General Guidelines:

1) Make sure that the following errors / alarms are not present before running the T1 test.

a) Power Failure

b) Dialines not conn

c) Shunt cover open

d) Connect conc.Line

e) Blood sensed by OD

f) Flow alarm

g) Water alarm

2) When T1 test fails, press Test key again and note the displayed error code before proceeding
to technical manualfor further clues.

3) If more than 1 test fails select Test Service ON (PCB LP 632 DIP switch array 1, SW 4 ON) & do the
test individually to get the error codes of the concerned test fail.

Test Equipments:

a) Hemodialysis tester

b) Pressure guage

c) Graduated Measuring jar (100ml ,1000ml)

d) Flashlight.

e) Digital multimeter

Pressure Units:
1Bar = 14.5 psi =749.94mmHg

The Order of Troubleshooting :

1) Water Alarms

2) Flow Alarms

3) Upper Flow Alarm

4) Temperature

5) Conductivity

6) Blood Leak

7) Pressure Holding Test Failures.

8) TMP Continuous Alarms

9) Hydraulic Test Fails

10) Blood Pump Errors / Level Detector Alarms

/Arterial Alarms/ Venous Alarms

/Heparin Pump Errors/ Diasafe Errors

11) Ultra Filtration Errors

12) Battery Fail


S NO. Error Possible Causes Tips Remarks

1 Water Alarm a) No Water a) Water UseDiagnostics


Supply/ Low InletPressure(0.95 Program to check
Pressure. to 1.05 Bar) the water Flow Path.

b) Pressure
Regulator (61) c)
Check the Inlet
Filter c) Check Valve 41
d) Water Inlet
Valve(41)

2 Flow Alarm a) Blocked filter a) Verify flow 300, a) Create flow from
500, 800 diagnostics.
b) Defective
valve/pump b) Calibrate if b)Verify Loading
required. pressure (1.2 1.45
(valve resistance Bar)
50 to 80 Ohms)

c) Calibration

3 Upper Flow Alarm a) Degassing filter

b) Degassing
pressure

c) Flow calibration

4 Temperature a) Incomingwater a)Incomplete hot


temperature disinfection can be a
i) High
reason!
temperature b) Temp Calib

a) Temp Calib
a) Flow errors can
ii) Low lead to Low Temp
temperature Alarm!
b) Heater Rod.
b) Heater rod
resistance around
30 Ohms
S NO. Error Possible Causes Tips Remarks

5 Conductivity

i) Low Conductivity a) Dilution Ratio. a) Verify Dilution a)Normally


Ratio in Set Up 1+35.83(Nacl26);
Menu
depends on the A &
B part
b)Concentrate or b) Wrong or b) Provided the
Bicarbonate Expired Solution?? concentrate and BIC
strokes and suction
is OK
c) Calibrate if c) Ref: Calibration-
c) Conc. & BIC required Mixing System
pump strokes and
output??
d) Calibrate the
d) Replace if Conc and BIC pumps
d) Conc and BIC after replacing the
required
pump spring set springs. Use the
broken?? graduated
measuring Jar to
measure the stroke
e) Conductivity volume.
(Display )
e) Temperature
Calibration
and Conductivity
(display)
Calibration

f) Faulty Conc or
BIC Pump

a)Flow Errors a) Verify flow


,Calibrate if
required
b) Verify Dilution

ii) High b)Dilution c) Ref: Set up-


Conductivity Dilution
c) Wrong or
Expired Solution??

d) Temperature
c) Concentrate or and Conductivity
Bicarbonate (display)
Calibration

d) Conductivity
(Display )
Calibration
S NO. Error Possible Causes Tips Remarks

6 TMP / Dial Pressure a) Dialyser a) Replace Dialyser.


membrane defects

b) Dial Pressure
Calibration b)Continuous TMP
Required. Alarm / Pressure
fail indicates Dial a)Adjust 0 mmHg (
Pressure LP 633 Pot
Calibration P3) and do dial
pressure
calibration.

7 Blood Leak Blood Leakand Adjust both to 5


Dimness Volts.
Calibration

8 Pressure Holding
Test

a) Leakage in a) Observe for a) Use a flash light.


Hydraulic system. leaks during
Positive pressure
i) Positive Pressure test or Induced
Errors b) Do Dial Pressure
b) Dial Pressure pressure test. calibration only if
Calibration the error code
indicates dial
pressure
calibration.

a) UF filter
blockage b) Use the
graduated
ii) Negative measuring Jar.
Pressure Errors b) UF Pump
Calibration c) Observe/
Replace Spring set
c) Broken UF Pump d) Check tightness
Spring set of hoses
d) Use Diagnostics
program

d) Minute Leakage
in Hydraulic system

S NO. Error Possible Causes Tips Remarks

9 Hydraulic Test Fail a) Rinse Filters 148 a) Replace Filters if a) 3 minutes before
& 149 needed. the end of cleaning
program.
b) Most cases
i)Rinse Failure pressure leak due
Alarm b)Rinse Valves to loose hose at
91,99,100 these valves

OR Leak from Conc


and BIC pick up line
rubber seals.

c) Check PS 134
from Diagnostics.

c) Bibag Pressure
switch 134

a) Clogged
a) 1 minute before
Filter 210 the end of cleaning
program.

ii) V39 Failure.


10 i) Blood Pump a) BP Power cord a) Replace power a) Power cord is
Errors cord the one connecting
control board to
motor

b) Jammed Motor
b) Replace bearing

c) Control Board c) Do complete


arterial calibration
and BP speed
calibration after
replacing
the control board

a) Air Detector

a) Calibrate Air
detector
ii)Level detector
a) Arterial Pressure
Alarm
calibration

iii) Arterial Pressure


Alarms
S NO. Error Possible Causes Tips Remarks

10 Contd iv) Venous Pressure a) Transducer Partially clogged


Alarms Protector ( P Ven) female connection.

b) Venous Pressure
calibration

c) Venous Line set.

a) Mechanical
problem of gear

v) Heparin Pump a) Manually rotate


Errors the gear to Top or
bottom end and
retest the pump
a)F 03/ F02
a) Dial Pressure
calibration
required
vi) Diasafe Errors
b)F 34/ 43 b) Damaged
Diasafe Filter
a)
c) Diasafe Pressure
Transducer

c) Replace Diasafe
filter.
11 Ultrafiltration
Errors.

i) Less fluid
a)UF Pump filter
removal
(74)

b) UF Pump
Calibration b) Verify b) Verify the stroke
Calibration volume(1ml/
c) UF pump spring stroke)
c)Replace if
set
required
d) UF Pump
d) Replace if
defective.

12 Battery Fail. a) Battery a) BatteryVol. 18 V


.

b) BatteryFuse
(3.15 A)

c) LP 647 c) LP 647 is in the


Power supply
module.
Diagnostics Program

1) On the service switch

2) Enter Diagnostics- Write outputs Write Digital outputs CPU 1 : Digital outputs.

Rinsing V 84 ( Disinfection Valve)

1) On V 87, 30, 32, 34, 36, 38, 84 (On - 1111)

2) Start Flow Pump (FP On - 0000)

Creating the Flow

1) Switch On V 24, 24 B, 30,31,32,33,34,35,36,37,38,87

2) Start Flow Pump and Degassing pump

3) Set On FL_ SW_ EN (Float Switch Enable)

Note: Balancing Chamber inlet /outlet valve combination can be switched off one group at a time
to locate the faulty BC valve, if any.