RIDAQUICK Malaria Control Control+

Art. No.: NP7004 C

1. Intended use 7. Reagents and storage conditions


For in vitro diagnostic use. The RIDAQUICK Malaria Control must be stored at 2-8 C and can be
The RIDAQUICK Malaria Control is intended for use in the corres- used up to the expiry date printed on the vial. We do not guarantee for the
ponding RIDAQUICK Malaria rapid test (N7006/N7007). product or results once the stated shelf-life has expired.
2. Explanation 8. Indications of instability or reagent deterioration

The RIDA QUICK Malaria Control is available as a separate accessory for Unspecifically coloured T1 or T2-band or eventually additional appearing
the RIDAQUICK Malaria rapid test. It contains recombinant HRP2 protein bands on the test membrane of the cassette indicate that the reagents
(reactive at T1-band) and plasmodial-specific LDH protein (reactive at T2 may have degraded.
band. Prior to use, the positive control as well as the respective test- 9. Preparing the positive control
specific reagents (reagents, test cassettes) must be left to stand until they
reach room temperature (20-25C). The test cassettes should be removed The RIDAQUICK Malaria Control is ready to use. It must be handled
from their packages only briefly before they are required for use. Take analogously to blood sample in the RIDAQUICK Malaria test cassette.
care to avoid direct exposure to sunlight while performing the test. 10. Test procedure
3. Test principle Using one of the micropipettes (MicrosafeTubes), provided with the test
The RIDAQUICK Malaria Control is ready to use and reacts specifically at kits (N7006 or N7007) 5 l of RIDAQUICK Malaria Control are with-
both bands, marked with T1 and T2 on test cassette and yields the drawed capillary by simply dipping and without pressing the pipette bulb
specifically coloured reaction bands (control- and test bands) after the end and then control are added in the funnel of the cassette which is marked
of the reaction as described in point 11. by Blood. The pipette tip must touch the membrane surface in order to
secure the transfer of 5 l of the Malaria control reliably. After then 4 drops
4. Material provided of reagent provided by one of the RIDAQUICK Malaria testkits (N7006 or
1 x vial RIDAQUICK Malaria Control (0.12 ml). Ca. 24 control reactions N7007) in the funnel which is marked by Reagent. The valuable result
can be carried out. can be read after 30 minutes.
5. Material required but not provided 11. Interpretation of results
Corresponding RIDAQUICK Malaria Kit (N7006 or N7007). Beside the control line C also the test line T1 and T2 must be visible
for valid result. If only control line C appears, it means that positive
6. Precautions control is declined. If beside the control line C only one of both testlines
The RIDAQUICK Malaria Control contains recombinant plasmodial- is visible, it may indicate that one of both recombinant proteins (HRP2 or
specific antigens. This notwithstanding, the product should be treated as pLDH) is deteriorated because of instability.
potentially infectious material.