AFTER SALES
* Upgrade
Upgrade the software (recommended) Main control CPU version old: ________, new:_________.
Control panel software version old: __________, new __________
* Delivery arm
Check the friction of the delivery arm joints Adjust, if needed.
Check the balance of the delivery arm Adjust, if needed.
* Suction arm
Check the friction of the suction arm
Check the function of the suction holder sensors In the service modes 50 and / or 51.
Check the function of the control panel In the service modes 48 and 49.
* Foot control
* Cuspidor
Replace the filter and membrane of the air regulator
Replace the filter, membrane and spindle of the water regulator Lubricate the spindle.
Adjust the internal air and water pressures In the service mode 38 (air 5,5 bar, water 2,52,8 bar).
Replace the o-ring in the bowl assembly.
* Patient chair
Clean and lubricate the lift motor worm screw Use vaseline.
Clean and lubricate the backrest motor worm screw Use vaseline.
Check the operation of the safety switches and
emergency stop button In the service mode 75.
AFTER SALES
* Upholsteries
Check the attachment of the upholsteries Replace Velcro tapes, if necessary. Yes / No
* Final inspection
Check the operation of the unit generally
Check the condition of all the surfaces and clean them Replace, if necessary.
* Notes