RECOMMENDATION FOR ANNUAL SERVICE OF


PLANMECA COMPACT i DENTAL UNIT
Unit maintenance kit order code 10010264, Microvac maintenance kit order code 10011453.

Customer: Serial number:


Installation location: Installation date:
Technicians name:

Comments from the user(s) :

* Upgrade
Upgrade the software (recommended) Main control CPU version old: ________, new:_________.
Control panel software version old: __________, new __________

* Delivery arm
Check the friction of the delivery arm joints Adjust, if needed.
Check the balance of the delivery arm Adjust, if needed.

* Suction arm
Check the friction of the suction arm
Check the function of the suction holder sensors In the service modes 50 and / or 51.
Check the function of the control panel In the service modes 48 and 49.

* Instrument console and instruments


Clean the oil cup Replace the return air filter and o-ring.
Check the function of the control panel In the service modes 47 and 49.
Replace the spray water proportional valve and check the
operation of the addtional water valve
Replace multiplexer control membranes
Check the operation of the syringe water valve Clean or replace the valve, if necessary. Yes / No
Check the condition of the instrument hose. If the outer surface is
damaged, the hose needs to be replaced.
Check the instrument hoses and the quick connectors for leaks Replace the sealing rubbers of the quick connectors.
Check that the hole in the instrument hose outer surface is open. Open/make a hole, if needed (refer to tech bulletin D-167).
Check the operation of the instruments and service them
according to the manufacturer's instructions Replace the o-rings, if needed. Yes / No

* Foot control

Replace the pedal knob, cover knob and bottom knobs


Check the operation of the pedal Adjust and calibrate, if needed. Yes / No

* Cuspidor
Replace the filter and membrane of the air regulator
Replace the filter, membrane and spindle of the water regulator Lubricate the spindle.
Adjust the internal air and water pressures In the service mode 38 (air 5,5 bar, water 2,52,8 bar).
Replace the o-ring in the bowl assembly.

* Patient chair
Clean and lubricate the lift motor worm screw Use vaseline.
Clean and lubricate the backrest motor worm screw Use vaseline.
Check the operation of the safety switches and
emergency stop button In the service mode 75.

Planmeca Oy Tel: +358-20 7795 500


Asentajankatu 6 Fax: +358 20 7795 555
00880 Helsinki, Finland E-mail: aftersales@planmeca.com 1 (2)
Customer Service Information

AFTER SALES

* Upholsteries
Check the attachment of the upholsteries Replace Velcro tapes, if necessary. Yes / No

* Suction system (all)


Replace bowl filter, coarse filter and amalgam collector
Replace the suction tubes
Check the condition of the suction nozzles Replace, if necessary. Yes / No
Clean the separator assembly For other than Microvac, refer to manufacturers instructions.

* Microvac II separating tank


Replace the intermediate membrane assembly
Replace the drain membrane assembly
Replace the pressure balancing membrane
Clean the Microvac II separating tank
Other suction systems: Refer to manufacturer's instructions

* Waterline Cleaning System (WCS)


Adjust the feeding pressure In the service mode 38 (water 1,5 bar).
Check the operation of the WCS system

* Waterline Management System (WMS)


Check and adjust the sensors
Check and adjust the water flow In the service mode 127

* Suction Tube Cleaning System (STCS)


Check the operation of the system
Calibrate the water flow and feeding time if needed Service mode n.150 for feeding time calibration

* Operating light (Delight)


Check the focus of the light field Adjust, if necessary.
Clean the protective fascia Replace, if necessary. Yes / No
Clean the reflector Replace, if necessary. Yes / No

* Final inspection
Check the operation of the unit generally
Check the condition of all the surfaces and clean them Replace, if necessary.

* Notes

Technicians signature: Date:


Customers signature: