1) Make sure that the following errors / alarms are not present before running the T1 test.
a) Power Failure
d) Connect conc.Line
e) Blood sensed by OD
f) Flow alarm
g) Water alarm
2) When T1 test fails, press Test key again and note the displayed error code before proceeding
to technical manualfor further clues.
3) If more than 1 test fails select Test Service ON (PCB LP 632 DIP switch array 1, SW 4 ON) & do the
test individually to get the error codes of the concerned test fail.
Test Equipments:
a) Hemodialysis tester
b) Pressure guage
d) Flashlight.
e) Digital multimeter
Pressure Units:
1Bar = 14.5 psi =749.94mmHg
1) Water Alarms
2) Flow Alarms
4) Temperature
5) Conductivity
6) Blood Leak
b) Pressure
Regulator (61) c)
Check the Inlet
Filter c) Check Valve 41
d) Water Inlet
Valve(41)
2 Flow Alarm a) Blocked filter a) Verify flow 300, a) Create flow from
500, 800 diagnostics.
b) Defective
valve/pump b) Calibrate if b)Verify Loading
required. pressure (1.2 1.45
(valve resistance Bar)
50 to 80 Ohms)
c) Calibration
b) Degassing
pressure
c) Flow calibration
a) Temp Calib
a) Flow errors can
ii) Low lead to Low Temp
temperature Alarm!
b) Heater Rod.
b) Heater rod
resistance around
30 Ohms
S NO. Error Possible Causes Tips Remarks
5 Conductivity
f) Faulty Conc or
BIC Pump
d) Temperature
c) Concentrate or and Conductivity
Bicarbonate (display)
Calibration
d) Conductivity
(Display )
Calibration
S NO. Error Possible Causes Tips Remarks
b) Dial Pressure
Calibration b)Continuous TMP
Required. Alarm / Pressure
fail indicates Dial a)Adjust 0 mmHg (
Pressure LP 633 Pot
Calibration P3) and do dial
pressure
calibration.
8 Pressure Holding
Test
a) UF filter
blockage b) Use the
graduated
ii) Negative measuring Jar.
Pressure Errors b) UF Pump
Calibration c) Observe/
Replace Spring set
c) Broken UF Pump d) Check tightness
Spring set of hoses
d) Use Diagnostics
program
d) Minute Leakage
in Hydraulic system
9 Hydraulic Test Fail a) Rinse Filters 148 a) Replace Filters if a) 3 minutes before
& 149 needed. the end of cleaning
program.
b) Most cases
i)Rinse Failure pressure leak due
Alarm b)Rinse Valves to loose hose at
91,99,100 these valves
c) Check PS 134
from Diagnostics.
c) Bibag Pressure
switch 134
a) Clogged
a) 1 minute before
Filter 210 the end of cleaning
program.
b) Jammed Motor
b) Replace bearing
a) Air Detector
a) Calibrate Air
detector
ii)Level detector
a) Arterial Pressure
Alarm
calibration
b) Venous Pressure
calibration
a) Mechanical
problem of gear
c) Replace Diasafe
filter.
11 Ultrafiltration
Errors.
i) Less fluid
a)UF Pump filter
removal
(74)
b) UF Pump
Calibration b) Verify b) Verify the stroke
Calibration volume(1ml/
c) UF pump spring stroke)
c)Replace if
set
required
d) UF Pump
d) Replace if
defective.
b) BatteryFuse
(3.15 A)
2) Enter Diagnostics- Write outputs Write Digital outputs CPU 1 : Digital outputs.
Note: Balancing Chamber inlet /outlet valve combination can be switched off one group at a time
to locate the faulty BC valve, if any.