Professional Documents
Culture Documents
01
MGA3000C
MULTI-GAS
ANALYSER
OPERATION MANUAL
Version 3.01
1
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
Protective Clothing
Protective clothing must always be worn when working in the vicinity of hot vessels or ducts.
Storage
The instrument should be stored in its packaging, in a dry and sheltered area.
Unpacking
Check all packaging for external signs of damage. Check the contents against the packing
note.
Any item returned to ADC Gas Analysis Ltd should be adequately packed to prevent damage
during transit.
You must include a written report of the problem together with your own name and contact
information, address, telephone number, email address etc.
Lifting Instructions
Where items are too heavy to be lifted manually, use suitably rated lifting equipment. Refer
to the Technical Specification for weights. All lifting should be done as stated in local
regulations.
2
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
The protection provided by both CE and IP classifications to this product may be invalidated if
alterations are made to the structural, electrical, mechanical or pneumatic parts of this
system. Such changes may also invalidate the standard terms of warranty.
Dimensions
Unless otherwise stated, all measurements are given in millimetres and inches.
Copyright
This manual is provided as an aid to owners of ADC Gas Analysis products and contains
information proprietary. This manual may not, in whole or part, be copied, or reproduced
without the expressed written consent of ADC Gas Analysis Ltd
3
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
If further assistance is required please contact your nearest Gas Analysis Led office the
following addresses or your local distributor (see our website for details http://www.adc-
service.co.uk)
ADC Gas Analysis Ltd has a comprehensive range of analysers and Environmental Monitoring
Instrumentation.
4
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
CONTENTS
Contents .............................................................................................................. 5
1. Introduction...................................................................................................... 8
1.1 About the Instrument .............................................................................................. 8
1.2 Instrument Identification.......................................................................................... 8
1.3 The Display & Operating Controls .............................................................................. 8
1.3.1 The Display ...................................................................................................... 8
1.3.2 Design ............................................................................................................. 9
1.3.3 The Operating Controls ...................................................................................... 9
1.4 Electrical Outputs .................................................................................................... 9
1.4.1 Analogue Outputs ............................................................................................. 9
1.4.2 Alarm(trips) outputs .......................................................................................... 9
1.4.3 RS232 port ....................................................................................................... 9
1.5 Gas Path .............................................................................................................. 10
1.5.1 Gas Flow Diagram ........................................................................................... 10
1.5.2 Key to gas path figures .................................................................................... 11
1.5.3 Gas connections .............................................................................................. 11
1.5.4 MGA 3000 19 rack, infrared component with oxygen measurement .................... 12
1.6 Dimension drawings .............................................................................................. 13
1.7 Circuit diagrams (electric and gas connections) ........................................................ 14
1.7.1 RS-232C I/O (see section 2.4).......................................................................... 14
1.7.2 Trip I/O(see section 2.6) .................................................................................. 14
1.7.3 Analogue Output(see section 2.5) ..................................................................... 15
1.8 Operating Principles ............................................................................................... 16
1.8.1 Infrared Absorption ......................................................................................... 16
1.8.2 The Optical Bench In Use ................................................................................. 16
1.8.3 The Oxygen Sensor in Use ............................................................................... 17
2. Installation ..................................................................................................... 19
2.1 Installation Arrangement........................................................................................ 19
2.2 Location & Ambient Conditions ............................................................................... 19
2.3 Power Supply Setting & Connections ....................................................................... 19
2.3.1 Power Supply Setting ...................................................................................... 19
2.3.2 Power Supply Connections ............................................................................... 20
2.4 Serial I/O (RS232C) .............................................................................................. 20
2.5 Analogue Outputs.................................................................................................. 21
2.6 Alarm (Trip) Outputs ............................................................................................. 22
2.7 Gas Arrangements................................................................................................. 22
2.7.1 The Gas Circuit ............................................................................................... 22
2.7.2 The Sample Gas.............................................................................................. 23
2.7.3 The Span Gas ................................................................................................. 23
2.7.4 The Zero Gas .................................................................................................. 23
2.7.5 The Purge/bypass Gas ..................................................................................... 24
2.7.6 The Exhaust Gas ............................................................................................. 24
2.7.7 Gas Pump(s) .................................................................................................. 24
2.8 Use of Chemicals (certain models only) ................................................................... 24
2.8.1 Water Vapour ................................................................................................. 24
2.8.2 Carbon Dioxide ............................................................................................... 25
2.8.3 Precautions .................................................................................................... 25
2.9 Pre-operation Checks ............................................................................................. 25
2.10 General Advice .................................................................................................... 25
2.10.1 Use of Gas Cylinders ...................................................................................... 25
2.10.2 Gas Leaks..................................................................................................... 26
2.10.3 Exhausting & Ventilation ................................................................................ 26
5
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
3. Setting Up ...................................................................................................... 27
3.1 Preview ................................................................................................................ 27
3.2 On Power Up......................................................................................................... 27
3.3 Zero & Span Operations. ........................................................................................ 27
3.4 Display & Keypad .................................................................................................. 28
3.4.1 Display........................................................................................................... 28
3.4.2 Function Keys ................................................................................................. 29
3.4.3 Menu tree....................................................................................................... 30
3.5 Analogue (Chart) Outputs ...................................................................................... 31
3.6 Alarm (Trip) Outputs ............................................................................................. 31
3.6.1 Setting Alarm Outputs ..................................................................................... 31
3.7 RS232C Serial Port ................................................................................................ 32
3.7.1 Protocol.......................................................................................................... 32
3.7.2 Serial Port Settings.......................................................................................... 32
3.8 Measurement ........................................................................................................ 33
4. Calibration ...................................................................................................... 34
4.1 Initial Checks ........................................................................................................ 34
4.1.1 Warm Up........................................................................................................ 34
4.2 Zero Adjustment ................................................................................................... 34
4.2.1 Manual Zero ................................................................................................... 35
4.2.2 Auto Zero ....................................................................................................... 35
4.3 Span Adjustment .................................................................................................. 35
4.3.1 Setting Span................................................................................................... 36
4.3.2 Auto Span ...................................................................................................... 36
5. Maintenance ................................................................................................... 38
5.1 Routine Checks ..................................................................................................... 38
5.1.1 Water Vapour ................................................................................................. 38
5.1.2 Dust .............................................................................................................. 38
5.2 Instrument Checks ................................................................................................ 39
5.2.1 Source Block Purge (when fitted) ...................................................................... 39
5.2.2 Filter Column (when fitted) .............................................................................. 39
5.2.3 Sample Flow ................................................................................................... 39
5.2.4 Throttles ........................................................................................................ 39
6. Optical Bench Descriptions ............................................................................... 40
6.1 Introduction .......................................................................................................... 40
6.2 GC Bench ............................................................................................................. 40
7. Electrical Specifications .................................................................................... 42
7.1 Power Supply ........................................................................................................ 42
7.2 Analogue Outputs.................................................................................................. 42
7.2.1 Set for current output ...................................................................................... 42
7.3 Alarm (Trip) Contacts ............................................................................................ 42
7.4 Serial (RS 232) Port .............................................................................................. 42
8. Spare Parts..................................................................................................... 43
9. AK Remote Control Protocol ............................................................................ 44
9.1 Introduction .......................................................................................................... 44
9.2 Protocol Overview ................................................................................................. 44
9.3 Address ................................................................................................................ 45
9.4 Commands and Parameters ................................................................................... 45
9.5 Response To Commands ........................................................................................ 45
9.5.1 Status Codes .................................................................................................. 46
9.6 Command descriptions........................................................................................... 46
9.6.1 Interrogation Commands ................................................................................. 46
9.6.2 Show model and version information ................................................................ 46
6
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
9.6.3 Show range information................................................................................... 46
9.6.4 Show Calibration Information ........................................................................... 47
9.6.5 Show sample gas inlet ..................................................................................... 48
9.6.6 Show analyser gas flow and sample pump state ................................................. 48
9.6.7 Show Alarm and Valve States........................................................................... 48
9.6.8 Show analyser's gas name(s) ........................................................................... 49
9.7 Remote Control Commands .................................................................................... 49
9.7.1 Set remote control mode ................................................................................. 49
9.7.2 Set local control mode ..................................................................................... 49
9.8 Measurement Enquiry Commands ........................................................................... 50
9.8.1 Get current readings........................................................................................ 50
9.9 Range Selection .................................................................................................... 50
9.9.1 Select active range .......................................................................................... 50
9.10 Gas Inlet Control Commands ................................................................................ 50
9.10.1 Select sample inlet ........................................................................................ 50
9.10.2 Select zero gas ............................................................................................. 51
9.10.3 Select span gas ............................................................................................. 51
9.10.4 Sample pump control .................................................................................... 51
9.11 Calibration Commands ......................................................................................... 51
9.11.1 Perform auto-zero ......................................................................................... 51
9.11.2 Set span gas concentration ............................................................................ 52
9.11.3 Perform span calibration ................................................................................ 52
10. Implementation notes.................................................................................... 54
11. Known Issues ............................................................................................... 55
7
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
1. INTRODUCTION
IMPORTANT
Before operating the Instrument for the first time, ensure all the relevant
installation checks have been carried out [refer to Section 2.9].
Single, dual, triple and 4 gas versions are available utilising one or more IRGA optical
benches and/or chemical cells.
Most instruments incorporate IRGA benches that generate a signal proportional to the
infrared absorption of the measured gas. Providing the gas concentration lies within the
designed range of the instrument, the signal is periodically compared with reference gas
(usually at zero and a convenient mid-span condition). Calibration is achieved manually via
simple front panel controls or fully automatically within the software configuration.
Some instruments may include a chemical cell which reacts with the specific gas and which
provides an electrical output in proportion to the reaction.
Gas concentrations are indicated on a front display panel together with both analogue and
digital signal outputs.
Gas connections to and from the instrument are via suitable fittings on the rear panel. The
gas flow is maintained via an internal sample pump.
The MGA3000's electrical supply can be with any voltage from 90-240V AC and 50/60Hz
Always quote the serial number and model number in any communications
requiring after sales support.
8
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
1.3.2 Design
Selecting a menu function (by pressing the appropriate key) will cause the function to be
selected, or the next menu to be displayed, as required. In general, the bottom menu key is
used to return to the previous menu, acknowledge a message, or cancel an operation.
9
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
1.5 Gas Path
10
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
1.5.2 Key to gas path figures
Analogue Output
11
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
1.5.4 MGA 3000 19 rack, infrared component with oxygen measurement
Trips (Alarm)
Digital output
Relay output(N/C, N/O)
Gas outlet
Connections 6mm or 1/4
12
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
1.6 Dimension drawings
480 66 664 44
266
13
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
1.7 Circuit diagrams (electric and gas connections)
14
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
1.7.3 Analogue Output(see section 2.5)
NOTES:
1. I is used to indicate pin functions when output is set to 4..20mA mode
2. Screened cable should be used for analogue output connections (screen to connector
shell).
15
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
1.8 Operating Principles
All heteroatomic gases absorb (or transmit) light energy in the infrared region at specific
wavelengths that are dependent upon the chemical composition of each type of gas.
Furthermore, the level of absorption is proportional to the mass of the gas present. This
characteristic is utilised in infrared gas analysis (IRGA) benches.
The benches incorporate a source of infrared energy (usually in the form of a heated
filament), which is then optically filtered to reduce the radiation spectrum to that of the
absorption band of the gas being measured.
A pyro-electric detector is used to measure the difference in energy between a reference zero
(no absorption) condition and that after some absorption in the gas has occurred. By
calibrating the bench at the zero condition and a known span concentration, a reliable
measure of concentration over the span range is obtained.
To improve the stability and to assist in the detection process, the infrared source is
modulated mechanically at 3-18Hz depending on the type of bench being used.
The relationship between absorption and concentration is not linear. For this reason the non-
linear output from the detector is modified electronically to give a linear output to the chart
recorder and display outputs.
The level of absorption measured is relative and therefore an IRGA provides a comparative
measurement, not an absolute one, against 'zero' and 'span' reference gases.
A 'zero' gas has no detectable concentration of the gas being measured. Depending on the
gas, fresh air, chemically stripped air or nitrogen may be used. A 'span' gas has a known
concentration of the measured gas and is used to set the gain (or span) of the analysis in the
mid-range region. Under normal circumstances, the concentration is equivalent to about 80%
of the displayed full-scale reading.
The IRGA bench is essentially an optical system and can suffer some deterioration in the
performance if it is internally contaminated by the condition of the incoming gases.
The gas sample should be reasonably dry. If the presence of water vapour is or is not
expected it is recommended that the instrument is continuously switched 'ON' to prevent
condensation.
The flow rate of the gas sample affects the response time (to change) of the measurement -
the measurement being on average the mix of the sample flowing through the cell. In
practice a sample flow rate of nominally 500 ml/min1 provides satisfactory performance for
16
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
any instrument. The instrument is fitted with a flow sensor and the flow rate is indicated on
the display. There is a potentiometer and throttle fitted internally that allows this flow rate to
be reset if necessary.
Generally the sample gas will be pumped to achieve the required flow rate. An internal pump
is fitted in the instrument and can be switched on or off via the key labelled pump on the
front panel. It is recommended that the pump is switched 'OFF' if the sample lines are being
attended to or the condition of the sample gas is suspect (for example it contains water).
The sample rate is factory pre-set for sample gas at ambient pressure.
The installation of pressurised cylinders must include suitable pressure regulators and
carefully controlled to avoid damage to pump or flow sensor due to over-pressure.
The instrument must be allowed to 'temperature stabilise' after first being switched on. The
display will show a thermometer symbol until the associated bench has reached operating
temperature. The instrument's enclosure plays a role in temperature control as well as
providing protection and the enclosure (top and front) should be securely in place before use.
Unless the instrument is only used occasionally it is advisable to leave the instrument
switched on, particularly over-night when reduced temperatures may result in condensation.
In standby conditions the sample pump should be switched off'.
Chemical oxygen sensor: This operates in a similar way to a battery in that a voltage is
generated between two plates immersed in electrolyte. The sample gas is passed over a
membrane which allows oxygen molecules to pass through it into the electrolyte, stimulating
the reaction, and resulting in a voltage proportional to the oxygen concentration.
Typically the sensor is calibrated by adjusting for zero reading with nitrogen applied, and for
span using a cylinder containing 20.9% O2, though fresh air can be used for a quick check.
Both span and zero adjustments are made via front panel controls.
The sensor fitted has a long operational life and is very stable. If the output falls (i.e. the gas
reading is low), the sensor should be replaced. Sensor lifetime depends primarily conditions
of use (e.g. average sample oxygen concentration, sample flow rate, temperature), and so
this part is excluded from the instrument warranty. For typical applications, ADC
recommends routine replacement of the sensor every two years as part of the maintenance
programme.
Paramagnetic oxygen sensor: This operates on the principle that oxygen molecules are
paramagnetic in nature and that exposing them to a high magnetic field induces alignment to
the field, creating a movement detectable by a simple infrared sensor reflecting on a mirror.
A very stable and long life sensor with high performance, this sensor offers the best results
for measurement in the range 0-25%. Replacement of the sensor is in excess of 10 years
Calibration of the sensor is identical to procedures used for the chemical oxygen cell
Lambda oxygen sensor: Primarily designed for use in the MGA to measure low levels of
oxygen down to 100ppb levels, the sensor uses a variation on the zirconia oxide ceramic
sensors developed by Nernst. The Lambda sensor and is in general use in motor vehicles.
17
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
Calibration within the MGA is similar to the other oxygen sensors, but this sensor does
require an annual service and calibration outside of the MGA to ensure continued use.
ADC Gas Analysis Ltd or their authorised local agents can supply replacement sensors with
the correct fittings and electrical characteristics. Fitting alternative parts may cause
problems and will invalidate the instrument warranty.
18
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
2. INSTALLATION
All power input, output and gas connections are via the rear panel.
Power connection is via a standard three-way IEC power plug. A suitable plug and cable is
supplied.
RS232C, analogue and alarm contact outputs are detailed in sections 2.4, 2.5 and 2.6
respectively.
It is not necessary to force ventilate the enclosure to cool the unit, but free circulation of air
is necessary for accuracy and reliability.
In all cases the ambient conditions must comply with the specifications. High
ambient conditions will affect reliability.
IMPORTANT: The MGA series analysers are not certified for operation in hazardous
environments.
19
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
2.3.2 Power Supply Connections
The power cable supplied must be connected to a suitable plug or distribution
board as follows:
The instrument is provided with an ON/OFF switch under the power socket.
WARNING
If the front, top or bottom sections of the enclosure are removed, BEWARE of
HIGH VOLTAGES near the power connector, switch and internal supply.
Avoid touching the electronic boards. If these are removed for any reason -
SWITCH OFF the instrument FIRST.
The serial port is configured from the front panel see section 3.7.2. The serial port protocol
is detailed in section 3.7.1.
In common with all RS232C ports, long cable runs can cause problems with data corruption
and noise. Keep cables as short as practicable certainly no longer than 30 meters at 9600
baud. Route signal cables well away from power cables, switchgear and other noise sources.
20
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
For long cable runs, and in electrically noisy environments, optical RS232C cable extenders
are recommended. Contact ADC Gas Analysis Ltd. or their local authorised agents for advice.
The outputs are arranged so that the reading shown at the top of the screen is presented on
analogue output 1, the reading on the next line down (if any) is presented on output 2 and
so on. Where there are fewer than four gas readings, unused analogue outputs will be
inactive or not fitted.
The analogue outputs are available via a circular DIN connector on the rear panel, labelled
OUTPUTS. A suitable connector is provided in the kit with each instrument, and further
supplies are available from ADC Gas Analysis Ltd. or their authorised local agent.
NOTES:
1. I is used to indicate pin functions when output is set to 4..20mA mode
2. Screened cable should be used for analogue output connections (screen to connector
shell).
Optional output cards can be specified to provide isolated mA, voltage and MODBUS RTU
21
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
2.6 Alarm (Trip) Outputs
The alarm outputs are isolated relay contacts which can be set to operate at certain gas
levels, or on sample flow error. Up to five contacts may be available depending on the
specification of the instrument. The number actually available is shown on the trip set-up
screen (see section 3.6.1).
Alarm contact outputs are available via a 15 way SUB-D 15 F connector on the rear panel.
A suitable connector is provided in the kit with each instrument, and further supplies are
available from ADC Gas Analysis Ltd. or their authorised local agent.
The connector pin functions are printed on the rear panel adjacent to the connector, and the
pin numbering scheme is shown in Fig 4.
All external connections are made with 6.35mm (1/4'') pipe fittings. The type of piping used is
recommended to be either polythene for general use or polytetraflurothalene (PTFE) or
stainless steel for corrosive gases.
All entries are fitted with in-line filters that are intended to stop any foreign matter entering
the instrument as a last resort. Unless the gas is perfectly clean and dry, particulate
filters and / or driers must be fitted externally.
Contact ADC Analysis Ltd. or their authorised local agents for advice.
The requirements for particular gases vary with the application and therefore these and the
gas circuit are specific to the delivered Instrument. Some guidelines follow, but it is the
responsibility of the installer to ensure that all safety and other requirements are met.
The Instrument Specification accompanying the Instrument defines the specific requirements
and this must be referred to during installation.
WARNING
All gas lines and connections must be leak-tight. Joints can be checked by
applying water containing some liquid soap and looking for bubbles.
22
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
2.7.2 The Sample Gas
The gas to be measured is the 'sample gas' labelled 'GAS IN'. After passing through the gas
circuit the measurement gases are exhausted at the entry labelled 'GAS OUT'. The analyser
is designed to receive sample gas at nominally atmospheric pressure, using the internal
sample pump to control the flow. Alternatively external pumps may be used
The Sample Gas supplied to the Instrument must be relatively clean, cool and of low
moisture content. Samples containing dust or particulates must be filtered externally and the
filters serviced at regular intervals.
Hot Samples with high moisture content should be passed through a water trap or dessicator
after cooling.
Samples may contain other gases which interfere with the gas being measured.
The extent to which known interferents can affect the measurement are defined in the
Instrument Specification.
Sample gas pipes shall be inert to and shall not contaminate the gas.
The response time for a change in concentration will be affected by the sample flow rate.
Instrument response times are normally quoted for the recommended 500ml min-1 flow rate,
ignoring the effect of sample pipe volume. Lower flow rates will increase the response time,
as will the volume of the sample piping. Higher flow rates will reduce the response time, but
excessive flow will pressurise the instrument which may cause damage or measurement
errors.
If it is necessary to draw higher flows than recommended (e.g. due to long sample gas lines),
it will be necessary to provide some means of bypassing the excess flow to exhaust.
The Regulator should be set to give a nominal output pressure of 0.2 Bar (3psig). This will
usually give a suitable gas flow during a span operation (i.e. flow indicator approximately
mid-position). If necessary, a small adjustment to the pressure will provide correct flow.
WARNING Application of gas pressures above 0.3 Bar / 5 psig will damage the
instrument.
The zero gas must not contain any significant traces of the gas measured or interferents. For
all types of instruments and gases measured, nitrogen (from a cylinder) is usually specified.
For some gases however, fresh air may be used (for example on high concentration CO and
23
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
CO2 Instruments). When fresh air is used, it shall be drawn from 'outside air and away from
any possible contamination from such as Exhaust Ducts, Chimneys, etc.
In most respects fresh air should be treated in the same way as the sample gas with regard
to filtering, moisture content etc. See span gas, section 2.7.3.
It is important to minimise back pressure at the GAS OUT port, as this will affect readings
and reduce the efficiency of the sample pump. Where long runs of pipe are necessary, larger
bore pipe should be used via a suitable adapter. Where the exhaust pipe is connected to a
forced air extraction duct, make sure that any suction is minimal.
The sample pump operates at fixed speed and at a flow rate of nominally 500 ml/m1n1.
If it should be necessary to adjust the flow, it can be set using the potentiometer VR2 on the
power supply board and / or the throttle (when fitted).
The internal sample pump is operated via the front panel 'PUMP' menu button.
Certain instrument configurations may be fitted with a second internal pump which is used to
circulate purge gas. This pump would be set to run continuously at a factory set rate.
The chemicals are contained in glass columns which are accessible from the front of the
instrument to aid inspection and refilling.
The chemicals that may be used are listed below under the gas application.
24
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
Silica gel is a good indicating drier, but can hang on to some gases, especially C02, giving
false readings, and is not generally recommended.
In use, soda lime changes from green to brown and must be changed when green is no
longer visible. It is not recoverable. Replacement supplies are available from ADC Gas
Analysis Ltd or their local authorised agents.
Other chemicals may produce corrosive by-products which can damage the instrument.
2.8.3 Precautions
Any chemical intended for use in the instrument is provided with it in a separate container,
labelled with international identification labels. Under normal use, the quantity supplied will
last for about 6 months to a year. Replacements can be obtained from ADC Gas Analysis Ltd
who will also advise on the use of chemicals obtainable by the User.
When handling chemicals, refer to the Safety Letter enclosed in the manual for precautions
against possible health risk. For all chemicals, avoid skin and eye contact and do not taste or
swallow them.
25
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
the cylinder tap is turned 'OFF'.
if the Instrument is in Standby, turn 'OFF' the Sample Pump.
Following a cylinder change, it is advisable to initiate a Span or Zero cycle a few minutes
after the supply is re-connected.
Applying water containing a mild soap solution can check connections under pressure. Do not
increase the pressure to check for leaks (or for any other reason), since this may damage the
Instrument.
Connections of pipes under nominally ambient pressure, such as the sample piping, are
difficult to check for leaks, but can be checked for tightness.
If Span cylinders contain hazardous or inflammable gases, they must carry suitable warning
labels, and if inflammable, must be located in a suitable Inflammable Store when in use or
in storage.
In all cases the local safety regulations must be consulted and followed.
26
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
3. SETTING UP
3.1 Preview
This section describes the settings available from the front panel. Factory settings have been
chosen to suit the general application of the instrument, or in accordance with the particular
instructions specified previously by the user.
3.2 On Power Up
IMPORTANT: Before operating the Instrument, ensure all the relevant installation
checks have been carried out [refer to Section 2.9], and that if a filter column is
fitted, it is properly charged with the chemical supplied.
Immediately after power is applied, the display presents a screen for 10 seconds. This page
contains the software version number.
The next screen is the main display, which will show the measured value(s), units and the
Gas type, sample flow status and key functions.
HCL Range
The measured value is initially replaced with a
NOX Calib
symbol of a thermometer until the instrument
has warmed up.
CO
Pressing the key labelled Set Up displays a Set Up
second screen showing the gas(s) measured SO2
Stp.Pump
together with the range and units.
Pressing More allows other gas measurement s present to be shown
The gases used for zero and span adjustment are supplied through the inlets at the back of
the instrument (see section 2.7), and electrically operated valves are used to route the
27
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
appropriate gas in place of the sample automatically when zero or span adjustment is
selected using the front panel controls.
'How often' to Zero or Span varies with each instrument. The interval required depends on
how much re-adjustment is being carried out: this can be established by noting the
discrepancy of the displayed value prior to resetting.
If the discrepancy is less than 0.5% of full scale, the intervals can be increased, or if this is
not accurate enough, decreased. Generally Zeroing is required more often than Spanning, so
that typically a daily Zero, with weekly Span is
adequate.
On the right, adjacent to the four function keys, are the key labels.
These describe the functions of the keys at any time, and the labels change as different
options are selected.
Between the gas readings and the function key labels, is the sample flow indicator, which is
designed to mimic a float flowmeter. Normally, the flow, as indicated by the float, should be
around the centre of the indicator. If the float is in a shaded region, the flow is too high or
low and a warning triangle is shown.(Flow display can be optional with related flow switch
range(currently, 200 /min is installed))
28
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
the range of the bar, press the Default key, then, using a 2mm flat-bladed screwdriver,
adjust the LCD control through the labelled hole in the back panel for best display. Note
that the rear panel control has a wide range and the display will not be visible for extreme
settings.
HCL 2.2 PP
M Range Range
NOX 16.2 PP
M
Calib
Measuring gas is the range selectable
CO 21.2 PP
M
Manual and Automatic range
Set Up
PP
SO2 18.1 M
Stp.Pump
Set Up
Stp.Pump
29
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
3.4.3 Menu tree
30
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
Fig. 8 shows the menu tree, starting from the main screen showing the sample gas
readings.
Sub-menus provide routes to more detailed functions, until the operator is required to 'DO'
an action. At this point, the display will present instructions, or more information, appropriate
to the action required.
For example, pressing the function key 'Zero adj' on the main menu, presents a Sub-menu
with the function keys on the right hand side of the display and instructions on the bottom of
the screen.
The outputs are arranged so that the gas shown on the top line of the screen is presented on
analogue output 1. The gas shown on the second line of the screen (if any) is presented on
output 2 and so on.
Analogue output scaling is fixed and corresponds to the full scale measurement range of the
associated gas. Analogue outputs do not indicate negative or over-range conditions.
The number of trip contacts fitted may vary depending on the specification ordered.
Change. To modify settings for the selected trip. The screen will display the selected trip,
and the function keys now offer items to change. From this screen use:
Gas to select the gas for which this trip is active, or alternatively select not set (trip not
used) or flow error (active if sample flow out of range).
31
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
Lo point to set a gas value below which the trip is active. If there is no reading shown
under this column on the display, or if the Off key is used when setting the low point, this
trip will not respond to low readings.
Done to save any changes made and return to the previous screen.
3.7.1 Protocol
Please refer to the Section 9.
Note that the data format is fixed at 8 bit data, no parity, 1 stop bit.
32
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
3.8 Measurement
The gas ranges, display format and units of measurement are factory set to order, and no
changes are possible. If changes to ranges or gases are required, contact ADC (Sales &
Service) Ltd. or their local authorised agents for a quotation.
Units of measurement are normally by volume or ppm though other options are possible to
order (e.g. mg m-1).
The display uses up to four digits, with values up to 3 decimal places, depending on range,
gas etc.
Over-range the reading is too high. Normally this indicates a gas concentration that
is above the measuring range.
Under-range the reading is too low. Normally, zero adjustment will correct the
problem.
HCL Range
NOX Calib
CO
Set Up
SO2
Stp.Pump
The measured value is initially replaced with a symbol of a thermometer until the instrument
has warmed up.
33
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
4. CALIBRATION
Gas readings made using IRGA optical benches are relative rather than absolute.
Before precise gas readings can be obtained, the analyser must be adjusted so that it reads
zero when there is no trace of the measured gas present, then, when a known concentration
of the gas is applied, the reading is adjusted to agree with this concentration.
These adjustments are known as zero and span respectively.
Routine calibration is important, not only to verify measuring accuracy, but also to spot
changes over time that may indicate problems.
Note: because the instructions will vary slightly from gas to gas, the instructions given here
are fairly general. ADC Gas Analysis Ltd. or their authorised local agent will be pleased to
answer any questions.
Before operating the Instrument, ensure all the relevant installation checks have
been carried out [refer to Section 2.9], and that if a filter column is fitted, it is
properly charged with the chemical supplied.
As one of the first operations concerns Zeroing and Spanning the Instrument, ensure that
external Zero and Span Gas supplies are available for connection to the Instrument.
If gas cylinders are used, check the Pressure Regulators are set correctly.
Make sure that the Power Settings [on the Rear Panel] are correct for the supply used [refer
to Section 2.3]
4.1.1 Warm Up
If previously removed for any reason, re-fit and secure the enclosure's front or top covers to
the Instrument. The covers are necessary to ensure that the internal temperature stabilises.
Switch on the Instrument via the Rear Panel mounted Power ON/OFF switch, and allow the
instrument about 30 to 60 minutes to warm up.
Until the instrument has reached operating temperature the display will indicate a symbol
which represents a thermometer. This symbol will be replaced by the gas measurement
when the correct temperature has been achieved.
34
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
4.2.1 Manual Zero
1. Select 'Zero Adj. ' on main menu, which will
display the zero screen and apply zero gas to HCL 0.0 PP
M
Zero Adj.
3. Wait for the gas reading to stabilise, Figure 13 - Zero screen (Multi gas model)
allowing any error to be noted.
4. If the reading(s) are satisfactory, press < Main to return to the main screen without
making changes. or ->
5. When the readings have stabilised, select the key labelled with the gas type you wish to
zero.
6. The microprocessor will then reset the reading to zero. If the change is to great and the
processor is unable to reset the zero, a warning message will appear on the screen.
Depending on the bench type, the message will either warn of failure, or offer to set
coarse zero.
7. Before selecting to perform coarse zero adjustment, double check that the zero gas is
correct coarse zero cannot be cancelled and may take a few minutes.
Set Up , Auto Z.
The interval for auto zero can be set in 1 steps selecting the + or - key until the required
time is set. Press Off to disable auto zero, and < Back to save settings and return to the
previous screen.
When an Auto Zero takes place the analyser will go through the same procedure as for a
manual Zero, but the processor will automatically reset the gas reading to Zero.
Auto zero will attempt adjustment for all gases, but will not initiate a coarse zero adjustment
if the standard zero fails.
When using Auto Zero the zero gas cylinder must be connected at all times
35
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
Before a Span adjustment is made ensure that the analyser is set to the higher or
maximum range for multirange analysers.
For multi-gas analysers, it is convenient if the span gas used is a mixture containing an
accurately known concentration for each of the gases measured. If this is not practicable,
then the span process must be repeated for each gas in turn, using the appropriate span gas.
Verify that the flow meter shows adequate flow span adjustment cannot be made if the
flow is incorrect. If necessary, make small adjustments to span gas pressure to correct the
flow.
If the readings are satisfactory, press < Main to return to the main screen without making
changes. or ->
If the span adjustment fails, carefully check the span gas cylinder. If possible, verify the gas
concentration on a second analyser, or try another bottle of span gas. If the span fails and
the span gas is correct, contact ADC Gas Analysis Ltd., or their local authorised
agents for service.
Set Up , Auto S.
36
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
selecting the + or - key until the required interval is set. Press Off to disable auto span, and
< Back to save settings and return to the previous screen.
When an Auto Span takes place the analyser will go through the same procedure as for a
manual span, but the processor will automatically reset the gas reading to the entered span
gas values. Auto span will attempt adjustment for all gases.
When using Auto Span the span gas cylinder must be connected at all times
37
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
5. MAINTENANCE
The User should be aware of any constituents of the gas that may directly or indirectly affect
the optical components within the instrument as described in Section 2.7.2.
If Water Traps and/or Particulate (dust) Filters are included in the installation these must be
periodically inspected and maintained to avoid overflows or blockage.
If transparent piping was used, water can often be seen in droplets on the inside walls or
collecting at a low point in the piping arrangement. With opaque or steel pipes, disconnect
them at a low point to see if water drips out.
If signs of water are found, it may eventually block the pipes or result in amounts being
drawn into the instrument which may affect the pumps, solenoids etc., as well as the optics.
In this case, it is recommended that some form of Water Trap be fitted, such that the
contents can be easily seen and regularly inspected.
If the Sample does contain water vapour, leave the Instrument switched on if it is not in use,
and with the Sample Pump off. This will help to prevent condensation within the instrument.
A service visit is advisable in cases of water ingress, so that the optics can be cleaned. Dirty
optics will reduce performance.
5.1.2 Dust
Long term problems may occur if dust is in the sample. Usually this cannot be seen, but
over a period of time it can build up and affect all parts of the gas path. Dust may also block
the particulate filters fitted in the inlets of the Instrument and may therefore result in a
blockage or a reduction in flow.
As part of the maintenance routine, inspect the Inlet Filters for signs of dust. If there is any
sign of dust, the sample line must be fitted with a suitable filter (down to 1O microns) and
the filter elements regularly replaced.
Where dust has penetrated the instrument, arrange for a service visit so that the optics can
be cleaned and full performance restored.
38
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
5.2 Instrument Checks
The instrument requires little maintenance, other than the filter column (if fitted), source
block purge (if fitted), sample flow adjustment (if necessary), and zero and span adjustment
(see section 4).
When replacing or handling chemicals, avoid physical contact and wash hands
afterwards.
Dispose of chemicals carefully and in accordance with any local Health & Safety
requirements.
When replacing or handling chemicals, avoid physical contact and wash hands
afterwards.
Dispose of chemicals carefully and in accordance with any local Health & Safety
requirements.
If the flow continues to fall off over a period of time, it may indicate that the lines are
becoming blocked, as indicated in the previous section.
5.2.4 Throttles
The sample and zero gas flows are normally set in the instrument with external throttles, as
indicated in the Gas Circuit. These paths can also become affected by blockages and they (or
some other reason) may require some re-adjustment.
Readjustment is not a routine procedure always check span and zero gas bottles, regulator
pressures etc. before attempting adjustment.
39
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
6.1 Introduction
This section describes the basic construction and operation of the various types of bench that
may be fitted in an instrument. All the infrared types rely on a characteristic of heteroatomic
gases to absorb energy at specific wavelengths in the infrared spectrum, and, follow the
operating principles described in Section 1.8.1.
Each bench type makes use of a different technique. To recommend a type of bench best
suited to the requirements will generally relate to the type of gas measured and/or the
accuracy and sensitivity of measurement.
6.2 GC Bench
The GC (Gas Correlation) infrared bench is a non-dispersive single beam analyser in which
the gas to be measured is passed through an optical cell continuously. Although the
instrument output is in terms of gas concentration, this type of analyser operates as a
comparator. That is, its output is not absolute, but is established by standardising with a
known gas mixture at a point on the calibrated scale. See section 4.
Source
Motor
Optical
Filter Gas Entries
Lens
A typical GC Bench layout is shown in Fig. 18. This reflects a single analysis cell. Variations
on this will be on the length of cell or the number of cells.
The length of the cell is dependent on the concentration of the sample measured and may
vary from 1mm (at very high concentration) up to about 250mm (at low concentration).
40
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
Infrared (heat) radiation from a small source is directed through a rotating filter wheel, a
collimating lens, an optical (Thin-Film) filter, the sample cell, a focussing lens and on to a
solid state detector. The wheel is rotated at 1100 RPM, to give a modulation frequency of
18.3 Hz.
Each cell element is sealed at both ends by a window that does not optically interfere with
the radiation.
The heart of the analyser is the gas filter wheel, which contains a sealed sample of gas of the
type to be measured, and a non-absorbing gas. As the wheel rotates it provides a ratio of
sample and reference signals, in a sequence detected by a wheel position sensor. The ratio
signals, together with the selective transmission of the optical filter, sensitises the analyser to
respond to that region of the infrared spectrum corresponding to the measured gas in the
sample.
When the gas to be measured enters the analysis cell, it absorbs some radiation, and alters
the ratio of the sample and reference signals. It is this change in energy level that is
amplified to give the output signal.
The optical path between the source and the first cell window, and between the last window
and the detector, will normally contain room air, or a gas from which any interfering gases
have been stripped. This air is referred to as Purge air, and if a chemical stripper is used, it
will be shown on the gas circuit. As Purge air has the properties required for the Zero Gas, it
may be used as such.
Whilst the bench can be stripped down to individual parts and parts replaced, this process will
require re-calibration of the bench and possibly some re-adjustments in the Pre-amplifier
and/or amplifier associated with the detector. Since this requires special training and
equipment, the User should not undertake re-calibration unless such training and equipment
has been approved.
41
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
7. ELECTRICAL SPECIFICATIONS
For specifications relating to gas measurements, refer to the Certificate of Conformity.
42
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
8. SPARE PARTS
The following table lists some of the more common spare parts. ADC (Spares & Service)
Limited provided total parts support and can quote for any item not listed just contact us or
our authorised local agent.
Always quote the Serial Number and Model Number in any communications
requiring spares or after sales support.
43
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
9.1 Introduction
The MGA 300 has extensive remote control functionality using a variant of the widely used
AK protocol.
This document details the commands, allowing full remote control and monitoring of the
analyser.
Before starting, ensure that the serial port speed, handshake mode and analyser address are
set as required to match the AK host, using the Serial option from the Set Up menu.
Note: spaces are not allowed between the start code, address and command.
44
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
9.3 Address
The address character is used to distinguish different pieces of equipment. The MGA will only
respond to a command if the address digit in the command matches the address set in the
analyser (in the serial port set up screen). If the command address is * however, the MGA
will always respond.
For example, the command SREM (set remote control mode) requires no parameters, while
SEGA (set span gas concentration) requires three (channel, range, span gas). In general,
the MGA analyser ignores unexpected parameters.
All commands recognised by the MGA analyser for remote control use consist of four upper
case letters. Case is important.
Certain commands require that the analyser is set for remote control mode, or cannot be
used while calibration is in progress, for example. If a command is inappropriate or badly
formed, it will be ignored, and the reply will contain a fault code in place of any parameters.
Here is an example:
45
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
Do not confuse SMAN when used as an error indication, with a response to a valid SMAN
command. Error indications are parameters, and appear after the command echo.
Examples:
Error code digit Error(s)
2 Sample flow error
A Sample flow error & last span failed
B Last zero failed & last span failed
Both zero and span fail indications remain set until a successful zero/span has been
completed, or the analyser is reset.
The flow error indication is set whenever a warning triangle would be shown under the flow
meter on the analyser display.
Command: GRRD
Parameters: <channel> <range>
Where: <channel> is an integer 1..4 selecting possible channel
(which equates to a specific gas)
<range> is an integer 0 or 1;
46
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
0 = least sensitive or only range
1 = more sensitive range, if fitted
This command returns the calibration factors used by the MGA analyser. By reading these
factors before and after a calibration, it is possible to see by how much the calibration has
moved the zero/span. Factors are held separately for each range and gas.
Command: GRCL
Parameters: <channel> <range>
Where: <channel> is an integer 1..4 selecting possible channel
47
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
(which equates to a specific gas)
<range> is an integer 0 or 1;
0 = least sensitive or only range
1 = more sensitive range, if fitted
Data returned: <span_factor> <zero_offset>
Where: <span_factor> Is an integer representing the span
factor multiplied by 10000.
e.g. a value of 10000 represents a span factor of 1.0, the
nominal value.
<zero_offset> Is an integer representing the zero offset
Command: GRMW
Parameters: none
Data returned: <I>
Where: <I> is a number representing the analyser gas as follows:
-1 = zero calibration gas
-2 = span calibration gas
0 = first or only sample inlet
1 .. 15 = other sample inlets (as options fitted)
Command: AFLO
Parameters: none
Data returned: <flow> <pump>
Where: <flow> is the sample flow, 0 .. 1023 (see above)
<pump> is sample pump state: 0=off, 1=on
Example reply: <S>1AFLO 0 800 1<E>
sample flow is approx. 600ml min-1, pump is on
The MGA has eight ports which may be used for trip relays or gas switching valves. The
assignment of relays and valves to ports depends on the specification of the specific analyser.
Normally, ports 1 to 4 correspond to trips 1 to 4 (as fitted). Where trip 5 is fitted, this
corresponds to port 8.
When using this command to interrogate the state of trips, bear in mind that a trip can be
configured to make the relay operate either when the trip condition is met, or when it is not
met. The trip setting will therefore determine whether a port state of 1 means tripped, or
not tripped.
48
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
Command: GRTP
Parameters: none
Data returned: <p1> <p2> <p3> <p4> <p5> <p6> <p7> <p8>
Where: <px> is the state of port x
0 = relay not operated or valve closed
1 = relay operated, valve closed
Example reply: <S>1GRTP 0 0 1 0 0 1 0 0<E>
ports 3 and 6 are in the active state.
Command: SREM
Parameters: none
Data returned: none
Command: SMAN
Parameters: none
Data returned: none
49
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
9.8 Measurement Enquiry Commands
These commands are not accepted during calibration.
Command: SEMB
Parameters: <channel> <range>
Where: <channel> is an integer 1..4 selecting gas / channel
<range> is an integer, 0 or 1;
0 = least sensitive or only range
1 = more sensitive range, if fitted
Data returned: none
Note: inlet numbers displayed on the analyser screen start at 1, not zero. e.g. if SMGA
command is used with inlet parameter 2, the analyser screen will show the inlet as 3.
50
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
9.10.2 Select zero gas
This command switches zero gas through the analyser, to allow for zero check. Calibration is
not attempted - concentrations can be read as normal. Error NF is returned if a zero gas
switching valve is not fitted.
Command: SNGA
Parameters: none
Data returned: none
Command: SEGA
Parameters: none
Data returned: none
Command: SRPS
Parameters: <on_off>
Where: <on_off> an integer to control the pump
0 to turn off the pump, <>0 to turn it on
Data returned: none
For either case, zero is performed for all gasses and ranges simultaneously.
An instant zero calibration attempts to set zero using the current contents of the measuring
cells. There are no gas circuit changes or delays. It is assumed that the analyser is fully
51
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
purged with zero gas before sending an Instant zero command. Instant zero completes in 1-
2 seconds.
No error is given if the zero adjustment fails for any gas. Perform a zero check to verify if
necessary. The SATZ command is not acknowledged until the zero cycle is completed.
Error WMUP is returned if this command is attempted while any channel is warming up.
Command: SATZ
Parameters: [<instant>]*
Where: <instant> is a letter I, and, when present causes an instant
zero calibration.
This parameter is ignored prior to firmware 2.07.
Data returned: none
Command: SPGA
Parameters: <channel> <range> <conc>
Where: <channel> is an integer 1..4 selecting gas / channel
<range> is an integer 0 or 1;
0 = least sensitive or only range
1 = more sensitive range, if fitted
<conc> is a real value specifying span gas concentration
as a proportion of the full scale range.
Accepted values are from 0.5 to 1.0. Values can be set to three decimal
places.
Data returned: none
Example command:
Range 1 on channel 2 is 5% CO2 full scale. We wish to set the analyser to expect 4.17%
span gas:
<S>*SPGA 2 1 0.8340<E>
Only the selected range and channel is calibrated. The status of the calibration is reported.
Error NF is returned if the given channel or range is not fitted. Error WMUP is returned if this
command is attempted while any channel is warming up.
Command: SATS
Parameters: <channel> <range>
Where: <channel> is an integer 1..4 selecting gas / channel
<range> is an integer 0 or 1;
52
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
0 = least sensitive or only range
1 = more sensitive range, if fitted
Data returned: none *
* Note: error ERR is returned if the calibration failed.
53
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
54
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
When switching to remote control using the SREM command, ensure that the main results
screen is displayed. Other screens may appear corrupted by remote control commands.
Sample gas may be restored on setting the analyser back to local control if span or zero gas
was previously selected.
Analyser fail indication (bit 0 in status code digit) is not yet implemented all other
indications work as described.
55
MGA3000 Multi-Gas Analyser Operation Manual Iss. 3.01
56