You are on page 1of 5

2009 IEEE 11th International Conference on Rehabilitation Robotics

Kyoto International Conference Center, Japan, June 23-26, 2009

Comparison of Surface EMG Monitoring Electrodes for Long-term


Use in Rehabilitation Device Control
C. Pylatiuk, M. Mller-Riederer, A. Kargov, S. Schulz, O. Schill, M. Reischl and G. Bretthauer,
Member of IEEE

Abstract In this paper different types of electrodes for evaluate different electrode materials and test their
long-term surface EMG recording are compared to a practicability in comparison with AG/AgCl electrodes and
reference electrode that is established for clinical use. The especially in vivo tests are still lacking.
electrode materials include four different polymers with
conductive load and a fabric of threads coated by a
II. OBJECTIVE
conductive layer. Different criteria are used to evaluate
surface EMG recording: the signal quality, including signal- Electrodes for surface EMG monitoring that can be
to-noise (SNR) ratio, and impedance in long-term integrated in a textile or in a silicone liner have to comply
monitoring. The aim of the study is to find an EMG electrode with different requirements. Ideally, they have to be
that allows for both silicone liner and textile integration for
biocompatible and sufficiently soft and flexible to provide
control of rehabilitation devices for quadriplegics with a
partial residual function of the upper limb and for a good wearing comfort, thus resulting in wide acceptance
multifunctional prosthetic hands. Besides electrical by the patients. They also have to ensure a good signal
properties, the biocompatibility and the wearing comfort quality during long-term use. Finally, they have to be
have to be considered to achieve a wide acceptance by the robust enough to be washed repeatedly together with the
patients. Except for one evaluated electrode, the signal carrier textiles.
quality of the four different surface electrodes is comparable
Appropriate sensors for surface EMG monitoring are
to commercial Ag/AgCl gel electrodes in long-term
monitoring. needed that comply with the above criteria. In a first step
different potential electrode materials have to be evaluated
I. INTRODUCTION clinically with regard to their signal quality.

P EOPLE with paraplegia or amputation of the upper


extremity suffer from the loss of motoric functions and
in particular of the gripping function. This may lead to
III. MATERIALS AND METHODS

A. Electrode Materials
life-long dependence on helping persons and means a
Five different materials were investigated and are
considerable restriction of the quality of life. The
depicted in table 1. Three of them consist of various types
paramount objective of modern rehabilitation medicine is
of Pt-catalyzed polysiloxane (silicone rubber) with a
to at least partly restore the individual functional defects.
biocompatibility certificate according to ISO 10993. By
Powered prosthetic and orthotic devices can provide the
loading with carbon or unspecified nanoparticles,
patient with autonomy. However, high fidelity signals
electrical conductivity was achieved [2]. Additionally, a
from monitoring electrodes are required that ensure
textile electrode (#4) consisting of Elitex threads [3]
unchanged signal quality for a long period and easy
(galvanically modified silvered threads made of polyamide
application. Ideally, the sensors and stimulation system
yarn) was evaluated. This electrode has to be saturated
should be integrated in silicone rubber liner or textiles to
with electrolyte dilution for operation as a surface EMG
provide high wearing comfort without skin irritation.
electrode. The electrode is supplied by the Textile
Commercial Ag/AgCl electrodes that are applied in
Research Institute Thuringia Vogtland (TITV), Greiz,
clinical practice typically comprise an electrolyte gel to
Germany. The material used to assemble electrode #5 is
ensure functioning. However, these electrodes may cause
based on a flexible thermoplastic elastomer loaded with
skin irritation and allergies and the electrolyte gel dries up
silver-coated glass microspheres, resulting in a highly
with time, causing a drastic decrease of the signal quality
conductive compound that is usually intended for
[1]. Thus, dry electrodes have been proposed as an
electromagnetic interference (EMI) gasket applications.
alternative for long-term monitoring of biopotentials like
All electrodes were connected via a snap fastener to the
ECG, EEG, and EMG [1, 2]. Nevertheless, studies to
leads of the data acquisition equipment and the
manufacturing process is illustrated in Fig. 1. The
I. Manuscript received February 5, 2009. This work was supported different types of electrodes are depicted in Table 1 and
by the Federal Ministry of Education and Research (BMBF) within the Fig. 2.
funding program Innovative aids. The project Orthojacket is a health
research cooperation between science and industry.
C. Pylatiuk, M. Mller-Riederer, A. Kargov, S. Schulz, O. schill, M.
Reischl and G. Bretthauer are with the Institute for Applied Computer
Science, Forschungszentrum Karlsruhe, Hermann-von-Helmholtz-Platz-
1, D-76344 Eggenstein-Leopoldshafen, GERMANY (corresponding
authors phone: +49-7247-82-2430; fax: +49-7247-82-5786; e-mail:
pylatiuk@iai.fzk.de).

9781-4244-3789-4/09/$25.00 2009 IEEE 300


SENIAM project [4]. Data acquisition of the EMG signals
was performed with a sampling frequency of 2 kHz using
a BioAmp, a Powerlab 4/20 data acquisition system,
and LabChart software (all from AD Instruments Ltd.,
Sydney, Australia). Pre-processing of the raw signals
included differential amplification, high-pass and low-pass
filtering with cut-off frequencies of 10 and 500 Hz,
respectively, notch filtering of 50 Hz, and rectification.

Six healthy colleagues (age 25-41 years) participated in


the experiment after signing an informed-consent form.
The subjects had no history or clinical signs of upper
Fig. 1. Schematic view of the electrodes manufacturing process. extremity injury or dysfunction. Two test routines were
Two layers of conductive silicone (black) enclose a copper mesh and performed to assess the electrodes monitoring
are bonded to a textile.
capabilities:

1) Signal quality: The test persons were asked to lift five


conductive resistance times their forearm against gravity until elbow flexion
# carrier material of 90 was achieved and to maintain the position for 5
load [/cm]
seconds. Then, the same movements had to be
1 polysiloxane nanoparticles N.N. performed with weights of 1, 2, 3, and 7 kg added to
2 polysiloxane carbon 9 the hand (Fig. 3).
3 polysiloxane carbon 2.8
4 polyamide thread silver coating 0.2
thermoplastic silver-coated 0.01
5
elastomer glass
Table 1. The five different electrode materials evaluated and their
conductive loads and resulting resistances.

Fig. 3. Data acquisition with different loads.

Fig. 2. Different electrodes used in this study. Left side: Dual


Ag/AgCl gel electrodes (=reference electrode). Center: Polysiloxane-
based electrodes, right side: Textile electrodes

Static muscle contraction signals and periods without


B. Evaluation intentional muscle contraction were detected
As a reference, adhesive dual-snap disposable Ag/AgCl automatically. Mean values were computed using the
gel electrodes for surface EMG applications (product Matlab toolbox Gait-CAD [5]. Signal-to-noise ratios
#272, Noraxon Inc., Scottsdale, USA) were utilized. The (SNR) were calculated for each measurement and
diameter of each of the two circular conductive areas is 1 correlations were calculated between the signals of
cm, inter-electrode distance is 2 cm. The other five the Ag/AgCl gel electrode and the other electrodes.
electrodes were applied consecutively with an elastic band
to the biceps muscle belly of six healthy test persons with 2) Electrode characterization by assessing electrode-to-
an inter-electrode distance of 3 cm. The diameter of each skin impedance in a two-electrode configuration. The
of the two circular conductive areas is 2 cm. A three- 50 A measurement signal was provided with a
electrode setup with a copper counter electrode held on frequency of 100 Hz by a precision impedance
the same side was applied. The measurements were analyzer (HP4294A from Agilent, Santa Clara, USA).
completed five minutes after the application of the Measurements of the impedance were repeated 16
electrodes and repeated after 60 min. Data acquisition was times within 3 h.
performed according to the recommendations for non-
invasive surface EMG recording provided by the

301
IV. RESULTS fifth electrode did not correlate with the corresponding
signals of the Ag/AgCl gel electrode.
A. Correlations with Reference Electrode
The signals of the five electrode materials were
B. Signal-to-noise Ratios (SNR)
evaluated in comparison with the signals provided by an
Ag/AgCl gel electrode as a reference. The correlations of As a second criterion to evaluate signal quality, the
the signals measured 5 and 60 min. after application of the ratios of the signal to the background noise were
electrodes are depicted in Figs. 4 and 5, respectively. calculated for the data recorded 5 and 60 min. after
application of the electrodes. The results are given in Figs.
1.0 6 and 7.
0.8
300
0.6
correlation coefficients

0.4 250 high load

signal-to-noise ratios (SNR)


low load
0.2
200
0.0
1 2 3 4 5 6 150
-0.2
test person #
-0.4 100
-0.6
50
-0.8

-1.0 0
1 2 3 4 5 Ref
Fig. 4. Correlations of the five different electrodes with the Ag/AgCl
electrode material #
gel electrode five minutes after application to the skin of the six test
persons. (black bar= material #1, gray bar with dots= material #2, etc.) Fig. 6. Average SNRs recorded with 5 different electrode materials
and the Ag/AgCl reference electrode (Ref) 5 min. after application of
the electrodes to the test persons with low load on the muscle (light
gray bar) and high load on the muscle (dark gray bar).
1.0

0.8

0.6
correlation coefficients

0.4 300

0.2
250
signal-to-noise ratios (SNR)

high load
0.0
low load
1 2 3 4 5 6 200
-0.2
test person #
-0.4
150
-0.6
100
-0.8

-1.0 50

Fig. 5. Correlations of the same five electrodes with the reference 0


electrode one hour after application. (black bar= material #1, gray bar 1 2 3 4 5 Ref
with dots= material #2, etc.) electrode material #

Fig. 7. Average SNRs recorded 60 min. after application of the


electrodes to the test persons, who exerted low muscle contraction
Five minutes after application of the electrodes, highest (light graybar) and high muscle contraction (dark gray bar).
correlations were found between the signals of the
reference electrode and those of the third electrode, which
is reflected by an average correlation coefficient of
=0.98. High average correlations were also found The average SNR of all electrode materials used to
between the data of the Ag/AgCl gel electrode and the record surface EMG signals, except for electrode #5, was
corresponding measurements of the first, second, and in the range of 76:1 to 264:1 when an extra load of 7 kg
fourth electrodes with correlation coefficients of in the had to be lifted (high load). Lower SNRs were calculated
range of 0.91 to 0.95. The signals from the fifth electrode when no extra load had to be exerted by the biceps muscle
did not correlate with the data from the Ag/AgCl gel (low load). The SNR of electrode #1 was highest with
electrode and correlated negatively in test person #4. 16:1, followed by electrode #4 with 14:1. The others,
Comparable correlations were found sixty minutes after including the Ag/AgCl reference electrode, ranged
application of the electrodes. The signals recorded with between 4:1 and 9:1.
the first, second, and third electrode correlated highly with Sixty minutes after application of the electrodes, the
those of the reference electrode, which is reflected by a average SNRs of electrodes #1 and #4 decreased, whereas
coefficient in the range of =0.97 to 0.98. However, the average SNRs of electrodes #2 and #3 and of the
decreased average correlations (=0.76) were found reference electrode increased compared to the results
between the signals of the fourth electrode material and found after five minutes. The SNRs of electrode #5
the reference electrode. Again, the data recorded with the remained on a low level.

302
C. Electrode-to-skin Interface Impedance significantly e.g. when fixation of the electrodes is
The sequential progression of the electrode-to-skin insufficient and motions between the electrode and the
impedances for the different electrode materials is given in patients skin cause disturbance. It was demonstrated by
Fig. 8. Five minutes after application of the electrodes, the [1] that as perspiration builds up and hydrates the skin
electrode-to-skin impedance of the textile electrode (#4) layer artefact levels from movements are lower in dry
and of the reference electrode were below 50 k only. electrodes than in Ag/AgCl electrodes.
Twenty minutes after application, the impedances of all The electrode-to-skin impedance has a major impact on
electrodes except for #2, were in the range of 10 and 50 signal quality in EMG surface recording. According to the
k. After one hour, the impedance of the textile electrode SENIAM recommendations [4], the impedance has to be
increased continuously, whereas the impedances of the determined prior to any EMG surface data acquisition.
other electrodes remained in the range of 30 and 70 k. The impedance is a complex resistance to alternating
currents, and a simplified models of an electrical
400
equivalent circuit are given in [1, 2]. During the first two
hours after application of the electrodes to the test
300 persons` skin, the lowest impedance was found for the
textile electrode (material #4), which was even lower than
Impedance [k]

200 that of the reference electrode. After one hour, however,


the impedance of electrode #4 increased steadily due to
100
drying up of the electrolyte dilution, resulting in a
significant decrease of signal quality.
The electrode-skin impedances of electrode materials
0
0 30 60 90 120 150 180
#1 and #3 decreased to below 50 k within the first 20
time [min] min. after electrode application, whereas the impedances
Fig. 8. Temporal behavior of electrode-skin impedance of 5 different of electrode materials #2 and #5 dropped down to a range
electrode materials within 3 h. (=electrode #1, =electrode #2,
=electrode #3, o=electrode #4, = electrode #5, *= reference
of 40 to 80 k. Nevertheless, signal quality of all three
electrode) Polysiloxane electrodes was good already five minutes
after attaching them to the test persons` biceps muscles,
indicating that sufficient sweat was produced between the
skin and the electrodes that served as an electrolyte. The
V. DISCUSSION
findings comply with the results found by [1, 2] and the
Evaluation of the signal amplitudes depending on the impedances of all electrodes are much lower than those of
load added to the forearm revealed that all three steel or aluminum electrodes [1] which are in the range of
polysiloxane-based electrode materials exhibited high 0.5 M. Additionally, inter-individual differences of
correlations with the signals obtained with an Ag/AgCl gel electrode-to-skin impedances depending on different skin
electrode as a reference. High correlations were found types have to be considered [1].
both five and 60 minutes after application of the
electrodes, indicating robust signal conduction. However, VI. CONCLUSION
high electrical conductivity does not ensure good signal
It was demonstrated in this study that electrodes made
quality. For example, signals of the highly conductive
of three different material combinations of Polysiloxane
elastomer material used for electrode #4 did not correlate
and conductive load produce signals in EMG surface
with the signals measured with the reference electrode and
recordings that correlate highly with the corresponding
the SNRs were the lowest of all electrodes evaluated in
signals of Ag/AgCl gel electrodes. Moreover, two of the
this study. In contrast to this, the conductive textile
assessed electrodes proved to have higher SNRs than the
material delivered a good signal quality as long as the
reference electrode. A textile electrode made of
electrode was saturated with electrolyte dilution. Sixty
conductive yarn also delivered a good signal quality as
minutes after application of the electrodes to the test
long as the electrode was saturated with electrolyte
persons skin, however, signal quality decreased
dilution. All three Polysiloxane electrode materials
significantly due to evaporation of the electrolyte dilution.
assessed in this study are flexible and biocompatible and
Hence, EMG surface monitoring is limited to a short time
can be used for dry signal acquisition without the use of
unless the textile electrodes are drenched repeatedly.
any electrolyte gel or dilution. It is expected that they can
The signals recorded with Ag/AgCl gel electrodes were
be used for monitoring EMG signals in multifunctional
assumed to be the gold standard in this study. However,
prosthetic hands [6]. Additionally these electrodes are
some of the other electrodes evaluated reached better
ideally suited for textile integration being part of a control
SNRs than those calculated for the reference electrode.
system for powered upper limb rehabilitation system [7].
This applied especially when no extra load had to be lifted
Further measurements with different electrode designs and
by the test persons and their biceps muscles exerted only
patients suffering from paraplegia are needed for
low forces resulting in lower EMG signals.
identifying the best suitable surface EMG electrode.
Additionally, it has to be considered that the signals
were recorded under ideal conditions in this study. In
clinical practice the EMG signal quality may decrease

303
ACKNOWLEDGMENT
The authors gratefully acknowledge the valuable help of
Rdiger Rupp of the Orthopedic University Hospital,
Heidelberg, Germany, Christian Rotsch of the Textile
Research Institute Thuringia Vogtland (TITV), Greiz,
Germany, Niels Wessel of the University of Potsdam,
Germany and Roman Ruff of the Fraunhofer Institute for
Biomedical Engineering, St. Ingbert, Germany.
This work is supported by the German Federal Ministry of
Education and Research BMBF grant 01EZ-0774.

REFERENCES
[1] J. Mhlsteff, O. Such: Dry electrodes for monitoring of vital signs
in functional textiles. Proc. of the 26th Annual International
Conference of the IEEE Engineering in Medicine and Biology
Society (EMBS `04), San Francisco, CA, USA, September 1-5,
2004, 2212-2215.
[2] K.-P. Hoffmann, R. Ruff, W. Poppendieck: Long-Term
Characterization of Electrode Materials for Surface Electrodes in
Biopotential Recording. Proc. of 28th Annual International
Conference of the IEEE Engineering in Medicine and Biology
Society (EMBS '06), Aug 30-Sept 3 2006, NY, USA, 2239-2242.
[3] D. Obenauf, H. Hieber, W. Scheibner: Flat textile structure of
bonded insulating textile layers contains microstructured wiring.
German Patent DE19755792, 1999.
[4] H. J. Hermens et al.: SENIAM 8: European Recommendations for
Surface ElectroMyoGraphy. Roessingh Research & Development
b.v., ISBN: 90-75452-15-2, 1999. http://www.seniam.org .
[5] R. Mikut, O. Burmeister, S. Braun, M. Reischl: The Open Source
Matlab Toolbox Gait-CAD and Its Application to Bioelectric Signal
Processing. Proc. of the DGBMT Workshop Biosignalverarbeitung,
July 16-18, 2008, Potsdam, Germany, 109-111; 2008.
[6] C. Pylatiuk, S. Schulz, A. Kargov, G. Bretthauer: Two Multi-
articulated Hydraulic Hand Prostheses. Artificial Organs
28(11):980-986, 2004.
[7] R. Rupp, U. Eck, O. Schill, M. Reischl, S. Schulz: OrthoJacket
An active FES-hybrid orthosis for the paralyzed upper extremity.
2nd European Conference Technically Assisted Rehabilitation
(TAR 2009), Berlin, March 18-19, 2009.

304

You might also like