 1
What Can You Find in This Manual........................................................................................ 1
Conventions Used in This Manual ......................................................................................... 1
Safety Precautions ................................................................................................................. 4
Precautions on Use................................................................................................................ 6

Contents ........................................................................................................................................ i

1 System Description .......................................................................................................... 1-1


1.1 Hardware Introduction ......................................................................................... 1-1
1.1.1 Analyzing Unit ...................................................................................... 1-2
1.1.2 Operation Unit .................................................................................... 1-16
1.1.3 Output Unit ......................................................................................... 1-16
1.2 Software Introduction......................................................................................... 1-17
1.2.1 Software Interface .............................................................................. 1-17
1.2.2 Main Interface Components ............................................................... 1-18
1.2.3 Mouse Operations .............................................................................. 1-21

2 Installation ......................................................................................................................... 2-1


2.1 Unpacking............................................................................................................ 2-1
2.2 Installation Requirements .................................................................................... 2-1
2.2.1 Installation Environment Requirements................................................ 2-1
2.2.2 Power Requirements............................................................................ 2-2
2.2.3 Temperature and Humidity Requirements............................................ 2-2
2.2.4 Water Supply and Drain Requirements................................................ 2-3
2.2.5 Space and Accessibility Requirements................................................. 2-3
2.3 Connecting the Tanks .......................................................................................... 2-4
2.3.1 Connecting the Deionized Water Tank ................................................. 2-4
2.3.2 Connecting the Waste Tank.................................................................. 2-4
2.4 ISE Unit (optional) Installation ............................................................................. 2-5
2.4.1 Installing Electrodes ............................................................................. 2-6
2.4.2 Installing Reagent Module.................................................................... 2-8

3 Basic Operations .............................................................................................................. 3-1


3.1 Daily Procedure ................................................................................................... 3-1
3.2 Preparing for Analysis.......................................................................................... 3-2
3.2.1 Checking before Startup....................................................................... 3-2
3.2.2 Power-on .............................................................................................. 3-3
3.2.3 Starting the Operating Software ........................................................... 3-3
3.2.4 Setting Up the BS-300.......................................................................... 3-4
3.2.5 Preparing for Analysis .......................................................................... 3-5
3.3 Starting Analysis .................................................................................................. 3-5
3.3.1 Calibration ............................................................................................ 3-5
3.3.2 QC ........................................................................................................ 3-6
3.3.3 Samples................................................................................................ 3-6
3.3.4 Editing Results of Sample Runs........................................................... 3-7
3.3.5 Printing Results of Sample Runs.......................................................... 3-7
3.4 Finishing Analysis ................................................................................................ 3-7
3.4.1 Exiting the Operating Software............................................................. 3-7
3.4.2 Shutdown.............................................................................................. 3-8

i
Contents

3.4.3 Checking after Shutdown ..................................................................... 3-8

4 Advanced Operations....................................................................................................... 4-1


4.1 Routine ................................................................................................................ 4-1
4.1.1 Samples................................................................................................ 4-1
4.1.2 Calibration .......................................................................................... 4-11
4.1.3 QC ...................................................................................................... 4-13
4.1.4 Status.................................................................................................. 4-14
4.1.5 Start .................................................................................................... 4-34
4.1.6 Probe Stop.......................................................................................... 4-37
4.1.7 Stop .................................................................................................... 4-38
4.1.8 Exit...................................................................................................... 4-38
4.1.9 Emergency Exit .................................................................................. 4-39
4.2 Parameters ........................................................................................................ 4-39
4.2.1 Assay.................................................................................................. 4-39
4.2.2 Reagent .............................................................................................. 4-61
4.2.3 Calibration .......................................................................................... 4-67
4.2.4 QC ...................................................................................................... 4-87
4.3 Data ................................................................................................................. 4-108
4.3.1 Export ............................................................................................... 4-108
4.3.2 Import ............................................................................................... 4-112
4.3.3 Demographics .................................................................................. 4-114
4.3.4 Result ............................................................................................... 4-116
4.4 System............................................................................................................. 4-130
4.4.1 System Status................................................................................... 4-130
4.4.2 Setup ................................................................................................ 4-133
4.4.3 Hospital............................................................................................. 4-141
4.4.4 User .................................................................................................. 4-146
4.4.5 Log.................................................................................................... 4-150
4.4.6 Maintenance ..................................................................................... 4-152
4.4.7 Print .................................................................................................. 4-165
4.4.8 Temperature ..................................................................................... 4-171
4.4.9 Database .......................................................................................... 4-173
4.4.10 Bar Code Scanner............................................................................ 4-175
4.4.11 ISE Unit ............................................................................................ 4-176
4.5 Help .................................................................................................................4-181
4.5.1 Version.............................................................................................. 4-181
4.5.2 Guide ................................................................................................ 4-181

5 Maintenance ...................................................................................................................... 5-1


5.1 Preparation .......................................................................................................... 5-1
5.2 Daily Maintenance ............................................................................................... 5-2
5.2.1 Checking Sample/Reagent Syringes ................................................... 5-2
5.2.2 Checking/Washing Sample Probe........................................................ 5-3
5.2.3 Checking/Washing Reagent Probe ...................................................... 5-4
5.2.4 Checking/Washing Mixing Bar ............................................................. 5-4
5.2.5 Checking Detergents............................................................................ 5-4
5.2.6 Checking Connection of Deionized Water............................................ 5-5
5.2.7 Checking Connection of Wastewater ................................................... 5-6
5.2.8 Checking Remaining Deionized Water................................................. 5-8
5.2.9 Emptying Waste Tank........................................................................... 5-8
5.2.10 Emptying Used-Cuvettes Bucket ......................................................... 5-9
5.2.11 Cleaning ISE Unit (optional) ............................................................... 5-10
5.2.12 Pump Calibration (optional)................................................................ 5-10
5.3 Weekly Maintenance ......................................................................................... 5-11
5.3.1 Washing Sample Probe...................................................................... 5-11
5.3.2 Washing Reagent Probe .................................................................... 5-13
5.3.3 Washing Mixing Bar............................................................................ 5-15
5.3.4 Washing Deionized Water Tank ......................................................... 5-16

ii
Contents

5.3.5 Washing Waste Tank .......................................................................... 5-17


5.3.6 Cleaning Sample Disk/Compartment ................................................. 5-18
5.3.7 Cleaning Reagent Disk/Compartment................................................ 5-19
5.3.8 Cleaning Panel of Analyzing Unit ....................................................... 5-20
5.4 Monthly Maintenance ........................................................................................ 5-21
5.4.1 Cleaning Wash Well of Sample Probe ............................................... 5-21
5.4.2 Cleaning Wash Well of Reagent Probe.............................................. 5-21
5.4.3 Cleaning Wash Well of Mixing Bar ..................................................... 5-22
5.4.4 Cleaning Sample Probe Rotor............................................................ 5-23
5.4.5 Cleaning Reagent Probe Rotor .......................................................... 5-23
5.4.6 Cleaning Mixing Bar Rotor ................................................................. 5-24
5.5 Maintenance Every Six Months ......................................................................... 5-25
5.6 Irregular Maintenance........................................................................................ 5-25
5.6.1 Unclogging Sample Probe.................................................................. 5-25
5.6.2 Unclogging Reagent Probe ................................................................ 5-32
5.6.3 Replacing Sample Probe.................................................................... 5-38
5.6.4 Replacing Reagent Probe .................................................................. 5-38
5.6.5 Replacing Mixing Bar ......................................................................... 5-39
5.6.6 Replacing Plunger Assembly of Sample/Reagent Syringe ................ 5-43
5.6.7 Removing Air Bubbles ........................................................................ 5-45
5.6.8 Replacing Lamp ................................................................................. 5-46
5.6.9 Replacing Electrode (optional) ........................................................... 5-48
5.6.10 Replacing Reagent Module (optional)................................................ 5-49
5.6.11 ISE Unit Storage (optional)................................................................. 5-50
5.6.12 ISE Unit Re-activation (optional) ........................................................ 5-51
5.7 Maintenance Log ............................................................................................... 5-51

6 Troubleshooting................................................................................................................ 6-1
6.1 Classification of Error Messages ......................................................................... 6-1
6.2 Corrective Measures ........................................................................................... 6-4
6.2.1 Corrective Measures for General Errors .............................................. 6-4
6.2.2 Corrective Measures for ISE Unit Errors............................................ 6-48

7 Calculation Methods......................................................................................................... 7-1


7.1 Reaction Types .................................................................................................... 7-1
7.1.1 Endpoint ............................................................................................... 7-1
7.1.2 Fixed-Time............................................................................................ 7-1
7.1.3 Kinetic................................................................................................... 7-3
7.2 Calculation Process............................................................................................. 7-3
7.2.1 AD Value Absorbance...................................................................... 7-4
7.2.2 Absorbance Response .................................................................... 7-5
7.2.3 Response Calibration Parameter..................................................... 7-9
7.2.4 Calibration Parameter Test Result/QC Result ............................... 7-13
7.2.5 QC Result QC Conclusion ............................................................. 7-15
7.3 Calculation Method of ISE Unit (optional) ......................................................... 7-17

Appendix A Specifications............................................................................................ A-1


A.1 Technical Specifications.......................................................................................A-1
A.2 Power Requirements ...........................................................................................A-2
A.3 Environmental Requirements ..............................................................................A-2
A.4 Dimension and Weight ........................................................................................A-2
A.5 Other Specifications.............................................................................................A-2
A.6 Input/Output Devices ...........................................................................................A-2
A.7 Interface...............................................................................................................A-3
A.8 Built-in Sample Bar Code Scanner (optional)......................................................A-3
A.9 ISE Module (optional) ..........................................................................................A-3

Appendix B Supplies ..................................................................................................... B-1

iii
Contents

Appendix C Index........................................................................................................... C-1

iv
1 System Description
This chapter includes the following two sections:

Hardware Introduction
Software Introduction
The BS-300 Chemistry Analyzer is a chemistry system that is designed for the in
vitro quantitative determination of clinical chemistries in serum, plasma, urine, CSF
and other samples.

NOTE:
Not all the parameters the reagents claim capable of testing can be
tested on the analyzer. Consult the reagent suppliers for detailed
information.

1.1 Hardware Introduction


The BS-300 consists of the analyzing unit, operation unit, output unit, replacing parts
and consumables. The analyzing unit and the operation unit are as shown in Figure
1-1.

Figure 1-1 Overview of the analyzing unit and the operation unit

Analyzing Unit Operation Unit

1-1
System Description

1.1.1 Analyzing Unit


The analyzing unit handles all the analyzing operations, which include dispensing
samples and reagents, mixing, reaction and measurement, automatically replacing
reaction cuvettes, etc.

The analyzing unit consists of the following major parts:

Sample disk assembly


Built-in sample bar code scanner (optional)
Sample dispenser
Reagent disk assembly
Reagent dispenser
Mixer
Reaction cuvette loader
Reaction disk assembly
Photometric system
ISE module (optional)

Figure 1-2 Top view

1-2
System Description

Figure 1-3 Front view

Connect the power cord to the power plug shown in Figure 1-3.

Figure 1-4 Rear view

DEIONIZED WATER: If you choose to feed deionized water from the back of the
analyzing unit, pull the deionized water tubing into the unit from here;
D-SENSOR: If you choose to feed deionized water from the back of the analyzing
unit, pull the cable of the deionized water level detector into the unit from here;

1-3
System Description

WASTE: If choose to discharge waste from the back of the analyzing unit, pull the
waste tubing out of the unit from here;
W-SENSOR: If you choose to discharge the waste from the back of the analyzing
unit, pull the cable of the waste level detector out of the unit from here;
COM: It is the serial port for communication with the operation unit;
RESET: You can press this button to restart the analyzing unit when it stops
responding.

WARNING:
Press the RESET button only when the analyzing unit stops
responding. Do not press this button when the unit is working normally.
Otherwise, it may lead to system failure, unreliable analysis results or
serious unexpected consequences.

1.1.1.1 Sample Disk

The sample disk (as Figure 1-5 shows) holds sample tubes and rotates
(counter-clockwise only) the designated tube to the position for the sample probe to
aspirate.

Figure 1-5 Sample disk

The sample disk is composed of two circles an inner circle and an outer circle. It
provides 60 tube positions, 30 on the inner circle and 30 on the outer circle. Every
tube position has an identification number. Positions 1-45 are for samples, E1-E5 for
emergency samples, S1-S6 for calibrators, C1-C3 for controls, and W for distilled
water. All the other positions except the position W can be used for samples, too.

The sample disk can hold six types of tubes micro sample cup, collection tube,
12mm68.5mm, 12mm 99mm, 12.7mm75mm and 12.7mm100mm.

To install the sample disk, keep the handle at the vertical position, align the central
hole of the handwheel to the rotor, gently lower the sample disk all the way down
(rotate the disk a bit as necessary) and move the handle back to the horizontal
position to secure the disk to the rotor. When you have finished installing the disk,

1-4
System Description

grab the handwheel and lightly pull it upward. If the disk remains secured, it means it
is well installed.

To remove the sample disk, first shift the handle from the horizontal position to the
vertical position. Then grab the handle or handwheel and pull the disk upward to
remove it from the rotor.

To load sample tubes, insert the tube into the tube holder until the bottom of the tube
contacts the groove on the tube rack.

To remove sample tubes, grab the tube and pull it upward to remove it from the tube
holder.

WARNING:
Before installing/removing the sample disk, make sure the analyzing
unit is turned off and the sample disk has stopped.
Before installing/removing the sample tubes, make sure the analyzing
unit is turned off and the sample disk has stopped.
Do not use sample tubes other than the specified.
Before running the analyzing unit, make sure that the sample disk
cover is closed and the round red mark on the cover is aligned with its
counterpart on the panel. Otherwise the sample probe may be
damaged.
If samples spill in the sample compartment or on the sample disk, be
sure to wipe the spills with cloth soaked with water or disinfector after
turning off the analyzing unit.

1.1.1.2 Built-in Sample Bar Code Scanner (optional)

Figure 1-6 Built-in bar code scanner (optional)

Bar Code Scanner


Window

The built-in sample bar code scanner (optional) can read the bar code labels on the
sample tubes.

The bar code labels can be applied to the following sample tubes.

12mm 99mm
12.7mm75mm
12.7mm100mm

1-5
System Description

The bar code labels applied to the sample tubes must meet the following
requirements.

The bottom of the sample tube fully contacts the tube rack of the sample disk.
There is no obstruction between the scanner and the bar code labels.
Stick the bar code label to the sample tube so that the lower end of the bar code
label is 8-10mm above the bottom of the tube.
The length of the bar code label is no more than 55mm with blank areas of 5 mm
are provided at both sides of barcode.

Tube

Bar code label 5mm

no more than 55mm

5mm

8-10mm

Black print on white background.


Stick the bar code label to the sample tube so that the inclination angle is no more
than 5.

Inclination angle

Stick the entire surface of the label to the surface of the sample tube so it will not
peel.
When inserting the tube with bar code label into the sample disk, ensure the
center of the label is within the 2mm range, as shown in the figures below.

1-6
System Description

+2mm

Center of bar code label


-2mm

Bar code label

NOTE:
Mindray recommends printing coding information beside barcode.

Bar code symbol specifications:

The bar code scanner can read the following bar code symbologies: Code 128 (Set A, B,
C), Code 39, Codabar, ITF (Interleaved 2 of 5), UPC/EAN, Code93
Feature size: 0.19-0.50mm
Length: 3-16
Thin bar: thick bar: 1:2.5-3
Height: 10mm or more
Printing: black print on white background; the minimum acceptable symbol grade
is Class C as defined in the ANSI MH10.8M Print Quality Specification.

NOTE:
To ensure the stability of reading Codabar, the length of it is no less
than 4.
The maximum number of digits will be determined after confirming that
the actual printed labels can be read properly.

1.1.1.3 Sample Dispenser

The sample dispenser is composed of a sample probe, probe arm, probe rotor
(Figure 1-7), syringe (Figure 1-8) and a fluid system. It aspirates certain amount of

1-7
System Description

sample from the designated sample tube and dispenses the aspirated sample to the
designated reaction cuvette.

Open the middle door of the analyzing unit and you will see the sample syringe (right)
and the reagent syringe (left). The sample syringe can aspirate/dispense 3-45l of
sample for clinical chemistry tests or 70l of sample for ISE tests (if the ISE unit is
connected) and its minimum scale is 0.5l.

Figure 1-7 Sample dispenser

Figure 1-8 Sample syringe

Sample Syringe

To dispense the sample, the sample probe first moves to the sample tube to aspirate
certain amount of the sample, then to the reaction disk to dispense the aspirated
sample to a reaction cuvette and finally to the wash well for cleaning.

The sample probe can automatically detect the sample level. When the sample
probe contacts obstacles in the vertical direction, its safeguard will automatically
protect the probe from damage.

1-8
System Description

WARNING:
When the analyzing unit is running, do not place any part of your body
or any obstacle in the route the arm moves. Otherwise, it may lead to
personnel injury or equipment damage.

1.1.1.4 Reagent Disk Assembly

The reagent disk assembly is composed of a reagent disk and a refrigerator.

The reagent disk (as Figure 1-9 shows) holds the reagent bottles and rotates
(counter-clockwise only) the designated bottle to the position for the reagent probe to
aspirate.

The refrigerator keeps the reagents in a low temperature environment to keep them
stable and minimize volatilization.

Figure 1-9 Reagent disk

The reagent disk has two circles, inner circle and outer circle. There are 25 reagent
positions in each circle and the reagent disk has totally 50 reagent positions. The
reagent disk can hold the Mindray inner-circle bottles, Mindray outer-circle bottles,
Hitachi 7060 bottles, Hitachi 7170 bottles and Hitachi 20ml bottles.

The refrigerator can run 24 hours a day to keep the temperature at 4-10.

NOTE:
The refrigerator has a power supply independent of that of the
analyzing unit. It will be put into service once the POWER is turned on.

To install the reagent disk, keep the handle at the vertical position, align the central
hole of the handwheel to the rotor, gently lower the reagent disk all the way down
(rotate the disk a bit as necessary) and move the handle back to the horizontal
position to secure the disk to the rotor. When you have finished installing the disk, grab

1-9
System Description

the handwheel and lightly pull it upward. If the disk remains secured, it means it is
well installed.

To remove the reagent disk, first shift the handle from the horizontal position to the
vertical position. Then grab the handle or handwheel and pull the disk upward to
remove it from the rotor.

To load reagent bottles, insert the bottle into the bottle holder until the bottom of the
bottle contacts the groove on the bottle rack.

To remove the reagent bottle, grab the bottle and pull it upward to remove it from the
bottle holder.

WARNING:
Before installing/removing the reagent disk, make sure the analyzing
unit is turned off and the sample disk has stopped.
Before installing/removing the reagent bottles, make sure the
analyzing unit is turned off and the sample disk has stopped. When
installing the bottle, make sure the bottle bottom is in full contact with
the bottle rack.
Before running the analyzing unit, make sure that the reagent disk
cover is closed and the round red mark on the cover is aligned with its
counterpart on the panel. Otherwise the sample probe may be
damaged and cooling capacity of the refrigerator will be degraded.
If reagents spill in the reagent compartment or on the reagent disk, be
sure to wipe the spills with cloth soaked with water or disinfector after
turning off the analyzing unit.

1.1.1.5 Reagent Dispenser

The reagent dispenser is composed of a reagent probe, probe arm, probe rotor
(Figure 1-10), syringe (Figure 1-11) and fluid system. It aspirates certain amount of
reagent from the designated reagent bottle and dispenses the aspirated reagent to
the designated reaction cuvette. The reagent probe arm can preheat the aspirated
reagent.

Open the middle door of the analyzing unit and you will see the reagent syringe
(right) and reagent syringe (left). The reagent syringe can aspirate/dispense
30-450l of reagent and its minimum scale is 1l.

1-10
System Description

Figure 1-10 Reagent dispenser

Figure 1-11 Reagent syringe

To dispense the reagent, the reagent probe first moves to the reagent bottle to
aspirate certain amount of the reagent, then to the reaction disk to dispense the
aspirated reagent to a reaction cuvette and finally to the wash well for cleaning.

The reagent probe can automatically detect the reagent level. When the reagent
probe contacts obstacles in the vertical direction its safeguard will automatically
function to protect the probe from damage.

WARNING:
When the analyzing unit is running, do not place any part of your body
or any obstacle in the route the arm moves. Otherwise, it may lead to
personnel injury or equipment damage.

1-11
System Description

1.1.1.6 Mixer

The mixer is composed of a mixing bar, bar arm and rotor, as Figure 1-12 shows.
The mixing bar thoroughly stirs the reaction mixture (reagents and sample) in the
reaction cuvette.

Figure 1-12 Mixer

For the single-reagent test, the mixer starts to work after the sample is dispensed
into the reaction cuvette. For the double-reagent test, the mixer starts to work after
the sample or the second reagent is dispensed into the reaction cuvette. The mixing
bar moves to a position above the reaction disk, lowers into the cuvette, and rotates
to mix the mixture in the cuvette. After that, the bar moves to the wash well for
cleaning and de-watering.

1.1.1.7 Reaction Cuvette Loader

The cuvette loader is composed of a cuvette feeder, manipulator and used-cuvette


bucket.

This loader loads new cuvettes to the reaction disk and dumps used cuvettes into a
bucket. The feeder is located in the cuvette compartment and transfers cuvettes to
the manipulator, which grabs the cuvettes and loads them to the reaction disk and
then unloads the used cuvettes and dumps them into the used-cuvette bucket. Open
the lower door of the analyzing unit and you will see the used-cuvette bucket in the
middle.

NOTE:
Normally, the used-cuvette bucket can hold 80 segments of reaction
cuvettes (10 cuvettes per segment). Be sure to empty the bucket in
time so that excessive cuvettes will not overflow and contaminate the
cabinet.

Disposable cuvettes are adopted as the reaction cuvettes. Ten cuvettes are bound
together as one cuvette segment. The cuvette compartment can house 30 cuvette
segments, which are sufficient for 300 continuous tests.

1-12
System Description

When all the cuvettes of a cuvette segment have been used, the reaction disk
rotates the segment to the unloading position, and then the manipulator dumps the
used segment into the used-cuvette bucket and loads the new segment. When a
segment is taken from the cuvette compartment, the cuvette feeder pushes the next
segment to the loading position.

When there are less than 10 segments in the compartment, the analyzing unit,
though keeping working, will give an alarm. Seeing the alarm, you should, if you still
have more tests to do, add more segments to the compartment.

When all the segments in the compartment are used, the analyzing unit will stop first
loading cuvette segments, and then stop dispensing samples when all the segments
on the reaction disk are used. But it will keep working until all the scheduled tests are
done.

Present the cuvette segment to the entry of the cuvette feeder as Figure 1-13 shows
and gently push it toward inside. Load all the segments one by one until they are all
loaded. When you are done, click the Load Cuvette button (refer to 4.1.4.2 Reaction
Disk tab for details) to command the cuvette feeder to press the newly added
segments forward. Be sure not to load the segments as Figure 1-14 shows.

NOTE:
When loading cuvette segments to the entry of the cuvette
compartment, be sure to keep them upright. Remove the inclined
segments immediately upon seeing them and reload them the right
way.
Be sure to load the segments to the cuvette compartment in the right
direction (Figure 1-13). Reversely loaded segments (Figure 1-14) may
jam the loader and halts the analyzing unit.
Do not click the Load Cuvette button before completing adding all the
cuvette segments to the compartment. Otherwise, the loading
operation may fail.

Figure 1-13 Loading cuvette segments the right way

1-13
System Description

Figure 1-14 Loading cuvette segments the wrong way

Open the middle door to load/remove the used-cuvette bucket.

BIOHAZARD:
Be sure to load the used-cuvette bucket to the designated position on
the bottom plate. Otherwise, the used-cuvette may not be dumped into
the bucket and the spilled reaction mixture may contaminate and
corrode the cabinet.
Be sure to wear gloves, goggles and protective clothing when loading
or removing the used-cuvette bucket.
Be sure to dispose of the used-cuvette in compliance with the local
regulations.

1.1.1.8 Reaction Disk Assembly

The reaction disk assembly is composed of a reaction disk and a


temperature-controlled chamber, both of which are inside the analyzing unit.

The reaction disk holds the reaction cuvettes, in which the sample reacts with the
reagents and colorimetric readings are taken.

The temperature-controlled chamber provides for the reaction an environment of


consistent temperature.

The reaction disk can hold 8 cuvette segments (80 cuvettes). The cuvettes adopted
are

Disposable;
5mm6mm25mm (5mm optical path);
750l (capable of holding 180-500l of the reaction mixture).

The temperature chamber keeps the temperature at 370.3 with 0.1


fluctuation.

1-14
System Description

BIOHAZARD:
Be sure to dispose of the used cuvettes in compliance with the local
regulations.

CAUTION:
The reaction cuvettes should not be re-used. Otherwise, the system
performance may be degraded.

During the analyzing process, the reaction disk rotates (counter-clock wise only) to
cuvette-loading position, sample-dispensing position, reagent-dispensing position, or
mixing position as needed. The colorimetric readings are taken when the specified
cuvette passes through the optical axis.

1.1.1.9 Photometric System

The photometric system measures the absorbance of the reacting liquid in the
reaction cuvettes. It consists of a measurement optical system and a reference
optical system, the former providing 9 monochromatic lights and the latter
compensating the former.

Specifications
9 optical paths with fixed interference filters.
Wavelengths: 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm,
700nm
Half band-width: 12nm
Wavelength accuracy: 2nm
Measurement range: 0.1-4.0 Abs
Lamp: 12V 50VA tungsten-halogen

Movement
The measurement optical system consists of 9 optical paths. The absorbance is
taken when the reaction cuvette rotates to the correspond path. The analyzer can
measure the absorbance of 9 cuvettes simultaneously.

1.1.1.10 ISE Unit (optional)

The ISE unit (optional) consists of ISE module, pump module and reagent module,
as shown in the figure below.

1-15
System Description

Figure 1-15 ISE unit (optional)

Pump module

ISE module

Reagent module to be
placed here

The ISE unit (optional) measures the concentration of Li+, Na+, K+ and Cl- in serum,
plasma and diluted urine.

The volume of the serum or plasma sample is 70l and that of the diluted urine
sample is 140l. The dilution ratio of the urine sample is 1:10 (1 part of urine sample
and 9 parts of urine diluent).

There are five electrodes including Li+, Na+, K+, Cl- and reference electrodes in the
ISE module. If the Li+ electrode is not selected, a spacer electrode is substituted for
it.

Reagent module is integrated with Calibrant A, Calibrant B, waste containers and a


chip which indicates the volume of the reagents.

1.1.2 Operation Unit


The operation unit is a computer with the Operating Software of the analyzer
installed. It manages test applications, registration, generation of the work list,
monitoring of reactions, calculation and data management (entering, storing and
searching).

1.1.3 Output Unit


The output unit is a printer that prints out the test results.

1-16
System Description

1.2 Software Introduction

1.2.1 Software Interface


The interface of the Operating Software of BS-300 Chemistry Analyzer (hereinafter
referred to as the operating software) is shown in Figure 1-16.

Figure 1-16 Interface of the operating software

Header area
Displays the names of the operating software and the manufacturer.

Title bar
Displays the title of the current display area.

Menu area
Displays: [Routine], [Parameters], [Data], [System], [Help]. You can move the
mouse to any of the listed menu and click the left button of the mouse (hereinafter
referred to as click) to display its pull-down menu.

Pull-down menu
You can click the listed item of the pull-down menu to perform certain operation.

Small buttons
Click the small button to view the user guide. You can also click Help Guide
to view the user guide. Refer to 4.5.2 Guide for details.

Click the small button to exit the operating software. You can also click Routine
Exit to exit the software. Refer to 4.1.8 Exit for details.

1-17
System Description

Shortcut buttons area

Totally 13 buttons are available, which are: , , , ,

, , , , , , , , . You can
click any of the buttons to perform certain operation quickly.

Display area
Displays settings, results, graphics and so on.

System status area


Displays the system status and reaction temperature.

Warning messages area


Displays the warning messages, including malfunction alarm messages and prompt
messages for mis-operation. When seeing the warning messages, refer to 6
Troubleshooting for solutions.

Period area
Displays how many periods (every period lasts 12 seconds) of test have passed
since the beginning of the first test after the latest power-on.

Time area
Displays current date and time.

Operator area
Displays the current operator who operates the system.

1.2.2 Main Interface Components


Dialog box
The dialog box is one of the most common interfaces. See the following example:

Tab
See the figure below for examples of the tab. Click the tab and you can access the
screen that it indexes.

1-18
System Description

Combo box
Click and a pull-down list will display, as the figure below shows. Click the
desired item to select it.

Combo box

Option button
Click the option button to select the option it represents. Note that for a given group
of option buttons, you can only select one of them.

Option button

Check box
Click the check box to select the option it represents and click it again to de-select it.
Note that for a given group of check boxes, you can choose more than one of them
at one time.

1-19
System Description

Check box

Edit box
You can enter characters from the keyboard in the edit box.

Edit box

Button
Click a button and you can access the function it indexes.

Scroll bar
When the contents to be displayed are too many for one screen, you can use the
scroll bar to see the hidden contents. Move the cursor to the scroll bar, press left
button of the mouse and hold it, then you can drag the scroll bar upward or
downward to see the hidden contents. Or you can press the key such as Page Up,
Page Down,oron the keyboard to see the hidden contents.

Scroll bar

1-20
System Description

1.2.3 Mouse Operations


Clicking
In this manual, clicking refers to moving the pointer of the mouse to the desired
item and click the left button of the mouse.

Double-clicking
In this manual, double-clicking refers to moving the pointer of the mouse to the
desired item and click the left button of the mouse twice swiftly.

Working with the keyboard


When you wan to select multiple targets, you need to use the mouse together with
the keyboard.

To select discontinuous targets:


Press and hold CTRL and click them one by one. Release CTRL when you have
selected all the targets.

To select continuous targets:


Click the first (or last) of the desired targets, and then press and hold SHIFT and
click the last (or the first) of the desired targets. Release SHIFT when you have
selected all the targets.

1-21
2 Installation

WARNING:
The BS-300 should only be installed by Mindray-authorized personnel
only.

To facilitate the installation, you should prepare a place meeting the requirements
specified in 2.2 Installation Requirements. If you need to move the analyzer to
another site, please contact Mindray Customer Service Department or your local
distributor, who are the appropriate people for the moving job.

2.1 Unpacking
When you receive your analyzer, carefully inspect the package. If you see any signs
of mishandling or damage, file a claim immediately with Mindray Customer Service
Department or your local distributor.

After opening the package, check the delivered goods against the packing list as
well as the appearance of the analyzer. If you find anything missing or damaged,
alert Mindray Customer Service Department or your local distributor immediately.

2.2 Installation Requirements

CAUTION:
Make sure the analyzer is installed in a place meeting the following
requirements. Otherwise, it will not perform as promised.

2.2.1 Installation Environment Requirements


The BS-300 is for indoor use only.
The bearing platform (or ground) should be level (gradient less than 1/200).
The bearing platform (or ground) should be able to bear 500Kg weight.
The installation site should be well ventilated.

2-1
Installation

CAUTION:
The analyzer radiates heat from its rear side when operating. A
well-ventilated environment helps keep the room temperature stable.
Use ventilation equipment if necessary. But if so, be sure not to
expose the analyzer to the direct draft that may lead to unreliable
results.

The site should be free of dusts as much as possible.


The site should not be in direct sun.
The site should not be near a heat or draft source.
The site should be free of corrosive gas and flammable gas.
The bearing platform (or ground) should be free of vibration.
The site should not be disturbed by large noise or power supply.
The analyzer should not be placed near brush-type motors and electrical contacts
that are frequently turned on and off.
Do not use such devices as mobile phones or radio transmitters that generate
electromagnetic waves near the analyzer.
The above-sea-level height of the site should be lower than 2000 meters.

2.2.2 Power Requirements


Power supply: 100-130V/200-240V~, 50/60Hz, three-wire power cord and
properly grounded.
The analyzer should be connected to a properly-grounded power socket, whose
maximum power consumption should be 1kVA and neutral-to-ground voltage
6V. If possible, connect the analyzer to a power line specially designed for
medical instrumentation.
The distance between the power socket and the analyzer should be less than 5
meters.

WARNING:
Make sure the power socket is grounded correctly. Improper grounding
may lead to electric shock and/or equipment damage.
Be sure to connect the analyzer to a power socket that meets the
above-mentioned requirements and has a proper fuse installed.

2.2.3 Temperature and Humidity Requirements

Ambient temperature: 15-30, with fluctuation less than 2/H.


Relative humidity: 35%RH-80%RH, no condensation.

2-2
Installation

CAUTION:
When used in conditions other than the specified, the analyzer may
not provide reliable test results. If the temperature or relative humidity
does not meet the above-mentioned requirements, be sure to use
air-conditioning equipment.

2.2.4 Water Supply and Drain Requirements


The water must meet requirements of the CAP Type II water.
The water temperature should be within 5-32.

BIOHAZARD:
Be sure to dispose of the waste according to the local regulations.

CAUTION:
The water to be used on the analyzer must meet the requirements of
the CAP Type II water. Insufficiently purified water may result in
misleading measurement.

2.2.5 Space and Accessibility Requirements


See the figure below for the space and accessibility requirements.

Figure 2-1 Space and accessibility requirements

WALL
Minimum500

Maximum2500

Operation
Unit
710

Analyzing Unit

980
F
Minimum500

R
O
N
T

Minimum500 Minimum500
Unit: mm

2-3
Installation

2.3 Connecting the Tanks

BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the below checks.

2.3.1 Connecting the Deionized Water Tank


1 Place the ANALYZING UNIT POWER to OFF.
2 Open the lower door of the analyzing unit.
3 Pull out the extractable plate.
4 Put the pickup tubes and the sensor into the deionized water tank, then
close the cap of the deionized water tank clockwise.
5 Put the deionized water tank on the right of the extractable plate.
6 Push the extractable plate back in.
7 Plug the green connector to its counterpart on the analyzing unit and turn
the connector clockwise until secure.

Deionized Water Connector Deionized Water Connector


(Green) (Red) BNC Connector

Pickup Tube

8 Plug the red connector to its counterpart on the analyzing unit and turn
the connector clockwise until secure.
9 Plug the BNC connector to its counterpart on the analyzing unit and turn
the connector clockwise until secure.
10 Close the lower door.

2.3.2 Connecting the Waste Tank


1 Place the ANALYZING UNIT POWER to OFF.
2 Open the lower door of the analyzing unit.
3 Pull out the extractable plate.
4 Put the waste tube and the sensor into the waste tank, then close the cap
of the waste tank clockwise.
5 Put the waste tank on the left of the extractable plate.

2-4
Installation

6 Push the extractable plate back in.


7 Keep pressing the pin on the connector and grab the waste tap and insert
it to the connector.
BNC Connector Waste Connector

Pin

Waste Tap
Waste Tube

8 Plug the BNC connector to its counterpart on the analyzing unit and turn
the connector clockwise until secure.
9 Close the lower door.

2.4 ISE Unit (optional) Installation

BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the below checks.

CAUTION:
Use the consumables recommended by Mindray. Other consumables
may degrade the system performance.

2-5
Installation

2.4.1 Installing Electrodes

NOTE:
The electrode storage for Li and K is different from the other
electrodes. For there is some certain kind of solution in the lumen of
the electrode, a tape covering both ends of the lumen can be seen. So
whenever unpacking these electrodes, remove the tape first. If some of
the solution leaks outside of the electrode, it should be wiped before
installation.
Each electrode including the reference electrode has an O-ring on one
side of its lumen, so ensure the side with the O-ring towards up when
installing the electrodes. If the O-ring is missing, replace with a new
one. Extra two O-rings are supplied in the electrode boxes.

Connection Pins

O-ring

1 Place the POWER to OFF.


2 Open the ISE unit (optional) door.
3 Unscrew the screw of the ISE module cover by hand.

2-6
Installation

4 The reference electrode is the first to be installed.


Open the electrode from its protective packaging and remove the insert
from the lumen of the reference electrode.
Make sure that the red sphere in the reference electrode floats on the top
of the internal fill solution in the reservoir.

Red sphere

Insert

Place the reference electrode inside the housing by pressing down the
compression plate and push it straight against the back of the housing.
Release the compression plate and ensure the electrode cannot be easily
moved.

Sample entry port

Li+ electrode (or Spacer electrode)


Na+ electrode
K+ electrode
Cl- electrode
Reference electrode
Compression plate

Housing

5 Remove the Chloride electrode from its protective packaging and place it
in the ISE module housing in the same way as the reference electrode.
6 Repeat the process for the Potassium electrode.
7 Repeat the process for the Sodium electrode.
8 Repeat the process for the Lithium electrode. If you do not select Lithium
electrode, repeat the process for the spacer electrode.
9 Push all the electrodes simultaneously to ensure they are in correct
alignment.
All the electrodes for Li (spacer electrode included), Na, K and Cl are the
same size and shape. Connection pins at the rear of each electrode are
different which ensures that the electrodes are inserted in the correct
order. So if one of the electrodes can not be easily push into the housing,
check the electrode first then repeat the installation process.
10 Screw the screw of the ISE module cover.
11 Close the ISE unit door.

2-7
Installation

2.4.2 Installing Reagent Module


1 Place the POWER to OFF.
2 Open the ISE unit (optional) door.
3 Take off the three red caps from the reagent module. Put the wand on the
module, press it down to connect to the module firmly. Then put the
module with the wand into the ISE unit.
4 Place the POWER to ON.
5 Restart the operating software, or if it is running, enter the System
Maintenance screen and click the Download Settings button at the
MotionOthers tab.
6 Enter the ISE screen and click the Replace button.

7 Enter the ISE screen of the operating software.

2-8
Installation

8 Enter 20 into the edit box to the right of Time(s) and click the Purge AB
button.
9 Close the ISE unit door.

2-9
3 Basic Operations
This chapter provides step-by-step procedures for operating the analyzer for basic
tasks.

3.1 Daily Procedure

3-1
Basic Operations

3.2 Preparing for Analysis

3.2.1 Checking before Startup


Do the followings checks before starting the analyzer.

BIOHAZARD:
Wear gloves, goggles and protective clothing when performing the
below checks.

Check the power supply and make sure it can supply proper voltage for the
analyzer.
Check the connections among the analyzing unit, operation unit and printer. Make
sure the connections are right and secure. Check the power cords of the
analyzing unit, operation unit and printer and make sure they are well connected
to power sockets.
Check and make sure sufficient printing paper is prepared for the printer.
Check the sample and reagent syringes for leakage as instructed by 5.2.1
Checking Sample/Reagent Syringes.
Check the sample and reagent probes and make sure there is no fluid hanging on
the probe tips. If there is, please contact Mindray Customer Service Department
or your local distributor.
Check the connections of the deionized water and wastewater as instructed by
5.2.6 Checking Connection of Deionized Water and 5.2.7 Checking
Connection of Wastewater.
Check the sample probe and make sure it is neither dirty nor bent. If it is dirty,
wash it as instructed by 5.3.1 Washing Sample Probe; if it is bent, replace it as
instructed by 5.6.3 Replacing Sample Probe.
Check the reagent probe and make sure it is neither dirty nor bent. If it is dirty,
wash it as instructed by 5.3.2 Washing Reagent Probe; if it is bent, replace it as
instructed by 5.6.4 Replacing Reagent Probe.
Check the mixing bar and make sure it is neither dirty nor bent. If it is dirty, wash it
as instructed by 5.3.3 Washing Mixing Bar; if it is bent, replace it as instructed
by 5.6.5 Replacing Mixing Bar.
Check the cuvette compartment and make sure there are sufficient cuvettes in it.
If the cuvettes are not sufficient, add more cuvettes as instructed by 1.1.1.7
Reaction Cuvette Loader.

CAUTION:
You must prepare enough cuvettes for the upcoming analysis. If you
find the cuvettes are not enough, add more immediately.
Sometimes the new cuvettes are not as smooth as they should be.
Check the surface of the cuvettes for bumps and take necessary
measures to remove the found bumps before loading them to the
compartment. Otherwise, they may jam the analyzing unit.

3-2
Basic Operations

Check and make sure there is enough deionized water in the deionized water
tank as instructed by 5.2.8 Checking Remaining Deionized Water.

CAUTION:
You must prepare enough deionized water. If you find the deionized
water is not enough, add more deionized water to the tank.

Check the waste tank and make sure it is empty. If not, empty the tank as
instructed by 5.2.9 Emptying Waste Tank.
Check the used-cuvettes bucket and make sure it is empty. If not, empty the
bucket as instructed by 5.2.10 Emptying Used-Cuvettes Bucket.

3.2.2 Power-on
Power up the analyzer in the sequence presented below:

Place the POWER to ON.


Place the ANALYZING UNIT POWER to ON.
Press the power button of the monitor of the operation unit.
Press the power button of the computer of the operation unit.
Press the power button of the printer.

3.2.3 Starting the Operating Software

NOTE:
Ensure the hibernation function of the display is unavailable:
After entering the display properties screen, select Never for Turn off
monitor, Turn off hard disks and System standby on the Power
Schemes tab and do not select Enable hibernation on the Hibernate
tab.

After you have logged on the WINDOWS operating system, you may

Double-click the icon on the desktop to start the operating software; or


Click [START] [PROGRAM] or [ALL PROGRAMGS][BS-300 Chemistry
Analyzer Operating Software][BS-300 Chemistry Analyzer Operating Software]
to start the operating software.

The operating software will do the initialization procedure in the sequence presented
below:

Initializing system environment.


Initializing database.
Initializing printer.
Downloading system settings from the analyzing unit.

3-3
Basic Operations

Shaking hands with the analyzing unit.

The system does the startup and warm-up procedures in the sequence presented
below and the analyzer will then enter the standby state:

Initializing database, system environment and printer; connecting to the LIS


(Laboratory Information System. If the LIS server is configured.); downloading
settings; shaking hands with the modules; communicating with the ISE unit (if
configured) and the built-in bar code scanner (if configured); running self-tests
and resetting the modules. A dialog box will then pop up to ask you to enter your
username and password.

NOTE:
You need to enter proper username and password to access the
control system. Use the initial username and password (username:
administrator; password: bs300) if this is the first time the control
system being used and then you may edit your own username and
password as instructed by 4.4.4 User.

Measuring dark currents;


Turning on the lamp of the photometer and loading the first cuvette segment to
measure the backgrounds, which reflects the status of the lamp;
Loading new cuvettes, taking their absorbance and marking the dirty ones;
Checking the status of the deionized water tank and waste tank and checking the
status of the cuvette compartment;
Performing the default washing procedure to wash the exteriors and interiors of
the sample probe, reagent probe and mixing bar with deionized water;
Checking the status of the reaction disk repeatedly and displaying its temperature
on the main interface. When the warm-up process is still under way, the status
box in the upper right corner displays Incubation and when the expected
temperature is reached it displays Idle, indicating the analyzing unit is on standby.

CAUTION:
You may start the analysis once the initialization is finished. However,
Mindray recommends you wait until the temperature and photometric
system is fully stabilized, which is when the status box displays Idle
and the analyzer has been turned on for 30 minutes.

3.2.4 Setting Up the BS-300


The analyzer will not function properly unless it is properly set up.

You must complete all the following settings if this is the first time the analyzer being
used.

Before requesting the tests, you must finish the following settings:

To set the options regarding the basic parameters of the system, refer to 4.4.2
Setup.

3-4
Basic Operations

To set the options regarding the hospital information, refer to 4.4.3 Hospital.
To set the options regarding test parameters and reference, refer to 4.2.1.1
Parameters tab and 4.2.1.2 Reference Range tab.
To set the options regarding calibrators and calibration, refer to 4.2.3.1 Calibrator
tab and 4.2.3.2 Rule tab.
To set the options regarding controls, QC rule and QC test, refer to 4.2.4.1
Control tab, 4.2.4.2 Rule tab and 4.2.4.3 Setup tab.
To set the options regarding the reagent parameters, refer to 4.2.2.1 Reagent
Disk tab.
To set the options regarding the carryover information among tests, refer to
4.2.1.5 Carryover tab.
To set the options regarding the printing parameters, refer to 4.4.7 Print.

3.2.5 Preparing for Analysis

3.2.5.1 Preparing Reagents

Load reagent bottles to their assigned positions on the reagent disk. Open the
bottles.

3.2.5.2 Preparing Distilled Water

Load enough distilled water to position W on the sample disk.

Load enough distilled water to position 49 on the reagent disk.

Load enough urine diluent to position 50 on the reagent disk if ISE tests to be run.
Note that the urine diluent recommended by Mindray is used.

3.2.5.3 Checking Remaining Reagents

Check remaining reagents if necessary. Refer to 4.2.2.2 Inventory tab for details.

3.3 Starting Analysis

3.3.1 Calibration
You must calibrate if this is the first time the analyzer being used.

CAUTION:
You need to run the calibration again when you change reagent lots,
test parameters, lamp or other conditions.

3-5
Basic Operations

To request calibrations, refer to 4.1.2 Calibration.

After requesting calibrations, you should load corresponding calibrators to their


assigned positions on the sample disk.

To run calibrations, refer to 4.1.5 Start.

A calibration run experiences 4 states: Requested, Waiting, Running, Finished.


You can check the results of the finished runs in the following ways.

To view the results at the StatusCalib. Test, refer to 4.1.4.4 Calib. Test tab.
To view the results at the StatusSample Disk, refer to 4.1.4.1 Sample Disk
tab.
To view the results at the CalibrationResult, refer to 4.2.3.3 Result tab.
If the ISE unit is connected, refer to 4.4.11.1 Calibration for detailed information
about running ISE calibration and check its results.

3.3.2 QC
To request QCs, refer to 4.1.3 QC.

After requesting QCs, you should load corresponding controls to their assigned
positions on the sample disk.

NOTE:
If the ISE unit (optional) is connected, ISE QC should be performed.
Please use Mindray-recommended controls for ISE QC.

To run QCs, refer to 4.1.5 Start.

A QC run experiences 4 states: Requested, Waiting, Running, Finished. You can


check the results of the finished runs in the following ways.

To view the results at the StatusQC Test, refer to 4.1.4.5 QC Test tab.
To view the results at the StatusSample Disk, refer to 4.1.4.1 Sample Disk
tab.
To view the results at the QC Real-time Result /Daily Result /Day-to-day
Result, refer to 4.2.4.4 Real-time Result tab, 4.2.4.5 Daily Result tab or 4.2.4.6
Day-to-day Result tab.

3.3.3 Samples
To request samples, refer to 4.1.1 Samples.

You can request a STAT sample run in the way similar to an ordinary run, except that
you have to tick the check box to the right of STAT, as Figure3-1 shows. The
requested STAT run will be inserted into the current work list as Waiting and has
higher priority than other tests.

3-6
Basic Operations

Figure3-1 Requesting a STAT Sample Run

After requesting, load the samples to their corresponding positions on the sample
disk.

To run samples, refer to 4.1.5 Start.

A sample run experiences 4 states: Requested, Waiting, Running, Finished. You


can check the results of the finished runs in the following ways.

To view the results at the StatusSample Test, refer to 4.1.4.3 Sample Test
tab.
To view the results at the StatusSample Disk, refer to 4.1.4.1 Sample Disk
tab.
To view the results at the DataResultResults, refer to 4.3.4.1 Results tab.

3.3.4 Editing Results of Sample Runs


Skip this section if you do not need to edit the sample run results.

CAUTION:
Sample run results can only be edited by authorized personnel.

To edit one or more sample run results, refer to 4.1.4.3 Sample Test tab or 4.3.4.1
Results tab for details.

To revise one or more sample run results, refer to 4.1.4.3 Sample Test tab or 4.3.4.1
Results tab for details.

3.3.5 Printing Results of Sample Runs


To print sample results, refer to 4.1.4.3 Sample Test tab or 4.3.4.1 Results tab for
details.

3.4 Finishing Analysis

3.4.1 Exiting the Operating Software


To exit the operating software, refer to 4.1.8 Exit for details.

3-7
Basic Operations

3.4.2 Shutdown
After exiting the Windows operating system, switch off the following powers in the
presented order:

Turn off the printer;


Turn off the monitor of the operation unit;
Place the ANALYZING UNIT POWER to OFF.

NOTE:
The refrigerator still functions after the ANALYZING UNIT POWER is
placed to OFF. To shut down the refrigerator, place the POWER to
OFF.

3.4.3 Checking after Shutdown


Cap every reagent bottle on the reagent disk.
Remove the calibrators, controls, distilled water and samples from the sample
disk.
Check the surface of the analyzing unit for stains and wipe them off with clean
soft cloth.
Check the waste tank and make sure it is empty. If not, empty the tank as
instructed by 5.2.9 Emptying Waste Tank.
Check the used-cuvette bucket and make sure it is empty. If not, empty the
bucket as instructed by 5.2.10 Emptying Used-Cuvettes Bucket.

BIOHAZARD:
Exercise with caution. Wear gloves, goggles and protective clothing.

3-8
4 Advanced Operations
The chapter presents an introduction of the BS-300 operating software by menus.

The main interface of the operating software has five major menu groups:

Routine
Parameters
Data
System
Help

4.1 Routine

4.1.1 Samples
Click Routine Sample, or press F1 to enter the Request Sample screen, as
Figure4-1 shows. You can also enter the Request Sample screen by clicking the

short-cut button .

4-1
Advanced Operations

Figure4-1 Request Sample screen

The Request Sample screen consists of three zones:

Sample zone displays information regarding the samples;


Assays and Profiles zone displays information regarding the assays and
profiles;
Operating buttons zone displays such buttons as Browse, Edit, etc.

The Assays zone lists all available tests. You cannot request the assays followed by

a forbidden mark, such as . The double-reagent assays, whose reagent


positions have already been set, are yellow-framed; the double-reagent assays,
whose reagent positions have not been set yet, are gray-framed. The selected assay
is followed by a tick. Only if an assay, whose dilution parameters have been set, is
selected after Dilution is selected, the tick is red. Otherwise, the tick is green.

The sample parameters are interpreted as follows.

Parameter What it does How to set


Sample Defines which virtual sample disk Click to the right of
Disk the sample is located on. Sample Disk and select the
desired disk from the pull
down list.
Sample No. Defines the identification number Use the system-generated
of the sample or samples. You can ID(s) or enter the desired
enter the desired integers (1-9000) ID(s) manually.
into the edit boxes respectively as
the start (left) and end (right) IDs.
If you are entering only one
samples ID ensure the two IDs are
the same.
Index A system-generated number for
fast search.

4-2
Advanced Operations

Parameter What it does How to set


Sample ID Bar code of the sample. Enter the bar code from the
bar code scanner
automatically, enter it
manually, or leave it blank.
Position Defines the samples position on Use the system-generated
the selected sample disk. For number, click to the
STAT samples, Mindray right of Position to select
recommends they be placed on the desired position, or
positions E1-E5. obtain it by the bar code
scanner.
For multiple samples, the
selected position represents
that of the first sample and
the analyzer will generate
positions for the rest of the
samples.
Tube Defines which type of tube holds Click to the right of
the sample. Tube and select either
Large tube or Small tube
from the pull down list.
Type Defines the sample type. Click to the right of
Type and select one of the
following four types: Serum,
Plasma, Urine and Others.
Replicates Defines for how many times the Enter the desired integer
analyzer should run this sample.
(1-10) or click the button
The default value is 1.
to the right of Replicates
until the desired integer is
reached.
STAT Defines whether the sample is a Tick the check box to the
stat sample. The analyzer will add right of STAT.
the stat sample to the current work
list.
S. Blank Applies to the tests that use Tick the check box to the
Endpoint method. right of Blank to enable this
function and click it again to
Defines whether to replace the disable it.
reagents with distilled water and
take the absorbance of the mixture
before formally analyzing the
reagents-sample mixture.
This parameter is only available for
clinical chemistry tests.
Dilution Defines whether to dilute the Tick the check box to the
sample before analyzing it. right of Dilution to enable
this function and click it
This parameter is only available for again to disable it.
clinical chemistry tests.

4-3
Advanced Operations

To scan bar codes

NOTE:
If the built-in bar code scanner (optional) is not connected, the function
is unavailable.

1 Click the Scan button and a dialog box pops up.

2 At the Scan Sample ID screen, select sample(s) you want to be scanned.


If you select All Positions, the scanner will scan all the positions in the
sample disk; if you select Selected Positions, you need to enter position
No. into the edit boxes to the right of Range and the scanner will scan the
positions you just entered.
3 Click the Start button and the scanner will begin to work. The scanning
results will be displayed at the Result field.

4-4
Advanced Operations

4 If you want to download information of the scanned samples from the LIS
server, click the Download button.

NOTE:
The Download button is only available if you select
Real-time Download at the System SetupHost
Communication screen.

5 Click the Close button.


If there is a sample whose bar code is the same with the other one or
which is placed in a position where bar code should not be scanned, the
Repeated Sample ID dialog box will pop up to remind you.
In this case, you need to click the Close button and check the bar code or
sample position and take corresponding measures.

To download all sample information from LIS server

NOTE:
If there is no LIS server connected, the function is unavailable.

1 Click the Download All button and a dialog box pops up.

4-5
Advanced Operations

2 Click the Start button and the system will download all the sample
information from the LIS server.
If you want to cancel the downloading, click the Cancel button.

3 After the downloading is completed or cancelled, click the Close button.


The system will save the information that has been downloaded.

If there are samples that have been requested but not able to test, a
dialog box which is shown as the figure below will pop up to remind you.
You can click the Close button to exit the dialog box. Refer to the
following content To view requested but untestable samples for details.

To download single sample information from LIS server

NOTE:
If there is no LIS server connected, the function is unavailable.

4-6
Advanced Operations

1 Enter the bar code into the edit box to the right of Sample ID for the
current sample, or click the Browse button and find the sample whose
sample information you want to get from the LIS server.
2 Click the Download Single button to get sample information
corresponding to the bar code from the LIS server.

NOTE:
The system will download information only for the sample
that meets any of the following conditions:
(1) It has not been requested before;
(2) All the tests requested for it before are at Requested
status and Overwrite has been selected at the Host
Communication screen (refer to 4.4.2.4 Host
Communication tab).

To request samples

1 Set the parameters as instructed above.


2 Click the desired test to select it; click it again to de-select it. The selected test will
be followed by a green tick. You may select multiple tests for a sample run.

You may also request a pre-defined profile or profiles and the selected profile or
profiles will also be followed by a green tick.

3 Click the Request button to request a sample run.

4-7
Advanced Operations

NOTE:
Routine sample can be requested on any virtual sample disk.
STAT sample can only be requested on the default virtual sample disk.
After the system is powered on, the sample disk 1 is the default at that
time.
If there are any sample runs in Waiting or Running status on a sample
disk, the disk is the default at that time.
If the system just finishes the runs on a sample disk, the disk is the
default at that time.
If you request a sample whose sequence number has been used on
the current day (the same No., but the different index) and youve set
the Sample ID and Type (sample type) for it, the system will consider
the Sample ID and Type of the former one as those of the current
sample. You can change the two parameters only at the
Demographics screen.

To view requested but untestable samples

1 Click the Abnormal Sample button and a dialog box pops up. If there are
samples that have been requested but not able to test, they will be listed
in the dialog box.

2 Select a sample to see the tests it has been requested for.

4-8
Advanced Operations

3 If you want to run the untestable test, you have to find the cause why it is
untestable and take some actions. For example, if the test ALB has not
been calibrated, it will become testable after being calibrated.
4 If you want to add the sample to the requested test list, tick the combo
box to the left of the sample and click the Test button. If you want to
select all the samples, tick the combo box to the left of Select All.
You can see the selected sample at the Test Status screen.
If there are untestable tests for the selected sample, the tests will be
discarded and you will not see them at the Sample Test screen.
5 Click the Close button.

To view a requested sample run

1 Click the Browse button.


2 Click or until the desired sample is reached.

NOTE:
Click to go to the previous sample run.

Click to go to the next sample run.

3 After you have finished browsing samples, click the Browse button again
to quit the browsing mode.

To edit a requested sample run

1 Click the Browse button.


2 Click or until the desired sample is reached.

NOTE:
Click to go to the previous sample run.

Click to go to the next sample run.

3 Click the Edit button.

4-9
Advanced Operations

4 If you do not want to change the sample information, go directly to the


next step;
Otherwise, edit the sample information.

NOTE:
You can edit routine sample information here only if the
sample is in Requested status.
You cannot edit STAT sample information here because the
STAT sample is in Waiting status after it is requested.
Sample Disk, Sample No. and Replicates cant be edited.
You can edit Sample Type and Sample ID only at the
Demographics screen. Refer to 4.3.3 Demographics for
details.

5 If you do not want to change the assays for the sample, go directly to the
next step.
Otherwise, select tests and profiles in the Assays area or Profiles area.
If you want to add an assay or a profile, select it; if you want to delete an
assay or a profile, de-select it.

NOTE:
You can change the tests or profiles for the sample in any
status.
You can only delete a test or a profile for the sample in
Requested status.

6 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.
7 Repeat from step 2 to edit another sample run; or click the Browse button
again to quit the browsing mode.

To delete a sample run in Requested status

1 Click the Browse button.


2 Click or until the desired sample is reached.

NOTE:
Click to go to the previous sample run.

Click to go to the next sample run.

4-10
Advanced Operations

3 Click the Delete button to delete the selected sample.

NOTE:
You can only delete the routine sample that is in Requested
status here.
You cannot delete the STAT sample here because the STAT
sample is in Waiting status after it is requested. If you want
to delete a requested STAT sample, refer to 4.1.4.1 Sample
Disk tab or 4.1.4.3 Sample Test tab for details.

4 Repeat from step 2 to delete another sample run; or click the Browse
button again to quit the browsing mode.

To enter the patient information screen

1 Click the Demographics button and the Demographics window will pop
up.
Refer to 4.3.3 Demographics for details.

To exit the Request Sample screen

1 Click the Close button.

4.1.2 Calibration

Click Routine Calibration, or press F2, or click the short-cut button to


enter the Request Calibration screen, as Figure4-2 shows.

Figure4-2 Request Calibration screen

4-11
Advanced Operations

At the Request Calibration screen, all the available tests are listed. The tests
followed by forbidden marks, such as , are those whose calibration
parameters have not been set yet and cannot be requested for calibrations. The
yellow-framed tests are the double-reagent tests whose reagent positions have
already been set. The gray-famed tests are the double-reagent tests whose reagent
positions have not been set yet. The selected test will be followed by a green tick.

You can select the desired sample disk (virtual disk) from the pull down list of the
Sample Disk filed. The default sample disk is Sample Disk1. The adopted
calibration rule and calibrator will be displayed in the Method and Calibrator fields
respectively.

To request a calibration run

1 Click below Sample Disk and select the desired sample disk from the
pull down list.
2 Click the desired test to select it; click it again to de-select it. The select
test will be followed by a green tick. You can select multiple tests for a
calibration run.

3 Click the Request button to request a calibration run.


4 Repeat steps 1 - 3 to request another calibration run, or click the Close
button to exit the Request Calibration screen.

CAUTION:
You must run the calibration if this is the first time the analyzer using
Endpoint or Fixed-time method for double-reagent tests. For the
Kinetic method, you can directly input the K Factor without running the
calibration first (see 4.2.3.2 Rule tab).
You need to run the calibration when you have changed the reagents,
reagent lots, test parameters, lamp or other analysis conditions.
Before calibrating a double-reagent test that uses Endpoint method,
the system will run reagent blank test for it automatically.

4-12
Advanced Operations

4.1.3 QC
Click Routine QC, or press the F3 to enter the Request QC screen, as Figure4-3
shows. You can also enter the Request QC screen by clicking the shortcut button

Figure4-3 Request QC screen

The screen lists all available tests. The tests whose reagent positions have not been

set yet are followed by a forbidden symbol, such as , and you cannot
request to run them. The double-reagent tests, whose reagent positions have
already been set, are yellow-framed; the double-reagent tests, whose reagent
positions have not been set yet, are gray-framed. The selected test is followed by a
green tick.

When any test is selected, its corresponding control information will be displayed in
the Controls filed. The default sample disk (virtual disk) is disk 1 and you should
select the appropriate sample disk to the actual arrangement.

NOTE:
If the auto QC function is enabled (see 4.4.2.1 Routine tab for details)
and the QC interval is not 0 (see 4.2.4.3 Setup tab for details), the
analyzer will automatically insert QC runs among sample runs and you
do not have to request additional QC runs.
If the ISE unit (optional) is connected, the ISE will be displayed in the
Assays field.
Please use Mindray-recommended controls for ISE QC.

To request a QC run

1 Click below Sample Disk and select the desired sample disk from the
pull down list.

4-13
Advanced Operations

2 Click the desired test to select it; click it again to de-select it. The selected
test will be followed by a green tick. You may select multiple tests for one
run.

.
3 Click the Request button to request a QC run.
4 Repeat steps 1 - 3 to request another QC run, or click the Close button to
exit the Request QC screen.

4.1.4 Status
Click Routine Status or press F4 to enter the Status screen, as Figure 4-4

shows. You can also enter the Status screen directly by clicking the
button.

Figure 4-4 Status screen

At the Status screen, you can see five tabs:

4-14
Advanced Operations

Sample Disk
Reaction Disk
Sample Test
Calib. Test
QC Test

4.1.4.1 Sample Disk tab

The Sample Disk tab displays current status of the sample disk.

Figure 4-5 Sample disk tab

As Figure 4-5 shows the Sample Disk tab consists of three major parts, a sample
disk diagram on the left, sample information and tests on the right and operating
buttons on the bottom.

The sample disk diagram illustrates two circles. Every position consists of two
color-coded rings, the inner ring and the outer ring:

positions with blue inner rings represent routine samples;


positions with red inner rings represent STAT samples;
positions with pink inner rings represent controls;
positions with purple inner rings represent calibrators;
positions with dark green outer rings indicate the sample is in Requested status;
positions with orange outer rings indicate the sample is in Waiting status;
positions with green outer rings indicate the sample is in Running status ;
positions with blue outer rings indicate the sample is in Finished status;
positions with gray inner and outer rings represent unoccupied positions;
position W with brown inner rings represent reagent blanks;
position W with green inner and outer rings indicates there is no reagent blank.

4-15
Advanced Operations

To check information about a sample and its included tests

1 Select the desired sample disk from the pull down list in the Sample Disk
field to check the corresponding sample and sample run information.
2 Click the desired position on the selected disk to check the corresponding
sample and its included tests.

To delete a sample in Requested or Waiting status

1 Select the desired sample disk from the pull down list in the Sample Disk
field.
2 Click the desired position on the selected disk. Note that you can only
select a sample which test(s) is (are) all in Requested or Waiting status.
3 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.
4 Click the OK button to confirm the deletion; click the Cancel button, for
any reason, to cancel deletion.

To release a position (except for the control, calibrator positions


and position W) whose scheduled run has finished

1 Select the desired sample disk from the pull down list in the Sample Disk
field.
2 Click the desired position on the selected disk. Note that you can only
select a sample which test(s) is (are) all in Finished status.
3 Click the Release button and a dialog box will pop up to ask you to
confirm the release.
4 Click the OK button to confirm the release; click the Cancel button, for
any reason, to cancel the release.

To release all the positions (except for the control, calibrator


positions and position W) whose scheduled run has finished

1 Select the desired sample disk from the pull down list in the Sample Disk
field.
2 Click the Clear button and a dialog box will pop up to ask you to confirm
the clearing.
3 Click the OK button to confirm the clearing; click the Cancel button, for
any reason, to cancel the clearing.

To rerun a sample

1 Select the desired sample disk from the pull down list in the Sample Disk
field.
2 Click the desired position on the selected disk. Note that you can only
rerun a routine or STAT sample.

4-16
Advanced Operations

3 Select the test or tests need to be included in the rerun. Click the check
box to select it. Click the check box again to de-select it.

4 To rerun the selected sample without dilution, go directly to the next step.
To dilute the selected sample before the rerun, tick the Diluted field. Only
if the dilution parameters of a test have already been set, the sample will
be diluted before the rerun.

5 Click the Rerun button to request the selected sample again.

To check the reaction curve

1 Select the desired sample disk from the pull down list in the Sample Disk
field.
2 Click the desired position on the selected disk to check the corresponding
sample and its included tests.
3 Click the desired test. Note that you can only select a Finished test.

4 Click the Reaction Curve to check the reaction curve and data of the
selected blank runs.

There are two tabs at the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
5 If you have no intention to view other results, go directly to the next step;
Otherwise, click the Previous or Next button at the Reaction Curve
window; or click the Close button and repeat from step 2.

4-17
Advanced Operations

6 Click the Close button to close the Reaction Curve window.

To exit the Status screen

1 Click the Close button to exit the Status screen.

4.1.4.2 Reaction Disk tab

The Reaction Disk tab displays current status of the reaction disk.

Figure 4-6 Reaction Disk tab

As Figure 4-6 shows, the Reaction Disk tab consists of three major parts, a reaction
disk diagram on the left, test information, status and results on the right and
operating buttons on the bottom.

The reaction disk illustrates 80 reaction cuvette positions, all of which are
color-coded.

To check the parameter, status and results

1 Click the desired cuvette position and the corresponding parameters,


status and results (for finished runs) will be displayed on the right of the
screen.

To check the reaction curve and data

1 Click the desired cuvette in the reaction disk diagram. Note that you can
only check the reaction curve and data of the Finished runs.

4-18
Advanced Operations

2 Click the Reaction Curve to check the reaction curve and data of the
selected blank runs.
There are two tabs at the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
3 If you have no intention to view other results, go directly to the next step;
Otherwise, click the Previous or Next button at the Reaction Curve
window; or click the Close button and repeat from step 1.
4 Click the Close button to close the Reaction Curve window.

To command feeder to press segments

1 After loading the segments into the cuvette compartment, click the Push
Cuvettes button to command the cuvette feeder to press the newly added
segments forward.

NOTE:
The Push Cuvettes button is only to command the cuvette feeder to
press the cuvette segments forward.

To exit the Status screen

1 Click the Close button to exit the Status screen.

4.1.4.3 Sample Test tab

Figure 4-7 Sample Test tab

As Figure 4-7 shows, the Sample Test tab consists of the following tabs:

Unfinished Samples

4-19
Advanced Operations

Results by Samples
Results by Tests

Unfinished Samples

The Unfinished Samples tab lists all the Requested, Waiting and Running samples
of the current day since the latest power-on.

Figure 4-8 Unfinished Samples tab

To view tests requested for a sample

1 Select a virtual sample disk in the Sample Disk field.


2 Select a sample in the Samples field, then you can view the tests
requested for the selected sample.

To delete a sample in Requested or Waiting status

1 Select a virtual sample disk in the Sample Disk field.


2 Click the Delete button and the Delete Sample dialog box will pop up.
3 Select one or more samples in the Delete Sample dialog box.
Click the check box to the left of the sample to select it and click the box
again to de-select it.
Click the check box to the left of All to select all the displayed samples
and click the box again to de-select them.
4 Click the Delete button to delete the select samples. After that, you can
see the status of the samples changed to Deleted.
5 Click the Close button to close the Delete Sample dialog box.

4-20
Advanced Operations

NOTE:
In the Delete Sample dialog box, you can click the Refresh button to
view current samples in Requested or Waiting status.

To view or modify patient information

1 Select a virtual sample disk in the Sample Disk field.


2 Select a sample in the Samples field.
3 Click the Demographics button to pop up the Demographics dialog box,
and you can view or modify the patient information in the dialog box (refer
to 4.3.3 Demographics for details).

To view reaction curve of a finished test

1 Select a virtual sample disk in the Sample Disk field.


2 Select a sample in the Samples field.
3 Select a test in the Finished status in the Assays field.
4 Click the Reaction Curve button to pop up the Reaction Curve screen
that shows the reaction curve of the selected test.
There are two tabs at the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
5 If you have no intention to view other results, go directly to the next step;
Otherwise, click the Previous or Next button at the Reaction Curve
screen; or click the Close button and repeat from step 2.
6 Click the Close button to close the Reaction Curve window.

To exit the Status screen

1 Click the Close button to exit the Status screen.

Results by Samples tab

The Results by Samples tab lists all the Finished samples of the current day since
the latest power-on. You can print patient reports and edit test results through this
tab.

4-21
Advanced Operations

Figure 4-9 Results by Samples tab

To select certain samples

1 Click the Select button. A dialog box will pop up.

2 Enter the number of the start sample into the first dialog box that pops up
and then click the OK button. A second dialog box will then pop up.

3 Enter the number of the end sample into the second dialog box that pops
up and click the OK button to select the two samples and those between
them.

To de-select certain samples

1 Click the Deselect button and a dialog box will pop up.

4-22
Advanced Operations

2 Enter the number of the start sample into the dialog box that pops up and
then click the OK button. Another dialog box will then pop up.

3 Enter the number of the end sample into the second dialog box that pops
up and click the OK button to de-select the two samples and those
between them.

To view reaction curve

1 Select a virtual sample disk in the Sample Disk field.


2 Select a sample in the Samples field.
3 Select a test in the Assays field.
4 Click Reaction Curve to check the reaction curve and data of the
selected test.
There are two tabs at the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
5 If you have no intention to view other results, go directly to the next step;
Otherwise, click the Previous or Next button at the Reaction Curve
screen; or click the Close button and repeat from step 2.
6 Click the Close button to close the Reaction Curve screen.

To print out a patient report

1 Select a virtual sample disk in the Sample Disk field.


2 Select the desired sample or samples in the Samples field. You can use
the Select button or select them one by one.
3 Click the Print Pat. Report button and a print preview window will pop up.
At the Print Preview window, you may choose Pr. Series (series printing)
for your impact printer; you may choose Pr. Current (printing current
page) or Pr. All (printing all pages) for other types of printers.

NOTE:
You can click (to the first page), (to the
previous page), (to the next page) or (to the
last page) to go to the desired page.
Pr. Current refers to printing the page currently displayed on
the screen.
Pr. All refers to printing all the pages of the selected sample
runs.

4-23
Advanced Operations

4 When the printing is done, click the Close button to exit the Print Preview
window.

To edit sample results

NOTE:
Sample results can only be edited by authorized personnel.

1 Select a virtual sample disk in the Sample Disk field.


2 Select a sample in the Samples field.

3 Select a test in the Assays field.

4 Click the Edit button.


5 Enter the new value in the edit box to the right of Result.

6 If you want to save the new results, click the Save button; otherwise, click
the Cancel button.

NOTE:
You can only edit sample results of the current day since the latest
power-on here.

To print daily summary

To print daily summary is to print out all the patient results of a certain day. You can
only use an impact printer for this purpose. Before printing, ensure the analyzer is
not doing anything related to an analysis.

1 Click the Print Daily Summary button and a dialog box will pop up to ask
you to confirm the printing.

4-24
Advanced Operations

2 Click the OK button to confirm the printing; click the Cancel button, for any
reason, to abort the printing.
Once you click the OK button, another dialog box will pop up to ask you to
select a printer.

3 If the printer is not an impact one, click the Cancel button;


Otherwise, click the OK button and another dialog box will pop up to ask you
to enter the date of which the patient results you want to print.

4 Enter the date (the default is the current date) and click the OK button to
print out the patient results; click the Cancel button, for any reason, to abort
the printing.
If there is no result of the entered date, a dialog box will pop up to alert you.
Click the OK button to close the dialog box and abort the printing.

To exit the Status screen

1 Click the Close button to exit the Status screen.

Results by Tests

The Results by Tests tab lists all the finished sample tests of the current day since
the latest power-on. You can print the test report and revise results through this tab.

Figure 4-10 Results by Tests tab

4-25
Advanced Operations

To select certain tests

1 Click the Select button. A dialog box will pop up.

2 Enter the number of the start test into the first dialog box that pops up and
then click the OK button. A second dialog box will then pop up.

3 Enter the number of the end test into the second dialog box that pops up
and click the OK button to select the two tests and those between them.

To de-select certain tests

1 Click the Deselect button and a dialog box will pop up.

2 Enter the number of the start test into the dialog box that pops up and
then click the OK button. Another dialog box will then pop up.

3 Enter the number of the end test into the second dialog box that pops up
and click the OK button to de-select the two tests and those between
them.

To view reaction curve

1 Select a virtual sample disk in the Sample Disk field.

4-26
Advanced Operations

2 Select the desired test.

3 Click Reaction Curve to check the reaction curve and data of the
selected test.
There are two tabs at the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
4 If you have no intention to view other results on the current virtual sample
disk, go directly to the next step;
Otherwise, click the Previous or Next button at the Reaction Curve
screen; or click the Close button and repeat from step 2.
5 Click the Close button to close the Reaction Curve screen.

To print out a test report

1 Select a virtual sample disk in the Sample Disk field.


2 Select the desired test or tests. You can use the Select button or select
them one by one.
3 Click the Print Test Report button and a print preview window will pop
up.
At the Print Preview window, you may choose Pr. Series (series printing)
for your impact printer; you may choose Pr. Current (printing current
page) or Pr. All (printing all pages) for other types of printers.
4 When the printing is done, click the Close button to exit the Print Preview
window.

To revise sample results

NOTE:
Sample results can only be revised by authorized personnel.
If you are not authorized to revise the results, a dialog box will pop up
to ask you to enter the authorized user name and password.

1 Select a virtual sample disk in the Sample Disk field.

4-27
Advanced Operations

2 Select a test or tests. You can use the Select button or select them one
by one.

3 Click the Correct button.


4 Enter values in the edit boxes to the right of Factor a: and Factor b:.

5 If you want to revise the tests with the formula Y=aX+b (Y stands for the
new result and X stands for the old result), click the Save button;
otherwise, click the Cancel button.
After clicking the Save button, you can see the test result in the Finished
Tests field have been changed to the new ones.

NOTE:
You can only revise the sample results of the current day since the
latest power-on here.

To exit the Status screen

1 Click the Close button to exit the Status screen.

4.1.4.4 Calib. Test tab

The Calib. Test tab lists the calibration runs of the current day since the latest
power-on.

4-28
Advanced Operations

Figure 4-11 Calib. Test tab

To view calibration test status

1 Select a calibration test in the Assays field. You can view the basic
information and the status of the selected test and the results of the
Finished tests.

To delete a Requested or Waiting calibration run

1 Select the desired calibration run in Requested or Waiting status. You can
only delete one run at a time.

2 Click the Delete button to delete the selected run.

To check reaction curve of calibration runs

1 Select a Finished calibration run in the Assays field.

4-29
Advanced Operations

2 Select a calibrator from the Calibration Status zone.

3 Click the Reaction Curve and a window will pop up to display the
corresponding reaction curve.

There are two tabs in the Reaction Curve screen: the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.

4 Go directly to the next step if you have no intention to check other results;
Otherwise, click the Previous or Next button to view other results (not
only the results of calibrations) or click the Close button and then repeat
from step 1.
5 Click the Close button to close the Reaction Curve window.

4-30
Advanced Operations

To view calibration curve

1 Select a Finished calibration test in the Assays field.

2 Click the Calib. Curve button to display the calibration curve and the
calibration result.

To print the displayed curve and result, click the Print button.
3 Click the Return button to exit the Calibration Curve window.

To exit the Status screen

1 Click the Close button to exit the Status screen.

4.1.4.5 QC Test tab

The QC Test tab lists the QC runs of the current day since the latest power-on.

4-31
Advanced Operations

Figure4-12 QC Test tab

To view status of QC run

1 Select a QC run in the Assays field. You can view the parameters and the
status of the selected run and the results of the Finished runs.

To view reaction curve of QC runs

1 Select a finished QC run from the Assays field.

2 Select a control from the QC Status field.

3 Click the Reaction Curve and a window will pop up to display the
corresponding reaction curve.

4-32
Advanced Operations

There are two tabs in the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.

4 Go directly to the next step if you have no intention to check other results;
Otherwise, click the Previous or Next button to view other results (not
only the results of QCs) or click the Close button and then repeat from
step 1.
5 Click the Close button to close the Reactive Curve window.

To delete a requested QC run

1 Select the desired Requested QC run. You can only delete one run at a
time.

2 Click the Delete button to delete the selected run.

4-33
Advanced Operations

To exit the Status screen

1 Click the Close button to exit the Status screen.

4.1.5 Start
Start is used to run the requested tests.

If the current tests are paused, Start will change to Resume that is used to continue
the paused tests.

Start tests

1
Click Routine Start or press F5, or click the short-cut button to
pop up the following dialog box.

2 In the Select Samples dialog box, select a virtual sample disk to the right
of Sample Disk.

NOTE:
If there are any requested STAT samples on the current
default virtual sample disk, you can only select the disk.

4-34
Advanced Operations

3 In the Select Samples dialog box, select samples to be run in the Select
sample to be run field.
If you select All Samples, go directly to the step 5; if you select Selected
Samples, go to the next step.

NOTE:
If you select All Samples, the system will run all the
Requested samples (Waiting STAT samples included),
calibrations and QC tests.
If you select Selected Samples, the system will run some
Requested samples, calibrations and QC tests. When
Selected Samples is selected, the Select Samples field is
available.

4 In the Select Samples dialog box, select tests to be run in the Select
Samples field.

If you want to run sample tests, select the check box to the left of Sample
No. and enter the sample No. in the edit boxes to the right of it. The
system will run the Requested samples whose IDs are between the two
numbers (including the two numbers).
If you want to run calibration tests, select the check box to the left of
Calibration and select tests below it.
If you want to run QC tests, select the check box to left of QC and select
tests below it.

4-35
Advanced Operations

5 If you want to continue, click the OK button; otherwise, click the Cancel
button.
When you click the OK button, a dialog box will pop up.

6 Select a virtual reagent disk in the Select Reagent Disk dialog box.
7 If you want to run the selected tests, click the OK button; otherwise, click
the Cancel button and the system will not run the selected tests.
When you click the OK button, the status of the selected tests changes to
Waiting and the system will run the tests that use the reagents on the
selected virtual disk immediately.

NOTE:
If the system finishes the tests that use the reagents on the selected
virtual reagent disk, and there are waiting tests that use reagents on
the other disk, a dialog box will pop up to remind you.
Click the OK button to exit the dialog box.
After replacing the reagent disk, click the Resume button to continue
the tests that use reagents on the other disk. Refer to the following
contents for details.

Resume tests

1 When the system is paused, click Routine Resume or press F5, or

click the short-cut button to continue the tests. The system will
take different operations according to different conditions. Refer to the
following steps for details.
2 If there is no test in Requested status and no need to replace the reagent
disk, the system will continue the tests directly.
3 If there is no test in Requested status but you need to replace the reagent
disk, the following dialog box will pop up.

After selecting a virtual reagent disk in the dialog box, click OK to


continue the tests or click Cancel to abort running.

4-36
Advanced Operations

4 If there are tests in Requested status, the following dialog box will pop up.

If you want to continue the tests currently being paused, click Cancel
button;
If you want not only to continue the tests currently being paused but also
to run the tests in Requested status, select tests and click the OK button.
After that, if there is no need to replace the reagent disk, the system will
continue the tests; if you want to replace the reagent disk, a dialog box
will pop up.

After selecting a virtual reagent disk in the dialog box, click OK to


continue the tests or click Cancel to abort running.

4.1.6 Probe Stop

Click Routine Probe Stop, or press F6, or click the short-cut button to
stop the sample and reagent probes and the mixing bar without stopping the reaction
disk. You may use this function to add more samples or reagents without stopping
the analysis. This menu is only functional when the system is testing.

NOTE:
Do not put the probes and the bar on hold for too long. Otherwise,
certain analyses may be affected.

4-37
Advanced Operations

4.1.7 Stop

Click Routine Stop, or press F8, or click the short-cut button to stop the
current run. This menu is only functional when the system is testing.

NOTE:
Mindray recommends this stop function not be used unless it is
necessary (for instance the analyzer is experiencing problems). When
the analyzer is experiencing a serious problem, the analyzer will be
stopped automatically.
After the analyzer is stopped, all the remaining runs will be invalidated
and return to the Requested status, but the finished runs will be

saved. To continue these runs, click the short-cut button .

4.1.8 Exit
When you have finished all analyses and the system is in the idle status, click

Routine Exit, or press F10, or to click the short-cut button to exit the
control system. A dialog box will pop up to ask you to confirm the exiting operation.

Click the OK button and another dialog box will pop up to ask you to load detergents.

NOTE:
Please use Mindray-recommended detergents: 0.1mol/l hydrochloric
acid, or 0.5% (V/V) Javel water.
Use of other detergents may lead to inaccurate results. Do not use
other detergents.

4-38
Advanced Operations

Ensure acid and alkaline detergents are loaded to positions 45 and 47 respectively
and then click the OK button and the analyzer will start the following shutdown
procedure:

1 Back up the data in database and unloads the used cuvettes.


2 Carry out the washing procedure: washing the sample probe, reagent probe
and mixing bar for three times respectively with the acid and alkaline
detergents.
3 Wash the analyzing unit with deionized water as defined by Times of Auto
Washing(Startup/Shutdown) at the System SetupRoutine screen.
4 Unloads the first cuvette segment.

When the shutdown procedure is done, a dialog box will pop up to remind you that
you can shutdown the ANALYZING UNIT POWER to OFF safely. Click the OK
button to exit the control software.

4.1.9 Emergency Exit


When the analyzer is experiencing a serious problem, you need to click Routine
Emergency Exit, or press CTRL+F12 to exit the control system without doing the
shutdown procedure.

NOTE:
Use this emergency exit function only when the analyzer is
experiencing a serious problem.

4.2 Parameters

4.2.1 Assay
The Assay Parameters screen is where you set parameters regarding the tests,
calculation tests, profiles, carryover and off-system tests.

To enter the Assay Parameters screen, click Parameters Assay, or press the
F11 key on the keyboard.

4-39
Advanced Operations

NOTE:
If you are not authorized to edit the test parameters, a dialog box will
pop up to ask you to enter the authorized user name and password.

Figure 4-13 Assay Parameters screen

At the Assay Parameters screen, shown in Figure 4-13, you can see the following
tabs:

Parameters
Reference Range
Calculation
Profile
Carryover
Off-system

Follow the instructions given below to set them.

4-40
Advanced Operations

4.2.1.1 Parameters tab

Figure 4-14 Parameters tab

The Parameters tab consists of three main parts, the upper part being the
parameters to be set for the selected test, the middle part being all the tests whose
parameters are set already, the lower part being the operating buttons.

In the middle part, yellow-framed tests are the double-reagent tests whose reagent
positions are set already; the gray-framed tests are the double-reagent tests whose
reagent positions are not set yet; the non-framed tests are the single-reagent tests.

Follow the instructions given below to set the parameters for the selected test.

Parameter What it does How to set


No. It is the system-generated It is system-generated and you cant
number assigned to the set it.
current test.
Assay Defines the identification of Enter maximum 20 letters, digits for
the test to be set. the identification.
Type Defines the analysis method Select one of the three listed
to be used on the current methods: Endpoint, Fixed time and
test. Kinetic, from the pull-down list.
Reaction Defines how the absorbance Select Ascending or Descending
changes during the reaction from the pull-down list.
process, ascends (the
absorbance increasing with
the reaction) or descends
(the absorbance decreasing
with the reaction).
Units Defines the unit in which the Select the desired unit from the
test result is to be pull-down list.
expressed.

4-41
Advanced Operations

Parameter What it does How to set


Precision Defines the decimal places Enter the desired integer (0-5) into
of the test result. the edit box to the right of Precision;
or click on the right of the edit
box until the desired integer is
reached.
Prim. Defines the primary Select one of the seven wavelengths
Wave. wavelength to be used on (340, 405, 450, 510, 546, 578, 630,
the test. 670 and 700) from the pull-down list.
Sec. Defines the secondary Select one of the seven wavelengths
Wave. wavelength to be used on (340, 405, 450, 510, 546, 578, 630,
the test. 670 and 700) from the pull-down list.
Sample Defines the sample volume Enter the desired number (must be
Vol. (3-45l) to be dispensed for the multiple of 0.5) into the edit box
the reaction. to the right of Sample Vol.; or click
on the right of the edit box until
the desired number is reached.
R1 Vol. Defines the volume Enter the desired integer (180-450)
(180-450 l) of the first into the edit box to the right of R1
reagent to be dispensed for Vol..
the reaction.
Note the entered Sample Vol. + R1
Vol. must be 183l -500l.
R2 Vol. Defines the volume (30-450 Enter the desired integer (30-450)
l) of the second reagent to into the edit box to the right of R2
be dispensed for the Vol..
reaction.
Note the entered Sample Vol. + R1
Vol. + R2 Vol. must be 183l -500l.
Lin. Limit Applies to the Kinetic type Enter the desired value (1-100) into
only. It ranges from 1 to 10 the edit box.
and the default value is 20.
The analyzer will
automatically calculate the
linearity within the
measurement time and
compare the result to the
line limit and flag the result
exceeding the defined limit.
The formula to calculate the
linearity is as follows:
(1)When more than 9 points
are measured
Linearity = (absorbance
change of the first 6 points
absorbance change of the
last 6 points)/(absorbance
change of all the points)
(2)When the 4 number of
measured points 8
Linearity = (absorbance
change of the first 3 points

4-42
Advanced Operations

Parameter What it does How to set


absorbance change of the
last 3 points)/(absorbance
change of all the points)
Inc. Time Applies only to Enter the desired integer (1-60) into
double-reagent tests. the edit box to the right of Inc. Time.
Defines the interval (counted
in periods. One period
equals 12 seconds) between
the point the sample is
dispensed and the point the
second reagent is
dispensed.
R. Time Defines the reaction time Find the corresponding integers
(counted in periods. One listed on the instructions of the
period equals 12 seconds). reagent and enter them into the edit
For the Endpoint method, boxes to the right of R. Time.
the reaction time refers to
the interval between the start Enter the start time into the first box.
of the reaction and the end For the single-reagent test to be
of the reaction; for the analyzed by the Endpoint method,
Kinetic or Fixed time the start time is 0; For the
method, the reaction time double-reagent test to be analyzed
refers to the interval by the Endpoint method, the start
between the point when the time can be negative (-incubation
reaction becomes stabilized time-100); For any test to be
and the point when the analyzed by the Kinetic or Fixed time
reaction is no longer method, the start time must be
monitored. positive.

The entered integers must Enter the end time, which is no less
be no less than -40 and no than 0 and no greater 100, into the
greater than 100. second box. The end time must be
greater than the start time. For any
The analyzer defines the test to be analyzed by the Kinetic
start time of the reaction as method, the end time must be
0. For the single-reagent test greater than the start time 7 at least.
the start time refers to the
point when the sample is
dispensed; For the
double-reagent test, the start
time refers to the point when
the second reagent is
dispensed.
Antigen Defines whether the Click the check box to the right of
analyzer should check for Antigen to enable the check; Click it
surplus antigen. again to disable the check.
When the antibody reacts
with the antigen, the amount
of the product is closely
related to their proportion.
Only the appropriate
proportion can introduce the
maximum amount of the
product and when this
happens, the transmitted
light is minimized and the

4-43
Advanced Operations

Parameter What it does How to set


largest absorbance is
obtained. Any other
proportion will lead to a less
absorbance and when this
happens, samples of
significantly different antigen
concentrations may have the
same test results.
The surplus antigen
threshold (Antigen) refers to
the maximum or minimum
PC when there is no surplus
antigen.

It only applies to the


Endpoint tests to be
analyzed using the
antigen-antibody reaction.
When 1/3 of the reaction
time has elapsed, at least
80% of the reaction should
be finished. Otherwise, there
is surplus antigen.

Calculation:

(1)For the calibrator of the


maximum concentration:

threshold = 2*(absorbance
of the last point - absorbance
taken at 1/3 reaction time) /
(absorbance of the last point
+ absorbance taken at 1/3
reaction time)
(2)The above formula
applies to samples(including
controls) as well
(3)For a descending
reaction, if:

the calculated threshold of


the sample (including
controls) is less than that of
the maximum-concentration
calibrator, and
the absorbance of the last
point of the sample
(including controls) is
greater than that of the
maximum-concentration
calibrator
then there is surplus
antigen.
(4)For a ascending reaction,

4-44
Advanced Operations

Parameter What it does How to set


if :
the calculated threshold of
the sample (including
controls) is greater than that
of the
maximum-concentration
calibrator, and
the absorbance of the last
point of the sample
(including controls) is less
than that of the
maximum-concentration
calibrator
then there is surplus
antigen.
Substrate Defines whether the Enter the desired number (0-2.5).
analyzer should check for When the entered number is 0, the
surplus substrate and analyzer does not check for surplus
applies only to the Kinetic substrate.
and Fixed time method.
Certain highly concentrated
(active) sample may
consume the substrate
shortly. When it happens,
the reaction is no longer a
0th-order (kinetic) or
first-order (fixed-time) one,
which may lead to wrong
results. This parameter is
defined to prevent that from
happening. It refers to the
absorbance corresponding
to the critical point of the
linear and non-linear regions
(for the Kinetic) or of the
first-order and multiple-order
regions. It is the minimum
(descending curve) or
maximum (ascending curve)
absorbance within the given
reaction time and there is
still substrate left.
Response Defines allowed maximum Find the corresponding values on
(for ascending reaction) or the instructions of the reagent and
minimum (for descending enter them respectively to the edit
reaction) absorbance (for boxes to the right of L Limit and U
the Endpoint method) or Limit in the Response field.
absorbance change rate (for
other methods). Enter 0s into both the edit boxes if
there is no restriction on the
response.

4-45
Advanced Operations

Parameter What it does How to set


S. Volume Defines the volume of the Enter the desired integer (3-45).
sample to be diluted.
S. Volume and Ratio must meet the
following requirement:

120 (SampleVol. (Dilution + 1)) 500

Ratio Defines the dilution ratio. Enter the desired integer (2-150).
S. Volume and Ratio must meet the
following requirement:

120 (SampleVol. (Dilution + 1)) 500

R1 Blank Defines the allowed Find the corresponding values on


absorbance range of the R1 the instructions of the reagent and
blank (R1 refers to the enter them into the edit boxes to the
reagent of a single-reagent right of L Limit and U Limit in the
test or the first reagent of the R1 Blank field.
double-reagent test). The
analyzer will flag the blank Entering 0s into both the edit boxes
result that is beyond the means there is no need to check the
defined range. R1 blank.

Mix. R. Defines the allowed Find corresponding values on the


Blank absorbance range of the instruction of the reagent and enter
mixture (reagents + sample) them into the edit boxes to the right
of the double-reagent test. of L Limit and U Limit in the Mix. R.
The analyzer will flag the Blank field.
blank result that is beyond
the defined range. Entering 0s into both the edit boxes
means there is no need to check the
reagent blank of the mixture.
Linearity Defines the range in which Find the corresponding values on
the reaction result is linear the instructions of the reagent and
with the response (R). The enter them into the edit boxes to the
analyzer will flag the blank right of L Limit and U Limit in the
result that is beyond the Linearity field.
defined range.
Entering 0s into both the edit boxes
means there is no need to check the
linearity.
Full Name Defines the full name of the Enter maximum 60 letters, digits
current test. and/or symbols into the edit box to
the right of Full Name.
Print No. Defines a printing number Enter the desired integer into the
for the current test. edit box to the right of Print No.

NOTE:
Please define the parameters as instructed by this operation manual
and the instructions of the reagent suppliers. Improper settings may
lead to unreliable test results.

4-46
Advanced Operations

To view the parameters of a specific test

1 Go to the middle part of the tab and click the desired test and then the
upper part will show the parameters of the specific test.

To add a new test

1 Click the Add button.


2 Enter the parameters as instructed by the instruction of the reagents.
3 Click the Save button to save the setting; click the Cancel button, for any
reason, to ignore the setting.

NOTE:
If there are any obligatory parameters missing, once you
click the Save button, the analyzer will automatically
red-frame the missing parameters (or pops up a dialog box
to warning you about it) and disables the Save button. You
need to set the missing parameters and try to save the
settings again.

To delete a test

1 Go to the middle part of the tab and click the desired test.

2 Click the Delete button and a dialog box will pop up to ask you to confirm the
deletion.

3 Click the OK button to delete the test; click the Cancel button, for any reason,
to ignore the deletion.

NOTE:
Once you have confirmed to delete the selected test, the
analyzer will delete all the parameters, including the reagent
parameters, calibration parameters, and QC parameters,
related to the test. Ensure the right test is selected before
confirming the deletion.

4-47
Advanced Operations

To edit a test

1 Go to the middle part of the tab and click the desired test.

2 Click the Edit button.


3 Go back to the upper part of the tab to edit the test as you need.
4 Click the Save button to save the change; click the Cancel button, for any
reason, to ignore the change.
Once you have clicked the Save button, if the changed parameters change
analysis conditions, a dialog box will pop up to remind you to re-calibrate.
Click the OK button. If the changed parameters do not change analysis
conditions, this dialog box will not pop up.

A dialog box will pop up to ask you whether to delete other parameters related
to this test.

If you click the OK button, the analyzer will save the changed parameters and
clear all other unchanged ones, including the corresponding reagent
parameters, calibration parameters, and QC parameters; if you click the
Cancel button, the analyzer will save the changed parameters and keep the
other related parameters unchanged.

NOTE:
Once you have clicked the OK button, the analyzer will
delete all the parameters, including the reagent parameters,
calibration parameters, and QC parameters, related to the
test. Ensure the right test is selected before confirming the
deletion.

To exit the Assay Parameters screen

1 Click the Close button to exit the Assay Parameters screen.

4-48
Advanced Operations

4.2.1.2 Reference Range tab

Figure 4-15 Reference Range tab

Click the Reference Range tab to edit or view the expected means of the test, as
Figure 4-15 shows.

Parameter What it does How to set


Assay Selects the test to be Directly select the desired test from
set. the pull down list or click (to
check the previous test) or
button (to check the next test) until the
desired test is found.
Sample Type Defines to which Select one of the following four types
type the sample from the pull down list: Serum,
belongs. Plasma, Urine and Others.
Sex Defines the gender Select one of the following four types
of the patient. from the pull down list: Male, Female,
Others and blank.
Patient Defines expected Tick the patient group sequentially
Demographics means for patients of from No.1 and enter the desired year
(Figure 4-16) different ages. and month in the Years and Months
Maximum 6 patient columns to define the upper and lower
groups can be limits of the patient ages. Enter the
defined. desired upper and lower limits for the
age groups. You must not leave the
any box blank.
The first Years box is for the lower
limit of the age and the second Years
box is for the upper limit. Both the
entered values must be 0-150 and the
upper limit must be greater than or
equal to the lower limit.
For adult or pediatric patients older

4-49
Advanced Operations

Parameter What it does How to set


than one year, you can only enter 0s
to both the Months boxes. In case of
patients younger than one year, enter
the lower limit into the first Months
and the upper limit into the Months
limit. Both the entered values must be
0-12 and the upper limit must be
greater than or equal to the lower limit.
Default (Figure For a given test and Enter the lower limit into the first box
4-17) sample type and and the upper limit into the second.
patient gender, You can leave either of the boxes, but
defines the expected not both the boxes, empty. If you
range of the test entered values into both the boxes,
results for patients of ensure that 0 lower limit <upper
all ages. limit.
The expected range
defined for a
particular patient
group supersedes
this default range.

Figure 4-16 Patient demographics

Figure 4-17 Setting default expected range

To view the expected means

1 Click to the right of Assay and select the desired test from the pull
down list.
2 Click to the right of Sample Type and select the desired type from the
pull down list.

4-50
Advanced Operations

3 Click to the right of Sex and select the desired gender from the pull
down list.
The pre-set expected means will then be displayed.

To set expected means

1 Click to the right of Assay and select the desired test from the pull
down list.
2 Click to the right of Sample Type and select the desired type from the
pull down list.
3 Click to the right of Sex and select the desired gender from the pull
down list.
4 Click the Set button.
5 Tick the check box of the first patient group and set the Years, Months,
Lower Limit and Upper Limit field as instructed above.
6 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.

NOTE:
In case of inappropriate settings, a dialog box will pop up to
alert you once you have clicked the Save button.

To exit the Assay Parameters screen

1 Click the Close button to exit the Assay Parameters screen.

4.2.1.3 Calculation tab

Calculate certain tests to derive certain new tests of clinical purpose, such as A/G,
TBil-DBil and so forth.

The Calculation tab is where you set the parameters related to such calculation.

4-51
Advanced Operations

Figure 4-18 Calculation tab

As Figure 4-18 shows, the Calculation tab consists of four major parts: the
parameters on the upper left, all the available tests on the upper right, calculation
formula on the middle and operating buttons on the bottom.

The parameters are interpreted as follows.

Parameter What it does How to set


No. Assigns a Use the system-generated
system-generated number number.
to the current calculated
test.
Assay Defines the name of the Enter the desired name
test. (maximum 20 letters) into
the edit box.
Units Selects one of the units Select the desired unit from
for the test. the pull down list.
Precision Defines how many Enter the desired integer or
decimal places (0 - 6) the click the button until the
test result should have. desired number is reached.
Ref. Range Defines the expected Enter the lower limit into the
range of the test result. first edit box and the upper
into the second. If you enter
0s into both the boxes, it
means no specific range
defined for the test result.
Full Name Defines full name of the Enter maximum 60 letters
test. into the edit box.
Print No. Defines the user-assigned Enter the desired integer
printing order. into the edit box.

4-52
Advanced Operations

To check the parameters of a calculated test

1 Select the desire test from the Assays field, or click or


until the desired test is reached.

NOTE:
Click to go to the previous test.

Click to go to the last test.

The related parameters will be displayed on the screen.

Add a new calculated test

1 Click the Add button.


2 Enter the related parameters as instructed above.

3 Select the tests to be involved in the calculation from the Assays field.

4-53
Advanced Operations

4 Click the Add to List button to add the selected tests to the calculation list.

NOTE:
In case you add inappropriate tests to the calculation list by
mistake, click the Clear button to clear the list and then
re-select appropriate tests to add to the list.

5 Edit the calculation formula by clicking a selected test from the calculation
list and click the Add to Formula button to add the test to the Formula
field. Clicking the digit buttons (0-9) and operator buttons (+, -, and so
forth) to associate the test with appropriated digits and operators as
needed by the desired calculation formula. Repeat this step on every
selected test until the desired calculation formula is completed.

NOTE:
To clear a mistake, click the Clear Formula button to clear
the formula and re-edit it the right way.

Example: to edit ALT-AST+TP, first select ALT from the calculation list and
then click the Add to Formula button to add the test to the Formula field.
Click the button to associate the ALT with the minus sign. Select AST
from the calculation list and click the Add to Formula button to add the
test to the Formula field, following the minus sign. Click the + button to
associate the AST test. Select TP from the calculation list and click Add to
Formula. Thus, the desired formula is completed.

6 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.

To delete a calculated test

1 Select the test to delete from the pull down list on the middle left of the
screen.

4-54
Advanced Operations

2 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.

3 Click the OK button to confirm the deletion; click the Cancel button, for
any reason, to cancel the deletion.

To edit a calculated test

1 Select the test to edit from the pull down list on the middle left of the
screen.
2 Edit the formula as needed.

NOTE:
In case you add inappropriate tests to the calculation list by
mistake, click the Clear button to clear the list and then
re-select appropriate tests to add to the list.
To clear a mistake, click the Clear Formula button to clear
the formula and re-edit it the right way.

3 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.

To exit the Assay Parameters screen

1 Click the Close button to exit the Assay Parameters screen.

4.2.1.4 Profile tab

Tests grouped together for certain clinical purpose (for instance liver function)
constitute a profile.

4-55
Advanced Operations

Figure 4-19 Profile tab

As the figure shows, the Profile tab consists of three major parts: existing profiles
(groups) on the left, available tests (tests have no assigned reagent positions are
followed by a forbidden mark) on the right and operating buttons on the bottom.

To check the tests included in a profile (group)

1 Select the desired group from the pull down list of the Profile Name field
or from the Profile List field. The included tests will be followed a green
tick.

To add a new profile

1 Click the Add button.


2 Enter an appropriate profile name into the edit box of the Profile Name
field.
3 Select the tests to be included. The selected tests will be followed by a
green tick. Clicking the ticked test again may de-select the test.

4 Click the Save button to save the settings; click the Cancel button, for any
reason, to ignore the settings.

4-56
Advanced Operations

To delete a profile

1 Select the desired group from the pull down list of the Profile Name field
or from the Profile List field.

2 Click the Delete button to delete the selected profile.

3 Click the OK button to confirm the deletion; click the Cancel button, for
any reason, to cancel the deletion.

To edit a profile

1 Select the desired group from the pull down list of the Profile Name field
or from the Profile List field.
2 Click the Edit button.
3 Select or de-select the tests as you need. The selected tests will be
followed by a green tick. Clicking the ticked test again may de-select the
test.
4 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.

To exit the Assay Parameters screen

1 Click the Close to exit the Assay Parameters screen.

4.2.1.5 Carryover tab

Various reagents are sampled by the reagent probe and carryover between the
adjacent reagents, though minimized by the washing process, is still a factor to be
taken into account. This carryover may impose serious effect on certain reagents
and consequently on the related test results. This Carryover tab is designed so that
you can minimize this effect by keeping the tests whose reagents may affect each
other as far away from each other as possible, or, if this is not an option, ordering
extra probe washing before analyzing those tests.

4-57
Advanced Operations

Figure 4-20 Carryover tab

As the figure shows, the Carryover parameters screen consists of three major parts:
current reagents listed in the current field on the upper, the reagents that may be
affected (contaminated) by the current reagents listed in the contaminated field on
the lower, and operating buttons on the bottom.

Note that the yellow reagents are the first reagents of the double-reagent tests and
the blue reagents are the second reagents of the double-reagent tests.

To check the carryover parameters

1 Click the desired reagent from the current field and the selected reagent
will be followed by a green tick. Check the contaminated field and the
tests followed by a red tick are those that may be affected by the current
reagent.

To set the carryover parameters for a reagent

1 Click the Set button.


2 Select the desired reagent from the current field. The selected reagent will
be followed by a green tick. Clicking the ticked reagent again may
de-select it.
3 Check the contaminated field and click the reagents that may be affected
by the current test. The selected reagent will be followed by a red tick.
Clicking the ticked reagent again may de-select it.
4 Click the Save button to save the settings; click the Cancel button, for any
reason, to ignore the settings.

4-58
Advanced Operations

To clear the carryover settings of a reagent

1 Click the desired reagent from the current field and the selected reagent
will be followed by a green tick. Check the contaminated field and the
tests followed by a red tick are those that may be affected by the current
reagent.
2 Click the Clear button and you will see all the red ticks have disappeared.

To exit the Assay Parameters screen

1 Click the Close button to exit the Assay Parameters screen.

4.2.1.6 Off-system tab

All the tests that are not run by the BS-300 are referred to as the off-system tests,
given you have to manually enter the test results into the analyzer if you want to use
it to print out the results.

You can set basic information of off-system tests at the Off-system tab. Refer to
4.3.4.1 Results tab for details regarding entering the test results manually.

Figure 4-21 Off-system tab

As the figure shows, the Off-system tab consists of mainly three parts: the
parameters on the upper left, the tests on the upper right and the operating buttons
on the bottom.

The parameters are interpreted as follows.

Parameter What it does How to set


No. Assigns a system-generated Use the system-generated
number to the current number.
calculated test.

4-59
Advanced Operations

Parameter What it does How to set


Assay Defines the name of the test. Enter the desired name
(maximum 20 letters) into the
edit box.
Precision Defines how many decimal Enter the desired integer or click
places (0 - 6) the test result the button until the desired
should have. number is reached.
Units Selects one of the units for Select the desired unit from the
the test. pull down list.
Lower Limit Defines the lower limit of the Enter the desired into the edit
expected range of the test box.
result.
If you enter 0s for both the lower
and upper limits, it means no
specific range defined for the
test result.
Upper Limit Defines the upper limit of the Enter the desired into the edit
expected range of the test box.
result.
If you enter 0s for both the lower
and upper limits, it means no
specific range defined for the
test result.
Print No. Defines the user-assigned Enter the desired integer into
printing order. the edit box.
Full Name Defines full name of the test. Enter maximum 60 letters into
the edit box.
Property Defines property of the test. Select Quantitative or
Qualitative from the pull down
list.

To check the parameters of an off-system test

1 Select the desire test from the test list and its parameters will be displayed
on the screen.

To add a new off-system test

1 Click the Add button.


2 Enter the parameters.
3 Click the Save button to save the settings; click the Cancel button, for any
reason, to ignore the settings.

To delete an off-system test

1 Select the desire test from the test list.

4-60
Advanced Operations

2 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.

3 Click the OK button to confirm the deletion; click the Cancel button, for
any reason, to cancel the deletion.

To edit an off-system test

1 Select the desired test from test list.


2 Click the Edit button.
3 Edit its parameters as needed.
4 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.

To exit the Assay Parameters screen

1 Click the Close button to exit the Assay Parameters screen.

4.2.2 Reagent
To enter the Reagent Parameters screen, click Parameters Reagent, or press
F12.

NOTE:
If you are not authorized to change the reagent parameters, a dialog
box will pop up to ask you to enter the authorized user name and
password.

4-61
Advanced Operations

Figure 4-22 Reagent Parameters screen

At the Reagent screen, Figure 4-22, you can see two tabs:

Reagent Disk
Inventory

The Reagents Disk tab is where you set the parameters regarding the reagents.
The Inventory tab is where you check the remaining reagents and print the checking
results.

4.2.2.1 Reagent Disk tab

Click the Reagent Disk tab to edit or view the reagent disk status, as Figure 4-23
shows.

4-62
Advanced Operations

Figure 4-23 Reagent Disk tab

There are two virtual reagent disks, disk 1 and disk 2.

As the reagent disk diagram shows, the reagent disk has 50 reagent positions, 25 on
the inner circle and marked even number and another 25 on the outer circle and
marked odd numbers.

All the reagent positions are color-coded: white for the empty positions; green for the
positions occupied by reagents for the single-reagent tests; red for the positions
occupied by reagents for the double-reagent tests; blue for the positions (49 and 50)
occupied by distilled water. Note that you cannot use the two blue positions for other
purposes.

The list to the right of the reagent disk diagram displays all the tests that have been
set.

The reagent parameters are interpreted as follows.

Parameter What it does


Rgt. Disk Displays the current virtual reagent disk.
R1 Defines the position for reagent 1.
R2 Defines the position for reagent 2. Note that this parameter is
not available for single-reagent tests.
R1 Inventory Uses a color bar to display how much reagent left in the R1
bottle. The color can represent maximum 100ml of the reagent.
When the remaining reagent is still more than half of the bottle,
the color bar appears green; When the remaining reagent is
between 1/4 and 1/2 of the bottle, the color bar appears yellow;
When the remaining reagent is less than 1/4 of the bottle, and
the color bar appears red.
The box to the right of the color bar displays how many tests
the remaining reagent is enough for.
R2 Inventory Uses a color bar to display how much reagent left in the R2
bottle. The color can represent maximum 100ml of the reagent.

4-63
Advanced Operations

Parameter What it does


When the remaining reagent is still more than half of the bottle,
the color bar appears green; When the remaining reagent is
between 1/4 and 1/2 of the bottle, the color bar appears yellow;
When the remaining reagent is less than 1/4 of the bottle, and
the color bar appears red.
The box to the right of the color bar displays how many tests
the remaining reagent is enough for.
R1 Bottle Defines the bottle used by reagent 1.
R2 Bottle Defines the bottle used by reagent 2. Note that this parameter
is not available for single-reagent tests.

To view the reagent parameters of a specific test

1 Click to the right of Rgt. Disk and select the desired virtual reagent
disk from the pull down list.
2 Click the desired test in the test list.

To view the reagent parameters of a specific reagent position

1 Click to the right of Rgt. Disk and select the desired virtual reagent
disk from the pull down list.
2 Click the desired position in the reagent disk diagram.

To set reagent parameters for a new test

1 Click to the right of Rgt. Disk and select the desired virtual reagent
disk from the pull down list. The positions of the reagents will be set on
the selected virtual reagent disk.
2 Click the desired test from the test list.
3 Click the Set button.
4 Click to the right of R1 Bottle or R2 Bottle and select the desired
bottle from the pull down list. Note that for the single-reagent test, you
cannot select R2 Bottle.
5 Click to the right of R1 or R2 and select the desired position from the
pull down list. Note that for the single-reagent test, you cannot select R2
position.
6 Click the Save button to save the settings; click the Cancel button, for any
reason, to ignore the settings.

NOTE:
When you have set (not changed) the reagent position for a
test, the analyzer will automatically take the reagent bottle
volume as the volume of the remaining reagent.
If any field is inappropriately set, the analyzer will warn you
about the wrong settings once you have clicked the Save
button.

4-64
Advanced Operations

To clear the reagent positions set for a test

1 Click to the right of Rgt. Disk and select the desired virtual reagent
disk from the pull down list.
2 Select the desired test from the test list.
3 Click the Clear button to clear the reagent positions set for the selected
test.

NOTE:
Once you have clicked the Clear button, the analyzer will
immediately clear all the reagent positions set for the
selected test without asking you to confirm the clearing.
Ensure the right test is selected before clicking the Clear
button.

To change reagent parameters set for a test

1 Click to the right of Rgt. Disk and select the desired virtual reagent
disk from the pull down list.
2 Select a test.
3 If the new positions for the selected test are on the current virtual reagent
disk, go directly to the next step;
Otherwise, click the Clear button and then select the other virtual reagent
disk to the right of Rgt. Disk.
4 Click the Set button.
5 Select the bottle types to the right of R1 Bottle and R2 Bottle for the
reagents. For the single reagent test, R2 Bottle is unavailable.
6 Select the positions to the right of R1 and R2 for the reagents. For the
single reagent test, R2 is unavailable.
7 If you want to save the settings, click the Save button; otherwise, click the
Cancel button.

NOTE:
If some parameters are not set correctly, after you click the
Save button, the system will remind you.

To set the default reagent disk

1 Click the Def. Rgt. Disk button to pop up the following dialog box.

2 Select a virtual reagent disk in the dialog box.

4-65
Advanced Operations

3 If you want to set the selected disk as the default reagent disk, click the
OK button; otherwise, click the Cancel button.
After you set the default reagent disk, the Reagent Disk tab and the
Inventory tab will display its reagent information.

To exit the Reagent Parameters screen

1 Click the Close button to exit the Reagent Parameters screen.

4.2.2.2 Inventory tab

Click the Inventory tab to check the remaining reagents and print the checking
results, as Figure 4-24 shows.

Figure 4-24 Inventory tab

To check remaining reagents

NOTE:
Ensure you have opened the reagent bottles before checking the
remaining reagents.

1 Click the Reagent Disk tab, and then click the to the right of Rgt.
Disk and select the desired virtual reagent disk from the pull down list.
Refer to 4.2.2.1 Reagent Disk tab for details.

4-66
Advanced Operations

2 To select one reagent, click the desired reagent;


to select multiple discontinued reagents, press and hold CTRL and click
the desired reagents one by one;
to select multiple continue reagents, click the first reagent of the desired
ones then press and hold SHIFT and click the last reagent of the desired
ones.
3 Click the Check button to view how much reagent is left in the selected
bottle and how many tests the remaining reagent is enough for.

To clear reagent information

NOTE:
The function is available only when an external reagent bar code
scanner is used.

1 Select the reagent(s) whose information you want to clear from the
reagent list.
2 Click the Delete button.

To print out the remaining reagents

1 Click the Print button to print the currently displayed reagent leftover
information.

4.2.3 Calibration
The Calibration screen is where you set the parameters regarding calibrators and
calibrations and view the calibration results.

To enter the Calibration screen, click Parameters Calibration, or press CTRL +


K.

NOTE:
If you are not authorized to edit the calibration parameters, a dialog
box will pop up to ask you to enter the authorized user name and
password.

4-67
Advanced Operations

Figure 4-25 Calibration screen

At the Calibration screen, Figure 4-25, you can see the following tabs:

Calibrator
Rule
Result

4.2.3.1 Calibrator tab

The Calibrator tab is where you set parameters of the calibrators.

Figure 4-26 Calibrator tab

Click the Calibrator tab to display its contents, as Figure 4-26 shows. The tab
consists of 4 zones:

4-68
Advanced Operations

Calibrators zone displays such calibrator parameters as lot number, expiration


date, etc.
Positions zone (Figure 4-27) displays the positions of the calibrators.
Assays zone displays tests that related to the selected calibrator in Calibrators
zone.
Operating buttons such buttons as Add, Edit, etc.

Figure 4-27 Calibrator positions

Five virtual sample disks (Sample Disk1 - Sample Disk5) are displayed in the
Positions zone and every virtual disk has 6 calibrator positions (S1 - S6). The 0
position (default) means this virtual disk has no position set for the calibrator yet. You
cannot request a calibration run for a calibrator whose corresponding calibrator
positions on all the five virtual disks are 0s.

To view positions and tests of a specific calibrator

1 Click the desired calibrator in the Calibrators zone and check its
positions and tests in the Positions and Assays zones respectively.

4-69
Advanced Operations

To add a new calibrator

1 Click the Add button and the Calibrator window will pop up.

2 The analyzer will automatically assign a number to the new calibrator and
you can only view in the No. field without altering it;
Enter calibrator name into the edit box to the right of Name;
Enter the lot number into the edit box to the right of Lot No.;

Click to the right of Conc. Level and select Low, Normal or High
from the pull down list as the concentration level of the calibrator;

Click to the right of Exp. Date and select the appropriate as the
expiration date of the calibrator.

NOTE:
Ensure the right expiration date is set so that the analyzer
can correctly judge whether the calibrator has expired.

3 Click the OK button to save the settings; Click the Cancel button, for any
reason, to ignore the settings.

NOTE:
If any field is inappropriately set, the analyzer will warn you
about the wrong settings once you have click the OK button.

To change the settings of a specific calibrator

1 Select the desired calibrator from the Calibrators zone.

4-70
Advanced Operations

2 Click the Edit button and the Calibrator window will pop up.

3 The analyzer will automatically assign a number to the new calibrator and
you can only view in the No. field without altering it;
Enter calibrator name into the edit box to the right of Name;
Enter the lot number into the edit box to the right of Lot No.;

Click to the right of Conc. Level and select Low, Normal or High
from the pull down list as the concentration level of the calibrator;

Click to the right of Exp. Date and select the appropriate as the
expiration date of the calibrator.
4 Click the OK button to save the changes; click the Cancel button, for any
reason, to ignore the changes.

NOTE:
If any field is inappropriately set, the analyzer will warn you
about the wrong settings once you have clicked the Save
button.

To delete a specific calibrator

1 Select the desired calibrator from the Calibrators zone.

4-71
Advanced Operations

2 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.

3 Click the OK button to confirm the deletion; Click the Cancel button, for
any reason, to abort the deletion.

NOTE:
Once you have confirmed the deletion, the analyzer will
delete all the parameters, including calibrator positions, tests
included, etc., related to the selected calibrator. Ensure the
right calibrator is selected before confirming the deletion.
You cannot delete a calibrator involved in calibrations that
are in process or have been requested.

To set or change positions of a specific calibrator

1 Select the desired calibrator from the Calibrators zone.

2 Click the Set Position button.

4-72
Advanced Operations

3 To select a position fromS1 - S6 or 0, click the desired position button; To


select from other positions, click the Other button and select the desired
position from the corresponding pull down list of the Other Position
combo box.
The calibrator positions that appear gray are occupied by other
calibrators.

4 Click the Save Position button to save the changes.

To set or delete calibrator concentration for a specific test

1 Select the desired calibrator from the Calibrators zone.

4-73
Advanced Operations

2 Click the Concentration button and a Concentrations window will pop


up.

As the figure above shows, the window mainly consists of 4 zones:


(1) Calibrator zone displays the calibrator and its lot number. You can
click button to the right of Calibrator and select the desired calibrator
from the pull down list or click or until the desired
calibrator is found.
(2) Assays zone lists all the available tests. The blue-framed tests are
those whose standard concentrations have already been set.
(3) Concentration zone defines the standard concentration for the
selected test. The standard concentration cannot be negative and if it
appears -1, it indicates you have not set the standard concentration for
the selected test yet.
(4) Operating buttons zone displays such buttons as Set, Delete, etc.
3 To set or change the standard concentration for a specific test, click the
Set button.
To delete the standard concentration of a specific test, click the Delete
button.

4-74
Advanced Operations

4 Click the desired test in the Assays zone. The selected test will be
marked a green tick and click it again to de-select it. You can choose
multiple tests and set or delete their standard concentrations altogether.

5 To delete the standard concentration of the selected test or tests, go


directly to the next step;
To set standard concentration of the selected test or tests, enter the
desired value into the edit box below Concentration.
6 Click the OK button to save the changes; click the Cancel button, for any
reason, to ignore the changes.
7 If you do not need to set or delete standard concentration for another
calibrator, go directly to the next step;

To set or delete standard concentrations for another calibrator, click


button to the right of Calibrator and select the desired calibrator from the
pull down list or click or until the desired calibrator is
found. Then go back to steps 3.
8 Click the Close button to exit the Concentration window.

To exit the Calibration screen

1 Click the Close button to exit the Calibration screen.

4.2.3.2 Rule tab

The Rule tab is where you set calibration parameters for the tests.

4-75
Advanced Operations

Figure 4-28 Rule tab

Click the Rule tab to display its contents, as Figure 4-28 shows. The tab consists of
4 zones:

Operating buttons zones lists such buttons as Save, Cancel, Clear, etc.
Assays zone lists all the available tests. Those tests followed by forbidden
mark, such as , have no reagent positions yet and you cannot set the
calibration parameters for them. The blue-framed tests are those whose
calibration parameters have already been set.
Rule zone defines the nine parameters regarding calibration requests.
Calibrators zone lists the calibrators that have already been set. The selected
calibrator will be preceded by a red tick; the expired calibrator will be preceded by
a red block.

Parameter What it does How to set


Method Defines the calibration rule. Click to the right of
Method and select
The rules include One-point Linear
from the pull down list.
(K-factor), Linear (Two-point), Linear
(Multi-point), Logistic 4P, Logistic
5P, Exponential 5P, Polynomial 5p,
Parabola and Spline. One-point linear
(K-factor) is the default.

4-76
Advanced Operations

Parameter What it does How to set


K Factor Applies to One-point linear (K-factor) Enter the desired value
only. into the edit box.
When you set calibration parameters for
a test:
If the entered value is 0, a calibrator
whose concentration is not 0 must be
selected;
If the entered value is not 0 and no
calibrator is selected, the system will
calculate the concentration of sample
with the formula C=K*R (K is the K
factor and R is the response of sample)
without calibrating;
If the entered value is not 0 and a
calibrator whose concentration is 0 is
selected, the system will calculate the
concentration of sample with the
formula C=K*(R-Rs1) (K is the K factor;
R is the response of sample; Rs1 is the
response of the 0-concentration
calibrator) after calibrating.
Rerun Defines for how many times (1-5) the Enter the desire
Error Limit calibration test should be rerun. The integer(1-5) into the
default value is 1. edit box to the right of
Rerun Error Limit or
If the entered value is 2, the analyzer
will take average of the two runs as the click button on the
final result; right of the edit box
until the desired value
If the entered value is 3, the analyzer is reached.
will first remove the one with the largest
deviation and then take average of the
remaining two runs as the final result;
If the entered value is 4 or 5, the
analyzer will first remove the largest
and the smallest results and take
average of the remaining results as the
final result.
Difference Defines the maximum difference Enter the desired value
between the largest response and the (0-2.5) into the edit box
smallest response. Once the actual to the right of
difference exceeds this value, the Difference.
system will alert the user. The entered
value should be no less than 0 and no
greater than 2.5. The default value is 0,
meaning you do not want to do this
check.

4-77
Advanced Operations

Parameter What it does How to set


Interval Defines the interval (0-99 days) Enter the desired
between two adjacent auto calibrations. integer (0-99) into the
The default value is 0, indicating you do edit box to the right of
not need the analyzer to remind you of Interval or click on
the next calibration. the right of the edit box
until the desired value
is reached.
Blank Defines the response upper limit (0-2.5) Enter the desired value
Response for the 0-concentration calibrator. The (0-2.5) into the second
system will alert the user once the blank edit box to the right of
result exceeds this limit. The default Blank Response.
value is 0, meaning you do not want to
do this check.
Sensitivity Defines the maximum difference Enter the desired value
between the response of the into the edit box to the
largest-concentration calibrator and that right of Sensitivity.
of the smallest-concentration calibrator.
The system will alert the user once the
actual difference exceeds this limit.
The default value is 0, meaning you do
not want to do this check.
Difference Defines the difference between the Enter the desired value
Limit calibration factors (slope of the into the edit box to the
calibration curve) of current and last right of Difference
calibrations ((Current-Last)/Last). It only Limit.
applies to multi-point linear calibrations.
The default value is 0, meaning you do
not want to do this check.
Coefficient Applies only to multi-point linear or Enter the desired value
non-linear calibrations. The valid value into the edit box to the
is 0-1. The default value is 0, meaning right of Coefficient.
you do not want to do this check.
SD Applies only to non-linear calibrations. Enter the desired value
The fault value is 0, meaning you do not into the edit box to the
want to do this check. right of SD.

To view the calibration parameters of a specific test

1 Select the desired calibration rule.

4-78
Advanced Operations

2 Select the desired test to view its calibration settings.

To set or change the calibration parameters for a specific test

1 Select the desired calibration rule.

2 Click the Set button.


3 Click the test or tests. The selected tests will be followed by a green tick.

4 Set the request parameters as instructed above.

4-79
Advanced Operations

5 Select the calibrator or calibrators and the selected calibrator or


calibrators will be preceded by a red tick.
If selecting a nonlinear calibration rule in the step 1, you can select either
several calibrators or only one calibrator whose concentration is not 0.
If selecting a nonlinear calibration rule and only one calibrator, go to the
next step to set parameters of auto dilution; otherwise, go to the step 9.

NOTE:
You must determine the number of the calibrators to be used
according to the selected calibration rule. You can only
select calibrators whose standard concentrations have
already been set.

6 Click the Auto Dilution button and the Auto Dilution Setup window will
pop up.

4-80
Advanced Operations

7 Enter concentrations in the ascending order into the edit boxes in the
Auto Dilution Setup window.
You must determine the number of the edit boxes to be entered as you
need. Leave the edit boxes that are not to be used empty.
Conc. 1 must be 0. It refers to the concentration of the distilled water in
position W of the sample disk.
The entered second concentration must be greater than or equal to 0.6%
of the concentration of the calibrator selected in the step 5.
The entered penultimate concentration must be less than or equal to
37.5% of the concentration of the calibrator selected in the step 5.
The entered last concentration must be the same with the concentration of
the calibrator selected in the step 5.
Following values are recommended where N is the number of
concentrations and C is the concentration of the calibrator selected in the
step 5.
N=3 N=4 N=5 N=6 N=7 N=8
Conc. 1 0 0 0 0 0 0
Conc. 2 0.375*C 0.25*C 0.125*C 0.0625*C 0.03125*C 0.015625*C
Conc. 3 C 0.375*C 0.25*C 0.125*C 0.0625*C 0.03125*C
Conc. 4 C 0.375*C 0.25*C 0.125*C 0.0625*C
Conc. 5 C 0.375*C 0.25*C 0.125*C
Conc. 6 C 0.375*C 0.25*C
Conc. 7 C 0.375*C
Conc. 8 C
8 Click the Save button to save the settings; Click the Cancel button to
ignore the settings. Then click at the top-right corner to exit the
window.
9 Click the Save button to save the changes; Click the Cancel button, for
any reason, to ignore the changes.

NOTE:
If any field is inappropriately set, the analyzer will warn you
about the wrong settings once you have clicked the Save
button.

To clear calibration settings of a specific test

1 Select the desired calibration rule.

4-81
Advanced Operations

2 Select the test or tests whose calibration settings you want to clear.

3 Click the Clear button to clear the settings.

NOTE:
Once you have clicked the Clear button, the analyzer will
immediately clear all the calibration settings related to the
selected test without asking you to confirm the clearing.
Ensure the right test is selected before clicking the Clear
button.

To exit the Calibration screen

1 Click the Close button to exit the Calibration screen.

4.2.3.3 Result tab

The Result tab, as Figure 4-29 shows, is where you search for the calibration results
and edit them.

Figure 4-29 Result tab

4-82
Advanced Operations

To view the calibration result

1 To specify the search conditions.

Click to the right of Assay, or click or to select previous or


next test until the desired one is reached;

Click to the right of From (or To) and select the desired date to specify the
dates to be searched. The analyzer will search all the calibration runs taking
place during this specified period.

2 Click the Lock button to lock the selected search conditions.


3 Click the Search button to start the search and the matches will be displayed in
the Results zone.

NOTE:
If no calibration run matches the specified conditions, a
dialog box will pop up to remind you no match is found.
Under this circumstance, go directly to step 6.

4 If you do not intend to check the calibration curve, go directly to step 6;


Otherwise, click the Calib. Curve button to enter the Calibration Curve
window to view the calibration result.

To print out the calibration curve, click the Print button.


5 Click the Close button to close the Curve window.
6 If you do not intend to check other calibration results, the operation is over;
otherwise, click the Unlock button and return to step 1.

4-83
Advanced Operations

To re-calculate the calibration parameters using other methods

1 To specify the search conditions:

Click to the right of Assay, or click or to check previous


or next test until the desired one is reached;

Click to the right of From (or To) and select the desired date to
specify the dates to be searched. The analyzer will search all the
calibration runs taking place during this specified period.
2 Click the Lock button to lock the selected search conditions.
3 Click the Search button to start the search and the matches will be
displayed in the Results zone. You cant go on unless there are matches.
4 Select the desired calibrations (except for those whose Description
displays Calib. Edit) from the Calibrations field.

5 Click the Recalculate button.


6 Select the desired calibrator from the Results field.

7 Select the desired calibration rule from the pull down list to the left of
Method.

8 Click the Calculate button to start the calculation; click the Cancel button,
for any reason, to cancel the calculation.

To delete the calibration settings

1 To specify the search conditions:

Click to the right of Assay, or click or to check previous


or next test until the desired one is reached;

Click to the right of From (or To) and select the desired date to
specify the dates to be searched. The analyzer will search all the
calibration runs taking place during this specified period.
2 Click the Lock button to lock the selected search conditions.
3 Click the Search button to start the search and the matches will be
displayed in the Results zone. You cant go on unless there are matches.

4-84
Advanced Operations

4 Select the desired calibrations from the Calibrations field.

NOTE:
You cannot select the calibration whose Default field
displays Yes.

5 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.

6 Click the OK button to confirm the deletion; click the Cancel button, for
any reason, to cancel the deletion.

To define the default settings

1 To specify the search conditions:

Click to the right of Assay, or click or to check previous


or next test until the desired one is reached;

Click to the right of From (or To) and select the desired date to
specify the dates to be searched. The analyzer will search all the
calibration runs taking place during this specified period.
2 Click the Lock button to lock the selected search conditions.
3 Click the Search button to start the search and the matches will be
displayed in the Results area. You cant go on unless there are matches.
4 Select the desired record from the Calibrations field.

4-85
Advanced Operations

5 Click the Default button to set the current calibration settings as the
default settings of the selected test. The Default field of the calibration will
display Yes.

NOTE:
This analyzer uses the default calibration settings to
calculate the sample concentrations.
This analyzer will automatically set the latest settings
(include the settings of calibration test, edit, calculation) as
the default.

To view the calibration data

1 To specify the search conditions:

Click to the right of Assay, or click or to check previous


or next test until the desired one is reached;

Click to the right of From (or To) and select the desired date to
specify the dates to be searched. The analyzer will search all the
calibration runs taking place during this specified period.
2 Click the Lock button to lock the selected search conditions.
3 Click the Search button to start the search and the matches will be
displayed in the Results area. You cant go on unless there are matches.
4 Select the desired record from the Calibrations field.
5 Click the Data button. The Test Data dialog box is displayed. You can
view the data of selected calibration.

4-86
Advanced Operations

6 Select one result.

7 Click the Reaction Curve. A window will pop up to display the


corresponding reaction curve.

8 Click the Close button to close the Reaction Curve window.


9 Click the Close button to close the Test Data window.

To exit the Calibration screen

1 Click the Close button to exit the Calibration screen.

4.2.4 QC
The QC screen is where you set the parameters regarding controls and QC runs and
view QC results.

To enter the QC screen, click Parameters QC, or press CTRL + I.

4-87
Advanced Operations

NOTE:
If you are not authorized to change the QC parameters, a dialog box
will pop up to ask you to enter the authorized user name and
password.

Figure 4-30 QC screen

At the QC screen, Figure 4-30, you can see the following tabs:

Control defines parameters regarding the controls.


Rule defines parameters regarding the QC rule.
Setup defines parameters regarding the QC test.
Real-time Result displays information regarding real-time control.
Daily Result displays information regarding daily control.
Day-to-day Result displays information regarding day-to-day control.

4.2.4.1 Control tab

The Control tab is where you set parameters for the controls.

4-88
Advanced Operations

Figure 4-31 Control tab

Click the Control tab to display its contents, as Figure 4-31 shows. The tab mainly
consists of 4 zones:

Controls zone displays such control parameters as lot number, expiration date,
etc.
Positions zone displays the positions of the controls.
Assays zone displays available tests that related to the selected control in
Controls zone.
Operating buttons such buttons as Add, Edit, etc.

Five virtual sample disks (Sample Disk1 - Sample Disk5) are displayed in the
Positions zone and every virtual disk has 3 control positions (C1 - C3). The 0
position (default) means this virtual disk has no position set for the control yet. The
Other means you can set other positions for the controls on the disk.

You cannot request a QC run for a control whose corresponding control positions on
all of the five virtual disks are 0s.

4-89
Advanced Operations

To view positions and tests of a specific control

1 Click the desired control in the Controls zone and check its positions and
tests in the Positions and Assays zones respectively.

To add a new control

1 Click the Add button and the Control window will pop up.

2 The analyzer will automatically assign a number to the new control and
you can only view in the No. field without altering it;
Enter control name into the edit box to the right of Name;
Enter the lot number into the edit box to the right of Lot No.;

Click to the right of Conc. Level and select Low, Normal or High
from the pull down list as the concentration level of the control;

Click to the right of Exp. Date and select the appropriate date as the
expiration date of the control;

Click to the right of Tube Type and select Large Tube or Small Tube
for the control.

NOTE:
Ensure the right expiration date is set so that the analyzer
can correctly judge whether the control has expired.

4-90
Advanced Operations

3 Click the Save button to save the changes; Click the Cancel button, for
any reason, to ignore the change.

NOTE:
If any field is inappropriately set, the analyzer will warn you
about the wrong settings once you have clicked the Save
button.

To change the settings of a specific control

1 Select the desired control from the Controls zone.

2 Click the Edit button and the Control window will pop up.

3 The analyzer will automatically assign a number to the new control and
you can only view in the No. field without altering it;
Enter control name into the edit box to the right of Name;
Enter the lot number into the edit box to the right of Lot No.;

Click to the right of Conc. Level and select Low, Normal or High
from the pull down list as the concentration level of the control;

Click to the right of Exp. Date and select the appropriate date as the
expiration date of the control;

Click to the right of Tube Type and select Large Tube or Small Tube
for the control.
4 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes

NOTE:
If any field is inappropriately set, the analyzer will warn you
about the wrong settings once you have click the Save
button.

4-91
Advanced Operations

To delete a specific control

1 Select the desired control from the Controls zone.

2 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.

3 Click the OK button to confirm the deletion; Click the Cancel button, for
any reason, to abort the deletion.

NOTE:
Once you have confirmed the deletion, the analyzer will
delete all the parameters, including control positions, tests
included, etc., related to the selected control. Ensure the
right control is selected before confirming the deletion.
You cannot delete a control involved in QC runs that are in
process or have been requested.

To set or change position of a specific control

1 Select the desired control from the Controls zone.

2 Click the Set Position button.


3 To select a position from C1 - C3 or 0, click the desired position button;
To select other position, click the Other button and select the desired
position from the corresponding pull down list of the combo box.

The control positions that appear gray are occupied by other controls.
4 Click the Save Position button to save the changes.

4-92
Advanced Operations

To set standard concentration and SD for a specific test

1 Select the desired control from the Controls zone.

2 Click the Concentration button and a Concentration window will pop up.

ISE appears only when the ISE unit (optional) is connected.


As the figure above shows, the window mainly consists of 4 zones:
(1) Control zone displays the control to be set. You can select the
desired calibrator from the pull down list or click or until
the desired calibrator is found.
(2) Assays zone lists all the available tests. The blue-framed tests are
those whose standard concentrations have already been set.
(3) Concentration and SD zone defines the standard concentration and
SD for the selected test. Neither the standard concentration nor the SD
can be negative. When the Concentration or SD field displays -1, it
indicates the standard concentration or SD has not been set yet.
(4) Operating buttons zone displays such buttons as Set, Delete, etc.
3 If you want to set concentrations and SDs for ISE tests, select the check
box to the left of ISE.
To set or change concentration or SD, click the Set button.
To delete concentration or SD, click the Delete button.

4-93
Advanced Operations

4 If you enter concentrations and SDs for ISE tests, go directly to the next
step.
Click the desired test in the Assays zone. The selected test will be
marked a green tick and click it again to de-select it. You can choose
multiple tests and set or delete their standard concentrations or SD
altogether.

5 To delete concentrations and SDs, go directly to the next step; Otherwise,


enter the desired values into the edit boxes below Concentration (or
Conc.) and SD respectively.
6 Click the OK button to save the changes; Click the Cancel button, for any
reason, to ignore the change.
7 If you do not want to set or delete standard concentration or SD for
another control, go directly to the next step;

Otherwise, click button to the right of Control and select the desired
control from the pull down list or click or until the desired
control is found. Then go back to step 3.
8 Click the Close button to exit the Concentration window.

NOTE:
Please use Mindrayrecommended controls for ISE QC.

To exit the QC screen

1 Click the Close button to exit the QC screen.

4.2.4.2 Rule tab

The Rule tab is where you set QC rules for the tests.

4-94
Advanced Operations

Figure 4-32 Rule tab

Click the Rule tab to display its contents, as Figure 4-32 shows. The tab consists of
3 zones:

Independent zone
Control-related zone
Operating buttons zone

The QC rules can be divided into two categories: independent rules, such as
Cumulative Sum Check and Westgard Multi-rule, and control-related rules (not
related to controls), such as Accumulative Error and X-R. Note that you cannot set
the parameters of the control-related rules unless you have specified the controls to
be used.

Figure 4-33 Independent zone

As Figure 4-33 shows, all the available tests are listed on the left and the rules,
Cumulative Sum Check and Westgard Multi-rule, are listed on the right. The
Cumulative Sum Check rule has three sub-rules and you can choose one of the
sub-rules. The Westgard Multi-rule rule has six sub-rules and you can choose one
or more sub-rules.

4-95
Advanced Operations

Figure 4-34 Control-related zone

As Figure 4-34 shows, all the available tests are listed on the left and the QC rule
parameters are listed in the Control, Accumulated Error and X-R zones.

Note that any test whose standard concentration and SD have not been set yet will
be followed by a forbidden mark, such as .

NOTE:
If the ISE unit (optional) is connected, the ISE tests will be displayed in
the test list.

To set the independent parameters

1 Tick the check box to the left of Independent.

2 Click the Set button.


3 Select the desired test or tests. The selected test will be followed by a green
tick.

4-96
Advanced Operations

4 Click the option to the left of the desired sub-rule.

5 Click the Save button to save the settings; click the Cancel button, for any
reason, to ignore the settings.
6 Repeat from step 2 to set the parameters for other tests, or click the check
box to the left of Independent again to de-select the field.

To set the control-related parameters

1 Tick the check box to the left of Control-related.

2 Click the Set button.


3 Select the desired test or tests. The selected test will be followed by a
green tick.

4-97
Advanced Operations

4 To set the parameters: click button to the right of Control and select
the desired control from the pull down list; enter the desired E (limit of the
accumulated error) into the edit box to the right of Limit(E); enter the
desired SDR value in the edit box to the right SDR (for the X-R rule).

5 Click the Save button to save the settings; click the Cancel button, for any
reason, to ignore the settings.
6 Repeat from step 2 to set the parameters for other tests, or click the check
box to the left of Control-related again to de-select the field.

To exit the QC screen

1 Click the Close button to exit the QC screen.

4.2.4.3 Setup tab

The Setup tab is where you set QC parameters for the tests.

Figure 4-35 Setup tab

As Figure 4-35 shows, the Setup tab consists of four zones:

Assays zone lists all the available tests.

4-98
Advanced Operations

Controls zone lists all the available controls.


Auto QC zone defines parameters regarding auto QC.
Operating buttons lists such buttons as Save, Clear, etc.

The tests, whose QC parameters have already been set, are blue-framed. The tests
whose reagent positions have not been set yet are followed by a forbidden mark,
such as , and you cannot set QC parameters for the test.

All the controls that have been set are listed in the Controls zone. The selected
tests will be preceded by a red tick and the expired test will be preceded by a red
mark.

You can set the interval of the automatic QC runs by entering the desired number
into the edit box below Count or clicking on the right of the edit box until the
desired number is reached. To disable the auto QC function, enter 0 into the edit
box.

To view the QC settings of a selected test

1 Click the desired test.

To set or change the QC parameters of a selected test

1 Click the Set button.


2 Select the desired test or tests. The selected parameter will be followed
by a green tick.

4-99
Advanced Operations

3 Select the controls to be used.

4 Set the interval for the automatic QC runs.

5 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.

NOTE:
If any parameter is inappropriately set, the analyzer will alert
you after you have clicked the Save button.

To clear QC settings for a test

1 Select the desired test.

2 Click the Clear button to clear the settings.

NOTE:
Once you have clicked the Clear button, the analyzer will
immediately clear the QC settings related to the selected test
without asking you to confirm the clearing. Ensure the right
test is selected before clicking the Clear button.

To exit the QC screen

1 Click the Close button to exit the QC screen.

4-100
Advanced Operations

4.2.4.4 Real-time Result tab

The Real-time Result tab is where you view and print the real-time QC results and
delete the real-time QC data.

Figure 4-36 Real-time Result tab

NOTE:
If the ISE unit (optional) is connected, ISE tests will be displayed in the
test list.

To view the results of a real-time control

1 Select the desired test from the Assays zone. The selected test will be
followed by a green mark.

2 Tick the desired control or controls.

4-101
Advanced Operations

3 Click the Search button to display the information, including the results,
QC graph, statistic data and QC status, of the last 10 control runs of the
selected test using the selected control or controls.
To print the displayed data and graph, click the Print button.

To print out the result of a control run

1 Click the Print button to print out the result and QC graph after checking
the result of a control run.

To delete data of a real-time control

1 Select the desired test in the Assays field. The selected test will be
marked with a green block to the right of it.

2 Select the desired control in the Controls field.

3 Click the Data button. The Delete QC Data dialog box is displayed. You
can view and delete the measurement data of selected control.

4 Select one or more QC results you want to delete from the Delete QC
Data dialog box.

4-102
Advanced Operations

5 Select Delete. The Enter User and Password dialog box pops up.

6 Enter correct user name and password and click OK. The selected results
are deleted.
Please note that the username you are entering should have the authority
to set up quality control. Otherwise you cannot delete the QC results.
7 Click the Close button to exit the Delete QC Data window.

To exit the QC screen

1 Click the Close button to exit the QC screen.

4.2.4.5 Daily Result tab

The Daily Result tab is where you view and print the daily QC results and delete the
daily QC data.

Figure 4-37 Daily Result tab

4-103
Advanced Operations

NOTE:
If the ISE unit (optional) is connected, ISE tests will be displayed in the
test list.

To view the results of a daily control

1 Click to the right of Date and select the desired date from the pull
down list to search the QC runs taking place within this date.
2 Click to the right of Rule and select the desired QC rule from the pull
down list.

3 Select the desired test from the Assays zone. The selected test will be
followed by a green mark.

4 Tick the desired control or controls.

5 Click the Search button to display the information, including the results,
QC graph, statistic data and QC status, of the last 10 control runs of the
selected test using the selected control or controls.
To print the displayed data and graph, click the Print button.

To print out the result of a control run

1 Click the Print button to print out the result and QC graph after checking
the result of a control run.

4-104
Advanced Operations

To delete the data of a daily control

1 Select the desired test in the Assays field. The selected test will be
marked with a green block to the right of it.

2 Select the desired control in the Controls field.

3 Click the Data button. The Delete QC Data dialog box is displayed. You
can view and delete the measurement data of selected control.

4 Select one or more QC results you want to delete from the Delete QC
Data dialog box.
5 Select Delete. The Enter User and Password dialog box pops up.

4-105
Advanced Operations

6 Enter correct user name and password and click OK. The selected results
are deleted.
Please note that the username you are entering should have the authority
to set up quality control. Otherwise you cannot delete the QC results.
7 Click the Close button to exit the Delete QC Data window.

To exit the QC screen

1 Click the Close button to exit the QC screen.

4.2.4.6 Day-to-day Result tab

The Day- to-day Result tab is where you view and print the day-to-day QC results.

Figure 4-38 Day-to-day Result tab

NOTE:
If the ISE unit (optional) is connected, ISE tests will be displayed in the
test list.

To view the results of a day-to-day control

1 Click to the right of From (or To) and select the desired date to
specify the dates to be searched. The analyzer will search all the QC runs
taking place during this specified period.
2 Click to the right of Rule and select the desired QC rule from the pull
down list.

4-106
Advanced Operations

3 Select the desired test from the Assays zone. The selected test will be
followed by a green mark.

4 Tick the desired control or controls.

5 Click the Search button to display the information, including the results,
QC graph, statistic data and QC status, of the last 10 control runs of the
selected test using the selected control or controls.
To print the displayed data and graph, click the Print button.

To print out the result of a control run

1 Click the Print button to print out the result and QC graph after checking
the result of a control run.

To view the data of a day-to-day control

1 Select the desired test in the Assays field. The selected test will be
marked with a green block to the right of it.

2 Select the desired control in the Controls field.

3 Click the Data button. The Delete QC Data dialog box is displayed. You
can view the measurement data of selected control.

4-107
Advanced Operations

4 Click the Close button to delete fromexit the Delete QC Data dialog box.
window.

To exit the QC screen

1 Click the Close button to exit the QC screen.

4.3 Data

4.3.1 Export
Click Data Export or press CTRL + F to enter the Data exported screen, as
Figure 4-39 shows.

NOTE:
This analyzer automatically backs up its settings and test results.
Exporting data refers to saving some data to a file (.SHD file)
temporarily so that the data can be transmitted or updated. It is not the
same as backing up the database.
Mindray recommends this exporting function not be used unless it is
necessary.

4-108
Advanced Operations

Figure 4-39 Data exported screen

As Figure 4-39 shows, the Data exported screen consists of the following fields:

Assay
Calibration
QC
Sample
Others
Assays
Separator

The item in blue font listed in the Assay, Calibration, QC and Sample fields is
test-related and you can select test in the Assays field for these items.

NOTE:
If the data to be exported includes the data of last year, Mindray
recommends the Start date be not before November 1. Otherwise,
you may not get the complete data.

4-109
Advanced Operations

To select all the items

1 Click the All button to select all the items.

To de-select all the items

1 Click the Clear button to de-select all the items.

To export data

1 Select the items to be included in the exported data as instructed below.


Tick the check box to the left of the desired item to include it.
For the items in black font, they can be included into the exported data
simply by ticking the corresponding check boxes.
For the items in blue font, ticking their corresponding check boxes means
they apply to all the available tests. If you only intend to apply them to
certain tests, first specify the corresponding fields in the Applied to field and
then select the desired tests from the Assays field. See the examples below
for details.
Example1:
Follow the steps below to select: a) calculated tests (Calculation of the
Assay field) and units (Units of the Assay field) for all the tests; b) test
parameters (Parameters of the Assay field) for AST and ALP only.
(1) Tick the check boxes to the left of Parameters, Calculation, and Units
in the Assay field respectively.
(2) Tick the check box to the left of Assay in the Applied to field.
(3) Tick the check boxes to the left of AST and ALP in the Assays field
respectively.

4-110
Advanced Operations

Example2:
Follow the steps below to select: a) calibration parameters (Results in the
Calibration field) for all the tests; b) calibration results (Test Data in the
Calibration) for the tests analyzed on February 3, 2007 and February 4,
2007 only.
(1) Tick the check boxes to the left of Results and Test Data in the
Calibration field respectively.
(2) Select February 3, 2007 from the pull down list to the left of the From in
the Calibration field.
(3) Select February 4, 2007 from the pull down list to the left of To in the
Calibration field.

Example3:
Follow the steps below to select the UR results of the samples: a) sample
IDs are 1 - 10; b) analyzed on February 3, 2007 and February 4, 2007.
(1) Tick the check boxes to the left of Results in the Sample field
(2) Enter 1 and 10 into the edit boxes to the right of Sample No.
respectively.
(3) Select February 3, 2007 from the pull down list to the left of the From in

4-111
Advanced Operations

the Sample field.


(4) Select February 4, 2007 from the pull down list to the left of To in the
Sample field.
(5) Tick the check box to the left of Sample in the Applied to field.
(6) Tick the check box to the left of UR in the Assays field.

2 Click the Export button and a dialog box will pop up.

3 Select the desired directory and file and click the Save button to save the
data. You may also click the Cancel button to cancel the exporting.

To exit the Export screen

1 Click the Close button to exit the Export screen.

4.3.2 Import
Click Data Import or press CTRL + E to enter the Import Data screen, as Figure
4-40 shows. Note that you need a user name and password of the administrator
authority to enter the screen.

4-112
Advanced Operations

Figure 4-40 Import Data screen

NOTE:
The imported data will overwrite the corresponding data in the
analyzers database. Before importing data, ensure the corresponding
data is exported to another file. You can only import data from .SHD
files.
Mindray recommends this importing function not be used unless
necessary.

To import data

1 Click the File button and a dialog box will pop up.

2 Select the desired file and click the Open button to open the selected file.

4-113
Advanced Operations

3 Click the Import button to import the data into the database. A dialog box
will pop up when then importing is done. Click the OK button to close the
box.

To exit the Import Data screen

1 Click the Close button to exit the Import Data screen.

4.3.3 Demographics
Click Data Demographics or press CTRL + A to enter the Demographics screen,
as Figure 4-41 shows. You can set patient information and add manual samples at
the screen.

Figure 4-41 Demographics screen

As Figure 4-41 shows, you can click (to the first patient), (to the
previous patient), (to the next patient) or (to the last patient) to
select the desired patient.

To view the information of a patient

1 Select the date when the patient sample was requested from the pull
down list to the right of Date. The current date is the default.

4-114
Advanced Operations

2
You can click , , or to select the desired
patient, or you can enter the ID of the desired sample into the edit box to
the right of Move To and click the Move To button to go to the desired
sample.

To add a new manual sample

1 Select the date from the pull down list to the right of Date. The current
date is the default.
2 Click the Add button to add a new manual sample whose No. (greater
than 9000) is generated by the system automatically.
3 Enter the patient information of the sample.
The Date and the Sample No. cannot be changed.
The Sample ID can be entered manually or by the bar code reader.
4 If you want to save the new manual sample, click the Save button;
otherwise, click the Cancel button.

To edit the patient information

1 Select the date when the patient sample was requested from the pull
down list to the right of Date. The current date is the default.
2
You can click , , or to select the desired
patient, or you can enter the ID of the desired sample into the edit box to
the right of Move To and click the Move To button to go to the desired
sample.
3 Click the Edit button.
4 Edit the patient information as needed.
The Date and the Sample No. cannot be changed.
5 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.

To define the default setting

1 Select the date from the pull down list to the right of Date. The current
date is the default.
2 Click the buttons , , , , or after
entering the sample ID to the right of Move To, click the Move To, to
select a sample.
3 Click the Edit button.
4 Click the Set as Default button to set the Sample Type (sample type),
Sender (doctor who sends the sample), Sent by (department from which
the sample is sent), Tester (doctor who analyzes the sample) and Tested
by (the laboratory to analyze the sample) as the default settings.
5 Click the Save or Cancel button.

4-115
Advanced Operations

NOTE:
The default setting is only available at the current screen. If you click
the Exit button to close the Demographics screen, the default setting
disappears and you need to set it again when you enter the screen the
next time.

To obtain the default settings

If you add a new manual sample, after you click the Add button, the Sample Type,
Sender, Sent by, Tester and Tested by are set as the default.

If you edit the patient information of a sample, you can insert a step between the
step 3 and step 4 of To edit the patient information: click the Load Default button.

To exit the Demographics screen

1 Click the Exit button to exit the Demograhics screen.

4.3.4 Result
Click Data Result or press Ctrl + R to enter the Results & Statistics screen, as
Figure 4-42 shows.

Figure 4-42 Results & Statistics screen

At the Result & Statistics screen, you can see the following tabs: Results, Result
Statistics and Test Statistics.

4-116
Advanced Operations

4.3.4.1 Results tab

The Results tab is where you can view, edit, revise and print the sample results and
send them to LIS in batch mode.

Figure 4-43 Results tab

To view sample results

1 Click to the right of From (or To) and select the desired date to specify
the dates to be searched. The analyzer will search all the sample runs
taking place during this specified period.

NOTE:
In case of a year-to-year search, the start date must not be
earlier than November 1st of the previous year.

2 Specify other desired search conditions or go directly to the next step.

4-117
Advanced Operations

3 Click the Search button to start the search and the matches will be
displayed.

NOTE:
If no sample run matches the specified conditions, a dialog
box will pop up to remind you no match is found.

4 Click the By Sample, and then select a sample to view its results.

Click the By Test to view the results of all sample runs.

4-118
Advanced Operations

To view reaction curve

1 Search for the sample results. Refer to the previous section To view
sample results for details.
2 Select a sample run.
If you select By Sample, select a sample and then a run.

If you select By Test, select a sample run directly.

4-119
Advanced Operations

3
NOTE:
You will see the Curve button except that you select LIS in
the Option field.

Click the Reaction Curve button to check the reaction curve and data of
the selected blank runs.
There are two tabs at the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
4 If you do not want to check the results of other runs, go directly to the next
step;
Otherwise, click the Previous or Next button to check other runs (blanks,
samples, controls, or calibrations); or click the Close button and repeat
from step 2.
5 Click the Close button to close the Reaction Curve window.

To set printing options

1 Click the option button to the left of Print in the Option field.

2 Click the Print Setup button to enter the Printing Setup screen. Refer to
4.4.7 Print for details.

To print out patient reports

1 Search for the sample results. Refer to the previous section To view
sample results for details.
2 Select By Sample.

4-120
Advanced Operations

3 Click the option button to the left of Print in the Option field.

4 Select the sample or samples you want to print. You can select them by
clicking the Select button or ticking the check boxes directly.

NOTE:
Click the Select button, and a dialog box will pop up to ask
you to enter the start number of the sample.

Enter the number as required and click the OK button, and


another dialog box will pop up to ask you enter the end
number of the sample.

Enter the number as required and click the OK button. The


analyzer will select the two samples and those between
them.
If you want to de-select samples, click the Deselect button,
and the system will de-select two samples and those
between them.

5 Click the Print button to pop up the Print Preview screen.


At the Print Preview screen, you may choose Pr. Series (series printing)
for your impact printer; you may choose Pr. Current (printing current
page) or Pr. All (printing all pages) for other types of printers.
6 When you finish printing, click the Close button to exit the Print Preview
screen.

To print out a patient report in a combined mode

This function will help you print different sample results of the same patient in one
report.

1 Search for the sample results. Refer to the previous section To view
sample results for details.
2 Select By Sample.

4-121
Advanced Operations

3 Click the option button to the left of Print in the Option field.

4 Select the sample or samples (5 is the maximum) you want to print. You
can select them by clicking the Select button or ticking the check boxes
directly.
5 Click the Combined Print button to pop up the Print Preview screen.
6 If you have intended to print out all the results, go directly to the next step;
Otherwise, click the Selected Assay button in the preview window to
select the tests to be printed out. Then after clicking the OK button, click
the Hide Assays button.
7 At the Print Preview screen, you may choose Pr. Series (series printing)
for your impact printer; you may choose Pr. Current (printing current
page) or Pr. All (printing all pages) for other types of printers.
8 When you finish printing, click the Close button to exit the Print Preview
screen.

To print out test reports

1 Search for the sample results. Refer to the previous section To view
sample results for details.
2 Select By Test.
3 Click the option button to the left of Print in the Option field.

4-122
Advanced Operations

4 Select the sample run or runs you want to print. You can select them by
clicking the Select button or ticking the check boxes directly.

NOTE:
Click the Select button, and a dialog box will pop up to ask
you to enter the start number of the run.

Enter the number as required and click the OK button, and


another dialog box will pop up to ask you enter the end
number of the run.

Enter the number as required and click the OK button. The


analyzer will select the two runs and those between them.
If you want to de-select runs, click the Deselect button, and
the system will de-select two runs and those between them.

5 Click the Print button to pop up the Print Preview screen.


At the Print Preview screen, you may choose Pr. Series (series printing)
for your impact printer; you may choose Pr. Current (printing current
page) or Pr. All (printing all pages) for other types of printers.
6 When you finish printing, click the Close button to exit the Print Preview
screen.

To edit manual results

1 Search for the sample results. Refer to the previous section To view
sample results for details.

NOTE:
The To (end date) must be in the current year. Otherwise,
you cannot edit manual results.

2 Select By Sample.

4-123
Advanced Operations

3 Click the option button to the left of the middle Edit in the Option field.

4 Click the Edit button.


5 Select a sample or samples. If you select more than one sample, you will
set the same manual results for them.

6 Edit manual results in the Edit results field.

7 If you want to save the setting, click the Save button; otherwise, click the
Cancel button.

To edit sample results

NOTE:
Sample results can only be edited by authorized personnel.
If an unauthorized personnel wants to edit sample results, a dialog box
will pop up to ask you to enter authorized user name and password.

1 Search for the sample results. Refer to the previous section To view
sample results for details.

NOTE:
The To (end date) must be in the current year. Otherwise, you
cannot edit sample results.

2 Select By Sample.
3 Click the option button to the left of the right Edit in the Option field.

4-124
Advanced Operations

4 Select a sample.

5 Select a sample run.

6 Click the Edit button.


7 Enter new result to the right of Result.

8 If you want to save the setting, click the Save button; otherwise, click the
Cancel button.

To correct sample results

NOTE:
Sample results can only be revised by authorized personnel.
If an unauthorized personnel wants to revise sample results, a dialog
box will pop up to ask you to enter authorized user name and
password.

1 Search for the sample results. Refer to To view sample results for details.

NOTE:
The End (end date) must be in the current year. Otherwise,
you cannot revise sample results.

2 Select By Test.
3 Click the option button to the left of Correct in the Option field.

4 Click the Correct button.

4-125
Advanced Operations

5 Select a sample run or runs. You can select them by clicking the Select
button or ticking the check boxes directly. If you select more than one runs,
you will set the same factors for them.
6 Enter factor a and b to the right of Factor a: and Factor b:. The system will
revise the results of the selected runs according to the formula Y=aX+b (Y
stands for the new result and X stands for the old result).

7 If you want to revise the results, click the Save button; otherwise, click the
Cancel button.

To send sample results to LIS in batch mode

1 Search for the sample results. Refer to the previous section To view
sample results for details.

NOTE:
The To (end date) must be in the current year. Otherwise, you
cannot send sample results to the LIS in batch mode.

2 Select By Sample.
3 Click the option button to the left of LIS in the Option field.

4 Select a sample or samples. You can select them by clicking the Select
button or ticking the check boxes directly.
5 Click the Send button to send the results of the selected samples to the
LIS.
When the results are sent, a dialog box will pop up.

Click the Close button to exit the dialog box.

4-126
Advanced Operations

NOTE:
There are two modes to send information to the LIS, the real-time
mode and the batch mode.
You can send information to the LIS in batch mode at any time only
except that when you select the Real Time Send at the System
SetupHost Communication screen and the system is testing.

To exit the Results & Statistics screen

1 Click the Close button to exit the Results & Statistics screen.

4.3.4.2 Result Statistics tab

The Result Statistics tab is where you view the statistic information of sample
results.

Figure 4-44 Result Statistics tab

As Figure 4-44 shows, the search conditions are listed on the upper left of the
screen, such statistic data as SD and Mean on the lower left, expected range on the
upper right and statistic graph on the middle right.

4-127
Advanced Operations

To view statistic information of sample results

1 Specify the desired search conditions.

NOTE:
To specify the age range of the patients, enter the lower limit
of the range into the edit box to the right of Age Lower Limit
and the upper limit into the edit box to the right of Age
Upper Limit.
If you enter 0s into both the boxes, it means there is no
specific requirement on the ages.

2 Click the Search button and the analyzer will search for the matches and
present a statistic graph of the results found.

To exit the Results & Statistics screen

1 Click the Close button to exit the Results & Statistics screen.

4.3.4.3 Test Statistics tab

The Test Statistics tab is where you view the summary of the tests and the
reagents.

4-128
Advanced Operations

Figure 4-45 Test Statistics tab

To view the reagent status of a test

1 Click the Refresh button to refresh the displayed work list.


2 Click the desired test and the corresponding reagent status will be displayed
on the right of the screen.

To exit the Results & Statistics screen

1 Click the Close button to exit the Results & Statistics screen.

4-129
Advanced Operations

4.4 System

4.4.1 System Status


Click System System Status or press CTRL + P to enter the System Status
screen, as Figure 4-46 shows.

4-130
Advanced Operations

Figure 4-46 System Status screen

At the System Status screen, you can see two tabs:

Status
Others

4.4.1.1 Status tab

Figure 4-47 Status tab

As Figure 4-47 shows, you can check the status of the major components of the
analyzer.

4-131
Advanced Operations

Note that the light intensity of the lamp is expressed in Photometer Lamp. The bar
to the right of the Photometer Lamp represents the light intensity of the lamp in
term of percentage, light intensity of a new lamp being 100%. When the light
intensity is normal, the bar is green. When the light intensity is more than 50%, the
bar is yellow. When the lamp is to be changed, the bar is red. If the system does not
detect the light of the lamp, the bar is white.

The Network displays if the system is connected to the LIS.

Click the Close button to close the System Status screen.

4.4.1.2 Others tab

Figure 4-48 Others tab

As Figure 4-48 shows, you can not only view the dark currents of various
wavelengths, but also check the movement of the cuvette loader and wash the
sample and reagent probes and mixing bar.

The list shows the dark currents and backgrounds of the nine wavelengths and one
reference wavelength. The dark current refers to the AD output of the wavelength
when the lamp of the photometer is not switched on. The backgrounds refer to the
AD outputs generated by the lights when there are no cuvettes in the optical paths.

To test the dark current

1 When the analyzer is on standby, click the Test Dark Cur. button and the
readings will be displayed on the screen.

4-132
Advanced Operations

To load/unload cuvette segment to/from a specific position

1 When the analyzer is on standby, select the desired position and click the
Load Cuv. button to load/unload the segment to/from the selected
position.

To wash the sample and reagent probes and the mixing bar

1 When the analyzer is on standby, click the Enhanced Wash button and
the analyzer will use acid and alkaline detergents to wash the sample and
reagent probes and the mixing bar.

To exit the System Status screen

1 Click the Close button to exit the System Status screen.

4.4.2 Setup
The System Setup screen is where you set options regarding the routine settings,
reruns, test units and LIS connection.

To enter the System Setup screen, click System Setup, or press Ctrl + F3. A
dialog box shown as the figure below will pop up.

Click Cancel and a dialog box will then pop up to ask you to enter your user name
and password. You must have the corresponding authority to enter the System
Setup screen.

NOTE:
Only authorized personnel can enter the System Setup screen.

At the System Setup screen, shown in Figure 4-49, you can see the following tabs:

Routine
Rerun
Units
Host Communication

Follow the procedures given below to set them respectively.

4-133
Advanced Operations

4.4.2.1 Routine tab

Figure 4-49 Routine tab

As Figure 4-49 shows, there are several fields you can set at the Routine tab.
Follow the instructions given below to do so.

Field What it does How to set


Times of Auto Defines how many times the Enter the desired digit
Washing analyzer should wash the into the edit box to the
(Startup/Shutdow sample probe, reagent probe right of it.
n) and the mixing bar during every
startup or shutdown.
Alarm When Defines when the analyzer Enter the desired digit
Available Reagent should alert you about (must be greater than 2)
Not Enough For insufficient reagents. Also into the edit box to the
referred to as the insufficient right of it.
reagents alarm hereinafter.
Alarm Volume Defines the alarm volume when Drag the drag bar.
the analyzer gives alarm for
insufficient reagents. Six
volume levels, increasing from
left to right, are available.
Auto QC Defines whether the analyzer Click the check box to
should automatically insert QC the left of Yes to tick this
tests among routine tests. option and click it again
to cancel the selection.

4-134
Advanced Operations

Field What it does How to set


Tests Arranged Select the sequence mode of Select an option button.
the runs. Selecting by Samples
means arranging runs by the
sequence of samples; selecting
by Assays means arranging
runs by the sequence of tests.
Manual Bar Code If it is selected, you must enter Click the check box to
Scanner the bar code information when tick this option and click
requesting a sample. it again to cancel the
selection.

NOTE:
Set a suitable value for the insufficient reagents alarm so that the
analyzer can alert you about insufficient reagents in time.

When you have finished setting the desired fields, click the Save button to save the
changes. A dialog box will then pop up to remind you the settings are saved, as
Figure 4-50 shows. Click the OK button to close the dialog box.

Figure 4-50 A dialog to remind you the settings are saved

If you entered an improper value for any field, a dialog box will pop up to warn you
about the wrong setting after you clicked the Save button, as Figure 4-51 shows.
Seeing the dialog box, click OK button to close it and re-enter a proper value for the
field in question (insufficient reagents alarms, in case of Figure 4-51) and then save
the settings again.

Figure 4-51 A dialog box to warn you about wrong settings

When you have finished setting, you can click the Close button to exit the System
Setup screen.

4-135
Advanced Operations

4.4.2.2 Rerun tab

Figure 4-52 Rerun tab

Click the Rerun tab to view the 9 available fields in it, as Figure 4-52 shows. Follow
the instructions given below to set them.

Field What it does How to set


Rerun Defines how the rerun is Click the option button to
to be conducted, the left of Auto to select the
automatically or manually. auto rerun; Click the option
If automatically, the button to the left of Manual
analyzer will judge to select the manual rerun.
whether a rerun is
necessary based on the Note that the rest 8 fields
specified conditions (the are selectable only if you
other 8 fields in this tab); if have selected Auto.
manually, you should
decide whether to rerun
the test yourself and do it
manually.
Beyond Upper Limit Instructs the analyzer to Click the check box to the
of Reference Range re-dilute and then rerun right of Beyond Upper
the sample if the result of Limit of Reference Range
the previous run exceeds to select it; re-click the
the upper limit of the check box to de-select it.
expected range.

4-136
Advanced Operations

Field What it does How to set


Beyond Upper Limit Instructs the analyzer to Click the check box to the
of Linearity Range re-dilute and then rerun right of Beyond Upper
the sample if the result of Limit of Linearity Range
the previous run exceeds to select it; re-click the
the upper limit of the check box to de-select it.
linear range specified by
the instructions of the
reagent.
Beyond Response Instructs the analyzer to Click the check box to the
Range re-dilute and then rerun right of Beyond Response
the sample if the result of Range to select it; re-click
the previous run is the check box to de-select
beyond the allowed it.
response range.
Substrate out Instructs the analyzer to Click the check box to the
(Kinetic and Fixed re-dilute and then rerun right of Substrate out
Time) the sample if the Kinetic and Fixed Time) to
substrate ran out during select it; re-click the check
the previous sample run. box to de-select it.
It only applies to the runs
employing Kinetic or
Fixed Time methods.
Surplus Antigen Instructs the analyzer to Click the check box to the
re-dilute and then rerun right of Surplus Antigen to
the sample if the surplus select it; re-click the check
antigen was found during box to de-select it.
the previous sample run.

No Balance Point Instructs the analyzer to Click the check box to the
(Endpoint) re-dilute and then rerun right of No Balance Point
the sample if no balance (Endpoint) to select it;
point was detected during re-click the check box to
the previous sample run. de-select it.
It only applies to the
sample runs employing
Endpoint method.
If no dilution parameters
are set for a test, the
sample run of it will not be
rerun automatically even if
no balance point is
detected.

When you have finished setting the desired fields, click the Save button to save the
changes. A dialog box will then pop up to remind you the settings are saved, as
Figure 4-53 shows. Clicks OK button to close the dialog box.

4-137
Advanced Operations

Figure 4-53 A dialog to remind you the settings are saved

If you entered an improper value for any filed, a dialog box will pop up to warn you
about the wrong setting after you clicked the Save button. Seeing the dialog box,
click OK to close it and then re-enter a proper value for the field in question and then
save the settings again.

When you have finished settings, you can click Close to exit the System Setup
screen.

4.4.2.3 Units tab

Figure 4-54 Units tab

Click the Units tab (Figure 4-54) to set the units that will be used for the tests. Note
that all the units to be selected at other screens must be entered here.

Follow the instructions given below to add, edit or delete test units.

To add a new test unit

1 Click the Add button.

4-138
Advanced Operations

2 Enter the desired unit into the edit box to the right of Units. In case of
tests without units, enter blank (press the Space key) into the edit box.
3 Click the Save button to save the entered unit; Click the Cancel button,
for any reason, to discard the entered unit.

To edit an existing test unit

1 Click to the right of Units to display the pull-down list and select the
desired unit.
2 Click the Edit button.
3 Edit the unit in the edit box.
4 Click the Save button to save the changes; Click the Cancel button, for
any reason, to ignore the change.

To delete an existing test unit

1 Click to the right of Units to display the pull-down list and select the
desired unit.
2 Click the Delete button to delete the selected unit.

NOTE:
Once you click the Delete button, the software will immediately delete
the selected unit without asking you to confirm the deletion. Ensure the
right unit is selected before clicking the Delete button.

To exit the System Setup screen

1 Click the Close button to exit the System Setup screen.

4.4.2.4 Host Communication tab

The Host Communication tab is where you set the LIS connection mode and the
network parameters.

NOTE:
If you have no LIS server, the function of the screen is unavailable.

4-139
Advanced Operations

Figure 4-55 Host Communication tab

To set mode and network parameters

1 In the Mode field, if you select Real-time Send, the system will
automatically send sample results to the LIS server as soon as the tests
of a sample are all finished; otherwise, you can only send sample results
to the LIS server in batch mode at the Results & Statistics screen. Refer
to 4.3.4.1 Results tab for details.
If you select Real-time Download, the system will automatically
download sample information from the LIS server when it is scanning bar
code labels; otherwise, you can only download sample information from
the LIS server manually.
2 For the same sample, if the information downloaded from the LIS server is
different from the existing, selecting Neglect means the system will ignore
the information downloaded repeatedly, selecting Overwrite means the
system will overwrite the existing information with the new, and selecting
Add means the system will add the newly-downloaded test information to
the existing ones.
3 Enter the IP address of the server on the right of IP Address in the Host
Computer field.
4 Enter the port on the right of Port in the Host Computer field.
5 Click the Save button to save the settings.

4-140
Advanced Operations

NOTE:
There are two modes to send information to the LIS, the real-time
mode and the batch mode.
You can send information to the LIS in batch mode at any time only
except that when you select the Real Time Send at the System
Setup screen and the system is testing.

To connect/disconnect to the LIS

NOTE:
If the system is not connected to the LIS, the Test and the Connect
buttons are available; otherwise, the Disconnect button is available.

1 If you do not want to check the connection between the system and the
LIS, go directly to the next step;
Otherwise, click the Test button to check the network connection. The
system will pop up a dialog box to tell you whether the connection is
successful.
2 If you want to connect the system to the LIS, click the Connect button;
Otherwise, click the Disconnect button.

To exit the System Setup screen

1 Click the Close button to exit the System Setup screen.

4.4.3 Hospital
The Hospital Information screen is where you set the options regarding the
hospital and doctors.

To enter the Hospital Information screen, click System Hospital, or press CTRL
+ H. A dialog box will then pop up to ask you to enter your user name and password.
You must have the administrator authority to enter the Hospital Information screen.

NOTE:
You need the administrator authority to enter the Hospital
Information screen.

At the Hospital Information screen, shown in Figure 4-56, you can see three tabs:

Hospital
Doctors
All Doctors

4-141
Advanced Operations

Follow the instructions given below to set them.

4.4.3.1 Hospital tab

Figure 4-56 Hospital Information screen (Hospital tab)

As Figure 4-56 shows, you can set hospital name and departments at this tab. The
hospital name you set here will appear on the patients assay reports, and all the
departments to be selected at other screens must be entered here.

To edit the hospital name

1 Click the Edit button in the Name field.


2 Enter the desired hospital name into the edit box to the right of Name.
3 Click the Save button in the Name field to save the entered name.

To add a new department

1 Click the Add button in the Department field.


2 Enter the desired department name in the edit box to the right of
Department.
3 Click the Save button in the Department field to save the entered name;
Click the Cancel button, for any reason, to discard the entered name.

To edit an existing department name

1 Click to the right of Department to display the pull-down list and


select the desired department.
2 Click the Edit button in the Department field.
3 Edit the selected department name in the edit box to the right of
Department.

4-142
Advanced Operations

4 Click the Save button in the Department field to save the change; Click
the Cancel button, for any reason, to ignore the change.

To delete an existing department name

1 Click to the right of Department to display the pull-down list and


select the desired department.
2 Click the Delete button in the Department field to delete the selected
department.

NOTE:
Once you click the Delete button, the analyzer will immediately delete
the selected department, as well as all the doctors (see the next
section) belonging to this department without asking you to confirm the
deletion. Ensure the right department is selected before clicking the
Delete button.

To exit the Hospital Information screen

1 Click the Close button to exit the Hospital Information screen.

4.4.3.2 Doctors tab

Figure 4-57 Doctors tab

Click the Doctors tab to show the available fields, as Figure 4-57 shows. All the
doctors to be selected at other screens must be entered here.

There are four buttons at the middle part of the Doctors tab, as Figure 4-58 shows.

4-143
Advanced Operations

Figure 4-58 Buttons at the middle of the Doctors tab

Button Function
Displays the information of the first doctor of the selected
department.
Displays the information of the previous doctor of the selected
department.
Displays the information of the next doctor of the selected
department.
Displays the information of the last doctor of the selected
department.

If there is no existing doctor information in the selected department, the arrows will
appear gray, indicating they are not usable at this moment.

Follow the instructions given below to add, edit or delete doctors to a selected
department.

To add a new doctor

1 Click to the right of Dept. to display the pull-down list and select the
desired department.

2 Click the Add button and the analyzer will automatically assign a number
to the new doctor. Note that you cannot change this number.

3 Enter the doctors name in the edit box to the right of Name.

4-144
Advanced Operations

4 Enter the doctor s duty into the edit box to the right of Duty.
5 Click the Save button to save the change; Click the Cancel button, for
any reason, to ignore the change.

To edit information of an existing doctor

1 Click to the right of Dept. to display the pull-down list and select the
desired department.
2 Click the arrows to find the desired doctor information if necessary.
3 Click the Edit button.

4 Change the doctors name or duty as needed.


5 Click the Save button to save the change; Click the Cancel button, for
any reason, to ignore the change.

To delete the information of an existing doctor

1 Click to the right of Dept. to display the pull-down list and select the
desired department.
2 Click the arrows to find the desired doctor information if necessary.
3 Click the Delete button to delete the information of the selected doctor.

NOTE:
Once you click the Delete button, the analyzer will immediately delete
doctor from the selected department without asking you to confirm the
deletion. Ensure the right doctor is selected before clicking the Delete
button.

To exit the Hospital Information screen

1 Click the Close button to exit the Hospital Information screen.

4-145
Advanced Operations

4.4.3.3 All Doctors tab

Figure 4-59 All Doctors tab

Click the All Doctors tab to view all the saved doctor information, as Figure 4-59
shows.

To view the information of all the doctors, you can click to the right of Dept. and
select All departments.

To view the information of doctors of a specific department, click to the right of


Dept. and select the desired department.

You can click the Close button to exit the Hospital Information screen.

4.4.4 User
Click System User or press CTRL + B and a dialog box will pop up, as Figure
4-60 shows, to ask you to enter the appropriate user name and password.

Figure 4-60 A dialog box to ask you enter the appropriate user name and password

Enter a user name and a password of the administrator authority to enter the User
Management screen, as Figure 4-61 shows.

4-146
Advanced Operations

NOTE:
Only users of the administrator authority can enter the User
Management screen.

Figure 4-61 User Management screen

As Figure 4-61 shows, you can check the existing users or user groups at the User
Management screen.

To add a new user

1 Click User Management Add User to enter the Add User window.

2 Enter the necessary user information as instructed below.


User: Enter the maximum 15 letters or digits into the edit box as the user

4-147
Advanced Operations

name.
Description: Enter additional information regarding the users duty, title or
the like.
Password: Enter the password to be used by the user.
Confirm: Re-enter the password here to confirm the entry.
Authority: Assign the new user to an existing user group or groups. Click
button to add the new user to all the existing user groups; click
to remove the new user from all the existing groups; click
button to add the new user to a selected group or groups; click to
remove the new user from a selected group or groups.
Department: Select the department the new user belongs to.
Operator: Select the new users name from the pull down list.

3 Click the OK button to save the changes; click the Cancel button, for any
reason, to ignore the changes.

To delete a user (except for administrator)

1 Select the desired user from the User field.


2 Click User Management Delete User and a dialog box will pop up to
ask you to confirm the deletion.

3 Click the OK button to confirm the deletion; click the Cancel button to
abort the deletion.

4-148
Advanced Operations

To edit the user information of a user (except for administrator)

1 Select the desired user from the User Name field.


2 Click User Management Edit User to enter the Edit User window.

3 Edit the user information as needed.


4 Click the OK button to save the changes; click the Cancel button, for any
reason, to ignore the changes.

To change the user name and password of the administrator

1 Click User Management Change Admin. Password and a dialog box


will pop up to ask you to enter the user name and password of the
administrator authority.

4-149
Advanced Operations

2 Enter the user name and password as required and click the OK button
and another dialog box will pop up to ask you to enter the new user name
and password.

3 Enter the new username and password as required.

4 Click the OK button to save the changes; click the Cancel button, for any
reason, to ignore the changes.

To exit the User Management screen

1 Click on the upper right corner of the screen, or press ALT+ X, or click
User Management Close to exit the User Management screen.

4.4.5 Log
Click System Log or press CTRL + F2 to enter the Log Management screen, as
Figure 4-62 shows.

4-150
Advanced Operations

Figure 4-62 Log Management screen

As Figure 4-62 shows, the Log Management screen lists as many as 300 alarm
messages. Once the saved alarm messages reach 300, the analyzer will
automatically save the messages into a .log file and the file name is the saving time.

To check the current log

1 Click Log Current to list the latest 300 (or less) alarm messages.

To check the previously saved log file

1 Click Log View to enter the View Log File window.

Look in:

File Name: Open

Files of Type:
Cancel

2 Select the desired log file from the Log folder under the directory the
BS-300 operating software is installed and click the Open button to open
the selected log file.

4-151
Advanced Operations

To clear the current log

1 Click Log Clear and a dialog box will pop up to ask you to save the
current log to a .log file.
2 Click the OK button to save the log and then clear it; click the Cancel
button to clear the log without saving it.

To exit the Log Management screen

1 Click Log Exit or on the upper right corner of the screen to exit the
Log Management screen.

4.4.6 Maintenance
Click System Maintenance or press CTRL + Q and a dialog box will pop up, as
Figure 4-63 shows, to ask whether you are maintenance personnel.

Figure 4-63 A dialog box to ask you whether you are maintenance personnel

Click the Cancel button and another dialog box will pop up to ask you to enter the
appropriate username and password, as Figure 4-64 shows.

Figure 4-64 A dialog box to ask you to enter username and password

Enter a username and password of the administrator authority to enter the System
Maintenance screen, as Figure 4-65 shows.

4-152
Advanced Operations

Figure 4-65 System Maintenance screen

At the System Maintenance screen, the Sent Instructions field and the Received
Data field respectively display the instruction codes sent to the Analyzing Unit and
received by the Operation Unit.

NOTE:
Do not perform any maintenance action unless you are certain the
analyzer is on standby.

To start a maintenance operation

1 Click the desired tab.


2 Click the desired button.

To lock data

1 If you feel the codes displayed in the Sent Instructions and Received
Data fields refresh too fast, click the Lock button, which then will become
the Unlock button, and the displayed data will stop changing. You can
click the Unlock button to resume the data refreshing.

To clear data

1 Click the Clear button to clear the Sent Instructions and Received Data
fields.

To exit the System Maintenance screen

1 Click the Close button to exit the System Maintenance screen.

4-153
Advanced Operations

4.4.6.1 S. Probe tab

Figure 4-66 S. Probe tab

Button What it does


Above Inner Circle Click this button and the sample probe will move to a
position above inner circle of the sample disk.
Into Inner Circle Click this button and the sample probe will lower into
the sample tube in the inner circle and stop when it
contacts the liquid surface or the tube bottom.
Above Outer Circle Click this button and the sample probe will move to a
position above the outer circle of the sample disk.
Into Outer Circle Click this button and the sample probe will lower into
the sample tube in the outer circle and stop when it
contacts the liquid surface or the tube bottom.
Above Reaction Disk Click this button and sample probe will move to a
position above the dispensing hole of the reaction
disk.
Into Cuvette Click this button and the sample probe will lower into
the reaction cuvette and stop when it contacts the
liquid surface or the reaction cuvette bottom.
Above Wash Well Click this button and the sample probe will move to a
position above the wash well.

4-154
Advanced Operations

NOTE:
Ensure the sample disk is placed in the home position or stopped at
certain tube position before clicking Into Inner Circle or Into Outer
Circle button. Otherwise, it may lead to probe collision.
Ensure the reaction disk is placed in the home position or stopped at
certain cuvette position before clicking Into Cuvette button.
Otherwise, it may lead to probe collision.

4.4.6.2 R. Probe tab

Figure 4-67 R. Probe tab

Button What it does


Above Wash Well Click this button and the reagent probe move to a
position above the wash well.
Above Inner Circle Click this button and the reagent probe will move to a
position above inner circle of the reagent disk.
Into Inner Circle Click this button and the reagent probe will lower into
the reagent tube in the inner circle and stop when it
contacts the liquid surface or the tube bottom.
Above Outer Circle Click this button and the reagent probe will move to a
position above the outer circle of the reagent disk.

4-155
Advanced Operations

Button What it does


Into Outer Circle Click this button and the reagent probe will lower into
the reagent tube in the outer circle and stop when it
contacts the liquid surface or the tube bottom.
Above Reaction Disk Click this button and reagent probe will move to a
position above the dispensing hole of the reaction
disk.
Into Cuvette Click this button and the reagent probe will lower into
the reaction cuvette and stop when it contacts the
liquid surface or the reaction cuvette bottom.

NOTE:
Ensure the reagent disk is placed in the home position or stopped at
certain reagent position before clicking Into Inner Circle or Into Outer
Circle button. Otherwise, it may lead to probe collision.
Ensure the reaction disk is placed in the home position or stopped at
certain cuvette position before clicking Into Cuvette button.
Otherwise, it may lead to probe collision.

4.4.6.3 Mixing Bar tab

Figure 4-68 Mixing Bar tab

4-156
Advanced Operations

Button What it does


Above Wash Well Click this button and the mixing bar will move to a
position above the wash well.
Into Wash Well Click this button and the mixing bar will lower into the
wash well.
Above Reaction Disk Click this button and the mixing bar will move to a
position above the mixing hole of the reaction disk.
Into Cuvette Click this button and the mixing bar will lower into the
reaction cuvette.
Mix Enter the desired integer into the edit box to the right
of Time and click the Mix button to lower the mixing
bar into the reaction cuvette and mix the specified
time (expressed in terms of units and every unit
stands for 20ms).

NOTE:
Ensure the reaction disk is placed in the home position or stopped at
certain cuvette position before clicking Into Cuvette or Mix button.
Otherwise, it may lead to bar collision.

4.4.6.4 Reac. Disk tab

Figure 4-69 Reac. Disk tab

4-157
Advanced Operations

Button What it does


Rotate Given Circles to Enter the desired number into the edit box below
Given Position Circles to specify how many circles the reaction
disk should run. Enter the desired number (1-80)
into the edit box below Position to specify at
which cuvette position the reaction disk should
stop.
Click this button and the reaction disk will rotate
the given circles and stop at the given position.
Rotate Given Positions Enter the desired number (1-80) into the edit box
below Position to specify how many cuvette
positions the reaction disk should rotate.
Click this button and the reaction disk will rotate
the given positions.

NOTE:
Before operating the reaction disk, ensure the sample and reagent
probes and mixing bar are away from it. Otherwise, the moving disk
may bend the probes or bar.

4.4.6.5 Rgt. Disk tab

Figure 4-70 Rgt. Disk tab

4-158
Advanced Operations

Button What it does


Rotate Given Circles to Enter the desired number into the edit box
Given Position below Circles to specify how many circles the
reagent disk should run. Enter the desired
number (1-50) into the edit box below Position
to specify at which position the reagent disk
should stop.
Click this button and the reagent disk will rotate
the given circles and stop at the given position.
Rotate Given Positions Enter the desired number (1-50) into the edit
box below Position to specify how many
positions the reagent disk should rotate.
Click this button and the reagent disk will rotate
the given positions.

NOTE:
Before operating the reagent disk, ensure the reagent probe is away
from it. Otherwise, the moving disk may bend the probes.

4.4.6.6 S. Disk tab

Figure 4-71 S. Disk tab

4-159
Advanced Operations

Button What it does


Rotate Given Circles to Enter the desired number into the edit box
Given Position below Circles to specify how many circles the
sample disk should run. Enter the desired
number (1-60) into the edit box below Position
to specify at which position the sample disk
should stop.
Click this button and the sample disk will rotate
the given circles and stop at the given position.
Rotate Given Positions Enter the desired number (1-60) into the edit
box below Position to specify how many
positions the sample disk should rotate.
Click this button and the sample disk will rotate
the given positions.

NOTE:
Before operating the sample disk, ensure the sample probe is away
from it. Otherwise, the moving disk may bend the probe.

4.4.6.7 Fluid Path tab

Figure 4-72 Fluid Path tab

4-160
Advanced Operations

Button What it does


Wash Sample Probe Select Inner Surface, Outer Surface or Both
Surfaces from the pull down list to the right of this
button and click it to wash the exterior or interior of
the sample probe.
Wash Reagent Probe Select Inner Surface, Outer Surface or Both
Surfaces from the pull down list to the right of this
button and click it to wash the exterior or interior of
the reagent probe.
Wash Mixing Bar Click it to wash the exterior of the mixing bar.
S. Syringe Aspirate Enter the desired number (3-45) into the edit box
below Volume(S. Syringe).
Click this button and the sample syringe aspirates
the given volume (l).
S. Syringe Dispense Enter the desired number (3-45) into the edit box
below Volume(S. Syringe).
Click this button and the sample syringe dispenses
the given volume (l).
Reset S. Syringe Click this button and reset the sample syringe.
R. Syringe Aspirate Enter the desired number (30-450) into the edit box
below Volume(R. Syringe).
Click this button and the reagent syringe aspirates
the given volume (l).
R. Syringe Dispense Enter the desired number (30-450) into the edit box
below Volume(R. Syringe).
Click this button and the reagent syringe dispenses
the given volume (l).
Reset R. Syringe Click this button and reset the reagent syringe.

NOTE:
When you have clicked the S. Syringe Dispense button, if there is no
enough liquid inside the sample syringe, the plunger will stop at the
home position.
When you have clicked the R. Syringe Dispense button, if there is no
enough liquid inside the reagent syringe, the plunger will stop at the
home position.

4-161
Advanced Operations

4.4.6.8 Cuv. Comp. tab

Figure 4-73 Cuv. Comp. tab

Button What it does


Push Cuvettes Click the button and the trolley of the cuvette loader
should move to the loading position to load a new
cuvette segment.

4-162
Advanced Operations

4.4.6.9 Lamp tab

Figure 4-74 Lamp tab

Button What it does


Lamp On Click this button to turn on the lamp of the
photometer.
Lamp Off Click this button to turn off the lamp of the
photometer.

NOTE:
If nothing is to be run for a while and you have no intention to exit the
control system, you can turn off the lamp to maximize its service life.

4-163
Advanced Operations

4.4.6.10 Autoloader tab

Figure 4-75 Autoloader tab

Button What it does


Load Select the desired segment position from the pull
down list below Segment Position and click this
button to load a cuvette segment to the given
position in the reaction disk.
Start Select the desired motions from the pull down list
below Autoloader Motion and click this button to
start the manipulator.
Unload Select the desired segment position from the pull
down list below Segment Position and click this
button to unload a cuvette segment from the given
position in the reaction disk.
To Ver. Initial Position Click this button to move the manipulator vertically to
its initial point.
To Hor. Initial Position Click this button to move the manipulator horizontally
to its initial point.

4-164
Advanced Operations

4.4.6.11 Others tab

Figure 4-76 Others tab

Button What it does


Download Settings If the communication between the analyzing unit and
the operation unit is failed, you can download the
settings to the analyzing unit by clicking this button.
The operation unit will try to resume the
communication with the analyzing unit.
Reset Mechanical Parts Click this button to reset all the mechanical parts of
the analyzing unit.

4.4.7 Print
The Printing Setup screen is where you set the parameters regarding printing
reports.

To enter the Printing Setup screen, click System Print, or press CTRL + O.

4-165
Advanced Operations

Figure 4-77 Printing Setup screen

At the Printing Setup screen, Figure 4-77, there are four tabs:

Patient Report defines parameters regarding the format of the patient report.
Test Report defines items to appear on the printed test report.
Others defines printing order and printer.
Report Header defines the report header.

4.4.7.1 Patient Report tab

Figure 4-78 Patient Report tab

4-166
Advanced Operations

As Figure 4-78 shows, the items listed in the Required field are required to appear
on the patient report; the items listed in the Optional field are selectable items and
you can select any interested items to print on the patient report.

To set the items to appear on the patient report

1 Tick the check boxes to the left of the desired items. Clicking the ticked
check box again can cancel the selection.
2 Click the Save button and a dialog box will pop up to tell you settings are
saved. Click the OK button to close the box.

To set number of reports printed on each page (A4)

1 In the Number of Reports on Each Page(A4) field select 2 or 1 to print 2


or 1 report on each page.

To exit the Printing Setup screen

1 Click the Close button to exit the Printing Setup screen.

4.4.7.2 Test Report tab

Figure 4-79 Test Report tab

4-167
Advanced Operations

As Figure 4-79 shows, the items listed in the Required field are required to appear
on the test report; the items listed in the Optional field are selectable items and you
can select any interested items to print on the report.

To set the items to appear on the test report

1 Tick the check boxes to the left of the desired items. Clicking the ticked
check box again can cancel the selection.
2 Click the Save button and a dialog box will pop up to tell you settings are
saved. Click the OK button to close the box.

To exit the Printing Setup screen

1 Click the Close button to exit the Printing Setup screen.

4.4.7.3 Others tab

Figure 4-80 Others tab

To set the printing order

1 Select the desired test.

4-168
Advanced Operations

2 Select the desired sequence number from the pull down list above Set
Printing Order.

3 Click Set Printing Order.

To set the default printer

1 Select the desired printer from the pull down list in the Default Printer
area.

2 Click the Default Printer button to set the selected printer as the default
printer.

To enable auto print and edited result flag

1 Select the Print after Sample Run checkbox.

The system will automatically print reports of the sample runs whose
results are within or beyond reference range. A ! symbol will be printed
next to the sample ID of abnormal sample runs.

2 Select the Flag Edited Results checkbox to flag the edited results on
patient report.

To exit the Printing Setup screen

1 Click the Close button to exit the Printing Setup screen.

4-169
Advanced Operations

4.4.7.4 Report Header tab

Figure 4-81 Report Header tab

NOTE:
The format of the printout may not be exactly the same as that
displayed in the print preview window.

To set the content and font of the report header

1 Enter the desired header into the edit box to the right of Header.
2 Select the desired font from the pull down list to the right of Font.
3 Click the Setup button and a dialog box will pop up to tell you the settings
that have been saved. Click the OK button to close the box.

To exit the Printing Setup screen

1 Click the Close button to exit the Printing Setup screen.

4-170
Advanced Operations

4.4.8 Temperature
Click System Temperature or press CTRL + T and a dialog box will pop up, as
Figure 4-82 shows, to ask you whether you want to debug the temperature control
unit or check the temperature curve.

Figure 4-82 A dialog box to ask you whether to debug or to view curves

Click the Cancel button and another dialog box will pop up to ask you to enter
appropriate username and password, as Figure 4-83 shows.

CAUTION:
Only service engineers are authorized to debug the temperature
control unit.

Figure 4-83 A dialog box to ask you to enter username and password

Enter the administrator username and password as required to enter the


Temperature Control screen, as Figure 4-84 shows.

4-171
Advanced Operations

Figure 4-84 Temperature Control screen

As Figure 4-84 shows, you can see two tabs at the Temperature Control screen:

Preheating Temp. Curve tab - displays temperature curve of preheating the


reagents.
Reaction Temp. Curve tab - displays the temperature curve of the reaction disk.

For every temperature curve, its x-axis represents the time and y-axis the
temperature.

The parameters are interpreted as follows.

Parameter What it does How to set


Precision Defines the display Select one of the nine
precision of the options (0.01, 0.02, 0.05,
temperature. 0.1, 0.2, 0.5, 1, 2 and 5)
from the pull down list.
Range Defines the display range Enter the lower limit into the
of the y-axis of the first edit box and the upper
temperature curve. limit into the second.

To check the current temperature curve

1 Select the desired temperature precision in the right of Precision.


2 Enter the desired display range in the right of Range.
3 Click the Save button.

To check the previous temperature curve

1 Click the History button.

4-172
Advanced Operations

2 Click or button until the desired curve is reached.

NOTE:
Click to check the previous curve;

Click to check the next curve.

3 Enter the desired display range in the right of Precision.


4 Enter the desired display range in the right of Range.
5 Click the Save button to check the curve.
6 When you have viewed the curve, click the History button to quit viewing
the history curve.

To lock a curve

1 Click the Lock button to lock the currently displayed curve.

To unlock a curve

1 Click the Unlock button to unlock the currently locked curve.

To exit the Temperature Control screen

1 Click the Close button to exit the Temperature Control screen.

4.4.9 Database
Click System Database or press CTRL + L and a dialog box will pop up, as
Figure 4-85 shows, to ask you to enter the appropriate user name and password.

Figure 4-85 A dialog box to ask you enter the user name and password

Enter the user name and password of the administrator authority to enter the
Database Maintenance screen as Figure 4-86 shows.

4-173
Advanced Operations

Figure 4-86 Database Maintenance screen

NOTE:
You can only maintain the database when the analyzer is on standby.
This analyzer automatically backs up data. You only need to maintain
the database when the system alerts you that the database is
damaged.
Usually you do not need to delete the reaction curves. Deleting the
curves (only curves, not the results) only when you find insufficient
space left on the hard disk.

To repair and compress current database

1 When the analyzer warns you about database damage, click the Repair &
Compress button to repair and compress current database.

To delete certain reaction curve

1 Select the desired start date from the pull down list to the right of the
From.
2 Select the desired end date from the pull down list to the right of the To.
3 Click the Delete Reaction Curve And Data button to delete the curve of
the selected period.

To delete all the reaction curves

1 Click the Delete All Reaction Curve And Data button to delete all the
curves.

4-174
Advanced Operations

To exit the Database Maintenance screen

1 Click the Close button or the button at the upper right corner to exit
the Database Maintenance screen.

4.4.10 Bar Code Scanner

NOTE:
If the system has no built-in bar code scanner (optional), the sub menu
is gray and unavailable.

Click System Bar Code Scanner or press Ctrl+L and a dialog box shown as the
figure below will pop up.

Click the Cancel button and another dialog box shown as the figure below will pop
up.

Enter authorized user name and password to enter the Bar Code Scanner screen.

4-175
Advanced Operations

Figure 4-87 Bar Code Scanner screen

The screen displays the bar code symbols that can be read by the built-in scanner.
You can select the ones that you need.

To set bar code symbols

1 Select a symbology.
2 If you want the system to check the bar code of the selected symbology
after reading it, select the corresponding check box in the Check column.

NOTE:
To prevent from misreading, Mindray recommends you to
select the check box.

3 Enter the bar code length of the selected symbology into the
corresponding edit box in the Symbol Length column. The length is
between 3 and 16.
4 Repeat from step 1 to step 3 until all the symbologies you need are set.
5 Click the Save button and a dialog box will pop up to remind you that the
setup is completed.
6 Click the OK button.
7 Click the button at the upper right corner of the Bar Code Scanner
screen to exit the screen.

4.4.11 ISE Unit


Click System ISE Unit or press Ctrl+D to enter the ISE screen.

4-176
Advanced Operations

Figure 4-88 ISE screen

There are several tabs at the screen. They are:

Calibration
Setup
ISE Unit

4.4.11.1 Calibration

The Calibration tab displays the calibration results and limits of the ISE tests and
you can calibrate them.

NOTE:
An ISE calibration should be run in case of any of the following
accidents:
(1) More ISE tests are to be run after the system is powered on;
(2) It has been eight hours since the last ISE calibration is run after the
latest power-on;
(3) More ISE tests are to be run after the daily cleaning procedure. But
Mindray recommends running an ISE calibration after the daily
cleaning procedure.

To calibrate ISE unit

1 Click the Calibrate button and the system will calibrate the ISE unit.

4-177
Advanced Operations

4.4.11.2 Setup

The Setup tab is where you can set parameters for electrodes and tubing, set
coefficients for results correction and view information of the reagent module.

Figure 4-89 Setup tab

Parameter What it does How to set


Alarm Limit The system will remind you to Enter a value into the edit
(Days) replace the electrode (or tubing) box.
the days entered by you before the
electrode (or tubing) should be
replaced next time.
Alarm Limit The system will remind you to Enter a value into the edit
(Times) replace the electrode (or tubing) box.
when the electrode (or tubing) is
only enough for how many
samples the number of which is
entered by you.
(Electrode) Records the date when you Tick the combo box and
Replaced on replace the electrode. After then select the date from
replacing an electrode each time, the pulldown list box.
remember to record the date.
(Tube) Records the date when you Tick the combo box and
Replaced on replace the tubing. After replacing then select the date from
tubing each time, remember to the pulldown list box.
record the date.
Expiration Expiration date of the reagent -
Date module.

4-178
Advanced Operations

Parameter What it does How to set


Installation Installation date of the reagent -
Date module.
Total Total volume of the reagent -
Volume module.
Inventory(%) Ratio between the volume of -
remaining reagents and the total
volume.
Lot No. Lot No. of the reagent module. -
Expired Expired mode of the reagent -
Mode module.

To set parameters for electrodes, reagents and tubing

1 Enter parameters for electrodes and tubing as you need.


2 Click the Save button.

To set coefficients for ISE results correction

1 Enter coefficients a, b for ISE tests as you need.

2 Click the Save button. The system will correct the results of the tests you
have set according to the formula Y=aX+b (Y means new result and X
means original result).

NOTE:
The system will save the new results only.

To exit the ISE screen

1 Click the Close button.

4-179
Advanced Operations

4.4.11.3 ISE Unit

The ISE Unit tab is where you can maintain the ISE unit.

Figure 4-90 ISE Unit tab

Parameter What it does How to set


Times Defines the times the system Enter a value into the edit
runs A purge and B purge for box.
after click the Purge AB button.

Button What it does


Handshake Click this button to send an instruction to make the ISE unit
handshake with main unit.
Pump Calib. Click this button to send an instruction to correct the pumps.
Maintenance Click this button to send an instruction to maintain the ISE
unit.
Bubble Calib. Click this button to send an instruction to check bubbles.
Clean Click this button to send an instruction to clean the ISE unit
with ISE cleaning solution.
Before cleaning, place the cleaning solution to the position
45 on the sample disk.
Purge A Click this button to send an instruction to purge with
Calibrant A.

4-180
Advanced Operations

Button What it does


Purge B Click this button to send an instruction to purge with
Calibrant B.
Purge AB Click this button to send an instruction to purge with
Calibrant A and B alternately for given times.
Save Sent Click this button to save the instructions displayed in the
Sent instruction window to a .txt file in the directory where
the executable file of the operating software locates.
Save Received Click this button to save the instructions displayed in the
Received instruction window to a .txt file in the directory
where the executable file of the operating software locates.
Clear Click this button to clear the displayed in the upper two
windows.
ISE Reset If there is something wrong with the ISE unit and ISE tests
cannot proceed, click this button to reset the ISE unit.

4.5 Help

Help Version

Guide

4.5.1 Version

Click Help Version or press CTRL + S to enter the Software version screen.

Click the Close button to exit the Software Version screen.

4.5.2 Guide
Click Help Guide or press CTRL + F11 to view the on-line help document.

4-181
5 Maintenance
To ensure reliability, good performance and long service life of the system, regular
maintenance is required. Be sure to follow the instructions given below to maintain
the system. In case of problems beyond your ability or not covered in this chapter, be
sure to call Mindray or your local distributor in time.

WARNING:
Do not perform any maintenance procedures that are not described
in this chapter.
Do not touch the components other than the ones specified in this
chapter.
Performing unauthorized maintenance procedures can damage your
system, void any applicable warranty or service contract and even
cause personnel injuries.
After performing any maintenance actions or procedures, ensure
that the system runs normally.
For some maintenance actions or procedures, be sure to place the
POWER to OFF first.
Do not spill water or detergent on the mechanical or electrical
components of the system.

BIOHAZARD:
Wear gloves, goggles and protective clothing.

To ensure maintenance is performed periodically, Mindray recommends the copy of


the maintenance log in 5.7 Maintenance Log be used to keep the maintenance
records.

5.1 Preparation
The following things may facilitate your maintenance.

Tools

M1.5, M3 and/or M4 hex wrenches


Cross-headed screwdrivers (small, medium and large)
Syringe
A clean cup
Tweezers
Clean gauze
Clean cotton stick

5-1
Maintenance

Brush (to wash tanks)

Detergents

Acid: 0.1mol/l hydrochloric acid


Alkaline: 0.5% (V/V) javel water

WARNING:
Poisonous gas will be produced if acid wash solution is mixed with
alkaline wash solution. Do not mix the acid wash solution with the
alkaline one.

CAUTION:
Mindray has specified the following enhanced wash solutions:
Acid wash solution: 0.1mol/l hydrochloric acid;
Alkaline wash solution: 0.5%(V/V) javel water.
Be sure to use the detergents recommendedenhanced wash solution
specified by Mindray. Otherwise, proper result may not be obtained.
Mindray recommends the acid and alkaline wash solutions be used
alternately. For instance, if the acid wash solution is used at current
startup, the alkaline one should be used at next startup.

Others

Ethanol
Disinfector

5.2 Daily Maintenance

5.2.1 Checking Sample/Reagent Syringes


Follow the procedure given below to check the sample or reagent syringe. The
purpose of this check is to ensure the syringe does not leak.

1 Place the ANALYZING UNIT POWER to OFF.

5-2
Maintenance

2 Open the middle door of the analyzing unit and you will see both of the
syringes, the sample syringe on the right and the reagent syringe on the
left.

T-piece

Plunger Guide Cap

3 Check whether the T-piece leaks.


If not, proceed to the next step.
If so, contact Mindray Customer Service Department or the distributor.
4 Check whether the plunger guide cap leaks.
If not, proceed to the next step.
If so, replace the cap as instructed by 5.6.6 Replacing Plunger
Assembly of Sample/Reagent Syringe .
5 Close the middle door.

5.2.2 Checking/Washing Sample Probe


At every startup, the system will automatically wash the interior and exterior of the
sample probe with deionized water. At every shutdown, the system will automatically
wash the interior and exterior of the sample probe with the acid and alkaline
detergents.

During the washing process, check whether the flow from the interior of the sample
probe is continuous and in the direction of the probe; check the exterior of the
sample probe to see whether the flow is continuous and normal.

If the flow appears abnormal, wash the probe as instructed by 5.3.1 Washing
Sample Probe.

If the flow remains abnormal after washing, clean the probe as instructed by 5.6.1
Unclogging Sample Probe.

5-3
Maintenance

If the flow remains abnormal after cleaning, contact Mindray Customer Service
Department or the distributor.

5.2.3 Checking/Washing Reagent Probe


At every startup, the system will automatically wash the interior and exterior of the
reagent probe with deionized water. At every shutdown, the system will automatically
wash the interior and exterior of the reagent probe with the acid and alkaline
detergents.

During the washing process, check whether the flow from the interior of the reagent
probe is continuous and in the direction of the probe; check the exterior of the
reagent probe to see whether the flow is continuous and normal.

If the flow appears abnormal, wash the probe as instructed by 5.3.2 Washing
Reagent Probe.

If the flow remains abnormal after washing, clean the probe as instructed by 5.6.2
Unclogging Reagent Probe.

If the flow remains abnormal after cleaning, contact Mindray Customer Service
Department or the distributor.

5.2.4 Checking/Washing Mixing Bar


At every startup, the system will automatically wash the mixing bar with deionized
water. At every shutdown, the system will automatically wash the mixing bar with the
acid and alkaline detergents.

During the washing process, check whether the bar rotates normally. If not, contact
Mindray Customer Service Department or the distributor.

5.2.5 Checking Detergents

WARNING:
Always wear gloves when checking the detergents.

1 Make sure the system is on standby status, and ensure no part of the
system will motion.
2 Remove the cover from the reagent disk.
3 Check how much detergent left in the bottle on position 45. If there is not
much left, add more detergent or install a new bottle of detergent.
4 Check how much detergent left in the bottle on position 47. If there is not
much left, add more detergent or install a new bottle of detergent.
5 Replace the cover.

5-4
Maintenance

Every wash consumes about 450l detergent when the system uses detergent to
clean the probes and the mixing bar.

5.2.6 Checking Connection of Deionized Water


1 Place the ANALYZING UNIT POWER to OFF.
2 Open the lower door of the analyzing unit and you will find the deionized
water tank on the right.

Deionized
Waste Tank
Water Tank

Used-Cuvettes
Bucket

Extractable
Plate

3 Check the connections between the two deionized water connectors (green
and red) and their counterparts on the analyzer.

Deionized Water Connector Deionized Water Connector


(Green) (Red) BNC Connector

Pickup Tube

If not, jump to step 5.


If you see leaks, wipe off the water with clean gauze and proceed to the next
step.
4 Check whether the connectors are loose.
If not, proceed to the next step.
If so, pinch the connector with fingers and screw it counter-clockwise to
remove it and then screw it back on.
5 Check the connections between the pickup tubes and their connectors.
If not, go directly to the next step.
If you see leaks, wipe off the water with clean gauze and tighten the
connections and proceed to the next step.
6 Pull out the extractable plate.

5-5
Maintenance

7 Check the connection between the pickup tube and the tank cap.
If not, go directly to the next step.
If you see leaks, wipe off the water with clean gauze and tighten the
connection and proceed to the next step.
8 Push the extractable plate back in.
9 Close the lower door.

CAUTION:
Make sure the deionized water pickup tubes are neither blocked nor
bent.

NOTE:
If leaking remains, please contact Mindray Customer Service
Department or the distributor.

5.2.7 Checking Connection of Wastewater

BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the following checks.
Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place the ANALYZING UNIT POWER to OFF.


2 Open the lower door of the analyzing unit and you will find the waste tank
on the left.

5-6
Maintenance

3 Check the connections between waste connector and the waste tap of the
analyzing unit.

BNC Connector Waste Connector

Pin

Waste Tap
Waste Tube

If no leaks, go directly to the next step.


If you see leaks, wipe off the water with clean gauze and then press the
pin on the tap and grab the tap and pull it off the connector. Keep pressing
the pin and re-insert the connector back to the tap and proceed to the next
step.
4 Check the connection between the waste tube and the tap.
If no leaks, go directly to the next step.
If you see leaks, wipe them off with clean gauze and tighten the waste
tube and then proceed to the next step.
5 Pull out the extractable plate.
6 Check the connection between the waste tube and the tank cap.
If no leaks, go directly to the next step.
If you see leaks, wipe them off with clean gauze and tighten the
connection and proceed to the next step.
7 Push the extractable plate back in.
8 Close the lower door.

CAUTION:
Ensure the waste tube is neither blocked nor bent. A blocked or bent
waste tube may lead to waste overflow that may damage the analyzing
unit.

NOTE:
If leaking remains, please contact Mindray Customer Service
Department or the distributor.

5-7
Maintenance

5.2.8 Checking Remaining Deionized Water

CAUTION:
The deionized water to be used on the BS-300 must meet the CAP
Type II water requirements and the resistance rate must be no less
than 0.5Mcm.

1 Place the ANALYZING UNIT POWER to OFF.


2 Open the lower door of the analyzing unit and you will see the deionized
water tank on the right.
3 Check how much deionized water left in the tank.
If there is still enough left, jump to step 9.
If not, proceed to the next step.
4 Pull out the extractable plate.
5
CAUTION:
After removing the cap of the deionized water tank (together
with the pickup tube and level sensor), place it on a clean
table.

Unscrew (counter-clockwise) the tank cap and remove the cap together
with the pickup tube and the level sensor.
6 Add deionized water to the tank.
7 Screw (clockwise) the cap together with the pickup tube and the level
sensor back on to the tank until secure.
8 Push the extractable plate back in.
9 Close the lower door.

CAUTION:
Ensure the deionized water pickup tube is neither blocked nor bent.

5.2.9 Emptying Waste Tank

BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the below checks.
Dispose of the wastewater in accordance with your local or national
guidelines for biohazard waste disposal and consult the manufacturer
or distributor of the reagents for details.

5-8
Maintenance

1 Place the ANALYZING UNIT POWER to OFF.


2 Open the lower door of the analyzing unit and you will see the waste tank
on the left.
3 Pull out the extractable plate.
4
BIOHAZARD:
After removing the cap of the waste tank (together with the
tube and level sensor), place it on an appropriate place to
avoid biohazard contamination.

Unscrew (counter-clockwise) the tank cap and remove it together with the
waste tube and the level sensor from the tank.
5 Take out the tank and empty it.
6 Replace the tank back into the cabinet.
7 Screw (clockwise) the cap (together with the waste tube and the level
sensor) back onto the tank until secure.
8 Push the extractable plate back in.
9 Close the lower door.

CAUTION:
Ensure the waste tube is neither blocked nor bent. A blocked or bent
waste tube may lead to wastewater overflow that may damage the
analyzer.

5.2.10 Emptying Used-Cuvettes Bucket

BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the below checks.
Dispose of the used cuvettes in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place the ANALYZING UNIT POWER to OFF.


2 Open the lower door of the analyzing unit and you will see the
used-cuvettes bucket in the middle.
3 Take out the bucket and empty it.
4 Replace the bucket back into the cabinet.
5 Close the lower door.

5-9
Maintenance

5.2.11 Cleaning ISE Unit (optional)

BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the below checks.
The cleaning solution is irritating to eyes and skin. Avoid contact with
skin and eyes. In case of contact with eyes, rinse immediately with
plenty of water and seek medical advice.

CAUTION:
Use the consumables recommended by Mindray. Other consumables
may degrade system performance.
Add solution supplied in the cleaning solution kit to top of label on the
powder bottle that is also supplied in the same kit and shake well to
prepare the cleaning solution.
The cleaning solution must be stored at 2-8C and discarded two
weeks after being prepared.

NOTE:
The maintenance is necessary to be performed when the ISE unit
(optional) is connected.
You should perform the maintenance once a day after all the samples
analyzed. Besides, if the samples of a day requested for the ISE tests
are 50 or more, you should perform the maintenance after 50 samples
are analyzed.
If you give the electrodes some time to stabilize after cleaning, you will
experience slightly better performance.

1 Select SystemISE Unit screen of the operating software.


2 Select the ISE Unit tab.
3 Click the Clean button and a dialog box will pop up to remind you to put
ISE cleaning solution to the position 45 on the sample disk.
4 Load about 200l of ISE cleaning solution to the position 45 on the sample
disk and click the OK button.
5 After cleaning, if there are samples requested for the ISE tests to be run,
calibration should be run first. But Mindray recommends running an ISE
calibration after cleaning.

5.2.12 Pump Calibration (optional)


1 Select SystemISE Unit screen of the operating software.
2 Select the ISE Unit tab.
3 Click the Pump Calib. button.

5-10
Maintenance

5.3 Weekly Maintenance

5.3.1 Washing Sample Probe

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when washing the probe.
Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place the ANALYZING UNIT POWER to OFF.


2 Remove the cover from the sample disk. Shift the disk handle from
horizontal position to vertical position. Grab the handle or handwheel and
pull the disk upward to remove it.

3 Pull the sample probe arm to the highest point by hand. Rotate the probe
arm to a position above the sample compartment and convenient to
operate.

5-11
Maintenance

4
CAUTION:
The tweezers may scratch the probe. Exercise caution when
using it to wash the probe. Avoid direct contact between the
tweezers and the probe. Do not use excessive force when
washing the probe. Otherwise it may bend.

Pinch acid or alkaline detergent-soaked (ethanol-soaked also applicable)


gauze with tweezers and gently clean the exterior of the sample probe
until it is clean and smooth.

NOTE:
Mindray recommends the acid and alkaline detergents be
used alternately for this purpose. For instance, if the acid
detergent was used for last cleaning, the alkaline detergent
would better be used for this time.

5 Pinch deionized water-soaked gauze to wash the sample probe.


6 After washing, gently pull the probe arm to its highest point and rotate the
probe arm to move the probe to a position above the wash well.

CAUTION:
After washing the probe, be sure to rotate the probe to a
position above the wash well.

7 Keep the sample disk handle in the vertical position. Align the central hole
of the handwheel to the rotor and lower the sample disk all the way down.
Rotate the disk a bit as necessary during the lowering process. When the
disk is in place, shift the handle back to horizontal position to lock the disk
to the rotor.
To ensure the disk is secured, you may grab the handwheel and gently
pull it upward. If the disk does not loose, it means it is secured and you
can replace the cover; otherwise, re-install the disk.
8 Place the ANALYZING UNIT POWER to ON and the system will
automatically reset the sample probe and wash it with deionized water.

5-12
Maintenance

5.3.2 Washing Reagent Probe

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when washing the probe.
Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place the ANALYZING UNIT POWER to OFF.


2 Remove the cover from the reagent disk. Shift the disk handle from
horizontal position to vertical position. Grab the handle or handwheel and
pull the disk upward to remove it.

3 Pull the reagent probe arm to the highest point by hand. Rotate the probe
arm to a position above the reagent compartment and convenient to
operate.

5-13
Maintenance

4
CAUTION:
The tweezers can scratch the probe. Exercise caution when
using it to wash the probe. Avoid direct contact between the
tweezers and the probe. Do not use excessive force when
washing the probe. Otherwise it may bend.

Pinch acid or alkaline detergent-soaked (ethanol-soaked also applicable)


gauze with tweezers and gently clean the exterior of the reagent probe
until it is clean and smooth.

NOTE:
Mindray recommends the acid and alkaline detergents be
used alternately for this purpose. For instance, if the acid
detergent was used for last cleaning, the alkaline detergent
would better be used for this time.

5 Pinch deionized water-soaked gauze to wash the reagent probe.


6 After washing, gently pull the probe arm to its highest point and rotate the
probe arm to move the probe to a position above the wash well.

CAUTION:
After washing the probe, be sure to rotate the probe to a
position above the wash well.

7 Keep the reagent disk handle in the vertical position. Align the central hole
of the handwheel to the rotor and lower the reagent disk all the way down.
Rotate the disk a bit as necessary during the lowering process. When the
disk is in place, shift the handle back to horizontal position to lock the disk
to the rotor.
To ensure the disk is secured, you may grab the handwheel and gently
pull it upward. If the disk does not loose, it means it is secured and you
can replace the cover; otherwise, re-install the disk.
8 Place the ANALYZING UNIT POWER to ON and the system will
automatically reset the reagent probe and wash it with deionized water.

5-14
Maintenance

5.3.3 Washing Mixing Bar

BIOHAZARD:
Wear gloves, goggles and protective clothing when washing the probe.
Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place the ANALYZING UNIT POWER to OFF.


2 Pull the mixing bar arm to the highest point by hand. Rotate the bar arm to
a position convenient to operate.

3
CAUTION:
The tweezers can scratch the bar. Exercise caution when
using the tweezers to wash the bar. Avoid direct contact
between the tweezers and the bar. Do not use excessive
force when washing the bar. Otherwise it may bend.

Pinch acid or alkaline detergent-soaked (ethanol-soaked also applicable)


gauze with tweezers and gently clean the exterior of the mixing bar until it
is clean and smooth.

NOTE:
Mindray recommends the acid and alkaline detergents be
used alternately for this purpose. For instance, if the acid
detergent was used for last cleaning, the alkaline detergent
would better be used for this time.

5-15
Maintenance

4 Pinch deionized water-soaked gauze to wash the mixing bar.


5 After washing, gently pull the bar arm to its highest point and rotate the bar
arm to move the bar to a position above the wash well.

CAUTION:
After washing the bar, be sure to rotate the bar to a position
above the wash well.

6 Place the ANALYZING UNIT POWER to ON and the system will


automatically reset the mixing bar and wash it with deionized water.

CAUTION:
The mixing bar is precisely fabricated. Replace the scratched or bent
bar once finding it. Refer to 5.6.5 Replacing Mixing Bar for details. .

5.3.4 Washing Deionized Water Tank

CAUTION:
The deionized water to be used on the analyzer must meet the CAP
Type II water requirements and the resistance rate must be no less
than 0.5Mcm.

1 Place the ANALYZING UNIT POWER to OFF.


2 Open the lower door of the analyzing unit and you will see the deionized
water tank on the right.
3 Pull out the extractable plate.
4
CAUTION:
After removing the cap of the deionized water tank (together
with the pickup tube and level sensor), place it on a clean
table.

Unscrew (counter-clockwise) the cap (together with the deionized water


pickup tube and the level sensor).
5 Take out the deionized water tank.
6 Wash the tank interior with deionized water. Use a clean brush to clean the
interior if necessary.
7 Wash the pickup tube and the level sensor with deionized water. Use
clean gauze to wash them if necessary.
8 Wipe water off the tank exterior, pickup tube and sensor cable with clean
gauze.

5-16
Maintenance

9 Place the deionized water tank on the extractable plate.


10 Add deionized water into the tank.
11 Screw (clockwise) the cap (together with the pickup tube and level sensor)
back onto the tank until secure.
12 Push the extractable plate back in.
13 Close the lower door.

CAUTION:
Ensure the deionized water pickup tube is neither blocked nor bent.

5.3.5 Washing Waste Tank

BIOHAZARD:
To prevent biohazard contamination, wear gloves, goggles and
protective clothing when washing the tank.
Exercise caution and do not spill the waste onto other people or things.
Dispose of the waste in accordance with your local or national
guidelines for biohazard waste disposal and consult the manufacturer
or distributor of the reagents for details.
Dispose of used gauze in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place the ANALYZING UNIT POWER to OFF.


2 Open the lower door of the analyzing unit and you will see the waste tank
on the left.
3 Pull out the extractable plate.
4
BIOHAZARD:
After removing the cap of the waste tank (together with the
tube and level sensor), place it on an appropriate place to
avoid biohazard contamination.

Unscrew (counter-clockwise) the cap (together with the waste tube and the
level sensor).
5 Take out the waste tank.
6 Empty the waste tank.
7 Wash the tank interior with clean water. Soak the tank with disinfector if
necessary.
8 Wash the waste tube and the level sensor with clean water.

5-17
Maintenance

9 Wipe water off the tank exterior, waste tube and sensor cable with clean
gauze.
10 Place the deionized water tank back onto the extractable plate.
11 Screw (clockwise) the cap (together with the waste tube and level sensor)
back onto the tank until secure.
12 Push the extractable plate back in.
13 Close the lower door.

CAUTION:
Ensure the waste tube is neither blocked nor bent. A blocked or bent
waste tube will lead to waste overflow that may damage the analyzing
unit.

5.3.6 Cleaning Sample Disk/Compartment

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the
sample disk/compartment.
Dispose of used gauze in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place the ANALYZING UNIT POWER to OFF.


2 Remove the cover from the sample disk. Shift the disk handle from
horizontal position to vertical position. Grab the handle or handwheel and
pull the disk upward to remove it.

3 Wash the disk with clean water and wipe it dry with clean gauze.

5-18
Maintenance

4 Use clean gauze (water or disinfector-dipped gauze if necessary) to clean


the inside of the compartment.

5 If the bar code scanner(optional) is connected, use ethanol-dipped gauze


to clean the window of the bar code scanner.

6 Keep the sample disk handle in the vertical position. Align the central hole
of the handwheel to the rotor and lower the sample disk all the way down.
Rotate the disk a bit as necessary during the lowering process. When the
disk is in place, shift the handle back to horizontal position to lock the disk
to the rotor. To ensure the disk is secured, you may grab the handwheel
and gently pull it upward. If the disk does not loose, it means it is secured
and you can replace the cover; otherwise, re-install the disk.

5.3.7 Cleaning Reagent Disk/Compartment

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the
reagent disk/compartment.
Dispose of used gauze in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place the ANALYZING UNIT POWER to OFF.

5-19
Maintenance

2 Remove the cover from the reagent disk. Shift the disk handle from
horizontal position to vertical position. Grab the handle or handwheel and
pull the disk upward to remove it.

3 Wash the disk with clean water and wipe it dry with clean gauze.
4 Use clean gauze (water or disinfector-dipped gauze if necessary) to clean
the inside of the compartment.

5 Keep the reagent disk handle in the vertical position. Align the central hole
of the handwheel to the rotor and lower the reagent disk all the way down.
Rotate the disk a bit as necessary during the lowering process. When the
disk is in place, shift the handle back to horizontal position to lock the disk
to the rotor. To ensure the disk is secured, you may grab the handwheel
and gently pull it upward. If the disk does not loose, it means it is secured
and you can replace the cover; otherwise, re-install the disk.

5.3.8 Cleaning Panel of Analyzing Unit

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the panel
of analyzing unit.
Dispose of used gauze in accordance with your local or national
guidelines for biohazard waste disposal.

5-20
Maintenance

1 Place the ANALYZING UNIT POWER to OFF.


2 Wipe the panel of the analyzing unit with clean gauze (water or
disinfector-dipped gauze if necessary).

5.4 Monthly Maintenance

5.4.1 Cleaning Wash Well of Sample Probe

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the wash
well.
Dispose of used cotton sticks in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place the ANALYZING UNIT POWER to OFF.


2 Pull the sample probe arm to its highest point. Rotate the arm away from
the wash well.
3 Clean the inside of and the place around the wash well with cotton sticks.

4 Rotate the arm back to a position above the wash well.

5.4.2 Cleaning Wash Well of Reagent Probe

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

5-21
Maintenance

BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the wash
well.
Dispose of used cotton sticks in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place the ANALYZING UNIT POWER to OFF.


2 Pull the reagent probe arm to its highest point. Rotate the arm away from
the wash well.
3 Clean the inside of and the place around the wash well with cotton sticks.

4 Rotate the arm back to a position above the wash well.

5.4.3 Cleaning Wash Well of Mixing Bar

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the wash
well.
Dispose of used cotton sticks in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place the ANALYZING UNIT POWER to OFF.


2 Pull the mixing bar arm to its highest point. Rotate the arm away from the
wash well.

5-22
Maintenance

3 Clean the inside of and the place around the wash well with cotton sticks.

4 Rotate the arm back to a position above the wash well.

5.4.4 Cleaning Sample Probe Rotor

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the rotor.

1 Place the ANALYZING UNIT POWER to OFF.


2 Wipe the rotor with clean gauze.

5.4.5 Cleaning Reagent Probe Rotor

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

5-23
Maintenance

BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the rotor.

1 Place the ANALYZING UNIT POWER to OFF.


2 Wipe the rotor with clean gauze.

5.4.6 Cleaning Mixing Bar Rotor

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the rotor.

1 Place the ANALYZING UNIT POWER to OFF.


2 Wipe the rotor with clean gauze.

5-24
Maintenance

5.5 Maintenance Every Six Months

Washing Dust Screens

1 Place the POWER to OFF.


2 Open the middle door.
3 Use a screwdriver to unscrew the screws fixing the dust screens on the
middle door and remove the screens.
4 Use a screwdriver to unscrew the three screws fixing the back cover and
open the back cover.

5 Use a screwdriver to unscrew the four screws fixing the screens on the
back cover and remove the screens.
6 Wash the screens with clean water.
7 Wipe the screens dry with clean gauze.
8 Install the screens back to the back cover and the middle door with
screws.
9 Close the middle door.
10 Close the back cover and fix it with three screws.

5.6 Irregular Maintenance

5.6.1 Unclogging Sample Probe


When the sample probe is clogged, the fluid flow will become abnormal. Follow the
steps given below to remove, unclog and install the sample probe.

5-25
Maintenance

5.6.1.1 Removing Sample Probe

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when removing the
probe.

1 Place the ANALYZING UNIT POWER to OFF.


2 Remove the cover from the sample disk. Shift the disk handle from the
horizontal position to vertical position. Grab the handle or handwheel and
pull the disk upward to remove it.

3 Pull the sample probe arm to its highest point. Rotate the probe to a
position above the sample disk and convenient to operate.

5-26
Maintenance

4 Grab the lower part of the arm cover with two hands and pull them slightly
outwards and remove the cover upward from the arm base.

5 Hold the probes fluid connector of probe with one hand and the tubing
connector with the other. Rotate tubing connector counter-clockwise until it
disconnects from the probe. Remove the tubing from the probe.

5-27
Maintenance

6 Press the circuit board with one hand and disconnect the probes circuit
connector from the board with the other hand.

CAUTION:
The probe arm is delicate. Exercise caution when
disconnecting the connector. Excessive force may damage
the connector and/or the circuit board.

7 Use a screwdriver to remove the retaining screw and take out the spring.

5-28
Maintenance

8
WARNING:
Save the removed probe at a safe place where it will neither
endanger people working around the area nor be damaged.

NOTE:
Exercise caution when pulling the probe away from the arm
so that the probe tip will not contact or even damage the
probe arm.

Slowly pull the probe away from the probe arm. Exercise caution so that
the gasket inside the probe will not drop out and if it does drop out, save it
at a clean place for later installation.

NOTE:
The probe tip is precisely fabricated. A bent or damaged probe will lead
to unreliable test results and should be replaced immediately upon
seeing. Refer to 5.6.3 Replacing Sample Probe for details.

5.6.1.2 Unclogging Sample Probe

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when unclogging the
probe.
Dispose of the used needle in accordance with your local or national
guidelines for biohazard waste disposal.

1 Use a needle to unclog the probe from the tip.

CAUTION:
The probe tip is precisely fabricated. A bent or damaged probe will lead
to unreliable test results and should be replaced immediately upon
seeing. Refer to 5.6.3 Replacing Sample Probe for details.

5-29
Maintenance

5.6.1.3 Installing Sample Probe

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when installing the probe.

1 Place the ANALYZING UNIT POWER to OFF.


2 Insert the probe back into the arm.
3 Replace the spring and screw the retaining screw back onto the probe.
4 Pinch the probe by the part near the probe arm. Gently push the probe
upward and then release the probe to see if it can move freely. If so, proceed
to the next step; if not, check for errors and try again after removing the
errors.
5 Connect the probes circuit connector back to the circuit board.
6 Ensure the gasket is inside the probe.
7 Screw the probes fluid connector back to the tubing connector.

CAUTION:
Exercise caution when connecting the probe. Excessive
force may bend the probe.

8 Add deionized water to a clean cup. Immerse the probe tip into the water by
2-3mm and the LED on the circuit board should be lighted. Take the probe tip
out of water, and the LED should go out. If the test succeeds, proceed to the
next step; if not, please contact Mindray Customer Service Department or
the distributor.

5-30
Maintenance

9 Check the marks inside the probe arm cover to see the orientation of the
cover. Install the cover back to the probe arm.

CAUTION:
The marks inside the probe arm cover are shown in the figure
below.

10 Pinch the probe by the part near the probe arm. Gently push the probe
upward and then release the probe to see if it can move freely. If so, proceed
to the next step; if not, the cover is not properly installed. Remove the cover
and re-install it and try this step again.

11 Pull the sample probe arm to its highest point. Rotate the probe to a position
above the wash well.

CAUTION:
After installing the sample probe, be sure to rotate the probe
to a position above the wash well.

12 Keep the sample disk handle in the vertical position. Align the central hole of
the handwheel to the rotor and lower the sample disk all the way down.
Rotate the disk a bit as necessary during the lowering process. When the
disk is in place, shift the handle back to horizontal position to lock the disk to
the rotor. To ensure the disk is secured, you may grab the handwheel and
gently pull it upward. If the disk does not loose, it means it is secured and
you can replace the cover; otherwise, re-install the disk.
13 Place the ANALYZING UNIT POWER to ON. The system will automatically
reset the probe and wash it with deionized water.

5-31
Maintenance

CAUTION:
The probe tip is precisely fabricated. A bent or damaged probe will lead
to unreliable test results and should be replaced immediately upon
seeing.

5.6.2 Unclogging Reagent Probe


When the reagent probe is clogged, the fluid flow will become abnormal. Follow the
steps given below to remove, unclog and install the reagent probe.

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when removing the
probe.

5.6.2.1 Removing Reagent Probe

1 Place the ANALYZING UNIT POWER to OFF.


2 Remove the cover from the reagent disk. Shift the disk handle from the
horizontal position to vertical position. Grab the handle or handwheel and
pull the disk upward to remove it.

5-32
Maintenance

3 Pull the reagent probe arm to its highest point. Rotate the probe to a
position above the reagent disk and convenient to operate.

4 Grab the lower part of the arm cover with two hands and pull them slightly
outwards and remove the cover upward from the arm base.

5 Hold the probes fluid connector with one hand and the tubing connector
with the other. Rotate tubing connector counter-clockwise until it
disconnects from the probe. Remove the tubing from the probe.

5-33
Maintenance

6 Press the circuit board with one hand and disconnect the probes circuit
connector from the board with the other hand.

CAUTION:
The probe arm is delicate. Exercise caution when
disconnecting the connector. Excessive force may damage
the connector and/or the circuit board.

7 Use a screwdriver to remove the retaining screw and take out the spring.

8
WARNING:
Save the removed probe at a safe place where it will neither
endanger people working around the area nor be damaged.

NOTE:
Exercise caution when pulling the probe away from the arm
so that the probe tip will not contact or even damage the
probe arm.

Slowly pull the probe away from the probe arm. Exercise caution so that
the gasket inside the probe will not drop out and if it does drop out, save it
at a clean place for later installation.

5-34
Maintenance

NOTE:
The probe tip is precisely fabricated. A bent or damaged probe will lead
to unreliable test results and should be replaced immediately upon
seeing. Refer to 5.6.4 Replacing Reagent Probe for details.

5.6.2.2 Unclogging Reagent Probe

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when unclogging the
probe.
Dispose of the used needle in accordance with your local or national
guidelines for biohazard waste disposal.

1 Use a needle to unclog the probe from the tip.

NOTE:
The probe tip is precisely fabricated. A bent or damaged probe will lead
to unreliable test results and should be replaced immediately upon
seeing. Refer to 5.6.4 Replacing Reagent Probe for details.

5.6.2.3 Installing Reagent Probe

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when installing the probe.

1 Place the ANALYZING UNIT POWER to OFF.


2 Insert the probe back into the arm.
3 Replace the spring and screw the retaining screw back onto the probe.

5-35
Maintenance

4 Pinch the probe by the part near the probe arm. Gently push the probe
upward and then release the probe to see if it can move freely.
If so, proceed to the next step.
If not, check for errors and try again after removing the errors.
5 Connect the probes circuit connector back to the circuit board.
6 Ensure the gasket is inside the probe.
7
CAUTION:
You must shape the tubing inside the reagent probe arm into
a ring.

Exercise caution when connecting the probe. Excessive


force may bend the probe.

Screw the probes fluid connector back to the tubing connector.


8 Add deionized water to a clean cup. Immerse the probe tip into the water by
2-3mm and the LED on the circuit board will be lighted. Take the probe tip
out of water, and the LED will go out.
If the test succeeds, proceed to the next step.
If not, please contact Mindray Customer Service Department or the
distributor.

5-36
Maintenance

9 Check the marks inside the probe arm cover to see the orientation of the
cover. Install the cover back to the probe arm.

CAUTION:
The marks inside the probe arm cover are shown in the figure
below.

10 Pinch the probe by the part near the probe arm. Gently push the probe
upward and then release the probe to see if it can move freely.
If so, proceed to the next step.
If not, the cover is not properly installed. Remove the cover and re-install it
and try this step again.

11 Pull the reagent probe arm to its highest point. Rotate the probe to a position
above the wash well.

CAUTION:
After installing the reagent probe, be sure to rotate the probe
to a position above the wash well.

12 Keep the reagent disk handle in the vertical position. Align the central hole of
the handwheel to the rotor and lower the reagent disk all the way down.
Rotate the disk a bit as necessary during the lowering process. When the
disk is in place, shift the handle back to horizontal position to lock the disk to
the rotor. To ensure the disk is secured, you may grab the handwheel and
gently pull it upward. If the disk does not loose, it means it is secured and
you can replace the cover; otherwise, re-install the disk.
13 Place the ANALYZING UNIT POWER to ON. The system will automatically
reset the probe and wash it with deionized water.

5-37
Maintenance

NOTE:
The probe tip is precisely fabricated. A bent or damaged probe will lead
to unreliable test results and should be replaced immediately upon
seeing. Refer to 5.6.4 Replacing Reagent Probe for details.

5.6.3 Replacing Sample Probe


If the sample probe is bent or damaged, it must be replaced immediately. Follow the
procedure given below to replace the damaged or bent sample probe.

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when replacing the
probe.

CAUTION:
Please use consumables recommended by our company. Other
consumables may degrade the system performance.

1 Remove the bent or damaged probe as instructed by 5.6.1.1 Removing


Sample Probe.

BIOHAZARD:
Dispose of the bent or damaged sample probe in accordance
with your local or national guidelines for biohazard waste
disposal.

2 Install a new probe as instructed by 5.6.1.3 Installing Sample Probe.

CAUTION:
After installing the sample probe, be sure to rotate the probe
to a position above the wash well before installing the
sample disk.

5.6.4 Replacing Reagent Probe


If the reagent probe is bent or damaged, it must be replaced immediately. Follow the
procedure given below to replace the damaged or bent reagent probe.

5-38
Maintenance

WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.

BIOHAZARD:
Wear gloves, goggles and protective clothing when replacing the
probe.

CAUTION:
Please use consumables recommended by Mindray. Other
consumables may degrade the system performance.

1 Remove the bent or damaged probe as instructed by 5.6.2.1 Removing


Reagent Probe.

BIOHAZARD:
Dispose of the bent or damaged reagent probe in
accordance with your local or national guidelines for
biohazard waste disposal.

2 Install a new probe as instructed by 5.6.2.3 Installing Reagent Probe.

CAUTION:
You must shape the tubing inside the reagent probe arm into
a ring.

After installing the reagent probe, be sure to rotate the probe


to a position above the wash well before installing the
reagent disk.

5.6.5 Replacing Mixing Bar


If the mixing bar is damaged, it must be replaced immediately. Follow the procedure
given below to replace the damaged mixing bar.

5-39
Maintenance

WARNING:
The sample probe tip is sharp and can cause puncture wounds. To
prevent injury to your hands or arms, exercise caution when working
around the probe.
When replacing the bar, pinch the bar only by the knurled part and do
not touch the other part of the bar. Protect the flat par of the mixing bar
from scratches.

BIOHAZARD:
Wear gloves, goggles and protective clothing when replacing the
mixing bar.
Dispose of the damaged mixing bar in accordance with your local or
national guidelines for biohazard waste disposal.

CAUTION:
Please use consumables recommended by Mindray. Other
consumables may degrade the system performance.

1 Place the ANALYZING UNIT POWER to OFF.


2 Prepare a new mixing bar. Wash the flat part of the new bar with
detergent-dipped gauze or cotton sticks and then wipe it dry with
deionized water-dipped gauze.
3 Gently pull the bar to its highest point and rotate it to a position convenient
to operate.

5-40
Maintenance

4 Pinch the bar by the knurled part with one hand and unscrew
(counter-clockwise) the retaining nut with the other until the mixing bar
looses. Pull the bar downward to remove it and remove the nut.

CAUTION:
When trying to pull out the bar, concentrate your force in the
direction of the axis on the bar arm. Biased force may
damage the bar and/or the axis.

5 Align the mixing bar to the bigger hole of the retaining nut and gently
screw it into the nut until the end of the bar is in line with the smaller hole
of the nut.
6 Pinch the mixing bar by the knurled part and align the hole of the nut to the
axis of the mixing bar and push the nut onto the bar until it reaches the
end of the bar. Tighten the nut (screw clock-wise) with the other hand.

CAUTION:
When trying to push the bar into the arm, concentrate your
force in the direction of the axis on the bar arm. Biased force
may damage the bar and/or the axis.
Ensure the bar is all the way pushed to the end.

5-41
Maintenance

7 After replacing the bar, visually check whether the bar is vertical to the bar
arm.
If yes, proceed to the next step.
If not, repeat step 5 to remove the bar and re-install it.
8 Pull the bar arm to its highest point and rotate it back to a position above
its wash well.

CAUTION:
After installing the mixing bar, be sure to rotate the mixing
bar to a position above its wash well before turning on the
analyzing unit.

9 Place the ANALYZING UNIT POWER to ON. The system will automatically
reset the bar and wash it with deionized water.
10 Follow the steps given below to do the position checks:
Enter the System Maintenance screen and select the Mixing Bar tab as
instructed by 4.4.6.3 Mixing Bar tab.

Click Into Wash Well and observe whether the mixing bar is aligned to the
center of the wash well. If not, please contact Mindray Customer Service
Department or the distributor.
Set the Time to 100 and click Mix. The bar should start to rotate after
enters the reaction cuvette. Observe whether the rotation is normal and
pay attention to abnormal sound. If the rotation is not normal or the
abnormal noise is heart, please contact Mindray Customer Service
Department or the distributor.

5-42
Maintenance

5.6.6 Replacing Plunger Assembly of Sample/Reagent Syringe


The plungers of sample and reagent syringes may be worn out after a certain period
of service. A worn-out plunger may lead to leaks, which will consequently result in
inaccurate aspiration and unreliable test results. You should check the syringes
everyday and replace the old plunger assembly with a new one when

The old one has served for three months; or


The old one has been used for over 100,000 tests; or
The old one is apparently damaged.

CAUTION:
Please use consumables recommended by Mindray. Other
consumables may degrade the system performance.
Exercise caution when installing the syringe. Excessive force may
crack the syringe.

Follow the procedure given below to replace the plunger assembly of sample or
reagent syringe.

1 Place the ANALYZING UNIT POWER to OFF.


2 Open the middle door and you will see both of the syringes, the sample
syringe on the right and the reagent syringe on the left.

T-piece

Syringe Connector

Upper Retaining Screw

Space Bar

Syringe

Plunger Guide Cap

Holder
Plunger Button

Lower Retaining Screw

5-43
Maintenance

3 Prepare a new plunger assembly (shown in the figure below) and soak the
plunger tip in deionized water to eliminate bubbles.

Plunger Tip Plunger Rod Plunger Button


Plunger Guide Cap

4 Unscrew (counter-clockwise) the two upper retaining screws of the syringe


and remove the screws and the space bar.
5 Unscrew (counter-clockwise) the lower retaining screw of the syringe and
remove the syringe from the holder.
6
CAUTION:
There may be residual water in the syringe. Exercise caution
when removing it. Do not drop water onto the analyzing unit.

Grab the T-piece with one hand and the syringe connector with the other
hand and unscrew (counter-clockwise) the syringe. Exercise caution so that
the gasket on the syringe does not drop out and if it does, store it in a clean
place for later installation. Replace the gasket if it has been disassembled
for 2 to 3 times. Otherwise leakage may occur or sampling precision be
affected.
7
CAUTION:
There may be residual water in the syringe. Exercise caution
when pulling the plunger assembly out. Do not drop water onto
the analyzing unit.
The plunger rod of the sample syringe is slender. Exercise
caution when working on it. Excessive force may bend it.

Unscrew (counter-clockwise) the plunger guide cap and pinch the plunger
button to gently pull the plunger assembly from the syringe.

Syringe Connector Syringe Plunger Guide Plunger Rod Plunger Button


Cap

8
CAUTION:
The plunger rod of the sample syringe is slender. Exercise
caution when working on it. Excessive force may bend it.

Pinch the new plunger assembly by the plunger button and carefully insert
the plunger tip into the syringe and push it all the way to the end. Screw
(clockwise) the plunger guide cap until secure.

5-44
Maintenance

9 Immerse the syringe connector into deionized water. Pinch the plunger
button, pull it to aspirate half syringe of deionized water and then push it to
expel the deionized water and the air from the syringe.
10 Grab the T-piece with one hand and the syringe connector with the other
hand. Screw (clockwise) the syringe into the T-piece until secure.
11 Place the syringe on the holder.
12 Install the space bar and two upper retaining screws. Note that do not
tighten the screws now.
13 Align the plunger button to the lower retaining screw of the plunger and
screw (clockwise) the screw until secure.
14 Pinch the plunger guide cap to adjust the syringe. Our company
recommends the upper edge of the space bar reach the fourth line (the 4l
line of the sample reagent, or the 40l line of the reagent syringe) of the
scale on the syringe.
15 Tighten the two upper retaining screws.
16 Place the ANALYZING UNIT POWER back to ON.
17 Enter the System Maintenance screen and set the Volume (SS. Syringe)
or Volume (R. Syringe) to the maximum value(45 for the sample syringe
and 450 for the reagent syringe ). Click S. Syringe Aspirate or R. Syringe
Aspirate and the syringe should start aspirating. After the syringe finishes
the motion, click S. Syringe Dispense or R. Syringe Dispense and the
syringe should start expelling. You may repeat this action several times.
Refer to 4.4.6.7 Fluid Path tab for details. Pay attention to bubbles during
the aspiration/dispensing process.
Go to the next step if no bubble is found.
If there are bubbles observed during the process, they may be caused by
the air leak between the syringe and the T-piece. Uninstall the syringe and
re-install it. If the bubbles are found again, please contact Mindray
Customer Service Department or the distributor.
18 Close the middle door.

5.6.7 Removing Air Bubbles


When you see air bubbles in the syringe, follow this procedure to remove them.

BIOHAZARD
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the maintenance.
Dispose of the waste in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place the Power to OFF.


2 Loosen the screws on the syringe cover and remove the cover. You can
see the two reagent syringes on the left and sample syringe on the right.
3 Loosen the four upper retaining screws of the syringe, then remove the
screws and space bar.

5-45
Maintenance

4 Loosen the four lower retaining screws of the syringe and remove the
syringe from the holder.
5 Pull the plunger gently outwards until you can not proceed any more, and
then push it quickly. Repeat this pull-push operation until the air bubbles
are removed from the syringe.

CAUTION:
Be sure not to push the plunger to the end tip; otherwise the
syringe may be damaged.

6 Place the syringe on the holder. Install the space bars and tighten the
upper retaining screws.

NOTE:
The upper edge of the upper space bar must reach the 7th scale
on the syringe.
When fixing the retaining screws, be sure to tighten them
alternately with equilibrium force.

7 Tighten counterclockwise the lower retaining screw until secure.

5.6.8 Replacing Lamp


The lamp of the photometric system ages after a certain period of service. An aged
lamp may introduce extra noise to the analyzing process. Replace the lamp with a
new one when the system reminds you to do so, or the service time of the lamp has
added up to 1,000 hours.

CAUTION:
Please use consumables recommended by Mindray. Other
consumables may degrade the system performance.

NOTE:
Do not touch either the light entrance of the lamp housing or the lens in
front of the lamp. In case the entrance is dirty, clean it with
ethanol-soaked defatted cotton.

1 Place the POWER to OFF. Wait at least 15 minutes for the lamp and its
housing to cool down.

WARNING:
After working for a while, the lamp and its housing are
usually hot enough to burn you. Do not proceed with this
procedure until they have cooled down.

5-46
Maintenance

2 Unscrew (counter-clockwise) the three retaining screws on the upper part


of the back of the analyzing unit and open the upper part.

3 Unscrew the two lower screws on the back of the lamp housing and
remove the radiator from the housing.

4 Pinch the lamp base with one hand and press the lever on the left of the
base.

5 Pinch the new lamp by its base and insert the lamp pins into the mounting
holes, the larger pin into the larger hole and the smaller pin into the
smaller hole, on the base.
6 Replace the radiator and tighten the screws.

5-47
Maintenance

7 Close the upper part of the back of the analyzing unit and tighten the
screws.

5.6.9 Replacing Electrode (optional)

BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the below maintenance.

CAUTION:
Use the consumables recommended by Mindray. Other consumables
may degrade system performance.

NOTE:
The maintenance is necessary to be performed when the ISE unit
(optional) is connected.

If you run no more than 100 samples requested for the ISE tests a day, replace the
electrodes according to the following recommended schedule:

Li+ Electrode 6 months


+
Na Electrode 6 months
K+ Electrode 6 months
Cl- Electrode 6 months
Reference Electrode 6 months

If you run more than 100 samples requested for the ISE tests a day, replace the
electrodes according to the following recommended schedule:

Li+ Electrode 3,000 samples


+
Na Electrode 10,000 samples
K+ Electrode 10,000 samples
Cl- Electrode 10,000 samples
Reference Electrode 10,000 samples

1 Place the POWER to OFF.


2 Open the ISE unit door.
3 Unscrew the screw of the ISE module manually.

5-48
Maintenance

4 Take the electrode(s) out from the housing by pressing down the
compression plate(s).

NOTE:
Because the electrodes must be installed sequentially, you
have to take out the electrode to be replaced and those (or
that) over it from above to below. For example, there is no
other electrode over the Li electrode, so you just need to
take it out; There are four electrodes over the reference
electrode, so you have to take out all the five electrodes from
above to below.

5 Install the new electrode. Refer to the 2.4.1 Installing Electrodes for
details.

5.6.10 Replacing Reagent Module (optional)

BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the below maintenance.
Dispose of the waste in accordance with your local or national
guidelines for biohazard waste disposal.

CAUTION:
Use the consumables recommended by Mindray. Other consumables
may degrade system performance.

NOTE:
The maintenance is necessary to be performed when the ISE unit
(optional) is connected.

1 Place the POWER to OFF.


2 Open the ISE unit door.
3 Take out the reagent module. Press the yellow button on the wand and
take off the wand from the module. Then put the three red caps back to
the module that were taken off when the module was installed.
4 Install a new reagent module. Refer to 2.4.2 Installing Reagent Module
for details.

5-49
Maintenance

5.6.11 ISE Unit Storage (optional)

BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the below maintenance.
Dispose of the waste in accordance with your local or national
guidelines for biohazard waste disposal.

CAUTION:
Use the consumables recommended by Mindray. Other consumables
may degrade system performance.

NOTE:
The maintenance is necessary to be performed when the ISE unit
(optional) is connected.
The ISE unit (optional) should be on power all the time. In some cases
that the POWER will be shut down for a long time more than half an
hour, the following steps should be performed.

1 Enter the ISE Maintenance screen of the operating software.


2 Click the Clean button and a dialog box pops up to remind you to place
ISE cleaning solution to the position 45 on the sample disk.
3 Load about 200l of the cleaning solution to the position 45 on the sample
disk.
4 Click the OK button.
5 Click the Purge A button and after the instruction is executed, repeat this
step for another two times.
6 Place the POWER to OFF.
7 Take out the reagent module. Press the yellow button on the wand and
take off the wand from the module. Then put the three red caps back to
the module that were taken off when the module was installed.
8 Place the POWER to ON.
9 Enter the System Maintenance screen of the operating software.
10 Click the Download Settings at the Others tab.
11 Select SystemISE UnitISE Unit.
12 Click the Maintenance button.
13 Place the POWER to OFF.
14 Take all the electrodes out from the housing from above to below by
pressing down the compression plates.
15 Put the reference, Na+ and Cl- electrodes into their individual sealed bags.

5-50
Maintenance

16 Aspirate a small volume of Calibrant A from the port of the reagent module
with a syringe and inject it into the lumens of the K+ and Li+ electrodes until
the lumens are full.
Cover both ends of the lumens with tapes to prevent the Calibrant A flows
from the lumens.
Put the K+ and Li+ electrodes into their individual sealed bags.
17 Discard the reagent module.

5.6.12 ISE Unit Re-activation (optional)

BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the below maintenance.

CAUTION:
Use the consumables recommended by Mindray. Other consumables
may degrade system performance.
After completing the following steps, ensure to calibrate the ISE tests
before testing.

NOTE:
Before reusing the ISE unit (optional) that has been stored, you must
complete the following steps.

1 Take out all the electrodes from the sealed bags.


2 Remove the tapes from the K+ and Li+ electrode.
3 If necessary, soak the reference electrode in warm water until the lumen of
the electrode has been cleared of salt build-up.
4 Install electrodes into the housing sequentially. Refer to the 2.4.1
Installing Electrodes for details.
5 Install a new reagent module to the ISE unit. Refer to 2.4.2 Installing
Reagent Module for details.

5.7 Maintenance Log


See the following tables for the parts to be maintained and the maintenance
schedules. Please copy them every month and place a check mark in each day
column for the maintenance items in the list after performing maintenance.

5-51
Maintenance

Daily Maintenance Month: Year:


Maintenance Record
Daily Maintenance
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
1 Checking Sample Syringe
2 Checking Reagent Syringe
3 Checking/Washing Sample Probe
4 Checking/Washing Reagent Probe
5 Checking/Washing Mixing Bar
6 Checking Detergents
7 Checking Connection of Deionized Water
8 Checking Connection of Wastewater
9 Checking Remaining Deionized Water
10 Emptying Waste Tank
11 Emptying Used-Cuvettes Bucket
12 Cleaning ISE Unit (optional)
13 Pump Calibration (optional)

Weekly Maintenance Month: Year:


Maintenance Record
Weekly Maintenance
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
1 Washing Sample Probe
2 Washing Reagent Probe
3 Washing Mixing Bar
4 Washing Deionized Water Tank
5 Washing Waste Tank
6 Cleaning Sample Disk/Compartment
7 Cleaning Reagent Disk/Compartment
8 Cleaning Panel of Analyzing Unit

5-52
Maintenance

Monthly Maintenance Month: Year:


Maintenance Record
Monthly Maintenance
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
1 Cleaning Wash Well of Sample Probe
2 Cleaning Wash Well of Reagent Probe
3 Cleaning Wash Well of Mixing Bar
4 Cleaning Sample Probe Rotor
5 Cleaning Reagent Probe Rotor
6 Cleaning Mixing Bar Rotor

Maintenance Every Six Months Month: Year:


Maintenance Record
Maintenance Every Six Months
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
1 Washing Dust Screens

Irregular Maintenance Month: Year:


Maintenance Record
Irregular Maintenance
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
1 Unclogging Sample Probe
2 Unclogging Reagent Probe
3 Replacing Sample Probe
4 Replacing Reagent Probe
5 Replacing Mixing Bar
6 Replacing Plunger Assembly
7 Removing Air Bubbles
8 Replacing Lamp
9 Replacing Electrode (optional)
10 Replacing Reagent Module (optional)
11 ISE Unit Storage (optional)
12 ISE Unit Re-activation (optional)

5-53
6 Troubleshooting
This chapter presents all error messages and recommended measures, which
should be taken in time once any error occurs.

If the recommended measures fail to solve the problems, contact Mindray or your
local distributor.

When an error occurs, its corresponding error message will be displayed on the
lower left side of the screen, as Figure 6-1 shows.

The log will record the time, level, code and message of every error (except for the
level-11 error) to facilitate you to troubleshoot the analyzer. Refer to 4.4.5 Log for
details about the log.

Figure 6-1 Error messages are displayed on the lower left side of the screen

6.1 Classification of Error Messages


The error messages are classified as various levels according to their severity.

6-1
Troubleshooting

Severity: Warning

Level Definition Actions taken by the analyzer


0 Negligible errors The analyzer will alert the operator without
further treatment in most circumstances
except that sometimes it will invalidate the test
and even pause.
1 Errors to be flagged The analyzer will flag the affected analysis so
that the operator knows something went
wrong during the analysis.
2 Errors to invalidate an The analyzer will invalidate the abnormal
analysis analysis and rerun the test.
20 Errors caused by The analyzer will ignore the related analyses
unavailable samples and proceed to other ones. When other
analyses are finished, the analyzer will pause
so that the operator can load the missing
samples to the sample disk.
Seeing this message, you can also press the

button and load the missing samples


before continuing the analysis.
3 Errors to invalidate The analyzer will invalidate the analysis and
reagents rerun the test.
30 Errors caused by The analyzer will ignore the related analyses
unavailable reagents and proceed to other ones. When other
analyses are finished, the analyzer will pause
so that the operator can load the missing
reagents to the sample disk.
Seeing this message, you can also press the

button and load the missing reagents


before continuing the analysis.
31 Errors to invalidate all /
the analyses related
to the reagent or
sample associated
with the instruction

Severity: Putting certain units on hold

Level Definition Actions taken by the analyzer


4 Errors caused by the The analyzer will pause dispensing the
sample disk assembly samples. All the analyses whose samples
and the sample have already been dispensed will be analyzed
dispenser as programmed. The other analyses will be
invalidated. When all the analyses are
finished, the analyzer will pause.

6-2
Troubleshooting

Level Definition Actions taken by the analyzer


5 Errors caused by the The analyzer will pause dispensing the
reagent disk reagents. All the analyses whose reagents
assembly and the (both the first and second reagents) have
reagent dispenser already been dispensed will be analyzed as
programmed. The other analyses will be
invalidated. When all the analyses are
finished, the analyzer will pause.
6 Errors caused by The analyzer will use all the remaining
cuvette loader only cuvettes and finish all the analyses on the
(reaction disk not reaction disk. When all the analyses are
affected) finished, the analyzer will pause.
7 Errors affecting the The analyzer will put all the sample and
sample disk and reagent dispensing on hold. The analyses
reagent disk only whose samples or reagents have been
(reaction disk not dispensed will be conducted as programmed
affected) and the other analyses will be invalidated.
When the analyses are finished, the analyzer
will pause.
13 Errors caused by The analyzer will pause dispensing R1 right
unavailable deionized now. All the analyses whose R1 has already
water or full waste- been dispensed will be analyzed as
tank programmed. When the analyses whose R1
has already been dispensed have finished,
the analyzer will pause so that operator can
add deionized water or empty the waste tank.

Severity: Stopping the analyzer

Level Definition Actions taken by the analyzer


8 Errors caused by the The analyzer will send the self-test
reaction disk instructions and place the mechanical parts in
assembly their home positions in the next period. If the
analyzer can be reset within the period, the
analyzer will proceed with the remaining
analyses; otherwise, the analyzer will stop.

Severity: Emergency stop

Level Definition Actions taken by the analyzer


9 / The analyzer will not make any movement in
the next period. The error status will be
reserved. All the remaining analyses will be
invalidated. You can operate the analyzer for
other purposes such as printing and reviewing
results.

Severity: Prohibiting analyses

Level Definition Actions taken by the analyzer


10 / The analyzer will prohibit any operation
regarding analyses. But you can use the
control system for such purposes as

6-3
Troubleshooting

Level Definition Actions taken by the analyzer


reviewing data.
12 / Only for the ISE unit (optional). The analyzer
will prohibit ISE tests. But you can run other
tests normally.

Severity: Prohibiting startup

Level Definition Actions taken by the analyzer


11 / The control system cannot be started or is
terminated. The operation unit (computer)
exits the control system to the Windows
operating system.

NOTE:
All errors except for those of level 11 will be recorded in the log. When
a level-11 error occurs, the operating system will display the error
message and exit the operating software after the error is confirmed by
the operator.

6.2 Corrective Measures

6.2.1 Corrective Measures for General Errors


In case of an error, enter the log to check the error code. Based on the error code,
check the table below for recommended corrective measures.

WARNING:
When troubleshooting the analyzer, first find out whether it is
necessary to switch off the POWER or ANALYZING UNIT POWER.
Wear gloves, goggles and lab coat if necessary.

6-4
Troubleshooting

Error code Error Message Level Corrective Measure


100000001 Parity error. 9 Contact Mindray Customer Service Department or your local distributor.

100000002 Instruction too long. 9 Contact Mindray Customer Service Department or your local distributor.
100000003 Incomplete instruction. 9 Contact Mindray Customer Service Department or your local distributor.
100000004 Check sum error. 9 Contact Mindray Customer Service Department or your local distributor.

100000005 Downloading parameter. Can't respond to other 9 Contact Mindray Customer Service Department or your local distributor.
instructions.
100000006 Data at byte 00-0x7F exceed 0x7F. 9 Contact Mindray Customer Service Department or your local distributor.

100960100 Wrong command. Illegal frame type. 9 Contact Mindray Customer Service Department or your local distributor.

100960101 Wrong command. No this unit's command. 9 Contact Mindray Customer Service Department or your local distributor.
100960102 Wrong command. Illegal command identity. 9 Contact Mindray Customer Service Department or your local distributor.

100960103 Wrong command. Parameter significant bits wrong 9 Contact Mindray Customer Service Department or your local distributor.

100960104 Wrong command. Instruction length wrong. 9 Contact Mindray Customer Service Department or your local distributor.
100960105 Wrong command. Sample position not within 1-60 9 Contact Mindray Customer Service Department or your local distributor.
100960106 Wrong command. Reagent position not within 1-50 9 Contact Mindray Customer Service Department or your local distributor.

100960107 Wrong command. Other parameter error. 9 Contact Mindray Customer Service Department or your local distributor.

100960200 MUI self-test error. 10 Contact Mindray Customer Service Department or your local distributor.

100960300 MUI handshaking with sub-unit. Can't respond to other 9 Contact Mindray Customer Service Department or your local distributor.
instructions.
100960500 MUI calling sub-unit to reset and self-test. Can't 9 Contact Mindray Customer Service Department or your local distributor.
respond to other instructions.

6-5
Troubleshooting

Error code Error Message Level Corrective Measure


100960900 MUI work status not supporting this command. 9 Contact Mindray Customer Service Department or your local distributor.
Self-test.

100960901 MUI work status not supporting this command. Failed. 9 Contact Mindray Customer Service Department or your local distributor.
100960902 MUI work status not supporting this command. 9 Contact Mindray Customer Service Department or your local distributor.
Handshake.
100960903 MUI work status not supporting this command. Reset. 9 Contact Mindray Customer Service Department or your local distributor.
100960904 MUI work status not supporting this command. 9 Contact Mindray Customer Service Department or your local distributor.
Shutdown.
100960905 MUI work status not supporting this command. Free. 9 Contact Mindray Customer Service Department or your local distributor.

100960906 MUI work status not supporting this command. 9 Contact Mindray Customer Service Department or your local distributor.
Mechanical reset.
100961000 MUI busy. Can't respond to current instruction. 9 Contact Mindray Customer Service Department or your local distributor.

100961100 Disassemble instruction wrong. 9 Contact Mindray Customer Service Department or your local distributor.

100961200 Attempted undefined system operation. 9 Contact Mindray Customer Service Department or your local distributor.
100961400 Attempted undefined inquiry. 0 Contact Mindray Customer Service Department or your local distributor.

100961500 Wrong parameter for inquiry. 0 Contact Mindray Customer Service Department or your local distributor.

100961600 Attempted to configure undefined parameter. 0 Contact Mindray Customer Service Department or your local distributor.
100961800 Attempted to configure undefined procedure. 0 Contact Mindray Customer Service Department or your local distributor.

100961900 Wrong parameter for procedure configuration. 0 Contact Mindray Customer Service Department or your local distributor.
100962000 E2PROM R/W wrong. Read error. 10 Contact Mindray Customer Service Department or your local distributor.
100962001 E2PROM R/W wrong. Write error. 10 Contact Mindray Customer Service Department or your local distributor.

6-6
Troubleshooting

Error code Error Message Level Corrective Measure


100962002 E2PROM R/W wrong. Check sum error. 10 Contact Mindray Customer Service Department or your local distributor.

100962003 E2PROM R/W wrong. Download error. 10 Contact Mindray Customer Service Department or your local distributor.
100969000 Execution result not received within specified time. 9 Contact Mindray Customer Service Department or your local distributor.

100969100 Command response wrong or no response 9 Contact Mindray Customer Service Department or your local distributor.

101280100 Wrong command. Illegal frame type. 31 Contact Mindray Customer Service Department or your local distributor.
101280101 Wrong command. Not this unit's instruction. 31 Contact Mindray Customer Service Department or your local distributor.

101280102 Wrong command. Illegal command identity. 31 Contact Mindray Customer Service Department or your local distributor.
101280103 Wrong command. Instruction length wrong. 31 Contact Mindray Customer Service Department or your local distributor.
101280104 Wrong command. Parameter significant bits wrong 31 Contact Mindray Customer Service Department or your local distributor.

101280105 Wrong command. Other parameter wrong. 31 Contact Mindray Customer Service Department or your local distributor.

101280200 Control unit self-test wrong 10 Contact Mindray Customer Service Department or your local distributor.
101280300 Mechanical device reset wrong. 2 Contact Mindray Customer Service Department or your local distributor.

101280400 Control unit not supporting this command. Self-test. 2 Contact Mindray Customer Service Department or your local distributor.

101280401 Control unit not supporting this command. Malfunction. 4 Contact Mindray Customer Service Department or your local distributor.
101280402 Control unit not supporting this command. Wait for 2 Contact Mindray Customer Service Department or your local distributor.
handshake.
101280403 Control unit not supporting this command. Shutdown. 2 Contact Mindray Customer Service Department or your local distributor.
101280404 Control unit not supporting this command. Processing 2 Contact Mindray Customer Service Department or your local distributor.
command.
101280405 Control unit not supporting this command. Free. 2 Contact Mindray Customer Service Department or your local distributor.

6-7
Troubleshooting

Error code Error Message Level Corrective Measure


101280406 Control unit not supporting this command. Mechanical 2 Contact Mindray Customer Service Department or your local distributor.
reset.

101280500 Control unit busy. Can't respond to current instruction. 2 Contact Mindray Customer Service Department or your local distributor.
101280600 Attempted to configure un-defined speed. 2 Contact Mindray Customer Service Department or your local distributor.

101280700 Speed configuration parameter beyond limit or 0 Contact Mindray Customer Service Department or your local distributor.
unreasonable.
101280800 Attempted to configure undefined parameter. 0 Contact Mindray Customer Service Department or your local distributor.
101280900 Parameter value beyond limit or unreasonable. 0 Contact Mindray Customer Service Department or your local distributor.
101281000 Attempted undefined inquiry. 0 Contact Mindray Customer Service Department or your local distributor.

101281100 Inquiry parameter beyond limit or unreasonable. 0 Contact Mindray Customer Service Department or your local distributor.

101281200 Attempted undefined system instruction. 0 Contact Mindray Customer Service Department or your local distributor.
101281300 System instruction parameter beyond limit or 0 Contact Mindray Customer Service Department or your local distributor.
unreasonable.

101281400 Attempted to commission undefined position. 0 Contact Mindray Customer Service Department or your local distributor.
101281500 Commissioning parameter beyond limit or 0 Contact Mindray Customer Service Department or your local distributor.
unreasonable.

101281600 System not commissioned or confirmation not 0 Contact Mindray Customer Service Department or your local distributor.
consistent with instruction.

6-8
Troubleshooting

Error code Error Message Level Corrective Measure


101281701 Tubing control failed. Rinse surface not detected during 9 Turn off the analyzing unit and then
power-on.
(1) Check the sample syringe for leakage as instructed by 5.2.1 Checking
Sample/Reagent Syringes.
(2) Check the sample probe to see if there are drops hanging on the probe
tip. If there are, contact our company or your local distributor.
(3) Check the deionized water tank as instructed by 5.2.8 Checking
Remaining Deionized Water.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101281800 Sample disk rotation failed. Can't reach initial point. 4 Ensure the sample disk rotates well as instructed by 4.4.6.6 S. Disk tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101281801 Sample disk rotation failed. Can't leave initial point. 4 Ensure the sample disk rotates well as instructed by 4.4.6.6 S. Disk tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101281802 Sample disk rotation failed. Can't detect weight disk 4 Ensure the sample disk rotates well as instructed by 4.4.6.6 S. Disk tab.
signal.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101281803 Sample disk rotation failed. Sample probe left in 4 Turn off the analyzing unit and check if the sample tubes are placed too
Sample disk. high on the sample disk.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-9
Troubleshooting

Error code Error Message Level Corrective Measure


101281804 Sample disk rotation failed. Needle not at top point 4 Other errors have higher priority.
during rotation.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101281902 Syringe A/D error. Can't reach initial point. 4 Check the sample syringe as instructed by 4.4.6.7 Fluid Path tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101281903 Syringe A/D error. Can't leave initial point. 4 Check the sample syringe as instructed by 4.4.6.7 Fluid Path tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101282000 Sample probe rotation error. Can't reach initial point. 4 (1) Turn off the analyzing unit and clean the sample disk rotor.
(2) Turn on the analyzing unit and check the sample probe as instructed
by 4.4.6.1 S. Probe tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101282001 Sample probe rotation error. Not at v-position for 4 (1) Turn off the analyzing unit and clean the sample disk rotor.
rotation.
(2) Turn on the analyzing unit and check the sample probe as instructed
by 4.4.6.1 S. Probe tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101282002 Sample probe v-move wrong. Can't reach initial 4 (1) Turn off the analyzing unit. Clean the sample disk rotor.
(Sample disk/wash) point.
(2) Turn on the analyzing unit and check the sample probe as instructed
by 4.4.6.1 S. Probe tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-10
Troubleshooting

Error code Error Message Level Corrective Measure


101282100 Sample probe v-moved wrong. Can't leave initial 4 Check the sample probe as instructed by.
(Sample disk/wash) point.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101282101 Sample probe v-moved wrong. Can't leave initial 4 Check the sample probe as instructed by 4.4.6.1 S. Probe tab.
(Sample disk/wash) point.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101282102 Sample probe v-move wrong. Not at initial point 4 Check the sample probe as instructed by 4.4.6.1 S. Probe tab.
(Sample disk/wash) for d-move start.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101282107 Sample probe v-move wrong. Collision. 31 Turn off the analyzing unit and clear the obstacles from the path of the
vertical movement of the sample probe.
101282108 Sample probe v-move wrong. Collide with transducer 9 Turn off the analyzing unit and pinch the sample probe and push it upward
holder. to check whether the spring can move freely.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101282110 Sample probe v-move wrong. Current position prohibits 4 Check the sample probe as instructed by 4.4.6.1 S. Probe tab.
v-move (Sample disk/wash).
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101282200 Attempted undefined simple operation. 0 Other errors have higher priority.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101282300 Wrong cuvette position parameter in command. 2 Other errors have higher priority.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-11
Troubleshooting

Error code Error Message Level Corrective Measure


101282400 Illegal surface detection. 31 Turn off the analyzing unit.
(1) Check whether the sample tubes are placed too high on the sample
disk.
(2) Check the sample syringe for leakage as instructed by 5.2.1 Checking
Sample/Reagent Syringes.
(3) Check the sample probe to see if there are drops hanging on the probe
tip. If there are, contact our company or your local distributor.
(4) Check whether the sample probe is bent or scratched. Replace the
damaged probe as instructed by 5.6.3 Replacing Sample Probe if
necessary.
(5) Clean the sample probe as instructed by 5.3.1 Washing Sample
Probe.
If this error occurs frequently, contact our company or your local distributor.

101282600 Syringe not initialized. 4 Contact Mindray Customer Service Department or your local distributor.

101289000 No execution result received within specified time. 0 Contact Mindray Customer Service Department or your local distributor.
101289100 Command response wrong or no response. 0 Contact Mindray Customer Service Department or your local distributor.

101289200 Sending analyzed instructions failed. 9 Contact Mindray Customer Service Department or your local distributor.

101440100 Wrong command. Illegal frame type. 31 Contact Mindray Customer Service Department or your local distributor.

101440101 Wrong command. Not this unit's instruction. 31 Contact Mindray Customer Service Department or your local distributor.
101440102 Wrong command. Illegal instruction identity. 31 Contact Mindray Customer Service Department or your local distributor.
101440103 Wrong command. Instruction length wrong. 31 Contact Mindray Customer Service Department or your local distributor.

101440104 Wrong command. Parameter significant bits wrong. 31 Contact Mindray Customer Service Department or your local distributor.

101440105 Wrong command. Other parameter wrong. 31 Contact Mindray Customer Service Department or your local distributor.

6-12
Troubleshooting

Error code Error Message Level Corrective Measure


101440200 Control unit self-test wrong. 10 Contact Mindray Customer Service Department or your local distributor.

101440300 Mechanical device reset wrong 0 Contact Mindray Customer Service Department or your local distributor.
101440400 Control unit not supporting this command. Self-test. 3 Contact Mindray Customer Service Department or your local distributor.

101440401 Control unit not supporting this command. Malfunction. 5 Contact Mindray Customer Service Department or your local distributor.

101440402 Control unit not supporting this command. Wait for 3 Contact Mindray Customer Service Department or your local distributor.
handshake.
101440403 Control unit not supporting this command. Shutdown. 3 Contact Mindray Customer Service Department or your local distributor.

101440404 Control unit not supporting this command. Processing 3 Contact Mindray Customer Service Department or your local distributor.
instruction.
101440405 Control unit not supporting this command. Busy. 3 Contact Mindray Customer Service Department or your local distributor.

101440406 Control unit not supporting this command. Mechanical 3 Contact Mindray Customer Service Department or your local distributor.
reset.

101440500 Control unit busy. Can't respond to current instruction. 3 Contact Mindray Customer Service Department or your local distributor.

101440600 Attempted to configure undefined speed. 0 Contact Mindray Customer Service Department or your local distributor.

101440700 Speed configuration parameter beyond limit or 0 Contact Mindray Customer Service Department or your local distributor.
unreasonable.

101440800 Attempted to configure undefined parameter. 0 Contact Mindray Customer Service Department or your local distributor.
101440900 Parameter configuration beyond limit or unreasonable. 0 Contact Mindray Customer Service Department or your local distributor.
101441000 Attempted undefined inquiry. 0 Contact Mindray Customer Service Department or your local distributor.

101441100 Inquiry parameter beyond limit or unreasonable. 0 Contact Mindray Customer Service Department or your local distributor.
101441200 Attempted undefined system instruction. 0 Contact Mindray Customer Service Department or your local distributor.

6-13
Troubleshooting

Error code Error Message Level Corrective Measure


101441300 System instruction parameter beyond limit or 0 Contact Mindray Customer Service Department or your local distributor.
unreasonable.

101441400 Attempted to commission undefined position. 0 Contact Mindray Customer Service Department or your local distributor.
101441500 Commissioning parameter beyond limit or 0 Contact Mindray Customer Service Department or your local distributor.
unreasonable.
101441600 Not commissioned or confirmation not consistent with 0 Contact Mindray Customer Service Department or your local distributor.
instruction.

101441701 Tubing control failed. Washing solution surface not 9 Turn off the analyzing unit and
detected.
(1) Check whether the deionized water tank is empty and add more
deionized water to the tank as necessary.
(2) Check the sample syringe for leakage as instructed by 5.2.1 Checking
Sample/Reagent Syringes.
(3) Check the sample probe to see if there are drops hanging on the probe
tip. If there are, contact our company or your local distributor.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101441800 Reagent disk rotation failed. Can't reach initial position. 5 Check the reagent disk as instructed by 4.4.6.5 Rgt. Disk tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101441801 Reagent disk rotation failed. Can't leave initial position. 5 Check the reagent disk as instructed by 4.4.6.5 Rgt. Disk tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-14
Troubleshooting

Error code Error Message Level Corrective Measure


101441802 Reagent disk rotation failed. Can't detect weight disk 5 Check the reagent disk as instructed by 4.4.6.5 Rgt. Disk tab.
signal.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101441803 Reagent disk rotation failed. Reagent probe left in 5 Turn off the analyzing unit and check the types of the reagent bottles.
Reagent disk.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101441804 Reagent disk rotation failed. Needle not at highest 5 Other errors have higher priority.
position for disk rotation.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101441900 The aspiration/dispensing volume is wrong. The 31 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab. Check the
reagent volume exceeds the upper limit. reagent syringe as instructed by 4.4.6.7 Fluid Path tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101441902 Syringe A/D error. Can't reach initial position. 5 Check the reagent syringe as instructed by 4.4.6.7 Fluid Path tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101441903 Syringe A/D error. Can't leave initial position. 5 Check the reagent syringe as instructed by 4.4.6.7 Fluid Path tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101442000 Reagent probe rotation error. Can't reach initial 5 (1) Turn off the analyzing unit. Clean the sample disk rotor.
position.
(2) Turn on the analyzing unit and check the reagent probe as instructed
by 4.4.6.2 R. Probe tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-15
Troubleshooting

Error code Error Message Level Corrective Measure


101442001 Reagent probe rotation error. Can't leave initial 5 (1) Turn off the analyzing unit. Clean the sample disk rotor.
position.
(2) Turn on the analyzing unit and check the reagent probe as instructed
by 4.4.6.2 R. Probe tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101442002 Reagent probe rotation error. Not at v-position for 5 (1) Turn off the analyzing unit. Clean the reagent disk rotor and.
h-rotation.
(2) Turn on the analyzing unit and check the reagent probe as instructed
by 4.4.6.2 R. Probe tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101442100 Reagent probe v-move error. Can't reach initial 5 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab.
(Reagent disk/wash) position.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101442101 Reagent probe v-move error. Can't leave initial 5 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab.
(Reagent disk/wash) position.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101442102 Reagent probe v-move error. Not at v-initial (Reagent 5 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab.
disk/wash) position.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101442107 Reagent probe v-move error. Collision. 31 Turn off the analyzing unit and clear the obstacles in the path of the
vertical movement of the reagent probe.

6-16
Troubleshooting

Error code Error Message Level Corrective Measure


101442108 Reagent probe v-move error. Collision transducer hold. 9 Turn off the analyzing unit and pinch the reagent probe and push it upward
to see whether it can move freely.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101442110 Reagent probe v-move error. Current position prohibits 5 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab.
v-move (Reagent disk/wash).
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101442200 Attempted undefined simple operation. 0 Other errors have higher priority.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101442300 Wrong cuvette position parameter in command. 0 Other errors have higher priority.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-17
Troubleshooting

Error code Error Message Level Corrective Measure


101442400 Illegal surface detection. 31 Turn off the analyzing unit and
(1) Check the types of the reagent bottles.
(2) Check the reagent syringe for leakage as instructed by 5.2.1 Checking
Sample/Reagent Syringes.
(3) Check the sample probe to see if there are drops hanging on the probe
tip. If there are, contact our company or your local distributor.
(4) Check whether the sample probe is bent or scratched. Replace the
damaged probe as instructed by 5.6.4 Replacing Reagent Probe if
necessary.
(5) Clean the reagent probe as instructed by 5.3.2 Washing Reagent
Probe.
If this error occurs frequently, contact our company or your local distributor.
101442600 Syringe not initialized (unknown syringe current 5 Contact Mindray Customer Service Department or your local distributor.
position)
101449000 Execution result not received within specified time. 0 Contact Mindray Customer Service Department or your local distributor.

101449100 Command response wrong or no response. 0 Contact Mindray Customer Service Department or your local distributor.

101449200 Sending analyzed instructions failed. 9 Contact Mindray Customer Service Department or your local distributor.
101600100 Wrong command. Illegal frame type 31 Contact Mindray Customer Service Department or your local distributor.

101600101 Wrong command. Not this unit's instruction 31 Contact Mindray Customer Service Department or your local distributor.

101600102 Wrong command. Illegal command identity. 31 Contact Mindray Customer Service Department or your local distributor.
101600103 Wrong command. Instruction length wrong 31 Contact Mindray Customer Service Department or your local distributor.
101600104 Wrong command. Parameter significant bits wrong 31 Contact Mindray Customer Service Department or your local distributor.

101600105 Wrong command. Other parameter wrong 31 Contact Mindray Customer Service Department or your local distributor.

6-18
Troubleshooting

Error code Error Message Level Corrective Measure


101600200 Control unit self-test wrong 10 Contact Mindray Customer Service Department or your local distributor.

101600300 Mechanical device reset wrong 0 Contact Mindray Customer Service Department or your local distributor.
101600400 Control unit not supporting this instruction. Self-test. 6 Contact Mindray Customer Service Department or your local distributor.

101600401 Control unit not supporting this instruction. Malfunction 6 Contact Mindray Customer Service Department or your local distributor.

101600402 Control unit not supporting this instruction. Wait for 6 Contact Mindray Customer Service Department or your local distributor.
handshake.
101600403 Control unit not supporting this instruction. Shutdown. 6 Contact Mindray Customer Service Department or your local distributor.

101600404 Control unit not supporting this instruction. Processing 6 Contact Mindray Customer Service Department or your local distributor.
instruction.
101600405 Control unit not supporting this instruction. Free. 6 Contact Mindray Customer Service Department or your local distributor.

101600406 Control unit not supporting this instruction. Mechanical 6 Contact Mindray Customer Service Department or your local distributor.
reset.

101600500 Control unit busy. Can't respond to current instruction. 6 Contact Mindray Customer Service Department or your local distributor.

101600600 Attempted to configure undefined speed. 0 Contact Mindray Customer Service Department or your local distributor.

101600700 Speed configuration parameter beyond limit or 0 Contact Mindray Customer Service Department or your local distributor.
unreasonable.

101600800 Attempted to configure undefined parameter. 0 Contact Mindray Customer Service Department or your local distributor.
101600900 Parameter value beyond limit or unreasonable. 0 Contact Mindray Customer Service Department or your local distributor.
101601000 Attempted undefined inquiry. 0 Contact Mindray Customer Service Department or your local distributor.

101601100 Inquiry parameter beyond limit or unreasonable. 0 Contact Mindray Customer Service Department or your local distributor.
101601200 Attempted undefined system instruction. 0 Contact Mindray Customer Service Department or your local distributor.

6-19
Troubleshooting

Error code Error Message Level Corrective Measure


101601300 System instruction parameter beyond limit or 0 Contact Mindray Customer Service Department or your local distributor.
unreasonable.

101601400 Attempted to commission undefined position. 0 Contact Mindray Customer Service Department or your local distributor.
101601500 Commissioning parameter beyond limit or 0 Contact Mindray Customer Service Department or your local distributor.
unreasonable.
101601600 System not commissioned or confirmation not 0 Contact Mindray Customer Service Department or your local distributor.
consistent with instruction.

101601700 Mixing bar v-move error. Can't reach initial (wash) 7 (1) Turn off the analyzing unit. Clean the mixing bar rotor.
position.
(2) Turn on the analyzing unit and check the mixing bar as instructed by
4.4.6.3 Mixing Bar tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101601701 Mixing bar v-move error. Can't leave initial (wash) 7 (1) Turn off the analyzing unit. Clean the mixing bar rotor.
position.
(2) Turn on the analyzing unit and check the mixing bar as instructed by
4.4.6.3 Mixing Bar tab .
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101601702 Mixing bar v-move error. Not at v-initial position (wash) 7 (1) Turn off the analyzing unit. Clean the mixing bar rotor.
for d-move.
(2) Turn on the analyzing unit and check the mixing bar as instructed by
4.4.6.3 Mixing Bar tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-20
Troubleshooting

Error code Error Message Level Corrective Measure


101601707 Mixing bar v-move error. Current position prohibits 7 Other errors have higher priorities.
v-move (wash).
Check the mixing bar as instructed by 4.4.6.3 Mixing Bar tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101601800 Mixing bar h-move error. Can't reach initial position. 7 (1) Turn off the analyzing unit. Clean the mixing bar rotor.
(2) Turn on the analyzing unit and check the mixing bar as instructed by
4.4.6.3 Mixing Bar tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101601802 Mixing bar h-move error. Not at h-initial position for 7 (1) Turn off the analyzing unit. Clean the mixing bar rotor.
h-rotation.
(2) Turn on the analyzing unit and check the mixing bar as instructed by
4.4.6.3 Mixing Bar tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101602000 Robotic arm v-move error. Can't reach initial (other) 9 Turn on the analyzing unit. Check the manipulator as instructed by
position 4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101602001 Robotic arm v-move error. Can't leave initial (other) 9 Turn on the analyzing unit and check the manipulator as instructed by
position. 4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-21
Troubleshooting

Error code Error Message Level Corrective Measure


101602100 Robotic arm h-move error. Can't reach initial position. 6 Turn on the analyzing unit and check the manipulator as instructed by
4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101602101 Robotic arm h-move error. Can't leave initial position. 6 Turn on the analyzing unit and check the manipulator as instructed by
4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101602200 Robotic arm hold error. Upper hand can't close when 6 Contact Mindray Customer Service Department or your local distributor.
taking cuvette.
101602201 Robotic arm hold error. Upper hand can't open when 6 Contact Mindray Customer Service Department or your local distributor.
placing cuvette.
101602202 Robotic arm hold error. Lower hand can't close when 6 Contact Mindray Customer Service Department or your local distributor.
taking cuvette.

101602203 Robotic arm hold error. Lower hand can't open when 6 Contact Mindray Customer Service Department or your local distributor.
placing cuvette.

101602204 Robotic arm hold error. Upper hand accidentally open 6 Contact Mindray Customer Service Department or your local distributor.
during movement.
101602205 Robotic arm hold error. Upper hand accidentally closed 6 Contact Mindray Customer Service Department or your local distributor.
during movement.
101602206 Robotic arm hold error. Lower hand accidentally 6 Contact Mindray Customer Service Department or your local distributor.
opened during movement.
101602207 Robotic arm hold error. Lower hand accidentally closed 6 Contact Mindray Customer Service Department or your local distributor.
during movement.

6-22
Troubleshooting

Error code Error Message Level Corrective Measure


101602208 Robotic arm hold error. No cuvette in compartment. 0 Check the remaining cuvettes and add more cuvettes as needed.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101602209 Cuvette compartment not ready. 0 If there is no cuvette left in the cuvette compartment, add more cuvettes to
the compartment as instructed by 1.1.1.7 Reaction Cuvette Loader and

press the button.


If there is cuvette in the compartment, ignore the message.
If this error occurs frequently, contact our company or your local
distributor.

101602210 Cuvette compartment trolley moving. 0 If there is no cuvette left in the cuvette compartment, add more cuvettes to
the compartment as instructed by 1.1.1.7 Reaction Cuvette Loader and

press the button.


If there is cuvette in the compartment, ignore the message.
If this error occurs frequently, contact our company or your local
distributor.

101609000 Execution result not received within specified time. 5 Contact Mindray Customer Service Department or your local distributor.

101609100 Command response wrong or no response. 5 Contact Mindray Customer Service Department or your local distributor.
101609200 Sending analyzed instructions failed. 9 Contact Mindray Customer Service Department or your local distributor.

101760100 Wrong command. Illegal frame type 31 Contact Mindray Customer Service Department or your local distributor.

101760101 Wrong command. Not this Unit's instruction 31 Contact Mindray Customer Service Department or your local distributor.
101760102 Wrong command. Illegal command identity. 31 Contact Mindray Customer Service Department or your local distributor.

6-23
Troubleshooting

Error code Error Message Level Corrective Measure


101760103 Wrong command. Instruction length wrong 31 Contact Mindray Customer Service Department or your local distributor.

101760104 Wrong command. Parameter significant bits wrong 31 Contact Mindray Customer Service Department or your local distributor.
101760105 Wrong command. Other parameter wrong 31 Contact Mindray Customer Service Department or your local distributor.

101760200 Control unit self-test wrong 10 Contact Mindray Customer Service Department or your local distributor.

101760400 Control unit not supporting this instruction. Self-test. 0 Contact Mindray Customer Service Department or your local distributor.
101760401 Control unit not supporting this instruction. Malfunction 0 Contact Mindray Customer Service Department or your local distributor.

101760402 Control unit not supporting this instruction. Wait for 0 Contact Mindray Customer Service Department or your local distributor.
handshake.
101760403 Control unit not supporting this instruction. Shutdown. 0 Contact Mindray Customer Service Department or your local distributor.

101760404 Control unit not supporting this instruction. Processing 0 Contact Mindray Customer Service Department or your local distributor.
instruction.
101760405 Control unit not supporting this instruction. Free. 0 Contact Mindray Customer Service Department or your local distributor.
101760406 Control unit not supporting this instruction. Mechanical 0 Contact Mindray Customer Service Department or your local distributor.
reset.
101760500 Control unit busy. Can't respond to current instruction. 0 Contact Mindray Customer Service Department or your local distributor.

101760600 Attempted undefined inquiry. 0 Contact Mindray Customer Service Department or your local distributor.
101760700 Attempted undefined temperature parameter inquiry. 0 Contact Mindray Customer Service Department or your local distributor.
101761100 Attempted undefined temperature inquiry. 0 Contact Mindray Customer Service Department or your local distributor.

101761200 Attempted target temperature beyond limit or 0 Contact Mindray Customer Service Department or your local distributor.
unreasonable.
101761300 Attempted undefined system instruction. 0 Contact Mindray Customer Service Department or your local distributor.

6-24
Troubleshooting

Error code Error Message Level Corrective Measure


101769000 Execution result not received within specified time. 0 Contact Mindray Customer Service Department or your local distributor.

101769100 Command response wrong or no response. 0 Contact Mindray Customer Service Department or your local distributor.
101769200 Sending analyzed instructions failed. 9 Contact Mindray Customer Service Department or your local distributor.

101920100 Wrong command. Illegal frame type. 31 Contact Mindray Customer Service Department or your local distributor.

101920101 Wrong command. Not this unit's instruction. 31 Contact Mindray Customer Service Department or your local distributor.
101920102 Wrong command. Illegal command identity. 31 Contact Mindray Customer Service Department or your local distributor.

101920103 Wrong command. Instruction length wrong. 31 Contact Mindray Customer Service Department or your local distributor.
101920104 Wrong command. Parameter significant bits wrong. 31 Contact Mindray Customer Service Department or your local distributor.
101920105 Wrong command. Other parameter wrong. 31 Contact Mindray Customer Service Department or your local distributor.

101920200 Control unit self-test wrong. 10 Contact Mindray Customer Service Department or your local distributor.

101920300 Mechanical device reset wrong. 0 Contact Mindray Customer Service Department or your local distributor.
101920400 Control unit not supporting this instruction. Self-test. 8 Contact Mindray Customer Service Department or your local distributor.

101920401 Control unit not supporting this instruction. Malfunction. 9 Contact Mindray Customer Service Department or your local distributor.

101920402 Control unit not supporting this instruction. Wait for 8 Contact Mindray Customer Service Department or your local distributor.
handshake.

101920403 Control unit not supporting this instruction. Shutdown. 8 Contact Mindray Customer Service Department or your local distributor.
101920500 Control unit busy. Can't respond to current instruction. 8 Contact Mindray Customer Service Department or your local distributor.
101920600 Attempted undefined speed. 0 Contact Mindray Customer Service Department or your local distributor.

101920700 Speed configuration beyond limit or unreasonable. 0 Contact Mindray Customer Service Department or your local distributor.

101920800 Attempted to configure undefined parameter. 0 Contact Mindray Customer Service Department or your local distributor.

6-25
Troubleshooting

Error code Error Message Level Corrective Measure


101920900 Parameter configuration beyond limit or unreasonable. 0 Contact Mindray Customer Service Department or your local distributor.

101921000 Attempted undefined inquiry. 0 Contact Mindray Customer Service Department or your local distributor.
101921100 Inquiry parameter beyond limit or unreasonable. 0 Contact Mindray Customer Service Department or your local distributor.

101921200 Attempted undefined system instruction. 0 Contact Mindray Customer Service Department or your local distributor.

101921300 System instruction parameter beyond limit or 0 Contact Mindray Customer Service Department or your local distributor.
unreasonable.
101921400 Re-disk rotation error. Can't reach initial position. 8 Check the reaction disk as instructed by 4.4.6.4 Reac. Disk tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101921401 Re-disk rotation error. Can't leave initial position. 8 Check the reaction disk as instructed by 4.4.6.4 Reac. Disk tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101921402 Re-disk rotation error. Can't detect weight disk signal. 8 Check the reaction disk as instructed by 4.4.6.4 Reac. Disk tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

101921500 Photoelectrical data overflow. 9 Other errors have higher priorities.


If the error remains, contact Mindray Customer Service Department or
your local distributor.

101929000 Execution result not received within specified time. 0 Contact Mindray Customer Service Department or your local distributor.

101929100 Command response wrong or no response 0 Contact Mindray Customer Service Department or your local distributor.
101929200 Sending analyzed instructions failed. 9 Contact Mindray Customer Service Department or your local distributor.

6-26
Troubleshooting

Error code Error Message Level Corrective Measure


102409000 Bar Code Scanner handshake failed 12 Shut down the analyzing unit and start it again. Enter the System
Maintenance screen, select the Others tab and click the Download
Settings button.
If the error still appears, contact our company or your local distributor.

111282100 Sample probe v-move error. Can't reach initial position. 9 Check the sample probe as instructed by 4.4.6.1 S. Probe tab.
(Re-disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.

111282101 Sample probe v-move error. Can't leave initial position. 9 Check the sample probe as instructed by 4.4.6.1 S. Probe tab.
(Re-disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.

111282102 Sample probe v-move error. Not at v-initial position for 9 Check the sample probe as instructed by 4.4.6.1 S. Probe tab.
d-move. (Re-disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-27
Troubleshooting

Error code Error Message Level Corrective Measure


111282106 Sample probe v-move error. Illegal surface detection. 2 Turn off the analyzing unit and
(Re-disk)
(1) Check whether the sample tubes are placed too high on the sample
disk.
(2) Check the sample syringe for leakage as instructed by 5.2.1 Checking
Sample/Reagent Syringes.
(3) Check the sample probe to see if there are drops hanging on the probe
tip. If there are, contact our company or your local distributor.
(4) Check whether the sample probe is bent or scratched. Replace the
damaged probe as instructed by 5.6.3 Replacing Sample Probe if
necessary.
(5) Clean the sample probe as instructed by 5.3.1 Washing Sample
Probe.
If this error occurs frequently. Contact Mindray Customer Service
Department or your local distributor.

111282110 Sample probe v-move error. Current position prohibits 9 Check the sample probe as instructed by 4.4.6.1 S. Probe tab.
v-move. (Re-disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.

111282400 V-position not initialized. (Re-disk) 9 Check the sample probe as instructed by 4.4.6.1 S. Probe tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

111442100 Reagent probe v-move error. Can't reach initial 9 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab.
position. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-28
Troubleshooting

Error code Error Message Level Corrective Measure


111442101 Reagent probe v-move error. Can't leave initial 9 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab.
position. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
111442102 Reagent probe v-move error. Not at v-initial position for 9 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab.
d-move. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
111442106 Reagent probe v-move error. Illegal surface detection. 0 Turn off the analyzing unit and
(Reagent disk)
(1) Check the types of the reagent bottles.
(2) Check the reagent syringe for leakage as instructed by 5.2.1 Checking
Sample/Reagent Syringes.
(3) Check the reagent probe to see if there are drops hanging on the probe
tip. If there are, contact our company or your local distributor.
(4) Check whether the reagent probe is bent or scratched. Replace the
damaged probe as instructed by 5.6.4 Replacing Reagent Probe if
necessary.
(5) Clean the reagent probe as instructed by 5.3.2 Washing Reagent
Probe.
If this error occurs frequently. Contact Mindray Customer Service
Department or your local distributor.

111442110 Reagent probe v-move error. Current position prohibits 9 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab.
v-move. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-29
Troubleshooting

Error code Error Message Level Corrective Measure


111442400 Illegal surface detection. (Reagent disk) 9 Turn off the analyzing unit and
(1) Check the types of the reagent bottles.
(2) Check the reagent syringe for leakage as instructed by 5.2.1 Checking
Sample/Reagent Syringes.
(3) Check the reagent probe to see if there are drops hanging on the probe
tip. If there are, contact our company or your local distributor.
(4) Check whether the reagent probe is bent or scratched. Replace the
damaged probe as instructed by 5.6.4 Replacing Reagent Probe if
necessary.
(5) Clean the reagent probe as instructed by 5.3.2 Washing Reagent
Probe.
If this error occurs frequently, contact our company or your local distributor.
111601700 Mixing bar v-move error. Can't reach initial position. 9 (1). Turn off the analyzing unit. Clean the mixing bar rotor
(Re-disk)
(2) Turn on the analyzing unit. Check the mixing bar as instructed by
4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

111601701 Mixing bar v-move error. Can't leave initial position. 9 (1) Turn off the analyzing unit. Clean the mixing bar rotor
(Re-disk)
(2) Turn on the analyzing unit. Check the mixing bar as instructed by
4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-30
Troubleshooting

Error code Error Message Level Corrective Measure


111601702 Mixing bar v-move error. Not at v-initial position for 9 (1) Turn off the analyzing unit. Clean the mixing bar rotor
d-move. (Re-disk)
(2) Turn on the analyzing unit. Check the mixing bar as instructed by
4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

111601707 Mixing bar v-move error. Current position prohibits 9 Other errors have higher priorities. Check the movement of the mixing bar
v-move. (Re-disk) at the maintenance screen.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

111602000 Robotic arm v-move error. Can't reach initial position. 9 Turn on the analyzing unit and check the manipulator as instructed by
(Re-disk) 4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

111602001 Robotic arm v-move error. Can't leave initial position. 9 Turn on the analyzing unit and check the manipulator as instructed by
(Re-disk) 4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
121442100 Reagent probe v-move error. Can't reach initial 5 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab.
position. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.

121442101 Reagent probe v-move error. Can't leave initial 5 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab.
position. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-31
Troubleshooting

Error code Error Message Level Corrective Measure


121442102 Reagent probe v-move error. Not at v-initial position for 5 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab.
d-move. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
121442110 Reagent probe v-move error. Current position prohibits 5 Check the reagent probe as instructed by 4.4.6.2 R. Probe tab.
v-move. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.

131282100 Sample probe v-move error. Can't reach initial position. 4 Check the sample probe as instructed by 4.4.6.1 S. Probe tab.
(Sample disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
131282101 Sample probe v-move error. Can't leave initial position. 4 Check the sample probe as instructed by 4.4.6.1 S. Probe tab.
(Sample disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.

131282102 Sample probe v-move error. Not at v-initial position for 4 Check the sample probe as instructed by 4.4.6.1 S. Probe tab.
d-move. (Sample disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.

131282110 Sample probe v-move error Current position prohibits 4 Check the sample probe as instructed by 4.4.6.1 S. Probe tab.
v-move. (Sample disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-32
Troubleshooting

Error code Error Message Level Corrective Measure


162240000 ISE unit error 12 Enter the ISE Maintenance screen of the operating software and click the
Handshake button.
If the error remains, shut down the analyzer unit and start it again. Enter
the System Maintenance screen, select Others tab and click the
Download Settings button.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

200000001 Re-disk actual temperature exceeds top limit. 0 Turn off the analyzing unit and check whether the ambient temperature
falls into the specified range.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

200000002 Reagent actual pre-heat temperature exceeds top limit. 0 Turn off the analyzing unit and check whether the ambient temperature
falls into the specified range.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

200000003 Re-disk did not reach expected temperature within 0 Check whether the ambient temperature falls into the specified range.
specified time.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

200000004 Reagent preheat didn't reach set temperature within 0 Check whether the ambient temperature falls into the specified range.
specified time.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

200000005 Re-disk temperature fluctuates over given range under 0 Check whether the ambient temperature falls into the specified range.
normal operation
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-33
Troubleshooting

Error code Error Message Level Corrective Measure


200000006 Reagent preheat temperature fluctuate over given 0 Check whether the ambient temperature falls into the specified range.
range under normal operation.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
200000007 Re-disk temperature control shutdown for overheating. 0 Turn off the analyzing unit and check whether the ambient temperature
falls into the specified range.
Contact Mindray Customer Service Department or your local distributor.
200000008 Preheat temperature shutdown for overheating. 0 Turn off the analyzing unit and check whether the ambient temperature
falls into the specified range.
Contact Mindray Customer Service Department or your local distributor.
201280005 Sample probe syringe full. 0 If the error remains, contact Mindray Customer Service Department or
your local distributor.
201280006 Sample probe syringe empty. 0 If the error remains, contact Mindray Customer Service Department or
your local distributor.

201280007 Sample probe didn't detect fluid surface in Sample 20 Check the sample volumes in the sample tubes.
disk.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

201280009 Sample probe didn't detect fluid surface in Re-disk. 2 Reagent unit errors have higher priority.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201280010 Insufficient dispensing of Sample probe. 0 Check the dispensing ability of the sample syringe as instructed by 4.4.6.7
Fluid Path tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-34
Troubleshooting

Error code Error Message Level Corrective Measure


201440005 Reagent probe syringe full. 0 If the error remains, contact Mindray Customer Service Department or
your local distributor.

201440006 Reagent probe syringe empty. 0 If the error remains, contact Mindray Customer Service Department or
your local distributor.

201440007 Reagent probe didn't detect fluid surface in Reagent 30 Check the reagent volumes in the reagent bottles.
disk.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201440009 Reagent probe didn't detect fluid surface in Re-disk. 3 Turn off the analyzing unit and
(1) Check the remaining reagents in the reagent bottles.
(2) Check the reagent syringe for leakage as instructed by 5.2.1 Checking
Sample/Reagent Syringes.
(3) Check the reagent probe to see if there are drops hanging on the probe
tip. If there are, contact our company or your local distributor.
(3) Check the deionized water tank as instructed by 5.2.8 Checking
Remaining Deionized Water.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

201440010 Insufficient dispensing of Reagent probe. 0 Check the dispensing ability of the reagent syringe as instructed by 4.4.6.7
Fluid Path tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

201600009 Robotic arm can't release held Re-cuvette. Unloading 6 Check the manipulator as instructed by 4.4.6.10 Autoloader tab.
failed.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-35
Troubleshooting

Error code Error Message Level Corrective Measure


201600010 Robotic arm can't grab used Re-cuvette from Re-disk. 6 Check the manipulator as instructed by 4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201600011 Re-cuvette dropped from robotic arm. 0 Check the manipulator as instructed by 4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

201600012 Robotic arm didn't place used Re-cuvette. 0 Check the manipulator as instructed by 4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201600013 Robotic arm didn't grab new Re-cuvette from 6 Check the manipulator as instructed by 4.4.6.10 Autoloader tab.
compartment.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

201600014 Re-cuvette dropped from robotic arm. 0 Check the manipulator as instructed by 4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

201600015 Robotic arm didn't place new Re-cuvette. 6 Check the manipulator as instructed by 4.4.6.10 Autoloader tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

201600016 Collision occurred when unloading used cuvette. 6 Check the manipulator as instructed by 4.4.6.10 Autoloader tab.
Operation cancelled.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

6-36
Troubleshooting

Error code Error Message Level Corrective Measure


201600017 Collision occurred when loading new cuvette. 6 Check the manipulator as instructed by 4.4.6.10 Autoloader tab.
Operation cancelled.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201600018 Used cuvette compartment full. Storing segments may 0 Turn off the analyzing unit and empty the used-cuvette bucket as
cause problem. instructed by 5.2.10 Emptying Used-Cuvettes Bucket.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

201600019 Compartment empty. Can't load new segment. 0 Check the remaining cuvettes in the cuvette compartment and add more
cuvettes as needed.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

210000001 Waste tank full 13 Check and empty the waste tank as instructed by 5.2.9 Emptying Waste
Tank. If the error remains, contact Mindray Customer Service Department
or your local distributor.

210000002 Deionized water tank empty. 13 Check and refill the DI water tank. If the error remains, contact Mindray
Customer Service Department or your local distributor.
400000001 System initialization error. Can't connect main unit. 10 (1) Turn off the POWER and check the connection between the analyzing
unit and the operation unit.
(2) Use the emergency exit function to exit the control system.
If this error remains, contact our company or your local distributor.
400000002 System initialization error. Main unit self-test abnormal. 10 Use the emergency exit function to exit the control system and then restart
it.
If this error remains, contact our company or your local distributor.

6-37
Troubleshooting

Error code Error Message Level Corrective Measure


400000010 System initialization error. Serial port initialization error. 10 Use the emergency exit function to exit the control system and then restart
it.
If this error remains, Contact Mindray Customer Service Department or
your local distributor.

400000012 System initialization error. Can't connect printer. 0 (1) Ensure the printer is well connected to the printer.
(2) Use the emergency exit function to exit the control system and then
restart it.
If this error remains, Contact Mindray Customer Service Department or
your local distributor.
400000016 System initialization error. Main unit connection 10 (1) Turn off the POWER and check the connection between the analyzing
abnormal. unit and the operation unit.
(2) Use the emergency exit function to exit the control system and then
restart it.
If this error remains, contact our company or your local distributor.
400000100 Power-on pretreatment error. Lamp of insufficient 1 Other errors have higher priority.
illumination. Replacement needed.
If the error remains, change the lamp or contact our company or your local
distributor.

400000204 Updating database failed. 0 Contact Mindray Customer Service Department or your local distributor.

400000209 Can't open help file. 0 Contact Mindray Customer Service Department or your local distributor.

400000216 Wrong check sum for received data 0 Contact Mindray Customer Service Department or your local distributor.

400000220 Serial port error. Instruction can't be transmitted 10 Check the connection between the analyzing unit and the operation unit,
properly. or change the serial port. Then run the control system again.
If this error remains, Contact Mindray Customer Service Department or
your local distributor.

6-38
Troubleshooting

Error code Error Message Level Corrective Measure


400000221 Serial port error. Can't receive test data. 10 Check the connection between the analyzing unit and the operation unit,
or change the serial port. Then run the control system again.
If this error remains, Contact Mindray Customer Service Department or
your local distributor.

400000225 Insufficient remaining reagent. 0 Check the remaining reagents for the test and add reagents to the bottles
as needed.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
400000251 Calibration sensitivity error. 0 Check the calibration settings and calibrators. Then recalibrate.
If this error occurs frequently. Contact Mindray Customer Service
Department or your local distributor.

400000252 Calibration factor difference exceeds limit. 0 Check the calibration settings and calibrators. Then recalibrate.
If this error occurs frequently. Contact Mindray Customer Service
Department or your local distributor.
400000253 Multi-point linear or non- linear correlated calibration 0 Check the calibration settings and calibrators. Then recalibrate.
factor (R2) exceeds limit.
If this error occurs frequently. Contact Mindray Customer Service
Department or your local distributor.

400000254 Standard difference of calibration curve exceeds limit. 0 Check the calibration settings and calibrators. Then recalibrate.
If this error occurs frequently. Contact Mindray Customer Service
Department or your local distributor.

400000255 Can't calculate calibration parameter by given method 0 Check the calculation method.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

400000276 Received data overflow. 10 Contact Mindray Customer Service Department or your local distributor.

6-39
Troubleshooting

Error code Error Message Level Corrective Measure


400000281 Calibration zero concentration absorbency exceeds 1 Check the calibration settings and calibrators. Then recalibrate.
limit.
If this error remains, Contact Mindray Customer Service Department or
your local distributor.
400000282 Incomplete calibration repeat data. Can't calculate. 1 Re-calibrate.

400000283 No calibrator. Can't calculate. 1 Re-edit the calibrator settings and re-calibrate.
400000284 Reagent out. 30 Check the remaining reagents of the test and add reagents as needed.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
400000286 Test cycle overtime. Can't continue. 9 Other errors have higher priority.
Exit the control system and then restart it.
If the error remains, contact Mindray Customer Service Department or
your local distributor.

400000287 Excessive dark current. Test prohibited. 10 Other errors have higher priority.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
400000288 Reaction calculation error. 0 Re-run the sample.
If this error occurs frequently. Contact Mindray Customer Service
Department or your local distributor.

400000289 Response exceeds calibration range. Can't calculate 0 Dilute the sample and rerun it.
result concentration.

400000290 Main unit can't recover from wrong status. 8 Contact Mindray Customer Service Department or your local distributor.

6-40
Troubleshooting

Error code Error Message Level Corrective Measure


400000291 Startup pretreatment error. Wrong blank test result. 10 Use the emergency exit function to exit the control system and then restart
it.
If this error remains, Contact Mindray Customer Service Department or
your local distributor.

400000292 Downloading parameter failed. 10 Other errors have higher priority.


Exit the control system and then restart it.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
400000293 Pretreatment not finished properly. 10 This error is triggered by other errors. Check the causes of the root errors
and take the recommended corrective measures. Use the emergency exit
function to exit the control system and then restart it.
If this error remains, contact our company or your local distributor.

400000294 Shutdown. Not finished properly. 10 This error is triggered by other errors. Check the causes of the root errors
and take the recommended corrective measures. Use the emergency exit
function to exit the control system and then restart it.
If this error remains, contact our company or your local distributor.

400000296 Concentration can't be calculated. 0 Rerun or recalculate

500000001 LIS: Unconnected 0 Check if the LIS server is power on and can work properly.
Check if the line between the system and the LIS can work properly.
If this error remains, contact our company or your local distributor.
500000002 LIS: Sending HL7 message failed 0 Check if the LIS server is power on and can work properly.
Check if the line between the system and the LIS can work properly.
If this error remains, contact our company or your local distributor.

6-41
Troubleshooting

Error code Error Message Level Corrective Measure


500000005 LIS: Something wrong with the network. Check the 0 Check if the LIS server is power on and can work properly.
network connection.
Check if the line between the system and the LIS can work properly.
If this error remains, contact our company or your local distributor.

500000010 LIS: Segment sequence error in HL7 message 0 Downloading or sending the message again.
If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.

500000011 LIS: Required field missing in HL7 message 0 Downloading or sending the message again.
If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.
500000012 LIS: Data type error in HL7 message 0 Downloading or sending the message again.
If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.
500000013 LIS: Table value not found in HL7 message 0 Downloading or sending the message again.
If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.

6-42
Troubleshooting

Error code Error Message Level Corrective Measure


500000014 LIS: Unsupported message type in HL7 message 0 Downloading or sending the message again.
If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.

500000015 LIS: Unsupported event code in HL7 message 0 Downloading or sending the message again.
If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.

500000016 LIS: Unsupported processing id in HL7 message 0 Downloading or sending the message again.
If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.

500000017 LIS: Unsupported version id in HL7 message 0 Downloading or sending the message again.
If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.

500000018 LIS: Unknown key identifier in HL7 message 0 Downloading or sending the message again.
If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.

6-43
Troubleshooting

Error code Error Message Level Corrective Measure


500000019 LIS: Duplicate key identifier in HL7 message 0 Downloading or sending the message again.
If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.

500000020 LIS: Application record locked 0 Check the LIS server.


If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.
500000021 LIS: Application internal error 0 Check the LIS server.
If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.

500000031 LIS: Application error 0 Check the LIS server.


If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.

500000032 LIS: Application reject 0 Check the LIS server.


If the error remains, consult your LIS developer first.
If your LIS developer cannot resolve this problem, contact our company or
your local distributor.
500000040-50 LIS: Position , no data found 0 Input the sample information manually.
0000098

6-44
Troubleshooting

Error code Error Message Level Corrective Measure


500000099 LIS: No data found 0 Ensure there is no sample information of the current day in the LIS server.
If there is, check the LIS server.

500000101 Bar Code Scanner communication is abnormal 0 Shut down the analyzing unit and start it again. Enter the System
Maintenance screen, select the Others tab and click the Download
Settings button.
If the error still appears, contact our company or your local distributor.
500000102 Bar Code Scanner communication failed 0 Shut down the analyzing unit and start it again. Enter the System
Maintenance screen, select the Others tab and click the Download
Settings button.
If the error still appears, contact our company or your local distributor.

600000001 ISE communication is abnormal 12 Enter the ISE Maintenance screen of the operating software and click the
Handshake button.
If the error remains, shut down the analyzing unit and start it again. Enter
the System Maintenance screen, select the Others tab and click the
Download Settings button.
If the error still appears, contact our company or your local distributor.

600000002 ISE communication failed 12 Enter the ISE Maintenance screen of the operating software and click the
Handshake button.
If the error remains, shut down the analyzing unit and start it again. Enter
the System Maintenance screen, select the Others tab and click the
Download Settings button.
If the error still appears, contact our company or your local distributor.

600000003 ISE: results overtime. 0 Rerun the ISE tests.


If the error remains, restart the system.
If the error still remains, contact our company or your local distributor.

6-45
Troubleshooting

Error code Error Message Level Corrective Measure


600000010 Na electrode goes beyond its service days. 0 Replace the Na electrode with a new one.

600000011 K electrode goes beyond its service days. 0 Replace the K electrode with a new one. If the error remains, contact
Mindray Customer Service Department or your local distributor.

600000012 Li electrode goes beyond its service days. 0 Replace the Li electrode with a new one. If the error remains, contact
Mindray Customer Service Department or your local distributor.
600000013 Cl electrode goes beyond its service days. 0 Replace the Cl electrode with a new one. If the error remains, contact
Mindray Customer Service Department or your local distributor.

600000014 The reference electrode goes beyond its service days. 0 Replace the reference electrode with a new one. If the error remains,
contact Mindray Customer Service Department or your local distributor.
600000015 The pump tubing goes beyond its service days. 0 Contact Mindray Customer Service Department or your local distributor.

600000016 The fluidic tubing goes beyond its service days. 0 Contact Mindray Customer Service Department or your local distributor.

600000020 Na electrode goes beyond its service times. 0 Replace the Na electrode with a new one. If the error remains, contact
Mindray Customer Service Department or your local distributor.

600000021 K electrode goes beyond its service times. 0 Replace the K electrode with a new one. If the error remains, contact
Mindray Customer Service Department or your local distributor.

600000022 Cl electrode goes beyond its service times. 0 Replace the Cl electrode with a new one. If the error remains, contact
Mindray Customer Service Department or your local distributor.

600000023 Li electrode goes beyond its service times. 0 Replace the Li electrode with a new one. If the error remains, contact
Mindray Customer Service Department or your local distributor.

600000024 The reference electrode goes beyond its service times. 0 Replace the reference electrode with a new one. If the error remains,
contact Mindray Customer Service Department or your local distributor.

600000030 Reading information of reagent module error. 0 Restart the system.


If the error remains, replace with a new reagent module.
If the error still remains, contact our company or your local distributor.

6-46
Troubleshooting

Error code Error Message Level Corrective Measure


600000031 Writing information of reagent module error. 0 Restart the system.
If the error remains, replace with a new reagent module.
If the error still remains, contact our company or your local distributor.

600000032 Reagent module is expired. 0 Replace with a new reagent module. If the error remains, contact Mindray
Customer Service Department or your local distributor.
600000033 Reagents in reagent module exhausted. 0 Replace with a new reagent module. If the error remains, contact Mindray
Customer Service Department or your local distributor.

6-47
Troubleshooting

6.2.2 Corrective Measures for ISE Unit Errors


The error codes of the ISE unit (optional) are different with other ones. The error
code of the ISE unit has a format of 7 characters, the format of which is <1234567>.

NOTE:
Only one error code of the ISE unit has a format of 8 characters that is
<S0000000>. You can look it as <S000000>.

You can find all the seven numbers in the table below digit by digit and then know
what is wrong with the ISE unit. Each digit represents a certain error. Each row
represents where the error locates in.

When the system reminds you of ISE error, you can open the log to find the error
code.

For example, when you see 0001000, because the number 0 means no error, you
need only to find where the 1 locates in the Digit 4 column. You will find it locates in
the Na+ row and you will know there is something wrong with the Na+ electrode.

Error Digit Digit Digit Digit Digit Digit Digit


1 2 3 4 5 6 7
Air in S 0 0 0 0 0 0
Sample/Urine
Air in Calibrant A A 0 0 0 0 0 0
Air in Calibrant B B 0 0 0 0 0 0
Air in Cleaner C 0 0 0 0 0 0
Air in Segment M 0 0 0 0 0 0
Pump Calibration P 0 0 0 0 0 0
No Flow F 0 0 0 0 0 0
Bubble Detector D 0 0 0 0 0 0
Reagent Module N 0 0 0 0 0 0
Reagent Chip R 0 0 0 0 0 0
Read
Reagent Chip W 0 0 0 0 0 0
Write
Store Calibration Q 0 0 0 0 0 0
Value Error
Command Error T 0 0 0 0 0 0
No Error 0 0 0 0 0 0 0
Na+ 0 1 1 1 1 1 1
K+ 0 2 2 2 2 2 2
Na+, K+ 0 3 3 3 3 3 3

6-48
Troubleshooting

Error Digit Digit Digit Digit Digit Digit Digit


1 2 3 4 5 6 7
Li+ 0 4 4 4 4 4 4
Li+, Na+ 0 5 5 5 5 5 5
Li+, K+ 0 6 6 6 6 6 6
Li+, K+, Na+ 0 7 7 7 7 7 7
Cl- 0 8 8 8 8 8 8
Cl-, Na+ 0 9 9 9 9 9 9
Cl-, K+ 0 A A A A A A
Cl-, Li+ 0 B B B B B B
Cl-, K+, Na+ 0 C C C C C C
Cl-, Li+, Na+ 0 D D D D D D
Cl-, Li+, K+ 0 E E E E E E
Cl-, Li+, K+, Na+ 0 F F F F F F

The following table lists the corrective measures of the ISE unit errors.

NOTE:
The measures are all sequential, that means only after you perform the
first step and the error remains, you need to go to the second step.
Contact Mindray Customer Service Department or your local distributor if
the errors not listed in the table below occur.
If the corrective measures in the table below have been performed but
the error remains, contact our company or your local distributor.

Digit No. Character Corrective Measure


Digit 1 S Place sufficient sample in sample tube to account for all
tests programmed.
Replace the electrode leaking fluid.
Take out the electrodes and install them again. Ensure
that all electrodes and o-rings are in place.
Digit 1 A Reseat electrodes. Ensure that all electrodes and
o-rings are in place.
Click the Clean button at the ISE Unit screen.
Digit 1 A Replace reagent module, purge and recalibrate.
(S or B not
displayed)
Digit 1 B Reseat electrodes. Ensure that all electrodes and
o-rings are in place.
Click the Clean button at the ISE Unit screen.

6-49
Troubleshooting

Digit No. Character Corrective Measure


Digit 1 N Move the reagent module and install it again.
Click the ISE Reset button at the ISE Unit screen.
Replace the reagent module with a new one.
Click the ISE Reset button at the ISE Unit screen
again.
Digit 4 1,2,4,8 Replace problem electrode and retest.
Digit 4 3,5-7,9,A-F Replace reference electrode and retest.
Digit 5 1,2,4,8 Replace problem electrode and retest.
Digit 5 3,5-7,9,A-F Replace reference electrode and retest.
Digit 6 1,2,4,8 If the electrode or the Calibrant A is new, click the
Purge A button at the ISE Unit screen and recalibrate
the module. If the electrode is new it may initially drift
as it rehydrates over the course of 15 minutes.
Replace the problem electrode and retest.
Digit 6 3,5-7,9,A-F If an electrode or the Calibrant A is new, click the
Purge A button at the ISE Unit screen and recalibrate
the module. If the electrode is new it may initially drift
as it rehydrates over the course of 15 minutes.
Replace the reference electrode and retest.
Digit 7 1-7 If low slope, replace the reagent module and retest.
Remove the reference electrode, tap to dislodge
bubble, install it back, and recalibrate.
Replace the reference electrode.
Replace the problem electrode.

6-50
7 Calculation Methods

7.1 Reaction Types


The analyzer can perform calculations for three reaction types:

Endpoint
Fixed-time
Kinetic

7.1.1 Endpoint
The endpoint reaction is the most ideal. The reaction reaches equilibrium after
certain period. Because the equilibrium constant is very large, it can be considered
that all substrates have changed into products, and the absorbance of the reacting
liquid does not change any more. The absorbance change is proportional to the
amount of the product present.

Figure 7-1 Endpoint reaction

t1 t2 t3 t

As shown in Figure 7-1, t1 is the time when the reagent is added, and t 2 is the

time when the sample is added. The reaction starts when they are mixed. At t 3 the

reaction reaches the equilibrium and the absorbance reading is taken. The reaction
time is t 3 t 2 .

The Endpoint reaction is not subject to such condition changes as enzyme


concentration, pH value and temperature, as long as the changes are not significant
enough to affect the reaction time.

7.1.2 Fixed-Time
For the fixed-time reaction, the reaction velocity (v), within a specific period, is
proportional to the substrate concentration [S], namely, v=k[S]. As the substrate is

7-1
Calculation Methods

consumed continuously, the reaction velocity becomes smaller and smaller, and so
does the change rate of the absorbance. It takes much time for such a reaction to
reach the equilibrium. Theoretically, the absorbance reading can be taken at any
time. The reaction can, however, become steady only after a delay because it is
complicated at the beginning and there are miscellaneous reactions due to the
complex serum compositions. For any first order reaction, the substrate
concentration [S] at a specific time (t) during the reaction is:

[S ] = [S 0 ] e kt
Where,

[S0] - original substrate concentration

e - base of the natural logarithm

k - velocity constant

Within the interval between t1 and t 2 , the relationship between [S] (delta
substrate concentration) and [S0] is:

[ S ]
[ S 0] = kt1 kt 2
e e

That is, within a fixed time interval, the delta substrate concentration is proportional
to the original substrate concentration, which is the generality of the first order
reactions. Within this interval, absorbance change is proportional to the substrate
concentration. See Figure 7-2.

Figure 7-2 Fixed-time reaction

t1 t2 t3 t4 t

As shown in Figure 7-2, t1 is the time when the reagent is added and t 2 is the
time when the sample is added. From t 3 the reaction becomes steady and t 4 is
the time to stop monitoring the reaction. The absorbance readings are respectively
taken at t 3 and t 4 .

Methodologically, the fixed-time method requires more than the endpoint method
does. All factors that have effects on the reaction velocity, such as pH, temperature
and enzyme concentration, must keep constant in the two-point analysis. In addition,
the timing must be precise and the calibration is necessary.

7-2
Calculation Methods

7.1.3 Kinetic
For the kinetic reaction, the reaction velocity is not related to the substrate
concentration and remains constant in the reaction process. As a result, for a given
wavelength, the absorbance of the reacting liquid changes evenly, and the change
rate (A/min) is proportional to the activity or concentration of the subject, which is
usually the enzyme.

In fact, it is impossible for the substrate concentration to be large enough, and the
reaction will no longer a zeroth order reaction when the substrate is consumed to a
certain degree. Therefore, the theory only stands within certain period. In addition,
the reaction can become steady only after a certain period because the reaction is
complicated at the beginning and there are miscellaneous reactions due to the
complex serum compositions. All reagent manufacturers have strict specifications for
these two periods. See Figure 7-3.

Figure 7-3 Kinetic reaction

t1 t2 t3 tn
t

As shown in Figure 7-3, t1 is the time when the reagent is added. t 2 is the time
when the sample is added. From t 3 the reaction becomes steady. t n is the time to
stop testing the reaction. t 3 t 2 is the time delay, and t n t 3 is the monitoring
time.

7.2 Calculation Process


The figure below shows the measurement and calculation flow of the analyzer.

AD Value

Absorbance

Response

Calibration Parameter

Test Result QC Result

QC Conclusion

7-3
Calculation Methods

7.2.1 AD Value Absorbance

Figure 7-4 How the liquid absorbs light

I0 It

As shown in Figure 7-4, when a parallel monochromatic light beam whose intensity
is I 0 goes through a flow cell (whose length is L ) containing a solution (whose
concentration is C ), some photons are absorbed, and the intensity is attenuated
from I 0 to I t , so the absorbance A of this solution is:

It
A = Lg
I0

Where,

It
- transmissivity
I0

The analyzer measures the light intensity through the photo-electric conversion,
linear amplification and A/D conversion. For the light intensity signal I i of Channel i,
the AD output Di is:

Di = K pe K a K ad I i

Where,

K pe - photo electric conversion factor

Ka - linear amplification factor

K ad - A/D conversion factor

Di - data of Channel i

Ii - light intensity of Channel i

So,

Ii0 D
Ai = lg = lg i 0
Ii Di

7-4
Calculation Methods

Where,

Ai - absorbance of Channel i

Di 0 - background AD output

Di - AD output after the substrate is added.

In theory, when the lights are off, the AD output of each channel will be zero. In
practice, because of the existence of dark currents, there is still a background
output Dibackground , which should be deducted. Then, the complete absorbance
formula should be:

Di 0 Dibackground
Ai = lg
Di Dibackground

7.2.2 Absorbance Response

For the analyzer, the response (R) is defined as the absorbance change before and
after the reaction, or the absorbance change rate during the reaction process.

The formula for calculating the response (R) is closely related to the reaction type
(kinetic, fixed-time, endpoint), the number of reagents (single, double), and the
number of wavelengths (single or double). They are respectively detailed in the
following sections.

7.2.2.1 Calculating the response of the endpoint reaction

Single-reagent and single-wavelength

Figure 7-5 Endpoint reaction with single-reagent and single-wavelength

t1 t2 t3 t

As shown in Figure 7-5, t1 is the time when the reagent (volume: V) is added. t2
is the time when the sample (volume: S) is added. The reaction starts when they are
mixed. At t 3 the reaction reaches the equilibrium and the absorbance reading is
taken. t 3 t 2 is the reaction time. So,

R = Rs RSMPB

7-5
Calculation Methods

Both of Rs and RSMPB are calculated with the formula RAt3 K1 At2 1 .

Where,

Rs - response of reaction mixture

RSMPB - response of sample blank. If the sample blank has not been requested,
RSMPB =0.

At 3 - absorbance at t3

At2 1 - absorbance at the previous point before t2

K1 - single-reagent test volume coefficient. If the start time of reaction is 0,


V
K1 = ; otherwise, K1 1.
V +S

Single-reagent and double-wavelength


The calculation method is similar to that for the single-reagent and single-wavelength
reaction, except for every measurement period, the absorbance is A1 A 2 ,
where, A 1 is the absorbance at the primary wavelength, and A 2 is the
absorbance obtained under the secondary wavelength.

Double-reagent and single-wavelength

Figure 7-6 Endpoint reaction with double-reagent and single-wavelength

t1 t2 t3 t4 t

As shown in Figure 7-6, t1 is the time when reagent 1 (volume: V1) is added. t 2 is
the time when the sample (volume: S) is added. Then, reagent 1 and the sample are
mixed. t 3 is the time when the reagent 2 (volume: V2) is added. Then, reagent 2 is
mixed into the reaction liquid. At t4 the reaction reaches equilibrium. t3 t2 is the
incubation period, and t 4 t 3 is the reaction time. So,

R = Rs RRB

7-6
Calculation Methods

Both of Rs and RRB are calculated with the formula R = At4 K 2 An .

Where,

Rs - response of reaction mixture

RRB - response of double-reagent blank

At4 - absorbance at t 4

An - absorbance at the start time of reaction

K 2 - double-reagent volume coefficient. If the start time of reaction is 0, K 2 = 0; if


V1 + S
the start time of reaction is more than 0, K 2 = 1; otherwise, K 2 = .
V1 + S + V2

Double-reagent and double-wavelength


The calculation method is similar to that for double-reagent and single-wavelength
reaction, except for every measurement period, the absorbance is A1 A 2 ,
where, A1 is the absorbance obtained at the primary wavelength, and A 2 is the
absorbance obtained at the secondary wavelength.

7.2.2.2 Calculating the response of the fixed-time reaction

Single-wavelength (for both single-reagent and double-reagent)

Figure 7-7 Fixed-time reaction with single-wavelength (single reagent)

t1 t2 t3 t4 t

As shown in Figure 7-7, t1 is the time when the reagent (volume: V) is added and
t 2 is the time when the sample (volume: S) is added. t 3 is the time from which the
reaction becomes steady. t4 is the time the test on the reaction is stopped.
Absorbance is taken respectively at t 3 , t4 and the response R is:

At4 At3
R=
t4 t3

7-7
Calculation Methods

Double-wavelength
The response is:

( At41 At42 ) ( At31 At3 2 )


R=
t4 t3

Where,

At41 - primary wavelength absorbance at t4

At42 - secondary wavelength absorbance at t4

At31 - primary wavelength absorbance at t 3

At3 2 - secondary wavelength absorbance at t 3

7.2.2.3 Calculating the response of the kinetic reaction

Single-wavelength (for both single-reagent and double-reagent)

Figure 7-8 Kinetic reaction with single-wavelength (single reagent)

t1 t2 t3 tn
t

As shown in the figure, t1 is the time when the reagent (volume: V) is added. t 2 is
the time when the sample (volume: S) is added. t 3 is the time from which the
reaction becomes steady. tn is the time when the test on the reaction is stopped.
t 3 t 2 is the time delay, and t n t 3 is the reaction time. The response ( R ) is
equal to the slope of the linear section between t 3 and tn .

Double-wavelength
The calculation method is similar to that for the single-wavelength reaction. The only
difference is that the absorbance at a certain point is equal to the primary
wavelength absorbance minus the secondary wavelength absorbance.

7-8
Calculation Methods

7.2.3 Response Calibration Parameter

The analyzer provides two calibration methods: linear calibration and nonlinear
calibration.

The linear calibration includes the single-point linear calibration (also called factor
method), two-point linear calibration (also called linear method), and multi-point (over
3 points) linear calibration (also called linear regression method). They are mainly
used for tests determined by colorimetry.

The nonlinear calibration includes Logistic-Log 4P, Logistic-Log 5P, Exponential5P,


Polynomial5P, Parabola and Spline. They are mainly used for tests determined by
turbidimetry.

7.2.3.1 Calculations of linear calibration parameters

Single-point linear calibration


Calibration formula: R = KC

Where,

R - calibrator response

C - calibrator concentration (or activity)

K - calibration parameter

R
This calibration method adopts only one calibration parameter K , and K= .
C

This calibration method requires only one calibrator. For most enzyme tests, you can
input the factor directly (input value F=1/ K ) without any calibration.

Two-point linear calibration


Calibration formula: R = KC + b

Where,

R - calibrator response

C - calibrator concentration (or activity)

K , b - calibration parameters

This calibration method adopts two calibration parameters: K and b.

R2 R1
K=
C 2 C1

7-9
Calculation Methods

R2 R1
b = R1 ( ) C1
C 2 C1

This calibration method requires two calibrators. C 1 and C 2 are respectively the
concentrations of calibrator 1 and calibrator 2. R 1 and R 2 are respectively the
responses of calibrator 1 and calibrator 2.

Multi-point linear calibration


Calibration formula: R = KC + b

Where,

R - calibrator response

C - calibrator concentration (or activity)

K , b - calibration parameters

This calibration method adopts two calibration parameters: K and b.


n n n

CiRi ( Ci)( Ri) / n


K= i =1
n
i =1
n
i =1

Ci
i =1
2
( Ci ) 2 / n
i =1
n n n
n CiRi ( Ci)( Ri) / n n
b = ( Ri) / n [ i =1 i =1 i =1
]( Ci ) / n
n n
i =1
Ci 2 ( Ci) 2 / n i =1
i =1 i =1

This calibration method requires n (n3) calibrators. C i is the concentration of


calibrator i. R i is the response of calibrator i.

7.2.3.2 Calculations of nonlinear calibration parameters

Logistic-Log 4P
1
Calibration formula: R = R0 + K
1 + exp[(a + b ln C )]

Where,

R - calibrator response

C - calibrator concentration (activity)

K , R 0, a , b - calibration parameter

7-10
Calculation Methods

This calibration method adopts four parameters: K , R 0, a and b .

This calibration method requires at least four calibrators. The concentration (or
activity) of calibrator 1 is 0, and the corresponding R is equal to R 0.

This calibration method is applied to the calibration curve that the response becomes
smaller and smaller with the concentration increase. See Figure 7-9.

Figure 7-9 LogisticLog 4P calibration curve

C1 C2 C3 C4 C

Logistic-Log 5P
1
Calibration formula: R = R0 + K
1 + exp[ (a + b ln C + cC )]

Where,

R - calibrator response

C - calibrator concentration (activity)

K , R 0, a , b , c - calibration parameter

This calibration method adopts five parameters: K , R 0, a , b and c .

This calibration method requires at least five calibrators. The concentration (or
activity) of calibrator 1 is 0, and the corresponding R is equal to R 0.

The applications of the calibration method are the same with that of Logistic-Log 4P,
but this method has a higher fitting.

Exponential 5P
Calibration formula: R = R0 + K exp[a ln C + b(ln C ) + c(ln C ) ]
2 3

Where,

R - calibrator response

C - calibrator concentration (activity)

K , R 0, a , b , c - calibration parameter

This calibration method adopts five parameters: K , R 0, a , b and c .

7-11
Calculation Methods

This calibration method requires at least five calibrators. The concentration (or
activity) of calibrator 1 is 0, and the corresponding R is equal to R 0.

This calibration method is applied to the calibration curve that the response
increases sharply when the concentration reaches a specific value. See Figure 7-10.

Figure 7-10 Exponential 5p calibration curve

C1 C2 C3 C4 C5
C

Polynomial 5P
R R0 R R0 2 R R0 3
Calibration formula: ln C = a + b ( ) + c( ) + d( )
100 100 100
Where,

R - calibrator response

C - calibrator concentration (activity)

R 0, a , b , c , d - calibration parameter

This calibration method adopts five parameters: R 0, a , b , c and d .

This calibration method requires at least five calibrators. The concentration (or
activity) of calibrator 1 is 0, and the corresponding R is equal to R 0.

This calibration method is applied to the calibration curve that the response
increases sharply when the concentration reaches a specific value.

Parabola
Calibration formula: R = aC 2 + bC + c

Where,

R - calibrator response

C - calibrator concentration (activity)

a, b, c - calibration parameter

7-12
Calculation Methods

This calibration method adopts three parameters: a , b and c .

This calibration method requires at least three calibrators.

Spline
Calibration formula: R = R0i + a i (C C i ) + bi (C C i ) + ci (C C i )
2 3

Where,

R - calibrator response

C - calibrator concentration (activity)

R0i , ai , bi , ci - calibration parameter, i =1, 2,, n-1, n - number of calibrators

This calibration method adopts 4(n-1) parameters: R0i , ai , bi and ci .

This calibration method requires 2-6 calibrators. Because this method adopts the
segment fitting, its fitting degree is the highest among all calibration types.

7.2.4 Calibration Parameter Test Result/QC Result

The calculations of test results vary with different calibration types.

7.2.4.1 Result calculations after linear calibrations

Single-point linear calibration


R
C=
K

Where,

K - calibration parameter

Or,

C = RF

Where,

F - input factor.

Two-point linear calibration

7-13
Calculation Methods

Rb
C=
K

Where,

K, b - calibration parameters

Multi-point linear calibration


Rb
C=
K

Where,

K, b - calibration parameters

7.2.4.2 Result calculations after nonlinear calibrations

Logistic-Log 4P
K
a ln( 1)
R R0
C = EXP( )
b

Where,

R0, K, a, b - calibration parameters

Logistic-Log 5P
Obtain the positive real root with the dichotomy method.

Exponential 5P
Obtain the positive real root with the dichotomy method.

Polynomial5P
R R0 R R0 R R0
C = exp( a + b ( ) + c( )2 + d ( )3 )
100 100 100

Where,

R0, a, b, c, d - calibration parameters

Parabola

7-14
Calculation Methods

Obtain the positive real root of the following linear quadratic equation:

aC 2 + bC + c R = 0 .

Spline
Obtain the positive real root with the dichotomy method.

7.2.5 QC Result QC Conclusion



The analyzer provides four QC rules: Westgard multi-rule, X -R, Cumulative sum
check and cumulative error.

7.2.5.1 Westgard multi-rule

Westgard multi-rule includes ten sub-rules. The clinical meanings of the ten
sub-rules are as follows:

Symbol Definition QC Conclusion


12S There is 1 point falling out of the range from Warning
the average+2SD to the average-2SD.
13S There is 1 point falling out of the range from Out-of-control (random
the average+3SD to the average-3SD. error, system error)
22S There are 2 adjacent points falling out of Out-of-control (system
the range from the average+2SD to the error)
average-2SD.
R4S There are 2 continuous values whose Out-of-control (random
difference exceeding 4SD. error)
41S There are 4 adjacent points falling out of Out-of-control (system
the range from the average+1SD to the error)
average-1SD.
10X There are 10 adjacent points falling on the Out-of-control (system
same side of the average. error)

7-15
Calculation Methods

QC conclusion flow:

QC Data

No
12S within control

Yes No
No No No No
22 S R4 S 41 S 10 x
13 S

Yes Yes Yes Yes Yes

out-of-control

7.2.5.2 X -R

This rule obtains the difference (R) between two tested values (X1 and X2) of a same
control, and then compares it with the input SDR. If R is larger than 2SDR or smaller
than -2SDR, the rule considers it out-of-control. The value of SDR should be obtained
in the following procedures: After pre-testing the control, test it continuously for 20
days. Test it twice before and after adding the sample everyday. Obtain the SDR of
the difference (R) between two tested values (X1 and X2).

7.2.5.3 Cumulative sum check

Rule Threshold (k) Control Limit (h)


CS-(1.0SD: 2.7SD) 1.0SD 2.7SD
CS-(1.0SD: 3.0SD) 1.0SD 3.0SD
CS-(0.5SD: 5.1SD) 0.5SD 5.1SD

1 This rule obtains the value of k and the control limit h.


2 When the QC value is no larger than the value of k, no processing is
required.
3 When the QC value overruns the value of k (larger than the upper limit or
smaller than the lower limit), the system starts to calculate the cumulative
sum.
4 For subsequent data, the system continues to calculate the cumulative
sum.
5 When the symbol changes (from positive to negative, or from negative to
positive) with the cumulative sum, the system stops the calculation and
resets the cumulative sum. For subsequent data, the system continues to
calculate the cumulative sum.

7-16
Calculation Methods

6 When the cumulative sum overruns the control limit h (larger than the
upper limit or smaller than the lower limit), the system considers it
out-of-control.

7.2.5.4 Cumulative error

This rule calculates the difference (error, expressed with the symbol plus or minus)
between the tested value and target value of a test of a control.

1st day cumulative error1 = error1

2nd day cumulative error2 = error1 + error2

3rd day cumulative error3 = cumulative error2 + error3

nth day cumulative errorn = cumulative errorn-1 + errorn

When the cumulative error overruns the cumulative error limit E or E, the system
considers it out-of-control.

7.2.5.5 Twin-plot

+3SD
+2SD

-2SD
-3SD

-3SD -2SD Y +2SD +3SD

The QC chart, which can sensitively indicate system errors and random errors, is present to
help you make a QC conclusion.

7.3 Calculation Method of ISE Unit (optional)


An ion selective electrode develops a voltage that varies with the concentration of the ion to
which it responds. The relationship between the voltage developed and the concentration of
the sensed ion is logarithmic, as expressed by the Nernst equation:

Ex = Es + RT/nF log ( C)

Where,

Ex - the potential of the electrode in sample solution

7-17
Calculation Methods

Es - the potential developed under standard conditions

RT/nF - a temperature dependent constant, termed the slope(s)

Log - base ten logarithm function

- activity coefficient of the measured ion in the solution

C - concentration of the measured ion in the solution

The calculation process of the concentration of the ion in the sample is as the following steps.

First, the module measures the potential of the calibrant A and calibrant B and calculates the
slope S of the electrode with the formula below.

EA EB
S=
C
log A
CB

where,

S the slope of the electrode

EA the potential of the calibrant A

EB the potential of the calibrant B

CA the concentration of calibrant A

CB the concentration of calibrant B

Second, the module calculates the concentration of the ion in the sample with the formula
below.

E X E A

C X = C A 10 S

Where,

Cx the concentration of the ion in the sample

Ex the potential of the sample

CA the concentration of the calibrant A

EA the potential of the calibrant A

S the slope of the electrode

7-18
Appendix A Specifications

A.1 Technical Specifications


Throughput: maximum 300tests/h; if the ISE unit (optional) is connected,
maximum 420tests/h (Na, K and Cl included) or maximum 480tests/h (Na, K, Cl
and Li included)
Tests analyzed simultaneously: 48 (single-reagent) / 24 (double-reagent); if the
ISE unit (optional) is connected, 3 (Na, K and Cl included) or 4 (Na, K, Cl and Li
included) tests are added
Test types: Endpoint, Fixed-time and Kinetic. All supporting monochromatics and
bichromatics

Sample volume: 3l-45l; for the ISE unit (optional), 70l serum, 70l plasma,
140l diluted urine; precision: 0.5l
Sample Disk: 60 sample tube positions; 5 virtual disks for maximum 300 samples
Compatible sample containers: micro-tubes, collection tubes and test tubes of the
following sizes: 12mm68.5mm, 12mm99mm, 12.7mm75mm,
12.7mm100mm
STAT samples: can be inserted immediately for analysis with higher priority
Sample probe: with a built-in level detector; equipped with bump collision
safeguard; capable of tracking sample level
Washing function: automatically washing interiors and exteriors of sample probe
Auto rerun

Reagent volume: 30l -450l; precision 1l


Reagent disk: 50 bottle positions; 2 virtual disks
Compatible reagent bottles: Mindray inner-circle bottles, Mindray outer-circle
bottles, Hitachi 7060 bottles, Hitachi 7170 bottles and Hitachi 20ml bottles
Reagent probe: with a built-in level detector; equipped with bump collision
safeguard; capable of tracking reagent level
Washing function: automatically washing interiors and exteriors of reagent probe
Refrigerator: capable of running 24 hours a day to keep the temperature at 4-10

Mixing bar: for single-reagent tests, it functions immediately (within the same
period) after sample dispensing; for double-reagent tests, it functions immediately
(within the same period) dispensing of the second reagent.

Reaction disk: 80 reaction cuvettes


Reaction temperature: 37
Compatible cuvettes: 5625; 5mm optical length; disposable; 750l
Reaction volume: 180-500l
Auto cuvette load/unloading mechanism

A-1
Specifications

Photometric system: interference filters of various wavelengths; static fiber optics


Wavelength: 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm,
700nm
Measurement period: 12 seconds
Measurement range: 0.1-4.0 Abs
Lamp: tungsten-halogen; 12V, 50VA

A.2 Power Requirements


Power supply: 100-130V/200-240V~, 50/60Hz, three-wire power cord and
properly grounded
Input power: 1000VA

A.3 Environmental Requirements


Temperature: 15- 30
Humidity: 35% - 80%, no condensation
Atmospheric pressure: 800 hPa - 1,060 hPa
Above-sea-level height: -1,300feet- 6,500 feet
-400m-2,000 m

A.4 Dimension and Weight


Dimension: 980mm710mm1200mm (WDH)
Weight: 175kg

A.5 Other Specifications


The analyzing unit meets the requirements of EN 61326:2003, Class A
Sound data: 70dB
Fuse: 250VAC T1.6A
250VAC T2A
250VAC T6.3A
250VAC T10A

A.6 Input/Output Devices


PC Keyboard
PC Mouse
Printer
Display

A-2
Specifications

A.7 Interface
LIS interface: HL7 interface developed by Mindray
Interface between Analyzing Unit and Operation Unit: RS-232

A.8 Built-in Sample Bar Code Scanner (optional)


The scanner can read the bar code that meets the following requirements:

Symbologies: Code 128, Code 39, Codabar, ITF, UPC/EAN, Code93


Feature size: 0.19-0.50mm
Length: 3-16
Thin bar: thick bar: 1:2.5-3
Height: 10mm or more
Printing: black print on white background; the minimum acceptable symbol grade
is Class C as defined in the ANSI Print Quality Specification.

The bar code labels can be applied to the following sample tubes.

12mm 99mm
12.7mm75mm
12.7mm100mm

A.9 ISE Module (optional)


Electrodes: Na+ electrode, K+ electrode, Cl- electrode, Li+ electrode and reference
electrode
Sample: serum, plasma, urine
Sample volume: 70 l serum, 70 l plasma, 140 l diluted urine
Urine dilution: automatically performed by the system
Test ranges:
Analyte (serum, plasma) Test range limit (mmol/l)
K+ 1.00-8.00
Na+ 100.0-200.0
Cl- 50.0-150.00
Li+ 0.2-3.50

Analyte (urine) Test range limit (mmol/l)


K+ 5-200
Na+ 10-500
Cl- 15-400

A-3
Appendix B Supplies
To ensure personal safety and system performance, use only supplies manufactured
or recommended by Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Contact
Mindray Customer Service Department or your local distributor for details.

Replacing Parts and Consumables

Description Part Number Location Remarks


Lamp BA30-10-06365 Lamp box Replacing part
Replace it when
1) It serves for
over 1000h; or
2) The system
prompts.
Plunger assembly of 0040-10-32317 Reagent Replacing part
reagent syringe syringe
Replace it when
1) It serves for 3
months;
2) It runs for
100,000 times; or
3) It is seriously
damaged.
Sample syringe gasket 0040-10-32303 Connection Replacing part
part
between the Replace it when
sample the sample
syringe and syringe is
the T-piece disassembled for
2 - 3 times.
Reagent syringe gasket 0040-10-32303 Connection Replacing part
part
between the Replace it when
reagent the reagent
syringe and syringe is
the T-piece disassembled for
2 - 3 times.
Reagent probe assembly BA30-30-06753 Reagent Replacing part
probe arm
Replace it when
1) It serves for 1
year; or
2) It is damaged
or bent.

B-1
Supplies

Description Part Number Location Remarks


Sample probe assembly BA30-30-14978 Sample Replacing part
probe arm
Replace it when
1) It serves for 1
year; or
2) It is damaged
or bent.
Sample probe gasket 0040-10-32307 Nut on the Replacing part
sample
probe Replace it when
1) The sample
probe is
disassembled for
2 - 3 times; or
2) The sample
probe is replaced
with a new one.
Reagent probe gasket 0040-10-32307 Nut on the Replacing part
reagent
probe Replace it when
1) The reagent
probe is
disassembled for
2 - 3 times; or
2) The reagent
probe is replaced
with a new one.
Mixing bar BA30-20-06775 Mixing bar Replacing part
arm
Replace it when
it is damaged.
A4 copy paper 0150-10-00381 Printer Consumable
Reaction cuvette BA30-20-15179 Reaction Consumable
disk
Mindray inner-circle bottle BA30-20-15058 Reagent Consumable
disk
Mindray outer-circle bottle BA30-20-15059 Reagent Consumable
disk
+
K electrode (optional) BA34-10-63641 ISE unit Consumable
(optional)
Na+ electrode (optional) BA34-10-63642 ISE unit Consumable
(optional)
Cl- electrode (optional) BA34-10-63639 ISE unit Consumable
(optional)
Li+ electrode (optional) BA34-10-63644 ISE unit Consumable
(optional)
Reference electrode BA34-10-63640 ISE unit Consumable
(optional) (optional)

B-2
Supplies

Description Part Number Location Remarks


Spacer electrode BA34-10-63643 ISE unit Consumable
(optional) (optional)
Cleaning Solution Kit BA34-10-63645 ISE unit Consumable
(optional) (optional)
Reagent Module (optional) BA34-10-63673 ISE unit Consumable
(optional)
Urine Dilution (optional) BA34-10-63671 ISE unit Consumable
(optional)
ISE control BA34-10-63669 ISE unit Consumable
(optional)

B-3
Appendix C Index

calibration result............. 4-31, 4-83



Calibrator ......1-4, 4-12, 4-15, 4-16,
absorbance ...... 1-15, 3-4, 4-3, 4-41, 4-30, 4-44, 4-45, 4-69, 4-70,
4-42, 4-44, 4-45, 4-45, 4-46 4-71, 4-72, 4-73, 4-74, 4-75,
4-76, 4-78, 4-80, 4-93
acid .................................4-38, 4-39
carryover ................................8, 3-5
Acid wash solution .................... 5-2
check box1-19, 3-6, 4-3, 4-17, 4-43,
administrator...................3-4, 4-141 4-51, 4-96, 4-97, 4-98, 4-134,
4-135, 4-136, 4-137, 4-167,
administrator authority .......... 4-141 4-168

Air bubble................................ 5-45 Circle

alarm........ 1-13, 1-18, 4-134, 4-135 inner .......................................... 1-4

alkaline .................................... 4-39 outer .......................................... 1-4

Alkaline wash solution .............. 5-2 clinical...........................................1

analyzing unit . 7, 1-1, 1-2, 1-3, 1-4, Combo box...............................1-19


1-5, 1-8, 1-9, 1-10, 1-12, 1-13,
1-14, 3-2, 3-3, 3-4, 3-8, 4-39 Communication......................... 1-4

ANALYZING UNIT POWER.. 3-3, control 7, 1-4, 3-4, 4-13, 4-16, 4-32,
3-8, 4-39 4-38, 4-39, 4-88, 4-89, 4-90,
4-91, 4-92, 4-93, 4-94, 4-95,
Assembly 4-97, 4-98, 4-101, 4-102, 4-104,
4-106, 4-107
reagent disk................................ 1-9
control system 7, 3-3, 3-4, 4-1, 4-38

cuvette loader ................... 1-2, 1-12
backgrounds............................... 3-4

Biohazard waste disposal......... 5-45
dark currents.............................. 3-4
bottle used by reagent 1 ........... 4-64
Data..... 9, 1-17, 3-6, 3-7, 4-1, 4-17,
bottle used by reagent 2 ........... 4-64 4-19, 4-21, 4-23, 4-27, 4-30,
4-33, 4-82, 4-120

decimal.....................................4-42
calibration .................................. 3-5
Deionized Water........................ 1-3
calibration curve ... 4-31, 4-78, 4-83
deionized water tank ................. 3-3
calibration parameter .....4-12, 4-47,
department name ........ 4-142, 4-143
4-48, 4-67, 4-76, 4-78, 4-79

C-1
Index

Detector Help ................................. 1-17, 4-1

deionized water level ................. 1-3 horizontal position...... 1-4, 1-5, 1-9,
1-10
waste level ................................. 1-4
hospital name .........................4-142
dialog box ..... 1-18, 3-4, 4-16, 4-22,
4-23, 4-26, 4-27, 4-38, 4-39, Humidity ................................... 2-2
4-40, 4-47, 4-48, 4-51, 4-61,
4-67, 4-72, 4-83, 4-88, 4-92, hydrochloric .............................4-38
4-118, 4-133, 4-135, 4-137,
4-138, 4-141, 4-167, 4-168,
4-170

Diluents ..................................... 1-4 Incubation ....................... 3-4, 4-43

dilution ratio ............................ 4-46 inner circle ....................... 1-4, 4-63

Door
middle........................................ 1-8
Javel water ................................ 5-2
double-reagent1-12, 4-2, 4-12, 4-13,
4-41, 4-43, 4-46, 4-63

Kinetic4-12, 4-41, 4-42, 4-43, 4-45,



4-137
edit box 1-20, 4-2, 4-42, 4-43, 4-45,
4-46, 4-70, 4-71, 4-75, 4-77,
4-78, 4-90, 4-91, 4-94, 4-98,
4-99, 4-134, 4-139, 4-142, 4-144, lamp .......................................... 3-5
4-145, 4-170
linearity .......................... 4-42, 4-46
Endpoint ..... 4-12, 4-41, 4-43, 4-44,
4-45, 4-137 lower door ................................1-12

Equilibrium.............................. 5-46 lower limit............. 4-49, 4-50, 4-50


Finished . 3-6, 3-7, 4-16, 4-17, 4-18 MAIN POWER .............................7

first reagent.....................4-46, 4-67 maintenance ..................................7

Fixed Time ............................ 4-137 menu............................. 2, 1-17, 4-1

middle door ............. 1-8, 1-10, 1-14

Gasket...................................... 5-44 mixer ........................................1-12

mixing bar4, 7, 1-12, 3-2, 3-4, 4-39,


4-134

handle ............ 5, 1-4, 1-5, 1-9, 1-10



Handwheel......... 1-4, 1-5, 1-9, 1-10
operation unit ........ 7, 1-1, 1-4, 1-16

C-2
Index

optical path .....................1-14, 1-15

option button.................1-19, 4-136 QC parameter ....... 4-47, 4-48, 4-88,


4-99
outer circle ........................1-4, 4-63
QC rules ...................................4-95
output unit.........................1-1, 1-16
QC test ......................... 4-88, 4-134


Parameters ..... 1-17, 3-5, 4-1, 4-39,
4-40, 4-41, 4-61, 4-67, 4-96, R1 blank ...................................4-46
4-97
reaction curve....... 4-17, 4-18, 4-19,
password3-4, 4-27, 4-40, 4-61, 4-67, 4-23, 4-26, 4-27, 4-30, 4-32,
4-88, 4-133, 4-141 4-87, 4-120

patient report.......4-23, 4-27, 4-166, reaction cuvette1-2, 1-8, 1-10, 1-11,


4-167 1-12, 1-14, 1-15, 4-18

photometric system...........1-15, 3-4 reaction disk ..1-8, 1-11, 1-12, 1-13,


1-14, 1-15, 3-4, 4-18
Position
reaction disk diagram ...............4-18
tube ............................................ 1-4
reagent blank .................. 4-15, 4-46
position for reagent 1............... 4-63
reagent bottle...1-9, 1-10, 1-11, 3-5,
position for reagent 2............... 4-63 3-8, 4-64, 4-66

Power..................................3-3, 3-8 reagent disk ...7, 1-9, 1-10, 3-5, 3-8,


4-62, 4-63, 4-64
cord.................................... 2-2, A-2
reagent dispenser......................1-10
socket......................................... 2-2
reagent syringe .......... 1-8, 1-10, 3-2
power socket.......................2-2, 3-2
Refrigerator ............................... 1-9
power supply .............. 1-9, 2-2, 3-2
relative humidity ................ 2-2, 2-3
primary wavelength ................. 4-42
report header ..........................4-166
printing order ..............4-166, 4-168
Requested,.......................... 3-6, 3-7
probe...............................5-31, 5-37
re-runs ....................................4-133
probe arm..........................1-7, 1-10
response.......4-45, 4-46, 4-77, 4-78,
probe rotor ........................1-7, 1-10 4-137

probe tip..................................... 3-2 response range........................4-137

profile ........................................ 4-7 Routine 1-17, 4-1, 4-11, 4-13, 4-14,


4-38, 4-133, 4-134

Running............................. 3-6, 3-7

C-3
Index

sample ....................................... 1-7

Sample System 2, 1-1, 1-15, 1-17, 1-18, 3-3,


3-7, 4-1, 4-133, 4-141, 4-165
disk ............................................ 1-4

dispenser .................................... 1-7
tab...... 1-18, 4-15, 4-17, 4-18, 4-19,
probe...................................1-4, 1-7
4-20, 4-21, 4-22, 4-23, 4-25,
4-27, 4-28, 4-29, 4-30, 4-31,
probe arm................................... 1-7 4-32, 4-33, 4-41, 4-47, 4-48,
4-49, 4-62, 4-63, 4-66, 4-68,
probe rotor ................................. 1-7 4-76, 4-88, 4-89, 4-94, 4-95,
4-98, 4-120, 4-134, 4-135, 4-136,
sample cup ................................. 1-4 4-138, 4-142, 4-143, 4-144,
4-146, 4-166, 4-167, 4-168,
sample disk7, 8, 1-4, 1-5, 1-10, 3-5, 4-170
3-6, 3-7, 3-8, 4-3, 4-12, 4-13,
4-15, 4-16, 4-17, 4-69, 4-89, temperature1-9, 1-14, 1-18, 2-2, 2-3,
4-120 3-4

sample disk diagram ................ 4-15 test report ................... 4-166, 4-168

sample dispenser........................ 1-7 Tube .......................................... 1-4

sample probe...4, 7, 8, 1-4, 1-5, 1-7,


1-8, 1-10, 3-2, 3-4, 4-39, 4-134

sample syringe ........................... 1-8 Unit .. 1-1, 1-3, 1-4, 1-11, 1-13, 3-2,
3-3, 3-8, 4-41, 4-138, 4-139
sample tube................. 1-4, 1-5, 1-8
Analyzing.................................. 1-1
sample volume................. 8, 9, 4-42
Operation................................... 1-1
Sampling.................................. 5-44
Output ....................................... 1-1
scroll bar .................................. 1-20
upper limit...4-49, 4-50, 4-50, 4-78,
second reagent ................1-12, 4-43 4-136, 4-137

secondary wavelength.............. 4-42 urine ..............................................8

serum ............................................ 8 used-cuvettes bucket ................. 3-3

short-cut button.......4-1, 4-11, 4-13, username ................................... 3-4


4-38

single-reagent ...... 1-12, 4-41, 4-43,
4-46, 4-63, 4-64
vertical position..1-4, 1-5, 1-9, 1-10
surplus antigen..... 4-43, 4-44, 4-45,
4-137 volume (180~450 l) of the first
reagent..................................4-42
surplus substrate ...................... 4-45
volume (30~450 l) of the second
Syringe reagen...................................4-42

reagent ....................................... 1-8

C-4
Index

waste tank.................... 3-3, 3-4, 3-8

Waiting ...............................3-6, 3-7 Weight ....................................... 2-1

wash well .................1-8, 1-11, 1-12 Worklist Status ..... 4-14, 4-18, 4-19,
4-21, 4-25
Waste ......................................... 1-4

C-5
P/N: BA34-20-63687 (1.8)

You might also like