Professional Documents
Culture Documents
Abstract In this paper different types of electrodes for evaluate different electrode materials and test their
long-term surface EMG recording are compared to a practicability in comparison with AG/AgCl electrodes and
reference electrode that is established for clinical use. The especially in vivo tests are still lacking.
electrode materials include four different polymers with
conductive load and a fabric of threads coated by a
II. OBJECTIVE
conductive layer. Different criteria are used to evaluate
surface EMG recording: the signal quality, including signal- Electrodes for surface EMG monitoring that can be
to-noise (SNR) ratio, and impedance in long-term integrated in a textile or in a silicone liner have to comply
monitoring. The aim of the study is to find an EMG electrode with different requirements. Ideally, they have to be
that allows for both silicone liner and textile integration for
biocompatible and sufficiently soft and flexible to provide
control of rehabilitation devices for quadriplegics with a
partial residual function of the upper limb and for a good wearing comfort, thus resulting in wide acceptance
multifunctional prosthetic hands. Besides electrical by the patients. They also have to ensure a good signal
properties, the biocompatibility and the wearing comfort quality during long-term use. Finally, they have to be
have to be considered to achieve a wide acceptance by the robust enough to be washed repeatedly together with the
patients. Except for one evaluated electrode, the signal carrier textiles.
quality of the four different surface electrodes is comparable
Appropriate sensors for surface EMG monitoring are
to commercial Ag/AgCl gel electrodes in long-term
monitoring. needed that comply with the above criteria. In a first step
different potential electrode materials have to be evaluated
I. INTRODUCTION clinically with regard to their signal quality.
A. Electrode Materials
life-long dependence on helping persons and means a
Five different materials were investigated and are
considerable restriction of the quality of life. The
depicted in table 1. Three of them consist of various types
paramount objective of modern rehabilitation medicine is
of Pt-catalyzed polysiloxane (silicone rubber) with a
to at least partly restore the individual functional defects.
biocompatibility certificate according to ISO 10993. By
Powered prosthetic and orthotic devices can provide the
loading with carbon or unspecified nanoparticles,
patient with autonomy. However, high fidelity signals
electrical conductivity was achieved [2]. Additionally, a
from monitoring electrodes are required that ensure
textile electrode (#4) consisting of Elitex threads [3]
unchanged signal quality for a long period and easy
(galvanically modified silvered threads made of polyamide
application. Ideally, the sensors and stimulation system
yarn) was evaluated. This electrode has to be saturated
should be integrated in silicone rubber liner or textiles to
with electrolyte dilution for operation as a surface EMG
provide high wearing comfort without skin irritation.
electrode. The electrode is supplied by the Textile
Commercial Ag/AgCl electrodes that are applied in
Research Institute Thuringia Vogtland (TITV), Greiz,
clinical practice typically comprise an electrolyte gel to
Germany. The material used to assemble electrode #5 is
ensure functioning. However, these electrodes may cause
based on a flexible thermoplastic elastomer loaded with
skin irritation and allergies and the electrolyte gel dries up
silver-coated glass microspheres, resulting in a highly
with time, causing a drastic decrease of the signal quality
conductive compound that is usually intended for
[1]. Thus, dry electrodes have been proposed as an
electromagnetic interference (EMI) gasket applications.
alternative for long-term monitoring of biopotentials like
All electrodes were connected via a snap fastener to the
ECG, EEG, and EMG [1, 2]. Nevertheless, studies to
leads of the data acquisition equipment and the
manufacturing process is illustrated in Fig. 1. The
I. Manuscript received February 5, 2009. This work was supported different types of electrodes are depicted in Table 1 and
by the Federal Ministry of Education and Research (BMBF) within the Fig. 2.
funding program Innovative aids. The project Orthojacket is a health
research cooperation between science and industry.
C. Pylatiuk, M. Mller-Riederer, A. Kargov, S. Schulz, O. schill, M.
Reischl and G. Bretthauer are with the Institute for Applied Computer
Science, Forschungszentrum Karlsruhe, Hermann-von-Helmholtz-Platz-
1, D-76344 Eggenstein-Leopoldshafen, GERMANY (corresponding
authors phone: +49-7247-82-2430; fax: +49-7247-82-5786; e-mail:
pylatiuk@iai.fzk.de).
301
IV. RESULTS fifth electrode did not correlate with the corresponding
signals of the Ag/AgCl gel electrode.
A. Correlations with Reference Electrode
The signals of the five electrode materials were
B. Signal-to-noise Ratios (SNR)
evaluated in comparison with the signals provided by an
Ag/AgCl gel electrode as a reference. The correlations of As a second criterion to evaluate signal quality, the
the signals measured 5 and 60 min. after application of the ratios of the signal to the background noise were
electrodes are depicted in Figs. 4 and 5, respectively. calculated for the data recorded 5 and 60 min. after
application of the electrodes. The results are given in Figs.
1.0 6 and 7.
0.8
300
0.6
correlation coefficients
-1.0 0
1 2 3 4 5 Ref
Fig. 4. Correlations of the five different electrodes with the Ag/AgCl
electrode material #
gel electrode five minutes after application to the skin of the six test
persons. (black bar= material #1, gray bar with dots= material #2, etc.) Fig. 6. Average SNRs recorded with 5 different electrode materials
and the Ag/AgCl reference electrode (Ref) 5 min. after application of
the electrodes to the test persons with low load on the muscle (light
gray bar) and high load on the muscle (dark gray bar).
1.0
0.8
0.6
correlation coefficients
0.4 300
0.2
250
signal-to-noise ratios (SNR)
high load
0.0
low load
1 2 3 4 5 6 200
-0.2
test person #
-0.4
150
-0.6
100
-0.8
-1.0 50
302
C. Electrode-to-skin Interface Impedance significantly e.g. when fixation of the electrodes is
The sequential progression of the electrode-to-skin insufficient and motions between the electrode and the
impedances for the different electrode materials is given in patients skin cause disturbance. It was demonstrated by
Fig. 8. Five minutes after application of the electrodes, the [1] that as perspiration builds up and hydrates the skin
electrode-to-skin impedance of the textile electrode (#4) layer artefact levels from movements are lower in dry
and of the reference electrode were below 50 k only. electrodes than in Ag/AgCl electrodes.
Twenty minutes after application, the impedances of all The electrode-to-skin impedance has a major impact on
electrodes except for #2, were in the range of 10 and 50 signal quality in EMG surface recording. According to the
k. After one hour, the impedance of the textile electrode SENIAM recommendations [4], the impedance has to be
increased continuously, whereas the impedances of the determined prior to any EMG surface data acquisition.
other electrodes remained in the range of 30 and 70 k. The impedance is a complex resistance to alternating
currents, and a simplified models of an electrical
400
equivalent circuit are given in [1, 2]. During the first two
hours after application of the electrodes to the test
300 persons` skin, the lowest impedance was found for the
textile electrode (material #4), which was even lower than
Impedance [k]
303
ACKNOWLEDGMENT
The authors gratefully acknowledge the valuable help of
Rdiger Rupp of the Orthopedic University Hospital,
Heidelberg, Germany, Christian Rotsch of the Textile
Research Institute Thuringia Vogtland (TITV), Greiz,
Germany, Niels Wessel of the University of Potsdam,
Germany and Roman Ruff of the Fraunhofer Institute for
Biomedical Engineering, St. Ingbert, Germany.
This work is supported by the German Federal Ministry of
Education and Research BMBF grant 01EZ-0774.
REFERENCES
[1] J. Mhlsteff, O. Such: Dry electrodes for monitoring of vital signs
in functional textiles. Proc. of the 26th Annual International
Conference of the IEEE Engineering in Medicine and Biology
Society (EMBS `04), San Francisco, CA, USA, September 1-5,
2004, 2212-2215.
[2] K.-P. Hoffmann, R. Ruff, W. Poppendieck: Long-Term
Characterization of Electrode Materials for Surface Electrodes in
Biopotential Recording. Proc. of 28th Annual International
Conference of the IEEE Engineering in Medicine and Biology
Society (EMBS '06), Aug 30-Sept 3 2006, NY, USA, 2239-2242.
[3] D. Obenauf, H. Hieber, W. Scheibner: Flat textile structure of
bonded insulating textile layers contains microstructured wiring.
German Patent DE19755792, 1999.
[4] H. J. Hermens et al.: SENIAM 8: European Recommendations for
Surface ElectroMyoGraphy. Roessingh Research & Development
b.v., ISBN: 90-75452-15-2, 1999. http://www.seniam.org .
[5] R. Mikut, O. Burmeister, S. Braun, M. Reischl: The Open Source
Matlab Toolbox Gait-CAD and Its Application to Bioelectric Signal
Processing. Proc. of the DGBMT Workshop Biosignalverarbeitung,
July 16-18, 2008, Potsdam, Germany, 109-111; 2008.
[6] C. Pylatiuk, S. Schulz, A. Kargov, G. Bretthauer: Two Multi-
articulated Hydraulic Hand Prostheses. Artificial Organs
28(11):980-986, 2004.
[7] R. Rupp, U. Eck, O. Schill, M. Reischl, S. Schulz: OrthoJacket
An active FES-hybrid orthosis for the paralyzed upper extremity.
2nd European Conference Technically Assisted Rehabilitation
(TAR 2009), Berlin, March 18-19, 2009.
304