You are on page 1of 3

Dynamic and real-time display of running status

Running status of reagent tray, sample tray and reaction tray


BS-380
Real-time monitoring of reagent residual volume
Chemistry Analyzer
Real-time diagnosis of system working status
Technical Specifications
System Function: Reagent probe: Liquid level detection, vertical and horizontal
Fully automated, Discrete, Random access for collision protection
routine, STAT, Urine and homogeneous immuno Probe cleaning: Interior and exterior automatic washing
assays; Carry-over< 0.1%
STAT sample priority
Throughput: 300 tests / hour, up to 450 tests / hour with ISE Reaction System:
Measuring principles: Reaction rotor: Rotating tray, 72 cuvettes with automatic washing
Absorbance photometry, Turbidimetry Cuvette: Optical length of 5mm
Methodology: End-point, Fixed-time, Kinetic, optional ISE Reaction volume: 150~360l
Single / Dual reagent chemistries, Operating temperature: 37C
monochromatic / bichromatic Temperature fluctuation: 0.1C
Linear / non-linear calibration Mixing system: Integrated mixers system
Programming: User defined profiles and calculation
Optical System:
Sample Handling: Light Source: Halogen-tungsten lamp
Sample tray: 75 positions for primary or secondary tubes and Photometer: Reversed optics, grating photometry
Flexible and convenient software function
sample cups Wavelength: 340nm380nm412nm450nm505nm
Template Modifying Software Sample volume: 2~45l, step by 0.1l 546nm570nm605nm660nm700nm
Sample probe: Liquid level detection, vertical and horizontal 740nm800nm
Flexible to define various report templates for laboratory, and easy to import new template from Mindray or other users. collision protection Absorbance range: 0~3Abs
Probe cleaning: Interior and exterior automatic washing Resolution: 0.001Abs
Monitoring of various Samples
Carry-over< 0.1%
It is method to perform continuous glucose monitoring in subjects with impaired glucose tolerance or other multiple samples monitoring. Automatic sample dilution: Control and Calibration:
Pre-dilution and post-dilution ratio up to 1:150 Calibration mode: Linear (one-point, two-point and
Fast Emergency Detection Dilution vessel: Cuvette multi-point), Logit-Log 4P, Logit-Log 5P, Spline,
Exponential 5P, Polynomial 5P, Parabola
One-key STAT function makes it easy to perform an emergency test for operator.
Internal Bar Code Reader (optional): Control rules: Westgard multi-rule, Cumulative sum check,
Used for sample and reagent programming; Twin plot
applicable to various bar code systems
including CodabarITFCode128 Operation Unit:
Code 39UPC / EANCode 93. Operation system: Windows XP Professional/Home SP2 or above
Bi-directional LIS interface Windows Vista Home/Business, Windows 7
Interface: RS-232
ISE Module (optional):
Optional selection of K+, Na+, Cl
Throughput: Up to 225 tests per hour
Working Conditions:
Power Supply:
Temperature:
AC 100~240V, 50 / 60Hz, 1000VA
15~30C
BS-380
Reagent Handling:
Reagent tray: 60 positions in refrigerated compartment (2~8C)
Humidity: 35~85%
Water consumption: 10L/hour
Chemistry Analyzer
Reagent volume: R1: 150~350l, step by 1l; Dimension: 990mm x 693mm x 1135mm (WxDxH)
R2: 20~200l, step by 1l; Weigh: 200 Kg

Original reaction data record

Real-time monitoring of reactions

Simultaneously display reaction curves under primary and secondary wavelengths is a trademark of Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Mindray Building, Keji 12th Road South,
2013 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. Specifications subject to changes without prior notice.
Detailed profile of alert messages High-tech Industrial Park, Nanshan, Shenzhen 518057, P.R. China
P/N: ENG-BS380-21285x8-20130303
Tel: +86 755 8188 8998 Fax: +86 755 26582680
E-mail: intl-market@mindray.com www.mindray.com
Mindray is listed on the NYSE under the symbolMR
BS-380 BS-380
Chemistry Analyzer Chemistry Analyzer

Discrete, random access, fully automated


Constant 300 tests per hour, up to 450 tests per
Multi-functional sample tray Multi-functional sampling probes
hour with ISE
75 sample positions Interior and exterior probe washing
Onboard capacity of 75 sample positions, up to Can be programmed into 5 virtual sample disks Liquid level detection
Primary tubes and various sample cups can be used Collision protection
58 chemistries plus 3 ions
Automatic dilution of high concentration sample Inventory monitoring
Refrigerated reagent compartment Pre-dilution for samples Probe depth adjusted automatically
Internal sample bar code reader (optional)
Automatic probe cleaning, liquid level detection,
vertical and horizontal collision protection
8-step auto washing system
12 wavelengths: 340~800nm
Automatic dilution for abnormal sample
Internal bar code reader (optional)
Bi-directional LIS interface

Reaction tray Washing station


Test sequence optimization High-tech washing station to ensure
150 L minimum total reaction volume accurate results and valid diagnostics
Contains 72 reaction cuvettes Concentrated detergents to reduce carry-over
Maintenance free heating system 8-step auto washing station

High performance mixer design Refrigerated reagent tray


Integrated mixers system 58 reagent positions for R1, R2
Optimized homogenization 24 hour non-stop cooling with Peltier element
High precision mixer station to ensure excellent Ready-to-use liquid stable reagents
reaction conditions Internal reagent bar code reader (optional)
Standardized mixing procedures
Separate mixer for individual reaction steps
Mindray solution for clinical chemistry
Chemistry Reagents
After more than 10 years of research and development on reagents, Mindray can now provide 48
Hepatic Lipids
parameters of dedicated reagents(more than 17 others are coming), covering hepatic, renal, cardiac, Alanine Aminotransferase (ALT) Total Cholesterol (TC)
Aspartate Aminotransferase (AST) Triglycerides (TG)
lipids, diabetes, pancreatitis, inorganic ions and immunalassays, etc.,together with original calibrators
Alkaline Phosphatase (ALP) HDL-Cholesterol (HDL-C)
with metrological traceability as well as controls for BS-380 chemistry analyzer. -GlutamylTransferase (-GT) LDL-Cholesterol (LDL-C)
Direct Bilirubin (D-Bil) DSA Method Apolipoprotein A1 (ApoA1)
Direct Bilirubin (D-Bil)VOX Method Apolipoprotein B (ApoB)
Total Bilirubin (T-Bil) DSA Method Lipoportein(a) [LP(a)]
Total Bilirubin (T-Bil)VOX Method
Total Protein (TP) Pancreatitis
Albumin (ALB) -Amylase (-AMY)
Total Bile Acids (TBA) Lipase (LIP)
Auto Chemistry
Prealbumin (PA)
Analyzer
Original Quality Cholinesterase (CHE) Diabetes
Control Adenosine deaminase (ADA) * Glucose (Glu) GOD-POD Method
-L-fucosidase (AFU) * Glucose (Glu) HK Meth
5'-nucleotidase (5'-NT) * Hemoglobin A1c (HbA1c)
Mindray Fructosamine (FUN)
solution for Renal
clinical chemistry Urea (UREA) Inorganic ions
Creatinine (CREA) Modified JaffMethod Calcium (Ca)
Creatinine (CREA)Sarcosine OxidaseMethod Magnesium (Mg)
Uric Acid (UA) Phosphate Inorganic (P)
Carbon dioxide (CO2)
Microalbumin* Rheumatism
2-Microglobulin (2-MG) * High sensitivity C-reactive protein (hs-CRP) *
Cystatin C (CysC) * Rheumatoid Factor (RF)
Original System
Antibodies Against Streptolysin O (ASO)
Pack Reagent Original
Cardiac
Calibrators
Creatine Kinase (CK) Immune
with Traceability
Creatine Kinase-MB (CK-MB) Immunoglobulin A (IgA)
Lactate Dehydrogenase (LDH) Immunoglobulin G (IgG)
Original Calibrators with traceability:
-Hydroxybutyrate Dehydrogenase(-HBDH) Immunoglobulin M (IgM)
Homocysteine (HCY) Immunoglobulin E (IgE) *
Reference Method (Certified by Joint Committee for Traceability in Laboratory Medicine (JCTLM))
Myoglobin* Complement C3 (C3)
International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)
Complement C4 (C4)
National Institute of Standards and Technology(NIST)
Ferrum C-Reactive Protein (CRP)
Centers for Disease Control and Prevention (CDC, USA) Iron (Fe)
American Association for Clinical Chemistry (AACC) Ferritin (FER) * Others
Transferrin (TRF) * Glucose-6-phosphate dehydrogenase (G6PD) *
Reference Material Total iron binding capacity / unsaturated ironBinding D-dimer*
Institute for Reference Materials and Measurements (IRMM) standards capacity (TIBC/UIBC) * Angiotensin converting enzyme (ACE) *
National Institute of Standards and Technology (NIST) standards Retinol binding protein (RBP) *
World Health Organization (WHO) standards D3-hydroxybutyric acid (D3-HB) *

Japan Committee for Clinical Laboratory(JCCLS) standards


* Coming soon

You might also like